Macurco GS-1 User Manual

Gasman
Personal Single Gas Monitor
User Manual
M07630
Sep 2015 Issue 11
Instructions specific for use in hazardous areas
Baseefa04ATEX0383 Flammable Gas
IECEx BAS 05.0038 Flammable Gas
Baseefa04ATEX0384 Oxygen or Toxic Gas
IECEx BAS 05.0039 Oxygen or Toxic Gas
The following information covers all relevant points listed in clause 1.0.6 of the EHSR’s of the ATEX directive.
The certification marking is as follows:
1. The equipment may be used in Zones 1 and 2 for flammable versions and Zones 0, 1, and 2 for toxic and oxygen versions, for Group IIA, IIB, and IIC gases and vapours for Temperature Classes T1, T2, T3 and T4
2. The equipment is certified for use in ambient temperatures in the range –20°C
to +65°C (-4 to +149 F). The equipment should not be used outside these ranges.
3. Compliance with the Essential Health and Safety Requirements has been assured by compliance with EN60079-0:2004 + Amds 1 & 2, EN60079-1:2004 and EN60079­11:2006, as certified by Baseefa.
Compliance with gas detection performance
standards EN50054, EN50057, EN61779-1, EN50104 and EN50270 has been certified by Lloyd’s Register
4. Repair of this equipment and gas sensor replacement shall be carried out by the manufacturer or in accordance with the procedures in the manual and by trained personnel.
5. If the equipment is likely to come into contact with aggressive substances, then it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected, which may compromise the certification level.
6. The rechargeable battery must only be charged in non-hazardous (safe) areas by connection to the specified Crowcon charger.
7. Only the following cell types may be fitted in the battery compartment of the non-rechargeable battery pack: CR2 of the following brands only: Panasonic, GP, Energiser and Maxell. Do not fit a Duracell battery.
The cells must only be changed in a non-
hazardous (safe) area.
8. The equipment is not certified for use in atmospheres containing more than 21% oxygen.
From the 1st November 2010, EN60079­29 part 1 has been harmonised under the ATEX directive 94/9/EC. Therefore to comply with the ATEX directive, portable apparatus sensing flammable gases should have a functional check with gas before each day of use. Other testing regimes may be employed depending on local circumstances.
Area Classifications: -
Zone 0: An area classified as Zone 0 is an
area in which an explosive mixture of gas is continuously present or present for long periods.
Zone 1: An area classified as Zone 1 is likely
to have ignitable concentrations of flammable gases, vapours or liquids present under normal operating conditions.
Zone 2: An area classified as Zone 2 is not
likely to have ignitable concentra­tions of flammable gases, vapours or liquids present under normal operating conditions.
© Copyright Crowcon Detection Instruments Ltd 2015. All rights are reserved. No part of the docu­ment may be photocopied, reproduced, or translated to another language without the prior written consent of Crowcon Detection Instruments Ltd.
Safety information:
 • Read and understand all instructions in
the operation section of this manual before use.
 • Donotsubstitutecomponents asthismay
impair intrinsic safety and invalidate war­ranty.
 • Observe all warnings and instructions
marked on the unit and within this manual.
 • Observesitehealthandsafetyprocedures
for gases being monitored and evacua­tion procedures.
 • Understandthe screendisplayand alarm
warnings prior to use.
 • If this product is not working properly,
read the troubleshooting guide or contact your local Crowcon Office or agent..
 • Ensurequalifiedservicepersonnelchange
sensors and operating system.
 • Ensuremaintenance,service andcalibra-
tion are carried out in accordance with the procedures in the manual and by trained personnel.
Gasman
Personal Single Gas Monitor
Contents
Unpacking ...................................1
I. Introduction .................................7
II. Operation ..................................9
III. Batteries .................................14
IV. Alarm indications ...........................16
V. Fixing Accessories ...........................18
VI. Flow sampling .............................19
VII. Maintenance and Calibration .................23
VIII. PC interface and software ...................26
IX. i-module replacement .......................28
X. Specification ...............................30
XI. Accessories and spare parts ...................31
XII. Troubleshooting guide ......................34
Appendix: Limitations of sensors ..................35
Gasman Unpacking
Gasman
Personal Single Gas Monitor
Thank you for purchasing the new Gasman Personal Gas Monitor. Gasman has redefined portable gas monitoring and will give you years of unparalleled service and reliability.
Please read the instructions carefully before use. Keep the manual for future reference.
Unpacking
Remove the Gasman Personal Gas Monitor from the packaging. The Gasman accessories will be located in the bottom of the box. Check the contents are complete, you should have:
• Gasman unit with alligator pocket clip;
• A configuration report detailing the sensor installed, alarm
settings and a calibration certificate;
• Calibration flow cap and tubing;
• Usermanual.
Optional battery chargers and other accessories will be packed in a separate box.
Battery check
The Gasman Personal Gas Monitor has two battery options: Li-ion rechargeable or non-rechargeable batteries. The label on the back of the detector will have either NR (non-rechargeable) or R (rechargeable) to indicate which battery option is within the detector.
Rechargeable units
Gasman uses a Li-ion battery pack and should arrive with sufficient charge so that the unit can be used straight out the box. However, if this is the first time you have used the Gasman unit, you may need to charge the battery to attain the full operating time. (The actual operating time will depend on the types of sensor installed.) The flammable Gasman will operate for at least 12 hours on a fully charged battery.
1
Unpacking Gasman
Warning: rechargeable units
Only charge the rechargeable Gasman Personal Monitor with a Crowcon Gasman charger. Failure to comply could invalidate safety certification, warranty and safety; and may result in permanent damage to the unit.
Non-rechargeable units.
