Description HiTemp140-TSK HiTemp140-PT-TSK
Temperature Sensor 100 Ω platinum RTD
Temperature Range –200 °C to +260 °C (-328 °F to +500 °F) -200 °C to +350 °C (-328 °F to +662 °F)
Temperature Resolution 0.01 °C
Temperature Accuracy
Memory 32,700 readings
Reading Rate 1 reading every second up to 1 reading every 24 hours
Required Interface Package IFC400 or IFC406 USB docking station required; 125,000 baud
Typical Battery Life 1 year (1 minute reading rate at 25 °C/77 °F)
Operating Environment –40 °C to +140 °C (-40 °F to +284 °F), 0 %RH to 100 %RH
Material Data Logger: 316 stainless steel, Thermal Shield: PTFE
Dimensions: Body 2.7 in x 0.97 in dia. (67 mm x 25 mm dia.)
Dimensions: Shield
Dimensions:
HiTemp140-5.25-TSK
Dimensions:
HiTemp140-7-TSK
Dimensions:
HiTemp140-PT-1-TSK
Dimensions:
HiTemp140-PT-5-TSK
Weight
Maximum Exposure
Ambient Tempertature Exposure Time in Air Exposure Time In Liquid Exposure Time in Air Exposure Time In Liquid
Approvals CE
Time Chart
-200 °C (-328 °F) 12 minutes N/A 14 minutes N/A
-180 °C (-292 °F) 1
-160 °C (-256 °F) 15 minutes N/A 16 minutes N/A
-140 °C (-220 °F) 17 minutes N/A 18 minutes N/A
-120 °C (-184 °F) 19 minutes N/A 21 minutes N/A
-100 °C (-148 °F) 22 minutes N/A 24 minutes N/A
-80 °C (-112 °F) 27 minutes N/A 30 minutes N/A
-60 °C (-76 °F) 37 minutes 22 minutes 42 minutes 25 minutes
-40 °C to +140 °C
(-40 °F to +284 °F)
150 °C (302 °F) 59 minutes 34 minutes 66 minutes 40 minutes
160 °C (320 °F) 51 minutes 29 minutes 57 minutes 34 minutes
170 °C (338 °F) 43 minutes 25 minutes 48 minutes 29 minutes
180 °C (356 °F) 37 minutes 23 minutes 42 minutes 26 minutes
190 °C (374 °F) 34 minutes 20 minutes 38 minutes 23 minutes
200 °C (392 °F) 31 minutes 18 minutes 34 minutes 21 minutes
210 °C (410 °F) 29 minutes 17 minutes 32 minutes 19 minutes
220 °C (428 °F) 27 minutes 16 minutes 30 minutes 18 minutes
230 °C (446 °F) 25 minutes 15 minutes 27 minutes 17 minutes
240 °C (464 °F) 23 minutes 14 minutes 26 minutes 16 minutes
250 °C (482 °F) 22 minutes 13 minutes 24 minutes 15 minutes
5.25 in x 0.188 in dia.
(134 mm x 4.8 mm dia.)
7.0 in x 0.188 in dia.
(178 mm x 4.8 mm dia.)
3 minutes N/A 15 minutes N/A
Indefinitely Indefinitely Indefinitely Indefinitely
±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F)
±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F)
±0.4 °C/±0.72 °F (-40 °C to -20.01 °C/-40 °F to -4.02 °F)
Flush Top: 2.75 in x 2.0 in dia. (69.85 mm x 51 mm dia.)
Vented Top: 4.3 in x 2.0 in dia. (109.2 mm x 51 mm dia.)
n/a
n/a
n/a
n/a
Flush Top: 8.2 oz (232.5 g) (not including data logger)
Vented Top: 9.9 oz (280.7 g) (not including data logger)
Probe tip: 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.)
