Luminex Corporation (Luminex) reserves the right to modify its product s and services at any time. This guide
is subject to change without notice. Although prepared to ensure accuracy, Luminex assumes no liability for
errors or omissions, or for any damages resulting from the application or use of this information.
® 100™ IS User Manual Version 2.3
REF
CN-M017-01
PN 89-00002-00-071 Rev. C
October 2005
The following are trademarks of Luminex: Luminex, Luminex 100, Luminex 100 IS, LabMAP, xMAP,
LumAvidin, Luminex XYP, Luminex FlexMAP, and Luminex SD. All other trademarks, including Windows,
Cheminert, Pentium, and Dell
The Luminex 100 IS software uses the VideoSoft
ActiveX controls, which are copyrighted by VideoSof t, 2001.
The contents of this manual and the associated Luminex software are the proper ty of L uminex and are
copyrighted. Except as specified in the End User License Agreement, any reproduction in whole or in part is
strictly prohibited.
are trademarks of their respective companies.
® VsFlexGrid Pro 7.0, VsPrinter 7.0, and VsView 3.0
Standard Terms and Conditions For Use of Product
By opening the packaging containing this product ("Product") or by using such Product in any manner,
you are consenting and agreeing to be bound by the following terms and conditions. You are also agreeing that the following terms and conditions constitute a legally valid and binding contract that is enforceable against you. If you do not agree to all of the terms and conditions set forth below, you must promptly
return the Product for a full refund prior to using them in any manner.
1. Acceptance - ALL SALES ARE SUBJECT TO AND EXPRESSLY CONDITIONED UPON THE
TERMS AND CONDITIONS CONTAINED HEREIN, AND UPON BUYER'S ASSENT
THERETO. NO VARIATION OF THESE TERMS AND CONDITIONS SHALL BE BINDING
UPON LUMINEX CORPORATION ("LUMINEX") UNLESS AGREED TO IN WRITING AND
SIGNED BY AN AUTHORIZED REPRESENTATIVE OF LUMINEX. For purposes of this agreement, "Seller" shall mean the Luminex authorized reseller that sells the Product to Buyer. Buyer, by
accepting the Product shall be deemed to have assented to the terms and conditions set forth herein,
notwithstanding any terms contained in any prior or later communications from Buyer and whether or
not Seller shall specifically or expressly object to any such terms.
2. Warranties - Any warranty obligations for the Product shall be exclusively provided in writing to
Buyer directly by Seller. LUMINEX MAKES NO WARRANTY WHATSOEVER REGARDING
THE PRODUCT AND LUMINEX SPEFICALLY DISCLAIMS ALL WARRANTIES, EXPRESS
OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NEITHER SELLER NOR LUMINEX SHALL IN ANY
EVENT BE LIABLE FOR INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES OF ANY
KIND RESULTING FROM ANY USE OR FAILURE OF THE PRODUCT, EVEN IF SELLER OR
LUMINEX HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGE INCLUDING,
WITHOUT LIMITATION, LIABILITY FOR LOSS OF WORK IN PROGRESS, DOWN TIME,
LOSS OF REVENUE OR PROFITS, FAILURE TO REALIZE SAVINGS, LOSS OF PRODUCTS
OF BUYER OR OTHER USE OR ANY LIABILITY OF BUYER TO A THIRD PARTY ON
ACCOUNT OF SUCH LOSS, OR FOR ANY LABOR OR ANY OTHER EXPENSE, DAMAGE OR
LOSS OCCASIONED BY SUCH PRODUCT INCLUDING PERSONAL INJURY OR PROPERTY
DAMAGE UNLESS SUCH PERSONAL INJURY OR PROPERTY DAMAGE IS CAUSED BY
SELLER'S GROSS NEGLIGENCE.