Gasman uses a CR2 battery which can be easily field replaced. Ensure only the correct types of battery are fitted to maintain compliance with certification (see Battery check on page 1).
Gasman IR CO2 Personal Single Gas Monitor
The Gasman IR CO for CO
detection. This version of Gasman is not designed or certified
2
for use in a hazardous area, and the safety certification data in Section
personal single gas monitor an infrared gas sensor
2
X is not applicable. Operation and maintenance of the Gasman IR CO is essentially the same as for other Gasman units but with the following points to note.
2
CO2 is present in ambient air at a background concentration of just under 400ppm (0.04%). In any enclosed environment the ambient CO and if the room is poorly ventilated concentrations well in excess of 1000ppm (0.1%) can
levels in the outdoor air will also be enhanced by any emitted CO2 e.g. from vehicle
occur. CO
2
exhausts or boiler flues venting to atmosphere so it is normal to see fluctuating levels of CO the display whether indoors or outside.
concentration will increase as a result of respiration
2
2
2
on
Gasman Quickstart guide
Quickstart guide
1. Getting started
Gasman familiarisation
Gas
Gas Sensor
name label
Alarm LEDs
Sounder outlet
Operator Button
Operator LCD Display Screen
Switching on
Gasman requires little setting up. Follow these simple steps to get your unit ready for use.
1. Ensure the unit is in clean air.
2. Switch on. Press and hold the operator button for approximately 3 seconds until the red LEDs flash.
The operator display screen will light up and
the unit will begin a warm up sequence.
The user should now check the detector
goes through the warm up sequence detailed below. In the case of any deviations the Gasman should be returned to your local Crowcon office or Service agent.
2
HS
3
Quickstart guide Gasman
Gasman warm up sequence a) The Gasman will demonstrate
the alarm LEDs, sounder, vibration alerts, and the
Sounder Alarm test
HS
2
LEDs Alarm test
operator display screen. The sounder may be silenced by pressing the button.
Vibrator Alarm test
The Gasman screen will show the following display during the
warm up sequence. This will take approximately 20 seconds.
0.0
OK
ppb ppm
vv/v
ppm
Note: these screens may differ depending on what sensor is being used.
b) Autozero If Autozero is enabled (default), the unit will display the Autozero
confirm screen: the display will alternate between ‘ZERO’ and ‘????’. Press the operator button with a single click to confirm Zero. If the operator button is not pressed within the 10 second time out, Gasman will proceed directly to Run mode without performing Zero.
Autozero confirm
ppm
OK
No Autozero
OK
ppm
Screen Icons
OK
AL -1 AL -2
Warm up
Flashing icon ­Gasman running normally
Alarms
Battery life
OK
Autozero
0.0
4
Gasman Quickstart guide
Confidence signals
In normal Run mode, Gasman will emit a short beep accompanied by a blue LED flash every 10 seconds and the OK icon flashes to show operational health. These confidence signals can be disabled using the PC software.
2. In the event of an alarm
Alarm signals
In the event of gas concentrations exceeding the alarm thresholds for the gas being monitored, Gasman will activate the alarm signals.
Alarm signals
The red and blue alarm LEDs
HS
2
ppm
1. When the gas level returns to normal, press the operator button. This will reset your Gasman unit to normal Run mode. If the gas level is still within alarm limits, the button will have no effect.
The Gasman alarm is set to latch by default. The unit will still continue in alarm mode even when the gas level returns to normal, until the button is pressed and the alarm is cleared.
will flash, the sounder will emit a loud, fast series of beeps, the internal vibrator alarm will activate. The operator screen will display the alarm level and gas reading alternately. See the figure to the left.
AL - 1 — Alarm level one AL - 2 — Alarm level two
3. Switch off unit and storage
Switching off
1. Press and hold the button for 5 seconds until the display shows
OFF. The display will count down to switch off.
5
Quickstart guide Gasman
Storing conditions
In order to optimise sensor performance and lifetime, your Gasman unit should be stored in a safe, non-hazardous area, at 0-20°C and 20-90% RH.
4. Additional information
For battery recharging information go to section III.
For fixing accessories go to section V.
For sampling section go to section VI.
For calibration information go to section VII.
For troubleshooting guide go to section XII.
6
Gasman Introduction
I. Introduction
Thank you for purchasing the new Gasman Personal Single Monitor. Gasman is a portable Single Gas detector, designed to be carried or worn by individuals working in hazardous environments such as confined spaces. It is suitable for use in classified hazardous areas. Gasman monitors a single gas and displays its reading on a display screen. Alarm warnings are given through a combination of a loud audible alarm, a bright visual alarm of blue/red flashing LEDs and an internal vibrator. Gasman can be fitted with a wide range of modular, or plug and play gas sensors. The sensor carries an intelligent processor which contains calibration and sensor information.
Gasman is battery operated and is available with rechargeable batteries or dry cell battery options. The dry cell battery option is only available for toxics and oxygen Gasman units. A battery charger for single and multiple Gasman units is available for rechargeable units, see section XI for more information.
At Crowcon we recognised the need for a reliable and robust personal monitoring system, which is both small, light and easy to use. Gasman has a single operator button, and an intelligent user-friendly display with automatic backlight. The gas level is continuously monitored providing normal gas readings, peak readings and time weighted averages (TWA). Gasman is available as a diffusion sampling instrument, see section XI for sampling accessories. Configuration and data/event logging is handled by Crowcon Portables PC software, the PC communication link is provided as a link on the charger interface unit.
Gasman’s compact shape and design makes it comfortable to wear, with a non-slip grip for better handling. Extra accessories such as pocket clip, hard hat clip, shoulder strap and chest harness, can be purchased.
