Lutron LCR-9184 User Manual

100 Hz 120 Hz 1 KHz 10 KHz 100 KHz, Professional╱╱╱╱
Y
LCR METER
Model : LCR-9184
our purchase of this LCR METER marks a step forward for you into the field of precision measurement. Although this LCR METER is a complex and delicate instrument, its durable structure will allow many years of use if proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
OPERATION MANUAL
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
TABLE OF CONTENTS
1. FEATURES..................................................................1
2. SPECIFICATIONS........................................................1
2-1 General Specifications...........................................
2-2 Electrical Specifications..........................................3
3. FRONT PANEL DESCRIPTION......................................
3-1 Display................................................................. 6
3-2 Frequency Button..................................................6
3-3 Hold Button..........................................................
3-4 L/C/R Button.........................................................6
3-5 CAL Button...........................................................
3-6 Button, RS232 Button.......................................
6
3-7 < Button, D/Q/θ Button.......................................
3-8 REL/%/ Button..................................................
3-9 > Button, SER/PAL Button.................................... 6
3-10 Power Button..................................................... 6
3-11 Input terminals ( pin terminals )........................... 6
3-12 Input terminals ( banana terminals )....................
3-13 DC 9V Power Adapter Input Socket......................
3-14 RS-232/USB Output Terminal............................... 6
3-15 Stand.................................................................
3-16 Battery Cover's Screws........................................
3-17 Battery Compartment/Cover.................................6
3-18 LCD backlight Button........................................... 6
3-19 Enter Button......................................................
3-20 Sorting Button....................................................6
3-21 Setup Button......................................................6
4. MEASURING PROCEDURE...........................................
4-1 Measuring procedure............................................. 7
4-2 Data hold.............................................................
4-3 Serial/Parallel........................................................8
4-4 The select of test frequency...................................9
4-5 REL /%................................................................
4-6 Sorting.................................................................
4-7 Calibration............................................................ 10
4-8 Backlight..............................................................
4-9 RS232 output........................................................10
5. POWER SUPPLY from DC ADAPTER..............................11
6. BATTERY REPLACEMENT.............................................11
7. OPTIONAL ACCESSORIES............................................11
6
1
6
6
6
6
6 6
6 6
6
7
8
9 9
10
1. FEATURES
* 19,999/1,999 counts dual LCD display. * AutoLCR smart check and measurement. * Serial/Parallel modes are selectable. * Ls/Lp/Cs/Cp with D/Q/θ /ESR parameters. * Support DCR mode 1.00 to 200.0 MΩ . * Five different test frequency are available :
100 Hz/120 Hz/1 KHz/10 KHz/100 KHz.
* Test AC signal level : 0.6 V rms typically. * Test range : ( ex. F = 1 KHz )
L : 200.00 uH to 2000.0 H C : 2000.0 pF to 2.000 mF R : 20.000 to 200.0 M
* Multi-level battery detector. * LCD with green light backlighteasy reading. * RS232/USB PC Computer interface. * Can default auto power off.
2. SPECIFICATIONS
2-1 General Specifications
Display LCD size66.8 X 52.8 mm.
LCD with green backlight ( ON /OFF ) Test frequency 100 Hz/120 Hz/1 KHz/10 KHz/100 KHz Function L/C/R Function selector
Frequency selector
D/Q/θ /ESR selector
Sorting mode selector
Backlight Dissipation 0.000 to 1999 factor Quality factor 0.000 to 1999
1
θ ± 90°
A
measurement Sorting ±0.25%, ±0.5%, ±1%, ±2%, ±5% tolerance ± 10%, ± 20%, +80% -20% mode Calibration Open/Short calibration Data Hold Freeze the display reading Data output RS232/USB PC computer interface Power off Auto shut off saves battery life or
manual off by push button Operating 0 to 50℃℃ temperature Operating Less then 85% R.H. humidity Power Supply 006P DC 9V battery
* Alkaline or Heavy duty type
DC 9V adapter input
* AC/DC power adapter is optional.
Power DC 35 mA approximately consumption Dimension 193 x 88 x 41mm Weight 385 g
* meter only
Standard * Instruction manual....................... 1 PC
Accessories * Included Optional SMD tester, SMDA-22 Accessories SMD test clip, SMDC-21
lligator clips................................1 Pair
Holster, HS-03
AC to DC 9V adapter
Hard carrying case, CA-06
Soft carrying case, CA-05A
2
2-2 Electrical Specifications (23± 5 )
Resistance ( DCR )
Range Accuracy Remark
20 ± ( 0.5% + 5d ) After calibration 200 ± ( 0.5% + 5d ) 2 K ± ( 0.5% + 5d ) 20 K ± ( 0.5% + 5d ) 200 K ± ( 0.5% + 5d ) 2 M ± ( 0.5% + 5d ) After calibration 20 M ± ( 1% + 5d ) After calibration 200 M ± ( 2% + 5d ) After calibration
Resistance ( Rp/Rs )
Range Accuracy Accuracy Remark
100 Hz/120 Hz 1000 Hz 20 ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 200 ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2 M ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 20 M ± ( 1% + 5d ) ± ( 2% + 5d ) After calibration 200 M ± ( 2% + 5d ) ± ( 5% + 5d ) After calibration
Range Accuracy Accuracy Remark
10 KHz 100 KHz 20 ± ( 1% + 5d ) ± ( 2% + 5d ) After calibration 200 ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 K ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2 M ± ( 1% + 5d ) ± ( 2% + 5d ) After calibration 20 M ± ( 2% + 5d ) ------------------- After calibration
3
Remark :
y.
