This operating manual contains information and behaviour rules for the
safe and designated operation of the device.
Observe the following principles:
nRead the entire operating manual prior to starting-up the device.
nEnsure that everyone who works with or on the device has read the
operating manual and follows it.
nMaintain the operating manual throughout the service life of the de-
vice.
nPass the operating manual on to any subsequent owner of the device.
1.1 General non-discrimination
In this operating manual, only the male gender is used where grammar
allows gender allocation. The purpose of this is to make the text easy to
read. Men and women are always referred to equally. We would like to
ask female readers for understanding of this text simplification.
1.2 Explanation of the signal words
Different signal words in combination with warning signs are used in this
operating manual. Signal words illustrate the gravity of possible injuries if
the risk is ignored:
Signal wordMeaning
DANGER
WARNING!
CAUTION
Refers to imminent danger. Ignoring this sign may
lead to death or the most serious injuries.
Refers to a potentially hazardous situation.
Failure to follow this instruction may lead to death
or severe injuries.
Refers to a potentially hazardous situation.
Failure to follow this instruction may lead to
minor injury or damage to property.
1.3 Explanation of the warning signs
Warning signs represent the type and source of a danger:
Warning signType of danger
General danger
Danger from electrical voltage
Danger from poisonous substances
Danger of damage to machine or functional
influences
Table 2: Explanation of the warning signs
1.4 Identification of warnings
Warnings are intended to help you recognise risks and avoid negative
consequences.
This is how warnings are identified:
Warning sign
Description of danger.
Consequences if ignored.
The arrow signals a safety precaution to be taken to eliminate the
ð
danger.
SIGNAL WORD
NOTE
Table 1: Explanation of the signal words
Notes for the Reader
4
Identification of action instructions
Refers to a danger which, if ignored, may lead to
risk to the machine and its function.
1.5 Identification of action instructions
This is how pre-conditions for action are identified:
Pre-condition for action which must be met before taking action.
ü
@A resource such as a tool or auxiliary materials required to perform
the operating instructions.
This is how instructions for action are identified:
The following warnings are intended to help you eliminate the dangers
that can arise while handling the device. Risk prevention measures always apply regardless of any specific action.
Safety instructions warning against risks arising from specific activities
or situations can be found in the respective sub-chapters.
DANGER
Operating instructions
Hazards due to non-compliance with the safety instructions
Failure to follow the safety instructions may endanger not only persons,
but also the environment and the device.
The specific consequences can be:
nFailure of major unit und system functions.
nFailure of required maintenance and repair methods.
nDanger for individuals through dangerous dosing media.
nDanger to the environment from overdosing.
Mortal danger from electric shock!
Incorrectly connected, incorrectly placed and damaged cables can
cause injury.
Replace damaged cables immediately.
ð
Do not use extension cables.
ð
Do not bury cables.
ð
Secure cables to avoid being damaged by other equipment.
ð
CAUTION
Increased risk of accidents due to insufficient qualification of personnel!
The device may only be installed, operated and maintained by personnel with sufficient qualifications. Insufficient qualification will increase
the risk of accidents.
Ensure that all action is taken only by personnel with sufficient and
ð
corresponding qualifications.
Prevent access to the system for unauthorised persons.
ð
NOTE
Do not dispose of the device in the domestic waste!
Do not dispose of electric devices via the domestic waste.
The device and its packaging must be disposed of in accordance
ð
with locally-valid laws and regulations.
Dispose of different materials separately and ensure that they are
ð
recycled.
2.2 Working in a safety-conscious manner
Besides the safety instructions specified in this operating manual, further
safety rules apply and must be followed:
nAccident prevention regulations.
nSafety and operating provisions.
nSafety regulations on handling hazardous substances.
nEnvironmental protection provisions.
nApplicable standards and legislation.
2.3 Personnel qualification
Any personnel who work on the device must have appropriate special
knowledge and skills.
Anybody who works on the device must meet the conditions below:
nAttendance at all the training courses offered by the owner.
nPersonal suitability for the respective activity.
nSufficient qualification for the respective activity.
nTraining in how to handle the device.
nKnowledge of safety equipment and the way this equipment functions.
nKnowledge of this operating manual, particularly of safety instructions
and sections relevant for the activity.
nKnowledge of fundamental regulations regarding health and safety
and accident prevention.
All persons must generally have the following minimum qualification:
nTraining as specialists to carry out work on the device unsupervised.
nSufficient training that they can work on the device under the supervi-
These operating instructions differentiate between these user groups:
2.3.1 Specialist staff
Thanks to their professional training, knowledge, experience and knowledge of the relevant specifications, specialist staff are able to perform the
job allocated to them and recognise and/or eliminate any possible dangers by themselves.
2.3.2 Trained electricians
Due to their professional training, knowledge and experience as well as
knowledge of specific standards and provisions, trained electricians are
able to do the electrical work assigned to them and to recognise and
avoid any potential dangers by themselves.
They are specially trained for their specific working environment and are
familiar with relevant standards and provisions.
They must comply with the legally binding regulations on accident prevention.
2.3.3 Trained persons
Trained persons have received training from the operator about the tasks
they are to perform and about the dangers stemming from improper behaviour.
Trained persons have attended all trainings offered by the operator.
Operating instructions
2.3.4 Personnel tasks
In the table below you can check what qualifications are the pre-condition for the respective tasks. Only people with appropriate qualifications
are allowed to perform these tasks!
Any non-designated use of the device can impair its function and the protection provided. This leads to invalidation of any warranty claims!
Please note that liability is on the side of the user in the following cases:
nThe device is operated in a manner which is not consistent with these
operating instructions, particularly safety instructions, handling instructions and the section "Intended Use".
nInformation on usage and environment (section 5 “Technical data” on
page10) is not adhered to.
nIf people operate the device who are not adequately qualified to carry
out their respective activities.
nUnauthorised changes are made to the device.
3.2 Intended purpose
The controller monitors the current measured values during water treatment and controls the dosing systems connected for water treatment. In
this way, the controller ensures constant water values in various applications and can be deployed universally. One of the main applications is the
maintenance of water quality in industry and public swimming pools by
measuring and evaluating the water parameters chlorine value, pH value,
redox value, total chlorine as well as conductivity and the control of chlorine gas dosing systems.
3.2.3 Incorrect operation
nProtective equipment not functioning correctly or dismantled.
nUnauthorised modification of the controller.
nIgnoring of alarm or error messages.
nThe elimination of alarm or error messages by insufficiently-qualified
personnel.
nBridging the external fuse.
nDifficult operation due to insufficient lighting or poor access to the de-
vice.
nOperation not possible due to dirty or illegible display.
3.2.4 Incorrect maintenance
nCarrying out maintenance during ongoing operation.
nNo adequate or regular inspection of correct functioning.
nNo replacement of damaged parts or cables.
nNo securing against reactivation during maintenance work.
nUse of the wrong spare parts.
Foreseeable misuse
The following section provides information regarding the device applications which are classified as non-intended use. This section is intended to
allow you to detect possible misuse in advance and to avoid it.
Foreseeable misuse is assigned to the individual stages of the product
lifetime:
3.2.1 Incorrect assembly
nConnecting the mains voltage without a protective earth.
nNon-fused or non-standard mains voltage.
nNot possible to immediately or easily disconnect the power supply.
nWrong connecting cables for mains voltage.
nSensors and actors connected to the incorrect terminals or incorrectly
configured.
nProtective earth removed.
3.2.2 Incorrect start-up
nCommissioning with damaged or obsolete sensors.
nCommissioning without the establishment of all protective measures,
The stationary device measures the water values using sensors. Controlling actors such as dosing pumps controls the water values to the desired setpoint.
4.2.3 Main view
The main menu view will appear upon the start of the device or 5 minutes
after the last input. The main view shows the current values from up to
four sensors and further information.
1
6
4
Fig. 2: Main view with three sensors
ItemFunction
1Login/password settings
2Date/time
3Measured values
4Main menu
5File Browser
2
3
5
4.3 Rating plate
There is information on the equipment about safety or the product's way
of functioning. The information must stay legible for the duration of the
service life of the product.
Load capacity of the alarm relay250 V AC, 8 A (ohmic load)
Analogue outputs for remote transmission4 x 0/4 − 20 mA, max. 500 Ω
Operating instructions
Disturbance variable input
0/4 − 20 mA
Intervention of the disturbance variable, amplification between 0.1 − 10 times
Ethernet Modbus TCP/IP
Interfaces
RS485 Modbus RTU (optional)
Protection classIP65
Ambient temperature-5 to +45 °C
Air humiditymax. 95 %, non condensing
Control characteristic
Table 7: Technical data
P, PI, PID or PD behaviour, control direction selectable with disturbance variable feed
forward, 2-side control selectable
5.2 Input modules
All input modules consist of an input for temperature measurement via Pt100 and a second input for the measurement of a further water parameter. On
some input modules, this input will measure a number of different parameters.
Sampling systemValue
Temperature input-10 to +90 °C
Measurement input pH Redox
module
pH valuepH 0 – pH 14
Redox0 − 1000 mV
(Free chlorine)
Cl
Measurement input potentiostat
module
2
ClO
(Chlorine dioxide)
2
Excess chlorine measuring cell CS120Measuring range dependant on the slope up to c. 20 mg/l.
Measurement input current
module
Diaphragm-covered measuring cell,
conductivity or another 0/4−20mA
sensor
Voltage current module24 V
Table 8: Technical data
10
Technical data
Input modules
BA-46020-02-V01
Measuring range dependant on the slope up to c. 18 mg/l.
Make sure that the installation location complies with the following requirements:
nThe display is easily accessible and is visible.
nPlan to leave min. 20 cm free space for the installation of the cable
underneath the device. You must be able to install the cable without
kinking or damage.
nVarious lines (e.g. voltage supply, data cable and sensitive lines for
measuring purposes) must be installed separately. The different lines
should only cross at 90° so as to prevent falsifications.
nElectrical, magnetic and electromagnetic fields affect signal transmis-
sion and can destroy electronic components.
nCompliance with the permissible ambient temperatures (section 5
“Technical data” on page10).
7.2 Installation on the wall
Resources required:
@Assembly kit
@Drill
@Slotted screwdriver
Perform the following steps:
1. Drill the four drillholes for wall mounting. The exact dimensions are
stated in section 6 “Dimensions” on page11.
2. Unscrew the screw on the right-hand side of the device and pull out
the rod.
4You can now open the device.
3. Open the device and use the screws for wall mounting. Ensure that
the device is secured to the wall.
4. Close the device again using the rod.
@Wire end sleeves 0.75 − 2.5 mm²
DANGER
Mortal danger from electric shock!
Improper installation or damaged components of the electrical installation may result in injury.
Ensure that work on the electrical installation is only carried out by
ð
a qualified electrician.
Make sure that work on the eelectrical installation is only carried out
ð
when it is voltage-free.
Ensure that power supply is protected by a residual current circuit
ð
breaker.
Replace damaged cables or components immediately..
ð
1. Fit wire end sleeve to the cable ends if the supply cable does not
have them.
2. Open the device housing.
3. Lead the supply cable through a cable screw connection to the un-
derside of the device.
4. Turn the cable screw connection union nut until the line is fixed in the
screw connection so that the screws connection performs strain relief. Ensure that the feed cable is installed loosely.
5. Connect the voltage supply to the clamps 44 − 52. Observe the division into protective earth (PE), neutral conductor (N) and the phase (L)
on the circuit board.
The device is fitted on the wall.
ü
7.3 Electrical installation
The voltage supply to your device can now be performed via a normal
Schuko plug or a control cabinet. Perform the specifications of this section for devices without a pre-fitted Schuko plug.
Precondition for action:
The device was installed in accordance with section 7.2 Installation
ü
on the wall.
A voltage supply with 100 − 240 V AC (50/60 Hz) is available.
ü
The voltage supply is deactivated before the start and secured
ü
against reactivation.
The housing is open.
ü
Resources required:
@Grounding-type plug
Installation
12
Electrical installation
Fig. 6: Connected voltage supply
Electrically installation
ü
BA-46020-02-V01
Only 3 of 9 clamps are required for connection of the voltage
supply. You can use the free clamps to supply further devices
with voltage.
The output connects a servosmotor with or without feedback.
Fig. 16: Output module ATE (20 mA) with feedback
TerminalFunctionDescription
1-GND
2Input (feedback)4 − 20 mA
3Output4 − 20 mA
Table 21: Terminal connection 20 mA
ActorsConfiguration
C7700 chlorine control valve
Servomotor with 20 mA/
continuous output (20 mA)
3. Enter a value between 0 and 100 % and check whether the actor reacts as desired.
Actor has been tested.
ü
Testing the analogue outputs
3
2
1
You can also test the connection of terminals 13 to 20.
Precondition for action:
The device housing cover is closed.
ü
The voltage supply has been established and the device has been
ü
switched on.
Instruction:
1. Working in the main menu, navigate to System > Outputs > Analogue.
4You will now see all analogue outputs (terminals 13 – 20).
2. Press “Test signal”.
3. Set the mA value.
4. Press "Start”.
Analogue outputs tested.
ü
EASYZON chlorine dioxide plant
MAGDOS dosing pumps
MEMDOS dosing pumps
MEMDOS SMART dosing pumps
Table 22: Actors and configuration
7.6.6 Testing the outputs
You can use manual mode to test the correct connection of an actor.
Take the alarm chain into account before conducting the test and inform
any connection points or interrupt the alarm chain for the period of the
test.
Test the connected actors
Precondition for action:
The actors have been connected in accordance with section 7.6
ü
“Connecting actors” on page15.
The device housing cover is closed.
ü
The voltage supply has been established and the device has been
ü
switched on.
Instruction:
1. Working in the main menu, navigate to “Manual mode” (section 9.4
“Manual mode” on page27).
Continuous output (20 mA)
4You will now see all the outputs.
2. Select the output to which you have connected the actor and which
you wish to test.
7.7 Digital inputs
You can use up to 8 digital inputs to evaluate switching statuses and to
detect them as alarm message which are to be documented in the logfiles.
Further information about the settings of the digital inputs can be found in
section 8.2.1.6 “Digital inputs” on page21.
7.8 RC protection for relay
When connecting to the relay, note that inductive loads must be suppressed. If this is not possible, the relay contact on the device terminal
must be protected by an RC protective circuit/interference suppression
element.
If devices with inductive loads from a nominal current of 1 A are connected to a relay, the contacts in the relay may become bonded. Thus, the device will operate in an uncontrolled manner. To prevent bonding if the load
circuit suffers a short-circuit, the relays must be protected separately on
the maximum relay switching current.
Precondition for action:
You would like to connect an inductive load to the relay.
ü
Instruction:
1. Switch off the device.
2. Clamp the interference suppression element parallel to the inductive
load.
3. Should it prove impossible to perform point 2, clamp the interference
suppression element parallel to the relay output.
You can use the Ethernet connection for the following actions:
nReading / writing via Modbus TCP/IP protocol (PLC or Computer)
nAccess via web browser
nAccess via TFTP server
The device is fitted with a network input in the form or a 4-pole and
D-coded M12x1 socket. Lutz-Jesco GmbH offers different lengths of special twisted-pair network cables to make the typical Ethernet RJ-45 plug
connection. If you use third-party cables, choose a Category 3 cable with
an impedance of 100 Ω or above.
PinAssignmentWire colours
1TX-yellow
2TX+orange
3RX-white
4RX+blue
-Screen-
Table 23: Ethernet connection socket
Ethernet
port
Fig. 17: Ethernet socket
Installing a wired network
During installation, comply with the following points:
nThe Ethernet is cabled in a star topology. The maximum cable length
is 100 m.
nOnly use screened cables and connectors.
nOnly use CAT5 cables or better.
7.10 RS485-interface
ONOFF
A B
Fig. 18: Jumper position on RS485
When using multiple devices on a data line, you must activate
a 120 Ω resistance on the last device.
You can activate the resistance by setting the jumper to “ON”
as shown in Fig. 18.
Your device can have an optional RS485-interface. Using a two-paired
data cable you can connect up to 14 devices with a PC or a PLC. Modbus
RTU protocol serves as a protocol for data transfer. You can use the addresses 1 to 14. The addresses 0 and 15 are reserved for internal purposes and are not supported.
RS485 Modbus settings:
nBaud rate: 9600
nWord length: 8 Bit
nStop bit: 1 Bit
nParity: None
nYou can read out a maximum of 40 addresses at once.
The list of Modbus commands can be found in section 12 “Modbus addresses” on page36.
Perform the following steps:
1. Open the device housing.
2. Connect a two-paired data cable to terminals A and B of the RS485
nThe output switches if a value is exceeded.
nHysteresis can be set from 0.1 – 50 %.
nRelay: 10 – 100 pulses per minute
nOptocoupler: 10 – 350 pulses per minute
Pulse frequency or
2 sides pulse frequency
Pulse length or
2 sides pulse length
Servomotor with
potentiometer
Servomotor without
potentiometer
Servomotor with 20 mA
feedback
nThe pulse frequency is dependant on the control deviation and the set control
parameters.
nWith a control output power of Y = 25 % and a maximum pulse frequency of
100pulses/min., the controller would output 25 pulses/min.
n0 – 3600 seconds cycle duration
nRelay output (e.g. for solenoid valve)
nDepending on the control deviation and the defined control parameters, the relay
pulls in or drops out for the set cycle duration. If the cycle lasts 30 seconds and the
controller output power is 40% the relay applies for example for 12 seconds,
followed by 18 seconds of non-application.
nA feedback potentiometer can be connected (1 – 10 kΩ) for servomotors with
position feedback.
nCompensate the feedback potentiometer. During compensation, the servomotor is
first started and then stopped automatically.
nFor servomotors without feedback.
nMeasure and set the runtime of the servomotor.
nContinuous control output from 0/4 – 20 mA for the actuation of constant actors.
nServomotors which are controlled via 4 – 20 mA and have a 4 – 20 mA position
feedback.
Operating instructions
8.1 First steps
You need to make a number of basic settings before operating the device.
This section leads you through initial commissioning.
Precondition for action:
The device has been installed in accordance with section 7 “Installa-
ü
tion” on page12.
Configuration assistant
With initial commissioning, a configuration wizard will lead you through
the basic settings: Your preferred language, the measured values, controller assignments and switch outputs. With the exception of the language, the values configured here can only be set in the configuration assistant. The finer settings are made in the sub-menus.
Working in the configuration assistant, determine the tasks of the installed modules, the controller and the output modules.
The finer settings such as the behaviour of these modules are made later,
e.g in the “Outputs” menu item.
Commissioning
18
First steps
Perform the following steps:
1. Set the preferred language and press on the arrow.
2. Measured values: Determine the desired measured value for the in-
stalled input modules. Press the right-hand arrow.
Controller: You can assign inputs for up to four controllers in this tab.
Select a sensor, a virtual input or a timer. Set the centre row of the
control function (Table 24: Functions of the individual controllers) and
press the right-hand arrow. Controllers 1 − 4 must be assigned to the
output modules 1 − 4 in a fixed fashion.
3. Control output: You can assign a function to outsput modules in this
tab. Only the output modules which are still free are displayed. Press
the right-hand arrow.
4. Confirm the security query with “Yes” to save the configuration.
The configuration assistant has been ended.
ü
Start the configuration assistant manually in System > Setup
> Configuration > “Configuration assistant”.
The device is set up variably and can be individually adapted to meet your
requirements. As such, it is necessary to adjust the configuration of the
inputs and outputs to the sensors and actors used.
The following section leads you through the device configuration.
8.2.1 Inputs
You can connect up to four sensors for various water parameters and the
temperature to the device. You can also use eight digital inputs.
8.2.1.1 Sensors inputs
The sensors in the device must be configured individually to enable precise and error-free measurement of the water parameters. You can perform various settings.
Perform the following steps:
1. In the main menu, navigate from System > Inputs to the “Sensors”
tab.
2. In the “Sensors” tab, configure every connected sensor and state the
following information.
3. Input: Select the input module of the sensor which you wish to configure.
4. Signal: Enter the type of the sensor signal. Depending on the input
module, the signal type has been specified or you can select a signal
type.
