2.1 kg / 3.1kg 4.6lb/ 6.8lb
Standard carton: 515x415x120 (mm3)
5 Units in one box: 640×430×530 (mm3)
SPECIFICATIONS
MSS
INSTRUCTION MANUAL
MSS M-0612

HOLD FUNCTION
1) Place the item on the Scale platform.
2) Wait for the stable weight to be displayed.
3) Press the HOLD key. The "." hold indicator will turn on.
4) Remove the item from the Scale platform. He item's weight reading will
remain on the display. Now you can also select Unit ( Pounds or Kilograms) by
pressing [UNIT] key.
5) Quit HOLD function: Once an item is weighed and the hold function is
enabled, you want to quit HOLD function, press the HOLD key again to cancel
the "." hold indicator and return the Scale to weighing mode.
FUNCTION SETTINGS
1. Press and hold [HOLD] key then press [ON/OFF] key to turn on the scale, the
display will show A-ON or A-OFF, press [ZERO] key to select on or off (ON
means auto power off function activate, OFF means auto power off function
inactivate).
2. Press [HOLD] key, the display will show L-ON or L-OFF, press [ZERO] key to
select on or off (ON means auto backlight function activate, OFF means auto
backlight function inactivate).
3. Press [HOLD] key, the display will show H_ON or H_OFF, press [ZERO],
selection display H_ON, this means AUTO-HOLD function active, press [ZERO]
again, the display will show H_OFF, this means AUTO-HOLD function inactive.
4. Return to weighing mode.
Press [ON/OFF] to turn the power off, press [ON/OFF] again to turn on the
power and the scale return to the weighing mode.
4
PARTS DESCRIPTIONS:
·TOP
·Display
1. ON/OFF:
Turn the scale power on/off.
1. HOLD:
Press this key to store weight value.
2. UNIT:
Select weighing units:kg or lb .
3. ZERO: This key is “Zero” feature if load is below 4% of full capacity
or a negative value.
4. TARE:
between 0% and 100% is tare function
·BOTTOM
Battery Cover
AC Adaptor socket
1

When to calibrate
Note: Calibration is done in factory. Don't re-calibrate the Scale, unless it
is not accurate.
Calibration:
1. Turn the scale off.
2. Press and hold [ZERO] key then press [ON/OFF] key to turn on the
scale,, the display will show “X00.00”,
3. X is the flashing digit.
4. press [UNIT] key to select calibrationd unit kg or lb
5. Press [HOLD] key the flash digit will move to right;
6. Press [ZERO] key to increase the flash digit. (X is the calibration weight
which can be set according to your requirement – we recommend a
minimum weight of 50% of full capacity).
7. Press [TARE],the display will show 88888 then the AD value, press
[UNIT] key, the display will show flashing calibration weight, place the
shown weight on the platform, press [UNIT] key when the stable
indicator lights, the display will show ------- and the AD value
8. Now the Calibration is completed.
9. Press [ON/OFF], turn off the power.
10. Press [ON/OFF] turn on the power again.
11. Place the weight on the platform to make sure weighing is correct, if not,
repeat above steps.
2
Features:
1. Display options:5 digital LCD
2. Power:AC adapter 6V 100mA or 4xAA Size Batteries
3. Auto back light (selectable)
4. Auto Shut-off (selectable)
5. Auto Hold (Selectable)
6. Beeper(Non-selectable)
7. Operating Temperature:5~35℃
8. Operating Humidity:25%~95%RH
9. 5Keys: ON/OFF, ZERO, UNITS, TARE, HOLD
ZERO This key is “Zero” feature if load is below 4% of full
capacity or a negative value
UNITS Toggle among KG or LB unit
ON/OFF Turns the scale power on/off
TARE Press this key to tare weight value
HOLD Press this key to store weight value.
10. Zero Range (0 to 4% of full capacity)
11. Tare Range (0 to full capacity)
12. Error Message indication:
Err-0 Overload
Err-5 Unstable when power on
Err-c Calibration code is not enough or operation error.
Low battery indication
13 Two calibration weight units : kg , lb.
14. AnyCal calibration software
3