3. Buyer's Use of Product -Buyer agrees that no rights or licenses under Luminex's patents shall be
implied from the sale of the Product, except as expressly provided herein, and Buyer does not receive
any right under Luminex's patent rights hereunder. Buyer acknowledges and agrees that the Product
is sold and licensed only for use with Luminex's standard fluorescently dyed microspheres. Buyer
further acknowledges that the Product have not received approval from the United States Food and
Drug Administration or other federal, state or local regulatory agencies and have not been tested by
Seller or Luminex for safety or efficacy in food, drug, medical device, cosmetic, commercial or any
other use, unless otherwise stated in Seller's technical specifications or material data sheets furnished
to Buyer. Buyer expressly represents and warrants to Seller that Buyer will properly test and use any
Product in accordance with the practices of a reasonable person who is an expert in the field and in
strict compliance with the United States Food and Drug Administration and all applicable domestic
and international laws and regulations, now and hereinafter enacted.
BUYER HEREBY GRANTS TO LUMINEX A NONEXCLUSIVE, WORLDWIDE, UNRESTRICTED, ROYALTY-FREE, FULLY PAID-UP LICENSE, WITH THE RIGHT TO GRANT AND
AUTHORIZE SUBLICENSES, UNDER ANY AND ALL PATENT RIGHTS IN INVENTIONS
COMPRISING MODIFICATIONS, EXTENSIONS, OR ENHANCEMENTS MADE BY BUYER
TO THE PRODUCT OR TO THE MANUFACTURE OR USE OF THE PRODUCT ("IMPROVEMENT PATENTS"), TO MAKE, HAVE MADE, USE, IMPORT, OFFER FOR SALE OR SELL
ANY AND ALL PRODUCT; EXPLOIT ANY AND ALL METHODS OR PROCESSES; AND
OTHERWISE EXPLOIT IMPROVEMENT PATENTS FOR ALL PURPOSES. NOTWITHSTANDING THE FOREGOING, "IMPROVEMENT PATENTS" SPECIFICALLY EXCLUDES PATENT
CLAIMS CONCEIVED AND REDUCED TO PRACTICE BY BUYER CONSISTING OF METHODS OF SAMPLE PREPARATION, METHODS OF CONJUGATING PRODUCT TO ANALYTES,
THE COMPOSITION OF MATTER OF THE SPECIFIC CHEMISTRIES OF THE ASSAYS
DEVELOPED BY BUYER AND METHODS OF PERFORMING THE ASSAYS (I.E., THE PROTOCOL FOR THE ASSAY).
Buyer has the responsibility and hereby expressly assumes the risk to verify the hazards and to conduct any further research necessary to learn the hazards involved in using the Product. Buyer also has
the duty to warn Buyer's customers, employees, agents, assigns, officers, successors and any auxiliary
or third party personnel (such as freight handlers, etc.) of any and all risks involved in using or handling the Product. Buyer agrees to comply with instructions, if any, furnished by Seller or Luminex
relating to the use of the Product and not misuse the Product in any manner. Buyer shall not reverse
engineer, decompile, disassemble or modify the Product. Buyer acknowledges that Luminex retains
ownership of all patents, trademarks, trade secrets and other proprietary rights relating to or residing
in the Product.
4. Buyer's Representations, Release and Indemnity - Buyer represents and warrants that it shall use
the Product in accordance with Paragraph 2, "Buyer's Use of Product," and that any such use of Product will not violate any law, regulation, judicial order or injunction. Buyer agrees to release, discharge, disclaim and renounce any and all claims, demands, actions, causes of action and/or suits in
law or equity, now existing or hereafter arising, whether known or unknown, against Seller and
Luminex, and their respective officers, directors, employees, agents, successors and assigns (collectively the "Released Parties"), with respect to the use of the Product. Buyer agrees to indemnify and
hold harmless the Released Parties from and against any suits, losses, claims, demands, liabilities,
costs and expenses (including attorney, accounting, expert witness, and consulting fees) that any of
the Released Parties may sustain or incur as a result of any claim against such Released Party based
upon negligence, breach of warranty, strict liability in tort, contract or any other theory of law or
equity arising out of, directly or indirectly, the use of the Product or by reason of Buyer's failure to
perform its obligations contained herein. Buyer shall fully cooperate with the Released Parties in the
investigation and determination of the cause of any accident involving the Product which results in
personal injury or property damage and shall make available to the Released Parties all statements,
reports, recordings and tests made by Buyer or made available to Buyer by others.