5. Value: Here, you can check which water parameters are measured.
This setting can only be changed in the configuration assistant.
6. Unit: Select the appropriate unit.
7. Measuring range: Enter the maximum measuring range of the sensor.
8. Min-alarm: Activate or deactivate the minimum alarm and state a
value under which the alarm will be triggered.
9. Max-alarm: Activate or deactivate the maximum alarm and state that
will trigger the alarm if itis exceeded.
10. Delay: Set a time delay for the minimum and maximum alarm.
Configuration of the sensors completed.
ü
8.2.1.2 Temperature input
You can connect up to four temperature sensors to the device. This enables you to measure the temperatures at various positions.
8.2.1.3 Compensation
The water parameters which you measure with the device can be falsified by interference (e.g. with temperature or pH value).
The device can compensate these interferences automatically.
Perform the following steps:
1. Working in the main menu under System > Inputs, navigate to the
“Compensation” tab.
2. Working in the “Compensation” tab, configure every sensor connected for which the measured value is to be compensated and state the
following information.
3. Temperature: If it is possible to compensate for the influence of the
temperature, you can select a fixed reference value or one of the four
temperature inputs.
4. pH value: If it is possible to compensate for the pH value error, you
can select a fixed reference value or a sensor input.
Configuration of the compensation completed.
ü
8.2.1.4 Disturbance variable
You can connect the measurement of a disturbance variable (e.g. a flow
volume) to an analogue 4 − 20 mA input. The disturbance variable can
then be taken into account with a factor (0.1 to 10) during the calculation
of the control variable Y. The control variable Y will be multiplied with the
disturbance variable during the calculation.
Example: If the factor = 2 and the disturbance variable amounts to 42%,
the controller can be set to a maximum of the control variable Y = 84 %. If
the factor = 0.5 and the disturbance variable amounts to 42%, the controller can be set to a maximum of the control variable Y=21%.
Perform the following steps:
1. Working in the main menu under System > Controller, navigate to the
“Flow input” tab and state the following information.
2. Disturbance variable: Set the disturbance variable to an input signal
of 4–20mA or 0–20mA. You can also deactivate the disturbance
variable.
3. Unit: As a rule, the disturbance variable is the measurement of a flow.
Select the desired unit.
Configuration of the disturbance variable input completed.
ü
Perform the following steps:
1. In the main menu under System > Inputs, navigate to the “Temperature” tab.
2. In the “Temperature” tab, configure every connected temperature
sensor and state the following information.
3. Measurement: Chose between “On” and “Off”.
4. Min-alarm: Activate or deactivate the “Minimum-alarm” and enter a
temperature under which the alarm will be triggered.
5. Max-alarm: Activate or deactivate the “Maximum-alarm” and enter a
temperature to exceed which the alarm will be triggered.
Configuration of the temperature sensors completed.
You can calculate a new value from multiple measurements or reference
values using a virtual input. You can assign the new virtual value to a controller in the configuration assistant.
In this way, you can calculate a difference, the combined chlorine and the
effective chlorine and use them as a basis for controlling your actors.
Difference
You can calculate the difference between two measured values or the difference between a measured value and a fixed reference value.
Perform the following steps:
1. Working in the main menu under System > Inputs, navigate to the
“Virtual” tab.
2. State the following information.
3. Calculation: Select “difference”.
4. Select a sensor.
5. Select a second sensor or a reference value. The second sensor must
output the same measured value as the first. You will need to enter
the reference value manually.
6. Min-alarm: Activate or deactivate the minimum alarm and state a difference value which, if it falls below a value, will trigger an alarm.
7. Max-alarm: Activate or deactivate the maximum alarm and state a
difference value which, if exceeded, will trigger an alarm.
8. Delay: Set a time delay for the minimum and maximum alarm.
Configuration of the difference completed.
ü
Combined chlorine
Combined chlorine is calculated from the difference between the total
chlorine and the free chlorine:
Combined chlorine = total chlorine - free chlorine
At least one total chlorine measurement is required to calculate combined chlorine. The chlorine value can be entered manually as a single
reference value or a corresponding sensor input is selected.
Perform the following steps:
1. Working in the main menu under System > Inputs, navigate to the
“Virtual” tab.
2. Working in the “Virtual” tab, configure the desired calculation of the
combined chlorine and state the following information.
3. Calculation: Select “Combined chlorine” to calculate the combined
chlorine.
4. Total chlorine: Select the sensor which measures the total chlorine.
5. Free chlorine: Select the sensor which measures the free chlorine. If
no sensor is present, you can enter a reference value measured once
which can be used for the calculation.
6. Min-alarm: Activate or deactivate the minimum alarm and state a
value under which the alarm will be triggered.
7. Max-alarm: Activate or deactivate the maximum alarm and state a
value over which the alarm will be triggered.
8. Delay: Set a time delay for the minimum and maximum alarm.
Configuration of the combined chlorine completed.
ü
Effective chlorine
The disinfectant effect of the free chlorine is highly dependant on the pH
value of the process water. The pH value influences the reactivity of the
Chlorine ions. This relationship is underscored by the dissociation curve
(Fig. 19 “Dissociation curve of the effective chlorine” on page 20) of
the chlorine.
The actual disinfectant effect of the chlorine is generated by the hypochlorous acid (HClO). The figure shows that the proportion of the HClO
is largest between pH 2 and pH 7.5. The disinfectant effect is very low
outside this pH value.
For photometric measurements the pH value of the sample is buffered to
approx. pH 6.5. As a result the measurement has a higher effective chlorine content than is actually in the process water. For high pH-values significant differences will therefore occur between the expected and actual
disinfection if assessed by photometric analysis. The calculation of the effective chlorine can be used to display the proportion of the hypochlorous
acid (HCIO), i.e. the proportion which contributes to the disinfectant effect.
100 %
50
10
0,5
0,1
5
1
0
2
�
\
I
1
4
6 8
\
\
\
\
\
\
14
12
10
(pH)
+ +
+
Fig. 19: Dissociation curve of the effective chlorine
+
Perform the following steps:
1. Working in the main menu under System > Inputs, navigate to the
“Virtual” tab.
2. Working in the “Virtual” tab, configure the desired calculation of the
effective chlorine and state the following information.
3. Calculation: Select “effective chlorine” to calculate the effective chlorine.
4. Free chlorine: Select the sensor which measures the free chlorine.
5. pH value: Select the sensor which measures the pH value. If no sen-
sor is present, you can enter a reference value measured once which
can be used for the calculation.
6. Temperature: A temperature value is required to calculate the effective chlorine. Select the temperature input which can be used for the
calculation. If no temperature sensor is present, you can set a reference value measured once which can be used for the calculation.
7. Min-alarm: Activate or deactivate the minimum alarm and state a
value under which the alarm will be triggered.
8. Max-alarm: Activate or deactivate the maximum alarm and state a
value over which the alarm will be triggered.
9. Delay: Set a time delay for the minimum and maximum alarm.
Configuration of the calculation of the effective chlorine
ü
completed.
8.2.1.6 Digital inputs
You can use up to 8 digital inputs to evaluate switching statuses and to
detect them as alarm message which are to be documented in the logbook.
Perform the following steps:
1. In the main menu under System > Inputs, navigate to the “Digital”
tab.
2. In the “Digital” tab, configure the inputs and state the following information.
3. Action: Choose between “OK = open” (N.O., working contact) or
“OK=contact” (N.C., break contact).
4. Function: Select a function from Table 25 depending on the desired
reaction of your device to the input.
FunctionReaction
Off
Setpoint changeover
Sample water missingAll controller outputs will be switched off.
External stopAll controller outputs will be switched off.
Pre-alarm 1 − 4
The switching of the contact has no
influence on the measurement or control.
You can use the contact to switch between
setpoint sets.
Only display as an alarm message. Nothing
is switched off.
8.2.2 Outputs
Depending on the equipment, you can connect a range of actors to the
device and actuate them. Make sure that you actuate the actor with the
correct signal type and select an appropriate output module with the configuration.
An alarm relay, four analogue outputs and the possibility of connecting
external LEDs (e.g. for water sampling stations) are always available.
8.2.2.1 Controller outputs
You can configure and use up to four controllers.
Precondition for action:
You have used the configuration assistant to assign an input and a
ü
control function to a controller ( “Configuration assistant” on
page18).
Perform the following steps:
1. Working in the main menu under System > Outputs, navigate to the
“Controller” tab.
2. Working in the “Controller” tab, configure the controller output and
state the following information.
3. Y-alarm: Activate the Y alarm. The Y alarm is a safety cut-out. If the
controller output power amounts to more than 95% (e.g. through a
malfunction) over the set time, the Y alarm will be triggered and the
corresponding controller output will be set to 0%. You can set a time
between 0 and 200 minutes.
4. Baseload: Depending on the controller function, you can set a base
load which is always active independently of the control variable.
With a base load of 10 %, the actor is always actuated with a minimum of 10 %.
5. Limit: Depending on the controller function, you can set a limit of between 5 − 100 %. State the value at which the actor should be actuated as maximum.
6. Further settings are dependant on the function of the controller. Further information is available in Table 24 “Functions of the individual
controllers” on page18.
The appendant controller output is switched
off. The other outputs remain unaffected.
You can assign an individual name to this
digital input. The name is displayed in the
alarm messages during switching the
contacts.
8.2.2.2 Actuation via a timer
The output can be used for direct actuation via a timer. This is required
e.g. to run the flocculant pump or the peristaltic pumps over a certain
time.
Precondition for action:
ü
Perform the following steps:
1. Working in the System menu, navigate to System > Outputs and con-
BA-46020-02-V01
Using the configuration assistant, you have assigned a controller to
the “timer” input ( “Configuration assistant” on page18).
figure the output (8.2.2 “Outputs” on page21).
Commissioning
Configuration
21
Multi-Channel Controller TOPAX
®
MC
Operating instructions
2. Navigate to the setpoints menu. Here, you can set the desired set
control output directly from 0 to 100 %. Working in the “Setpoint set”
tab, you can set various control outputs and via the timers in the
“Switch-over” tab, you can determine when the control output
should be changed. The checkmark must be set against “Switch setpoints automatically”. Further information pertaining to switching is
found in section 9.6 “Setpoints and Setpoint sets” on page31.
