The following instructions apply to equipment
covered by certificate number:
Baseefa04ATEX0383 Flammable Gas
IECEx BAS 05.0038 Flammable Gas
Baseefa04ATEX0384 Oxygen or Toxic Gas
IECEx BAS 05.0039 Oxygen or Toxic Gas
The following information covers all relevant
points listed in clause 1.0.6 of the EHSR’s of
the ATEX directive.
The certification marking is as follows:
1. The equipment may be used in Zones 1
and 2 for flammable versions and Zones 0,
1, and 2 for toxic and oxygen versions, for
Group IIA, IIB, and IIC gases and vapours
for Temperature Classes T1, T2, T3 and T4
2. The equipment is certified for use in
ambient temperatures in the range –20°C
to +65°C (-4 to +149 F). The equipment
should not be used outside these ranges.
3. Compliance with the Essential Health and
Safety Requirements has been assured by
compliance with EN60079-0:2004 + Amds
1 & 2, EN60079-1:2004 and EN6007911:2006, as certified by Baseefa.
Compliance with gas detection performance
standards EN50054, EN50057, EN61779-1,
EN50104 and EN50270 has been certified
by Lloyd’s Register
4. Repair of this equipment and gas sensor
replacement shall be carried out by the
manufacturer or in accordance with the
procedures in the manual and by trained
personnel.
5. If the equipment is likely to come into
contact with aggressive substances, then
it is the responsibility of the user to take
suitable precautions that prevent it from
being adversely affected, which may
compromise the certification level.
6. The rechargeable battery must only be
charged in non-hazardous (safe) areas
by connection to the specified Crowcon
charger.
7. Only the following cell types may be
fitted in the battery compartment of the
non-rechargeable battery pack: CR2 of
the following brands only: Panasonic, GP,
Energiser and Maxell. Do not fit a Duracell
battery.
The cells must only be changed in a non-
hazardous (safe) area.
8. The equipment is not certified for use in
atmospheres containing more than 21%
oxygen.
From the 1st November 2010, EN6007929 part 1 has been harmonised under
the ATEX directive 94/9/EC. Therefore to
comply with the ATEX directive, portable
apparatus sensing flammable gases should
have a functional check with gas before
each day of use. Other testing regimes
may be employed depending on local
circumstances.
Area Classifications: -
Zone 0: An area classified as Zone 0 is an
area in which an explosive mixture
of gas is continuously present or
present for long periods.
Zone 1: An area classified as Zone 1 is likely
to have ignitable concentrations of
flammable gases, vapours or liquids
present under normal operating
conditions.
Zone 2: An area classified as Zone 2 is not
likely to have ignitable concentrations of flammable gases, vapours
or liquids present under normal
operating conditions.
XI. Accessories and spare parts ...................31
XII. Troubleshooting guide ......................34
Appendix: Limitations of sensors ..................35
Gasman Unpacking
Gasman
Personal Single Gas Monitor
Thank you for purchasing the new Gasman Personal Gas Monitor.
Gasman has redefined portable gas monitoring and will give you years
of unparalleled service and reliability.
Please read the instructions carefully before use. Keep the manual for
future reference.
Unpacking
Remove the Gasman Personal Gas Monitor from the packaging. The
Gasman accessories will be located in the bottom of the box. Check the
contents are complete, you should have:
• Gasman unit with alligator pocket clip;
• A configuration report detailing the sensor installed, alarm
settings and a calibration certificate;
• Calibration flow cap and tubing;
• Usermanual.
Optional battery chargers and other accessories will be packed in a
separate box.
Battery check
The Gasman Personal Gas Monitor has two battery options: Li-ion
rechargeable or non-rechargeable batteries. The label on the back of the
detector will have either NR (non-rechargeable) or R (rechargeable) to
indicate which battery option is within the detector.
Rechargeable units
Gasman uses a Li-ion battery pack and should arrive with sufficient
charge so that the unit can be used straight out the box. However, if
this is the first time you have used the Gasman unit, you may need
to charge the battery to attain the full operating time. (The actual
operating time will depend on the types of sensor installed.) The
flammable Gasman will operate for at least 12 hours on a fully charged
battery.
1
Unpacking Gasman
Warning: rechargeable units
Only charge the rechargeable Gasman Personal Monitor with a Crowcon
Gasman charger. Failure to comply could invalidate safety certification,
warranty and safety; and may result in permanent damage to the unit.
Non-rechargeable units.
Gasman uses a CR2 battery which can be easily field replaced. Ensure
only the correct types of battery are fitted to maintain compliance with
certification (see Battery check on page 1).
Gasman IR CO2 Personal Single Gas Monitor
The Gasman IR CO
for CO
detection. This version of Gasman is not designed or certified
2
for use in a hazardous area, and the safety certification data in Section
personal single gas monitor an infrared gas sensor
2
X is not applicable. Operation and maintenance of the Gasman IR CO
is essentially the same as for other Gasman units but with the following
points to note.
