Madgetech HITEMP140-CF Manual

Page 1
HITEMP140-CF
High Temperature Data Logger
The HiTemp140-CF securely attaches via MadgeTech Canning Fitting to jars, cans or other containers, allowing the probe to measure the heat penetration and temperature distribution of pasteurized and retorted foods..
The HiTemp140-CF can withstand temperatures up to 140 °C (284 °F), making it ideal for use in a variety of canning applications, in a wide range of industries and environments.
For Use With:
Glassware
Retort
Pouches
Plastic Cups
Recart
Packaging
Page 2
Specifications
Specications are subject to change without notice. Specic warranty remedy limitations apply. Call (603) 456-2011 or go to madgetech.com for details.
TEMPERATURE
Temperature Sensor 100 Ω Platinum RTD
Probe Measurement Range
Temperature Resolution
Calibrated Accuracy
-200 °C to +260 °C (-328 °F to +500 °F)
BODY OF LOGGER CANNOT EXCEED 140 °C
0.01 °C (0.02 °F)
±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to
+284 °F)
±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F)
±0.4 °C/±0.72 °F (-40 °C to –20.01 °C/-40 °F to
-4.02 °F)
GENERAL
Start Modes
Stop Modes Manual or Timed (specic date and time)
Real Time Recording
Password Protection
Memory 32,256 readings
Trigger Settings
Reading in Trigger Settings Mode
Wrap Around Yes
Reading Rate
Calibration Digital calibration through software
Calibration Date Automatically recorded within device
Software programmable immediate start Delay start up to 18 months in advance
May be used with PC to monitor and record data
in real time
An optional password may be programmed into
the device to restrict access to conguration
options. Data may be read out without the password.
High and Low limits may be set. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers.
6,451 readings
1 reading every second up to 1 reading every 24 hours
Battery Type
Battery Life 2 years typical (1 minute reading rate)
Data Format Date and time stamped °C, K, °F or °R
Time Accuracy
Computer Interface
Operating System Compatibility
Software Compatibility
Operating Environment
IP Rating IP68
Dimensions (Body)
Dimensions (Probe)
Weight 4.2 oz (120 g)
Material 316 Stainless Steel, Radel
Approvals CE
Battery Warning: re, explosion, and severe burn hazard. Do not short circuit, charge, force
over discharge, crush, penetrate, or incinerate. Battery may leak or explode if heated above
150 °c (302 °f).
3.6V high-temperature lithium battery included;
user replaceable
±1 minute/month at 20 °C to 30 °C (68 °F to 86 °F)
(Stand alone mode)
IFC400 OR IFC406 USB docking station required; 125,000 baud
Windows XP SP3 or later
Standard Software version 2.03.06 or later Secure Software version 4.1.3.0 or later
-40 °C to +140 °C (-40 °F to +284 °F)
0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA
1.55 in x 0.97 in (39.3 mm x 24.6 mm)
HiTemp140-CF-1.1:
1.1 in x 0.095 in (27.94 mm x 2.4mm)
HiTemp140-CF-2.1:
2.1 in x 0.095 in (53.34 mm x 2.4mm)
HiTemp140-CF-3.1:
3.1 in x 0.095 in (78.74 mm x 2.4mm)
HiTemp140-CF-3.9:
3.9 in x 0.095 in (99.05 mm x 2.4mm)
Ordering Information
HiTemp140-CF-1.1 902425-00 HiTemp140-CF with 1.1” probe
HiTemp140-CF-2.1 902426-00 HiTemp140-CF with 2.1” probe
HiTemp140-CF-3.1 902427-00 HiTemp140-CF with 3.1” probe
HiTemp140-CF-3.9 902428-00 HiTemp140-CF with 3.9” probe
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
6 Warner Road, Warner, NH 03278 (603) 456-2011 | madgetech.com
DOC-1450009-00 REV 2 2024.06.13