Actuation configured via a timer.
ü
8.2.2.3 Controller parameters
You can configure the behaviour of the individual controller channels. Explanations of the various functions can be found in Table 24 “Functions of
the individual controllers” on page18 and in section 15 “Glossary” on
page44.
Perform the following steps:
1. Working in the main menu under System > Controller, navigate to the
“parameter” tab.
2. Working in the “Parameter” tab, configure every control channel and
state the following information.
3. Control direction: Configure the direction of control. If a switch is to
be made between a 1- and a 2-side control, this must be set in the
configuration assistant.
4. Function: Set the desired controller function. Possible: P-, PI-, PDand PID controller.
5. Kp, Ki, and Kd: You can configure these parameters depending on the
control function that has been set.
6. Disturbance variable and factor: If you have activated a disturbance
variable (8.2.1.4 “Disturbance variable” on page 19), you can configure the influence of this disturbance variable on the selected controller channel. You can switch the influence on or off and set a factor
between 0.1 and 10.
Configuration of the controller parameters completed.
ü
8.2.2.4 Digital output signals
You can use digital output signals via the outputs of the alarm relay, an
optocoupler circuit board or a relay circuit board. The following sections
describe the configurations that you can perform.
6. Output triggers at: Select which alarms should trigger the alarm relay. The relay switches as soon as one of the selected alarms is active.
7. Alarm delay: Determine the earliest point (in seconds) at which the
relay should switch after activation of the alarm.
Configuration of the alarm relay completed.
ü
Further alarm outputs
In addition to the alarm relay, you can use the unused outputs of the optocoupler circuit board or relay circuit boards for further alarm messages.
Precondition for action:
You have used the configuration assistant to assign the “alarm out-
ü
put” function to a free output ( “Configuration assistant” on
page18).
Perform the following steps:
1. In the main menu under System > Outputs, navigate to the “Digital”
tab.
2. Working under “Digital output” select the desired output.
4The free outputs will be displayed which you have configured as an
“alarm output” in the configuration assistant. Example: “Relay 1.2”.
The first digit stands for the number of the output circuit board (1.X)
and the second digit stands for the number of the output on the circuit board (X.2).
3. Working under the “Function” display, check whether the function
with “alarm output” is displayed.
4. Configure the alarm output by entering the following data.
5. Action: Choose between “normal opened” (N.O., make contact) or
“normal closed” (N.C., break contact).
6. Latching: “On“= the alarm output is active until all alarms have been
manually confirmed. “Off” = the output will be deactivated automatically if the alarms are no longer active.
7. Output triggers at: Select from the list of all possible alarms those
which should trigger the output. The output switches as soon as one
of the selected alarms is active.
8. Alarm delay: Determine the earliest point (in seconds) at which the
output should switch after activation of the alarm.
Configuration of the alarm output completed.
ü
Alarm relay as an alarm output.
You can use the alarm relay (terminals 41 – 43) on the main board as an
output for alarm messages.
Perform the following steps:
1. In the main menu under System > Outputs, navigate to the “Digital”
tab.
2. Working under “Digital output”, select the “alarm relay” output.
3. Configure the alarm relay by entering the following data.
4. Action: Choose between “normal opened” (N.O., make contact) or
“normal closed” (N.C., break contact).
5. Latching: “On “= the alarm relay is active until all alarms have been
manually confirmed. “Off” = the relay is automatically deactivated if
the active alarms are no longer pending.
Commissioning
22
Configuration
BA-46020-02-V01
Limit value control
You can use unused outputs from optocoupler circuit boards or relay circuit boards as a limit value control (including “DIN contact”).
An output for limit value control switches if all measured values are located within the defined limits.
Precondition for action:
You have used the configuration assistant to assign the “limit value
ü
control” function to a free output ( “Configuration assistant” on
page18).
Perform the following steps:
1. In the main menu under System > Outputs, navigate to the “Digital”
tab.
2. Working under “Digital output” select the desired output.
4The free outputs will be displayed which you have configured as
“limit value control” in the configuration assistant. Example: “Relay
1.2”. The first digit stands for the number of the output circuit board
(1.X) and the second digit stands for the number of the output on the
circuit board (X.2).
3. Working in the “Function” display, check whether the function with a
“limit value control” is displayed.
4. Configure the limit value control by entering the following data.
5. Action: Choose between “normal opened” (N.O., make contact) or
“normal closed” (N.C., break contact).
6. Delay: The contact switches only if all water parameters are continually over the set lag time within the limits set. The lag time can be set
between 0 and 10.
7. Parameter: Set the measured values in which the water parameters
must be located so that the output switches. Press the minimum or
maximum value that you wish to change.
Configuration of the limit value control completed.
ü
8.2.2.5 Analogue output signals
The basic configuration of the device includes four analogue 4 – 20 mA
outputs. You can use the outputs to transmit the measured values to a
control room or a PLC (programmable logic controller).
8.2.2.6 Actuation of the LEDs for the water sampling stations
Some water sampling stations have multiple LEDs which display whether
a water parameter is in the desired range. You can configure the actuation
of the LEDs.
Perform the following steps:
1. Working in the main menu under System > Outputs, navigate to the
“External LEDs” tab.
2. Working under “LED”, select the desired LED.
3. Configure the LED by entering the following data.
4. Function: Select the reaction criteria for the LED. Selection of “Sam-
ple water missing” standardly results in a green LED. If the digital
contact “Sample water missing” is active, it will illuminate red. Further options are the measured values of the sensor inputs 1 – 4.
5. Minimum and maximum: If you have decided to use a measured value, you must establish a minimum and a maximum value. The LED
will illuminate red if the minimum value is undercut or the maximum
value exceeded. It illuminates green between the values.
6. Testsignal: You can test the configuration of the LEDs.
Configuration of the external LED outputs completed.
ü
Some actuators such as dosing pumps can be controlled directly via this
output.
Perform the following steps:
1. In the main menu under System > Outputs, navigate to the “Analogue” tab.
2. Working under “Analogue”, select the desired output.
3. Configure the analogue output by entering the following data.
4. 20 mA type: You can choose between the following signal types:
4–20mA, 0 – 20 mA, 20 – 4 mA or 20 – 0 mA.
5. Testsignal: You can test the configuration of the analogue outputs.
Check the actuation on the external device.
6. Output: Determine what should be outputted on the selected output.
You can choose between the measurement and temperature values
or outputs (controller outputs).
7. Minimum: Determine a minimum value. The minimum value indicates for which measured value the signal is the weakest.
8. Maximum: Determine a maximum value. The maximum value indicates for which measured value the signal is the strongest.
The values for “minimum” and “maximum” serve the scaling
of the analogue output signal. Example: Sensor 1 has a measuring range of 0 − 10 mg/l. 4 − 20 mA was selected as the
20mA type. If the complete sensor measuring range is to be
covered by the analogue signal, 0 mg/l should be selected for
“Minimum” and 10 mg/l for “Maximum”. At 0 mg/l, a 4 mA
signal will be issued; at 10 mg/l a 20 mA signal will be issued.
In the main view, the device shows the measured values of the installed
input modules. This display is pre-determined and cannot be changed.
However, you can activate a “second overview” and adapt your needs individually (Fig. 20 “Individual second overview” on page24).
8.2.3.1 Activating the second overview
Perform the following steps:
1. In the main menu under System > Setup, navigate to the “Display”
tab.
2. 2. Overview: Activate the second overview with “On”.
3. Number windows: Select how many individual and freely-configur-
able windows should be displayed.
Second overview activated.
ü
8.2.4 Colours of the alarm messages
You can edit the colours of the different messages.
Perform the following steps:
1. In the main menu under System > Setup, navigate to the “Alarm colour” tab.
2. Look in the “Alarm colour” tab for the message for which you wish to
edit the alarm colour.
3. Press on the row of the alarm and then on “Edit”.
4You can chose between four colour fields.
Alarm colour edited.
ü
8.2.5 Save the configuration
You can save your individual configuration and load it later to rectify problems quickly.
Recommendation: Leave the factory-set configuration file unchanged and save your personal configuration in a new file.
Given problems with the configuration, this enables you to return to a functioning configuration quickly.
Perform the following steps:
Fig. 20: Individual second overview
8.2.3.2 Configuring the second overview
Perform the following steps:
1. Press the “Main menu” button (bottom left on the display) until the
second overview with the individually-settable windows appears.
2. A gear wheel icon is displayed in the top right-hand corner of every
window. Press on the icon to configure the selected window and
state the following information.
3. Name: Give each window an individual name.
4. Display 1 – 5: Up to five displays can be configured depending on the
size of the window. You can choose between the measured values,
the virtual values, the outputs, the setpoints, the disturbance variable,
the digital inputs and the external LED outputs.
Configuration of the second overview completed.
ü
1. In the main menu under System > Setup, navigate to the “Configuration” tab.
2. Press “save” and enter an existing file name to overwrite the file or a
new name to generate a new configuration file.
3. Press on the green checkmark to confirm the entry.
Configuration saved.
ü
8.3 Password protection
The password protection of your device has been deactivated at the factory. You can provide your device with password protection against access to specific functions in three levels.
n1. Level: Only simple settings are accessible here. This level is suita-
ble for daily operation.
n2. Level: The configuration of the inputs and outputs and the adjust-
ment of the sensors are accessible here. This level is required for device configuration and should only be operated by experienced users.
n3. Level: The service menu is accessible here. This level is mainly re-
quired for maintenance work such as changing the sensor, performing
software updates or network settings.
1. Press the lock icon in the left-hand upper corner to configure the
password protection.
2. Password active: Select whether password protection should be activated or deactivated. Password protection can only be deactivated if
you are logged in to level 3.
4Password protection must be activated to unlock the following
steps.
3. Select one of the three password levels into which you wish to log on.
4. Login: Login with the password for the password level previously se-
lected.
5. Change password: You can edit the password of the level in which
you are logged in.
Operating instructions
Password protection configured.
ü
8.4 Network settings
You may need to perform settings in order to be able to use the device in a
network.