2
CO2 is present in ambient air at a background concentration of just under 400ppm (0.04%). In
any enclosed environment the ambient CO
and if the room is poorly ventilated concentrations well in excess of 1000ppm (0.1%) can
levels in the outdoor air will also be enhanced by any emitted CO2 e.g. from vehicle
occur. CO
2
exhausts or boiler flues venting to atmosphere so it is normal to see fluctuating levels of CO
the display whether indoors or outside.
concentration will increase as a result of respiration
2
2
2
on
Gasman Quickstart guide
Quickstart guide
1. Getting started
Gasman familiarisation
Gas
Gas
Sensor
name
label
Alarm
LEDs
Sounder
outlet
Operator
Button
Operator LCD
Display Screen
Switching on
Gasman requires little setting up. Follow these simple steps to get your
unit ready for use.
1. Ensure the unit is in clean air.
2. Switch on. Press and hold the operator button for approximately 3
seconds until the red LEDs flash.
The operator display screen will light up and
the unit will begin a warm up sequence.
The user should now check the detector
goes through the warm up sequence
detailed below. In the case of any deviations
the Gasman should be returned to your local
Crowcon office or Service agent.
2
HS
3
Quickstart guide Gasman
Gasman warm up sequence
a) The Gasman will demonstrate
the alarm LEDs, sounder,
vibration alerts, and the
Sounder
Alarm test
HS
2
LEDs Alarm test
operator display screen. The
sounder may be silenced by
pressing the button.
Vibrator Alarm test
The Gasman screen will show the following display during the
warm up sequence. This will take approximately 20 seconds.
0.0
OK
ppb
ppm
vv/v
ppm
Note: these screens may differ depending on what sensor is being used.
b) Autozero
If Autozero is enabled (default), the unit will display the Autozero
confirm screen: the display will alternate between ‘ZERO’ and
‘????’. Press the operator button with a single click to confirm
Zero. If the operator button is not pressed within the 10 second
time out, Gasman will proceed directly to Run mode without
performing Zero.
Autozero confirm
ppm
OK
No Autozero
OK
ppm
Screen Icons
OK
AL -1
AL -2
Warm up
Flashing
icon Gasman
running
normally
Alarms
Battery life
OK
Autozero
0.0
4
Gasman Quickstart guide
Confidence signals
In normal Run mode, Gasman will emit a short beep accompanied by a blue LED flash every 10
seconds and the OK icon flashes to show operational health. These confidence signals can be
disabled using the PC software.
2. In the event of an alarm
Alarm signals
In the event of gas concentrations exceeding the alarm thresholds for the
gas being monitored, Gasman will activate the alarm signals.
Alarm signals
The red and blue alarm LEDs
HS
2
ppm
1. When the gas level returns to normal, press the operator button.
This will reset your Gasman unit to normal Run mode. If the gas
level is still within alarm limits, the button will have no effect.
The Gasman alarm is set to latch by default. The unit will still continue in alarm mode even
when the gas level returns to normal, until the button is pressed and the alarm is cleared.
will flash, the sounder will emit
a loud, fast series of beeps, the
internal vibrator alarm will activate.
The operator screen will display
the alarm level and gas reading
alternately.
See the figure to the left.
AL - 1 — Alarm level one
AL - 2 — Alarm level two
3. Switch off unit and storage
Switching off
1. Press and hold the button for 5 seconds until the display shows
OFF. The display will count down to switch off.
5
Quickstart guide Gasman
Storing conditions
In order to optimise sensor performance and lifetime, your Gasman unit
should be stored in a safe, non-hazardous area, at 0-20°C and 20-90%
RH.
4. Additional information
For battery recharging information go to section III.
For fixing accessories go to section V.
For sampling section go to section VI.
For calibration information go to section VII.
For troubleshooting guide go to section XII.
6
Gasman Introduction
I. Introduction
Thank you for purchasing the new Gasman Personal Single Monitor.
Gasman is a portable Single Gas detector, designed to be carried or
worn by individuals working in hazardous environments such as confined
spaces. It is suitable for use in classified hazardous areas. Gasman
monitors a single gas and displays its reading on a display screen. Alarm
warnings are given through a combination of a loud audible alarm, a
bright visual alarm of blue/red flashing LEDs and an internal vibrator.
Gasman can be fitted with a wide range of modular, or plug and play
gas sensors. The sensor carries an intelligent processor which contains
calibration and sensor information.
Gasman is battery operated and is available with rechargeable batteries
or dry cell battery options. The dry cell battery option is only available for
toxics and oxygen Gasman units. A battery charger for single and multiple
Gasman units is available for rechargeable units, see section XI for more
information.
At Crowcon we recognised the need for a reliable and robust personal
monitoring system, which is both small, light and easy to use. Gasman
has a single operator button, and an intelligent user-friendly display with
automatic backlight. The gas level is continuously monitored providing
normal gas readings, peak readings and time weighted averages (TWA).
Gasman is available as a diffusion sampling instrument, see section XI for
sampling accessories. Configuration and data/event logging is handled by
Crowcon Portables PC software, the PC communication link is provided
as a link on the charger interface unit.
