MadgeTech RTDTemp101A Specifications

RTDTEMP101A
RTD BASED TEMPERATURE DATA LOGGER
Features
• 10 Year Battery Life
• ±0.04 °C Accuracy
• Multiple Start/Stop Function
• Ultra High Speed Download
• 1,376,256 Reading Storage Capacity
• Memory Wrap
• Precision RTD Sensing Element
• Battery Life Indicator
• Optional Password Protection
• Programmable High and Low Alarms
• Field Upgradeable
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Calibration Chamber Monitoring
• HVAC
• Medical/Pharmaceutical
• Environmental Studies
• Precision Temperature Recording
Don’t be fooled by the small size, the RTDTemp101A temperature data logger oers a wealth of features with a compact form factor about the size of a matchbox. When used with an external RTD probe, this data logger measures temperatures from
-200 °C to 850 °C.
The low power design of this data logger provides a battery life of up to 10 years but still delivers
ultra-fast download speeds. The RTDTemp101A can store over a million readings and oers a software configurable memory wrap option. In addition to having a pushbutton start stop, the device also can be programmed to a delay start up to 18 months in advance.
Using the MadgeTech Data Logger Software, data from the RTDTemp101A is easily down­loaded, displayed and formatted for reports or the data can be exported to Excel for further presentation customization.
This versatile device can be used for a wide variety of applications such as calibration chamber monitoring, HVAC applications, environmental studies or precision temperature recording for medical and pharmaceutical applications.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Data table view
• Automatic report generation
• Summary view
• Multilingual
RTDTEMP101A SPECIFICATIONS*
*SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE. SPECIFIC WARRANTY REMEDY LIMITATIONS APPLY. CALL (603) 456-2011 OR GO TO MADGETECH.COM FOR DETAILS.
Temperature Sensor:**
Measurement
Temperature Range:
Temperature Resolution:
Calibrated Accuracy:
Calibrated Accuracy Range***:
Input Connection****:
Temperature Eect on Span: < 2 ppm/ºC
Temperature Eect on Oset: < 10 ppm cumulative over entire range
Reading Rate:
Memory:
Wrap Around: Yes
Start Modes:
Stop Modes:
Multiple Start/Stop Mode:
Multiple Start/Stop
Mode Activation:
**Temperature specifications based on ideal 100 Ω Pt RTD compliant with IEC 751 (1983) and ITS-90, 5000 Ω FSR (accuracy based on 36 in lead wire RTD with 4 wire configuration)
***Calibrated accuracies based on standard MadgeTech calibrations for 0 to 200 Ω range.
****100 Ω, 2 or 4 wire RTD probes are recommended for the most accurate per formance. Most 100 Ω, 3 wire RTD probes will work, but MadgeTech cannot guarantee the accuracy. To determine whether or not the 3-wire RTD probe will work, the resistance between the two same colored wires should be less than 1 Ω. (Note: Please contact the manufacturer of the RTD probe for questions on the resistance)
External 100 Ω RTD (Sold Separately)
18 Ω to 400 Ω
-200 °C to +850 °C (-328 °F to +1562 °F)
(Probe Dependent)
0.0005 Ω
0.01 °C (0.018 °F) (Probe Dependent)
±0.015 Ω ±0.04 °C (±0.072 °F) (Probe Dependent)
18.5 Ω to 175.9 Ω
-200 °C to +200 °C (-328 °F to +392 °F)
Removable screw terminal; 2, 3 or 4 wire interface
1 reading every second up to 1 reading every 24 hours
• 1,376,256 readings; software configurable memory wrap
• 589,824 readings in multiple start/stop mode
• Immediate start
• Delay start up to 18 months
• Multiple pushbutton start/stop
• Manual through software
• Timed (specific date and time)
Start and stop the device multiple times without having to download data or communicate with a PC
To start the device:
Press and hold the pushbutton for 5 seconds, the green LED will flash during this time. The device has started logging.
To stop the device:
Press and hold the pushbutton for 5 seconds, the red LED will flash for three seconds and then the green LED will flash for two seconds. The device has stopped logging.
Real Time Recording:
Alarm:
The device may be used with PC to monitor and record data in real time
Programmable high and low limits; alarm is activated when temperature reaches or exceeds set limits
Green LED blinks:
• 10 second rate to indicate logging
• 15 second rate to indicate delay start mode
LED Functionality:
Red LED blinks:
• 10 second rate to indicate low battery and/or full memory
• 1 second rate to indicate an alarm condition
An optional password may be programmed into
Password Protection:
the device to restrict access to configuration options. Data may be read out without the password.
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6V lithium battery included; user replaceable
10 years typical at a 15 minute reading rate
Battery Life:
Graph display of the device recording in a 25 °C environment.
Data Format: Date and time stamped ºC, ºF, K, ºR
Time Accuracy:
±1 minute/month (at 20 ºC/68 ºF, stand alone data logging)
Computer Interface: USB (interface cable required); 115,200 baud
Software: XP SP3 or Later
Operating Environment:
-40 ºC to +80 ºC (-40 °F to +176 °F),
0 %RH to 95 %RH non-condensing
Dimensions: 1.4 in x 2.2 in x 0.6 in (36 mm x 56 mm x 16 mm)
Weight: 0.85 oz (24 g)
Materials: ABS Plastic
Approvals: CE
BATTERY WARNING: FIRE, EXPLOSION AND SEVERE BURN HAZARD. DO NOT RECHARGE,
DISASSEMBLE, HEAT ABOVE 100 °C (212 °F), INCINERATE, CRUSH, OR EXPOSE CONTENTS TO WATER.
ORDERING INFORMATION
MODEL PART NUMBER DESCRIPTION
RTDTEMP101A 901520-00 RTD Temperature Data Logger
IFC200 900298-00 Software, manual and USB interface cable
LTC-7PN 900352-00 Replacement battery for RTDTemp101A
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Temperature
ASK ABOUT OUR OTHER
DATA
LOGGERS
DOC 1216009-00 REV 9 2019.02.01
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion