Lutron Electronics SOIL PH METER, PH-220S Operation Manual

Pen type, separate electrode
SOIL PH METER
Model : PH-220S
Y
our purchase of this SOIL PH METER marks a step forward for you into the field o
f
precision measurement.
A
lthough this METER is a complex and delicate instrument, its durable structure will allow many years of use i
f
proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
OPERATION MANUAL
TABLE OF CONTENTS
1. FEATURES................................................................1
2. SPECIFICATIONS......................................................1
3. FRONT PANEL DESCRIPTIONS...................................3
3-1 Display............................................................... 3
3-2 Power button ( button )................................. 3
3-3 Hold button ...................................................... 3
3-4 Display reverse button ( button ).................... 3
3-5 pH electrode BNC socket..................................... 3
3-6 BNC protection cover...........................................3
3-7 Battery compartment/Cover................................ 3
3-8 pH electrode handle........................................... 3
3-9 pH electrode sensing head................................... 3
3-10 pH electrode BNC plug...................................... 3
3-11 pH electrode protection bottle............................
.
3
4. pH CALIBRATING PROCEDURE..................................
.
4
4-1 Calibrating Consideration.....................................
.
4
4-2 Requiring Equipment for Calibration......................4
4-3 Calibration procedures....................................... 4
5. MEASURING PROCEDURE..........................................5
6. REPLACEMENT OF BATTERY......................................5
7. OPTIONAL ACCESSORIES..........................................6
1. FEATURES
* Complete set with pen pH meter, soil pH electrode,
pH 4.0 /pH 7.0 standard buffer solution.
* Spear Tip pH electrode ( included ), ideal for
penetration pH measurement into soil, meats, cheese and other type of measurements requiring sample penetration. The electrode features a very durable glass measuring spear packaged in a rugged virtually unbreakable epoxy body.
* Separate soil pH electrode, easy for general purpose
usage and remote measurement.
* Auto calibration for pH 4, pH 7 or pH 10. * Build in reverse display button. * Data hold. * Auto power off. * Compact size, light weight. * Optional carrying case. * Water resistance on the front panel. * Application : Agriculture, Gardening, Food
industrial, Education, School, Colleges, Laboratory, Industrial and Quality control.
2. SPECIFICATIONS
2-1 Meter
Display LCD, size : 20 mm x 28 mm.
* User can reverse the display direction
by push button.
Measurement 0 to 14 pH x 0.01 pH Range Resolution 0.01 pH
1
Accuracy ± 0.07 pH(pH5 to pH 9)
± 0.1 pH(pH4 to pH4.9, pH9.1 to pH10) ± 0.2 pH(pH1 to pH3.9, pH10.1 to pH13)
* 23± 5 * After calibration.
Input 10^12 ohms. Impedance Temperature Not necessary to make adjustment. Compensation Sampling Approx. 0.8 second. Time Hold Function To freeze the display reading value. Calibration Auto calibration, method pH 7, pH 4 ( or pH 10 ) pH electrode BNC connector. connector Operating 0 to 50 ( 32 to 122 )℃℉ Temperature Operating Less than 80% RH. Humidity Power Supply DC 1.5V battery ( UM-4/AAA ) x 4 PCs. Power Approx. 4.8 mA. Consumption Dimension Meter :180 x 40 x 32 mm ( 7.1 x 1.6 x 1.3 inch ).
Electrode : Body size : 12 mm dia. x 160 mm. Electrode cable length : approx.. 1000 mm.
Weight 220 g/0.48 LB (included electrode).
2
Standard Instruction Manual......................... 1 PC.
Accessories Soil pH electrode, PE-06HD............. 1 PC.
pH 4.0 buffer solution, PH-04A.......
.
1 PC.
pH 7.0 buffer solution, PH-07A.......
.
1 PC. Optional * Soft carrying case with sash Accessories ( 210 x 110 x 55 mm ),
Model : CA-05A
* Hard carrying case
( 280 x 195 x 65 mm ), Model : CA-06
2-1 Soil pH Electrode ( included )
pH Measuring 1 to 13 pH ( typical 0 to 14 pH ). Range pH Operation 5 to 60 ( 41 to 140 ).℃℉ Temperature pH Electrode Combination type. Structure Reference Ag/AgCl, Double Junction. of pH Electrode Zero Potential 7± 1 pH. for pH Value Repeatability 0.05 pH. Response time 2 minutes. Body Material Epoxy. Connector BNC. Mechanical With protection bottle ( soaker Protection bottle ) on the electrode head. Weight 68 g ( including cable & socket ). Dimensions Body length - 160 mm.
Body Dia. - 12 mm. Cable length - 1000 mm.
3
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-2 Power button ( button ) 3-3 Hold button 3-4 Display reverse button ( button ) 3-5 pH electrode BNC socket 3-6 BNC protection cover 3-7 Battery compartment/Cover 3-8 pH electrode handle 3-9 pH electrode sensing head 3-10 pH electrode BNC plug 3-11 pH electrode protection bottle
4
4. pH CALIBRATING PROCEDURE
4-1 Calibrating Consideration
The pH meter already calibrated by mV signal that simulated from the ideal pH ELECTRODE mV output ( base on 25 environment ). However due to (a) An ideal electrode will produce 0 mV at pH 7.00, but most electrodes are slightly off. (b) The measuring environment temperature may not near 25 ( 15 to 35 ). ℃℃ If intend to keep instrument combined electrode within high accuracy, it is necessary to make the following calibration procedures ( 4-3 ).