Gasman’s compact shape and design makes it comfortable to wear, with
a non-slip grip for better handling. Extra accessories such as pocket clip,
hard hat clip, shoulder strap and chest harness, can be purchased.
Gasman has been designed from top to bottom to bring you a lighter,
compact design with single button operation for ease of use, maintenance
and extreme reliability.
7
Introduction Gasman
i-module gas sensor
Gasman uses unique plug and play i-module sensor technology. Each
sensor unit incorporates its own intelligent processor holding sensor
configuration and calibration data. Different sensors can be purchased,
and once inserted, are immediately ready to run. Flammable units are only
available with rechargeable batteries. Plug and play will ease maintenance
time and cost, and the intelligent modular system will remove the need
to calibrate the sensor. Gasman can be reconfigured by purchasing
additional pre-calibrated i-modules from Crowcon or your local office.
After fitting a new i-module, it is recommended that a gas test is carried
out on the detector before use.
Reliable, anti-shock mechanics and robust housing
The Gasman housing is built from resilient material, giving it strength and
flexibility to withstand the hardest of working conditions, water and dust
tight to IP65 with a non-slip grip. If the unit is dropped, there will be no
disruption of power or function, ensuring reliability and service for years
to come.
Software
The internal software in Gasman has been designed and written in
accordance with the requirement of IEC 61508 to ensure quality and
integrity of operation. Gasman has been designed to give a truly
reliable personal gas monitoring system. The internal circuitry includes
an independent watchdog. The software monitors for any malfunction
within the unit and will display an error warning to the user should they
occur.
8
Gasman Operation
II. Operation
2.1 Switch-on sequence
1. Ensure the unit is in clean air.
NB. For CO2 detectors see zeroing, Section VII.
2.Switch on
Press and hold the operator button for approximately 3 seconds
until the red LEDs flash.
The instrument begins with testing all the LCD segments on the
operator display screen, the red and blue alarm LEDs, sounder and
internal vibrator alert. The sounder may be silenced by pressing
the button. The unit enters a warm up mode and displays a
sequence of screens, see page 3 for more details. At the end of
warm up, the Autozero menu will be displayed.
The Autozero function can be disabled
or set to run automatically, without user
confirmation: Autozero will not appear.
See section VIII PC Interface and software.
Calibration check
During the warm up sequence, if the next calibration date is due in less than 31 days,
‘CAL - nn’ will be displayed (where nn is the number of days to the next calibration). If
calibration is already overdue, the Gasman unit will display a warning message ‘CAL’.
The instrument can still function, but it is strongly recommended the unit is sent for
calibration as soon as possible.
Press the button to continue to Run mode.
The Gasman can be set, using Portables PC software, to lock out in the event of the
calibration due date expiration in order to prevent further operation of the detector until
a calibration is preformed.
3.Autozero
Press the operator button with a single click to confirm
Autozero. Flammable and toxic units will be set to read zero
and oxygen units set to read 20.9%. If the operator button
is not pressed within 10 seconds Gasman will proceed
directly to normal operation without performing a zero.
Switch off
To switch off the unit, press and hold the operator button for 5
seconds. The display will count down from 5 until the unit counts
down to shut off.
Battery check
Usethistimetocheckthereis
!
sufficient charge in the battery
pack
NB. If Autozero
fails, a warning
message will be
displayed.
9
Operation Gasman
2.2 Run mode
The Gasman unit displays the gas reading on the operator
display screen. A typical display is shown below:
OK
ppm
The sensor channel will display the current value of the gas
being monitored and the unit of measurement, the ‘OK’
symbol will flash indicating unit is working correctly. Familiarise
yourself with the gas being monitored in your unit. Ensure you
understand site health and safety procedures. For information
on Peak and TWA readings, go to section 2.4.
Confidence signals
To reassure users the unit is working correctly, the Gasman unit will emit a short confidence
beep accompanied by a blue LED flash every 10 seconds and the OK icon will flash continuously.
Screen Icons
Warm up
Flashing
OK
icon Gasman
running
normally
AL -1
Alarms
AL -2
Battery life
Autozero
0.0
2.3 Display symbol guide
Battery
Full
TWA alarm
Gasman will display the TWA alarm for toxic gases when the 15 minute
or 8 hour Time Weighted Average alarm threshold is passed.
Gasman will display ‘LTWA’ and ‘STWA’. When a TWA alarm threshold is
reached the TWA alarm cannot be cleared.
A full battery is represented by a battery icon showing full
three bars. A low battery charge will show 1 bar. When
zero bars are shown the battery icon flashes and the
!
sounder will emit warning beeps.
If the battery becomes too low for operation the Gasman will
switch off. A low battery warning will alarm prior to switch off.
10
Gasman Operation
2.4 Display options
Gasman provides four additional selectable displays:
Peak display
When Peak mode is selected the instrument shows the
highest value for flammable and toxic gases, or the lowest
value for oxygen since switch on or last Peak Reset. This is
useful for vertical entry checks where the Gasman can be
lowered down the shaft rather than using a sampling tube
and for viewing peak exposure at the end of a shift.