5. Patent Disclaimer - Neither Seller nor Luminex warrants that the use or sale of the Product will not
infringe the claims of any United States or other patents covering the product itself or the use thereof
in combination with other products or in the operation of any process.
End-User License Agreement (EULA) for Luminex® Software
This Luminex End-User License Agreement (“EULA”) is a legal agreement between you (either an individual
or a single entity , also referred herein as “you”) the end-user and Luminex Corporation (“Luminex”) regarding
the use of the Luminex software product identified above, which includes computer software and online or
electronic documentation and may include associated media and printed materials (if any) (“SOFTWARE
PRODUCT” or “SOFTWARE”).
The SOFTWARE PRODUCT is protected by copyright laws and international copyright treaties, as well as
other intellectual property laws and treaties. The SOFTWARE PRODUCT is licensed, not sold.
1. GRANT OF LICENSE. Subject to the terms and conditions of this EULA, Luminex hereby grants to you a
nonexclusive, nontransferable, nonassignable license (without right to sublicense) under Luminex’s
copyrights and trade secrets to use the SOFTW ARE PRODUCT on a hardware platform purchased from
Luminex pursuant to Luminex’s terms and conditions of sale. You may make one (1) copy of the
SOFTWARE PRODUCT for backup or archival purposes only. Although no rights or licenses under any
of Luminex's patents are granted by or shall be implied from the license of the SOFTW ARE or the sal e of
Luminex instrumentation to you, the purchaser, you may obtain a license under Luminex’ s patents, if any,
to use this unit of Luminex instrumentation with fluorescently labeled microsphere beads authorized by
Luminex by purchasing such beads from Luminex or an authorized Luminex reseller.
2. RESTRICTIONS.
•You must maintain all proprietary notices on all copies of the SOFTWARE PRODUCT.
•You may not distribute copies of the SOFTWARE PRODUCT to third parties.
•You may not reverse-engineer, decompile, disassemble, or otherwise attempt to derive source
code from the SOFTWARE PRODUCT.
•You may not copy (other than one backup or archival copy), distribute, sublicense, rent, lease,
transfer or grant any rights in or to all or any portion of the SOFTWARE PRODUCT.
•You must comply with all applicable laws regarding the use of the SOFTWARE PRODUCT.
•You may not modify or prepare derivative works of the SOFTWARE PRODUCT.
•You may not use the SOFTWARE PRODUCT in a computer-based service business or publicly
display visual output of the SOFTWARE PRODUCT.
•You may not transmit the SOFTWARE PRODUCT over a network, by telephone, or
electronically by any means.
3. TERM AND TERMINA TION. Your rights under this EULA are effective until termination. You may
terminate this EULA at any time by destroying the SOFTWARE PRODUCT, including all computer
programs and documentation, and erasing any copies residing on your computer equipment. Luminex
may terminate this EULA upon thirty (30) days written notice to you. Your rights under this EULA
automatically terminate without further action on the part of Luminex if you do not comply with any of the
terms or conditions of this EULA. Upon any termination of this EULA, you agree to destroy the
SOFTWARE PRODUCT and erase any copies residing on your computer equipment.
4. RIGHTS IN SOFTWARE. All rights and title in and to the SOFTWARE PRODUCT and any copies
thereof are owned by Luminex or its suppliers. This EULA is not a sale and does not transfer to you any
title or ownership interest in or to the SOFTWARE or any p atent, copyright, trade secret, trade n ame,
trademark or other intellectual property right therein. You shall not remove, alter, or obscure any
proprietary notices contained on or within the SOFTWARE and shall reproduce such notices on any
back-up copy of the SOFTWARE. All title and intellectual property rights in and to the content which may
be accessed through use of the SOFTWARE PRODUCT is the property of the respective content owner
and may be protected by applicable copyright or other intellectual property laws and treaties. This EULA
grants you no rights to use such content.
