ThermoVault Max
Extreme Temperature Thermal Barrier
PRODUCT
USER GUIDE
To view the full MadgeTech product line,
visit our website at madgetech.com.

ThermoVault Max
PRODUCT USER GUIDE
Product Overview
The ThermoVault Max is an extreme, high temperature
thermal barrier designed for use with the HiTemp140
series. This durable thermal barrier can withstand
temperatures up to +400 °C in dry heat applications and
+250 °C in wet applications when used with the TMAX Wet
Seal Kit (optional accessory).
Compatible Data Loggers
• HiTemp140-M12
• HiTemp140-FP
• HiTemp140-PT
Getting Started
Dry Applications
The HiTemp140-FP, HiTemp140-PT HiTemp140-M12 can
be used for dry applications up to +400 °C.
1. Unscrew the lid of the
thermal barrier.
2. Feed the probe of the data
logger through the bottom
end of the lid.
3. If using the HiTemp140-M12
data logger, screw the
probe into the bottom end
of the barrier lid.
4. Slide the data logger and
lid back into the barrier
body and screw the lid
back on tightly.
Product User Guide | 2

ThermoVault Max
PRODUCT USER GUIDE
Wet Applications
For use in wet applications, the TMAX Wet Seal Kit (sold separately, optional accessory) should be used with the
HiTemp140-FP, and HiTemp140-PT data logger. When used in wet applications the maximum operating environment
cannot exceed 250 °C (482 °F).
The TMAX Wet Seal Kit Includes:
• PTFE Thread Sealing Tape
• Fitting Body
• Dome Nut
• Compression Ferrules: PTFE (5), Stainless Steel (1)
• O-Ring
Installing the O-Ring: The O-Ring ensures a tight seal which prevents liquid from entering the inside of the barrier. O-Ring
maintenance is a key factor when properly caring for the ThermoVault Max. Please refer to the application note “O-Rings
101: Protecting Your Data Logger,” found on the MadgeTech website, for information on how to prevent O-Ring failure.
1. Unscrew the endcap of the
thermal barrier.
2. Stretch the O-Ring over the
endcap so it sits in the groove just
beneath the knurled shoulder.
Product User Guide | 3

ThermoVault Max
PRODUCT USER GUIDE
Securing the Compression Fitting: The compression fitting is used with the HiTemp140-FP and HiTemp140-PT data
loggers to ensure a water tight seal.
1. Wrap the provided PTFE thread
sealing tape around the Fitting
Body threads.
4. Insert the Compression Ferrule
into the Fitting Body and hold it in
place loosely with the Dome Nut.
2. Screw the Fitting Body into the
end of the stopper assembly
(finger-tight).
Compression Ferrule Note: The
stainless steel Ferrule provides the
most reliable seal, but becomes
permanent crimped onto the probe.
The PTFE Ferrule will become more
deformed with multiple uses and will
eventually require replacement. It’s
benefit is that it allows easy removal
and repositioning of the probe.
3. Holding the smooth metal end of
the stopper, use a 7/16” wrench
or socket to tighten the body into
the stopper.
5. Feed the data logger probe throw
the Compression Fitting until
the base of the probe is tightly
against the Dome Nut.
6. Tighten the Dome Nut to secure
the data logger in place (fingertight).
7. Insert the data logger and stopper
assembly back into the thermal
barrier and screw the endcap
of the thermal barrier until the
O-Ring is no longer visible.
Product User Guide | 4

ThermoVault Max
NEED HELP?
Product Support & Troubleshooting
• Visit our Resource Library online at madgetech.com/resources.
• Contact our friendly Customer Support Team at (603) 456-2011 or support@madgetech.com.
6 Warner Road, Warner, NH 03278
(603) 456-2011
info@madgetech.com
madgetech.com
DOC-1322035-00 | REV 2 2020.04.28