TWA display
Shows the 15 minute and 8 hour TWA, for toxic gases,
monitored since last turn on.
Peak Reset
Before a Peak entry test is performed select this menu
option to clear any previously stored Peak value.
Zero
Performs Zero on your Gasman unit. (Please ensure you are
0.0
in clean air).
How to display the menu
1. To view the additional display option menu, double-click the button.
The menu icons will appear on the screen, as shown below.
2
HS
0.0
OK
ppm
0.0
Note: Only toxic gas
OK
monitoring units will
ppm
display the TWA menu
option.
2. Press the operator button with a single click to scroll across the
list. When the underscore appears under your choice, double-click
the operator button. This will select the option.
If Peak or TWA is selected, Gasman will display the icon on the
operator screen.
11
Operation Gasman
Peak test
When performing a Peak test, such as a vertical entry check,
previous readings can be cleared by selecting the Peak Reset menu
option
.
1. Scroll menus2. Select
0.0
OK
ppm
Peak display
Zero
To perform Autozero, select the Zero function from the menu.
When Zero is completed, the instrument will return to normal Run
mode.
3. Run mode
The Gasman is now ready to use.
Familiarise yourself with the gas being monitored in your unit and
make sure you understand site health and safety procedures in the
event of alarm conditions.
OK
ppm
12
Gasman Operation
2.5 Logging
Gasman incorporates both data and event logging which can be
accessed using the RS232 communications link available with the Single Way Charger Interface (part No.C01940), using Crowcon Portables PC
software. See section VIII.
Data is recorded every minute (this rate can be adjusted using the PC
software). The log can record 900 hours of data at 1-minute intervals.
Gasman also records the time and date for a number of operating and
diagnostic events including:
• Switch on and switch off;
• Level 1, Level 2 and TWA, alarm on, alarm off and the peak level
during the alarm;
• Zero, calibration and gas test with success or failure;
• Flammable sensor saver on and off;
• The battery condition is logged every switch on and switch off
while the instrument is operating, and certain configuration
changes are also logged;
• The Event log can record over 4800 events.
13
Batteries Gasman
III. Batteries
3.1 Rechargeable batteries
Recharge time for the Li-ion batteries is less then 6 hours (less, if they are
not fully discharged). Rechargeable batteries will typically last 12+ hours
for flammable units.
3.2 Gasman Charging Unit
There are 3 models of charging units available for Gasman; a single
way drop in charger unit, a single way charger unit with integral
PC Interface, and a multiway charger unit. The multiway charger unit
can accommodate up to 5 Gasman units. Note: the multiway charger
unit does not have a PC interface option.
The charger units are powered by a nominal 12 V dc input. Power
90-260 V power supply is available for any other configuration. The
multiway charger includes a universal power supply. A vehicle lighter
socket lead can also be supplied, please see section XI Parts and
Accessories.
Communications and charging are only permitted via Crowcon charger/
charger-interface C01940.
Charging the batteries
1. Ensure you are in a
safe area.
2. Plug the charger
power supply into
a mains socket,
and connect to the
charger lead.
3. To charge your
Gasman, simply drop the Gasman into the charger in an upright
position with the screen facing outwards, see diagram below:
14
12Vdc
Power
supply
Gasman Batteries
The Gasman should
normally be switched
off for charging. During
HS
2
Power
supply
charging the red LED on
the front of the charger
unit will be lit. When
charging is complete the
12Vdc
red LED is extinguished.
During charging, the display will show the battery icon sweeping
from empty to full. On removing Gasman from the charger unit,
if the Gasman is on, the display icon will update in 20 seconds to
show actual charge state.
Chargers with dual LEDs are no longer suitable for use with the
current Gasman. We recommend that such chargers be replaced
with the newer model shown above (see Section XI, Accessories
and spare parts, p32).
The charging time will be longer if the unit is switched on during charging.
When Gasman is fully charged and switched on the battery icon
will display three full bars.
3.3 Changing rechargeable batteries
It is recommended that rechargeable batteries should be changed by an
authorised Crowcon service centre only.
3.4 Non-rechargeable batteries
Gasman uses a Lithium Cell battery pack which will give up to two years
of operation.
Always switch the Gasman off before opening the case to change the battery.
To replace the battery, ensure you are in a safe, non-hazardous area.
Remove the back cover and remove the battery*. Insert the replacement
battery into the instrument and securely fasten the back cover.
Note: When a non-rechargeable detector is placed in the charger/charger
interface the red LED will not illuminate and the detector will not charge.
* Note: later versions of the Gasman will have a tab in place to aid
removal of the battery.
15
Alarm indications Gasman
IV. Alarm indications
Gasman provides two instantaneous alarm levels, designated level 1 and
level 2. For toxic gas sensors, there are also two TWA alarms, one for short
term exposure (STEL): based on a 15 minute time weighted average, and
the second TWA alarm is for long term exposure (LTEL): based on a 8 hour
time weighted average.
Alarm configurations are set via the Crowcon Portables PC software.
The following settings can be made:
Alarm thresholds for each sensor: Level 1 and level 2
instantaneous alarms.
