M2 PZ-4 User Manual

STEREO MICROSCOPE
User Manual
This manual is written for stereomicroscope.To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, it is recommended strongly that you study this manual thoroughly before using the microscope.
1
CONTENTS
USER NOTICE .............................................................................. 2
1. Components ................................................................................ 4
2. Assembly ..................................................................................... 5
2.1 Assembly diagram .............................................................. 5
2.2 Detailed Assembly Procedure ........................................... 6
3. Adjustment and Operation ....................................................... 9
3.1 Adjusting binocular head ................................................... 9
3.2 Focus adjustment .............................................................. 10
4Specification ........................................................................... 11
5. Troubleshooting ........................................................................ 11
6. Outfit ......................................................................................... 12
2
USER NOTICE
1. Safety Precaution
1. Carefully open the box, avoid the accessories, like lens, dropping to ground and being damaged.
2. Check the input voltage, be sure the input voltage which signed on the microscope is consistent with the power supply voltage, or it will bring a serious damage to the instrument.
3. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment. Make sure the stage is smooth, horizontal and firm enough.
4. To prevent the microscope body from turning over, its pivot angle must be limited to 30°.
5. .A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
6. Make sure the instrument is earthed, to avoid lighting strike.
7. Do not disassemble any part of the microscope, especially the electrical parts on the bottom side as this may result in electric shock.
8. Use the factory supplied power cord, please.
2. Maintenance and Storage
1. All the lenses have been well checked and adjusted. It is forbidden to disassemble them
yourself.
2. The binocular viewing head and focusing unit have a compact and precise frame; please
dont disassemble them as possibly as you can.
3. Keep the instrument clean, wipe dust regularly, and be attention to avoid contaminating the
optical elements especially.
4. The contaminations on the prism, as fingerprint and oil, could be gently wiped with a piece
of soft cloth or tissue paper, gauze which has been immersed in pure alcohol or ether.
(Note that the alcohol and ether are highly flammable, do keep them away from the fire or potential sources of electrical sparks, and use them in a drafty room as possible as you can.)
5. Do not attempt to use organic solvents to clean the microscope components other than the
glass components. To clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3
6. If the microscope is splashed by liquid during using, cut off the power at once, and wipe up
the moisture.
7. Do not disassemble any parts of the microscope, which will affect the function or decline
the performance of the microscope.
8. Place the instrument in a cool, dry position. When not using the microscope, keep it
covered with a dust cover.
4
1. Components
Stereo Microscope
Stereo Microscope
Adjustable WF10×eyepieces
Focusing Assembly
Binocular viewing head
Zoom body
Clamping Knob
Objectives
Glass stage
Standard base
Brightness adjusting knob
Power cord
Main Switch
Coarse focusing knob
Fine focusing knob
5
2. Assembly
2.1 Assembly diagram
The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly. When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces.
Power cord
Microscope main body
Zoom body
WF10× Diopter adjustable eyepiece
Binocular viewing head
Clamping knob
Clamping knob
6
2.2 Detailed Assembly Procedure
2.2.1 Mounting binocular viewing head (Fig.1 Fig.2)
1. Rotate the objective into the objective mount thread on the lower end of the zoom body. , as in fig.2.
2. Aligning the positioning pin of the zoom body③ to the positioning groove on the binocular viewing head , insert the dovetail mount at the bottom of the viewing head into the mounting port on the zoom body , and tighten the viewing head clamping screw with provided hexagonal driver.
Note:
Working Environment Requirement 1Ambient temperature: 5°C to 40°C . (41°F to 104°F)Maximum R.H85% 2.High Temperature will result in a mildewing, dew and even ruinous instrument. 3Avoid placing the instrument in a dusty environment. When ending your microscope operation, please cover it with the dust cover. 4Lay the microscope in a plan and stable position, please.
Fig.3
Fig.1
Fig.2
Positioning Pin
Mounting Port
,
7
2.2.2 Mounting binocular head (Fig.3)
Loose the clamping bolt on the pillar
and insert the binocular head and zoom body assemblyinto the bracket of the pillar, then screw down the bolt.
2.2.3 Mounting eyepieces (Fig.3 Fig.4 Fig.5)
1. Remove the eyepiece dust caps and gently insert two adjustable eyepieces into the eyepiece sleeves until they stops, as in Fig.4.
2. Using the Eye Shades
When Wearing Eyeglasses:
Use with eye shades in their normal folded-down position. This will prevent eyeglasses from being scratched by the eyepiece as in Fig 5.
When Not Wearing Eyeglasses:
Extend the folded eyeshades. This makes observation easier by preventing the inverse incidence of light from between the eyepiece and your eyes.
2.2.4Connecting the power cord (Fig.6)
1. Turn the main switch on “O” (off) state
before connecting the power cord.
2. Insert the power plug into the power jack
of the microscope; make sure the connection is well.
3. Plug the power cord into the power
supply receptacle safely. Make sure the connection is well.
2.2.5 Mounting glass stage plate (Fig.7)
If necessary, the original plastic stage
plate can be replaced by a glass one.
Both top and bottom illuminators are LEDS.
● Permitted power supply voltage:
100V240V
Fig.6
Fig.4
Fig.7
Fig.5
,
,
8
2.2.6 Mounting aperture iris
diaphragm attachment
(Fig.8 Fig.9)
Aperture iris diaphragm attachment
is optional.
1. Use the provided hexagonal driver
to loose the clamping bolt and
remove the binocular viewing head
.
2. Mounting the aperture iris
diaphragm attachment to the top end of the zoom body.
