MadgeTech PRTEMP1000 Operating Manual

Description PRTemp1000
Temperature Sensor Internal semiconductor
Temperature Range -40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution 0.1 °C (0.18 °F)
Calibrated Accuracy ±0.5 °C (±0.9 °F)
Pressure Range
*See Table for DetailsPressure Resolution
Pressure Accuracy
Memory 16,383/channel
Reading Rate 1 reading every 2 seconds up to 1 reading every 12 hours
Units PSIA(G), inches, altitude, Torr, mmHg, Pascals
Required Interface Package IFC200
Baud Rate 2,400
Typical Battery Life 1 year
Operating Environment -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH
Material 303 stainless steel
Dimensions 6.4 in x 1.25 in dia. (163 mm x 32 mm dia.)
Weight 12 oz (340 g)
Approvals CE
*PRTemp1000 Range, Accuracy and Resolution
Range (PSI)
Accuracy
Resolution (PSI)
0-30
PSIA/PSIG
0.002 0.005 0.02 0.05 0.05 0.2
0-100
PSIA/PSIG
0-300
PSIA/PSIG
2 %FSR, 0.25 % @ 25 °C typical
0-500
PSIA/PSIG
0-1000
PSIA
0-5000
PSIA
Product User Guide
PRTemp1000
PRTemp1000
Rugged Pressure and Temperature Data Logger
Battery Warning
WARNING: FIRE OR EXPLOSION RISK, DO NOT CHARGE, OPEN, HEAT OR DISPOSE OF IN FIRE.
Specifications subject to change.
See MadgeTech’s terms and conditions at madgetech.com
MadgeTech, Inc.
6 Warner Road
Phone (603) 456.2011
madgetech.com
l
l
Fax (603) 456.2012
l
info@madgetech.com
DOC-1029035-00 REV 16 2020.02.18
To view the full MadgeTech product line,
visit our website at madgetech.com.
PRTemp1000
Product User Guide
Product Notes
Getting Started
To access the COM Port for the interface cable, unscrew the key-ring end cap.
Pressure Sensor
To use the pressure sensor for gauge measurements, screw the 1/4 inch male NPT fitting into
the pipe to be measured with a 9/16 inch wrench. The NPT fitting can be used for absolute
and submersible applications, but in those cases it’s preferable to use the submersible fitting.
Submergibility
The PRTemp1000 is fully submergible and is rated IP68. It can be placed in environments
with up to 230 feet (70 m) of water.
O-Rings
O-ring maintenance is a key factor when properly caring for the PRTemp1000. The o-rings
ensure a tight seal and prevent liquid from entering the inside of the device.
Please refer to the application note “O-Rings 101: Protecting Your Data”, found on the
MadgeTech website, for information on how to prevent O-ring failure.
Installation Guide
Installing the Interface cable
- IFC200
Insert the device into a USB port. The drivers will install automatically.
Installing the software
The Software can be downloaded from the MadgeTech website at the following link at
madgetech.com. Follow the instructions provided in the Installation Wizard.
dataset as it records while connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to Manual Start”, depending upon your start method.
- Disconnect the data logger from the interface cable and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement
Materials: Small Needle Nose Pliers and a Replacement Battery (TLH-5902)
- Carefully unscrew the sensor end cap and pull the electronics out.
- The battery is the purple cylinder on the circuit board.
- Gently pull out the old battery.
- Insert the new battery one lead at a time, using pliers to fully push the leads into the
sockets. The battery should be flat against the circuit board, and the positive lead should
be closest to the communications jack.
- Ensure the circuit board is inserted into the white plastic bushing. The sensor cable
should not be twisted, or kinked. From the connection to the circuit board, it should run
up towards the battery, then down to the sensor.
- Insert the electronics back into the tube and carefully screw the cap on.
Recalibration
MadgeTech recommends annual recalibration. To send devices back for calibration, visit
madgetech.com.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “Custom Start” from the menu bar and choose the desired start method, reading rate and other parameters appropriate for the data logging application and click “Start”. (Quick Start” applies the most recent custom start options,
Batch Start” is used for managing multiple loggers at once, “Real Time Start” stores the
Call for custom calibration options to accommodate specific application needs. Prices and specifications subject to change. See MadgeTech’s terms and conditions at madgetech.com.
Loading...