HiTemp140-FP
High Temperature Data Logger with a
Flexible RTD Probe
The HiTemp140-FP is a durable, user friendly high temperature data logger featuring
a long, flexible RTD probe with a narrow diameter, making it ideal for use in steam
sterilization and lyophilization processes. Commonly used for mapping, validation and
monitoring of high temperature surfaces and environments, this stainless steel data
logger is available in several models. The flexible probe is coated with PFA insulation and
can withstand temperatures up to 260 °C. The body of the logger cannot exceed 140 °C.
The HiTemp140-FP probe design is narrow and lightweight making it ideal for placement
within small vials, tubing, test tube and other small diameter or delicate applications.
Using the latest MadgeTech Software, starting, stopping and downloading the
HiTemp140-FP is simple and easy. To use, simply place the HiTemp140-FP in the
IFC400 or IFC406 docking station (sold separately). Using the software, an immediate
or delay start can be chosen, as well as the reading interval. Start the data logger,
remove it from the docking station and the device is ready to be deployed. Graphical,
tabular and summary data is provided for analysis and data can be viewed in °C, °F, K
or °R. The data can also be automatically exported to Excel® for further calculations.
MadgeTech 4 Software Features
Features
• ±0.1 °C (0.18 °F) Accuracy
• Probe Measurement Range -60 °C to
+260 °C
• Operating Environment -40 °C to
+140 °C
• Submersible (IP68)
• User-Replaceable Battery
• Durable
• Programmable Start and Stop Time
• Four probe lengths: 6, 12, 36 and
72 inches
• Battery life indicator
• Multiple graph overlay
• Statistics
• Digital calibration
• Zoom in/ zoom out
• Lethality equations (F0, PU)
Graph View
• Mean Kinetic Temperature
• Full time zone support
• Data annotation
• Min./Max./Average lines
• Summary view
Cooling Flags
Export to Excel
Tabular Data View
Automation
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Autoclave Verification and Mapping
• Lyophilization
• Monitoring High-Temperature Surfaces
• Container Mapping
• Measurements Inside Small Vials
and Tubing
HiTemp140-FP
SPECIFICATIONS
Specifications are subject to change without notice. Specific warranty remedy limitations apply. Call (603) 456-2011 or go to madgetech.com for details.
TEMPERATURE
Temperature Sensor Flexible RTD Probe
Probe Measurement
Range
Temperature Resolution 0.01 °C (0.02 °F)
Calibrated Accuracy ±0.1 °C ( 0.18 °F)
-60 °C to +260 °C (-76 °F to +500 °F)
GENERAL
Start Modes
Stop Modes Manual or Timed (specific date and time)
Data Logger
Response Time
Trigger Settings
Readings in Trigger
Settings Mode
Real Time Recording May be used with PC to monitor and record data in real time
Password Protection
Memory 65,536 readings
Wrap Around Yes
Reading Rate 4 readings per second up to 1 reading every 24 hours
Soware programmable immediate start
Delay start up to 18 months in advance
In Air In Water
(hours : minutes : seconds : fractions of a second)
t - 00:00:30:00
t - 00:01:10:00
High and Low limits may be set. Once data meets or exceed
sets limits, the device will record to memory. Bi-level start
and stop triggers can also be programmed. Users can specify
the number of readings to take aer the device triggers
21,845 readings
An optional password may be programmed into the device
to restrict access to configuration options. Data may be read
out without the password
t - 00:00:03:50
t - 00:00:06:50
Calibration Digital calibration through soware
Calibration Date Automatically recorded within device
Battery Type
Battery Life 1 year typical (1 minute reading rate at 25 °C/ 77 °F)
Data Format Date and time stamped °C, K, °F or °R
Time Accuracy
Computer Interface
Operating System
Compatibility
Soware
Compatibility
Operating
Environment
IP Rating IP68
Dimensions (Body) 2.95 in x 0.97 in x 0.97 in (75 mm x 24.6 mm x 24.6 mm)
Dimensions (Probe)
Weight 3 oz (85 g)
Material
Approvals CE
BATTERY WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT
CIRCUIT, CHARGE, FORCE OVER DISCHARGE, CRUSH, PENETRATE, OR INCINERATE.
BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 150 °C (302 °F).
3.6V high-temperature lithium battery included; user
replaceable
±1 minute/month at 25 °C
Extended Operation: ±20 minutes/month at 140 °C (±450 ppm)
IFC400 OR IFC406 USB docking station required;
125,000 baud
Windows XP SP3 or later
MadgeTech Standard Soware version 4.2.1.1 or later
MadgeTech Secure Soware version 4.2.0.1 or later
-40 °C to +140 °C (-40 °F to +284 °F)
0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA
HiTemp140-FPST-6: 6 in x 0.13 in (152 mm x 3.2 mm)
HiTemp140-FPST-12: 12 in x 0.13 in (305 mm x 3.2mm)
HiTemp140-FPST-36: 36 in x 0.13 in (914 mm x 3.2mm)
HiTemp140-FPST-72: 72 in x 0.13 in (1829 mm x 3.2 mm)
Body: 316 Stainless Steel
Probe: PFA Insulated Cable with Stainless Steel Tip
Ordering Information
HiTemp140-FPST-6 PN 902330-00 High Temperature Data Logger with a 6 inch Flexible Probe and Stainless Steel Tip
HiTemp140-FPST-12 PN 902312-00 High Temperature Data Logger with a 12 inch Flexible Probe and Stainless Steel Tip
HiTemp140-FPST-36 PN 902313-00 High Temperature Data Logger with a 36 inch Flexible Probe and Stainless Steel Tip
HiTemp140-FPST-72 PN 902316-00 High Temperature Data Logger with a 72 inch Flexible Probe and Stainless Steel Tip
HiTemp140-FPST-6-KR PN 902339-00 High Temperature Data logger with a 6 inch Flexible Probe and Stainless Steel Tip with Key Ring
HiTemp140-FPST-36-KR PN 902336-00 High Temperature Data logger with a 36 inch Flexible Probe and Stainless Steel Tip with Key Ring
IFC400 PN 900319-00 Docking Station with USB Cable
IFC406 PN 900325-00 6 Port, Multiplexer Docking Station with USB Cable
ER14250MR-145 PN 900097-00 Replacement Battery for the HiTemp140-FP
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
6 Warner Road, Warner, NH 03278
(603) 456-2011 | madgetech.com DOC-1296009-00 REV 11 2021.01.05