Further information about using the device in a network can be found in
section 9.7 “Access via network” on page32.
Perform the following steps:
1. In the main menu under System > Service, navigate to the “Network”
tab.
2. In the “Network” tab, configure the output and state the following information.
3. IP address: Give the device a unique IP address through which it can
be reached in the network. If this IP address is already being used by
another device, errors can result.
4. Subnetmask: Enter the subnetmask.
5. TFTP server: “On” = Access via TFTP protocol activated on the device
memory. “OFF” = Access via TFTP protocol de-activated on the device memory.
6. Modbus RTU address: Enter a number between 1 and 14 in the device if your device is fitted with a RS485 network connection.
When in operation, the device will display the main view with the current
values and the status row with status messages.
Operating instructions
9.2.1 Display the files
You can display the logbook files on the device.
9.1 Confirming a message
You can view device alarm, error and service messages in the status row.
The status row flashes in the colour set for the message type. Inactive
messages are displayed white.
You must confirm a message on the device to end the display of inactive
messages in the status row. Active messages will still be displayed, even
if they have been cleared.
Perform the following steps:
1. When a message is displayed, press on the status row or go to the
“Reports” menu.
2. Select one or more messages and press either “Confirm” or “Confirm
all”.
4The confirmed message is marked with a green tick.
Message confirmed.
ü
History
You can follow the course of the messages in a history.
In the main menu, press “Reportss” and the tab “History”.
9.2 Logbook
Perform the following steps:
1. Press the USB icon in the status row (below right).
4The files saved on the USB flash drive will be displayed in a file
browser.
2. Open one of the file folders.
3. Select the desired file and press “Open file”.
File opened.
ü
9.2.2 Opening files externally
You can open the logbook files on an external device once you have removed the USB flash drive.
DANGER
Mortal danger from electric shock!
When the device housing is open, you can touch live parts. There is the
danger that you could suffer an electric shock.
Ensure that the device has been disconnected from the voltage sup-
ð
ply and is not live when conducting work with an open housing.
Secure the voltage supply against reactivation.
ð
The messages and service entries are saved in the device on a USB flash
drive. You can display the files on the device or connect the USB flash
drive to an external device.
The USB flash drive in the device contains the following logbook files:
LogFormatDescription
REPORTSCSVMessages
TRENDDATTrend data
SERVICECSVService entries
CHANGESCSVChanges to the configuration
SETUPSETConfiguration
ADJUSTMENTCSVAdjustment
Table 26: Logbook
You can open and analyse CSV files with a suitable programme
(e.g. MS Excel). You can also view all CSV files on the device.
Precondition for action:
The voltage supply is deactivated before the start and secured
ü
against reactivation.
Perform the following steps:
1. Open the housing.
2. Remove the USB flash drive.
3. Connect the USB flash drive with an external device and open it.
You can view the trend progression of up to four measured values in the
last 24 hours.
Perform the following steps:
1. In the main menu, press “Trend”.
4The device will show the trend.
2. You can configure the display in accordance with your wishes. Press
the “Display” tab and activate up to four values which are to be displayed in the trend.
3. You can set the scaling of the individual trend display for every value
individually. Press on the “Scaling” tab, select a sensor and define
"minimum” and “maximum”.
Trend display configured.
ü
9.4 Manual mode
You can manually control a controller output in the menu item “Manual
mode” and set an output capacity between 0 and 100 %.
If a controller is in manual mode, this is indicated by a blue Y display in
the main view and by a hand icon.
Working in the menu item “Manual mode”, you can also switch the automatic switching of the setpoints on or off (9.6 “Setpoints and Setpoint
sets” on page31).
9.5.1 pH
Calibration of the pH electrode can be performed as a 2-point calibration
with 2 buffer solutions or a single point calibration with subsequent input
of the slope of the sensor.
The actual voltage of the electrode and the ideal value of the set buffer
solution is displayed during calibration.
The reaction time of a new electrode amounts to a few seconds and has
been set when the physical value has become stable. Older electrodes
can have a longer reaction time.
To perform the calibration performed here, you will require a
buffer solution with a known pH value. Buffer solutions have a
restricted storage life and their pH value changes depending
on their duration of storage and the storage conditions.
Comply with the manufacturer’s instructions pertaining to the
correct storage; use buffer solutions only within the scope of
their life period and never submerge a sensor in a buffer solution immediately after removing it from a different buffer solution.
9.5.1.1 2-point calibration
Pre-conditions for actions:
Two different buffer solutions are ready.
ü
The sensor is clean.
ü
9.5 Adjustment
You must first adjustment the sensors before you can detect the correct
measured values. All adjustment processes in the device are monitored
for plausibility (zero point and slope) and the measured values documented. Measured values resulting from an unadjusted sensor are marked in
the main overview. In this case, the measured values are displayed in red.
If the measurement input is displayed individually in the main view, the
information “Improper adjustment” will be displayed.
The current calibration and slope can be found in the “Adjustment” menu
in the “Overview” tab.
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “sensor”, select the sensor that measures the pH and
which you wish to calibrate.
3. Press the “2 points” button.
42-point calibration begins.
4. Shut off the sample water of the sensor block.
5. Unscrew the pH electrode from the sensor block.
6. Rinse the electrode with water and dab it dry. Rubbing can cause
electrical discharge on the glass membrane, which results in a delayed display.
7. Hold the dry electrode in the first buffer solution. It is not important,
which of the two buffer solutions you start with.
8. Enter the pH value of the first buffer solution. This pH value serves as
a reference value for the device. The ideal voltage value and the current voltage value is displayed in mV. If these values deviate too
greatly from one another, the best value is displayed red. Too great a
level of deviation is an indication that the electrode needs to be replaced.
9. Wait until the value has stabilised.
10. Confirm the entry with the green checkmark.
11. Repeat point 6 to 10 for the second buffer solution.
12. A window with the actual slope of the sensor will open.
The slope of the electrode was measured in a laboratory beforehand.
ü
Perform the following steps:
1. Working in the main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “sensor”, select the sensor that measures the pH and
which you wish to calibrate.
3. Press the “1 point” button.
41-point calibration begins.
4. Shut off the sample water of the sensor block.
5. Unscrew the pH electrode from the sensor block.
6. Rinse the electrode with water and dab it dry. Rubbing can cause
electrical discharge on the glass membrane, which results in a delayed display.
7. Hold the electrode in the buffer solution and move back and forth for
a short time.
8. Enter the pH value of the buffer solution. This pH value serves as a
reference value for the device. The ideal voltage value and the current voltage value is displayed in mV. If these values deviate too
greatly from one another, the best value is displayed red. Too great a
level of deviation is an indication that the electrode needs to be replaced.
9. Wait until the value has stabilised.
10. Confirm the entry with the green checkmark.
11. You will be requested to set the slope. Enter the slope.
12. Confirm the entry with the green checkmark.
1-point calibration completed.
ü
9.5.1.3 Offset compensation
External influences can cause the pH value measured with the photometer to deviate from the electrometric measurement of the pH value by a
constant value. The offset compensation enables you to compensate for
this constant difference (zero-point deviation).
9.5.2 Redox
The Redox value is measured using the Redox sensor. The sensor measures the voltage present in the water due to oxidizing and reducing ions.
You must calibrate the sensor during commissioning.
Pre-conditions for actions:
A buffer solution is on hand for calibration.
ü
Perform the following steps:
1. Working in the main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “Sensor”, select the sensor that measures the Redox
and which you wish to calibrate.
3. Press the “1 point” button.
41-point calibration begins.
4. Shut off the sample water of the sensor block.
5. Unscrew the Redox electrode from the sensor block.
6. Rinse the electrode with water and dab it dry. Rubbing can cause
electrical discharge on the glass membrane, which results in a delayed display.
7. Hold the electrode in the buffer solution and move back and forth for
a short time.
8. Enter the voltage value in mV which is recorded on the buffer solution. The voltage value entered and the current measured voltage
value are displayed in mV. If these values deviate too greatly from
one another, the best value is displayed red. Too great a level of deviation is an indication that the electrode needs to be replaced.
9. Wait until the value has stabilised.
10. Confirm the entry with the green checkmark.
1-point calibration completed.
ü
With older sensors, the reaction time can increase or the
measured value can differ considerably from the buffer solution value. This indicates that the Redox sensor must be
checked and replaced if necessary.
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “Sensor”, select the sensor that measures the pH and
which you wish to calibrate.
3. Press the “Offset” button.
4. You can enter a pH offset from -0.30 pH to max +0.30 pH.
9.5.3 3-Electrode potentiostat and excess chlorine
measuring cell CS120
You should calibrate a 3-electrode potentiostat or the potentiostatic
measuring cells (chlorine sensors) as single-point calibration as a matter
of course. You will require a photometrically measured value in accordance with the DPD method as a reference value.
With operation in a hot water system, electrochemical processes on the measuring electrode can result in a displacement of the zero point. In this case, 2-point calibration is necessary in which you calibrate the zero point using
chlorine-free hot water. The second point is determined using
the DPD method with chlorinated water as usual.
9.5.3.1 1-point calibration
To calibrate chlorine sensors, you will require a photometer with which to
measure the reference value using the DPD method.
Pre-conditions for actions:
A measurement device for determining the DPD value is already
ü
present.
The sensor is operated with sample water.
ü
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “Sensor”, select the sensor that you wish to calibrate.
3. Press the “1 point” button.
41-point calibration begins.
4. Take sample water in immediate proximity to the measuring cell and
confirm with “OK”.
5. Determine the concentration in the sample water using the DPD
method.
6. Enter the measured concentration. This serves the device as a reference value with which to permit correct measurement.
7. Confirm the entry with the green checkmark.
4. You will be asked to set the first reference value; this is the zero point.
If the zero point was mal-set by accident, set it by pinching off the
measuring electrode and entering a value of zero. Should real 2-point
calibration be performed, e.g. due to hot water, you must first perform calibration with chlorine-free and then with chlorinated water.
Enter a value for the zero point.
5. Confirm the entry with the green checkmark.
6. Take sample water in immediate proximity to the measuring cell and
confirm with “OK”. This means that the current signal at the time of
the sample water extraction is saved to rule out signal fluctuation as
a measurement error during the DPD ascertainment.
7. Determine the concentration in the sample water using the DPD
method.
8. First enter the determined DPD value.
9. Confirm the entry with the green checkmark.
2-point calibration completed.
ü
9.5.4 Temperature
You can connect a temperature sensor to every input module. You can adjust the temperature sensor by setting a reference value.
When setting the reference value, the device will automatically correct
the measurement of the temperature sensor by the difference.
Pre-conditions for actions:
A thermometer is available.
ü
You have activated the measurement of the temperature (8.2.1.2
ü
“Temperature input” on page19).
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the
“Temperature” tab.
2. Working under “input”, select the input module for which you wish to
set a reference value.
3. Press the “Reference value” button.
4. Enter the reference temperature measured beforehand.
5. Confirm with the green checkmark.
1-point calibration completed.
ü
9.5.3.2 2-point calibration
Pre-conditions for actions:
A measurement device for determining the DPD value is already
ü
present.
The sensor is operated with sample water.
ü
Perform the following steps:
1. Working in the main menu under “Adjustment”, navigate to the “Sensors” tab.
2. Working under “Sensor”, select the sensor that you wish to calibrate.
You can connect servomotors with a position feedback to your device and
control them via an output. Before you can control the servomotor exactly, you must first adjust the control via your device and the position of the
servomotor.
During adjustment, the servomotor is first driven to the end position and
then back.
This section only applies to servomotors with a feedback potentiometer.
Pre-conditions for actions:
The servomotor is switched on and has been connected properly.
ü
The output has been configured correctly (Table 24 “Functions of the
ü
individual controllers” on page18).
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Outputs” tab.
2. Working under “Output”, select the output which is to be adjusted.
3. Press the “Setup servo motor” button.
4. A further window with a progress bar opens; this indicates the posi-
tion feedback.
5. Adjustment is started using the “Start” button.
4The motor now opens up completely and closes again. This can take
a number of minutes.
6. Completion of the adjustment is signalled with “Adjustment OK”.
Adjustment completed.
ü
9.5.5.2 Servomotor without position feedback
You can actuate servomotors without a position feedback. To this end, you
need to measure how quickly the motor starts and then set the controller
accordingly.
6. Deactivate manual mode with “Off”.
7. Working in the main menu under “Outputs” navigate to the “Control-
ler” tab.
8. Working under “Output”, select the output of the runtime which you
have just measured.
9. Working under “Runtime”, enter the time which you have just measured.
Runtime set.
ü
9.5.6 Servomotor mA
You can connect servomotors with a 20 mA actuation and a 20 mA feedback to the device. The servomotors must be calibrated with the actuation before commissioning. You can calibrate the 20 mA output signal
with ±1 mA.
9.5.6.1 Correction
Pre-conditions for actions:
The servomotor is switched on and has been connected properly.
ü
The output has been configured correctly (Table 24 “Functions of the
ü
individual controllers” on page18).
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Outputs” tab.
2. Working under “Output”, select the output which is to be adjusted.
3. Press the “Correction” button.
4A further window with a progress bar opens; this indicates the posi-
tion feedback.
4. Adjustment is started using the “Start” button.
5. The motor starts and opens up, then stops. This can take a number of
minutes.
6. Completion of the adjustment is signalled with “Adjustment OK”.
Correction completed.
ü
Precondition for action:
The servomotor is switched on and has been connected properly.
ü
The output has been configured correctly (Table 24 “Functions of the
ü
individual controllers” on page18).
@Timer
Perform the following steps:
1. Working in the main menu, navigate to the “Manual mode” menu
item.
2. Activate the manual mode of the output with “On” and set the control
output as 0 %.
4The servomotor closes completely.
3. Observe the drive and wait until it has closed completely.
4. Set the output to 100 % and at the same time, start to measure the
time for the complete opening of the drive.
4The servomotor now starts and opens up.
5. Observe the drive and wait until it opened up completely.
30
Operation
Adjustment
BA-46020-02-V01
9.5.6.2 Calibrating the 20 mA output
If, whilst performing the previous compensation, you have determined
that the output signal does not completely conform with the motor position, you can adjust the output signal. You can adjust both the lower signal
range (0/4 mA) and the upper signal range (20 mA) to the motor position.
Perform the following steps:
1. Working in the Main menu under “Adjustment”, navigate to the “Outputs” tab.
2. Working under “Outputs”, select the output that you wish to adjust.
3. Press the “1 point” button to calibrate just the upper area or the “2
points” button for the upper and lower areas.
4A further window will open in which you can set a tolerance. You can
change this by maximum -50 to +50 depending on whether you
wish to calibrate the upper or lower area. The maximum change corresponds c. to a current of 1 mA.
4. Check the output signal using a multimeter or a servomotor and
make sure that the motor is now under exact control.
5. Save the current offset by pressing the “Save” button.
Calibration completed.
ü
9.6 Setpoints and Setpoint sets
You can determine various setpoints which should actuate the device. You
have four different sets of setpoints; the controller can switch between
them automatically. The setpoint sets can be used to vary the control at
different times of the day or in different operating conditions.
The following section describes the possibilities of controlling using setpoints and their configuration.
9.6.1 Setting the setpoints
You can configure and save the setpoints via the menu item “Setpoints”.
Perform the following steps:
1. Working in the main menu under “Setpoints” navigate to the “Active”
tab.
2. Active: You can view the setpoint set currently active in the “Active”
tab.
4The individual setpoints are displayed. If you change one of the val-
ues, it will immediately be activated as a new setpoint.
3. Save: You can save the active setpoints as a setpoint set. Select a
setpoint set for this and press “Save".
4. Setpoints 1 − 4: The possible setpoints for the controller follow in sequence. The following information is displayed from left to right:
- S1 − S4/V1 − V4 indicates the sensor input or virtual input.
- O1 − O4 indicates the output. The “timer” indicates that the output
is actuated directly.
- This is followed by the measured value of this input, e.g. free chlorine, pH or the text “No output” if this controller is inactive.
An upwards or downwards arrow indicates the control direction.
Raise or lower.
- This is followed by the setpoint. Pressing on the setpoint enables
you to change it immediately. If it is a 2-side control, two setpoints
must be entered. Both for the control direction “Lift” or “Lower”.
Changing these setpoints does not have an impact on the savable
setpoint sets. To do so, you must perform step 3.
5. Flow rate: You can activate or deactivate the flow. This enables you to
reduce the flow by between 0 % − 100 %. The value of the flow is
multiplied with the output Y. If the output is e.g. 80 % but the flow is
only 50 %, this produces an output capacity of: Control variable
Y=80 % ∙ 50 % = 40 %.
2. Select the desired setpoint set.
3. Press "Load”.
4The desired setpoint set is active immediately.
Setpoint set loaded.
ü
9.6.3 Switching between setpoints
The setpoint sets set in the previous section can be switched manually or
automatically. You have two possibilities to activate automatic switching.
1. Working in the “Setpoints” menu item, navigate to the “Switch-over”
tab. Set a checkmark against “Switch setpoints automatically”.
2. Working in the “Manual mode” menu, set a checkmark against
“Switch setpoints automatically”.
You can use a digital input (9.6.3.1 “Switching via digital input” on
page 31) and multiple internal timers (9.6.3.2 “Switching via timer-switch” on page 32) for automatic switching. The switching via a
digital input has priority. Switching via a timer only occurs if no switching
is active via a digital input.
9.6.3.1 Switching via digital input
Before you can use a digital input for switching to a particular setpoint
set, you must configure the digital input in accordance with section
8.2.1.6 “Digital inputs” on page21.
Switching can be performed in three different forms: externally-controlled switching; switching via an internal timer and the “ECO control”
function, which includes the limit value control.
External switching
In the case of external actuation, switching to the desired setpoint set is
performed as long as the digital input has been activated.
If the digital input is deactivated, the device switches back to the previous
setpoint set.
To configure, working in the “Setpoints” menu item, select the “Switchover” tab and perform the following steps:
1. Configure the switching and state the following information.
2. Automatically switching the setpoint sets: Set a checkmark here.
3. Switch-over: Select “Digital input”.
4. Function: Select the point “External switching”.
5. Setpoint set: Select the setpoint set to which is to be switched.
6. Digital input: The digital input in use is indicated here.
Configuration completed.
ü
Setpoints set.
ü
9.6.2 Loading setpoint sets
You can load a setpoint set.
Perform the following steps:
1. Working in the main menu under “Setpoints”, navigate to the “Setpoint set” tab.
During switching via a timer, the desired setpoint set is active until the set
period has been completed. The previous setpoint set is re-activated after
the time has elapsed.
You can also start the timer manually, thereby e.g. triggering shock chlorination.
To configure, working in the “Setpoints” menu item, select the “Switchover” tab and perform the following steps:
Setpoints and Setpoint sets
Operation
31
Multi-Channel Controller TOPAX
i
®
MC
Operating instructions
1. Configure the switching and state the following information.
2. Automatically switching the setpoint sets: Set a checkmark here.
3. Switch-over: Select “Digital input”.
4. Function: Go to “Timer“.
5. Setpoint set: Select the setpoint set to which is to be switched.
6. Digital input: The digital input in use is indicated here.
7. Time: Configure the timer to the desired duration.
8. “Off” or “Active”: It will be displayed here whether the timer is cur-
rently active. If this is the case, the display will show how long it is
still active.
9. “Start” and “Stop” button: The timer can be started or stopped manually. For example, for a shock chlorination.
Configuration completed.
ü
Switching via ECO control
A limit value control is defined for the "Eco control”. The limit value control is generally used to reduce the circulation capacity.
If the measured values are located within the setpoints, the output is
closed. If the digital input is also switched for switching to another setpoint set, “Economy mode” is active and the switch will be made to the
desired setpoint set.
Precondition for action:
The limit value control is configured as described in section “Limit
ü
value control” on page22.
To configure, working in the “Setpoints” menu item, select the “Switching” tab and perform the following steps:
1. Automatically switching the setpoint sets: Set a checkmark here.
2. Switch-over: Select “Switching input”.
3. Function: Select “ECO control”.
4. Setpoint set: Select the setpoint set to which is to be switched.
5. Digital input: The digital input in use is indicated here.
Configuration completed.