Alarm trigger: This can be set to rising levels of gas
concentration, or as falling. Oxygen units are set to falling
for deficiency monitoring.
Alarm latching: Alarms can be set to be latched or unlatched. Latched
alarms will require the button to be pressed in order to clear the alarm
will clear automatically when the gas hazard has passed, but not until
then.
Alarm mute: The sounder can be set to mute for level 1 alarm only;
pressing the operator button during an alarm condition i.e. presence of
hazardous gas, will silence the sounder and stop the vibration alarm. The
alarm LEDs will continue to flash.
Alarm sounder tone: Different distinctive tones can be selected for the
each alarm condition to achieve the best performance for the monitoring
conditions available.
In the event of a Time Weighted Alarm (TWA)
In the event of the 15 minute or the 8 hour TWA being
triggered, Gasman will go into alarm and display a TWA
warning with the toxic gas readings. Neither the 15 minute
nor the 8 hour TWA alarm can be cleared.
16
Gasman Alarm indications
In the event of a Flammable overange alarm
If flammable gas concentrations exceed 100% LEL Gasman
locks into alarm and displays ‘9999’ showing an overange
condition. Gasman will temporarily cut off power to the
sensor to prevent burn out and display a progress bar for 200 seconds.
When the timeout is complete, either press the button to continue or
switch your unit off and on to restart. This option is programmable with
Crowcon Portables PC software and the default is switched on.
If the Gasman goes over range at any time it is recommended to carry out
a gas test before using it again.
17
Fixing Accessories Gasman
V. Fixing Accessories
Clip Accessories
Gasman is supplied with an alligator clip. Other clip options are:
Hard hat clip
Allows Gasman to be clipped to a hard hat, a method of wearing a
portable gas detector preferred by many users.
Pocket Clip
Allows Gasman to be clipped to a pocket, near the breathing zone of
the user.
Universal harness plate
Crowcon provide a universal harness plate which can be used with either
a chest harness or a shoulder strap.
How to wear your Gasman unit
Chest harness
Use the M3 fixing (usable once the
Shoulder/neck
connectors
clip has been removed) on the back
of your Gasman unit to attach the
M3 nut
chest harness plate. Create a chest
harness by attaching one strap to the
top connectors, to go around the neck,
and the other to link around the waist
Waist
connectors
using the side connectors. Adjust the
strap lengths until the Gasman unit is in
a comfortable working position.
Chest harness plate
Shoulder strap
With the universal harness plate in place on the belt clip, attach the
shoulder strap accessory onto the top connectors. Adjust to a comfortable
working position.
See Accessories, section XI, for full list.
18
Gasman Flow sampling
VI. Flow sampling
Attaching the flow cap
To perform manual sampling using Gasman, a flow cap must be fitted
onto the front of the unit, over the sensor. A flow cap is included with
your Gasman unit.
1. To fit the flow cap, firstly ensure the cap gasket is intact, slip the
flow cap over the sensor on the front of your Gasman unit, until
the cap is tightly fitted into place.
2. Attach the sampling tube or flow accessory onto a gas nozzle.
3. Attach the aspirator bulb onto the remaining gas nozzle.
4. To remove the flow cap, gently tease the cap from the sensor and
lift the cap away from the instrument.
When using the manual aspirator kit, adopt a consistent rhythm whilst
using the hand aspirator. Crowcon recommend squeezing once per
second to achieve a flow rate of approximately 0.5 - 1 litre/min. At least
10 pumps per sample are recommended.
The sampling tube supplied is normally a 2m (6ft) length. Longer lengths of sampling tube
can be provided, but will increase the time taken to get a sample from the point of sampling
to the Gasman instrument. When using an extended length of tubing a response time test is
recommended. Gas of known concentration should be sampled along the full length of tubing
to be used and the time taken for the sensor reading to reach the known gas levels should be
noted. This time should be used as the minimum for sampling before readings should be taken.
19
Gasman gas test accessory kit Gasman
Gasman gas test accessory kit
The Gas Test Accessory is a gas testing kit designed to enable gas testing
and one button calibration of your Gasman gas detector using a specially
formulated, high stability long life single gas mix. It can be used with
Gasman units having sensors for Flammable, Oxygen, Carbon Monoxide
and Hydrogen Sulphide gases.
6.1 Gas testing
Gas testing checks the sensor is responding within set limits to an applied
gas of known composition. This can be performed as often as desired, but
would usually be carried out each time the Gasman is issued for use. The
Gasman itself will determine Pass/Fail status for the gas test.
In order to perform successful gas tests you should ensure:
• The gas used has the correct gas concentration, and that it is
within the validity date specified by the supplier.
• The gas flow path is leak tight. It is important to check that the flow
cap is properly fitted to the Gasman unit, and the outlet tubing is
not restricted in any way, nor additional tubing length used.
The Gas Test Accessory Kit comprises a gas cylinder containing the gas, a
'Trigger' regulator with interconnect tubing, a magnet - used to activate
Test mode, a flow cap to attach to the Gasman, and a vent line. The kit is
supplied in a convenient carry case. The Trigger regulator can be operated
in two ways: (1) squeeze and hold - allows gas flow as long as the lever is
held in, or (2) by lifting the lever - the flow is locked on.
