QUADTCTEMPV2
4-CHANNEL THERMOCOUPLE
DATA LOGGER
Features
• Up to 4 Hz Reading Rate
• 261,888 Readings Per Channel
• Open Channel Detection
• Title Channels
• Accepts Thermocouple Types
J, K, T, E, R, S, B, N
• External Power or User
Replaceable Battery
• Pushbutton or Programmable Start
• Calibration Certificate included
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Warehouse Monitoring
• Refrigerator Monitoring
• Medical and Pharmaceutical
• Oven Monitoring
• Smoke Houses
• Food Processing
• HVAC
• Engine Studies
The QuadTCTempV2 is a 4-channel thermocouple data logger with a reading rate of up to 4 Hz.
It can measure and record 261,888 readings per channel. To maximize memory capacity, users can enable or disable channels. For easy identification, each channel can be named with
up to a ten digit title.
In addition, the QuadTCTempV2 features individual cold junction compensation for each
channel providing increased accuracy and response time, and if a probe is removed or
severed during logging, the software automatically annotates the data.
The QuadTCTempV2 is ideal for a variety of applications, whether it is remote temperature
monitoring, or multiple points in a central location. Data from all channels is simultaneously
logged. After the monitoring cycle is complete, data can be downloaded for analysis.
* Thermocouple Plugs/Probes
Sold Separately
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Data table view
• Automatic report generation
• Summary view
• Multilingual
QUADTCTEMPV2 SPECIFICATIONS*
* Specifications are subject to change without notice. Specific warranty remedy limitations apply.
Call (603) 456-2011 or go to madgetech.com for details.
4 Internal Channels
Temperature Range: -20 ºC to +60 ºC (-4 ºF to +140 ºF)
Temperature Resolution: 0.01 °C (0.018 °F)
Calibrated Accuracy: ±0.5 °C (0 °C to 50 °C)
4 Remote Channels
Remote Channel
Thermocouple Types:
J, K, T, E, R, S, B, N
Thermocouple Connection: Female subminiature (SMP)
Cold Junction
Compensation:
Maximum Thermocouple
Resistance:
Automatic, based on internal channel
1000 Ω, <100 Ω recommended
Thermocouple Range (ºC) Resolution Accuracy**
J -210 to +760 0.1 ºC ±0.5 ºC
K -270 to +1370 0.1 ºC ±0.5 ºC
T -270 to +400 0.1 ºC ±0.5 ºC
E -270 to +980 0.1 ºC ±0.5 ºC
R -50 to +1760 0.5 ºC ±2.0 º C
S -50 to +1760 0.5 ºC ±2.0 º C
B +50 to +1820 0.5 ºC ±2.0 º C
N -270 to +1300 0.1 ºC ±0.5 ºC
Software programmable immediate
Start Modes:
start or delay start up to six months
in advance
**Thermocouple accuracy is specified with a 24 AWG.
Real Time Recording:
Memory:
(all 4 channels enabled)
May be used with PC to monitor and
record data in real time
261,887 readings per channel
Reading Rate: 4 Hz up to 1 reading every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type:
9 Volt lithium or alkaline battery included,
user replaceable
Battery Life: 18 months typical
Data Format: Date and time stamped ºC, ºF, K, ºR, mV, V
Time Accuracy: ±1 minute/month
Computer Interface: USB (Interface cable required); 115,200 baud
Software:
Operating Environment:
Dimensions:
Windows XP SP3 or later
MadgeTech 4 software, version 4.2.12.0 or later
-20ºC to +60ºC (-4 ºF to +140 ºF),
0 %RH to 95 %RH non-condensing
2.7 in x 7.2 in x 1.0 in
(69 mm x 184 mm x 26 mm)
Material: Black anodized aluminum
Weight: 15 oz (420 g)
Approvals: CE
BATTERY WARNING: BATTERY MAY LEAK, FLAME OR EXPLODE IF DISASSEMBLED, SHORTED,
CHARGED, CONNECTED TOGETHER, MIXED WITH USED OR OTHER BATTERIES, EXPOSED TO
FIRE OR HIGH TEMPERATURE. DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF
CHILDREN.
ORDERING INFORMATION
MODEL PART NUMBER DESCRIPTION
QUADTCTEMPV2 902086-00 4-Channel Thermocouple Data Logger
IFC200 900298-00 Software, manual and USB interface cable
U9VL-J 901804-00 Replacement battery for QuadTemp
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Temperature
ASK ABOUT
OUR OTHER
DATA
LOGGERS
DOC-1358009-00 REV 2 2019.05.28
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current
Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion