MadgeTech Cryo-Temp Specifications

CRYOTEMP
ULTRA LOW TEMPERATURE DATA LOGGER
Features
• Ultra-Low Operating Temperature (-86 °C)
• Direct Contact with Dry Ice
• No External Probes or Wiring
• High Accuracy
• Programmable Alarm and Warn Ranges with Alarm Delay
• LED Status Indicators for Instant Alarm Notification
• Magnetic Start
• Optional Password Protection
• Splash Resistant (IP64)
• CE Compliant
Benefits
• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment
Applications
• Dry Ice Shipments
• Plasma and Red Cell Freezers
• Blood Banks/Frozen Tissue Storage
• Monitoring Frozen Vaccines
• Cold Chain Monitoring
• Medical and Pharmaceutical Shipping
• Ultra-Low Temperature Storage
MadgeTech’s popular ultra-low temperature data re­corder, records temperatures as low as –86 °C. The Cryo-Temp can be used in a broad line of applica­tions such as monitoring blood plasma, vaccines, pharmaceuticals, frozen foods and shipping con­tainers. This standalone device does not require any additional probes. The enclosure is designed with a built in handle for easy attachment and is IP64 splash resistant. It utilizes a USB docking station (sold separately), to communicate with a desktop PC or laptop.
The Cryo-Temp provides date and time stamped temperature readings and uses the MadgeTech Data
Logging Software. High and low alarm thresholds can be configured in the MadgeTech software to warn users if safe limits are exceeded. The Cryo-Temp is equipped with three LED lights to visually indicate when the logger is record­ing or if temperature thresholds are in danger of being exceeded or have been breached.
The Cryo-Temp is the ideal temperature monitoring solution for any application involving ultra-low temperature recording.
MADGETECH DATA LOGGER SOFTWARE
A
Key
A Graph View
B Tabular Data View
C Statistics
D Digital Calibration
E Copy to Excel®
B
C
E
D
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Software Features:
Multiple graph overlay
Statistics
Digital calibration
Zoom in/ zoom out
Lethality equations (F0, PU)
Mean Kinetic Temperature
Full time zone support
Data annotation
Min./Max./Average lines
Data table view
Automatic report generation
Summary view
Multilingual
CRYO-TEMP SPECIFICATIONS*
*SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE. SPECIFIC WARRANTY REMEDY LIMITATIONS
APPLY. CALL 1-603-456-2011 OR GO TO WWW.MADGETECH.COM FOR DETAILS.
Temperature Sensor: NTC Thermistor
-86 °C to +35 °C (-122.80 °F to +95 °F)
Temperature Range:
recommended, may be used up to +55 °C (131 °F)
Temperature Resolution: 0.1 °C (0.18 °F)
Calibrated Accuracy: ±1.0 °C
1 reading every 5 seconds up to 1
Reading Rate:
reading every 30 minutes, selectable in software
Memory:
32,767 readings; software configurable memory wrap
Software programmable immediate
Start Modes:
start, delay start up to 30 days in advance or manual start with magnetic switch.
Software programmable high and low
Temperature Alarm:
alarm limits, high and low warn limits each set point may be individually enabled or disabled
Software programmable time delay for warn/alarm limits. Logger will not
Alarm Delay:
indicate out of range status until the temperature has been out of range for a set period of time.
• Green: Blinks to indicate safe conditions during logging
Status Indicators:
• Yellow: Blinks to indicate warn limits have been exceeded
• Red: Blinks to indicate alarm criteria (limits/delay) have been exceeded
BATTERY WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT
CIRCUIT, CHARGE, FORCE OVER DISCHARGE, CRUSH, PENETRATE, OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 100 ºC (212 ºF).
An optional password may be programmed
Password Protection:
into the device to restrict access to configuration options. Data may be read out without the password.
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6V lithium battery (LTC-7PN)
Battery Life:
2 years typical @ 25 °C, 1 min. reading intervals 1 year typical @ -80 °C, 1 min. reading intervals
Data Format: Date and time stamped ºC, ºF, K, ºR
Time Accuracy:
Computer Interface:
Software:
±2 minutes/month max. at 20 °C to 30 °C ±5 minutes/month max at -80 °C
USB docking station to PC (IFC300) 38,400 baud
XP SP3/Vista/Windows 7/Windows 8 (MadgeTech 4 Only)
-86 °C to +35 °C (-122.80 °F to +95 °F)
Operating Environment:
may be used up to +55 °C (131 °F); 0 %RH to 100 %RH
Dimensions:
Ingress Protection/IP
Rating:
3.3 in x 2.1 in x 0.4 in (83 mm x 54 mm x 11 mm)
IP64 (splash resistant, not submersible)
Enclosure: ABS plastic, white, polycarbonate labeling
Weight: 1.4 oz (40 g)
Compliance:
CE, Approved for carriage on Delta® mainline flights
The Cryo-Temp exchange program provides the
Product Maintenance:
user with a new replacement device at a reduced cost when battery is depleted or device is in need of factory calibration. Call for details.
ORDERING INFORMATION
MODEL DESCRIPTION
Cryo-Temp Ultra Low Temperature Recorder
IFC300 Software, manual, docking station and USB connection cable
SVP–Secure Protocols for validating the 21 CFR Part 11 ready, secure logging system
Calibration Certificate Calibration Certificate available for data logger
Exchange Program
For Quantity Discounts call 603-456-2011 or email sales@madgetech.com
When product battery is depleted or device requires factory calibration, the product can be exchanged for a new Cryo-Temp Data Logger at a reduced cost. Call for details.
6 Warner Road · Warner, NH 03278
Phone: (603) 456-2011 · Fax: (603) 456-2012
Temperature
ASK ABOUT OUR OTHER
DATA
LOGGERS
DOC-1001009-00 REV 21 03.07.2017
Humidity
Pressure
pH
Level
Shock
LCD Display
Pulse/Event/State
Current Voltage
Wireless
Intrinsically Safe
Spectral Vibration
Motion
Loading...