Features
• Programmable Alarm and Warn
Ranges with Alarm Delay
• Instantaneous Out of Range
Indication
• Programmable or Manual Start
• Inexpensive
• Splash Resistant (IP64)
• Sleek Enclosure with Lanyard
Attachment
• Multi Use
Applications
• Verification of Safe Shipment and
Storage of Perishable Goods
• Implementation of HACCP
Programs
• Cold Chain Monitoring
• Medical and Pharmaceutical
Shipping
• Supermarkets, Warehouses,
and Cold Storage
The TransiTempII is a low cost, splash proof temperature data logger featuring a highly precise semiconductor temperature sensor and an ABS plastic enclosure. This affordable device is ideal for monitoring and recording the temperature of all types of perishable products, in­cluding food items and medical/pharmaceutical goods, both in stor­age and in transit. Program the TransiTempII for an immediate start or a delayed start for up to 30 days. The device can also be started by use of a magnetic wand. Because there are no buttons on this logger, there is no risk of it being accidentally stopped or started.
The reading rate can be set anywhere between once every five seconds to once every thirty minutes. The total reading capacity of the TransiTempII is 32,767 data points, one of the highest reading limits of all such loggers on the market today. For increased understanding of events that impact recorded temperatures, users can “mark” data points with a quick wave of the magnetic wand.
Three status LED indicators alert the user to the temperature status, with green signaling safe conditions, yellow indicating that warning limits have been exceeded, and red denoting that alarm limits have been exceeded.
Two of the most outstanding features of the TransiTempII are the available IFC300 docking station (sold separately) and the accompanying free software. To download data, simply slip the logger into the docking station and select “Download Data.” After download, the software’s clear, easy-to-read graphs and data tables are automatically created and ready to use. Temperature data may be shown in degrees Celsius, Fahrenheit, Kelvin or Rankine. More complex analyses, such as Mean Kinetic Temperature, can be performed with just one click of the mouse, while users needing to perform additional analyses can easily export data to MS Excel®.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
Multiple graph overlay
Statistics
Digital calibration
Zoom in/ zoom out
Lethality equations (F0, PU)
Mean Kinetic Temperature
Full time zone support
Data annotation
Min./Max./Average lines
Data table view
Automatic report generation
Summary view
Multilingual
TRANSITEMPII SPECIFICATIONS*
Temperature
Sensor: Semiconductor
Measurement Range: -40 °C to +80 °C (-40 °F to +176 °F)
Resolution: 0.1 °C (0.18 ºF)
Calibrated Accuracy: ±0.5 °C (±0.9 °F)
Specified Accuracy Range: -10 °C to +40 °C (14 °F to +104 °F)
Response Time:
Calibration: Digital calibration through software
Calibration Date:
Memory:
Reading Rate:
Start Modes:
Temperature Alarm:
Alarm Delay:
BATTERY WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT RECHARGE,
DISASSEMBLE, HEAT ABOVE 100 ºC (212 ºF), INCINERATE OR EXPOSE CONTENTS TO WATER.
90% of change in 30 minutes typical in slow moving air
Recorded within device; software notifies user when calibration is due
32,767 readings; software configurable memory wrap
5 seconds to 30 minutes, selectable in software
Immediate start, delay start up to 30 days in advance, or manual start with magnetic switch
Configurable temperature alarm with programmable high and low alarm limits, high and low warn limits. Each set point may be individually enabled or disabled
Programmable alarm delay inhibits alarm until alarm limits are exceeded for a total time from the start of recording
ORDERING INFORMATION
*SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE. SPECIFIC WARRANTY REMEDY LIMITATIONS
APPLY. CALL 1-603-456-2011 OR GO TO WWW.MADGETECH.COM FOR DETAILS.
Blinking Status Indicators:
Green: Indicates safe conditions during logging
Yellow: Indicates warn limits have been exceeded
Red: Indicates alarm criteria have been exceeded
Date and time stamp may be placed in the
Marking Feature:
recorded data with the magnetic switch. The marking feature may be configured to clear the alarm/warn indicators.
An optional password may be programmed
Password Protection:
into the device to restrict access to configuration options. Data may be read out without the password.
*Battery Type:
Battery Life:
3.6V lithium battery, *Must be sent to MadgeTech for battery replacement.
1 year typical at 25 °C (77 °F), 1 minute reading intervals
±2 minutes/month at 25 °C (77 °F),
Time Accuracy:
±5 mins./month at –10 °C to +40 °C (14 °F to 104 °F)
Data Format: Date and time stamped at °C, °F, K,°R
Computer Interface: USB docking station to PC; 38,400 baud
Software: XP SP3/Vista/Windows 7/Windows 8
Operating Environment:
-40 °C to +80 °C (-40 °F to 176 °F), 0 %RH to 100 %RH
Ingress Protection: IP64 (splash resistant)
Weight: 1.4 oz (40 g)
Dimensions: 3.3 in x 2.1 in x 0.4 in (83 mm x 54 mm x 11 mm)
Enclosure: ABS plastic, white
Approvals: CE
The TransiTempII exchange program provides the
Product Maintenence:
user with a new replacement device at a reduced cost when battery is depleted or device is in need of factory calibration. Call for details.
MODEL DESCRIPTION
TransiTempII Cold Chain/Perishables Temperature Recorder
IFC300 USB Docking Station, Software and Manual
Calibration Certificate Calibration Certificate available for data logger
MadgeTech 4 Secure
Software License
Exchange Program
For Quantity Discounts call 603-456-2011 or email sales@madgetech.com
License for MadgeTech Secure Software
(unlimited users, 1 license per computer, more licenses are available.)
When product battery is depleted or device requires factory calibration, the product can be exchanged for a new TransiTempII Data Logger at a reduced cost. Call for details
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Temperature
ASK ABOUT OUR OTHER
DATA
LOGGERS
DOC-1139009-00 REV 11 2014.11.26
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion
Loading...