6.2 How to perform a gas test
1. Ensure your Gasman unit is switched on and in normal operation.
2. Fit the flow cap onto the front of the sensor and attach the hose
from the Trigger regulator.
Attach the outlet hose to ‘vent gas away’ - do
not extend this hose and do not restrict or
allow kinks.
3. Touch the magnet, horizontally aligned onto
the case, to the left hand side of the display
(as shown left). Your Gasman unit will activate
the Gas Test and show ‘TEST’ on the display.
20
HS
2
OK
ppm
Gasman Gasman gas test accessory kit
4. Gasman will display a progress bar. Apply the gas
whilst the progress bar is counting down. The display
OK
ppm
will alternate between the screens shown right.
Gasman will then display ‘PASS’ or ‘FAIL’
In the event Gasman displays ‘FAIL’, please see the
OK
ppm
troubleshooting guide in the first instance or contact
Crowcon or your local service centre.
5. To abort the gas test press the button at any time whilst the test is
in progress.
6.3 How to perform a one button
calibration test
To perform a one button calibration test, you must first perform a Zero
on your Gasman unit.
1. Ensure you are in clean air.
Double click the button and select Zero from the Options menu.
Gasman will perform a Zero.
To perform a one button Calibration, complete the next steps within 15
minutes of completing the Zero.
2. Follow steps 1 to 3 given in 6.2, Gasman will display ‘CAL’ and ‘????’
alternately. Press the button to confirm one button Calibration.
If the button confirmation for calibration is not made within 10
seconds then the process will revert to test as in 6.2.
3. Apply calibration gas following step 4 in 6.2.
4. To abort the Calibration test press the button at any time whilst
the test is in progress.
Your Gasman unit will adjust the value for the gas channel to match the
stored calibration gas value within the sensor i-module. This will also
change the calibration due date which is set as a default of 182 days.
If your Gasman unit does not calibrate successfully, Gasman will display
‘FAIL’. Your Gasman unit must be sent to Crowcon or a local service centre
for re-calibration.
Any Gas Test ‘pass’ and ‘fail’, and calibration ‘pass’ and ‘fail’ and their
values are stored in the Event log.
21
Gasman gas test accessory kit Gasman
6.4 Gas test/calibration troubleshooting
SymptomPossible CauseAction
No response to gasGas cylinder emptyCheck gauge, replace
Hose blocked or kinkedEnsure no restriction to
Gasman fails gas
test
Gasman fails
calibration
Gasman passes gas
test but will not
enter calibration
mode
Gas cylinder emptyCheck gauge, replace
Gas cylinder out of dateCheck date and replace as
Hose blocked or kinkedEnsure no restriction to
Calibration driftedCalibrate Gasman
Gas flow not started
immediately
Gas cylinder emptyCheck gauge replace
Gas cylinder out of dateCheck date and replace as
Hose blocked or kinkedEnsure no restriction to
Calibration driftedCalibrate Gasman
Stabilisation time too shortReset using PC software
Menu Zero not performedSelect and perform Zero
Gasman not configured for
field button calibration
Note: Remove regulator from gas cylinder when not in use over a
prolonged period. This will cause the gas to leak.
For parts list, see section XI.
One button Calibration will check for small drifts of stored calibration
value, but Crowcon recommends Gasman is sent for full certified
calibration at six monthly intervals.
cylinder as needed
flow
cylinder as needed
needed
flow
Repeat test, starting gas
immediately
bottle as needed
needed
flow
from menu
Send for re-configuration
22
Gasman Maintenance and Calibration
VII. Maintenance and Calibration
Gasman is designed to operate almost maintenance free under most
conditions. However, some small items of routine maintenance are
recommended.
General
To keep the display panel and operator button free from dirt build up,
regularly wipe over your Gasman unit with a damp cloth.
Filter
Inspect the front filter at regular intervals for dirt or damage. Clean as
necessary.
Zero and calibration
Gasman is supplied with an Autozero function on start-up. This function
can be configured to operate automatically, on user confirmation (see
quick start guide), or can be disabled. This configuration can be set with
the Crowcon Portables PC software, see section VIII. Gasman also has a
Zero function in the menu, see section 2.4.
Crowcon recommends a monthly gas test to confirm sensor operation.
However, please check with your own Health and Safety regulations.
A test gas of known composition needs to be applied to verify sensor
response and alarm function, see section 6.1.
Instrument calibration of sensor should be performed at 6 month regular
intervals.
Calibration method
Gasman calibration can be performed using the Gas Test Kit as per 6.3
or with Crowcon Portables PC software and calibration gas mixtures.
Appropriate certified calibration gases should be used. Calibration gas is
applied using the appropriate flow cap. Refer to Crowcon Portables PC
software help file for further information.
