Lutron Electronics TC-424 Operation Manual

Pt-100 calibrator,Pt 100 thermometer,Type K,J,T,R,E,S thermocouple
THERMOMETER CALIBRATOR
Source & Measurement
400 ohm resistance calibrator, measurement DC mV calibrator(source) , measurement
Model : TC-424
Your purchase of this 4 in 1 THERMOMETER CALIBRATOR
marks a step forward for you into the field of precision measurement. Although this meter a complex and delicate instrument, its durable structure will allow many years of use if proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
OPERATION MANUAL
TABLE OF CONTENTS
.......................................................................
..........................................
......................................................................
..........................................................
4-1. Thermocouple SOURCE PROVIDE PROCEDURE ....................................................................................................................................................................................
4-2. VOLTAGE SOURCE PROVIDE PROCEDURE ..........................................................................................................................................
4-3. RTD Pt100 ohm SOURCE PROVIDE PROCEDURE .....................................................................................................................................................................
4-4. 400 ohm SOURCE PROVIDE PROCEDURE .....................................................................................................................................................................
..........................................................
5-1.Thermocouple Type MEASURE PROCEDURE .....................................................................................................................................................................................
5-2. VOLTAGE MEASURE PROCEDURE ..........................................................................................................................................
5-3. RTD Pt100 ohm MEASURE PROCEDURE ..........................................................................................................................................
5-4. 400 ohm MEASURE PROCEDURE ..........................................................................................................................................
5-5. LCD Backlight Operation PROCEDURE ..........................................................................................................................................
. ADVANCED SETTING.......................................................................................................
6-1 Auto power OFF management ..........................................................................................................................................
...........................................................................................................................................................................................
. POWER SUPPLY from DC ADAPTER........................................................................................
. BATTERY REPLACEMENT.................................................................................................
...........................................
1. FEATURES....................................................................................
2. SPECIFICATIONS....................................................................................................
3. FRONT PANEL DESCRIPTION....................
4. SOURCE PROVIDE PROCEDURE................................................................................
5. MEASUREMENT PROCEDURE................................................................................
6
6-2 Select the Temp. unit to or
7 8
9. SYSTEM RESET............................................................................................................
1
1
5
6
6
6
7
7
7
7
8
8
8
8
8
9
9
10
10
10
1. FEATURES
*
Four kinds Source output :
1. Pt-100 source, 2. Voltage( DC mV) source.
3. 400 ohm source , 4. Type KJRETS Temp. source .
*
Four kinds measurement inputer :
1. Pt-100 measure, 2. Voltage( DC mV) measure,
3. 400 ohm measure , 4. Type K/J/R/E/T/S Temp. measure .
* Portable instrument for calibrating process devices.
* Four kinds Source Adjustable output.
2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 30 mm.
LCD with white backlight ( ON/OFF ).
Measurement 1.Type k/J/R/E/T/S Temp. , 2. Voltage (DC mV).
3. Pt-100 ohm, 4. 400 ohm .
Source
Sampling Time Approx. 1 second. of Display Advanced setting
Type k/J/R/E/T/S Temp.,
Voltage (DC mV : - 10.00 to + 110.00 mV),
Pt-100 ohm , 400 ohm .
* Auto power OFF management.
*
Set unit or .
1
Operating 0 to 50 .
Instruction manual..............................................
Hard carrying case( CA-06 ).....................................................
( Type K source output line ), AC-DUK02.........................................
MTL-CP424.......................................................................
Temperature Less than 85% R.H. and Humidity Power Supply * Alkaline or heavy duty DC 1.5 V battery
( UM3, AA ) x 6 PCs, or equivalent.
* DC 9V adapter input. ( AC/DC power
adapter is optional ).
Power Current Normal operation ( w/o current source ) :
Approx. DC 18 mA. LCD Backlight is ON) : Approx. + DC 2 mA.
Weight Meter : 282 g/ 0.61 LB. (with cable) Dimension Meter : 198 x 68 x 45 mm Accessories * Included *
*
* Optional AC to DC 9V adapter. Accessories TP-01,TP-02A, TP-03, TP-04, TP-05
1 PC 1 PC 1 PC 1 PC
2
2-2 Electrical Specifications ( 23.0 ± 5℃ )
30 MHz only.
Thermocouple
Type
-200.0to 1370.0 >-100: >-100:
K
-328 to 2498 ±(0.05 %+ 1 ) ±(0.07%+1.5)0.1 or 1
-200.0 to 1200.0 <-100: <-100:
J
-328 to 2192 ±(0.05 %+ 2 ) ±(0.07%+ 2 )
-200.0 to 400.0
T
-328 to 752
-200.0 to 1000.0
E
-328 to 1832
R
S
* input impedence : 10 ^12 ohms. * measure protection : DC 60 V , AC 24 V.
* source protection : Reference CAUTION.
DC Voltage
DC Voltage
Range of source / measure Display Accuracy
-10.00mV to +110.00 mV 0.01 mV
* input impedence : 10 ^12 ohms. * measure protection : DC 60 V , AC 24 V. * source protection : Reference CAUTION.