Gasman has been designed from top to bottom to bring you a lighter, compact design with single button operation for ease of use, maintenance and extreme reliability.
7
Introduction Gasman
i-module gas sensor
Gasman uses unique plug and play i-module sensor technology. Each sensor unit incorporates its own intelligent processor holding sensor configuration and calibration data. Different sensors can be purchased, and once inserted, are immediately ready to run. Flammable units are only available with rechargeable batteries. Plug and play will ease maintenance time and cost, and the intelligent modular system will remove the need to calibrate the sensor. Gasman can be reconfigured by purchasing additional pre-calibrated i-modules from Crowcon or your local office. After fitting a new i-module, it is recommended that a gas test is carried out on the detector before use.
Reliable, anti-shock mechanics and robust housing
The Gasman housing is built from resilient material, giving it strength and flexibility to withstand the hardest of working conditions, water and dust tight to IP65 with a non-slip grip. If the unit is dropped, there will be no disruption of power or function, ensuring reliability and service for years to come.
Software
The internal software in Gasman has been designed and written in accordance with the requirement of IEC 61508 to ensure quality and integrity of operation. Gasman has been designed to give a truly reliable personal gas monitoring system. The internal circuitry includes an independent watchdog. The software monitors for any malfunction within the unit and will display an error warning to the user should they occur.
8
Gasman Operation
II. Operation
2.1 Switch-on sequence
1. Ensure the unit is in clean air.
NB. For CO2 detectors see zeroing, Section VII.
2. Switch on Press and hold the operator button for approximately 3 seconds until the red LEDs flash.
The instrument begins with testing all the LCD segments on the
operator display screen, the red and blue alarm LEDs, sounder and internal vibrator alert. The sounder may be silenced by pressing the button. The unit enters a warm up mode and displays a sequence of screens, see page 3 for more details. At the end of warm up, the Autozero menu will be displayed.
The Autozero function can be disabled or set to run automatically, without user confirmation: Autozero will not appear.
See section VIII PC Interface and software.
Calibration check
During the warm up sequence, if the next calibration date is due in less than 31 days, ‘CAL - nn’ will be displayed (where nn is the number of days to the next calibration). If calibration is already overdue, the Gasman unit will display a warning message ‘CAL’. The instrument can still function, but it is strongly recommended the unit is sent for calibration as soon as possible.
Press the button to continue to Run mode. The Gasman can be set, using Portables PC software, to lock out in the event of the
calibration due date expiration in order to prevent further operation of the detector until a calibration is preformed.
3. Autozero Press the operator button with a single click to confirm Autozero. Flammable and toxic units will be set to read zero and oxygen units set to read 20.9%. If the operator button is not pressed within 10 seconds Gasman will proceed directly to normal operation without performing a zero.
Switch off
To switch off the unit, press and hold the operator button for 5 seconds. The display will count down from 5 until the unit counts down to shut off.
Battery check
Usethistimetocheckthereis
!
sufficient charge in the battery pack
NB. If Autozero fails, a warning message will be displayed.
9
Operation Gasman
2.2 Run mode
The Gasman unit displays the gas reading on the operator display screen. A typical display is shown below:
OK
ppm
The sensor channel will display the current value of the gas being monitored and the unit of measurement, the ‘OK’ symbol will flash indicating unit is working correctly. Familiarise yourself with the gas being monitored in your unit. Ensure you understand site health and safety procedures. For information on Peak and TWA readings, go to section 2.4.
Confidence signals
To reassure users the unit is working correctly, the Gasman unit will emit a short confidence beep accompanied by a blue LED flash every 10 seconds and the OK icon will flash continuously.
Screen Icons
Warm up
Flashing
OK
icon ­Gasman running normally
AL -1
Alarms
AL -2
Battery life
Autozero
0.0
2.3 Display symbol guide
Battery
Full
TWA alarm
Gasman will display the TWA alarm for toxic gases when the 15 minute or 8 hour Time Weighted Average alarm threshold is passed. Gasman will display ‘LTWA’ and ‘STWA’. When a TWA alarm threshold is reached the TWA alarm cannot be cleared.
A full battery is represented by a battery icon showing full three bars. A low battery charge will show 1 bar. When zero bars are shown the battery icon flashes and the
!
sounder will emit warning beeps. If the battery becomes too low for operation the Gasman will switch off. A low battery warning will alarm prior to switch off.
10
Gasman Operation
2.4 Display options
Gasman provides four additional selectable displays:
Peak display When Peak mode is selected the instrument shows the highest value for flammable and toxic gases, or the lowest value for oxygen since switch on or last Peak Reset. This is useful for vertical entry checks where the Gasman can be lowered down the shaft rather than using a sampling tube and for viewing peak exposure at the end of a shift.
TWA display Shows the 15 minute and 8 hour TWA, for toxic gases, monitored since last turn on.
Peak Reset Before a Peak entry test is performed select this menu option to clear any previously stored Peak value.
Zero Performs Zero on your Gasman unit. (Please ensure you are
0.0
in clean air).
How to display the menu
1. To view the additional display option menu, double-click the button.
The menu icons will appear on the screen, as shown below.
2
HS
0.0
OK
ppm
0.0
Note: Only toxic gas
OK
monitoring units will
ppm
display the TWA menu option.
2. Press the operator button with a single click to scroll across the list. When the underscore appears under your choice, double-click the operator button. This will select the option.
If Peak or TWA is selected, Gasman will display the icon on the
operator screen.
11
Operation Gasman
Peak test
When performing a Peak test, such as a vertical entry check, previous readings can be cleared by selecting the Peak Reset menu option
.
