LUMAT LB 9507 Instruction

Page 1
LUMAT
LB 9507
Variable Injector
Firmware Version 5.03
Id. No.: 81 957 BA4 Rev. No.: 06 5.03.2001
Page 2
Page 3
LB 9507 Contents
i
Contents
Trademarks iv
1. SYSTEM DESCRIPTION 1-1
1.1 Application 1-1
1.2 Structure and Operation of the Software 1-3
1.3 The Principle of Measurement 1-8
1.4 Sample Tubes 1-9
1.5 Data Management and Software update 1-10
1.6 Simple and Reliable Operation 1-11
1.7 Instrument and software versions 1-11
2. DESIGN OF LUMAT LB 9507 2-1
2.1 Keyboard 2-4
2.2 Display 2-5
2.3 Printer 2-5
2.4 Measurement Unit 2-6
2.5 Reagent Storage Area 2-10
2.6 Variable injectors 2-11
2.7 Reagent Supply 2-13
2.8 Memory Management 2-14
3. GETTING STARTED 3-1
3.1 Instrument Setup 3-1
3.2 Power Supply 3-1
3.3 Checking the Paper Supply 3-1
3.4 Switching on Lumat 3-2
3.5 System Configuration Parameters 3-3
3.5.1 Description of the System Configuration Parameters 3-4
3.5.2 System Configuration Table 3-6
3.6 Instrument Parameters 3-7
3.6.1 Setting/Verifying the Instrument Parameters 3-7
3.6.2 Changing the Instrument Parameters 3-9
3.7 Reagent Supply 3-11
3.8 Safety Precautions for Operation 3-13
4. STRUCTURE AND OPERATION OF SOFTWARE 4-1
4.1 Operation 4-1
4.2 The Six Functions of the Main Menu 4-2
4.2.1 <MEASURE> Menu 4-3
4.2.2 <PROTOCOL> Menu 4-4
4.2.3 <OPERATOR FUNCTIONS> Menu 4-6
4.2.4 <SYSTEM CONFIGURATION> Menu 4-7
4.2.5 <INSTRUMENT PARAMETER> Menu 4-8
4.2.6 <TURN ONCE> function 4-8
4.2.7 Service functions 4-8
4.3 General Procedure 4-9
5. RAW DATA MEASUREMENTS 5-1
Page 4
Contents LB 9507
5.1 Introduction 5-1
5.2 Creating Raw Data Protocols 5-3
5.3 Editing/Deleting Protocols 5-9
5.4 Performing Raw Data Measurements 5-11
6. CUT-OFF MEASUREMENTS 6-1
6.1 Introduction 6-1
6.2 Creating CUT-OFF Protocols 6-4
6.3 Editing/Deleting Protocols 6-12
6.4 Running Cut-off Measurements 6-14
7. QUANTITATIVE MEASUREMENTS (LIA...) 7-1
7.1 Introduction 7-1
7.2 Setting Up Quantitative Protocols 7-3
7.3 Editing/Deleting Protocols 7-12
7.4 Entering a Master Curve 7-14
7.5 Performing Quantitative Measurements 7-17
7.5.1 Measurement with a Complete Standard Curve 7-17
7.5.2 Measurements with Master Curve 7-25
8. T-UPTAKE/FTI-MEASUREMENTS 8-1
8.1 Introduction 8-1
8.2 Creating T-uptake/FTI protocols 8-3
8.3 Editing/Deleting protocols 8-12
8.4 Performing T-uptake/FTI Measurements 8-14
9. DUAL-LUCIFERASE
TM
REPORTER GENE ASSAY 9-1
9.1 Creating Protocols for Dual-Luciferase
TM
Assay 9-3
9.1.1 With 2 injectors installed 9-3
9.1.2 With 1 injector installed 9-8
9.1.3 Without injectors 9-11
9.2 Editing/Deleting Protocols 9-14
9.3 Performing Dual-Luciferase
TM
Reporter Gene Assay
Measurements 9-16
10. KINETIC AND REPEATED MEASUREMENTS 10-1
10.1 Kinetic Option in Raw Data Protocol 10-1
10.2 Repeated Measurements 10-2
10.2.1 Creating Repeated Protocols 10-4
10.2.2 Editing/Deleting Protocols 10-9
10.2.3 Performing Repeated Measurements 10-11
10.3 Ratemeter 10-14
11. OPERATOR FUNCTIONS 11-1
11.1 <REAGENT> Function 11-2
11.1.1 Loading reagents 11-3
11.1.2 Refreshing reagents 11-4
11.1.3 Washing of the injector system 11-5
11.1.4 Removing reagents 11-7
11.2 Performance test 11-8
11.3 Manual Data Entry 11-9
Page 5
LB 9507 Contents
iii
11.4 Retrieval of Stored Data 11-10
11.4.1 Rerun of a Full Standard Protocol 11-11
11.4.2 Rerun of a 2-Point Calibrator Protocol 11-14
12. CONNECTION TO EXTERNAL COMPUTERS 12-1
12.1 General 12-1
12.2 RS 232 Data Output LB 9507 in Protocol-Specific Measuring Mode 12-2
12.3 Data Output in Raw Data Measurements 12-4
12.4 Data Output after Manual Input of RLU Values 12-4
12.5 Data Output of Stored Assay Data 12-4
12.6 Data Output of RLU/s in the Ratemeter Mode 12-4
13. MAINTENANCE 13-1
13.1 Supplying Paper 13-1
13.2 Replacing the Mains Fuse 13-2
13.3 Changing the Operating Voltage 13-3
13.4 Cleaning the Injector System 13-3
13.5 Cleaning Lumat 13-3
13.6 Changing the Air Filter 13-3
13.7 Checking the Hose Connections 13-3
14. TROUBLESHOOTING 14-1
14.1 Error Messages during Operation 14-1
14.2 Injector System Malfunction 14-1
14.3 Other Errors 14-2
14.4 Default Values 14-2
15. TECHNICAL SPECIFICATIONS OF LB 9507 15-1
16. FLOW DIAGRAM OF THE MEASUREMENT PROCEDURE 16-1
APPENDIX 1
A.1 Mathematical Basis 2 A.1.1 Overview of the Mathematical Data Reduction 2 A.1.2 Variable Zero Concentration Factor 4 A.1.3 Spline Function 5 A.1.4 Calculation of a Variable Reference Value 6 for Standardization of the Logit Transformation 6 A.1.5 Extrapolation of the Concentration 7 A.2 Data Acquisition of the Lumat LB 9507 8 via the Serial Interface (I/O-Port) and 8 Saving this Data in a MS-DOS ASCII-File 8 A.3 Connection Cable Lumat/Apple Macintosh Computer 9 A.4 Background measurement 10 A.5 Overload Detection 12 A.6 Glossary 13 A.7 Index I
Page 6
Contents LB 9507
iv
Trademarks
Dual-Luciferase and Stop and Glow are trademarks of Promega Corp. Lumat is a trademark of Berthold Technologies. Macintosh is a registered trademark of Apple Computer Inc. Excel is a registered trademark of Microsoft Corp.
Page 7
LB 9507 Preface
I
Preface
Use and Function The Lum at LB 9507 is a semi- automatic luminometer for universal
use in bio- and chemiluminescence, and is very easy to operate. It can work with up to 2 reagent injectors, and its rotating sampling method enables samples to be measur ed and loaded at the sam e time.
Special Instructions and Precautions The manufacturer has undertaken everything to ensure the safe
operation of the instrument (regarding electricity, electronics and mechanics). The user has t o make sure that the eq uipment is
set up and installed properly to guarantee safe operation:
Use grounded mains.
Do not open the instrument.
Disconnect Lumat from the mains before cleaning.
If the operational safety of the instrument is no longer ensur ed,
turn it off and disconnect it from the mains.
If liquid gets into the instrument, disconnect it from the mains.
Protect yourself from electrostatic charging (e.g. carpets),
since discharging may damage sensitive electronic components
Set up the instrument in a dry room only.