* If the impedance is larger than 10 K, Rp is shown on the displa * If the impedance is less than 10 K, Rs is shown on the display. * If intend to obtain the accurate value, please test the component
into the " Pin terminals " ( 3-11, Fig. 1 ) or tested via optional SMD tester, SMDA-22 or SMD test clip, SMDC-21.
Capacitance ( Cp/Cs ) : D 0.1
Range Accuracy Accuracy Remark
100 Hz/120 Hz 1000 Hz 20 pF ± ( 2% + 5d ) ± ( 1% + 5d ) After calibration 200 pF ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 2000 pF ± ( 0.8% + 5d ) ± ( 0.8% + 5d ) After calibration 20 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2000 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 uF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 uF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) After calibration 2000 uF ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 20 mF ± ( 2% + 5d ) ------------------- After calibration
Range Accuracy Accuracy Remark
10 KHz 100 KHz 20 pF ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 200 pF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) After calibration 2000 pF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) After calibration 20 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2000 nF ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 uF ± ( 0.8% + 5d ) ± ( 0.8% + 5d ) 200 uF ± ( 1% + 5d ) ------------------- After calibration
4
Remark :
.
.
* If the impedance is larger than 10 K, Cp is shown on the display * If the impedance is less than 10 K, Cs is shown on the display. * If intend to obtain the accurate value, please test the component
into the " Pin terminals " ( 3-11, Fig. 1 ) or tested via optional SMD tester, SMDA-22 or SMD test clip, SMDC-21.
Inductance ( Lp/Ls ) : D 0.1
Range Accuracy Accuracy Remark
100 Hz/120 Hz 1000 Hz 20 uH ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 200 uH ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 2000 uH ± ( 0.8% + 5d ) ± ( 0.8% + 5d ) 20 mH ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 mH ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 2000 mH ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 H ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 H ± ( 0.5% + 5d ) ± ( 0.8% + 5d ) 2000 H ± ( 1% + 5d ) ------------------- After calibration
Range Accuracy Accuracy Remark
10 KHz 100 KHz 20 uH ± ( 1% + 5d ) ± ( 1% + 5d ) After calibration 200 uH ± ( 0.8% + 5d ) ± ( 0.8% + 5d ) After calibration 2000 uH ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 20 mH ± ( 0.5% + 5d ) ± ( 0.5% + 5d ) 200 mH ± ( 0.5% + 5d ) ------------------- 2000 mH ± ( 0.5% + 5d ) -------------------
Remark :
* If the impedance is larger than 10 K, Lp is shown on the display * If the impedance is less than 10 K, Ls is shown on the display. * If intend to obtain the accurate value, please test the component
into the " Pin terminals " ( 3-11, Fig. 1 ) or tested via optional SMD tester, SMDA-22 or SMD test clip, SMDC-21.
5
3. FRONT PANEL DESCRIPTION
ld
/Q/θ
/%/
/
ls (
)
)
l
d
/
backligh
g
3-1 Display
3-2 Frequency Button
3-3 Ho
3-4 L/C/R Button
3-5 CAL Button
3-6 Button, RS232 Button
3-7 < Button, D
3-8 REL
3-9 > Button, SER
3-10 Power Button
3-11 Input termina
3-12 Input terminals ( banana terminals
3-13 DC 9V Power Adapter Input Socket
3-14 RS-232/USB Output Termina
3-15 Stan
3-16 Battery Cover's Screws
3-17 Battery Compartment
3-18 LCD
3-19 Enter Button
3-20 Sortin
3-21 Setup Button
Button
Button
Button
PAL Button
pin terminals
Cover
t Button
Button
6
4. MEASURING PROCEDURE
4-1 Measuring procedure
1)Power on by pressing the " power on/off key ", the all
segments of LCD will be ON for 2 seconds
2)The default mode is AUTO LCR smart mode and the
default test frequency is 1 KHz.
3)In order to extend the battery lifeit is recommended
to use the external power supply and the APO setting ( Auto power off setting ).When all function button do not be push within 5 minutesthe system will launch the alarm buzzer beep at three times before the power-off state
4)When AUTO/L/C/R function selection button ( L/C/R ) is
pushedthe main test mode could be selected sequentially
Auto-LCR mode Auto-L mode→→ Auto-C mode Auto-R mode→→ DCR mode Auto-LCR mode.
The default test mode is Auto LCR mode which could check the type of impedance smartly and enter to the L/C/R measurement mode automatically﹒The secondary parameter will follow the L/C/R measurement
When Auto-L or Auto-C mode is selected, the impedance measurement is auto ranging.