1. Scroll menus 2. Select
0.0
OK
ppm
Peak display
Zero
To perform Autozero, select the Zero function from the menu. When Zero is completed, the instrument will return to normal Run mode.
3. Run mode
The Gasman is now ready to use. Familiarise yourself with the gas being monitored in your unit and
make sure you understand site health and safety procedures in the event of alarm conditions.
OK
ppm
12
Gasman Operation
2.5 Logging
Gasman incorporates both data and event logging which can be accessed using the RS232 communications link available with the Single Way Charger Interface (part No.C01940), using Crowcon Portables PC software. See section VIII.
Data is recorded every minute (this rate can be adjusted using the PC software). The log can record 900 hours of data at 1-minute intervals.
Gasman also records the time and date for a number of operating and diagnostic events including:
• Switch on and switch off;
• Level 1, Level 2 and TWA, alarm on, alarm off and the peak level
during the alarm;
• Zero, calibration and gas test with success or failure;
• Flammable sensor saver on and off;
• The battery condition is logged every switch on and switch off
while the instrument is operating, and certain configuration changes are also logged;
• The Event log can record over 4800 events.
13
Batteries Gasman
III. Batteries
3.1 Rechargeable batteries
Recharge time for the Li-ion batteries is less then 6 hours (less, if they are not fully discharged). Rechargeable batteries will typically last 12+ hours for flammable units.
3.2 Gasman Charging Unit
There are 3 models of charging units available for Gasman; a single way drop in charger unit, a single way charger unit with integral PC Interface, and a multiway charger unit. The multiway charger unit can accommodate up to 5 Gasman units. Note: the multiway charger unit does not have a PC interface option.
The charger units are powered by a nominal 12 V dc input. Power
SuppliesareavailablewithUK,EURorUSpins,alternatelyauniversal
90-260 V power supply is available for any other configuration. The multiway charger includes a universal power supply. A vehicle lighter socket lead can also be supplied, please see section XI Parts and Accessories.
Communications and charging are only permitted via Crowcon charger/ charger-interface C01940.
Charging the batteries
1. Ensure you are in a safe area.
2. Plug the charger power supply into a mains socket, and connect to the charger lead.
3. To charge your Gasman, simply drop the Gasman into the charger in an upright position with the screen facing outwards, see diagram below:
14
12Vdc
Power supply
Gasman Batteries
The Gasman should normally be switched off for charging. During
HS
2
Power supply
charging the red LED on the front of the charger unit will be lit. When charging is complete the
12Vdc
red LED is extinguished.
During charging, the display will show the battery icon sweeping
from empty to full. On removing Gasman from the charger unit, if the Gasman is on, the display icon will update in 20 seconds to show actual charge state.
Chargers with dual LEDs are no longer suitable for use with the
current Gasman. We recommend that such chargers be replaced with the newer model shown above (see Section XI, Accessories and spare parts, p32).
The charging time will be longer if the unit is switched on during charging.
When Gasman is fully charged and switched on the battery icon
will display three full bars.
3.3 Changing rechargeable batteries
It is recommended that rechargeable batteries should be changed by an authorised Crowcon service centre only.
3.4 Non-rechargeable batteries
Gasman uses a Lithium Cell battery pack which will give up to two years of operation.
Always switch the Gasman off before opening the case to change the battery.
To replace the battery, ensure you are in a safe, non-hazardous area. Remove the back cover and remove the battery*. Insert the replacement battery into the instrument and securely fasten the back cover.
Note: When a non-rechargeable detector is placed in the charger/charger interface the red LED will not illuminate and the detector will not charge.
* Note: later versions of the Gasman will have a tab in place to aid removal of the battery.
15
Alarm indications Gasman
IV. Alarm indications
Gasman provides two instantaneous alarm levels, designated level 1 and level 2. For toxic gas sensors, there are also two TWA alarms, one for short term exposure (STEL): based on a 15 minute time weighted average, and the second TWA alarm is for long term exposure (LTEL): based on a 8 hour time weighted average.
Alarm configurations are set via the Crowcon Portables PC software. The following settings can be made:
Alarm thresholds for each sensor: Level 1 and level 2 instantaneous alarms.
Alarm trigger: This can be set to rising levels of gas concentration, or as falling. Oxygen units are set to falling for deficiency monitoring.
Alarm latching: Alarms can be set to be latched or unlatched. Latched alarms will require the button to be pressed in order to clear the alarm
oncethehazardhaspassed.Thisisthedefaultsetting.Unlatchedalarms
will clear automatically when the gas hazard has passed, but not until then.
Alarm mute: The sounder can be set to mute for level 1 alarm only; pressing the operator button during an alarm condition i.e. presence of hazardous gas, will silence the sounder and stop the vibration alarm. The alarm LEDs will continue to flash.
Alarm sounder tone: Different distinctive tones can be selected for the each alarm condition to achieve the best performance for the monitoring conditions available.
In the event of a Time Weighted Alarm (TWA)
In the event of the 15 minute or the 8 hour TWA being triggered, Gasman will go into alarm and display a TWA warning with the toxic gas readings. Neither the 15 minute nor the 8 hour TWA alarm can be cleared.
16
Gasman Alarm indications
In the event of a Flammable overange alarm
If flammable gas concentrations exceed 100% LEL Gasman locks into alarm and displays ‘9999’ showing an overange condition. Gasman will temporarily cut off power to the sensor to prevent burn out and display a progress bar for 200 seconds. When the timeout is complete, either press the button to continue or switch your unit off and on to restart. This option is programmable with Crowcon Portables PC software and the default is switched on.
If the Gasman goes over range at any time it is recommended to carry out a gas test before using it again.