Be careful when handling strong alkaline or acidic reagents.
Make sure you follow the reagent manufacturer’s instructions.
If there is a leak in the reagent tubing system you must abort the measurement series and fix the leak . If necess ary, call the manufacturer’s Technical Service. If a leak occurs there is a risk of reagent entering the inside of the instrument.
Only technicians authorized by Berthold should carry out repairs or modifications on the instrument.
The Lumat LB 9507 should only be used to perform the functions described in this manual.
The proper use and handling of the instrument r equires that the operating personnel be familiar with this manual. Please read it carefully before taking the instrum ent into use - even though the operation of Lumat may appear simple.
Page 8
Preface LB 9507
The tests and maintenance routines recommended by the manufacturer should be carried out to ensure the operator’s safety as well as the correct performance of the instrument.
All maintenance and service routines above and beyond those mentioned in this Operator’s Manual mus t only be performed by service technicians authorized by BERTHOLD TECHNOLOGIES or by the company itself.
The Structure of the operating manual
Because the Lumat LB 9507 is an universal luminometer, it is capable of a wide range of applications and different measurement modes (raw data, cut-off, kinetics, repeated, T­Uptake/FTI, Dual-Luciferase
TM
reporter gene assay and quantitative measurements ). To make it easy for the user who is working with only one measurement mode to find the required information quick ly, this operating m anual inc ludes all the relevant information on each measurement mode in separate chapters. This means you will not need to search through the m anual to f ind a different measurement mode.
The description of each measurement mode follows the same pattern: first a brief introduction, then a flow chart providing a summary of the procedure, and finally a detailed explanation of how to create a protocol and run a measurement, using all the messages shown on the instrument’s display.
Chapter 1 describes the system construction, the ways it can be used, and
the operating procedure
Chapter 2 provides detailed information on the individual system
components
Chapter 3 describes how to commission Lumat, from assembly to
configuration
Chapter 4 explains the structure and handling of the software and its six
menus
Chapter 5 describes the Raw data measurement mode Chapter 6 describes the Cut-off measurement mode Chapter 7 describes the Quantitative measurement mode (LIA..) Chapter 8 describes the T-Uptake/FTI measurement mode Chapter 9 describes the Dual-Luciferase
TM
Reporter Gene Assay mode
Chapter 10 describes the Kinetic and Repeated measurement mode
Page 9
LB 9507 Preface
III
Chapter 11 describes the special instrument functions such as washing,
pumping back reagents, performance testing, etc.
Chapter 12 explains how to connect an external PC and how to transfer data Chapter 13 summarizes the maintenance and service routines that can and
should be carried out by the user
Chapter 14 lists all possible error messages and tells you how to eliminate
problems
Chapter 15 lists technical specifications of Lumat Chapter 16 contains a diagram and a flow chart to help you find your way
around the menu structure in the Lumat software. The diagram shows the assignment of the functions to the menus, and the flow chart shows the procedure when creating measurement protocols
Appendix includes a description of the mathematical basis for calculations, a
glossary and an index to help you find your way around, and finally the data sheets.
Writing conventions The description of the sof tware and dialogs are bas ed on a writing
convention, which uses repeated symbols and makes reading easier.
The software dialog, i.e. the input sequences performed by the user via keyboard and display, are described a) by illustrating the displays one after another b) by explanatory texts.
How displays are shown In the displays shown, the functions in bold face are those that
can be selected using softkeys to move to the next display. The text and numerical entries are only exam ples, and ar e als o shown in bold face on the displays illustrated.
This allows you to see the difference immediately between things defined by the program and individual entries.
Explanatory text In the explanator y text, the menus and functions of the program
and the function keys are shown in pointed brackets <> and bold face. Text and numerical inputs are placed in inverted comm as and printed in bold face. Actions that the user needs to perform are immediately recognizable, e.g. enter ”15” and <enter>.
Page 10
Preface LB 9507
IV
Display numbering For reasons of clarity, the displays within one measurement type
are numbered. In addition, every measurem ent type and special function is given an identifying letter. This allows every display to be easily identified and immediately assigned to a certain procedure. The following identifying letters are used:
R = Raw data measurements (Chapter 5) C = Cut-off measurements (Chapter 6) L = Quantitative (LIA...) measurements (Chapter 7) T = T-uptake/FTI measurements (Chapter 8) D = Dual-Luciferase
TM
Reporter Gene Assay (Chapter 9) K = Kinetic measurements (Chapter 10) W = Washing function (Chapter 11.1)
S = Call up stored data (Chapter 11.4)
Page 11
LB 9507 Chapter 1: System Description
1-1
1. System Description
1.1 Application
The Lumat LB 9507 is a semi-automatic, user-friendly luminometer for universal use in bio- and chemiluminescence. Equipped with up to 2 reagent injectors, a highly sensitive photomultiplier, flexible software and a highly advanced, easy-to­handle mechanical system, Lumat is excellently suited to applications such as ATP measurement, immunoassays, DNA probe assays, calcium meas urements and reporter gene assays. Qualitative classification of unknown samples using the ”cut-off method” can also be performed.
A specially developed rotating sample holder and transport
system for 2 test tubes allows samples to be measured and loaded in parallel. This almost doubles the sample throughput
during the normal measurement duration of 5 to 10 seconds, thus equalling the speeds reached by automatic systems.
The Lumat software has been tailored to the user’s specific needs and the processes of lumines cence measurement. Essentially, it is divided into 6 basic parts:
1. Raw data measurement
In this measurement m ode, the quantity of light measured in RLU units over a user-defined measuring time is produced for each sample. The mean value and the coefficient of variation are generated for replicate groups. You have the option of printing out a high resolution (up to 2000 data points) curve and 20 data points with time scale for each sample, which enables you to analyze the kinetic behaviour of the light emission of a sample.
This measurement mode, which allows up to 2 injections per sample, is mainly used for ATP measurements or f or measuring luminescence reporter gene assays.
2. Cut-off measurements
Using negative and positive standards, unknown samples are classified quantitatively by interpreting the results of the measurements on the basis of predefined RLU limits.
3. Quantitative measurements (LIA, ILMA...)
Known standard concentrations are measured before the unknown samples. After transformation of the calibration curve (either log-log or logit-log), the program interpolates standards and unknown samples. Instead of using the standard concentrations, you can also enter the values from a master c ur ve which is defined by the reagent manufacturer.
Page 12
Chapterl 1: System Description LB 9507
1-2
4. Uptake/FTI measurements
In this measurem ent mode, a standard is used to convert patient samples into a %-uptake and then into an F T I- value (F ree Thyroid Index).
5. Dual-Luciferase
TM
Reporter Gene Assay
This measurement is used for automatic measurement of Promega's Dual-Luciferase
TM
Reporter Gene Assay (injection­measurement-injection-measurement). In this measurement mode, the quantity of light measured in RLU units over a user­defined measuring time is produced for both measurements of each sample. The m ean value and the coefficient of variation are generated for replicate groups. The measured data of both measurements will be evaluated to get the normalized response.
6. Kinetic and Repeated Measurements
These measurement modes are used to follow light emission from the sample in time. A kinetic protocol can be used to tak e up to 2000 data points from one measurement. During a repeated measurement the s ample will be measured several times. Each measurement results in one measurement point.
Page 13
LB 9507 Chapter 1: System Description
1-3
1.2 Structure and Operation of the Software
The Lumat software is designed for the applications and processes found in lum inesc enc e measurement. In addition to the five measurement modes, we also distinguish between routine and single measurements. Kinetic measurements can also be performed with Lumat.
Routine measurements
Measurement protocols Measurement protocols def ining the param eters for m easur em ent
and evaluation are generally set up once for routine measurements. To run a measurement, one simply uses an existing protocol and starts the measurement.
Lumat can store up to 40 diff erent meas urem ent protocols. When creating a protocol, select one of the measurement types described above and the entry sequence for this measurement type will appear on the instrument display. Each entry is automatically printed out on the integrated thermal printer , so that each protocol is documented.