ü
9.6.3.2 Switching via timer-switch
Up to ten switches can be configured parallel to the switching via a digital
switching input (9.6.3.1 “Switching via digital input” on page 31).
Times are defined for the point at which the switches should switch to a
certain setpoint set.
6. Setpoint set: Select the setpoint set to which is to be switched.
Configuration completed.
ü
9.7 Access via network
Accessing the device via a network requires that it is connected to an existing Ethernet or RS485 network.
Further information about connection to an existing network is specified
in sections 7.9 “Connecting Ethernet” on page 17, 7.10 “RS485-interface” on page17 and 8.4 “Network settings” on page25.
If connection problems are experienced during access via network, check the configuration of your security software.
Modbus
You can access certain data on the device via the Modbus protocol using
both Ethernet and the RS485. You need the Modbus protocol e.g. For the
connection with a control panel or a PLC. Modbus TCP/IP is supported for
Ethernet and Modbus RTU is supported for the RS485 interface.
The Modbus addresses of your device are stated in section 12 “Modbus
addresses” on page36.
Web browser (only Ethernet)
You can access the device data using all network devices which are fitted
with a web browser. You will require the IP address, subnetmask and possibly the MAC address of the device.
The network settings of your device are listed under Main menu > System > Service > Network.
Open the web browser of your end device and enter the IP address of the
device in the address row. The page of the device will open and provide a
range of information.
TFTP protocol (only Ethernet)
You can access the device memory via a TFTP client software as long as
TFTP is activated in the network settings. You need the device IP address
for access.
The network settings of your device are listed under Main menu > System > Service > Network.
To configure, working in the “Setpoints” menu item, select the “Switching” tab and perform the following steps:
1. Automatically switching the setpoint sets: Set a checkmark here.
2. Switch-over: Configure up to ten timer switches and state the follow-
ing information.
3. Off/On: Switch on the timer.
4. Time: Configure a time at which the switch-over is to be made. State
the hour and minutes.
5. Monday - Sunday: Set a checkmark against every weekday on which
the timer should be active.
Products by Lutz-Jesco are manufactured to the highest quality standards and have a long service life. However, some parts are subject to operational wear. This means that regular visual inspections are necessary
to ensure a long operating life. Regular maintenance will protect the device from operation interruptions.
10.1 Maintenance intervals
This table gives you an overview of maintenance work and the intervals
at which you must carry it out. The next few sections contain instructions
for carrying out this work.
IntervalMaintenance
nVisual check
Monthly
Table 27: Maintenance intervals
nTouchscreen function test
nCalibrating the measured values
10.2 Keeping logfiles
If you make an entry in the logfiles, the device will issue a reminder when
a sensor needs to be replaced.
Perform the following steps:
10.4 Battery change
The device is fitted with a button cell. Check the button cell within the
scope of the annual maintenance. The lifetime of the button cell is determined by the device usage and can vary considerably.
You will need to replace the battery more often with devices which are
switched off often or over a long period (e.g. over winter).
Fig. 22: CR1220 button cell
10.4.1 Checking the charge
You can check the battery charge easily using the device. Replace the
battery if the charge amounts to less than 2.9V.
1. Working in the main menu, navigate to System > Service > Service
entry and working under “Service entry”/“Sensor”, select the desired sensor.
2. Enter the serial number in the tab and the manufacturing company of
the sensor.
3. Activate the reminder function and enter a date for the next sensor
change.
Logfiles maintained.
ü
10.3 Updating software
The most up-to-date firmware version can be downloaded
from www.Lutz-Jesco.com. Copy this *.BIN file onto the device USB flash drive. The file must be saved in the root directory of the USB flashdrive and may not be stored in a sub-folder.
You can update the device software to a newer version.
Perform the following steps:
1. Working in the main menu, navigate to System > Service > Device.
2. Press “Software update”.
3. Select the *.BIN file with the newer version and press “Load“.
4The software is installed. The device will restart automatically during
this procedure.
Update performed.
ü
èWorking in the main menu, navigate to System > Information > Sys-
tem values and read the current charge state of the battery.
10.4.2 Replace the battery
You must remove the two input circuit boards to be able to replace the
battery (Fig. 22 “CR1220 button cell” on page33).
Precondition for action:
The voltage supply has been disconnected and protected against
ü
re-connection.
The housing is open.
ü
Resources required:
@Socket wrench 5.5mm (M3)
@New battery: CR1220, Ø12,5 mm, 3 V, 35 mAh
Perform the following steps:
1. Pull all cable connections from the input circuit boards which you
need to dismantle.
2. Using the socket wrench, unscrew the retaining nuts from the white
protective plate and remove the plate.
3. Using the socket wrench, unscrew the two nuts from the input circuit
boards which you need to remove.
4. Working carefully, slide the input circuit boards from their brackets.
4The battery is now easily accessible.
5. Lever the battery out of its holder without damaging the contact bow.
Your device is fitted with an electrical fuse to protect against short circuits
or over-voltage. You can change the fuse if it is defective.
44
Fig. 23: Position of the fuse
Precondition for action:
The voltage supply has been disconnected and protected against
ü
re-connection.
The housing is open.
ü
Resources required:
47
45
48
46
49 50 51
52
Internal fuse
@Slotted screwdriver
@New fuse: 5 x 20 mm, 3.15 A, 250 V (delay)
Perform the following steps:
1. The fuse holder in the form of a bayonet catch is located at the bottom right-hand side, above the clamps for PE, N and L with the marking “Fuse”. Use the slotted screwdriver to press the catch downwards and then turn it leftwards.
2. Remove the fuse.
3. Replace the fuse and fix it in place by turning the catch clockwise.
Fuse has been replaced.
ü
10.6 Resetting the settings
3. Confirm with “Yes”.
4The configuration will be deleted. You must proceed with the follow-
ing section.
All factory default settings will be reset.
ü
Reset the configuration
The device configuration will be saved in *.SET files. A factory-set configuration file with standard settings is already present. You can change
these or save your personal configuration in new files.
Recommendation: Leave the factory-set configuration file unchanged
and save your personal configuration in a new file. Given problems with
the configuration, this enables you to return to a functioning configuration
quickly.
Perform the following steps:
1. Working in the main menu, navigate to System > Settings > Configuration.
2. Select an existing configuration file.
3. Click “Load” to confirm.
4The device configuration returns to the saved state.
Load the old configuration.
ü
10.7 Finishing maintenance
Perform the following steps:
1. Make a note of the date and scope of the maintenance performed.
2. In the “Service” menu, navigate to the “Service entry” tab. Enter your
company name and notes about the maintenance. Activate the reminder function and enter a date for the next service. Confirm with
the “Save” button .
4Your service action has been saved in the logfiles.
3. To restart the system, proceed in accordance with the instructions in
section 8 “Commissioning” on page18.
The instructions differentiate between the internal factory settings and the device configuration.
The factory settings contain the basic configuration of the device hardware and cannot be changed.
The configuration file (*.SET) contains the individual device
configuration. You can change, save and load the individual
settings.
Reset to the factory settings
You can now reset the device to its factory settings. This deletes the configuration. You must then either load a configuration file or perform the
configuration manually.
Perform the following steps:
1. Working in the main menu, navigate to System > Service > Device.
See below for information about how to rectify faults on the device or the system. If you cannot eliminate the fault, please consult with the manufacturer
on further measures or return the device for repair.
FaultPossible causeRemedy
nCheck that the battery is really empty. To do
so, navigate to the menu System >
The device loses all settings after it has been
disconnected from the voltage network and
then reconnected.
The device is off.
The sensor error is displayed as an alarm.
Table 28: Troubleshooting
The battery is empty.
The power supply has been interrupted.Restore the power supply.
The device fuse is defective.
The sensor has not been installed correctly.
The signal cable to the sensor has a break.Replace the signal cable.
Information > System values. You can view
the battery voltage under “Battery”. If the
voltage is under 2.9 V, change the battery.
nReplace the battery (section 10.4 “Battery
change” on page33)
Replace the fuse (section 10.5 “Replacing the
fuse” on page34)
Make sure that the sensor has been connected
correctly (section 7.5 “Connecting sensors” on
page14)
With a DOUBLE-WORD, the HIGH-WORD is transferred first! Hexadecimal display is shown by a leading “0x”.
Operating instructions
Address
DescriptionMeaning
Read
Write
Data that are not channel-related.
4xDevice type + version
6 – 8xSoftware Version:
10 – 11xOperating hours
13xHardware version
2000 – 2002xSerial number
2003xStatus of the digital inputs
2004xStatus digital outputs
0x0500 + (number of channels -1)
A single-channel controller accordingly has the identification: 0x0500
Transfer: ASCII sign
e.g. 102 is the software version V1.02
The information consists of a ASCII sign in HIGH-BYTE and one in
LOW-BYTE
Serial number: 123456 will thus be transferred as
Address 2000: 0x3132
Address 2001: 0x3334
Address 2002: 0x3536
Displays the terminal logic (not the configured software function).
The individual bits are assigned directly to the input clamps.
Example: 0x01 means that the first digital inputs (clamps 21 + 22) are
actuated.
The individual bits of the output modules.
Example: 0x03 means that the upper relay or the upper optocoupler of
the second output module (from the top) is active.
2008 – 2017xxName of the device
Table 29: Modbus addresses
Max. of 20 characters
Caution!
The evaluation must stop at the first zero (string end).
The individual letters are located in the HIGH-BYTE and LOW-BYTE of
every address. “GW” thus produces:
Address 2008 = 0x4757
Address 2009 = 0x00??
The question marks are undefined.
In this case, all other addresses send undefined values.
Input-related data. 20 addresses will be held available for each of the 1 to 4 inputs. The address space for the inputs begins at 2020, 2040,
2060 and 2080.
500 = 50,0 % (2 Byte signed int)
Second side if 2-side control is active.
Number of positions after the decimal point, see unit (4 bytes signed
int).
2029 – 2030xxSetpoint 2If 2-side control is active (4 byte signed int).
2031xxKp
Number of positions after the decimal point, see unit (2 bytes unsigned
int).