0 to 1768
32 to 3214
source measure
<100: >100: ±(0.05 %+ 3 ) ±(0.07 %+ 3 )1 or 1 ≧100: 100: ±(0.05 %+ 2 ) ±(0.07 %+ 2 )
Resolution source / measure
Accuracy
± ( 0.05 % + 30 µV )
ResolutionRange of source / measure
Pt100
Pt100
Range of source / measure Display Accuracy
-200.0 to 850.0
-328.0 to 1562.0
* input impedence : 10 ^12 ohms. * protection : Reference CAUTION.
Resolution source measure
0.1 degree
0.1 degree
Remark :
The above specification are tested under the environment RF Field Strength less than 3 V/M & frequency less than the
3
±(0.05%+0.6)±(0.05%+ 1)
400 Ω
400 Ω
Range of source / measure Display Accuracy:
Resolution source measure
400.0 Ω 0.1 Ω
* input impedence : 10 ^12 ohms. * protection : Reference CAUTION. * FS : full scale
Remark :
The above specification are tested under the environment RF Field Strength less than 3 V/M & frequency less than the
30 MHz only.
±(0.1%+ 0.4Ω)
±(0.1%+ 0.6Ω)
CAUTION
* Do not apply a voltage exceeding the maximum input voltage.
otherwise the input part may be damaged. * Do not short_circuit or apply an external voltage to output terminals of the instrument or standard equipment, or else their internal circuitry may be damaged.
4
3. FRONT PANEL DESCRIPTION
3-11 DC 9V Power Adapter Input Socket
3-12 Battery Compartment/Cover
3-5 ▲ Button
3-6 ▼ Button
3-1 Display 3-2 ◄ Button( Backlight Button ) 3-3 Power Button 3-4 ► Button
3-7 FUNCTION Button/SET Button 3-8 Source/Measure Button / Type Button 3-9 Reset Button
3-10 Thermocouple Temp.Input Socket
Fig. 1
3-13 Battery Cover Screws 3-14 Tripod Fix Nut 3-15 Stand 3-16 Red(Hi) bananaTerminal 3-17 Black(Lo) bananaTerminal 3-18 Green(Lo) bananaTerminal
5
4. SOURCE PROVIDE PROCEDURE
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
M
odel
.
It is recommended to use the Thermocouple Temp.Input Socket (3-10 Fig 1)
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
M
odel
.
4-1. Thermocouple SOURCE PROVIDE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to SOURCE
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to Thermocouple Function.
4) Press and hold the " Source/Measure button " ( 3-7, Fig. 1 ) select to Thermocouple Type .
5) Press the " ► button "( 3-4, Fig. 1 )or " ◄ button"( 3-2, Fig. 1 ) select a digit for adjustment. The blinking digit identifies the digit selected.
6)
Press the " button "( 3-6, Fig. 1 ) or " button "( 3-5, Fig. 1 ) to adjust the value of the digit. Press and Hold the or button to rapidly adjust the value. And will generate the Temp. output same as the display value.
Remark :
output better , or Use The " Red banana Terminal " ( 3-16, Fig. 1 ) is for the positive Temp. output. The " Black banana Terminal " ( 3-17, Fig. 1 ) is the " ground " for negtive Temp. output.
4-2. VOLTAGE SOURCE PROVIDE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to SOURCE
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to Votage Function.
4) Press the " ► button "( 3-4, Fig. 1 )or " ◄ button"( 3-2, Fig. 1 ) select a digit for adjustment. The blinking digit identifies the digit selected.
5)
Press the " button "( 3-6, Fig. 1 ) or " button "( 3-5, Fig. 1 ) to adjust the value of the digit. Press and Hold the or button to rapidly adjust the value. And will generate the voltage output same as the display value.
Remark : Use The " Red banana Terminal" ( 3-16, Fig. 1 ) is for the positive Votage output. The " Black banana Terminal " ( 3-17, Fig. 1 ) is the " ground " for negtive Votage output.
6
4-3. RTD Pt100 ohm SOURCE PROVIDE PROCEDURE
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to SOURCE Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to Pt100 Function.
4) Press the " ► button "( 3-4, Fig. 1 )or " ◄ button"( 3-2, Fig. 1 ) select a digit for adjustment. The blinking digit identifies the digit selected.
5)
Press the " button "( 3-6, Fig. 1 ) or " button "( 3-5, Fig. 1 ) to adjust the value of the digit. Press and Hold the or button to rapidly adjust the value. And will generate the Pt100 ohm Temp. output same as the display value.
Fig. 1-A Fig. 1-B Remark : when carrying out calibration on a resistance temperature detector or a resistance measurement unit in three-wire connection, the supplide terminal adapter can be used to achieve a three-wire configuration (Fig. 1-A).2W Lo and LO terminals can be shorted (Fig. 1-B) Otherwise connect as Fig.1-A
4-4. 400 ohm SOURCE PROVIDE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to SOURCE
Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to 400 ohm Function.
4) Press the " ► button "( 3-4, Fig. 1 )or " ◄ button"( 3-2, Fig. 1 ) select
a digit for adjustment. The blinking digit identifies the digit selected.
