Safety Precautions: Read Operating Instructions Thoroughly Prior to Operation
Read Operating Instructions thoroughly prior to operation. Use only a grounded outlet that is rated for your
model's electrical requirement. Do not modify the oven or factory control settings to operate the oven above the
stated maximum operating temperature. Exterior surfaces on the 180E models may become hot to the touch
when operating at higher set temperatures. Conduct periodic maintenance as required.
Set-up & Installation
Position unit in its ultimate operating location. Keep a minimum of
3" of airspace around the unit and a minimum of 6" above the unit.
The port hole at the top of the unit will expel a small amount of
warm air through natural convection. This port can also be used as
an access way for external temperature measurement of a solution
for example.
Install adjustable shelf by placing the ends of the wire shelf bracket
into the corresponding holes located on the inner sides of the oven
at the desired height. Push the ends of the bracket into the holes
until the first bends in the bracket are against the wall, then rotate
the bracket down. Place the shelf on the brackets. (FIG 1)
Plug the unit into a grounded outlet for your unit's rated voltage.
FIG. 1
AB
CD
General Operation
The unit is ready for your immediate use. All control parameters,
calibration and tuning has been done at the factory, no adjustments
are necessary.
Push the illuminated power button. All LED's on the temperature
control will light up for 5 seconds until the current or actual chamber
temperature is displayed.
To view the set temperature press the star " " key. To change the
set temperature, hold the star key together with the up (raise temp)
or down (lower temp) arrow key until the desired temperature is
indicated on the LED display. (FIG 2)
set at the factory to read in 1/10 degree C (centigrade) units. To
change temperature units or display resolution see: Menu Level
Functions (page 3).
Once the unit reaches the set temperature, allow the unit to cycle
for 20 minutes at set point before temperature becomes fully stable.
NOTE: Upon each initial powering-up, the control may typically overshoot the set temp by 2 to 4 degrees,
especially if the temperature set point is close to the operating ambient temperature. After equilibrium is achieved
the control will hold set temperature within 1 unit degree.
*
The temperature control is
(PRESS)
FIG. 2
*
View Setpoint in oC / oF units
*
Increase Setpoint
*
Decrease Setpoint
*
(3 sec.) Enter/Exit Menu Levels
Chamber Loading
Article processing times and temperature uniformity are largely dependent on load density and positioning. Load
the incubator so that the air circulation within the incubator is not impaired. Here are some general guidelines:
Leave a space between articles on a shelf. Stagger articles from those on lower shelves.
Avoid placing articles or media against or within an inch of the walls, especially on the lower shelf. Heated air from
the lower plenum openings, designed to travel up the side walls, can have a slightly elevated temperature from set
point and the rest of the chamber.
Use of large solid trays or foil on shelves limits heat to any articles placed on shelves above.
Avoid extremely large (in quantity or size), or high-density loads. This will show by non-uniform processing and
long or impossible "heat-through" times. To help determine a load's suitability, use the set-point recovery time (the
time it takes for the temperature to recover to the original set temperature once the load is placed), as a guide. To
reduce recovery time, reduce load proportionally. When possible, measure large loads or solution temperatures
directly with an ancillary thermometer or probe. Probes can be inserted at the top port.
best processing performance for a single item, adjust one shelf so that the article is centered in the incubator
For
chamber.
PAGE 2
Page 3
)
Control Menu Functions
Access the menu levels for the following functions: (user applicable functions and
FUNCTION SETTING
their menu locations are high-lighted in white in the Functions Menu Guide below).