5. EXPORT RESTRICTIONS. You agree that you will not export or re-export the SOFTWARE PRODUCT
to any country , person, entity, or end-user subject to U.S.A. export restrictions. Y ou hereby warrant no
state or federal agency has suspended, revoked, or denied your export privileges.
6. NO WARRANTY. THE SOFTWARE PRODUCT IS LICENSED “AS IS.” ANY USE OF THE SOFTWARE
PRODUCT IS A T YOUR OWN RISK. THE SOFTWARE PRODUCT IS PROVIDED FOR USE ONLY
WITH LUMINEX PRODUCTS. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
LUMINEX AND ITS SUPPLIERS DISCLAIM ALL WARRANTIES, EITHER EXPRESS OR IMPLIED,
INCLUDING , BUT NOT LIMITED TO, IMPLIED W ARRANTIES OF MERCHANT ABILITY, FITNESS FOR
A PARTICULAR PURPOSE, AND NONINFRINGEMENT.
7. LIMITATION OF LIABILITY. IN NO EVENT SHALL LUMINEX OR ITS SUPPLIERS BE LIABLE FOR
ANY SPECIAL, INCIDENTAL, INDIRECT, OR CONSEQUENTIAL DAMAGES WHATSOEVER
(INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF BUSINESS PROFITS, BUSINESS
INTERRUPTION, LOSS OF BUSINESS INFORMA TION, OR ANY OTHER PECUNIARY LOSS)
ARISING OUT OF THE USE OF OR INABILITY TO USE THE SOFTWARE PRODUCT, EVEN IF
LUMINEX HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGE S.
MISCELLANEOUS. This EULA is governed by the laws of the State of Texas, U.S.A., without reference to
conflicts of laws principles. You shall not assign or sublicense or otherwise transfer the rights or license
granted hereunder, by agreement or by operation of law, without the prior written consent of Luminex, and all
assignments in violation of this prohibition shall be null and void. This EULA is the complete and exclusive
agreement of Luminex and you and supersedes all other communications, oral or written, relating to the subject matter hereof. No change to this EULA shall be valid unless in writing and signed by the party against
whom enforcement is sought. The waiver or failure of Luminex or you to exercise in any respect any right or
rights provided for herein shall not be deemed a waiver of any further right hereunder. If any provision of this
EULA is held unenforceable, the remainder of this EULA will continue in full force and effect.
About This ManualThis manual provides you with information to understand and use the
Luminex
appendices that take you from this introduction to complete system
operation.
The manual’s text and figures offer examples when necessary.
Procedures are presented as step-by-step instructions. Glossary and
index sections assist as references.
The conventions in this manual assume that the reader has a basic
familiarity with computers, xMAP
Microsoft
common methods of accessing a command, such as from the main
menu bar, from the toolbar, and from menus that appear when you
right-click an area of the screen. Refer to the Glossary appendix for
unfamiliar terminology.
® 100™ IS system. It consists of multiple chapters and
® technology, and a knowledge of
® Windows® software. We typically document the
The Luminex 100 IS
2.3 System
The Luminex 100 IS 2.3 system is a benchtop system consisting of
the Luminex 100 analyzer, computer, monitor, keyboard, mouse,
Luminex XY Platform instrument (Luminex XYP™), Luminex
Sheath Delivery System (Luminex SD™), software, barcode reader,
sheath and waste containers, and xMAP technology reagents.
Intended UseThe Luminex software is designed to use xMAP technology with
assay kits available through kit manufacturers.The Luminex 100 IS
2.3 system performs a wide range of xMAP technology-based
laboratory tests, measuring biomolecular reactions on the surface of
xMAP microspheres. This system is intended for general
laboratory use.
