MadgeTech HITEMP140-FR Specifications

HITEMP140-FR
HIGH TEMPERATURE DATA LOGGER WITH 2” FAST RESPONSE PROBE
HiTemp140-FR Features
• Fast Response Time
• ±0.1 °C (0.18 °F) Accuracy
• Measures from -200 ºC to +260 ºC (-328 ºF to +500 ºF)
• Submersible (IP68)
• NIST traceable
• User Replaceable Battery
• Trigger Settings
• Programmable Start and Stop Time
• 4 Hz Reading Rate
• Battery Life Indicator
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Autoclave Verification and Mapping
• Seafood Processing
- Smoked Lobsters, Shelfish & Shrimp
• Steam Sterilization
• Oven Profiling
• Roasting
• Flash Freezing
• Frying
HiTemp140-FR-TSK Features
• Withstands Temperatures between
-200 ºC up to 250 ºC
• Submersible
• Vented or Flush Enclosure Options
Benefits
• Validate a Wide Range of Temperature Processes Using One Data Logger
• Durable Thermal Shield Protects Probe and Allows for a Fast Response Time
MadgeTech has designed the HiTemp140-FR, a high temperature data logger
with an ultra-fast response time, to record temperature during rapidly
changing thermal processes. This high temperature data logger features
a 2 inch x 0.0625 inch diameter probe and is capable of recording
this class. This allows the temperature sensor to quickly adapt and
accurately record temperature variations in changing environments.
Applications include oven profiling, steam sterilization, chamber
mapping, seafood processing, flash freezing and more.
MadgeTech’s entire HiTemp140 data logger series is designed
with food grade stainless steel. The device can be placed in
environments up to 140 °C and the probe is capable of measuring
from -200 °C to +260 °C (-328 °F to +500 °F). The HiTemp140-FR
is also available with an optional thermal shield enclosure to extend
the operating range of the data logger to -200 °C to +250 °C
(-328 °F to +482 °F). The HiTemp140-FR-TSK (Thermal Shield Kit)
comes with either a vented or flush top enclosure to accommodate
a multitude of applications.
The Trigger Settings feature of the HiTemp140-FR allows users to configure high and low temperature thresholds
that when met or exceeded, will automatically start or stop recording data to memory. This data logger is capable
of storing up to 32,700 date and time stamped readings and features a nonvolatile solid state memory which retains
data even if the battery becomes discharged.
The HiTemp140-FR Fast Response High Temperature Data Logger is available for $499.00, or the Hi-Temp140-FR-TSK
Data Logger and Thermal Shield Kit is available for $549.00, with volume discounts available. To speak with our
knowledgeable sales staff, please call (603) 456-2011.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
Multiple graph overlay
Statistics
Digital calibration
Zoom in/ zoom out
Lethality equations (F0, PU)
Mean Kinetic Temperature
Full time zone support
Data annotation
Min./Max./Average lines
Data table view
Automatic report generation
Summary view
Multilingual
HITEMP140-FR SPECIFICATIONS*
Temperature
Temperature Sensor: 100 Ω Platinum RTD
Probe Measurement Range: -200 ºC to +260 ºC (-328 ºF to +500 ºF)
Temperature Resolution: 0.01 ºC (0.02 ºF)
Calibrated Accuracy: ±0.1 ºC/±0.18 ºF (20 ºC to +140 ºC/68 ºF to +284 ºF)
General
Data Logger Response Time:
Reading Rate: 4 readings per second up to 1 reading every 24 hours
Memory: 32,767 readings
Start Modes:
Stop Modes: Manual or Timed (specific date and time)
Trigger Settings:
Readings in Trigger
Settings Mode:
Real Time Recording: May be used with PC to monitor & record data in real time
Password Protection:
In Air In Water
t60 - 00:00:39:00 t90 - 00:02:71:00
• Software programmable immediate start
• Delay start up to 18 months in advance
High & Low limits may be set. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers.
10,922 readings
An optional password may be programmed into the device to restrict access to configuration options. Data may be read without the password.
t60 - 00:00:10:00 t90 - 00:00:12:00
*SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE. SPECIFIC WARRANTY REMEDY LIMITATIONS
APPLY. CALL 1-603-456-2011 OR GO TO WWW.MADGETECH.COM FOR DETAILS.
Memory Wrap Around: Yes
Battery Type: 3.6V high-temperature lithium battery included; user replaceable
Battery Life: 2 years typical (1 minute reading rate)
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Data Format: Date and time stamped °C, °F, °R, K
Time Accuracy:
Computer Interface: IFC400 or IFC406 USB docking station required; 125,000 baud
Operating System
Compatibility:
MadgeTech Software
Compatibility:
Operating Environment:
IP Rating: IP68
Dimensions (body):
Dimensions (probe):
Weight: 65 g (2.3 oz)
Materials: 316 Stainless Steel, PEEK
Approvals: CE
±1 minute/month at 25 ºC Extended Operation: ±20 minutes/month at 140 °C (±450 ppm)
XP SP3/Vista/Windows 7/Windows 8
• MadgeTech Standard Software version 4.2.3.0 or later
• MadgeTech Secure Software version 4.2.2.0 or later
-40 ºC to +140 ºC (-40 ºF to +284 ºF), 0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA
1.89 in x 0.970 in x 0.970 in (48 mm x 24.6 mm x 24.6 mm)
1.75 in x 0.0625 in dia. (0.188 in transitional dia.)
44.5 mm x 1.59 mm (4.78 mm Transitional dia.)
