BW/BWS Series Indicator Technical Manual

CONTENTS
1. SPECIFICATION..…………………………………………………………………. 2
2. INTRODUCTION…………………………………………………………………... 3
3. INSTALLATION……………………………………………………………………. 4
Precautions …………………………………………………………………………4
Parts………………………………………………………………………………… 4
Installation………………………………………………………………………….. 5
4. KEYS DESCRIPTION……………………………………………………………... 6
Key Board…………………………………………………………………………... 6
Secondary functions of the keys………………………………………………….7
Display……………………………………………………………………………….7
5. OPERATION……………………………………………………………………… 8
5.1. Basic Operation…………………………………………………………... ...8
1. Power On /Off………………..………………………………………………8
2. Zero…………………………………………………………………………...8
3. Tare…………………………………………………………………………...8
4. Select Unit……………………………………………………………………8
5.2. Check Weighing……………………………………………………………..9
5.2.1. Set Limits…...……………………………………………..………….9
5.2.2. Set Check Weighing………..……………………………………….9
5.3. Accumulation…………………………………………………………………10
5.3.1. Memory Recall……………...…………………….………………….10
5.3.2. Memory Clear………………………………………………………...10
5.3.3. Automatically Accumulation………………………………………...10
5.4. Parts Counting...……………………………………………………………..11
5.5. Animal Weighing……………………………………………………………..11
5.6. Key Board Lock………………………………………………………………11
5.7. Set Auto Power Off…………………………………………………………..12
5.8. Set Back Light………………………………………………………………..13
6. PARAMETERS…………………………………………………………………….14
7. CALIBRATION……………………………………………………………………..16
8. RS-232 OUTPUT…………………………………………………………………17
8.1. Specifications……………………….………………………………………17
8.2. Connecter……………………………………………………………………17
RS-232 Output………………………………………………………………17
Check weighing Output…………………………………………………….17
8.3. Continues output protocol………………………………………………….17
9. DRAWING…………………………………………………………………………19
9.1. BW Drawing………………..………………………………………………19
9.2. BW Parts List……………………………………………………………….20
9.3. BWS Drawing………………………………………………………………21
9.4. BWS Parts List……………………………………………………………..22

BW/BWS Series Indicator Technical Manual
DISCONNECT ALL POWER TO THIS UNIT
BEFORE INSTALLING, CLEANING, OR
SERVICING. FAILURE TO DO SO COULD
RESULT IN BODILY HARM OR DAMAGE THE
UNIT.
Permit only qualified persons to service the instrument
Before connecting or disconnecting any components, remove the
power.
Failure to observe these precautions bodily harm or damage to or
destruction of the equipment.
PRECAUTIONS
The weighing indicator is a precision electronic instrument,
handle it carefully.
Do not install the scale in direct sunlight.
Verify the local voltage and receptacle type are correct for
the scale.
Only use original adaptor, other could cause damage to the
scale.
Pluggable equipment must be installed near an easily
accessible socket outlet.
Avoid unstable power sources. Do not use near large users
of electricity such as welding equipment or large motors.
Avoid sudden temperature changes, vibration, wind and
water.
Avoid heavy RF noise.
Keep the indicator clean

BW/BWS Series Indicator Technical Manual
-10°C - 40°C / 14°F - 104°F
AC Adaptor (12V/500mA)/ Battery (6V/4Ah)
1. SPECIFICATION

BW/BWS Series Indicator Technical Manual
2. INTRODUCTION
The BW series weighing indicator that amplifies signals from a load cell,
converts it to digital data and displays it as a mass value.
It is suitable for general weighing or more specialized applications such as
check weighing, animal weighing and accumulation applications.
It can connect the indicator to a printer or a PC.
Large LCD with white LED back light displays

BW/BWS Series Indicator Technical Manual
3. INSTALLATION
Unpacking
When you receive the scale, inspect it to make sure that it is not damaged and
that all are parts are included:
Remove the Indicator from the carton.
Remove the protective covering. Store the packaging and to use if you
need to transport the scale later.
Inspect the indicator for damage.
Make sure all components are included.
1. Indicator
2. Adaptor
3. Manual
4. Indicator holder (Optional)
5. Load cell Output connecter (Optional)
6. RS-232 Output Connecter (Optional)
Parts Description

BW/BWS Series Indicator Technical Manual
Installation
Place the Indicator on a table or use indicator holder to connect with stand.
Connect the plat form load cell cable in to the indicator load cell connecter.
Load cell connecter is locating back side of the indicator.
Connect the adaptor pin in to the indicator adaptor jack.
Adaptor jack is locating, back side of the indicator.
Adaptor connects into your AC power socket.
Pluggable equipment must be installed near an easily accessible socket
outlet with a protective ground/ earth contact.
Turn on the On/Off key. If you want to turn off, press the key again.
Display will be show the scale capacity and will be starting self checking.
After self checking, display will be come to normal weighing mode.
Warm-up time of 15 minutes stabilizes the measured values after
switching on.
Calibrate with exact calibration weights, minimum 1/3 of the scale capacity
want to use for calibration. For calibration see details in parameter.
Then you can start your operation

BW/BWS Series Indicator Technical Manual
To perform a tare function, Subtracts weights.
Accumulator key, current values will store to the memory,
To send the data to printer or PC
Shift to Gross / Net Weight.
4. KEYS DESCRIPTION
Key Board