Flexible portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
Probe tip: 4.8 in x 0.125 in dia. with 1 in x 0.188 in dia. handle
(121 mm x 3.2 mm dia. with 25 mm x 4.8 mm dia. handle)
Flexible portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
HiTemp140-TS (Flush) HiTemp140-TS (Vented)
Product User Guide
HiTemp140-PT-TSK and
Ven ted
HiTemp140-PT-TSK
High Temperature Data Logger with a 24 inch Stainless Steel
Flexible Probe and Thermal Shield
HiTemp140-TSK
High Temperature Data Logger with Thermal Shield
HiTemp140-TSK
Flush
DOC-1238035-00
REV 12 2019.01.25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
To view the full MadgeTech product line,
visit our website

HiTemp140-TSK and HiTemp140-PT-TSK
Product User Guide
Product Notes
Getting Started
The HiTemp140-TSK consists of a HiTemp140 data logger in a thermal enclosure and the
HiTemp140-PT-TSK consists of a HiTemp140-PT dat a logger in a thermal enclosure. They can be
used in both wet and dry applications up to 140 °C indefinitely. When used with the thermal
enclosure the devices can withstand higher temperatures for certain durations of time. The chart
located on the back page, outlines the time versus temperature durations.
- To start the logger, unscrew the thermal enclosure and separate the thermal enclosure lid,
from the body of the data logger.
- The logger should be removed from the barrier immediately after removal fr
environment. Be extremely careful, the logger may be very hot.
om the heat
Submergibility
The HiTemp140 and the HiTemp140-PT are fully submergible and are rated IP68. They can be
placed in environments with up to 230 feet (70 m) of water.
Installation Guide
Installing the Interface cable
- IFC400 or IFC406
Refer to the “Quick Start Guide” included in the package.
Installing the software
Insert the Software USB stick in an open USB port. If the autorun does not appear, locate the drive
on the computer and double click on Autorun.exe. Follow the instructions provided in the Wizard.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the docking station.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger
- For most applications, select “Custom Start” from the menu bar and choose the desired start
method, reading rate and other parameters appropriate for the data logging application and
click “Start”. (“Quick Star t” applies the most recent custom start options, “Batch Start” is used
for managing multiple loggers at once, “Real Time Start” stores the dataset as it records while
connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to Manual
Start”, depending upon your star t method.
- Disconnect the data logger from the interface cable and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device
cannot be restarted until it has been re-armed by the computer.
.
Device Maintenance
Battery Replacement
Materials:
ER1425S-HT Battery
- Unscrew the bottom of the logger and remove the battery.
- Place the new battery into the logger. Note the polarity of the battery.
- Screw the cover back onto the logger.
Recalibration
The HiTemp140 and HiTemp140-PT standard calibrations are two points at 30 °C and 140 °C.
Prices and specifications subject to change. See MadgeTech’s terms and conditions
Disclaimer and T
Listed specifications can be used to determine maximum allowable exposure times for the
HiTemp140 with Thermal Shield at dierent temperatures beyond the normal operating range
of the logger. Both the data logger and Thermal Shield must be at ambient temperature
(approximately 25 °C) before being placed in the extreme temperature environment.
Immediately following exposure to high temperature, the data logger should be removed from
the thermal shield (using appropriate precautions, as it could be VERY hot) OR the data logger and
shield should be placed in a water bath (approximately 25 °C) for at least 15 minutes to allow it to
cool. Failing to do this may allow heat trapped in the Thermal Shield to continue to heat the data
logger to potentially unsafe levels.
If your application involves a ramp up to a temperature above 140 °C and/or any complex
temperature profile that isn’t simply a constant temperature, please contact MadgeTech to
determine whether the HiTemp140 with Thermal Shield is suitable.
Please provide MadgeTech with a detailed description of your temperature profile, including
temperatures, durations, ramp times, and process media (air, steam, oil, water, etc.)
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER
DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE
140 °C (284 °F).
erms of Use
Specifications subject to change.
See MadgeTech’s terms and conditions
Downloading data from a data logger
- Place the logger into the docking station.
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will be
prompted to name your report.
- Downloading will ooad and save all the recorded data to the PC.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com