17
Fixing Accessories Gasman
V. Fixing Accessories
Clip Accessories
Gasman is supplied with an alligator clip. Other clip options are:
Hard hat clip
Allows Gasman to be clipped to a hard hat, a method of wearing a portable gas detector preferred by many users.
Pocket Clip
Allows Gasman to be clipped to a pocket, near the breathing zone of the user.
Universal harness plate
Crowcon provide a universal harness plate which can be used with either a chest harness or a shoulder strap.
How to wear your Gasman unit
Chest harness
Use the M3 fixing (usable once the
Shoulder/neck connectors
clip has been removed) on the back of your Gasman unit to attach the
M3 nut
chest harness plate. Create a chest harness by attaching one strap to the top connectors, to go around the neck, and the other to link around the waist
Waist connectors
using the side connectors. Adjust the strap lengths until the Gasman unit is in a comfortable working position.
Chest harness plate
Shoulder strap
With the universal harness plate in place on the belt clip, attach the shoulder strap accessory onto the top connectors. Adjust to a comfortable working position.
See Accessories, section XI, for full list.
18
Gasman Flow sampling
VI. Flow sampling
Attaching the flow cap
To perform manual sampling using Gasman, a flow cap must be fitted onto the front of the unit, over the sensor. A flow cap is included with your Gasman unit.
1. To fit the flow cap, firstly ensure the cap gasket is intact, slip the flow cap over the sensor on the front of your Gasman unit, until the cap is tightly fitted into place.
2. Attach the sampling tube or flow accessory onto a gas nozzle.
3. Attach the aspirator bulb onto the remaining gas nozzle.
4. To remove the flow cap, gently tease the cap from the sensor and
lift the cap away from the instrument.
When using the manual aspirator kit, adopt a consistent rhythm whilst using the hand aspirator. Crowcon recommend squeezing once per second to achieve a flow rate of approximately 0.5 - 1 litre/min. At least 10 pumps per sample are recommended.
The sampling tube supplied is normally a 2m (6ft) length. Longer lengths of sampling tube can be provided, but will increase the time taken to get a sample from the point of sampling to the Gasman instrument. When using an extended length of tubing a response time test is recommended. Gas of known concentration should be sampled along the full length of tubing to be used and the time taken for the sensor reading to reach the known gas levels should be noted. This time should be used as the minimum for sampling before readings should be taken.
19
Gasman gas test accessory kit Gasman
Gasman gas test accessory kit
The Gas Test Accessory is a gas testing kit designed to enable gas testing and one button calibration of your Gasman gas detector using a specially formulated, high stability long life single gas mix. It can be used with Gasman units having sensors for Flammable, Oxygen, Carbon Monoxide and Hydrogen Sulphide gases.
6.1 Gas testing
Gas testing checks the sensor is responding within set limits to an applied gas of known composition. This can be performed as often as desired, but would usually be carried out each time the Gasman is issued for use. The Gasman itself will determine Pass/Fail status for the gas test.
In order to perform successful gas tests you should ensure:
• The gas used has the correct gas concentration, and that it is
within the validity date specified by the supplier.
• The gas flow path is leak tight. It is important to check that the flow
cap is properly fitted to the Gasman unit, and the outlet tubing is not restricted in any way, nor additional tubing length used.
The Gas Test Accessory Kit comprises a gas cylinder containing the gas, a 'Trigger' regulator with interconnect tubing, a magnet - used to activate Test mode, a flow cap to attach to the Gasman, and a vent line. The kit is supplied in a convenient carry case. The Trigger regulator can be operated in two ways: (1) squeeze and hold - allows gas flow as long as the lever is held in, or (2) by lifting the lever - the flow is locked on.
6.2 How to perform a gas test
1. Ensure your Gasman unit is switched on and in normal operation.
2. Fit the flow cap onto the front of the sensor and attach the hose
from the Trigger regulator. Attach the outlet hose to ‘vent gas away’ - do not extend this hose and do not restrict or allow kinks.
3. Touch the magnet, horizontally aligned onto the case, to the left hand side of the display (as shown left). Your Gasman unit will activate the Gas Test and show ‘TEST’ on the display.
20
HS
2
OK
ppm
Gasman Gasman gas test accessory kit
4. Gasman will display a progress bar. Apply the gas whilst the progress bar is counting down. The display
OK
ppm
will alternate between the screens shown right.
Gasman will then display ‘PASS’ or ‘FAIL’ In the event Gasman displays ‘FAIL’, please see the
OK
ppm
troubleshooting guide in the first instance or contact Crowcon or your local service centre.
5. To abort the gas test press the button at any time whilst the test is in progress.
6.3 How to perform a one button calibration test
To perform a one button calibration test, you must first perform a Zero on your Gasman unit.
1. Ensure you are in clean air.
Double click the button and select Zero from the Options menu. Gasman will perform a Zero.
To perform a one button Calibration, complete the next steps within 15 minutes of completing the Zero.
2. Follow steps 1 to 3 given in 6.2, Gasman will display ‘CAL’ and ‘????’ alternately. Press the button to confirm one button Calibration.
If the button confirmation for calibration is not made within 10
seconds then the process will revert to test as in 6.2.
3. Apply calibration gas following step 4 in 6.2.
4. To abort the Calibration test press the button at any time whilst
the test is in progress.
Your Gasman unit will adjust the value for the gas channel to match the stored calibration gas value within the sensor i-module. This will also change the calibration due date which is set as a default of 182 days.
If your Gasman unit does not calibrate successfully, Gasman will display ‘FAIL’. Your Gasman unit must be sent to Crowcon or a local service centre for re-calibration.
Any Gas Test ‘pass’ and ‘fail’, and calibration ‘pass’ and ‘fail’ and their values are stored in the Event log.