The primary LCD display will show the inductance or capacitance of DUT ( device under testing ). The secondary LCD display will show the quality ( Q ) or dissipation ( D ) factor.
7
The phase angel ( θ ) or equivalent resistance ( ESR ) can also be shown by pushing the PARAMETER ( D/Q/θ ) keypad to choose D/Q/θ /ESRWhen Auto-R(ACR mode) or DCR mode is selectedthe secondary parameter is omitted
Note 1 : When Auto LCR mode is activethe secondary parameter will show the equivalent resistance in parallel mode ( Rp ) to replace the D factor if the C measured value of DUT ( device under testing ) is less then 5 pF
Note 2 : At Auto LCR mode onlyDuring Auto R mode or DCR modethe secondary parameter is not available
4-2 Data hold
During the measurementpush the HOLD key to stop the reading of DUT ( device under testing ) on primary display. Push the HOLD key again to cancel the hold mode and original measurement mode
4-3 Serial/Parallel
When any L/C/R function mode is selected, the default measurement in serial or parallel mode is auto selected and the AUTO segment will be shown on LCD display. It depends on the total equivalent impedance measured. If the impedance is larger than 10KΩ , parallel mode will be set and Lp/Cp/Rp will be shown on the display. If it is less than 10K, serial mode will be set and Ls/Cs/Rs will be shown on the display. When SER/PAL key is pushed, the impedance measurement will be set in serial mode or in parallel
8
4-4 The select of test frequency
When FREQ key is pushed, the test frequency will be changed sequentially. There are five different test frequencies ( 100Hz/120Hz/1KHz/10KHz/100KHz ) can be selectedThe LCR impedance scale range are depended on the test frequency, See the specifications.
4-5 REL /%
The REL /% mode help the user to make a relative value of the component . At first, select the function to inductors, capacitors, resistors ( push L/C/R button ), then put under test DUT ( device under testing ) side, LCD will be shown the testing value. Until tested value is stable, push REL button, LCD will appear and 0.0%. This testing is the first analytic value as a bench mark, the next DUT are compared with the first percentage. Then push the REL/% button, the indicator will be flashing, LCD will be shown the first value and the comparison percentage.
* Press REL/% button 2 seconds continuously will
exit this function.
4-6 Sorting
The sorting mode could help the user to make a quick sort for a bunch of components. Push Sort key to the sorting mode which will be set to 2000 digits display automatically. If the LCD reading is OL or less than 200 counts, the SORT key is not available. The primary display to show PASS or FAIL status depends on whether the impedance measured exceeds tolerance range. The current measurement result will be shown on secondary display. When sorting mode is active, push SETUP keypad to modify the reference value, range and the tolerance setting sequentiallyIf the target is reached, push ENTER keypad to confirm it. Use the direction keypads(
///)to change the target data easily. The↑↓←→
reference value setting is available from 20 to 1999 counts.
9
The tolerance range setting selection :
± 0.25% ± 0.5% ± 1% ± 2% ± 5% →→→
± 10% ± 20% +80% -20%→→ →
The default tolerance is ± 1%
* Press Sorting button once will exit this function.
4-7 Calibration
In order to improve the accuracy for high/low impedance, it is recommended to do OPEN/SHORT calibration mode before the measurement. Push CAL key larger than 2 seconds to start the open/short calibration procedure :
OPEN ready OPEN calibration SHORT ready→→
SHORT calibration.
During open or short calibration processing, the 30-second countdown will be shown on LCD panels. If the calibration procedure is finished, the PASS or FAIL symbol will be shown on the primary display. If the PASS symbol appear on for both OPEN and SHORT modes, the calibration data will be saved to external EEPROM after push CAL key again
4-8 Backlight
When user push BKLIT button ( LCD backlight Button ), the backlight driver will be activePush the BKLIT button again to disable the backlight driver
4-9 RS232 output
When you want to send the measurement value to the PC ( computer ), push the RS232 key to start a 9600 bps RS232 transmission activePush the RS232 button again to cancel the transmissionWhen RS232 output port is transmitting, a RS232 indicator of LCD segment will be active.
10
5. POWER SUPPLY from DC
" "
ADAPTER
The meter also can supply the power supply from the DC 9V power adapter ( optional ). Insert the plug of power adapter into " DC 9V Power Adapter Input Socket ".
6. BATTERY REPLACEMENT
When the lower left of LCD display show necessary to replace the battery. However, in-specmeasurement may still be made for several hours after low battery indicator appears before the instrument become inaccurateWhen the display show " batt " and the power will be shut down.
it is
7. OPTIONAL ACCESSORIES
SMD TESTER Model : SMDA-22
* Optional SMD tester for LCR-9184, LCR-9183. * Useful tool for SMD components ( Resistor, Capacitor,Inductor ) LCR value measurement.
11
SMD TEST CLIP Model : SMDC-21
* Optional SMD test clip for LCR-9184,LCR-9183. * Useful test clip for SMD components ( Resistor, Capacitor,Inductor ) LCR value measurement.
12
1012-LCR9184