MadgeTech RHTemp1000IS User Manual

RHTemp1000IS Intrinsic Safety User Notice

Intrinsic Safety Approval:

The RHTemp1000IS has been certified by FM Approvals as Intrinsically Safe (IS) for use in Class I, Division 1, groups A, B, C, D, and Nonincendive (NI) for use in Class I, Division 2, groups A, B, C, D Hazardous (Classified) Locations. The rating listed in the Factory Mutual approval guide is as follows:
RHTemp1000is. Temperature and Humidity Recorder.
IS / I / 1 / ABCD T4A Ta=80°C; NI / I / 2 / ABCD / T4A Ta=80°C
These are the only safety ratings relevant to the use of this product. Use of this product in hazardous environments not specifically covered by this rating, is prohibited, unless the user takes the appropriate steps to ensure the safety of the product and assumes full responsibility for its safe use. Refer to the reference sections at the end of this document for further information on approval standards and environments.

Conditions of Use:

The following conditions must be satisfied to maintain the IS rating of the RHTemp1000IS:
When used in hazardous locations, the RHTemp1000IS is to be
installed prior to the location becoming hazardous, and removed only after the area is no longer hazardous.
The maximum allowed ambient temperature for the RHTemp1000IS
(under any circumstances) is 80°C. The minimum rated operating temperature is -20°C.
The RHTemp1000IS approved for use only with the Tadiran TL-
2150/S battery. Replacement with any other battery will void the safety rating.
Batteries are user replaceable, but are to be removed or replaced only
in locations known to be non-hazardous.
Tampering or replacement of non-factory components may adversely
affect the safe use of the product, and prohibited. Except for replacement of the battery, the user may not service the RHTemp1000IS. MadgeTech, Inc. or an authorized representative must perform all other service to the product.
DOC-1271036-IS
Standard
Date
Title
FM Class 3600
2011
Electrical Equipment for Use in Hazardous (Classified) Locations, General Requirements
FM Class 3610
2010
Intrinsically Safe Apparatus and Associated Apparatus for Use in Class I, II and III, Division 1 and Class I, Zone 0 and 1 Hazardous (Classified) Locations
FM Class 3611
2004
Non-incendive Electrical Equipment for Use in Class I and II, Division 2 and Class III, Divisions 1 and 2 Hazardous (Classified) Locations
FM Class 3810 Including Supplement #1
2005 1995
Electrical Equipment for Measurement, Control, and Laboratory Use
The RHTemp1000IS enclosure does not carry a NEMA rating, as there
are openings in the enclosure for measuring humidity. For this reason, the product is only suitable for use in indoor locations.

Required Labeling:

The following label must be engraved to the enclosure of the RHTemp1000IS. It contains critical information for the safe use of the product.

Reference Standards:

The RHTemp1000IS complies with the following standards: