OCTRTDTEMPV2
8-CHANNEL RTD TEMPERATURE
DATA LOGGER
Features
• ±0.04 °C Calibrated Accuracy
• Real time Operation
• Up to 4 Hz Reading Rate
• Pushbutton or Programmable
Start Time
• Title Channels
• Calibration Certificate included
• External Power or User
Replaceable Battery
• Accepts 2, 3 and 4-wire RTD’s
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Calibration Chamber Monitoring
• Process Verification/Validation
• Warehouse Monitoring
• Precision Temperature Monitoring
• HVAC
• Clean Rooms
• Medical/Pharmaceutical
• Museum Monitoring
• Environmental Studies
• Replace Costly Strip Chart
Recorders
The OctRTDTempV2 is an 8-channel, RTD based temperature data logger with a reading rate of
up to 4 Hz. It can measure and record up to 174,592 readings per channel. To maximize memory
capacity, users can enable or disable channels. For easy identification, each channel
can be named with up to a ten digit title. The storage medium is non-volatile solid state
memory, providing maximum data security even if the battery becomes discharged. The
OctRTDTempV2 makes data retrieval quick and easy.
Using the MadgeTech Software, starting, stopping and downloading from the OctRTDTempV2
is simple and easy. Graphical, tabular and summary data is provided for analysis and data
can be viewed in multiple units, using the Engineering Units function. The data can also be
automatically exported to Excel® for further calculations.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Data table view
• Automatic report generation
• Summary view
• Multilingual
OCTRTDTEMPV2 SPECIFICATIONS*
* Specifications are subject to change without notice. Specific warranty remedy limitations apply.
Call (603) 456-2011 or go to madgetech.com for details.
**Temperature
Measurement Range:
Resolution:
***Calibrated Accuracy:
Specified Accuracy Range:
18.5 Ω to 390.5 Ω
-200 ºC to +850 ºC (-328 ºF to +1562 ºF)
(Probe Dependent)
0.0005 Ω
0.01 °C (0.018 °F) (Probe Dependent)
±0.015 Ω @ 25 ºC (77 °F) ambient
±0.04 °C (±0.072 °F) (Probe Dependent)
18.5 Ω to 175.9 Ω
-200 °C to +200 °C (-328 °F to 392 °F)
Channels: 8
Input Connection:
Removable screw terminal; 2, 3 or 4
wire interface
Temperature Eect on Span: 2 ppm/°C typical
Software programmable immediate
Start Modes:
start or delay start up to six months in
advance, programmable in software
** Temperature specifications based on ideal 100 Ω Pt RTD compliant with IEC 751
(1983) and ITS-90, 5000 Ω FSR (accuracy based on 36 in lead wire RTD with
4 wire configuration)
*** Calibrated accuracies based on standard MadgeTech calibrations for 0 Ω to
200 Ω range.
Warning: Note the polarity instructions. Do not attach wires to the wrong terminals.
100 Ω, 2 or 4 wire RTD probes are recommended for the most accurate performance.
Most 100 Ω, 3 wire RTD probes will work, but MadgeTech cannot guarantee the accuracy. To determine whether or not the 3-wire RTD probe will work, the resistance
between the two same colored wires should be less than 1 Ω. (Note: Please contact the
manufacturer of the RTD probe for questions on the resistance.)
Real Time Recording:
Memory:
(all 8 channels enabled)
May be used with PC to monitor and record data
in real time
174,592 readings per channel
Reading Rate: 4 Hz up to 1 reading every 24 hours
Calibration: Digital calibration through the software
Calibration Date: Automatically recorded within device
Battery Type:
9V lithium or alkaline battery included,
user replaceable
Battery Life: 1 year typical
Data Format: Date and time stamped ºC, ºF, K, ºR; Ω
Time Accuracy: ±1 minute/month
Computer Interface: USB (Interface cable required); 115,200 baud
Software:
Operating Environment:
Dimensions:
Windows XP SP3 or later
MadgeTech 4 software, version 4.2.12.0 or later
-20 ºC to +60 ºC (-4 ºF to +140 ºF),
0 %RH to 95 %RH non-condensing
2.7 in x 7.2 in x 1.3 in
(69 mm x 184 mm x 32 mm)
Weight: 17 oz (490 g)
BATTERY WARNING: BATTERY MAY LEAK, FLAME OR EXPLODE IF DISASSEMBLED, SHORTED,
CHARGED, CONNECTED TOGETHER, MIXED WITH USED OR OTHER BATTERIES, EXPOSED TO
FIRE OR HIGH TEMPERATURE. DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF
CHILDREN.
ORDERING INFORMATION
MODEL PART NUMBER DESCRIPTION
OCTRTDTEMPV2 902073-00 8-Channel RTD Temperature Data Logger
IFC200 900298-00 Software, manual and USB interface cable
U9VL-J 901804-00 Replacement battery for OctRTDTempV2
ASK ABOUT
OUR OTHER
DATA
LOGGERS
Temperature
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current
Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
DOC-1346009-00 REV 4 2019.05.28