2032xxKdDerivative time in s (2 byte unsigned int).
2033xxKiReset time in s (2 byte unsigned int).
Read:
Bit 0 = minimum alarm is active
Bit 1 = maximum alarm is active
Read:
Bit 0 = Y alarm is inactive
Bit 1 = Y alarm is active
2034xxMinimum and maximum alarm
2035xxY alarm
Write:
0 = clear alarm
Write:
0 = clear alarm
2036xxManual mode
Bit 0: Manual mode on
Bit 1: Lower (with 2-side control)
Input-related data of the virtual inputs. 20 addresses will be held available for each of the 1 to 4 inputs. The address space for the virtual
inputs begins at 2100, 2120, 2140 and 2160.
If the bit is set, the associated alarm or
message is active.
Alarm status 2
If the bit is set, the associated alarm or
message is active.
Message or alarmBit
Sensor error 1
Sensor error 2
Sensor error 3
Sensor error 4
Sensor 1 maximum alarm
Sensor 2 maximum alarm
Sensor 3 maximum alarm
Sensor 4 maximum alarm
Virtual 1 maximum alarm
Virtual 2 maximum alarm
Virtual 3 maximum alarm
Virtual 4 maximum alarm
Sensor 1 minimum alarm
Sensor 2 minimum alarm
Sensor 3 minimum alarm
Sensor 4 minimum alarm
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Message or alarmBit
Virtual 1 minimum alarm
Virtual 2 minimum alarm
Virtual 3 minimum alarm
Virtual 4 minimum alarm
Controller 1 Y alarm
Controller 2 Y alarm
Controller 3 Y alarm
Controller 4 Y alarm
Temperature 1 maximum alarm
Temperature 2 maximum alarm
Temperature 3 maximum alarm
Temperature 4 maximum alarm
Temperature 1 minimum alarm
Temperature 2 minimum alarm
Temperature 3 minimum alarm
Temperature 4 minimum alarm
Sensor change sensor 1 due
Sensor change sensor 2 due
Sensor change sensor 3 due
Sensor change sensor 4 due
0
1
2
3
235 = 23.5 °C
With an inactive temperature, the return is -10000
(2 byte signed int)
Modbus addresses
41
Multi-Channel Controller TOPAX
(DE) EU-Konformitätserklärung
Hiermit erklären wir, dass das nachfolgend bezeichnete Gerät aufgrund seiner Konzipierung und Bauart sowie in der von uns in Verkehr gebrachten Ausführung
den einschlägigen grundlegenden Sicherheits- und Gesundheitsanforderungen der aufgeführten EU-Richtlinien entspricht. Bei einer nicht mit uns abgestimmten
Änderung am Gerät verliert diese Erklärung ihre Gültigkeit.
(FR) Déclaration de conformité UE
Nous déclarons sous notre propre responsabilité que le produit ci-dessous mentionné répond aux exigences essentielles de sécurité et de santé des directives UE
énumérées aussi bien sur le plan de sa conception et de son type de construction que du modèle que nous avons mis en circulation.
Cette déclaration perdra sa validité en cas d’une modification effectuée sur le produit sans notre accord explicite.
(EN) EU Declaration of Conformity
We hereby certify that the device described in the following complies with the relevant fundamental safety and sanitary requirements and the listed EU regulations
due to the concept and design of the version sold by us.
If the device is modified without our consent, this declaration loses its validity.
(ES) Declaración de conformidad UE
Por la presente declaramos que, dados la concepción y los aspectos constructivos del modelo puesto por nosotros en circulación, el aparato mencionado a continuación cumple con los requisitos sanitarios y de seguridad vigentes de las directivas de la U.E. citadas a continuación.
Esta declaración será invalidad por cambios en el aparato realizados sin nuestro consentimiento.
EU-Richtlinien:
EU directives:
Dokumentationsbevollmächtigter:
Authorized person for documentation:
Lutz-Jesco GmbH
Am Bostelberge 19
30900 Wedemark
Germany
Harmonisierte Normen:
Harmonized standards:
(PT) Declaração de conformidade UE
Declaramos pelo presente documento que o equipamento a seguir descrito, devido à sua concepção e ao tipo de construção daí resultante, bem como a versão
por nós lançada no mercado, cumpre as exigências básicas aplicáveis de segurança e de saúde das directivas CE indicadas.
A presente declaração perde a sua validade em caso de alteração ao equipamento não autorizada por nós.
Designação do aparelho:
Mehrkanalregler
Multi-Channel Controller
Régulateur multi-canaux
Controlador multi canal
Controlador multi-canal
Topax MC
2014/30/EU
2014/35/EU
2011/65/EU
Die Schutzziele der Niederspannungsrichtlinie 2014/35/EU wurden gemäß Anhang I, Nr. 1.5.1
der Maschinenrichtlinie 2006/42/EG eingehalten.
The protective aims of the Low Voltage Directive 2014/35/EU were adhered to in accordance
with Annex I, No. 1.5.1 of the Machinery Directive 2006/42/EC.
DIN EN ISO 12100:2011-03
DIN EN 61000-4-2:2009-12
DIN EN 61000-4-3:2006 + A1:2008 + A2:2010
DIN EN 61000-4-4:2012
DIN EN 61000-4-5:2014
DIN EN 61000-4-6:2014-08
DIN EN 61000-4-11:2005-02
DIN EN 61000-6-2:2016-05
DIN EN 61000-6-3:2011-09
DIN EN 55016-2-3:2010 + A1:2010
Manufacturer order no.: .......................................................................................... Date of delivery: .........................................................
Device type: ............................................................................................................ Serial number: ...........................................................
Description of fault:.....................................................................................................................................................................................
Point of use / system designation:...............................................................................................................................................................
Accessories used (suction line etc.):............................................................................................................................................................
Duty period (approx. operating hours): ........................................................................................................................................................
Please describe the specific installation and enclose a simple drawing or picture of the chemical feed system, showing materials of construction, diameters, lengths and heights of suction and discharge lines.
The limit value control is an output which switches when determined limit values are exceeded or undercut. This function is used to
control an ECO or Night mode in a swimming pool with reduced circulation. If there are no swimmers in the swimming pool, energy and dosing
media can be saved. The limit value control monitors the maintenance of
the parameters.
In Germany, the limit values of the national standard DIN 19643 apply. As
such, “DIN contact“ is a widespread name for this function.
Hysteresis
Hysteresis is the delayed response time of a two position controller,
when it reaches the “setpoint Y” and the control switches quickly between raising and lowering. As quick switching or control can have undesired effects, you can define hysteresis and achieve a more equal control.
Actual value X
The actual value ‘X’ is the continually measured value of a sensor.
Reset time Ki
The reset time ‘Ki’ is the integral proportion (I proportion) with PI and
PID controllers. ‘Ki’ is the time required by the controller to alter the control variable ‘Y’, which is generated by the proportional range ‘Kp’
immediately after the step change of the control deviation ‘X-W‘.
You can set a reset time ‘Ki’ of up to 200 minutes.
P controller
A P controller is defined by the proportional range ‘Kp’. The use of a
pure P controller means that a control deviation ‘X-W’ always remains.
The setpoint ‘W’ will thus never be reached.
PI controller
The PI controller is defined by the proportional range ‘Kp’ and the reset time ‘Ki’. The use of the PI controller means that the actual value ‘X’
can reach the setpoint ‘W’.
The PI controller is suitable for the majority of applications.
PD-controller
The PD controller is defined by the proportional range ‘Kp’ and the derivative time ‘Kd’. The use of a PD controller means that a control devia-
tion ‘X-W’ always remains. The setpoint ‘W’ will thus never be reached.
PID controller
Proportional range Kp
The proportional range ‘Kp’ (p proportion) of a P, PI or PID controller indicates the amount by which the actual value ‘X’ must deviate from the setpoint ‘W’ so that the variable Y = 100 %. If the control deviation
‘X-W’ is lower, the control variable is also lower.
The control variable ‘Y’ of a P controller is only affected by the control deviation ‘X-W’. The Kp value is stated in the unit of the variable to be controlled. If for example, during the control of the pH value, an Kp = 2 pH is
selected and the actual value is X = 9 pH and setpoint W = 7 pH, the control deviation is X-W = 9 pH – 7 pH = 2 pH.
In this case, the deviation X-W is as large as the Kp value. In this case, the
variable Y would be 100 %. With a decreasing deviation X-W, the control
variable decreases in a linear fashion to 0 % with an actual value X = setpoint W.
Control deviation X-W
The control deviation X-W is the difference between the actual value
‘X’ and the setpoint ‘W’. The control variable ‘Y’ results from the control
deviation.
Setpoint W
The setpoint ‘W’ of a control is the desired value.
Control variable Y
The control variable ‘Y’ is the value with which the controller actuates
the actor in accordance with its set parameters and the control deviation
‘X-W’. The value lies between 0 % and 100 %.
Derivative time Kd
With PD or PID controllers, the differential proportion (D proportion) is defined with the derivative time ‘Kd’. The D proportion ensures that the
control path already contains a correction factor at the point at
which the actual value "X" begins to differ from setpoint "W". The
control variable ‘Y’ depends on the speed with which the control deviation
‘X-W’ takes place. The duration of the correction is determined by the derivative time ‘Kd’. If the actual value ‘X’ does not change, i.e. the speed of
change is “0”, the correction factor effected by the D proportion with the
derivative time ‘Kd’ drops as far as “0” (even if the actual value ‘X’ does
not match the setpoint ‘W’, but consistently deviates from it). The fact that
the control system causes the actual value ‘X’ to match the setpoint ‘W’ is
due mainly to the I proportion of the controller. The D proportion often improves the controller behaviour because it acts against the trend to deviate.
The derivative time Kd can be set from 0 seconds to a maximum of 1200
seconds.
The PID controller is defined by the proportional range ‘Kp’, the reset time ‘Ki’ and the derivative time ‘Kd’. The integral gain means that the
actual value ‘X’ can reach the setpoint ‘W’.
44
Glossary
BA-46020-02-V01
Y alarm
You can activate an alarm which deactivates the controller if the control variable ‘Y’ amounts to over 95 % over a defined period.