4-2 Requiring Equipment for Calibration
1)Combination soil pH electrode ( PE-06HD, included ).
2)Buffer solutions :
pH 7.00 buffer solution ( PH-07A, included ). pH 4.00 buffer solution ( PH-04A, included ).
4-3 Calibration procedures
pH 7 calibration
1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC
socket ( 3-5, Fig. 1 ) and place electrode into pH 7.00 buffer solution ( PH-07A ).
2)Power on the instrument by pressing the " Power button "
( 3-2, Fig. 1 ).
3)Press the " Hold button " ( 3-3, Fig. 1 ) once.
the display will show the " HOLD " symbol. Press the " button " ( 3-3, Fig. 1 ) once, the display will show following screen in sequence.
5
screen 1
CAL flashing
screen 2
7.00
flash CAL
screen 3
7.00 flashing
CAL
screen 4
End
screen 5 ( return to normal screen )
7.00
6
pH 4 calibration
1)Rinse the electrode in distilled water.
Place electrode into buffer pH 4.00 solution.
2)All the calibration procedures are same as the
above pH calibration, except : a. Change the buffer solution to pH 4.00 ( PH-04A ). b. The calibration screen value will change to 7.00 instead of 4.00
Remark :
* Except calibrate pH 7, pH 4, meter also can calibrate
pH 10.
* The calibration procedures should start from the pH 7 at
first, then pH 4 ( or pH 10 ).
* During the calibration, the circuit will sensing solution
value automatically. If the buffer solution value already beyond 7± 1 pH. 4± 1 pH, 10± 1 pH, the LCD will show " Err ".
* When the LCD show the screen 2, it allow the user to
adjust another calibration value precisely ( for example
7.01, 6.99... ) by pressing the " button " ( 3-2, Fig. 1 ) , " button " ( 3-4, Fig. 1 )'
* Rinse the electrode in distilled water before change
the buffer solution.
* Repeat above pH, pH calibration procedures two
times at least.
* After calibration, the " Electrode sensing head "
( 3-9, Fig. 1 ) should insert into the " Protection bottle " ( 3-11, Fig. 1 ).
*
PE-06HD, PH-07A, PH-04A is the consumer accessory.
7
5. MEASURING PROCEDURE
After the meter along the pH electrode is calibrated ( section 4-3 ) , the unit is ready for measuring.
1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC
socket ( 3-5, Fig. 1 ) and place electrode into the measured solution.
2)Power on the meter by pressing the " Power button "
( 3-2, Fig. 1 ).
3)Place the electrode into the measured soil or other
type penetration material ( such as cheese, meats... ) or solution, then the instrument will display the pH value.
* Electrode ( attached ) is the
rough and durable penetration electrode. However user should operate the electrode very carefully.
* The electrode is the consumer
accessory, after the electrode is used, if the damage of the Electrode Sensing Head ( Spear Tip Head ), it is without warranty.
* After make the measurement, please rinse the
electrode in distilled water.
* After calibration, the " Electrode sensing head "
( 3-9, Fig. 1 ) should insert into the " Protection bottle " ( 3-11, Fig. 1 ).
* PE-06HD, PH-07A, PH-04A is the consumer accessory.
Data hold
During the measurement, press the " Hold button " ( 3-3, Fig. 1 ) once will hold the measured value & the LCD will display a " HOLD " symbol.
* Press the " Hold Button " once again will release the data hold function.
8
Display reverse
During the measurement, press the " Display reverse button " ( 3-4, Fig. 1 ) once will reverse the display direction for the user convenience.
* Press the " Display reverse button " once again will
return to the normal display.
Auto power off
The meter is built the " Auto power shut off " function to save battery life. If any function buttons are not pushed within 10 minutes, the power will be off automatically.
6. REPLACEMENT OF BATTERY
1)When the LCD display show " " ,
It is necessary to replace the battery. However, in-spec measurement may still be made for several hours after Low Battery Indicator appears before the instruments instrument become inaccurate.
2)To replace the battery, rotate and remove the "
Battery Cover " ( 3-7, Fig. 1 ), take out the old batteries, install new batteries :
DC 1.5V battery ( UM-4/AAA ) x 4 PCs.
3)When install the batteries, should make attention the
battery polarity.
4)After install the batteries, reinstall the battery cover
again.
9
7. OPTIONAL ACCESSORIES
SPEAR TIP pH ELECTRODE Model : PE-06HD
Spear tip pH electrode is ideal for penetration pH measurements into meats, cheese, soil and other types of measurements requiring sample penetration.
pH 7.00 buffer solution
Model : PH-07A
pH 4.00 buffer solution
Model : PH-04A
CARRYING CASE Model : CA-06
Hard carrying case. ( 280 x195 x 65 mm ).
CARRYING CASE Model : CA-05A
Vinyl soft case. ( 260 x110 x 55 mm ).
Hard carrying case. Soft carrying case. Model : CA-06 Model : CA-05A 280 x195 x 65 mm 260 x110 x 55 mm
10
0701-PH220S