HITEMP140-FR-TSK SPECIFICATIONS*
HiTemp140-FR-TSK
(Flush)
Data Logger
Response
Time with
Thermal
Shield:
Dimensions
(enclosure):
Weight:
Operating
Environment:
Material: Enclosure: PTFE
Disclaimer and Terms of Use
Listed specifications can be used to determine maximum allowable exposure times for the HiTemp140 with Thermal Shield at different temperatures beyond the normal operating range of the logger. Both the data logger and Thermal Shield must be at ambient temperature (approximately 25 °C) before being placed in the extreme temperature environment. Immediately following exposure to high temperature, the data logger should be removed from the thermal shield (using appropriate precautions, as it could be VERY
hot) OR the datalogger and shield should be placed in a water bath (approximately 25 °C) for at least 15 minutes to allow it to cool. Failing to do this may allow heat
trapped in the Thermal Shield to continue to heat the datalogger to potentially unsafe levels. If your application involves a ramp up to a temperature above 140 °C and/or any complex temperature profile that isn’t simply a constant temperature, please contact MadgeTech to determine whether the HiTemp140 with Thermal Shield is suitable. Please provide MadgeTech with a detailed description of your temperature profile, including temperatures, durations, ramp times, and process media (air, steam, oil, water, etc.) If MadgeTech is unable to definitively calculate the suitability of our product for your application, we can provide a test unit outfitted with a high temperature indicator sticker. This sticker has an indicator dot which will turn black if exposed to temperatures above 143 °C. Apply the sticker to the bottom of the data logger itself (not the thermal shield), remove the battery for safety, place the data logger into the thermal shield and run the assembly through the proposed temperature program. The first indicator dot on the sticker will turn black at 143 °C. If that happens, the HiTemp140 with thermal shield is not appropriate for the application and we will work to find a solution that is.
BATTERY WARNING: WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO
NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, CRUSH, PENETRATE, OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 150 ºC (302 ºF).
Exposure
Time in Air
t60 -
0:00:30
t90 -
0:13:10
2.75 in x 2.0 in dia.
(69.85 mm x 51 mm dia.)
6.7 oz (190 g)
not including data logger
-200 ºC to +250 ºC (-328 ºF to +482 ºF) (Time limited) 0 %RH to 100 %RH
Exposure
Time in
Water
t60 -
0:00:09
t90 -
0:00:12
HiTemp140-FR-TSK
(Vented)
Exposure
Time in Air
t60 -
0:00:59
t90 -
0:22:10
4.3 in x 2.0 in dia.
(109.2 mm x 50.8 mm dia.)
9.5 oz (270 g)
not including data logger
Exposure
Time in
Water
t60 -
0:00:10
t90 -
0:00:12
Maximum Exposure
Ambient Temperature
-200 °C (-328 °F) 12 minutes N/A 14 minutes N/A
-180 °C (-292 °F) 13 minutes N/A 15 minutes N/A
-160 °C (-256 °F) 15 minutes N/A 16 minutes N/A
-140 °C (-220 °F) 17 minutes N/A 18 minutes N/A
-120 °C (-184 °F) 19 minutes N/A 21 minutes N/A
-100 °C (-148 °F) 22 minutes N/A 24 minutes N/A
-80 °C (-112 °F) 27 minutes N/A 30 minutes N/A
-40 °C to +140 °C
(-40 °F to +284 °F)
150 °C (302 °F) 59 minutes 34 minutes 66 minutes 40 minutes
160 °C (320 °F) 51 minutes 29 minutes 57 minutes 34 minutes
170 °C (338 °F) 43 minutes 25 minutes 48 minutes 29 minutes
180 °C (356 °F) 37 minutes 23 minutes 42 minutes 26 minutes
190 °C (374 °F) 34 minutes 20 minutes 38 minutes 23 minutes
200 °C (392 °F) 31 minutes 18 minutes 34 minutes 21 minutes
210 °C (410 °F) 29 minutes 17 minutes 32 minutes 19 minutes
220 °C (428 °F) 27 minutes 16 minutes 30 minutes 18 minutes
230 °C (446 °F) 25 minutes 15 minutes 27 minutes 17 minutes
240 °C (464 °F) 23 minutes 14 minutes 26 minutes 16 minutes
250 °C (482 °F) 22 minutes 13 minutes 24 minutes 15 minutes
ORDERING INFORMATION
MODEL DESCRIPTION PRICE (U.S.)
Hitemp140-FR-TSK HiTemp140-FR data logger and thermal shield (Flush or Vented). $549.00
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com *To order the product with the NIST certificate add -CERT to the end of the part number and add $60.00 to the price.
Time Chart
-60 °C (-76 °F) 37 minutes 22 minutes 42 minutes 25 minutes
HiTemp140-FR High Temperature Data Logger with a 2 inch Fast Response Probe. $499.00
HiTemp140-TS Thermal Shield for the HiTemp140-FR data logger (Flush or Vented). $99.00
IFC400 Docking station with USB cable, software and manual. $120.00
IFC406 6 Port, Multiplexer docking station with USB cable, software and manual. $549.00
ER1425S-HT Replacement battery for the HiTemp140-FR. $50.00
*NIST NIST Calibration Certificate. $60.00
HiTemp140-FR-TSK (Flush) HiTemp140-FR-TSK (Vented)
Exposure Time in Air
(150 °C/302 °F)
Indefinitely Indefinitely Indefinitely Indefinitely
Exposure Time in Liquid
(150 °C/302 °F)
Exposure Time in Air
(150 °C/302 °F)
Exposure Time in Liquid
(150 °C/302 °F)
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
DOC-1300009-00 REV 2 2015.03.19
Loading...