23
Maintenance and Calibration Gasman
Gasman IR CO2 Personal Detector
The concentration of CO2 in fresh air is 0.04% by volume. A correctly
zeroed Gasman IR CO
ambient air.
monitor will read 0.04% in uncontaminated
2
Before zeroing the Gasman must be in fresh uncontaminated air well away
from a building or any CO
breathing zone of the operator – i.e. at arm’s length. When Gasman IR CO
is zeroed in fresh air it automatically sets the CO
Gasman IR CO
and Portables PC software. This is in effect a zero point calibration
can also be set to zero in nitrogen using the PC interface
2
emissions. Hold the unit well away from the
2
baseline level to 0.04%.
2
rather than a normal zero. To do this, after uploading the instrument
configuration file, note the existing calibration gas level, then set the
calibration value to 0.00%, apply nitrogen gas and click the Calibrate
button. After this zero point calibration is completed, remember to re-set
the calibration gas level to its previous level.
Nitrogen for zeroing should be provided from a suitable gas bottle at a
regulated flow of 0.5 l/min (1 SCFH), connected to the standard flow cap
clipped over the sensor aperture. Recommended calibration gas is 2%
CO
in a background of nitrogen.
2
Care should be taken to vent the nitrogen flow away from the operator
and not allow nitrogen flow for longer than necessary to complete the
zeroing. If indoors it is recommended to vent gas out of a window or into
an extracted fume cupboard.
Ozone sensor
Due to the reactive nature of ozone (O3) special procedures should be
followed when calibrating gas detectors that incorporate an ozone sensor.
By following the guidelines for calibration below users will maximise the
protection level that their gas detectors give them.
Always calibrate Crowcon ozone detectors with ozone only.
2
Fittings
For ozone calibrations all fittings and pipework should be stainless steel,
brass, aluminium or PTFE. Fittings or pipework in other plastic materials
such as Tygon must not be used.
24
Gasman Maintenance and Calibration
Gas flow rate
It is important that the flow rate is set correctly as otherwise pressure
effects may distort calibration values and prevent the gas detector from
working correctly. To calibrate Crowcon portable detectors the flow
rate should be set between 0.8-1.0 litres/minute (0.03-0.04 cubic feet/
minute).
Normal safety precautions for handling ozone should always be observed
along with any special instructions that accompany the calibration gas
cylinder or generator being used.
The standard flow plate should be used to supply gas at the flow rate
stated above.
If required your local Crowcon representative or service centre will be able
to provide further advice.
25
PC interface and software Gasman
VIII. PC interface and software
Gasman can be connected to a PC using the single way charger unit with
optional PC interface. The charger unit is fitted with a D-type 9 pin RS232
socket which is located at the rear of the charger, see diagram below. The
The software provides the user with access to reconfigure alarm levels,
operation, run calibrations, print reports and to access data and event
log files.
Set-up
1. Install Portables PC software on PC and attach the RS 232 cable to
the charger and PC.
2. Switch on the Gasman unit and drop it into the charger unit with
the display facing forward.
3. For information on using the Crowcon Portables PC software, see
installed help file.
26
Gasman PC interface and software
Data logging
Data is logged at a rate set by the log period which is configurable via
Portables PC software. This is set at 1 minute intervals as a default.
The Gasman is capable of storing 54,000 logs (>4,800 events). When the
memory is filled, new data then overwrites the oldest data.
A log can be extracted from the Gasman and saved using Portables PC
software.
The instrument will record the following events:
• Switch on/off
• Battery status
• Sensor channel over-range
• Alarms activated/cleared (including peak values of response)
• Gas test pass/fail
• Usertriggeredevent
For more information on using Crowcon Portables PC software see the
installed help file.
27
i-module replacement Gasman
IX. i-module replacement
Installing or replacing an i-module
1. Ensure you are in a non-hazardous (safe) area, with suitable ESD
protection.
Switch off the unit
2. Remove the back cover by unscrewing the four M2.5, 12 mm Torx
(T6) screws as shown the in the drawing, point . Do not touch
the charging elastomer connector with your fingers.
If replacing an i-module with one of the same type, instrument specific configuration will be
retained. If replacing with a different i-module its default configuration will be loaded.
28
Gasman i-module replacement
6. Ensure the gasket is in place on the sensor. Then slip the sensor
onto the sensor housing. Click the quick release fixings around the
i-module board, ensuring the i-module is held in place firmly and
the sensor is still tightly located on the module board.
7. Replace the back of the Gasman case and secure with screws.
8. Switch on the Gasman unit. The new i-module will be automatically
recognised.
9. Crowcon advise carrying out a calibration check when any new sensor
is fitted.
29
Specification Gasman
X. Specification
Dimensions90 x 48 x 24 mm (31/2 x 1.9 x 1 inches)
Weight138 g flammable
Housing, degree of
protection
Operating temperature -20°C to +55°C (-4°F to +131°F)
Humidity 20-99% RH, non-condensing for continuous operation
Display Custom LCD with backlight. Starburst characters for number
Warm up time 90 seconds maximum
Response time (typical) (T90) : approx 20 seconds for most toxic and flammable
Audible Alarms95 dBA (Multiple alarm sounds allow selection of distinctive
129 g oxygen
118 g toxic
Ingress protection IP65 (NEMA 4)
and text display, plus screen icons for status and mode.
sensors, 10 seconds for oxygen.
tones for different alarms).
