LW Scientific Dry Bath User Manual

USA Dry Bath Incubator
Instruction Manual
Designed, Engineered &
Assembled Proudly in USA
Introduction
The LW Scientific Dry Bath Incubator is designed to be used for a broad range of applications. Stable temperatures and even heating make this two block unit perfect for use in clinical, biological, and even industrial lab settings. The 12v electronics make it portable when the optional battery cords are purchased – perfect for use by medical and veteri­nary professionals in the field. Economical and dependable, the LW Scientific Dry Bath Incubator offers users the choice to mix and match two popular standard heat blocks, while numerous other block configurations are available. Plus, the LWS Dry Bath Incubator will accommodate many other common brands of heat blocks.
Do not operate unit in or near water or other liquids.
Notes for Safe Use
Note: All hard surfaces of the dry bath incubator may be cleaned with common laboratory disinfectants. Only the heat blocks are autoclavable.
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046
•Unplug unit before cleaning or repairing.
Do not use if electrical cord is damaged.
Unpacking & Set Up
Carefully unpack the dry bath incubator and included power adapter, saving packing materials in the unlikely event
1
you need to return the item. Immediately contact your distributor if you detect any damage. If you need to return the dry bath to LW Scientific, please use the original packing material.
2
Place dry bath on a flat, stable surface away from the edges. Ensure that the controls are set in the “OFF” position.
Connect the included power adapter to the dry bath, attach the included line cord, and then plug it into a properly
3
grounded electrical outlet. (Use only the supplied power adapter and plug.)
4
Congratulations! You are now ready to use your LW Scientific Dry Bath Incubator.
The dry bath includes 1 lift rod to make it easier to change the blocks.
5
Caution: As with any temperature controlled equipment, take care not to touch the blocks while the equipment is in
use. Hot surfaces, especially on the block, can cause serious injury or burns.
Do not put water or liquids into the dry bath well.
LW Scientific is not responsible for any injury caused by misuse or abuse of this equipment.
PID Controller
All programming functions of the PID controller have been preset by LW Scientific, and the temperature has been preset to 98.6°F (37°C). There are many complicated programming functions built into the PID controller, which should not be altered by the customer. The only two functions that may need to be adjusted by the customer are “Set Tem­perature” and “Change from Fahrenheit to Celsius” as described below. For any questions about altering other programming, please contact the LW Scientific technical support department at 800-726-7345 (770-270-1394).
AL- Relay output indicator
1 2
Value increment/Select next parameter
3
Value decrement /Select previous parameter
4
Auto tuning /Digit shift
5
Set/Confirm
6
Will not illuminate
The temperature of the blocks (Process Value, or PV) is displayed during operation. The target temperature (Set
7
value or SV) is displayed when flashing by pressing the V or key once.
V
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046