Once you have set up a measurement protocol with the parameters you need, you can use it as often as you choose for later measurements.
Measurements Measurements based on these stored protocols can be performed
by simply selecting one of them and starting the m easurem ent. In accordance with the instructions defined in the measurement protocol, instructions for the oper ations that are expected of the user and the sequence of tubes being m easured are displayed on the LCD.
Printout At the start of the measurement, all essential parameters are
printed out, followed by the results of the measurem ent obtained by the calculation methods defined by the selected measurement protocol.
Single measurements
You may often want to try out new parameters instead of the stored protocols. If this is the case, the Raw data measuremen t type allows you to enter measurement parameters immediately before starting a measur ement, and then start the measurement. The most recently used param eter set will rem ain stored between individual measurements, allowing for easy correction.
Page 14
Chapterl 1: System Description LB 9507
1-4
Kinetic measurements
There are three ways to perform kinetic measurements:
1. Using a “RAW DATA" measurement protocol or a "Direct
Entry” measurement which is not stored in the instrument memory. This k ind of measurement provides always a fixed measurement time. The results can be printed out, or received on the RS 232 port of a computer. The Berthold "WinTerm" program is suitable to receive and transfere the results into Microsoft Excel.
2. Using a “REPEATED” measurement pr otocol to set up the
injector and timing intervals in combination with a stored protocol. The results can be printed out, or received on the RS 232 port of a computer. T he Berthold "WinTerm " program is suitable to receive and transfer the results into Microsoft Excel.
3. Using continuous measurement in “RATEMETER” mode.In
this mode results are continuously produced until the measurement is stopped by the user. Results can be printed out, or received on the RS 232 port of a computer.The Berthold "WinT erm" program is suitable to receive the results and transfer them into Microsoft Excel.
The measurement procedure
The measurement procedure is defined by the user in measurement protocols. The integrated software controls the measurement and evaluation s teps, and displays messages and instructions for the user. Lum at is a semi-automatic system that requires only a few simple actions by the operator; this is supported by an ergonomically sensible arrangement of controls on the front of the instrument.
To perform a measurement you must obser ve a s imple procedure which is controlled by the software. If you do not follow this procedure, error messages will be produced accordingly.
The sample chamber and the sample holder form a single unit. The sample holder contains 2 holes, each for one sample tube, and sits on the round measurement cham ber as a rotating block. It is turned in 180° steps by the integrated software, moving one hole in front of the photomultiplier (measurement position) and the other into the loading position (see Fig. 1-1).
Page 15
LB 9507 Chapter 1: System Description
1-5
Fig. 1-1: Measurement unit
In the Lumat’s standby mode, the cap should be closed to protect the measurem ent chamber against dust. You can leave it open during measurement s o that tubes can be swapped without interruption; the measurement c hamber into which the tubes are inserted is sealed against the entry of light anyway. The cap only needs to be closed if you are measuring samples whose fluorescent light fades before measurement.
At the start of a measurement, load the 1st tube into the accessible hole (loading position) as instructed on the display, and press the <start> button to the right of the measurement chamber. When the system detects that a tube is loaded, the measurement cham ber rotor will turn through 180° and the tube will end up in front of the photomultiplier. The measurement procedure defined in the measurement protocol begins automatically in this position. The 2nd hole is now in the loading position on the side facing the user. Load another tube during the measurement, and pres s the <start> button onc e again. The rotor will not turn until the sample in front of the photomultiplier has been fully measured. The measured sample tube is transported back to the loading position and can be replaced while the 2nd tube is being measured, and so on.
tube
transport
rotor
Loading position
Mirrored measurement
chamber
Spring pin for lifting the tube in loading position
Injector head
Injector tips
Measurement position
Meas. chamber opening for photomultiplier
Measurement chamber cap Injector lines to the pumps
Page 16
Chapterl 1: System Description LB 9507
1-6
The program runs through the individual steps for each sam ple in accordance with the settings in the measurement protocol:
background measurement, injections, measurement.
Upon completion of the bac kground meas urement, the pum ps ar e activated one after the other. If required, pump # 1 injects a variable amount of the first r eagent into the sample tube. After a defined delay, pump # 2 also injects a variable amount (which may differ from the 1st injection) of the second reagent into the sample tube. Of cours e, measurements without or with only one injection are also possible.
The injection of the 2nd reagent typically initiates the chemical reaction resulting in the emission of light photons. The num ber of photons that are emitted during this reaction is indirectly related to the analyte concentration in the sample.
The photons hitting the cathode of the photomultiplier are transformed into electrons and amplified as single pulses. A threshold discriminator suppresses the noise signals of the photomultiplier. The computer c ounts the pulses and supplies the integral result of the light signal (photon counting method).
Every measurement step is automatically documented on the thermal printer together with the measurement results obtained.
Page 17
LB 9507 Chapter 1: System Description
1-7
Chart of measurement procedure
Display User action Instrument action
1 SAMPLE 1 TUBE 1 INSERT TUBE... Load the 1st tube
2 SAMPLE 1 TUBE 1
START
Press <start> button
Rotor turns the 1st tube
in measurement position 3 MEASURING... Measuring tube 1 4 MEASUREMENT (RLU) 133 Display results
5 SAMPLE 2 TUBE 2 INSERT TUBE... Load the 2nd tube
6 SAMPLE 2 TUBE 2
START
Press <start> button
Rotor turns the 2nd tube
in measurement position 7 MEASURING... Measuring tube 2
8 MEASUREMENT (RLU) 137 REMOVE TUBE Remove the 1st tube Displays results
9 SAMPLE 3 TUBE 3 INSERT TUBE... Load the 3rd tube
10 SAMPLE 3 TUBE 3
START
Press <start> button
Rotor turns the 3rd tube
in measurement position
Pressing the <exit> button finishes measurement.
Page 18
Chapterl 1: System Description LB 9507
1-8
1.3 The Principle of Measurement
The emitted light is m easured by a selected high-sensitivity, low­noise photomultiplier. Its spectral sensitivity covers a range between 390 and 620 nm. All well-established applications in bio­and chemiluminescence emit in this wavelength range.
The photomultiplier operates as an ultr a-fast photon counter . The photoelectrons released from the photocathode by the light quanta are multiplied via a dynode chain and trigger a fast pulse with a rise time of a few nanoseconds at the anode. T hese pulses are then amplified by a very fast amplifier. A threshold discriminator suppresses the low-energy pulses caused by the noise of the photomultiplier. The single pulses are counted digitally, their total number being directly proportional to the emitted quantity of light.
However, the ”relative light units” (= RLU) are used as unit of measurement f or the r aw data, and not the num ber of puls es. The number of RLUs is f ound by dividing the directly counted pulses by ten; moreover, the raw data are m ultiplied by the RLU-factor (see Instrument Parameters) which allows compensation of the inevitable individual fluctuations of the cathode sensitivity of various photomultipliers.
The kinetics of many chemilum inescence reactions is so fast that typical counting times are in the range of 1-5 s per m easurem ent. Thus, sample throughputs of 600 samples per hour are possible.
Bioluminescent samples usually require longer measurement times.
The individual measurement time can be found in method publications or in your kit description.
Page 19
LB 9507 Chapter 1: System Description
1-9
1.4 Sample Tubes
The Lumat LB 9507 uses sample tubes with the nominal dimensions 12
x 75 mm length with round bottoms. However,
due to the special requirements of luminesc ence measurements not every type of sample tube can be used.
The permissible diameter variation is lim ited to a range between
11.5 and 12.1 mm; the length of the tubes may vary from 75 to 76 mm.
Basically, all materials with good transparency, such as polystyrene, glass and polyethylene may be used. However, some tube types show a higher and, moreover, fluctuating back ground caused by electrostatic charging and phosphorescence.
To prevent electrostatic charging we recommend using sample tubes with antistatic additives. These can be purchased from Berthold Technologies.
