MadgeTech HiTemp140 Operating Manual

Specification HiTemp140
Probe Range –200 °C to +260 °C (-328 °F to +500 °F)
Temperature Resolution 0.01 °C
±0.1 ºC/±0.18 ºF (20 ºC to +140 ºC/68 ºF to +284 ºF)
Temperature Accuracy
Memory 65,536 readings
Reading Rate 1 reading every second up to 1 reading every 24 hours
Required Interface Package IFC400 or IFC406
Baud Rate 125,000 baud
Typical Battery Life 2 years (1 minute reading rate)
Operating Environment –40 °C to +140 °C, 0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA
Material
Dimensions (Body: HiTemp140-1) 1.6 in x 0.970 in dia. (40 mm x 24.6 mm dia.)
Dimensions (Body: HiTemp140-2,
HiTemp140-5.25, HiTemp140-7)
Weight 4.2 oz (120 g)
Submersible Yes (IP68)
Approvals CE
Model Number Dimensions (Probe)
HiTemp140-1
HiTemp140-2 2.0 in x 0.188 in dia. (51 mm x 4.8 mm dia.)
HiTemp140-2-TD
HiTemp140-5.25 5.25 in x 0.188 in dia. (133 mm x 4.8 mm dia.)
HiTemp140-5.25-TD
HiTemp140-7 7.0 in x 0.188 in dia. (178 mm x 4.8 mm dia.)
±0.3 °C/±0.54 ºF (-20 °C to +19.99 °C/-4 °F to +67.98 °F)
±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to -4.02 °F)
316 Stainless Steel
1.9 in x 0.970 in dia. (48 mm x 24.6 mm dia.)
1.1 in x 0.125 in dia. (0.188 in transitional diameter)
27 mm x 3.2 mm dia. (4.8 mm transitional diameter)
2.0 in x 0.125 in dia. (0.188 in transitional diameter)
51 mm x 3.2 mm dia. (4.8 mm transitional diameter)
5.25 in x 0.125 in dia. (0.188 in transitional diameter)
133 mm x 3.2 mm dia. (4.8 mm transitional diameter)
Product User Guide
HiTemp140 Series
IFC400 and IFC406 (sold separately)
HiTemp140 Series
High Temperature Data Loggers
Probe lengths up to 7 inches available
MadgeTech, Inc.
6 Warner Road
Phone (603) 456-2011
madgetech.com
l
Warner, NH 03278
l
info@madgetech.com
l
Fax (603) 456-2012
DOC-1211035-00 REV 13 2019.08.05
To view the full MadgeTech product line,
visit our website at madgetech.com.
HiTemp140
Product Notes
The HiTemp140 is a rugged, high precision, temperature data logger that is built for use with
autoclaves and harsh environments. This stainless steel device is submersible, can withstand
temperatures up to 140 °C (284 °F) and has an accuracy of +/-0.1 °C (0.18 °F).
Product User Guide
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to
Manual Start”, depending upon your start method.
- Disconnect the data logger from the docking station and place it in the environment to
measure.
The HiTemp140 can store up to 32,700 readings, and features a rigid external probe capable
of measuring extended temperatures, up to 260 °C (500 °F). Custom probe lengths up to
7 inches are available. The device records date and time stamped readings, and has non-
volatile solid state memory that will retain data even if the battery becomes discharged. The
HiTemp140 can be used in both wet and dry applications up to 140 °C indefinitely.
Submergibility
The HiTemp140 is fully submersible and is rated IP68. It can be placed in environments with
up to 230 feet (70 m) of water.
O-Rings
O-ring maintenance is a key factor when properly caring for the HiTemp140. The O-rings
ensure a tight seal and prevent liquid from entering the inside of the device. Please refer to
the application note “O-Rings 101: Protecting Your Data”, found on the MadgeTech website,
for information on how to prevent O-ring failure.
Note: This product is rated for use up to 140 °C. Please heed the battery warning. The product will explode if exposed to temperatures above 140 °C.
Installation Guide
Installing the Interface cable
- IFC400 or IFC406
Refer to the “Quick Start Guide” included in the package.
Installing the software
The Software can be downloaded from the MadgeTech website at madgetech.com. Follow
the instructions provided in the Installation Wizard to install the MadgeTech Software.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the docking station.
- Connect the USB end of the interface cable into an open USB port on the computer.
- Place the data logger into the docking station.
- The data logger will automatically appear under Connected Devices within the software.
- For most applications, select “Custom Start” from the menu bar and choose the desired
start method, reading rate and other parameters appropriate for the data logging application
and click “Start”. (“Quick Start” applies the most recent custom start options, “Batch Start”
is used for managing multiple loggers at once, “Real Time Start” stores the dataset as it records
while connected to the logger.)
Note: The device will stop recording data when the end of memory is reached or the device is stopped, unless user selectable memory wrap is enabled. At this point the device cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Place the logger into the docking station.
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will
be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement Warning
When replacing the battery, it is important to insert the battery with positive polarity pointing
upward towards the probe. Failure to do so could result in product inoperability or potential
explosion if exposed to high temperatures.
Recalibration
Recalibration is recommended annually. To send devices back for calibration, visit
madgetech.com.
Call for custom calibration options to accommodate specific application needs. Prices and specifications subject to change. See MadgeTech’s terms and conditions at madgetech.com. To send devices to MadgeTech for calibration, service or repair, please use the MadgeTech RMA Process by visiting madgetech.com.
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 140 °C (284 °F).
Specifications subject to change.
See MadgeTech’s terms and conditions at madgetech.com
Loading...