5)
Press the " button "( 3-6, Fig. 1 ) or " button "( 3-5, Fig. 1 ) to adjust the value of the digit. Press and Hold the or button to rapidly adjust the value. And will generate the 400 ohm. output same as the display value.
7
5. MEASUREMENT PROCEDURE
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
Type k temp
.
value.
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
V
oltage (mV)
, value.
5-1. Thermocouple Type MEASURE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to
MEASURE Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to Thermocouple Type
Measure Function.
4) Press and hold the " SOURCE/MESURE button " ( 3-8, Fig. 1 ) to select
Thermocouple Type .
5) Thermocouple Type Temp. probe into the "Thermocouple Type Temp. socket "
( 3-10, Fig. 1 ) .
The meter LCD display will show the measure value same as the input
6)
5-2. VOLTAGE MEASURE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to
MEASURE Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to Voltage
Measure Function.
4) Voltage (mV) input to the "RED banana Terminal " (3-16 ,Fig.1 ) and
"BLACK banana Terminal " (3-17 ,Fig.1) .
The meter LCD display will show the measure value same as the input
5)
Remark : Use The " Red banana Terminal " ( 3-16, Fig. 1 ) is for the positive Votage input. The " Black banana Terminal" ( 3-17, Fig. 1 ) is the " ground " for negtive Votage input.
8
5-3. RTD Pt100 ohm MEASURE PROCEDURE
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
Power on the meter by press
the " Power
button
" ( 3-
3,Fig. 1 ) > 2 sec.
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to
MEASURE Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to RTD Pt100 ohm
Measure Function.
4) Pt100 ohm Temp. probe into the Red (HI) banana Terminal(3-16,Fig1) and
Black (LO) banana Terminal(3-17,Fig1) .
The meter LCD display will show the measure value same as the input
5)
Pt 100 temp. value.
5-4. 400 ohm MEASURE PROCEDURE
1)
2) Press the " Source/Measure button " ( 3-8, Fig. 1 ) select to
MEASURE Model .
3) Press the " FUNCTION button " ( 3-7, Fig. 1 ) select to 400 ohm
Measure Function.
4) Measured resistance into the Red (HI) banana Terminal (3-16,Fig1) and
Black (LO) banana Terminal(3-17,Fig1) .
The meter LCD display will show the measure value same as the input
5)
Measured resistance, value.
5-5. LCD Backlight Operation
During the meter Power on
long press The" ◄ button"( 3-2, Fig. 1 ) > 2 sec. the LCD Backlight will be on ,and long press The" ◄ button"( 3-2, Fig. 1 ) > 2 sec. the LCD Backlight will be off.
9
6. ADVANCED SETTING
POFF.......
Under do not execute the Datalogger function, press the " Function Button " ( 3-7, Fig. 1 ) continuously at least 3 seconds will enter the " Advanced Setting " mode. Then press the " Function Button " (3-7, Fig. 1 ) once a while in sequence to select the two main function, the lower display will show :
Auto power OFF management
t-CF.......
Remark : During execute the " Advanced Setting " function, if press " Power Button " ( 3-3, Fig. 1 ) once will exit the " Advanced Setting " function, the LCD will return to normal screen.
6-1 Auto power OFF management
When the lower display show " PoFF "
Select the Temp. unit to or
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to select the upper value to " yES " or " no ".
yES - Auto Power Off management will enable. no - Auto Power Off management will disable.
2) After select the upper text to " yES " or " no ", press the
" Source/Measure Button " ( 3-8, Fig. 1 ) will save the setting function with default.
10
6-2 Select the Temp. unit to ℃ or
When the lower display show " t-CF "
1) Use the " ▲ Button " ( 3-5, Fig. 1 ) or " ▼ Button "
( 3-6, Fig. 1 ) to select the upper Display text to " C " or " F ".
C - Temperature unit is F - Temperature unit is
2) After Display unit is selected to " C " or " F ", press the
" Source/Measure " ( 3-8, Fig. 1 ) will save the setting function with default.
7. POWER SUPPLY from DC ADAPTER
The meter also can supply the power supply from the DC 9V Power Adapter ( optional ). Insert the plug of Power Adapter into " DC 9V Power Adapter Input Socket " ( 3-11, Fig. 1 ). The meter will auto select to DC 9V Power Adapter.
11
8. BATTERY REPLACEMENT
1) When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
2) Loose the screws of the " Battery Cover Screws " ( 3-13,
Fig. 1 ) and take away the " Battery Cover " ( 3-12, Fig. 1 ) from the instrument and remove the battery.
3) Replace with DC 1.5 V battery ( UM3, AA,
Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
4) Make sure the battery cover is secured after changing
the battery.
9. SYSTEM RESET
If the meter happen the troubles such as :
CPU system is hold ( for example, the key button can not be operated... ).
Then make the system RESET will fix the problem. The system RESET procedures will be either following
method :
During the power on, use a pin to press the " Reset Button " ( 3-9, Fig. 1 ) once a while will reset the circuit system.
12 171016-TC-424