PRTC210
THERMOCOUPLE TEMPERATURE
AND PRESSURE
DATA LOGGER
Features
• Up to 2 year battery life
• Wide temperature range
• Accepts thermocouple types
J, K, T, E, R, S, B, N
• High speed download
• Real time operation
• Low cost
• Programmable start time
• Miniature size
Applications
• Dry Ice Shipments
• Factory systems
• Laboratory testing
• HVAC
• Environmental studies
The PRTC210 is a pressure and thermocouple tempera-
ture data logger, ideal for use in factory settings.
The device is available from 30 to 5000 PSI and features
a 1/8 in NPT tting for quick connection.
The PRTC210 has a real-time clock and a battery
life of up to 2 years. This all-in-one compact,
portable, easy to use device has a high-speed
download and can measure and record up
to 10,922 readings per channel. The storage medium is non-volatile solid state memory,
providing maximum data security even if the battery becomes discharged. The PRTC210
can be started and stopped directly from your computer and its small size allows it to t
almost anywhere.
The PRTC210 makes data retrieval quick and easy. Simply plug it into an empty COM or USB
port and our user-friendly software does the rest.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Data table view
• Automatic report generation
• Summary view
• Multilingual
PRTC210 SPECIFICATIONS*
Internal Channel
Temperature Range: -20 ºC to +80 ºC (-4 ºF to +176 ºF)
Temperature Resolution: 0.1 ºC (0.18 ºF)
Calibrated Accuracy: ±0.5 ºC (±0.9 ºF)
Remote Channel
Thermocouple Types: J, K, T, E, R, S, B, N
Thermocouple Connection: Removable Mini Plug
Cold Junction Compensation: Automatic, based on internal channel
Maximum Thermocouple
Resistance:
**Thermocouple Type: Range (ºC) Resolution Accuracy
Voltage
Pressure Sensor: Semi Conductor Strain Gauge
Pressure Range:
Pressure Resolution:
Calibrated Accuracy: 2 % FSR, 0.25 % @ 25 ºC typical
Pressure Response Time: 0.1 ms (10 % to 90 % FSR)
Adapter: 1/8 in Male NPT
1000 Ω, <100 Ω recommended
-210 to +760 0.1 ºC ±0.5 ºC
J
-270 to +1370 0.1 ºC ±0.5 ºC
K
-270 to +400 0.1 ºC ±0.5 ºC
T
-270 to +980 0.1 ºC ±0.5 ºC
E
-50 to +1760 0.5 ºC ±2.0 ºC
R
-50 to +1760 0.5 ºC ±2.0 ºC
S
+50 to +1820 0.5 ºC ±2.0 ºC
B
-270 to +1300 0.1 ºC ±0.5 ºC
N
-75 mV to +75 mV 5 μV ±20 μV
0 to 30, 100, 300 500 PSIA/G; 1000
PSIA, 5000PSIA
0 to 1000 and 5000PSIA
0.002, 0.005, 0.02, 0.05PSIA/G;
0.05, 0.2PSIA
*SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE. SPECIFIC WARRANTY REMEDY LIMITATIONS
APPLY. CALL 1-603-456-2011 OR GO TO WWW.MADGETECH.COM FOR DETAILS.
Start Modes:
Real Time Recording:
Memory:
Reading Rate:
Software programmable immediate start or delay
start up to six months in advance
May be used with PC to monitor and record data
in real time
10,922 readings per channel
readings; software congurable memory wrap
1 reading every 2 seconds up to 1 reading every
12 hours, selectable in software
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6V lithium battery included, user replaceable
Battery Life: Up to 2 years
Date and time stamped ºC, ºF, ºK, ºR; ºC, ºF, ºK,
Data Format:
ºR, mV; inH2O, PSIG, inHg, mmHg, bar, atm,
Torr, Pa, kPa, and MPa.
Time Accuracy: ±1 minute/month (at 20 ºC to 30 ºC)
Computer Interface:
PC serial or USB (interface cable required);
57,600 baud
Software: XP SP3/Vista/Windows 7/Windows 8
Operating Environment:
Dimensions:
-20 ºC to +80 ºC (-4 ºF to +176 ºF),
0 %RH to 95 %RH non-condensing
2.2 in x 2.5 in x 0.9 in
(55 mm x 63 mm x 22 mm) plus NPT Fitting
Enclosure: Delrin
Weight: 4.2 oz (119 g)
BATTERY WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT RECHARGE,
DISASSEMBLE, HEAT ABOVE 212 °F, INCINERATE OR EXPOSE CONTENTS TO WATER
**Accuracy specications based on 3-meter 24 AWG thermocouple.
ORDERING INFORMATION
MODEL DESCRIPTION
PRTC210 Pressure and Thermocouple Temperature Recorder
IFC200 Software, manual and USB interface cable
Calibration Calibration Certicate
LTC-7PN Replacement battery for PRTC210
For Quantity Discounts call 603-456-2011 or email sales@madgetech.com
*To order the product with the calibration certicate add -CERT to the end of the part number and add $90.00 to the price.
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Temperature
ASK ABOUT
OUR OTHER
DATA
LOGGERS
DOC-1158009-00 REV 11 08.07.2017
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current
Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion