Macnaught IM012E Instruction Manual

IM213D-MC Issue 10 © 2017 Page 1 of 8
INSTRUCTION MANUAL
To the Owner
POSITIVE DISPLACEMENT FLOWMETER
Important Information
Read and retain this instruction manual to assist you in the operation and maintenance of this product.
If you have any problems with the meter, refer to the maintenance and trouble shooting sections of this manual.
This manual contains connection and operating instructions for meters with Liquid Crystal displays (LCD).
If you need further assistance, please contact your local representative or distributor for advice.
Operating Principle
WARNING
This Flow Meter has incorporated the oval rotor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow.
Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flow rates are features of the oval rotor design. With a low pressure drop and high pressure rating oval rotor flow meters are suitable for both gravity and pump (in line)
Before use, confirm the fluid to be used is compatible with the meter. Refer to Industry fluid compatibility charts or consult your local representative for advice.
When fluid passes through the meter the rotors turn, as shown below. The magnets which are located in the rotors will pass across the PCB sensors. A signal is generated which is then sent by the PCB to the relevant LCD, or receiving instrument as a Pulse Output.
IM012E (ELECTRONIC)
Include s m odels
IM012E-01 IM012E-02
Need Assistance? Macnaught USA Inc (813) 628-5506 info@macnaughtusa.com
IM213D-MC Issue 10 © 2017 Page 2 of 8
Operational Overview
1. During normal operation the 6 Figure LCD display will appear as per the example diagram below.
2. Pressing the blue ‘Mode’ button will enable the
operator to toggle between the following two Display options.
Batch
Total
3. The ’Batch’ total can be reset by pressing the red
Reset’ button.
Please note:
This function resets the ‘Batch’ only. The ‘Total’ displayed is unable to be reset.
4. A ‘Sleep’ mode has been incorporated in the
meter to prolong battery life. The unit will activate sleep mode after 30 secs without use.
2. Pressing the blue ‘Mode’ button will cycle through options available
dEC .1
= 1 Decimal Place
dEC .22
= 2 Decimal Places
dEC .333
= 3 Decimal Places
3. To move to the next section (Unit) press the red ‘Reset’ button
Setting the Units for both Batch & Total.
1. The LCD will now display ‘
UNIT
’.
See ‘Operational Overview’.
2. Pressing the blue ‘Mode’ button will cycle through the options of units that can be displayed for Batch.
L
GAL
Qt
Pt
Oz
dL
3. Next press the red ‘Reset’ button to move onto setting the ‘Total’ units. The availible unit options are as shown above.
4. Once the required ‘Units’ have been selected move to the next section (Calibration) by pressing the Red ‘Reset’ button.
Programming Instructions
Note
Any changes made during the programming phase will automatically be ‘Saved’ when the unit is returned to the operation mode.
Accessing ‘Programming’ Menu
To enter in the programming ‘Menu’, press the reset button for 5 secs.
Once in the programming menu the operator will be able to access (and adjust) 3 programming selections.
1- Setting Decimal Place 2- Display Units of measurement 3 - Calibration Mode
Setting the decimal place.
1.The unit will display the mode and the number of Decimal Places currently set.
E.g. ‘
dEC .22
Calibration.
The calibration mode enables, in the case the operator suspects the accuracy of the meter is in question, the operator to dispense a known volume of fluid through the meter (Test Volume) This Test Volume is compared to the volume measured by the meter (Measured Volume). The meter will perform an ‘Auto Calibration’ if applicable.
1. The unit will display ‘
CALIBRATE
’ in the lower left hand corner, and a number on the main display.
The following options can be scrolled through by pressing the blue Mode button
2
4
8
20
100
250
This number represents the ‘Test Volume’ to be dispensed through the meter during Calibration.
IM213D-MC Issue 10 © 2017 Page 3 of 8
2. On selecting the ‘Test Volume’ press the blue Mode button for 3 secs. The meter will display ‘PURGE’ and ‘CALIBRATE’ will also start to flash.
3. Purge the system of air by running fluid through the system.
4. Once purged of air the calibration process can be started by pressing the blue Mode button.
The unit will display
RUN
and the ’Test Volume’.
E.g.
RUN 100
5. Run the Test Volume through the meter until stipulated volume has been reached (e.g. 100).
6. Once this volume has been reached press the blue Mode button to stop the test. The unit will now compare the ‘Measured Volume’ to the ‘Test Volume’ and perform an ‘Auto Calibration’ if the difference between the two volumes are within ± 8% of each other.
Note: If the difference between the two volumes is greater than ± 8% of each other, the unit will display one of the following messages..
ERROR LOW
ERROR HIGH
if these messages are displayed please contact your Macnaught agent for advice.
Returning to ‘Operation’ Mode At any stage the unit can be returned to the ‘Operation’ mode by pressing the red ‘Reset’ button for approx. 3 secs. Any program changes will automatically be saved.
Maintenance Procedures.
4. Remove the rotors (item 6) and inspect the condition of each. Also investigate if there is the presence of any foreign material in the meter body, that may inhibit the rotors performance
Reassembly
1. Please note, the design of the rotor and shaft assembly ensures that the rotors can only be re-installed with the correct orientation. (i.e. with the magnets being in close proximity to the Electronic module). When replacing the rotors the top face of the rotors should be flush with the sealing face of the meter body. If they sit higher than the sealing face remove, turn over and replace.
2. Replace the rotors (Item 6) onto the shafts at 90 degrees to each other (as per diagram below) and check their operation by turning either of the rotors. If the rotors are not ‘in mesh’ correctly or do not move freely, remove one of the rotors and replace correctly at 90 degrees to the other rotor.
Programming Instructions
Disassembly
Ensure that the fluid supply to the meter is disconnected, and the line pressure is released before disassembly, with the exception for repair or maintenance to the LCD or PCB where it is not necessity to isolate the meter from flow. Refer to the exploded parts diagram on subsequent pages for item numbers.
1. Pull off protective boot (item 1) and unscrew the four retaining screws (item 2) next remove the Electronic Module (item 3)
2. Check for evidence of moisture into the electronic housing. If there is evidence of this, check the condition of the O-Ring (item 4)
3. Check the Rotors (item 6) rotate freely.
4. Replace the 0-Ring (item 7) into the groove of the Meter Cap (item 8).
5. Replace the Meter Cap onto Meter Body (item 5) Tighten Meter Cap screws (item 9) in a diagonal sequence E.g. 1, 5, 3, 7, 4, 8, 6, 2
6. Place the O-Ring (item 4) into the Electronic Module (item 3) and mount the Electronic module onto to the Meter Body
7. Replace and tighten the Retaining Screws (item 2) in diagonal sequence.
8. Align and push on the protective boot (item 1) onto the electronic module (item 3).
9. Before returning to service test the meter by turning the Rotors with your finger. Or applying a very low air pressure (no more than a good breath) to the meter
3. To access the Rotor assembly, remove the 8 Meter Cap screws (item 9)
Loading...
+ 5 hidden pages