Phosphorescence can be avoided by not exposing the tubes to short-wave light prior to measurement. In general, there is less interference if the tubes are m oistened prior to the m easurement, for example during the preceding sample preparation. Never change the tube type within an assay!
Page 20
Chapterl 1: System Description LB 9507
1-10
1.5 Data Management and Software update
The Lumat LB 9507 Software is based on two parts:
The kernel pr ovides all the basic or elementary functions to
control all the electronic functions, also the download manager and serial number.
The User interface which controls the us er functions and all
individual parameters eg. individual measurem ent times and measurement methods, reagent timing, data management for results and set up of evaluation procedures.
This arrangement allows you to load updates and upgrades with the minimum of expense.
If a power failure occurs, the measured data from the last 30 measurement runs remain stor ed in a battery-backed mem ory for at least 3 months.
The measurem ent protoc ols are s tored in the instr um ent’s current and battery independent internal memory, and can only be lost if the CPU develops a fault or during a program is update.
A RS232 interface allows the kernel or operating s oftware to be loaded from a PC to Lumat. The Update software will be provided only from Berthold Technologies directly or from by Berthold Technologies authorised service organisations.
Page 21
LB 9507 Chapter 1: System Description
1-11
1.6 Simple and Reliable Operation
A characteristic feature of the LB 9507 is its user-friendliness. Possible alternatives in a specific situation ar e displayed in plain language - either German, English or F rench - as chosen by the user. You just have to press the ”softkey” below the respective option to execute this function.
Numerous self-diagnos tic and service functions prevent incorrect measurements. For example:
Background measurement prior to the injection to detect
unacceptable phosphorescence.
Sample tube detection to rule out injections into the empty
measurement chamber.
Automatic wash cycle to fill the injector system with fresh
reagents if the luminometer has not been used for a long period of time.
Automatic printout of all entries and results.
Measurement and printout of the kinetics, so you can
determine the best counting time. The k inetics monitor is also of importance in designing new assays.
1.7 Instrument and software versions
Order number
Lumat LB 9507-0 without injectors
90V – 240V 81957-50
Lumat LB 9507-1 one variable volume injector
90V – 240V 81957-51
Lumat LB 9507-2 two variable volume injectors
90V – 240V 81957-52
Lumat LB 9507-LIA
90V – 240V 81957-53
Please notice that not all the instrument vers ions comprise exactly the same software environment described in this manual. The standard version of Lumat LB 9507 does not include the LIA software module.
Page 22
Page 23
LB 9507 Chapter 2: Design of the Lumat
2-1
2. Design of Lumat LB 9507
This chapter describes the various components of Lumat LB 9507.
Front view
1 Keyboard 2 Display 3 Thermal printer 4 Paper advance button 5 Reagent supply bottle (optional) 6 Injector line (optional) 7 Injector pumps (optional) 8 Sample holder cap (keeps out dust) 9 Measurement chamber with sample holder 10 <start> button
23 45678
1910
Fig. 2-1: Lumat LB 9507 (front view)
Page 24
Chapter 2: Design of the Lumat LB 9507
2-2
Inside view
Main switch; fuse holder Contrast for printer Host I /O HV; amplifier Air filter Injector pumps Combi element and display RS 232 and analyzer
Power supply CPU Auxiliary board Stepper m ot or Tube detector Measurement chamber
for turning the
tube transport rotor
Fig. 2-2: Inside view of Lumat
Do not open the instrument!
The Lumat LB 9507 must only be opened by service technicians authorised by Berthold Technologies!
Injector lines
Photomultiplier
Injector head
Page 25
LB 9507 Chapter 2: Design of the Lumat
2-3
Rear side The rear panel of Lumat:
1 Mains connection (115 or 230 V) 2 Fuse with selector for voltage, display for selecting voltage 3 Power switch 4 Computer port (I/O port) 5 Printer contrast adjust 6 Display contrast adjust
Air filter Power switch Fuse
0
Host I/O port Display Mains plug
RS 232 Print er
Fig. 2-3: Rear view of Lumat
Page 26
Chapter 2: Design of the Lumat LB 9507
2-4
2.1 Keyboard
The Lumat functions ar e directed through keyboard entries. The keyboard consists of twenty-six letter keys, twelve numerical keys, three softkeys, and five additional function keys (<enter>, <delete>, <last>, <exit> and <space>). Softkeys are keys with changing functions defined by software and shown on the display. The functions are selected using the corresponding buttons beneath the display.
The keyboard design utilizes pushbuttons which are all sealed to protect them from liquid spillage (Fig. 2- 4). Audio feedback of key action is provided by a short beep for correct entries and a longer beep for incorrect entries.
1 Alphanumeric keyboard 2<space> key 3 Function keys <exit>, <delete>, <last> and <enter> 4 Softkeys 5 Display 6 Softkey functions: left, center, right
6
exit delete last enter
1 2 3 4 5 6 7 8 9 0 .
Q W E R T Y U I O P
A S D F G H J K L -
space Z X C V B N M /
21
Fig. 2-4: Keyboard
5
4
3
Page 27
LB 9507 Chapter 2: Design of the Lumat
2-5
2.2 Display
The display located above the keyboard is a two-line LCD display, each line containing 40 characters. T he LED background lighting switches on when Lumat is powered up.
Functions:
Entries request and display. T he second line displays the functions of the sof tkeys, which
are activated using the buttons located directly below the display field
Results display Instrument status display User guidance during protocol entry and measurement
2.3 Printer
The thermal printer located on the top left corner of the ins tru ment has the following functions:
➤➤➤➤ to print the protocols entered ➤➤➤➤ to print the parameters of an individual protocol ➤➤➤➤ to print measurement results and cut-off classes ➤➤➤➤ to plot kinetics curves
As an alternative to the thermal printer you can also connect a PC to the RS 232 Interface to receive the measured data. A suitable software is "WinTerm", available from Berthold Technologies.
Page 28
Chapter 2: Design of the Lumat LB 9507
2-6
2.4 Measurement Unit
The measurem ent unit is located to the right of the keyboard (see Fig. 2-1, [9]). The sample c hamber is protected agains t dust by a semi-circular cap. The measurement chamber is located in the instrument itself and c an only be accessed by removing the cover of the instrument (see Fig. 2-2).
Structure and function The following diagrams (F igs. 2- 5 to 2- 8) ex plain the s truc tur e and
function of the measurement unit.
The measurement unit comprises the following components:
●●
a black, light-proof measurement chamber casing
●●
the tube transport rotor inside this casing, incorporating 2 holes
for holding sample tubes and a mirrored measurement chamber
●●
a tube detector
●●
a stepping motor for rotating the rotor
●●
an injector head in the measurement unit lid
●●
an opening for the photomultiplier at the rear.
Injector lines to the pumps
Measurement chamber cap
Fig. 2-5: The structure of the measurement unit (schematic illustration)
Loading position
Mirrored measurement
chamber
Spring pin for lifting the tube in loading position
Injector head
Injector tips
Measurement position
Meas. chamber opening for
p
hotomultiplier
tube
transport
rotor
Page 29
LB 9507 Chapter 2: Design of the Lumat
2-7
Tube transport rotor The tube transport rotor is m ade of a special conductive plastic
which prevents electrostatic charges from building up.
The rotor contains two holes positioned opposite each other, and each can hold a sample tube, m eaning that one hole is always in
loading position (at the front, facing the user) and one in measurement position. Pressing a button (<START> when
measuring, or the softkey <TURN ONCE>) causes the rotor to turn through 180°, thus switching the positions of the tubes. The tube in the loading position is lifted by a spring-pin so that it is easier to take out (see Figs. 2-5 and 2-6) . This rotation principle
allows samples to be loaded and measured at the same time.
Sealing lips (spring-loaded)
Fig. 2-6: Tube transport rotor
To distinguish between the two measurement chambers, one of them is marked with a red spot. This measurement chamber is called the "red" chamber, the oth er is called the "black" chamber.
The top of the tube transport rotor includes two spring-loaded sealing lips which shield the measurement chamber from light.