~ Change control to read in C or F temperature units.( in level 2)
~ Change to whole degree or 1/10th degree display resolution.( in level 2)
~ Run or Read temperature tracking.( or in level 3)
~ Lock Temperature setting against inadvertent adjustment.( in level 1)
~ Calibrate control temperature to an external standard.( in level 3)
FUNCTION PROMPT
FIG. 3
To access the control's function menu:
Press and hold both arrow keys for 3 seconds then release when the "tune" function prompt is displayed from
within LEVEL 1 (see "menu entry point" at the bottom of the Menu Guide below). When in the function menu
the LED display will alternate the function prompt with the function setting when keys are released. (FIG 3)
To navigate within the functions menu:
Use up and down arrow keys individually to move "right" or "left" within a level. Press and hold the star key
and then the up or down arrow keys to move "up" or "down" respectively through levels 1, 2 and 3 (Note: you
must be at prompt to move up or down levels).
To change a function setting:
Once at the desired level function prompt, press and hold the star key and press the up or down key to select
or change the function setting. Release star key to set the function. Press the up and down keys together to
return to temperature display or the control will auto-return in 60 seconds.
Functions Menu
Guide
(-10 to +10)
(oFF)
Alternating
Display
LEVEL 3
(
*
PRESS
LEVEL 2
*
(
PRESS
LEVEL 1
ON / OFF
VAR / HI / LOW
Control Calibration Adjust
Temperature Tracking Monitor
(
)
PRESS
)
Select Display Resolution
(
)
PRESS
)
Set Temperature Lock
(
)
PRESS
Read Tracking Monitor Results
1 / 0.1
OFF/ON
o
Select
NONE
o
F
o
C /
o
F, Bar, PSi, Ph
C,
,
(C)
MENU ENTRY
POINT
IMPORTANT NOTE: The grey-shaded control functions within the menu levels are factory
tuned and set for the optimum performance and the safe operation for your model
incubator. Care should be taken when navigating or changing user applicablesettings
so as not to alter these factory settings. Factory settings for each function are listed within
the parentheses "( )" on each function tab above. Use as a trouble-shooting reference.
PAG E 3
Page 4
Temperature Tracking Feature
This feature monitors the stability of the control during for any given length process. It will record and display: 1.
The total variation or spread between high and low, 2. The absolute or maximum high and 3. The absolute or
minimum low.
To start the tracking feature navigate to "CheK" prompt in menu level 3. Hold star key then up arrow key to select
ON. Return to temp display or control will auto-return in 60 seconds. The control will track the temperature
variation until "CheK" is turned off. Recorded readings are retained until next "CheK" ON.
You can view readings at any time during or after tracking feature has been turned off. But de-powering the unit
will reset "CheK" to OFF and "rEAd" to zero. To view readings navigate to "rEAd" prompt in menu level 3. Then:
Release orPress/Hold
Displays variance
(0.6o)
TIP: To avoid erroneous tracking data from run-up temperatures or door openings, start the Tracking feature after
articles have been placed and temperature steadies at set point.
*
Press/Hold
Press once
Displays maximum
*
Press/Hold
Press once more
Displays minimum
*
Control Self Diagnostics
Control prompts will only display when a fault or alarm condition exists.
until manually reset. Typically indicates
relay or temperature control failure.
Replace relay or temp control if persists.
Example for 140 Series
PRESS
(Press together to
clear alarm condition)
Maintenance / Control Calibration
To clean interior and exterior surfaces, use a damp cloth with or without an all-purpose cleaner. The acrylic door
should only be cleaned using a lint-free cloth, with or without water. Paper towels can mar the surface of the
acrylic door. Use of any commercial cleansers on the acrylic door will cause crazing and cracking of the surface of
the acrylic over time. Periodically, check the accuracy of the control's temperature display against a known
accurate or calibrated device. This should be done with an empty chamber after the set temperature becomes
steady (typically after 45 to 60 minutes). Calibrate the control in the control's functions menu, level 3 (see page 3).
Tech Support
If you have any questions or need technical assistance, please contact LW Scientific.
Warranty
Please read and complete the warranty form online at LWScientific.com/warranty_form. The warranty form
documents your purchase. Failure to fill out the warranty form may void any warranty claims on the unit.
MKT-7.5.3-L-134 | Rev 0
PAGE 4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.