Essential Health and Safety Requirement, clause 15.9
BASEEFA04ATEX0383 Flammable Gas
BASEEFA04ATEX0384 Oxygen or Toxic Gas
IECExBAS05.0038 Flammable Gas
ECExBAS05.0039 Oxygen or Toxic Gas
ATEX II 1G Ex ia IIC T4 Ga, (-20°C ≥Ta≥ +65°C)
Toxic/Oxygen
ATEX II 2G Ex ia d IIC T4 Gb, (-20°C ≥Ta≥ +65°C)
Flammable
Class I Division 1, Groups A, B, C and D
Class I Division 1, Groups A, B, C and D
Baseefa 04ATEX0383
EN 60079-0: 2012 EN 60079-1: 2007 EN 60079-11: 2012
94/9/EC
IEC 60079-0 : 2011 I EC 60079-1 : 2007-04 IEC 60079-11
: 2011
Baseefa 04ATEX0384
EN 60079-0: 2012 EN 60079-11: 2012 94/9/EC
IEC 60079-0 : 2011 IEC 60079-11 : 2011
CSA22.2, 152
EN50270, EN50271
Gasman Accessories and spare parts
XI. Accessories and spare parts
Accessory list
Crowcon
part number Description
Single way chargers
C01941 Single way charger 12Vdc input
C011318 Single way charger with multi-region power supply
C011011 Single way charger 230V in-line power supply, no plug
C011009 Single way charger 110V in-line power supply, no plug
C01945 Single way charger 90-260V in-line power supply
C01296 Charging lead for vehicle lighter socket lead
C01940 Single way charger/interface 12Vdc input
C011319 Single way charger/interface with multi-region power supply
C011012 Single way charger/interface 230V in-line power supply, no plug
C011010 Single way charger/interface 110V in-line power supply, no plug
C01950 Single way charger/interface with 90-260V in-line power supply
Multiway charger
C01951 5-way multiway charger without power supply
M04852 Flow cap
C01937 Aspirator Assembly
Calibration gas contact Crowcon - required gases depend on
sensor combination
Carrying and wearing:
C01952 Harness Plate
C01843 Shoulder Strap
C01844 Chest Harness Straps
C01953 Hard Hat Clip
M02362 Alligator Clip
C03329 Metal Clip
Communications:
E07532 PC Interface Lead
C01832 PC Software
C02097USBtoRS232adaptor
Spares / consumables:
S011818/1 Rear Moulding
32
Gasman Accessories and spare parts
M04973 Gasman front and rear moulding set
E01918 Non-rechargeable Battery
E07621 i-module Elastomer
E07620 LCD Elastomer
M04682 Sensor Sealing Ring
M03705 Clip retaining nut M3
M03793 Case screw
M05910 Crowcon domed label
E01552 Power Supply for charger - 230 V No plug
E01553 Power Supply for charger - 115 V No plug
reading area immediately. Return
instrument for recalibration
or sensor replacement.
Autozero failed Zeroing in Switch off and restart in
contaminated clean air.
atmosphere
Cannot Autozero due to Zeroing in Switch off and restart in
alarm contaminated clean air.
atmosphere
Calibration expired The calibration Send for calibration.
due date has
passed
Display shows empty Battery depleted Charge or change as
battery symbol on appropriate.
switch on
34
Gasman Appendix: Limitations of sensors
Appendix: Limitations of sensors
Sensor limitations
The sensors used in Gasman have limitations common to all such gas
sensors, and users should be aware of the points listed below. Crowcon
can advise on particular situations and suggest alternative sensors if the
instrument is likely to experience extreme conditions.
Gasman Flammable uses a catalytic gas sensor, which measures the
flammability of the gas. For this reason, readings displayed on the unit will
be unreliable over concentrations of approximately 120% LEL. Oxygen
is necessary for catalytic sensors to operate. A 'pellistor saver' is used to
disconnect power to the pellistor sensor in the event of over-range to
prevent burn out. This locks out the unit for 200 seconds after which a
button press will reconnect power to the pellistor. If the sensor power is
reconnected when the unit is exposed to an over-range gas concentration
there is a risk of damage to the pellistor sensor. Restart should be carried
out in a known fresh air environment. Depleted oxygen levels can reduce
the flammable gas reading, and if oxygen levels are below safe breathing
limits it should be assumed that the flammable reading is low.
Electrochemical gas sensors, toxic gases or oxygen, contain chemicals.
Extreme levels of humidity can also cause the sensor to be unstable. The
sensors are rated for an (average) ambient of 20-90% R.H. However they
are used from the tropics to deserts to tundra.
Water should not be allowed to collect on the sensor as this may impede
gas diffusion.
Persistent exposure to high levels of toxic gas will shorten the life of a toxic
sensor. If the high level gas is corrosive (e.g. hydrogen sulphide) damage
may occur over time to metal components.
Sensors may be cross sensitive to other gases. If unsure, contact Crowcon
or your local agent.