MadgeTech PRTC110 Reference Manual

Red 485 Dark Blue Pantone 289 Light Blue 58% of Pantone 289
Description PRTC110
Internal Channel Temperature Sensor Semiconductor
Internal Channel Temperature Range -20 °C to +80 °C
Internal Channel Temperature Resolution 0.1 °C
Internal Channel Temperature Accuracy +0.5 °C
Remote Channel Temperature Sensor
Remote Channel Temperature
Range, Resolution and Accuracy
Pressure Range (Nominal) 0 to 2 inH2O (differential) 0 to 5 inH2O (differential)
Pressure Range (Max) -2 to +4 inH2O (differential) -2 to +8 inH2O (differential)
Pressure Resolution 0.002 inH2O 0.005 inH2O
Pressure Accuracy 0.5%FSR, 0.25% @ 25 °C Typical
Cold Jct. Compensation Automatic
Channels 2 Internal & 1 Remote
Memory 21,844/channel
Sample Rate 2 seconds up to 12 hours
LED Indicator Red
Required Interface Package IFC110 or IFC200
Baud Rate 57,600
Typical Battery Life 2 years
Operating Environment –20 °C to +80 °C, 0 to 95%RH (non-condensing)
Dimensions 2.7” x 1.6” x 0.8” (68 mm x 41 mm x 20 mm)
Material ABS plastic
Weight 8.8 oz (250 g)
Approvals CE
* Remote Channel
Thermocouple Range (°C) Resolution Accuracy
J -210 to +760 0.1°C +0.5°C K -260 to +1370 0.1°C +0.5°C T -260 to +400 0.1°C +0.5°C E -260 to +980 0.1°C +0.5°C R -50 to +1760 0.5°C +2.0°C S -50 to +1760 0.5°C +2.0°C B +60 to +1820 0.5°C +2.0°C
N -260 to +1300 0.1°C +0.5°C
Thermocouple Types
J, K, T, E, R, S, B, N
*See Table Below for Details
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 80 °C (176 °F).
Specifications subject to change.
See MadgeTech’s terms and conditions at www.madgetech.com
Product Information Card
PRTC110
PRTC110
Differential Pressure and Thermocouple Temperature Data Logger
MadgeTech, Inc.
6 Warner Road
Phone 603.456.2011
www.madgetech.com
l
Warner, NH 03278
l
Fax 603.456.2012
l
info@madgetech.com
DOC-1170035-00 REV 11 2014.03.07
To view the full MadgeTech product line,
visit our website at www.madgetech.com.
PRTC110
Product Quick Reference Card
Product Notes
Pressure Sensor
The PRTC110 measures differential pressure in the range of 0 to 2inH2O or 0 to 5inH20,
depending upon the range purchased. The device is also equipped with one thermocouple
input. Two 0.080” diameter pressure ports are located on the side of the device. The
pressure ports will accept 1/16”ID flexible rubber tubing with up to 5/32”OD. The tubing
can be connected to two different pressure locations in which the device will measure the
difference between the two pressure’s.
LEDs
Once started, the LED will flash at the selected reading rate to indicate that the device is
running.
Thermocouple Type
-To change the thermocouple type in the MadgeTech software:
- In the Connected Devices panel, click the device desired.
- On the Device Tab, in the Information Group, click Properties. Or, right-click the device
and select Properties in the context menu.
- On the General Tab, change the Thermocouple type in the drop down menu.
- Apply these changes, there will be a prompt to reset the device, select yes.
Please note that the same thermocouple type must be used on all of the channels.
Installation Guide
Installing the Interface cable
- IFC200
Insert the device into a USB port. The drivers will install automatically.
- IFC110
Plug the serial cable into the port and verify it is secure.
Installing the software
Insert the Software USB Stick in an open USB port. If the autorun does not appear, locate
the drive on the computer and double click on Autorun.exe. Follow the instructions
provided in the Wizard.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “Custom Start” from the menu bar and choose the desired
start method, reading rate and other parameters appropriate for the data logging
application and click “Start”. (Quick Start” applies the most recent custom start options,
Batch Start” is used for managing multiple loggers at once, “Real Time Start” stores the
dataset as it records while connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to
Manual Start”, depending upon your start method.
- Disconnect the data logger from the interface cable and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Connect the logger to the interface cable.
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will
be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement
Materials:
Small Phillips Head Screwdriver
Replacement Battery (LTC-7PN)
- Puncture the center of the back label with the screw driver and unscrew the enclosure.
- Remove the battery by pulling it perpendicular to the circuit board.
- Insert the new battery into the terminals and verify it is secure.
- Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The PRTC110 standard calibration is two pressure points at 14.7 PSI and 30 PSI and one
point at 0mVs for the thermocouple channel.
Pricing:
Recalibration traceable to NIST $110.00
Recalibration $70.00
Additional Services:
Verification Point $15.00 per point
Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com To send the devices back, visit www.madgetech.com, select Services then RMA Process.