Tube in measurement position
Tube in loading position (pushed up)
Measurement chamber: the rear surface is mirrored to increase the light again
Spring pin: the tube sits on the spring pin; in the loading position it is ­ pushed up by a sloping base
Page 30
Chapter 2: Design of the Lumat LB 9507
2-8
In case of need the rotor can also be turned manually (see below), so that it can be placed in positions other than the exact measurement positions.
The software-controlled rotation movement positions the rotor exactly; this is vital for measur ement as all of the sam ple m ust be located in front of the photomultiplier, and the measurement chamber mus t rem ain sealed agains t light. T his position has been reached if:
the tube is slightly raised in the loading positionthe k idney-shaped holes in the measurem ent chamber cover
and the rotor are exactly aligned (see Fig. 2-7).
The spring-loaded sealing lips (see Fig. 2-6) of the tube transport rotor will only click into the groove in the lower side of the measurement unit cover when this position has been reached, thus preventing any light from entering the rear measurement chamber.
Fig. 2-7: Exact positioning of tube transport rotor
Position a
In position a, turn the rotor anticlockwise until position b has been reached
Position b
Page 31
LB 9507 Chapter 2: Design of the Lumat
2-9
Tube detector At the front of the instrum ent is a light barr ier for detecting tubes.
This prevents an unloaded sample holder from moving into the measurement position. In this case injection would contaminate the inside of the measurement chamber. The test is performed every time the <start> button is pressed during the meas urement procedure or when activating the washing function.
Stepping motor The tube transport rotor is driven by a stepping motor which can
be stopped by applying a slight amount of pressure. T his prevents finger injury and stops the tubes from getting damaged. If you want to clean the upper surface of the tube transport rotor, you can turn it manually to any position. To return the rotor to the correct position, press the sof tkey <TURN ONCE>. This will turn the rotor to the next position or the position after next, depending on the current position. You can reset the pos ition m anually if you like.
Injector head T he injector head (only on instruments with injectors) is fixed to
the measurement c hamber cover by 2 silver-coloured cross-head bolts. It incorporates a ring-shaped raised sec tion with a sealing ring which sits on the opening on the side of the measurement chamber cover (Fig. 2-8).
Do not open the small cover on the injector head.
Fig. 2-8: Placing the injector head on the sample holder
Injector head
Ring-shaped raised
section with sealing ring
Measurement
chamber
Tube detector
Page 32
Chapter 2: Design of the Lumat LB 9507
2-10
2.5 Reagent Storage Area
A maximum of 2 reagent bottles are stored in an eas ily acces s ible recess on the top of Lumat. The containers are filled with the reagents required for the applications in hand, and linked to connectors 1 and 2 next to the storage bottles on the instrument by KEL/F screw joints (Fig. 2-9). Inside the instrument, thes e lines lead to the injectors, and from there to the injector head with its two injector tips.
To clean the injector lines, fill the tank s with washing solution and select the program function <WASH>.
Injector line T-piece Sealing ring T-piece
Hexagon bolt Injector l i ne
Fig. 2-9: Connection of the injector lines
Page 33
LB 9507 Chapter 2: Design of the Lumat
2-11
2.6 Variable injectors
High-precision pumps inject the reagent into the chamber. The user can define the injector volume. The integrated software allows you to select the injectors as requir ed (no injec tion, inj ector 1, injector 2, or injector 1 and 2) and to define their volumes in 5 µL steps in the range of 25 µL to 300 µL. If the user enters the injector volume not within the 5 µL step range, the system rounds up or off. Values out of the range are adjus ted to the limits ( 25 µL or 300 µL).
Flexible delay times can also be selected between the injections and measurement. Injectors developed by Berthold and the precise arrangement of the injector tips ens ure that the reagent is mixed with the sample immediately upon injection, and that exactly the right volume of reagent is used.
The injector pumps are located on the right hand side of the instrument beneath the reagent bottles (Fig. 2-10), and are covered by a metal plate.
Reagent 1 Reagent 2
Injector pump 1 Injector pump 2
Fig. 2-10: Lumat injectors
Page 34
Chapter 2: Design of the Lumat LB 9507
2-12
Injector lines
The connections for the injector lines on the instrument are defined as 1 (injector 1) and 2 (injec tor 2) . Use thes e des ignations when programming measurement protocols.
The injector lines lead to the corresponding pumps, and from there to the injector head. The right hand connection on the injector head comes from injector pump 1 (see Fig. 2-11).
Fig. 2-11: The connections injector pumps / injector head
Injector pump 2
Injector pump 1
Injector head
Measurement chamber
Page 35
LB 9507 Chapter 2: Design of the Lumat
2-13
2.7 Reagent Supply
Lumat is available in different versions:
1. without injectors (with septum for manual injections)
2. with one variable injector
3. with two variable injectors.
Setting the injector volume
For instruments with one or two injectors , you enter the desired injector or injectors together with their volumes once in the <INSTRUMENT PARAMETERS> menu. The sof tware us es these settings when creating protocols and perform ing measurements. When creating protocols the settings entered in the instruments parameter can be changed.
Setting ”0” means the respective inj ector is dis abled and does not appear in the protocol prompts.
Injector vol umes: 25 µL - 300 µL in 5 µL steps If the user enters the injector volume not within the 5 µL step range, the system rounds up or off. Values out of the range are adjusted to the limits (25 µL or 300 µL).
Example: When entering 43 µL the system rounds up to 45 µL, when entering 102 µL the system rounds down to 100 µL.
Wash function and selected injectors:
If the wash function is selected in the <REAGENTS> menu, only the injector lines are filled and washed in the case in which an injector volume > 0 has been entered in the <INSTRUMENT PARAMETERS> menu.
This manual descr ibes the version with 2 injectors, s ince this text covers all other options. So, if your Lumat does not include an injector, you can skip all inform ation relating to injectors such as selecting injectors when creating protocols, injecting during measurement, etc.
Page 36
Chapter 2: Design of the Lumat LB 9507
2-14
2.8 Memory Management
The Lumat LB 9507 memory manager has the following structure:
PC
Operating software with or without custom modules
CPU
Internal Memory
Data from 30 most recent runs
Max. of 40 measure­ment protocols
Instrument parameters
Instrument
configuration Measurement electronic Working memory
Fig. 2-12: Structure of the memory manager in Lumat LB 9507
Memory structure
The internal memory contains the instrument and configuration parameters, as well as the measurem ent protocols and data fr om the 30 most recent runs. Par t of this memory is used as a work ing memory for measurements and evaluations.
Storage security
A maximum of 40 measurement pr otocols can be created by the user, and these are stored in the instrument’s internal memory independent of current and battery, as are the instrument and configuration parameters . Therefore, these can only be lost if the CPU develops a fault, or similar.
In the event of a power failure, the data from the 30 most recent runs are stored in a battery-backed memory for at least 3 months.
Page 37
LB 9507 Chapter 2: Design of the Lumat
2-15
Operating - software
Fig. 2-13: Data transfer: internal memory - PC
Loading operating software
The instrument’s RS232 interface enables you to load the operating software from a PC to Lumat. From the terminal pr ogram, loading the operating software via a serial cable is a matter of a few mouse-clicks. This transfer proc edure should be carried out when you receive an update or upgrade on diskette or by email.
Procedure for loading the operating software
Use a PC to load the operating software from the diskette or email to Lumat. The existing sof tware on the m em ory will be over written and all measurement protocols on the memory will be deleted, that new protocols can be stored.
Switch Lumat on.Connect Lumat (host I/O board) and the PC using a serial
cable.
Prepare the instrument for receiving data by entering the
service function “S4”.
Page 38
Chapter 2: Design of the Lumat LB 9507
2-16
Start the term inal program, which is along the new operating
software on the diskette or email, and load the new software to Lumat. This overwrites the m emory with the new operating software, deleting all measurement protocols. For better instructions see text file "READ.ME", which is inc luded in the software update package.