21
Gasman gas test accessory kit Gasman
6.4 Gas test/calibration troubleshooting
Symptom Possible Cause Action
No response to gas Gas cylinder empty Check gauge, replace
Hose blocked or kinked Ensure no restriction to
Gasman fails gas test
Gasman fails calibration
Gasman passes gas test but will not enter calibration mode
Gas cylinder empty Check gauge, replace
Gas cylinder out of date Check date and replace as
Hose blocked or kinked Ensure no restriction to
Calibration drifted Calibrate Gasman
Gas flow not started immediately
Gas cylinder empty Check gauge replace
Gas cylinder out of date Check date and replace as
Hose blocked or kinked Ensure no restriction to
Calibration drifted Calibrate Gasman
Stabilisation time too short Reset using PC software
Menu Zero not performed Select and perform Zero
Gasman not configured for field button calibration
Note: Remove regulator from gas cylinder when not in use over a prolonged period. This will cause the gas to leak.
For parts list, see section XI.
One button Calibration will check for small drifts of stored calibration value, but Crowcon recommends Gasman is sent for full certified calibration at six monthly intervals.
cylinder as needed
flow
cylinder as needed
needed
flow
Repeat test, starting gas immediately
bottle as needed
needed
flow
from menu
Send for re-configuration
22
Gasman Maintenance and Calibration
VII. Maintenance and Calibration
Gasman is designed to operate almost maintenance free under most conditions. However, some small items of routine maintenance are recommended.
General
To keep the display panel and operator button free from dirt build up, regularly wipe over your Gasman unit with a damp cloth.
Filter
Inspect the front filter at regular intervals for dirt or damage. Clean as necessary.
Zero and calibration
Gasman is supplied with an Autozero function on start-up. This function can be configured to operate automatically, on user confirmation (see quick start guide), or can be disabled. This configuration can be set with the Crowcon Portables PC software, see section VIII. Gasman also has a Zero function in the menu, see section 2.4.
Crowcon recommends a monthly gas test to confirm sensor operation. However, please check with your own Health and Safety regulations. A test gas of known composition needs to be applied to verify sensor response and alarm function, see section 6.1.
Instrument calibration of sensor should be performed at 6 month regular intervals.
Calibration method
Gasman calibration can be performed using the Gas Test Kit as per 6.3 or with Crowcon Portables PC software and calibration gas mixtures. Appropriate certified calibration gases should be used. Calibration gas is applied using the appropriate flow cap. Refer to Crowcon Portables PC software help file for further information.
23
Maintenance and Calibration Gasman
Gasman IR CO2 Personal Detector
The concentration of CO2 in fresh air is 0.04% by volume. A correctly zeroed Gasman IR CO ambient air.
monitor will read 0.04% in uncontaminated
2
Before zeroing the Gasman must be in fresh uncontaminated air well away from a building or any CO breathing zone of the operator – i.e. at arm’s length. When Gasman IR CO is zeroed in fresh air it automatically sets the CO
Gasman IR CO and Portables PC software. This is in effect a zero point calibration
can also be set to zero in nitrogen using the PC interface
2
emissions. Hold the unit well away from the
2
baseline level to 0.04%.
2
rather than a normal zero. To do this, after uploading the instrument configuration file, note the existing calibration gas level, then set the calibration value to 0.00%, apply nitrogen gas and click the Calibrate button. After this zero point calibration is completed, remember to re-set the calibration gas level to its previous level.
Nitrogen for zeroing should be provided from a suitable gas bottle at a regulated flow of 0.5 l/min (1 SCFH), connected to the standard flow cap clipped over the sensor aperture. Recommended calibration gas is 2% CO
in a background of nitrogen.
2
Care should be taken to vent the nitrogen flow away from the operator and not allow nitrogen flow for longer than necessary to complete the zeroing. If indoors it is recommended to vent gas out of a window or into an extracted fume cupboard.
Ozone sensor
Due to the reactive nature of ozone (O3) special procedures should be followed when calibrating gas detectors that incorporate an ozone sensor. By following the guidelines for calibration below users will maximise the protection level that their gas detectors give them.
Always calibrate Crowcon ozone detectors with ozone only.
2
Fittings
For ozone calibrations all fittings and pipework should be stainless steel, brass, aluminium or PTFE. Fittings or pipework in other plastic materials such as Tygon must not be used.
24
Gasman Maintenance and Calibration
Gas flow rate
It is important that the flow rate is set correctly as otherwise pressure effects may distort calibration values and prevent the gas detector from working correctly. To calibrate Crowcon portable detectors the flow rate should be set between 0.8-1.0 litres/minute (0.03-0.04 cubic feet/ minute).
Normal safety precautions for handling ozone should always be observed along with any special instructions that accompany the calibration gas cylinder or generator being used.
The standard flow plate should be used to supply gas at the flow rate stated above.
If required your local Crowcon representative or service centre will be able to provide further advice.
25
PC interface and software Gasman
VIII. PC interface and software
Gasman can be connected to a PC using the single way charger unit with optional PC interface. The charger unit is fitted with a D-type 9 pin RS232 socket which is located at the rear of the charger, see diagram below. The
PCrequiresCrowconPortablesPCsoftware.AUSB-RS232adaptorisalso
available from Crowcon.
HS
2
RS 232 link
The software provides the user with access to reconfigure alarm levels, operation, run calibrations, print reports and to access data and event log files.
Set-up
1. Install Portables PC software on PC and attach the RS 232 cable to
the charger and PC.
