Lutron Electronics PH-321M Operation Manual

pH MONITOR
Model : PH-321M
L11'i.u1ron PH-321 M
'I
ENTER
,,,,,
I.LI LI
••
PH METER:
Your purchase of this
pH
monitor marks a step forward for you into the
field of
precision
measurement. Although
this
pH monitor is a complex and delicate instrument, its durable structure will allow many years of use if proper
operating techniques are developed. Please read the following instructions
carefully and always keep this manual within easy
reach.
'
UKAS
Q
UA
L
I
T
Y
,
\
\
A
N
A
G
C
M
(
N
T
005
Your purchase of this pH monitor marks a step forward for you into
the eld of precision measurement. Although
this pH monitor is a
complex and delicate instrument, its durable structure will allow many
years of use if proper
operating techniques are developed. Please read the following instructions carefully and always keep this manual within
easy reach.
2 3
TABLE OF CONTENTS
FEATURES .......................................................................... 3
SPECIFICATIONS .............................................................. 3
FRONT PANEL DESCRIPTIONS ...................................... 5
Display ............................................................................... 5
CAL (Enter) Button ............................................................. 5
ESC Button ........................................................................ 5
▼ Button ............................................................................ 5
pH Electrode BNC Socket ................................................. 5
DC 9V Input Socket ........................................................... 5
Stick Sponge Fix Position .................................................. 5
Hanging Hall ...................................................................... 5
pH CALIBRATING PROCEDURE ........................................ 6
Calibrating Considration .................................................... 6
Requiring Equipment for Calibration .................................. 6
Calibration Procedures ...................................................... 6
Calibration Clear ................................................................ 9
MEASURING PROCEDURES ......................................... 10
OPTIONAL ACCESSORIES .............................................11
3
FEATURES
• Economical cost digital pH meter General purpose applications. Easy operation.
• Back light LCD display.
• Auto calibration.
• Compact size, light weight.
• Avaialble for wide applications, such as AQUARIUM,
BEVERAGE, FISH HATCHERIES, FOOD PROCESSING, PHOTOGRAPHY, LABORATORY, QUALITY CONTROL, SCHOOL & COLLEGES, SWIMMING POOLS, WATER CONDITONS.
SPECIFICATIONS
Display LCD size : 31 mm x 31 mm.
LCD with back light, easy read.
Measurement
Range
0 to 14 pH x 0.01 pH
Resolution 0.01 pH Accuracy ± 0.07 pH (pH 5 - pH 9).
± 0.1 pH (pH 4 - pH 10). ± 0.2 pH (pH 1 - pH 3.9, pH 10.1 - pH 13). * Main instrument only
Input Impedance
10^12 ohms.
Temperature
Compensation
Not necessary to make adjustment.
Sampling Time Approx. 1 second. Operating
Temperature
0°C to 50°C (32°F to 122°F)
Operating
Humidity
Less than 80% RH.
4 5
Calibration Automatically, calibrated by front buttons.
* pH 7 calibration. * pH 4 calibration.
* pH 10 calibration Power Supply DC 9V via optional external adapter Power
Consumption
Approx. 11 mA
Dimension 89mm x 57mm x 31 mm,
(3.5 in x 2.2 in x 1.2 in). Weight 117 g / 0.26 LB. pH electrode
(Optional)
Any connector pH Electrode with
BNC connector.
Standard Accessories
Instruction Manual - 1 PC
Stick sponges - 2 PC’s Optional
Accessories
pH electrodes :
pH electrode, pH electrode,
Standard bu󰀨er solutions :
pH 7.0 bu󰀨er solutions, pH 4.0 bu󰀨er solutions,
DC 9V adapter, AP-9VA
* Input voltage : 100 to 240 ACV,
50/60 Hz.
* Output voltage : Regulation DC 9V/1
Amp. rating max.
* Output plug : round 2.5 mm dia. plug.
plug converter, AP-GTU
* Convert the input plug of AP-9VA from
the Germany type to USA type
* Appearance and specifications listed in this brochure are
subject to change without notice.
5
FRONT PANEL DESCRIPTION
1. Display
2. CAL (Enter) button
3. ESC button
4. ▼ button
5. pH Electrode BNC Socket
6. DC 9V Input Socket
7. Stick Sponge Fix Position
8. Hanging Holes
1.
2.
3.
4.
5.
5.
6.
8.
7.
6 7
pH CALIBRATING PROCEDURE
Calibrating Consideration
These pH meter already calibrated by mV signal that simu­lated from the ideal pH Electrode mV output base on 25°C
environment. However due to
a. An ideal electrode will produce 0 mV at pH 7.00, but
most elctrodes are slightly o.
b. The measuring environment temperature may not near
25°C (15 to 35°C).
It is necessary to make the following calibration prpcedures if the user need to keep instrument combined electrode
within accuracy.