PN 89-00002-00-071 Rev. C 1 - 1
Luminex 100IS User Manual Version 2.3xMAP Technology
Technical SupportLuminex Technical Support representatives are ready to help you,
particularly when the system or software cause any questions or
problems. If the question or problem relates to materials from the
assay kit, you should contact the kit provider directly.
Luminex Technical Support is available to users in the U.S. and
Canada by calling 1-877-785-BEAD (-2323) between the hours of
7:00 a.m. and 7:00 p.m. Central T ime, Mon day through Fri day. Users
outside of the U.S., Canada, Europe can contact us at +1 512-3814397 between the hours of 7:00 a.m. to 7:00 p.m. Central Time,
Monday through Friday. Inquiries may also be sent by email to
support@luminexcorp.com.
Luminex Technical Support is also a vailable to users in Europe by
calling +31-162408333 between the hours of 8:30 and 5:30, Central
European Time, Monday through Friday. Email inquiries in Europe
can be sent to supporteurope@luminexcorp.com.
Luminex WebsiteAdditional information is available on the Luminex website. Search
on the desired topic or navigate through menus. Also, review the
website’s FAQ section.
To access Luminex website FAQ section:
In your browser’s address field, enter:
http://luminexcorp.custhelp.com. This address takes you
directly to the FAQ section.
1 - 2PN 89-00002-00-071 Rev. C
Safety
2
SymbolsPlease become familiar with the information in this chapter before
using the equipment. Do not perform procedures on your Luminex
100 IS 2.3 system that are not specifically contained in this manual,
unless you are directed to do so by Luminex Technical Support.
These symbols describe warnings, cautions, and general information
used in the operation of this instrument. These symbols are further
defined under “Safety Precautions.”
NumberSymbolDescriptionNumberSymbolDescription
1Alternating current (ac)7Warning (refer to manual)
2Protective ground8Warning (refer to manual)
3On9Warning (refer to manual)
4Off10Warning (refer to manual)
5SNSerial number11Warning (refer to manual)
6Warning (refer to manual)
PN 89-00002-00-071 Rev. C 2 - 1
Luminex 100IS User Manual Version 2.3xMAP Technology
Warnings and NotesInformational notes and warnings appear in this manual.
Note: A note provides general helpful information. No safety or
performance issues are involved.
Caution: This message is used in cases where the hazard is minor or
only potential hazard is present. Failure to comply with the caution
may result in potentially hazardous conditions.
Warning: This message is used in cases where danger to the
operator or to the performance of the instrument is present. Failure to
comply with the warning may result in incorrect performance,
instrument failure, invalid results, or hazard to the operator.
Danger: This message is used in cases where significant risk of
serious injury or death is present.
Safety PrecautionsRead the following safety information before setting up or using the
Luminex 100 IS 2.3 system. A user should be present during
operation. This system contains electrical, mechanical, and laser
components which, if handled improperly, are potentially harmful. In
addition, biological hazards may be present during system operation.
Therefore, we recommend that all system users become familiar with
the specific safety advisories below, in addition to adhering to
standard laboratory safety practices. The protection provided by the
equipment may be impaired or the warranty voided if the system is
used in a manner not specified by the instructions or by Luminex
Corporation.
This caution label appears on the back of the Luminex 100 analyzer
and on the Luminex XYP instrument.
.
Figure 2-1 Fuse Caution Label
2 - 2PN 89-00002-00-071 Rev. C
xMAP TechnologySafety
Do not perform any maintenance or cleaning of the system’s
electrical components, with the exception of replacing fuses.
This label appears on the back panel of the Luminex 100 analyzer
and on the back panel of the Luminex XYP instrument.
Figure 2-2 CE Label
The Luminex 100 analyzer and the Luminex XYP instrument
comply with European Union (EU) safety requirements and,
therefore, may be marketed in the Europe Single Market.
This voltage label appears on the back of the Luminex 100 analyzer:
Figure 2-3 Luminex 100 Serial Number Label
The Luminex 100 analyzer has been tested by Underwriter
Laboratories, Inc.® (UL).
PN 89-00002-00-071 Rev. C 2 - 3
Luminex 100IS User Manual Version 2.3xMAP Technology
The following label appears on the back of the Luminex XYP
instrument.
Figure 2-4 Luminex XYP Serial Number Label
The Luminex XYP instrument has been tested by UL.
FCC LabelThis equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and the
receiver.
• Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for
help.
FluidicsThis system contains fluidics. In the event of a fluid leak, turn off all
power to the system and disconnect all power cords. Remember that
the on/off switch is not a disconnect means; the power cord must be
removed from the outlet. Contact Luminex Corporation for further
information.
2 - 4PN 89-00002-00-071 Rev. C
xMAP TechnologySafety
You must monitor waste levels manually. Do not allow the waste
container to overflow! Empty the waste container each time the
sheath fluid container is filled. Do not place the waste container on
top of the instrument.
Warning: If biological samples have been tested with the system,
use your standard laboratory safety practices when handling
system waste.
Luminex 100 Analyzer
Laser
The Luminex 100 IS system classifies per FDA 21 CFR 1040.10 and
1040.11 as a Class II laser product consisting of a Class I laser
product (Luminex 100 analyzer) and a Class II laser product
(barcode reader).
Figure 2-5 Laser Product Class Label
United States and international regulations require the following
warnings to appear on the instrument during operation and
maintenance.
This label appears on the back panel of the Luminex 100 analyzer.
Figure 2-6 Laser Radiation Caution Label
Under NO circumstances should you remove the Luminex 100
analyzer cover! When performing routine maintenance, turn power
to the Luminex 100 analyzer OFF and the disconnect the power cord.
PN 89-00002-00-071 Rev. C 2 - 5
Luminex 100IS User Manual Version 2.3xMAP Technology
All laser apertures are located within the Luminex 100 analyzer and
are contained within a protective housing. This label appears on the
optics cover within the Luminex 100 analyzer.
Figure 2-7 Laser Class Label
Warning — Use of controls or adjustments or performance of
procedures other than those specified herein may result in
hazardous radiation exposure.
Attention — L’utilisation des commandes ou réglages ou
l’exécution des procédures autres que celles spécifiées dans les
présentes prescriptions peuvent entraîner d’une exposition à un
rayonnement dangereux.
This label appears above the laser apertures located inside the optics
enclosure inside the Luminex 100 analyzer.
Figure 2-8 Avoid Exposure Label
2 - 6PN 89-00002-00-071 Rev. C
xMAP TechnologySafety
Barcode Reader LaserThis label is attached to the barcode reader.
Figure 2-9 Barcode Reader Laser Label
Do not stare into the beam or shine it into other people’s eyes.
Mechanical
Warning: During operation, this system contains exposed,
moving parts. Risk of personal injury is present. Observe all
warnings and cautions.
Warning: During operation, this system contains exposed,
moving parts which could result in puncture hazard. Risk of
personal injury is present. Keep hands and fingers away from the
Luminex XYP instrument slot during operation.
Note: Access doors must be
closed while operating the
Luminex 100 analyzer; the
operator must be present during
operation.
Biological
Warning: During operation, this system contains exposed,
moving parts which could result in pinch point hazard. Risk of
personal injury is present. Keep hands and fingers away from the
Luminex XYP instrument slot during operation.
Warning: Hum an and animal samples ma y cont ain biohazardo us
infectious agents.
Where exposure (including aerosol) to potentially biohazardous
material exists, follow appropriate biosafety procedures and use
personal protective equipment, such as gloves, gowns, laboratory
coats, face shields, or mask and eye protection, and ventilation
devices.
Observe all local, state, and federal biohazard handling
regulations when disposing of biohazardous waste material.
