3.3.4 Control via Serial Ports!
SPECTRA III (L) light engines have two serial ports, labeled USB and RS-232 (Figure 3), which can be set to
receive either LEGACY or STANDARD mode commands. Connection to the computer requires a USB-A-to-
USB B cable (29-10058) or USB-to-RS-232 cable (29-10011). LEGACY commands are limited to controlling
on/off switching and intensity adjustment of individual color channels. Only one of the two ports can be set
to LEGACY mode at one time. The STANDARD mode command set gives access to an extensive panel of
operating status reports and configuration settings in addition to the the basic control functions of the
LEGACY command set. A complete listing of STANDARD mode commands is provided in Lumencor Light
Engine Command Reference (Document number 57-10018). Note that LEGACY and STANDARD mode
communications use different serial protocols (9600,8,N,1 and 115200,8,N,1 respectively). Changes to the
command mode setting for a serial port can be made via the Settings screen of Web GUI (Figure 5).
Changes are applied instantaneously and are retained between power cycles.
Select the command mode setting for the serial port that is compatible with the light engine device driver in
the control software. This selection is typically found under the “Devices” tab. The COM port address
assigned by the computer to the light engine USB serial port must be correctly registered in control software
at the same place.
3.3.5 Control From Light Engine Control Pod!
1. Connect SPECTRA III (L) to pod (p/n 83-10007) using USB A-to-USB B cable (p/n 29-10058)
2. Open the Web GUI control interface as described in Section 3.3.3.
3. Go to the SETTINGS page in the Web GUI. Make sure that the USB port configuration is set to LEGACY
mode and USB 5V is set to ENABLED.
4. The pod must be set in SPECTRA III (L) control mode. The light engine control mode setting is shown in
green letters at the bottom of the pod display screen. If the pod is not in SPECTRA III (L) control mode,
change the setting by holding down the MODE button on the pod until the light engine selection menu
appears. Move the cursor to “SPECTRA III (L)” by turning the pod control knob. Press the MODE button
again to select SPECTRA III (L) control mode and return to the main control screen.
5. Follow the instructions on Lumencor’s Light Engine Control Pod Operation sheet (p/n 54-10036). In brief,
press the COLOR button to select output light output channel, press the MODE button to toggle light output
on and off and turn the control knob to adjust intensity for the current light output channel selection.
3.3.6 TTL Control!
The TTL Interface provides users with a faster method of switching
color channel outputs on and off. Individual TTL inputs are provided
for each color channel, as shown in the adjacent table. As a
safeguard against unintended light output when the inputs are
initially connected, the light engine TTL port is disabled and must be
enabled by a serial command. To enable the TTL input, click the
Enabled button next to “TTL inputs” in the GUI under the Settings
tab (Figure 5) or send the standard mode command SET