Requiring Equipment for Calibration
1. Combination pH Electrode (optional).
2. Two bu󰀨er solutions (optional): pH 7.00 & pH 4.00 (or pH 10.00).
Calibration Procedures
pH 7.00 Calibration
1. Rinse the electrode in distilled water completely.
2. Connect the pH Electrode to the “BNC Socket” and place electrode onto standard bu󰀨er pH 7.00 solution.
3. Power on the instrument by plugging in the DC 9V adapter in the “DC 9V Input Socket”.
4. Press “CAL button” once, the display will show:
7
5. Press “ENTER button” once, the display will ashing several time, the show 7.00 (if show 6.99, 7.01 it is normal), now nish the pH 7.0 calibration procedures.
pH 4.00 Calibration
1. Rinse the electrode in distilled water completely.
2. Connect the pH Electrode to the “BNC Socket” and place electrode onto standard bu󰀨er pH 4.00 solution.
3. Power on the instrument by plugging in the DC 9V adapter in the “DC 9V Input Socket”.
4. Press “CAL button” once, the display will show:
Press “▼button” once, the display will show:
5. Press “ENTER button” once, the display will ashing several time, the show 4.00 (or 4.00 ± 0.01), now nish the pH 4.0 calibration procedures.
8 9
pH 10.00 Calibration
1. Rinse the electrode in distilled water completely.
2. Connect the pH Electrode to the “BNC Socket” and place electrode onto standard bu󰀨er pH 10.00 solution.
3. Power on the instrument by plugging in the DC 9V adapter in the “DC 9V Input Socket”.
4. Press “CAL button” once, the display will show:
Press “▼button” once a while with two times, the display will show:
5. Press “ENTER button” once, the display will ashing several time, the show 10.00 (or 4.00 ± 0.01), now nish the pH 10.0 calibration procedures.
Calibration Consideration
• During make the Calibration if intend to exit the calibration
procedures, then just press “ESC button” will return to normal screen.
• The calibration procedures should be executed two points
calibration (pH 7 & pH 4 or pH 7 & pH 10) and start from the pH7 at rst, then pH4 (or pH10).
9
• During the calibration, the circuit will sensing solution value
automatically. If the bu󰀨er solution value already beyond 7 ± 2 pH. 4 ± 2 pH, 10 ± 2 pH, the LCD will show “Err”.
• Rinse the Electrode in distilled water before change the
bu󰀨er solution.
• Repeat above calibration procedures two times at least
if possible.
• Rinse the electrode in distilled water after execute the
calibration procedures.
Calibration Clear
During execute the calibration procedures, if happen something wrong (display Error information), the execute the following “Calibration Clear” procedures will clear all the existing calibration data and return to default value.
However after execute the “Calibration Clear” then make the calibration again. If the display still show “Err” (error information), then the “Standard solution” or the “pH electrode” may existing problem.
1. Power on the instrument by plugging in the DC 9V adapter into the “DC 9V Input socket”
2. Press “CAL button” once, the display will show:
10 11
Press “▼button” once a while with three times, the display will show:
Press “ENTER button” once, the display will show following screen once a while,
Then return to normal screen and nish the “Calibration Clear” procedures.
MEASUREMENT PROCEDURES
After the instrument and pH electrode are calibrated, then the unit is now ready for measuring.
1. Connect the combination pH Electrode to the “BNC socket”.
2. Power on the instrument by plugging in the DC 9V adapter into the “DC 9v Input socket”.
3. Place the electrode into the measured solution, then the instrument will display the pH value.
4. After make the measurement, please rinse the electrode in
distilled water.
11
OPTIONAL ACCESSORIES
pH Electrode PE-03
General purpose pH electrode with BNC connector. Range: 1 to 13 pH (typical 0 to 14 pH). Epoxy body, Body size: 12 mm dia. x 160 mm. Cable length: 1 meter.
pH Electrode PE-11
General purpose & high quality pH electrode with BNC connector. Range: 1 to 13 pH (typical 0 to 14 pH). Epoxy body, Body size: 9.5 mm dia. x 120 mm. Cable length: 1 meter.
Spear Tip pH Electrode PE-06HD
The “Spear Tip ph Electrode” is perfect for those pH measurements in applications where sample piercing is required. Meat, sausage and cheese ideal applications. The electrode features a very durable glass measuring spear packaged in a rugged virtually unbreakable epoxy body. Range: 1 to 13 pH (typical 0 to 14 pH).
12 PB
DC 9V Power Adapter AP-9VA
Input voltage: 100 to 240 ACV, 50/60 Hz. Output voltage: Regulation DC 9V/1 amp. rating max Output plug: round 2.5 mm dia. plug.
Plug Converter AP-GTU
Convert the input of AP-9VA, AP-9VB from the Germany type
to the USA type.
Plug Converter AP-GTE
Convert the input plug of AP-9VA, AP-9VB from Germany type to the UK (England) type.