QUADTCTEMP2000V2
4-CHANNEL THERMOCOUPLE
DATA LOGGER
WITH LCD DISPLAY
Features
• Large Backlit LCD Display
• User-Friendly Front Panel
• Accepts thermocouple types
J, K, T, E, R, S, B, N
• High speed downloading
• On-screen statistics
• External power or user
replaceable battery
• Low battery indicator
Applications
• Monitor multiple points
• Warehouse monitoring
• Refrigerator monitoring
• Medical and Pharmaceutical
• Oven monitoring
• Smoke houses
• Food Processing
• HVAC
• Engine Studies
The QuadTCTemp2000V2 is a four channel thermocouple data logger with an LCD display.
The device features on-screen minimum, maximum and average statistics, as well as a user
configurable graphics screen that allows for any combination of channels to be displayed.
The device accepts J, K, T, E, R, S, B and N type thermocouples.
The QuadTCTemp2000V2 is ideal for a variety of applications, whether it is remote
temperature monitoring, or multiple points in a central location. Data from all channels
is simultaneously logged and displayed on-screen for real-time monitoring. It can then be
downloaded to a PC for further analysis.
*Thermocouple Plugs/Probes Sold Separately
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Data table view
• Automatic report generation
• Summary view
• Multilingual
QUADTCTEMP2000V2 SPECIFICATIONS*
* Specifications are subject to change without notice. Specific warranty remedy limitations apply.
Call (603) 456-2011 or go to madgetech.com for details.
4 Internal Channels
Temperature Range: -20 ºC to +60 ºC (-4 ºF to +140 ºF)
Temperature Resolution: 0.01 °C (0.018 °F)
Calibrated Accuracy: ±0.5 ºC (0 to +50 ºC)
4 Remote Channels
Remote Channel
Thermocouple Types:
Thermocouple Connection: Female subminiature (SMP)
Cold Junction
Compensation:
Maximum Thermocouple
Resistance:
Thermocouple Range (ºC) Resolution Accuracy**
**Thermocouple accuracy is specified with a 24 AWG.
J, K, T, E, R, S, B, N
Automatic, based on internal channel
1000Ω, <100Ω recommended
J -210 to +760 0.1 ºC ±0.5 ºC
K -270 to +1370 0.1 ºC ±0.5 ºC
T -270 to +400 0.1 ºC ±0.5 ºC
E -270 to +980 0.1 ºC ±0.5 ºC
R -50 to +1760 0.5 ºC ±2.0 ºC
S -50 to +1760 0.5 ºC ±2.0 ºC
B +50 to +1820 0.5 ºC ±2.0 ºC
N -270 to +1300 0.1 ºC ±0.5 ºC
Start Modes:
Real Time Recording:
Memory:
(all 8 Channels enabled)
Software programmable immediate start or delay
start up to six months in advance
May be used with PC to monitor and record data
in real time
261,888 Readings per channel
Reading Rate: 4 Hz up to 1 reading every 24 hours
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type:
Battery Life:
9 Volt lithium or alkaline battery included,
user replaceable
18 months typical with display o. 3 months
typical with continuous display use.
Data Format: Date and time stamped ºC, ºF, K, ºR, mV, V
Time Accuracy: ±1 minute/month
Computer Interface: USB (Interface cable required); 115,200 baud
Software:
Operating Environment:
Dimensions:
Windows XP SP3 or later
MadgeTech 4 software, version 4.2.12.0 or later
-20 ºC to +60 ºC (-4 ºF to +140 ºF),
0 %RH to 95 %RH non-condensing
7.24 in x 2.7 in x 1.14 in
(183 mm x 68 mm x 29 mm)
Material: Black anodized aluminum
Weight: 14.8 oz (420 g)
Approvals: CE
BATTERY WARNING: DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF
CHILDREN. DO NOT DISPOSE OF IN FIRE, RECHARGE, PUT IN BACKWARDS, DISASSEMBLE,
OR MIX WITH OTHER BATTERY TYPES. MAY EXPLODE, FLAME, OR LEAK AND CAUSE
PERSONAL INJURY.
ORDERING INFORMATION
MODEL PART NUMBER DESCRIPTION
QUADTCTEMP2000V2 902089-00 4-Channel Thermocouple Data Logger with LCD Display
IFC200 900298-00 Software, manual and USB interface cable
U9VL-J 901804-00 Replacement battery for QuadTemp2000
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
ASK ABOUT
OUR OTHER
DATA
LOGGERS
Temperature
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current
Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
DOC-1364009-00 REV 1 2019.05.28