MADGETECH, INC. REVISION CONTROLLED DOCUMENT
ANSI/ISA S82.02.01 Standard:
Electric and Electronic Test, Measuring,
Controlling, and Related Equipment: General
Title
Requirements
Specifies the safety requirements that are
generally applicable to all equipment within its
scope. For certain types of equipment, these
requirements will be supplemented or modified
by the special requirements of a particular
standard that must be read in conjunction with
Abstract
Part 1 requirements. This International Standard
specifies general safety requirements for
electrical equipment intended for professional,
industrial process, and educational use,
including equipment and computing devices for:
measurement and test, control, laboratory use,
and accessories intended for use with the above
(e.g., sample handling equipment).
PO Box 50 Warner, NH 03278
879 Maple Street Contoocook, NH 03229
Phone 603.456.2011 Fax 603.456.2012
www.madgetech.com support@madgetech.com
Document Number:
Page Number: 4 of 4
Revision: B
Date Revised: 09/22/2010
DOC-0080009-00
MADGETECH, INC. REVISION CONTROLLED DOCUMENT
Document Number:
Page Number:
Revision: B
Date Revised: 09/22/2010
DOC-0080009-00
of 4
PRTemp1000IS Intrinsic Safety User Notice
Intrinsic Safety Approval:
The PRTemp1000IS has been certified by FM Approvals as Intrinsically Safe
(IS) for use in Class I, Division 1, groups A, B, C, D, and Nonincendive (NI) for
use in Class I, Division 2, groups A, B, C, D indoor hazardous (classified)
Locations. The rating is listed as follows in the Factory Mutual approval guide:
PRTemp1000is. Pressure and Temperature Recorder.
IS / I / 1 / ABCD T4A Ta=80°C; NI / I / 2 / ABCD / T4A Ta =80°C
Special Condition of Use:
1. Use only with Tadiran Battery P/N TL-2150
These are the only safety ratings relevant to the use of this product. Use of this
product in hazardous environments not specifically covered by this rating is
prohibited, unless the user takes the appropriate steps to ensure the safety of
the product and assumes full responsibility for its safe use. Refer to the
reference sections at the end of this document for further information on
approval standards and environments.
Conditions of Use:
The following conditions must be satisfied to maintain the IS rating of the
PRTemp1000IS:
• When used in hazardous locations, the PRTemp1000IS must be
installed prior to the location becoming hazardous, and
removed only after the area is no longer hazardous.
• The maximum allowed ambient temperature for the PRTemp1000IS
(under any circumstances) is 80°C. The minimum rated operating
temperature is -40°C.
• The PRTemp1000IS is approved for use only with the Tadiran TL2150 battery. Replacement with any other battery will void the
safety rating.
• Batteries are user replaceable, but they may only be removed or
replaced in locations that are known to be non-hazardous.
• Tampering or replacement of nonfactory components may adversely
affect the safe use of the product, and is prohibited. Except for
replacement of the battery, the user may not service the
PRTemp1000IS. MadgeTech, Inc. or an authorized representative
must perform all other service to the product.
MADGETECH, INC. REVISION CONTROLLED DOCUMENT
Document Number:
Page Number: 2 of 4
Revision: B
Date Revised: 09/22/2010
DOC-0080009-00
• The PRTemp1000IS enclosure does not carry a NEMA rating. For this
reason, the product is only suitable for use in indoor locations.
• For enclosure requirements, please co nsult Standard ANSI/ISA
S82.02.01 (reference on page 4) or other applicable ordinary location
standards.
Required Labeling:
The following label must be affixed to the PRTemp1000IS. It contains critical
information for the safe use of the product.
Reference Standards:
The PRTemp1000IS has been found to comply with the following standards:
Standard Date Title
FM Class 3600 1998 Electrical Equipment for Use in
Hazardous (Classified) Locations,
General Requirements
FM Class 3610 1999 Intrinsically Safe Apparatus and
Associated Apparatus for Use in Class I,
II and III, Division 1 and Class I, Zone 0
and 1 Hazardous (Classified) Locations
FM Class 3611 2004 Nonincendive Electrical Equipment for
Use in Class I and II, Division 2 and
Class III, Divisions 1 and 2 Hazardous
(Classified) Locations
FM Class 3810
Including
1995
2005
Electrical and Electronic Test, Measuring
and Process Control Equipment
Supplement #1
MADGETECH, INC. REVISION CONTROLLED DOCUMENT
Document Number:
Page Number:
Revision: B
Date Revised: 09/22/2010
DOC-0080009-00
of 4
Protection and Environment Reference:
The environmental rating is per ANSI/NFPA 70 National Electric Code® (NEC®)
Article 500. The following information is excerpted from FM Approvals reference
documents.
Protection Concepts
Type of
Protection Code Permitted Use Standard
Intrinsic Safety (IS) Class I, Division 1 FM Class 3610
Non-Incendive (NI) Class I, Division 2 FM Class 3611
Apparatus Grouping Per NEC
Class Group Typical Gas
I
Area Classification Per NEC
Division Description
1
2 Flammable Material Present Abnormally
Temperature Class Per NEC
Temperature
Class
T4A 120 °C (with maximum 80 °C ambient)
Note that the T4A rating indicates the maximum surface temperature that may
be encountered in a fault conditio n. This is not the allowed operating
temperature. This temperature rating limits the maximum amb ient
temperature of 80 °C.
®
500:
A Acetylene
B Hydrogen
C Ethylene
D Propane
®
500:
Flammable Material Present Continuously
Flammable Material Present Intermittently
®
500:
Maximum Surface Temperature (of any
component under fault conditions)