In all cases, the contents of the inst rumen t’s in ternal memo ry remain unchanged (instrument and configuration parameters, measurement modes).
Page 39
LB 9507 Chapter 3: Getting Started
3-1
3. Getting Started
3.1 Instrument Setup
The place where the instrument is set up should have the following characteristics:
- free of dust
- no strong magnetic fields
- no direct exposure to sunlight
- ambient temperature +15 - +30°C
- relative humidity 10 - 90 %, no condensation
- power supply as specified on the type label
- distance of rear panel from the wall at least 10 cm to provide the necessary cooling
- operation area above the instrument at least 20 cm
- firm, smooth surface leaving the air filter inlet open
3.2 Power Supply
A selection of the line voltage is not necessary. The multi range power supply will detect the line voltage automatically , only make sure that the correct lin e fuse is fitted.
1 Amps slow blow for 230 V 2 Amps slow blow for 115 V
at a frequency of 47 - 62 Hz.
☞☞☞☞ Use only the cable supplied with the instrument to
connect it to the mains!
3.3 Checking the Paper Supply
Check if the paper supply is sufficient; for changing paper see section 13.1.
Page 40
Chapter 3: Getting Started LB 9507
3-2
3.4 Switching on Lumat
Please watch the display:
1. After the power is switched on, the following messages
should appear:
ROTATING TO INITIALISE
When this message appears, the tube transport rotor will perform a whole revolution. The ”ready” message then appears with the main menu:
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS-
If not, the microprocessor has detected an error (see Chapter
14.1).
2. If "MEMORY ERASED" is displayed together with the
"READY" message after the power is switched on, the stored data has been lost due to a memor y defect. This will also be indicated by a 3 second beep. You can work with the instrument after re-entering all configuration and instrument parameters.
Page 41
LB 9507 Chapter 3: Getting Started
3-3
3.5 System Configuration Parameters
The parameters listed in this chapter are valid f or all pr otoc ols and measurement types, and are therefore default values that are essential for operation of the instrument (s ee Chapter 14.4). The parameters entered rem ain stored in the lum inom eter, even if it is switched off (except for the time and date, which remain stored for approximately one month).
Please verify all configuration parameters when installing the instrument.
These parameters are entered in the <SYSTEM CONFIGURATION> menu. They describe the general mode of operation of the LB 9507.
The entry menu is accessed as follows: in the main menu ( with the READY display), press the k ey below <OTHERS>. Press this key again in the next display. The next display appears, containing the <SYSTEM CONFIGURATION> menu. Press the key below this option to display the configuration parameters.
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
SYST. CONFIG. INSTR.PARAM. -OTHERS-
Fig. 3-1: Operate the softkey typed in bold face to access the
configuration parameters.
Page 42
Chapter 3: Getting Started LB 9507
3-4
3.5.1 Description of the System Configuration Parameters
a) Calculating the Coefficient of Variation
<DATE> <TIME>
SYST. CONFIG. INSTR.PARAM. -OTHERS-
INSTRUMENT CONFIGURATION CV/CVD LANGUAGE -OTHERS-
COEFFICIENT OF VARIATION CALCULATED FROM RLU (C.V.) CONC. (CVD.)
<C.V./CVD.>: this option is available only when using the internal thermal printer. By selecting either <RLU (C.V.)> or <CONC. (CVD.)> you can specify here whether, in the assay mode, the calculated coefficient of var iation should be printed out in RLU or units of concentration.
b) Selecting the Language
INSTRUMENT CONFIGURATION CV/CVD LANGUAGE -OTHERS-
SELECT LANGUAGE ENGLISH GERMAN FRENCH
<LANGUAGE>: here you define the desired user language. German, English and French are available.
c) Setting the Date and Time
INSTRUMENT CONFIGURATION CV/CVD LANGUAGE -OTHERS-
INSTRUMENT CONFIGURATION DATE TIME -OTHERS-
<DATE>: enter date (DDMMMYY).
Page 43
LB 9507 Chapter 3: Getting Started
3-5
The date and time will continue to run for one month with the LB 9507 switched off. The built-in clock is supplied by a small rechargeable battery, making it independent of memor y activities in the instrument.
<TIME>: enter time and time format. You can select either 24­hours <24HR>, or <AM> and <PM>, which are displayed after you have selected <TIME> in the second line of the display. Separate the date and time by pressing the point key.
d) Beep
INSTRUMENT CONFIGURATION CV/CVD LANGUAGE -OTHERS-
INSTRUMENT CONFIGURATION DATE TIME -OTHERS-
INSTRUMENT CONFIGURATION PRINTER READY BEEP -OTHERS-
BEEP WHEN MEASUREMENT FINISHED : OFF OFF SHORT LONG
You can decide whether you want the luminometer to beep when the measurement is f inished. You can select SHORT, LONG or no (OFF) beep.
Page 44
Chapter 3: Getting Started LB 9507
3-6
3.5.2 System Configuration Table
Configuration Alternatives
Language German/English/French Coefficient of CV or CVD variation (only required for thermal printer) Time 0 to 24 hrs or AM/PM; entry Date entry Ready beeper short, long or off
Page 45
LB 9507 Chapter 3: Getting Started
3-7
3.6 Instrument Parameters
3.6.1 Setting/Verifying the Instrument Parameters
The instrument parameters are those instrument settings which are valid for all measurements. This is where the mechanical times for the injector pumps and the instrument sensitivity are standardised using a certain calibration factor.
For luminometers with one or two injectors installed, this is where you define which injector(s) is/are to be activated, along with their volumes. The activated injectors and injector volumes can be changed here (default values) or in the parameter protocols (protocol specified) when needed (see chapter 2.7 for volume ranges).
The best way to check the presently defined instrument parameters is by producing a param eter printout which is created as follows (press the softkeys in bold face):
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
SYST. CONFIG. INSTR. PARAM. -OTHERS-
ENTER CALIBRATION FACTOR FOR BLACK RLU 1.00 PRINT PARAM.
ENTER CALIBRATION FACTOR FOR RED RLU 1.00 PRINT PARAM.
These values are factory set and do not usually have to be changed. You will only have to enter them again in the event of a memory loss in which some or all of these data are des troyed. For your instrument, these parameters are:
Page 46
Chapter 3: Getting Started LB 9507
3-8
Cal. Factor for red RLU
Cal. Factor for black RLU
Injector volume Inj. 1 (µL):
Injector volume Inj. 2 (µL):
red: measurement chamber marked with a red spot black: measurement chamber unmarked.
Please enter these values listed in your test protocol before you start up your instrument and compare them with the printout. Correct the stored values only if the values in your printout and in the test protocol do not match (see chapter 14.4).
Page 47
LB 9507 Chapter 3: Getting Started
3-9
3.6.2 Changing the Instrument Parameters
The instrument parameters of the LB 9507 are edited as follows:
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
SYST.CONFIG. INSTR.PARAM. -OTHERS-
ENTER CALIBRATION FACTOR FOR RED RLU 1.00 PRINT PARAM.
ENTER CALIBRATION FACTOR FOR BLACK RLU 1.00 PRINT PARAM.
These values must match the RLU value of the test protocol. It refers to the inherent sensitivity of the photomultiplier . Confir m the value by pressing <enter>.
If a printout of the instrument parameters is required, press the button beneath <PRINT PARAM.>.
ENTER VOLUME INJ. 1 300 PRINT PARAM. LAST DISPLAY
Enter the volume of injector 1 in µL. This display appears only for instruments with at least one injector. If you enter “0”, injec tor 1 is disabled for all measurements.
ENTER VOLUME INJ. 2 [µl] 200 PRINT PARAM. LAST DISPLAY
Enter the volume of injector 2 in µL. This display appears only for instruments with 2 injectors. If you enter “0”, injector 2 is disabled.
Next, you are asked:
Page 48
Chapter 3: Getting Started LB 9507
3-10
CHANGES COMPLETE? YES NO PRINT PARAM.