2. Switch on the Gasman unit and drop it into the charger unit with the display facing forward.
3. For information on using the Crowcon Portables PC software, see installed help file.
26
Gasman PC interface and software
Data logging
Data is logged at a rate set by the log period which is configurable via Portables PC software. This is set at 1 minute intervals as a default.
The Gasman is capable of storing 54,000 logs (>4,800 events). When the memory is filled, new data then overwrites the oldest data.
A log can be extracted from the Gasman and saved using Portables PC software.
The instrument will record the following events:
• Switch on/off
• Battery status
• Sensor channel over-range
• Alarms activated/cleared (including peak values of response)
• Gas test pass/fail
• Usertriggeredevent
For more information on using Crowcon Portables PC software see the installed help file.
27
i-module replacement Gasman
IX. i-module replacement
Installing or replacing an i-module
1. Ensure you are in a non-hazardous (safe) area, with suitable ESD
protection. Switch off the unit
2. Remove the back cover by unscrewing the four M2.5, 12 mm Torx (T6) screws as shown the in the drawing, point . Do not touch the charging elastomer connector with your fingers.
i-module
Elastomer
3. Place the Gasman unit face down on a surface.
4. Unclipthei-modulefromitsretainingsupportclip.Easeonesideout
at a time. Ensure the elastomer which is retained within the body of the support clip remains in place and should not be touched by hand.
5. Unwrapthereplacementi-modulefromanypackagingandensure
the sensor is fully seated on the module board.
If replacing an i-module with one of the same type, instrument specific configuration will be retained. If replacing with a different i-module its default configuration will be loaded.
28
Gasman i-module replacement
6. Ensure the gasket is in place on the sensor. Then slip the sensor onto the sensor housing. Click the quick release fixings around the i-module board, ensuring the i-module is held in place firmly and the sensor is still tightly located on the module board.
7. Replace the back of the Gasman case and secure with screws.
8. Switch on the Gasman unit. The new i-module will be automatically
recognised.
9. Crowcon advise carrying out a calibration check when any new sensor is fitted.
29
Specification Gasman
X. Specification
Dimensions 90 x 48 x 24 mm (31/2 x 1.9 x 1 inches) Weight 138 g flammable
Housing, degree of protection
Operating temperature -20°C to +55°C (-4°F to +131°F) Humidity 20-99% RH, non-condensing for continuous operation Display Custom LCD with backlight. Starburst characters for number
Warm up time 90 seconds maximum Response time (typical) (T90) : approx 20 seconds for most toxic and flammable
Audible Alarms 95 dBA (Multiple alarm sounds allow selection of distinctive
Visible Alarms Dual colour red/blue flashing LEDs in gas hazard. Vibrating Alarm Internal vibrating alarm. Repeatability ±2% FSD, 6 months ATEX Safety certificate no.
IECEx
Approval codes Europe:
USA:
Canada: Standards Safety:
Canada: Operation
30
129 g oxygen 118 g toxic Ingress protection IP65 (NEMA 4)
and text display, plus screen icons for status and mode.
sensors, 10 seconds for oxygen.
tones for different alarms).
Essential Health and Safety Requirement, clause 15.9 BASEEFA04ATEX0383 Flammable Gas BASEEFA04ATEX0384 Oxygen or Toxic Gas
IECExBAS05.0038 Flammable Gas ECExBAS05.0039 Oxygen or Toxic Gas
ATEX II 1G Ex ia IIC T4 Ga, (-20°C ≥Ta≥ +65°C) Toxic/Oxygen ATEX II 2G Ex ia d IIC T4 Gb, (-20°C ≥Ta≥ +65°C) Flammable Class I Division 1, Groups A, B, C and D Class I Division 1, Groups A, B, C and D
Baseefa 04ATEX0383 EN 60079-0: 2012 EN 60079-1: 2007 EN 60079-11: 2012 94/9/EC IEC 60079-0 : 2011 I EC 60079-1 : 2007-04 IEC 60079-11 : 2011 Baseefa 04ATEX0384 EN 60079-0: 2012 EN 60079-11: 2012 94/9/EC IEC 60079-0 : 2011 IEC 60079-11 : 2011 CSA22.2, 152 EN50270, EN50271
Gasman Accessories and spare parts
XI. Accessories and spare parts
Accessory list
Crowcon part number Description
Single way chargers
C01941 Single way charger 12Vdc input C011318 Single way charger with multi-region power supply C011011 Single way charger 230V in-line power supply, no plug C011009 Single way charger 110V in-line power supply, no plug C01945 Single way charger 90-260V in-line power supply C01296 Charging lead for vehicle lighter socket lead C01940 Single way charger/interface 12Vdc input
C011305 Chargerinterfacekit(includessinglewaycharger/interface,USB-
C011319 Single way charger/interface with multi-region power supply C011012 Single way charger/interface 230V in-line power supply, no plug C011010 Single way charger/interface 110V in-line power supply, no plug C01950 Single way charger/interface with 90-260V in-line power supply
Multiway charger
C01951 5-way multiway charger without power supply
C011041 5-waymultiwaychargerwith90-260Vin-linepower supply,US
C011042 5-waymultiwaychargerwith 90-260Vin-line powersupply,UK
C011043 5-waymultiwaychargerwith90-260Vin-linepower supply,EU
i-modules:
S011424/M 0-100% LEL methane S011436/M 0-100% LEL propane S011437/M 0-100% LEL pentane S011439/M 0-100% LEL butane S011440/M 0-100% LEL ethylene
RS232 adaptor and PC lead, multi-region power supply )
power supply
power supply
power supply
31
Accessories and spare parts Gasman
S011460/M 0-100% LEL hydrogen
S011423/M 0-25%oxygen-add/USAforUSAalarmconfig.