PN 89-00002-00-071 Rev. C 2 - 7
Luminex 100IS User Manual Version 2.3xMAP Technology
Heat
Warning: The heater plate of the Luminex XYP instrument may
be hot and could cause personal injury if touched. Do not touch
the heater plate.
Blue Indicator Light
The blue light above the Luminex 100 analyzer sample arm indicates
the on/off status of the Luminex 100 analyzer, and is harmless. The
blue light-emitting diode (LED) does not emit light in the UV
spectrum.
2 - 8PN 89-00002-00-071 Rev. C
xMAP TechnologySafety
Decontaminating
the Luminex 100
Analyzer for Return
Shipment
Note: It is the user’s
responsibility to
decontaminate the
analyzer before
shipment.
Luminex Technical Support will give you a Return Material
Authorization (RMA) number if they direct you to return the system.
They will explain how to return the system according to Luminex
procedures.
The accessible surfaces and the internal fluidics system must be
sanitized and decontaminated before returning the analyzer. This is
particularly important when biohazardous samples have been run.
Make a copy of this page to fill out and return with the system.
Complete the following checklist, signed and dated, and return it
with the Luminex 100 analyzer.
1.Remove all specimens, disposables, and reagents from the system.
2.Disconnect the sheath line going from the SD system to the analyzer.
3.Connect a sheath bottle filled with a solution of 10% to 20% household
bleach and water to the analyzer.
4.Sanitize the system using the Sanitize command on the main screen of
the system. Follow this by washing twice with distilled water.
5.Disconnect the system from AC power by turning off the power switch
on the rear of the system, then unplugging the analyzer power cord
from the wall source.
6.Disconnect the SD system and waste and sheath containers.
7.Rinse the waste container with 10% to 20% household bleach solution
and drain.
8.Wash all exterior surfaces with a mild detergent, followed by a 10% to
20% bleach solution.
9.Open both front doors of the analyzer and clean all accessible surfaces
with mild detergent followed by a 10% to 20% bleach solution.
10. Pack the system within a biohazard bag, place it in the corrugated box,
then insert it in its original packaging or an approved shipping
container. Attach this checklist to the top of the corrugated box prior to
packaging in the crate.
Was there an internal leak in the system? Yes No
Print Name:
Signature:
Date: Instrument Serial No.
PN 89-00002-00-071 Rev. C 2 - 9
Luminex 100IS User Manual Version 2.3xMAP Technology
2 - 10PN 89-00002-00-071 Rev. C
The System
3
Theory of OperationLuminex 100 IS technology is based on flow cell fluorometry with
Luminex-developed innovations. The fluidics, optics, robotics,
temperature control, software, and xMAP microspheres work
together to enable simultaneous analysis of up to 100 analytes in a
single test sample. Assay analysis requiring temperature control is
provided through the Luminex XYP instrument heater block.
There are two fluidics paths in the Luminex 100 analyzer. The first
path involves a syringe-driven mechanism that controls the sample
uptake. This mechanism permits small sample uptake volumes from
small reaction volumes. The syringe-driven system transports a
specified volume of sample from a sample container to the cuvette.
The sample is injected into the cuvette at a steady rate for analysis.
Following analysis, the sample path is automatically purged with
sheath buffer by the second fluidics path. This process removes
residual sample within the tubing, valves, and probe. The second
fluidics path is driven by positive air pressure and supplies sheath
fluid to the cuvette and sample path.
Sheath fluid is the delivery medium of the sample to the optics
component. The analysis sample is acquired using a sample probe
from a 96-well microtiter plate via the Luminex XYP instrument and
injected into the base of the cuvette. The sample then passes through
with sheath fluid at a reduced rate resulting in a narrow sample core
to ensure that each microsphere is illuminated individually. The
sample injection rate is such that the xMAP microspheres are
introduced to the optics path as a series of single events. The
Luminex SD system lets you run samples continuously without
refilling sheath bottles. It automatically draws sheath from a non
pressurized bulk sheath container to constantly maintain a reservoir
PN 89-00002-00-071 Rev. C 3 - 1
Luminex 100 IS User Manual Version 2.3xMAP Technology
of pressurized sheath fluid. A single 20 liter sheath container
provides enough fluid for 48 hours or more of normal operation.