Entering <YES> stores the parameters and returns the program to the menu. Entering <NO> returns the program to the first instrument parameter.
Page 49
LB 9507 Chapter 3: Getting Started
3-11
3.7 Reagent Supply
Depending on the requirements for the luminescence reaction, the instrument may operate with 0, 1 or 2 inject ors. Accordingly, you have to install one or two reagent bottles. Two empty bottles are included with the instrument. Kit manufacturers supply their specific activator and starter reagents in suitable bottles.
Technical data about the injector system
Injector volume: 25µL - 300µL in 5µL steps Injection time: less than 0.6 seconds Precision: ± 3µL for the total range;
<0.5% (300µL <0.1%)
Handling of injectors
Injector lines have to be washed before starting an assay. Fill lines with reagents. After each experim ent the injector lines have to be cleaned and the liquid has to be removed from the system.
Filling of the injector lines
Proceed as follows:
a) Switch on the instrument.
b) Make sure that the setting of the injectors is correct (see
<INSTRUMENT PARAMETERS>, section 3.8.1).
c) From the <OPERATOR FUNCTIONS> menu, select the
function <REAGENT> <PRIME> to fill up the supply lines and pump(s).
Pumping back the reagents
After measurem ent, the reagents left in the injector system, can be pumped back into the reagent bottles (see section 4.2.3, <MANUAL UNLOAD> function in the <OPERATOR FUNCTIONS> <REAGENT> menu).
Washing of injector lines after the measurement
After finishing measurement the injector system should be cleaned. Injector lines are made of T eflon/T ef zel and are resistant to a variety of different chemicals. For cleaning the lines you should use a lye or an acid (e.g. 0.3 M NaOH or 0.3 M HCL). Then rinse lines several tim es with H
2
0. Remove the liquid from
Page 50
Chapter 3: Getting Started LB 9507
3-12
the injection lines if the instrument will not be used for a long period of time.
Page 51
LB 9507 Chapter 3: Getting Started
3-13
3.8 Safety Precautions for Operation
1. Operate the instrument in dry environments only.
2. Be careful when handling strong alkaline or acidic reagents. Be sure to follow the reagent manufacturer’s safety instructions (R&S-set).
3. If there is a leak in the reagent tubing system you must abort the measurement s eries and eliminate the leak. If nec essary, call the manufacturer’s T echnical Ser vice. If there is a leak the reagents may flow on the instrument desk.
4. Only service technicians authorized by Berthold Technologies are allowed to repair or modify the instrument.
Page 52
Page 53
LB 9507 Chapter 4: Structure and Operation of Software
4-1
4. Structure and Operation of Softw are
4.1 Operation
The software of the Lumat LB 9507 is menu-based. The main menu comprises 5 subm enus c ontaining a wide range of progr am functions. Commands or parameters are entered via the respective menus.
You can select a menu and run through the program by means of function keys and so-called softkeys. Softkeys are keys with changing functions defined by the software and shown on the display. The functions are selected using the corresponding buttons beneath the display.
There are three function keys used for operating the program, each with a fixed meaning:
<enter> to confirm entries <delete> to delete the most recently entered text or numerical
values
<exit> to abort a command or program procedure and to
return to the main menu
The keyboard includes a complete s et of alphanumeric characters and a spacebar.
6
exit delete last enter
1 2 3 4 5 6 7 8 9 0 .
Q W E R T Y U I O P
A S D F G H J K L -
space Z X C V B N M /
21
Fig. 4-1: Keyboard of Lumat LB 9507
5
4
3
Page 54
Chapter 4: Structure and Operartion of Software LB 9507
4-2
4.2 The Six Functions of the Main Menu
You can select various functions of the main menu from the READY screen on the Lumat LB 9507. By pressing the softkey <OTHERS> you can switch through a loop containing the various menu displays. If no entry follows in 10 seconds, the program automatically returns to the READY display with the <MEASUREMENT> and <PROTOCOL> menus.
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
SYST. CONFIG. INSTR. PARAM. -OTHERS-
Fig. 4-2: Main menu options
Pressing the other softkeys opens the associated menus and leads you to the options involved. The following sections provide an overview.
Page 55
LB 9507 Chapter 4: Structure and Operation of Software
4-3
4.2.1 <MEASURE> Menu
The <MEASURE> menu offers various options for running a measurement:
READY MEASURE PROTOCOL -OTHERS-
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
Fig. 4-2: Measurement options
<PROTOCOLS> This calls up a stored protocol which defines
the measurement mode (raw data measurement, LIA, Cut-of f, T UP/FTI, etc .) and the parameters for the following m eas urement, and starts the measurement. Depending on the selected protocol you will be given instructions on how to operate Lumat.
<DIRECT ENTRY>Here you can enter parameters f or a raw data
measurement and then star t the m eas urem ent directly. The most recently entered param eter set remains stored.
<RATEMETER> You can perform a continuous measurement
of any length you choose, and the measurement intervals can be selected as required. During the measurement, any number of injections can be performed using both injectors.
Page 56
Chapter 4: Structure and Operartion of Software LB 9507
4-4
4.2.2 <PROTOCOL> Menu
In the <PROTOCOL> menu you select the protocol types for the desired measurement mode and enter the parameters for the measurement. T he following protocol types can be selected after you have selected the option <CREATE> in this menu:
READY MEASURE PROTOCOL -OTHERS-
PROTOCOL MASTER CURVE CREATE CHANGE
06: PROTOCOL NAME
<enter>
06: PROTOCOL TYPE RAW DATA CUTOFF -OTHERS-
06: PROTOCOL TYPE QUANT TUP/FTI -OTHERS-
06: PROTOCOL TYPE DL-ASSAY REPEATED -OTHERS-
Fig. 4-4: Selection of protocol types
<RAW DATA> Protocol for raw data measurements. <CUTOFF> Protocol for qualitative measurements. Using negative and
positive standards, RLU-ranges are def ined which are marked in the printout of results.
<QUANT.> Protocol for LIA measurements with or without master curve. To
run an LIA measurement with m as ter cur ve you only have to enter a short protocol here. Select <MASTER CURVE> and enter the data for the Master Curve in this m enu. A master curve can only be attached to an existing LIA protocol without its standard series.
Page 57
LB 9507 Chapter 4: Structure and Operation of Software
4-5
<TUP/FTI> This protocol type is used to calculate the patient data using a
standard.
<DL-ASSAY> Automatic measurement of the Dual-Luciferase
TM
Reporter
Assay. (Injection-measure-injection-measure)
<REPEATED> Protocol for the meas urement of long time k inetics us ing repeated
measurements of the same sample within a fixed timed interval; with or without injection.
Page 58
Chapter 4: Structure and Operartion of Software LB 9507
4-6
4.2.3 <OPERATOR FUNCTIONS> Menu
This menu includes the following user functions (for a description see Chapter 11):
<REAGENT> Loading, refreshing and removing reagents and washing of
injectors <PERF. TEST> Instrument and reagent test <MANUAL DATA > Manual entry of RLU data for calculation with evaluation protocols <RERUN> Calls up stored data for re-calculation and printout.
READY MEASURE PROTOCOL -OTHERS-
READY OPER. FUNCT. TURN ONCE -OTHERS-
OPERATOR FUNCTIONS: REAGENT PERF.TEST -OTHERS-
OPERATOR FUNCTIONS: MANUAL DATA RERUN -OTHERS-
Fig. 4-5: Operator functions
The function <REAGENT> is only provided for instruments with at
least one injector installed.