S011422/M 0-500 ppm carbon monoxide
S011421/M 0-100ppmhydrogensulphide-add/USAforUSAalarmconfig.
S011425/M 0-10 ppm sulphur dioxide S011429/M 0-2000 ppm hydrogen S011426/M 0-20 ppm nitrogen dioxide S011428/M 0-20 ppm chlorine S011432/M 0-1 ppm ozone S011430/M 0-25 ppm hydrogen cyanide S011435/M 0-100 ppm ammonia S011438/M 0-1000 ppm ammonia S011431/M 0-5 ppm phosphine S011434/M 0-1 ppm fluorine S011433/M 0-10 ppm hydrogen fluoride S01217 1/M 0-5% Gasman IR (carbon dioxide) CO
only)
i-module (for use in safe areas
2
Sampling accessories:
M04852 Flow cap C01937 Aspirator Assembly Calibration gas contact Crowcon - required gases depend on
sensor combination
Carrying and wearing:
C01952 Harness Plate C01843 Shoulder Strap C01844 Chest Harness Straps C01953 Hard Hat Clip M02362 Alligator Clip C03329 Metal Clip
Communications:
E07532 PC Interface Lead C01832 PC Software
C02097 USBtoRS232adaptor
Spares / consumables:
S011818/1 Rear Moulding
32
Gasman Accessories and spare parts
M04973 Gasman front and rear moulding set E01918 Non-rechargeable Battery E07621 i-module Elastomer E07620 LCD Elastomer M04682 Sensor Sealing Ring M03705 Clip retaining nut M3 M03793 Case screw M05910 Crowcon domed label E01552 Power Supply for charger - 230 V No plug E01553 Power Supply for charger - 115 V No plug
E07693 PowersupplyforchargerUniversal90-260V C03580 Multi-region powersupply(includes UK, EU,US and Australian
plug adapters)
33
Accessories and spare parts Gasman
XII. Troubleshooting guide
Symptom/ Cause Action error message
Instrument won't switch on Flat battery Recharge or replace battery.
No confidence beep Function disabled Reconfigure with PC software.
Gas reading when no gas Zero drifted Restart instrument in clean present air.
Unstable/inaccurategas Sensorfailure Donotuse;exithazardous
reading area immediately. Return instrument for recalibration or sensor replacement.
Autozero failed Zeroing in Switch off and restart in contaminated clean air. atmosphere
Cannot Autozero due to Zeroing in Switch off and restart in alarm contaminated clean air. atmosphere
Calibration expired The calibration Send for calibration. due date has passed
Display shows empty Battery depleted Charge or change as battery symbol on appropriate. switch on
34
Gasman Appendix: Limitations of sensors
Appendix: Limitations of sensors
Sensor limitations
The sensors used in Gasman have limitations common to all such gas sensors, and users should be aware of the points listed below. Crowcon can advise on particular situations and suggest alternative sensors if the instrument is likely to experience extreme conditions.
Gasman Flammable uses a catalytic gas sensor, which measures the flammability of the gas. For this reason, readings displayed on the unit will be unreliable over concentrations of approximately 120% LEL. Oxygen is necessary for catalytic sensors to operate. A 'pellistor saver' is used to disconnect power to the pellistor sensor in the event of over-range to prevent burn out. This locks out the unit for 200 seconds after which a button press will reconnect power to the pellistor. If the sensor power is reconnected when the unit is exposed to an over-range gas concentration there is a risk of damage to the pellistor sensor. Restart should be carried out in a known fresh air environment. Depleted oxygen levels can reduce the flammable gas reading, and if oxygen levels are below safe breathing limits it should be assumed that the flammable reading is low.
Electrochemical gas sensors, toxic gases or oxygen, contain chemicals. Extreme levels of humidity can also cause the sensor to be unstable. The sensors are rated for an (average) ambient of 20-90% R.H. However they are used from the tropics to deserts to tundra.
Water should not be allowed to collect on the sensor as this may impede gas diffusion.
Persistent exposure to high levels of toxic gas will shorten the life of a toxic sensor. If the high level gas is corrosive (e.g. hydrogen sulphide) damage may occur over time to metal components.
Sensors may be cross sensitive to other gases. If unsure, contact Crowcon or your local agent.
35
UK Office
Crowcon Detection Instruments Ltd 172 Brook Drive, Milton Park, Abingdon Oxfordshire OX14 4SD Tel: +44 (0) 1235 557700 Fax: +44 (0) 1235 557749 Email: sales@crowcon.com Website: www.crowcon.com
USA Office
Crowcon Detection Instruments Ltd 1455 Jamike Ave, Suite 100 Erlanger KY 41018 Tel: +1 859 957 1039 or 1 800 527 6926 Fax: +1 859 957 1044 Email: salesusa@crowcon.com Website: www.crowcon.com
Netherlands Office
Crowcon Detection Instruments Ltd Vlambloem 129 3068JG, Rotterdam Netherlands Tel: + 31 10 421 1232 Fax: + 31 10 421 0542 Email: eu@crowcon.com Website: www.crowcon.com
Singapore Office
Crowcon Detection Instruments Ltd Block 194, Pandan Loop #06-20 Pantech Industrial Complex Singapore 128383 Tel: + 65 6745 2936 Fax: +65 6745 0467 Email: sales@crowcon.com.sg Website: www.crowcon.com
China Office
Crowcon Detection Instruments Ltd (Beijing)
Unit316,Area1,TowerB,Chuangxin
Building 12 Hongda North Road, Beijing Economic Technological Development Area Beijing, China 100176 Tel: +86 10 6787 0335 Fax: +86 10 6787 4879 Email: saleschina@crowcon.cn Website: www.crowcon.com
Loading...