3. Mounting the binocular head into the aperture iris diaphragm attachment and tighten the clamping bolt.
Adjusting aperture iris diaphragm
The aperture iris diaphragm adjustment can increase the image contrast and focus depth.
The size of aperture iris diaphragm can be adjusted. Switch the lever to left to open the aperture
iris diaphragm; to the right, to close the aperture iris diaphragm. Check the effect
during
observation.
Do not close the aperture iris diaphragm too much, or it will decrease the resolution and lead
to
lack of illumination.
Closing the aperture iris diaphragm to the minimum state will result in faintness in field of view.
In this case, adjust the aperture iris diaphragm to the half open state.
Closing the aperture iris diaphragm to the minimum state and increase the magnification of the
zoom body will result in lack of illumination around circumference of field of view.
Fig.8
Fig.9
Aperture iris diaphragm lever
,
9
3. Adjustment and Operation
3.1 Adjusting binocular viewing head
3.1.1Adjusting interpupillary distance (Fig.10)
Different users have different interpupillary distances. So when users change, do the interpupillary distance adjustment.
While looking through the eyepieces, hold the left and right of the binocular assembly①and adjust the eyepieces by opening or closing them for binocular vision until the left and right fields of view coincide completely.
3.1.2 Adjusting diopter ring (Fig.11 Fig.12)
1. Set the dioptor rings of both eyepieces to 0position.
(Do this when users change, because different users
have different diopters.) 2Place an easy-to-observe specimen on the stage plate. 3 Rotate the zoom body knob to the highest
magnification 5×(or 6.4× ,or 10×), and turn the focusing knob to focus the specimen.
4 Rotate the zoom body knob to the lowest
magnification 0.8 × , looking only into the left
eyepiece, adjust the diopter ring on left eyepiece to
focus the specimen. Then do the same thing for the
right one. 5Repeat item 3 and 4 until the specimen image can
always focus without any adverse effect of the changes of magnification on the definition of the image.
Fig.12 shows the focusing part of Stereo Microscope .There are coarse focusing knob and fine focusing knob in this kind of microscope.
The working distance of the objectives listed in
standard outfit is all 78mm for these three kinds of stereo microscopes.
When 2× objective is chosen , it is necessary to
mount the binocular head bracket onto the lower
position. Remove the clamping bolt with provided
hexagonal driver and move the bracket down to the
lower position and set it with original bolt (Fig.11).
Fig.10
Fig.12
Fig.11
,
10
3. 2 Focus adjustment
3.2.1 Adjusting the rotation tension of the focus
adjustment knob (fig.13 fig.14)
For these Microscopes ,the tension has been
adjusted well before delivery.
1Rotate the focus adjustment knob to move
the slider to a lower place for fear that the zoom body drops down suddenly if the tension is decreased too much.
2. Hold the focus adjustment knobs① with both
hands, stop the left knob and rotate the right knob to increase or decrease the focus knob tension.
3. Brightness adjustment: the right brightness adjustment knob is used to adjust the bottom illumination, and the left brightness adjustment knob is used to adjust the top illumination
3.2.2 Adjusting tension collarFig.15
1. If the microscope body falls down by its own
weight or the focus obtained by fine focusing is lost immediately, the rotation tension adjustment may be too light. In this case, rotate the tension collar with provided spanner to increase the rotation tension..
2. The brightness adjustment method is the same
as that of fig14
Spanner
Fig.15
7
Fig.13
Fig.14
,
11
4Specification
5. Troubleshooting
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local franchiser for assistance.
PROBLEM CAUSE SOLUTION PAGE
1 Incomplete binocular vision.
Interpupillary distance is not correctly adjusted.
Adjust it correctly
7
Diopter adjustment is incomplete.
Complete diopter adjustment
2Dirt is visible in field of
Dirt on specimen Clean specimen 2
Model NO.
Parameters Zoom Ratio
16 18 110
Objective Magnification
0.8×~5× 0.8×~6.4× 0.8×~8×
Eyepiece
Φ 24 10×WF Diopter adjustable
Binocular Viewing Head
Inclined at 20°, interpupillary distance range: 5075mm
Working Distance 78mm
Vertical Movement Range 105mm
Illumination System Adjustable illumination for both LEDs (top and bottom)
Input Rating
100240V 0.15A50/60HZ
SAFETY PRECAUTION
1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection.
2. Moving this microscope is accompanied with the risk of dropping the specimen and causing infections. Be sure to remove the specimen before moving this product.
3. Move the microscope with both hands; one hand catches the arm, and the other hold the base.
PZ401101 PZ401102
PZ401103
12
view
Dirt on eyepiece Clean eyepiece
3 The visibility of the image is poor.
Dust on objective front lens.
Clean lens surface
6. Specimen image blurs when zoom magnification is changed.
Diopter ring is not correctly adjusted.
Adjust it correctly.
7
Not in complete focus on specimen.
Focus specimen correctly at a high
magnification. 5Coarse focus adjustment knobs rotate with too much resistance.
Tension adjustment ring is too tight.
Loosen it properly. 8
6Zoom microscope body drops or specimen goes out of focus during observation.
Tension adjustment ring is too loose.
Tighten it properly. 8
6. Outfit
Components Specification Qauntity
Standard Outfit
Main body Main body 1
Observation system Binocular head and zoom body 1
Illuminator
Transmitted illuminator (LED) 1
Reflected illuminator (LED) 1
Annular illuminator
Eyepiece
10×Eyepiece
2
20×Eyepiece
Objectives
0.5×objective
1×objective
1
2×objective
Aperture iris
diaphragm
Stage plate
Black-and- white stage 1
Glass plate 1
Power cord Power cord 1
○:Optional