Macnaught 050 User Manual

Page 1 of 14
INSTRUCTION MANUAL
To the Owner
OVAL GEAR FLOWMETER
MECHANICAL MODEL 050 (2”)
PLEASE READ THIS SAFTEY INFORMATION CAREFULLY BEFORE USE.
Read and retain this instruction manual to assist you in the operation and maintenance of this product.
If you have any problems with the meter, refer to the maintenance and trouble shooting sections of this manual.
This manual contains connection and operating instructions for meters with Mechanical Displays
If you need further assistance, contact your local representative or distributor for advice.
This Flow Meter has incorporated the oval rotor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow.
Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flow rates are features of the oval rotor design. With a low pressure drop and high pressure rating oval rotor flow meters are suitable for both gravity and pump (in line) applications.
050 Mech
07.11
Page 2 of 14
CAUTION
Important Information
Operating Principle
Installation Procedure
WARNING
Before use, confirm the fluid to be used is compatible with the meter. Refer to Industry fluid compatibility charts or consult your local representative for advice.
To prevent damage from dirt or foreign matter it is recommended that a Y or Basket type 60 mesh strainer be installed as close as possible to the inlet side of the meter. Contact your local representative for advice.
When a strainer is installed it should be regularly inspected and cleaned. Failure to keep the strainer clean will dramatically effect flow meter performance.
To prevent damage caused by air purge slowly fill the meter with fluid. To reduce pressure build up turn off the pump at the end of each day.
Maintenance can be carried out to the liquid crystal displays and pulse units without removing or isolating the meter from the line. When maintenance to any other part of the meter is required, the meter must be isolated and the line pressure reduced.
The reed switch pulse unit can cause inaccurate rate counts when used with high speed counters. It is advised that a debounce circuit be used. Contact
CAUTION
CAUTION
When fluid passes through the meter the rotors turn, as shown below. The magnets which are located in the rotors will pass across the pulser circuit board (containing either Reed switches or Hall Effect sensors). A signal is generated which is then sent by the Pulse Circuit Board (PCB) to the relevant LC display or receiving instrument..
1) It is recommended that when setting up pipe work for meter installations a bypass line be included in the design. This provides the facility for a meter to be removed for maintenance without interrupting production. (see figure above)
2) Use thread sealant on all pipe threads.
3) For pump applications ensure pipe work has the appropriate working pressure rating to match the pressure output of the pump. See Meter Specifications section for further details.
4) Install a wire mesh strainer, Y or basket type 60 mesh (250 micron), as close as possible to the inlet side of the meter.
5) Ensure that the meter is installed so that the flow of the liquid is in the direction of the arrows embossed on the meter body.
6) The meter can be installed in any orientation as long as the meter shafts are in a horizontal plane. (Refer to figures to the right for correct installation) The register assembly may be orientated to suit the individual. Note: Incorrect installation can cause premature wear of meter components.
7) Do not over tighten meter connections. Note: Incorrect installation can cause premature wear of meter components.
8) It is important that after initial installation you fill the line slowly, high speed air purge could cause damage to the rotors.
9) Test the system for leaks.
10) Check the strainer for swarf or foreign material, after the first 200 litres check periodically, particularly if the flow rate decreases.
Page 3 of 14
Maintenance Procedures
Disassembly - Flow Meter
Ensure that the fluid supply to the meter is disconnected, and the line pressure is released before disassembly. Refer to the exploded parts diagram and parts list, for item numbers.
1) Remove the four screws located on the face of the register. Then remove the face plate cover including register assembly.
2) Remove the four register mounting screws and remove the lower half of the register housing.
3) Remove the six cover plate screws (Item 12) and remove the cover plate (Item 11).
4) Remove the six meter cap screws (Item 5) and remove the meter cap (Item 4).
5) Remove rotors (Item 3).
Reassembly - Flow Meter
1) Before reassembling check the condition of the rotors (replace if necessary).
2) Check that the smooth side of the rotors (not the plug side) is facing you when inserting the rotors, the smooth side of the rotor is the magnet side. There is no difference between rotor one or rotor two.
3) Replace the rotors (Item 3) onto the shafts at 90 degrees to each other (refer Fig) and check their operation by turning either of the rotors. If the rotors are not in mesh correctly or do not move freely, remove one of the rotors and replace correctly at 90 degrees to the other rotor.
4) Re-check the operation of the rotors
5) Inspect the gears (Item 6) in the meter cap (Item
4) for wear. (Replace if required, refer to parts on page 4)
6) Replace the o’ring (Item 2) into groove In the
meter cap, if the o’ring has grown or is damaged in any way replace it with a new part.
7) Replace the meter cap making sure that the gear on the rotor is meshing correctly with the gear in the meter cap (Item 4). Insert the cap head screws (Item 5) and tighten in a diagonal sequence 1, 4, 2, 5, 3, 6.
8) Replace the cover plate inspect the o-ring (item 10), bevel gear (Item 7), for wear or damage. (Replace if necessary).
Page 4 of 14
Product Specifications
Problem Cause Remedy
Fluid will not flow through meter
a) Foreign matter blocking rotors b) Line strainer blocked c) Damaged rotors d) Meter connections over tightened e) Fluid is too viscous
a) Dismantle meter, clean rotors (strainer must be fitted in line) b) Clean strainer c) Replace rotors (Strainer must be fitted in line) d) Re-adjust connections e) See specifications for maximum viscosity
Reduced flow through meter
a) Strainer is partially blocked b) Fluid is too viscous
a) Clean strainer b) See specifications for maximum viscosity
Meter reading inaccurate
a) Fluid flow rate is too high or too low b) Fluid is too viscous c) Excess wear caused by incorrect installation
a) See specifications for minimum and maximum flow rates b) Bleed air from system c) Check meter body and rotors. Replace as required. Refer to installation instructions
Fluid flows but no reading on meter
a) Bevel gear is loose on shaft b) Rotor drive gear is damaged c) Transmission gears damaged d) Register gears damaged
a) Tighten grub screws b) Replace rotor c) Replace gears d) Replace register assembly
Fluid leaks into register
a) Seal worn or damaged on the cover plate
a) Replace seal (Check seal compatibility with fluid)
Troubleshooting Guide
Flowmeter Mech
Metric US
F’ and ‘M’ Meters
Flow Range
Below 5 cP 15 to 500 L/min 4 to 130 G/min
5 to 1000 cP 15 to 500 L/min 4 to 130 G/min
S’ Meters
Flow Range
Below 5 cP 30 to 300 L/min 8 to 80 G/min
5 to 1000 cP 15 to 350 L/min 4 to 93 G/min
Maximum Operating Temperature (Standard / High Temperature) 80°C / 120°C 176°F / 248°F
Maximum Operating Pressure
1
3400 kPa 500 psi
Accuracy of Reading ±0.5%
1. Conforms to Directive 97/23/ECCat 1
Page 5 of 14
Exploded Diagram
Parts Identification
Item No. Part Description
1
Meter Body
2
Meter Cap 0-Ring
3
Rotor Set
4
Adaptor Bolts
5
Flanged Process Connection
6
Adaptor O-Rings
7
Threaded Process Connection
8
Meter Cap
9
Gear Set
10
Meter Cap Screws
11
Cover Plate O-Ring
12
Cover Plate and Gear Set
13
Cover Plate Screws
Loading...
+ 9 hidden pages