The optics assembly consists of two lasers. One laser excites the dye
mixture inside the xMAP microspheres and the second laser excites
the fluorosphere bound to the surface of the xMAP microspheres.
Avalanche photo diode detectors measure the excitation emission
intensities of the color coding classification dye mixtures inside the
xMAP microspheres and a photomultiplier tube detects the excitation
emission intensity of the reporter molecule bound to the surface of
the xMAP microspheres. High speed digital signal processors and
advanced computer algorithms provide analysis of the xMAP
microspheres as they are processed through the Luminex 100
analyzer. Results of the analyses are processed and provided in a
report format.
HardwareThe Luminex 100 IS system includes the following hardware:
•Luminex 100 analyzer
•Computer (PC), monitor, and accessories
•Luminex XYP instrument
•Luminex Sheath Delivery System (Luminex SD™)
•Power cables
•Alignment guide
•Two long sample probes
•Luminex XYP instrument sample probe
•Reservoir
•Shield
•Heater block
•Sheath fluid container
•Waste container
•Sheath fluid line
•Air line
•Sheath fluid intake line
•Communications: 1 serial communication cable
•Communications: 1 USB communication cable
•Communications: 1 CANBUS communication cable (short
cable)
•Barcode reader
•Sample probe height alignment kit
•3/32 Hexdrive, Balldriver wrench
3 - 2PN 89-00002-00-071 Rev. C
xMAP TechnologyThe System
xMAP Technology
Reagents
Required
Laboratory
Reagents
Luminex 100 IS 2.3
Software
•Classification calibration microspheres (CAL1)
•Reporter calibration microspheres (CAL2)
•Classification control microspheres (CON1)
•Reporter control microspheres (CON2)
•Sheath fluid
•Household bleach
•70% isopropanol or 70% ethanol
•Mild detergent
Luminex 100 IS 2.3 software provides complete control of the
system and performs data analysis. Your Luminex 100 IS 2.3 system
is preloaded with the Luminex software. However, we supply a
software CD should you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.
Luminex 100 IS
Performance
Specification
Speed•USB communications link for fast data transfer
•Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write CD
•Installation: < 4 hours
•System calibration: < 10 minutes
•System controls: < 10 minutes
•Barcode reader entry of sample IDs
•Automatic post-analysis
•Analyze one 96-well plate/hour depending on manufacturer’s kit
•System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
PN 89-00002-00-071 Rev. C 3 - 3
Luminex 100 IS User Manual Version 2.3xMAP Technology
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.
Accuracy and Precision•Sample uptake volume: ± 5%
•Classification of xMAP microspheres: > 80%
•Misclassification of xMAP microspheres: ≤ 2% - may vary by
xMAP microsphere product lines. Refer to the specific product
information sheet for further details.
•Temperature control: 0°C to + 2°C of target
•Internal sample carry over: < 0.9%
•Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity values
Sensitivity•Detect 1000 fluorochromes phycoerythrin (PE) per xMAP
microsphere
•Reporter channel dynamic range: 3.5 decades of detection
•Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)
•Automatic sampling from a 96-well plate
•Start sampling from any well position
•Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills
•M icrotiter plates with 96 wells must be compatible with the
Luminex XYP instrument plate holder. The following microtiter
plate types are compatible with the Luminex XYP instrument
plate holder: flatbottom, conical, round, filter bottom, half plates
[overall height no more than 0.75” (19 mm)], any color
•M icrotiter plates with 96 wells must be compatible with
Luminex XYP instrument heater block temperature from 35°C to
3 - 4PN 89-00002-00-071 Rev. C