Page 59
LB 9507 Chapter 4: Structure and Operation of Software
4-7
4.2.4 <SYSTEM CONFIGURATION> Menu
Here you define the instrument configuration parameters: <CV/CVD> Specify here whether the coefficient of variation is to be s pecified
in RLU (=C.V.) or in concentration units (CVD) on the thermal
printout. <LANGUAGE> Set the language which you want to use in the integrated
software’s user interface (German/English/French). <DATE> Enter the date. <TIME> Enter the time. <BEEP> Switching on and off the internal beep when the m easurement is
finished
READY MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
<DATE> <TIME>
CONFIG. INSTR. PARAMS -OTHERS-
INSTRUMENT CONFIGURATION CV/CVD LANGUAGE -OTHERS-
INSTRUMENT CONFIGURATION DATE TIME -OTHERS-
INSTRUMENT CONFIGURATION READY BEEP -OTHERS-
Fig. 4-6: Instrument configuration
Page 60
Chapter 4: Structure and Operartion of Software LB 9507
4-8
4.2.5 <INSTRUMENT PARAMETER> Menu
Enter and verify the instrument parameters, e.g. the volume of the
injectors used or the calibration factors (s ee section 3.8). Fig. 4-6
shows you how to select this menu.
4.2.6 <TURN ONCE> function
When you select this func tion, the measurem ent cham ber rotates
through 180°. This allows you to move a sample tube into the
measurement position or out of the measurem ent position while
not actually measuring (useful when cleaning).
READY MEASURE PROTOCOL -OTHERS-
<DATE> <TIME>
OPER. FUNCT. TURN ONCE -OTHERS-
Fig. 4-7: Selecting the <TURN ONCE> function
4.2.7 Service functions
The program contains the following service functions; these are
only accessible to Berthold Technologies’ customer service
engineers.
<INJECT. TEST> Injector test.
<PM TEST> Tests the photomultiplier.
<ROTOR TEST> Tests the measurement chamber rotor.
<MONITOR> Loads new operating software.
Page 61
LB 9507 Chapter 4: Structure and Operation of Software
4-9
4.3 General Procedure
The following flow chart provides an overview of the essential
functions of Lumat, together with the way in which they are
interrelated. More detailed information is available in the c hapters
listed.
1. Getting Started
The instrument parameters are defined and the instrument is
configured during the commissioning procedure.
2. Creating protocols for routine measurements
Up to 40 protocols can be stored in the <PROTOCOL> menu for
routine measurements. Four different types of protocols are
available.
3. Measurements
The <MEASURE> menu includes the following options:
Measurements based on stored protocols
Raw data measurement with direct entry of the parameters
Kinetics measurement
4. Operator Functions
The <OPERATOR FUNCTIONS> menu provides the following
options:
Data evaluation with manual data entry
Evaluation of stored data
Performing a system test
Washing of the injector system
Re-pumping of the reagents
Page 62
Chapter 4: Structure and Operartion of Software LB 9507
4-10
Fig. 4-8: Illustration of the general procedure
1. Getting started Chapter 3
Connect instrument
Setting the instrument parameters
Injector volume, calibration factors etc.
Instrument configuration
Language, date, time, printer
Calculating the coefficient of variation
Chap. 5 Chap. 6 Chap. 8 Chap. 7 Chap. 7 Chap. 9 Chap. 5 Chap. 10
2. Protocol setu
p
Raw data
Cut-off UP/FTI Quant. Master
curve
DLR
A
ssa
y
3. Measurements
Raw data
meas.
Cut-off
meas.
UP/FTI
meas.
Quant.
meas.
DLR
Assay
Raw data
direct meas.
Kinetic
meas.
4. Operator functions Chapter 11
Evaluation
of stored data
Reagents Performance
Test
Data evaluation
with manual data
input
Prime reagents
Page 63
LB 9507 Chapter 5: Raw Data Measurements
5-1
5. Raw Data Measurements
5.1 Introduction
In the Raw data measurement mode the quantity of light
measured over a user-defined measuring time is measured in
RLUs for each sample. You have the option of printing out a high
resolution (up to 2000 data point) curve and/or up to 50 data
points with measurement time for each sample, which enables
you to analyze the kinetic behaviour of the light emission of a
sample. Replicate groups can also be defined for which the aver age value
and the variation coefficient are calculated and printed out.
This measurement mode, which allows up to 2 injections per
sample, is used mainly for ATP measurements and for
measurement of luminescence reporter gene assays.
There are two ways of starting raw data measur ement (see Fig.
5-1):
1. Create and save protocols for routine measurements in the <PROTOCOL> menu (max. 40). W ith each of these protocols you can later start as many measurements as you want in the <MEASURE> menu.
2. Running a raw data direct measurement. Select the option < MEASURE > from the <DIRECT ENT RY>
menu to enter the measurement parameters. Then start the measurement.
In both cases the parameters you are asked to enter are the same.
The displays are numbered consecutively from R1... R34 for clarity.
The following illustration shows a flow chart of the protocol setup procedure and measurement start for routine and direct measurements.
Page 64
Chapter 5: Raw Data Measurements LB 9507
5-2
Direct measurements
R1
READY MEASURE PROTOCOL -OTHERS-
R2
PROTOCOL MASTER CURVE CREATE CHANGE
R3 R2a
R2b
R4 R2b
R2c
R5
06: OPERATOR ID.
(…)
R19
06: INPUT ALL CORRECT? YES YES NO
R1
READY MEASURE PROTOCOL -OTHERS-
R21
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
R22 R6
R19
Fig. 5-1: Flow chart showing protocol setup and measurement
Create protocol
Change protocol
06: PROTOCOL NAME
06: PROTOCOL NAME RAW DATA CUTOFF -OTHERS-
06: PROTOCOL NUMBER 6 PRINT LIST
PROTOCOL 6 CHANGE OTHERPROT.PRINT PROT.
DELETE PROTOCOL? NO YES
Routine measurements
ENTER PROTOCOL NUMBER 6
06: INPUT ALL CORRECT ? YES YES NO
06: ENTER INJECTOR VOLUME 1 [µL] PRINT LIST LAST DISPLAY
Start of measurement
Page 65
LB 9507 Chapter 5: Raw Data Measurements
5-3
5.2 Creating Raw Data Protocols
For routine measurements, start by entering the parameters for the measurement procedure and the evaluation strategy in a measurement protocol in the <PROTOCOL> menu, and then save them. Later, you can run a measurem ent simply by calling up one of these stored protocols in the <MEASURE> menu and pressing the <START> button.
You can save up to 40 different measurement protocols which include all information such as injector control, definition of replicates, measuring time etc., as well as information on data evaluation. Each new protocol is given a number which is dependent upon the time the protocol was set up and independent of the protocol type (raw data, cut-off...).
To create a protocol you have to run through displays R1 to R18. Most displays include the option <LAST DISPLAY> which takes you back to the previous display.
Page 66
Chapter 5: Raw Data Measurements LB 9507
5-4
R1
READY MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the button located directly below. The following display appears:
R2
PROTOCOL MASTER CURVE CREATE CHANGE
The <MASTER CURVE> option is not relevant to raw data measurements. Select <CHANGE> to open an existing protocol by entering its protocol number. You can then edit it. To create a new protocol, select <CREATE>. You will see this:
R3
06: PROTOCOL NAME ATP-10SEC-1INJ
The next free number is displayed for the new protocol (in our example “06:”). Enter a name for this protocol using the alphanumeric keyboard. You should use a name which specifies the protocol type and other important features, so that it is easy to identify later. For example “ATP-10SEC-1INJ”. Confirm your entry with <enter>.
R4
06: PROTOCOL TYPE RAW DATA CUTOFF -OTHERS-
Select <RAW DATA>. Now you have to enter the parameters which are the same for all protocol types:
R5
06: OPERATOR ID.
Enter the user name and press <enter>.
Page 67
LB 9507 Chapter 5: Raw Data Measurements
5-5
R6
06: ENTER INJECTOR VOLUME 1 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 1 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. Range: 25 µL - 300 µL in 5 µL steps.
If the user enters the injector volume not within the 5 µL step range, the system rounds up or down. Values out of the range are adjusted to the limits (25 µL or 300 µL). Example: W hen entering 43 µL the system rounds up to 45 µL, when entering 102 µL the system rounds down to 100 µL.
Press <enter> to use the response displayed in the upper-right corner of the LCD.
R7
06: ENTER INJECTOR VOLUME 2 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 2 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. <LAST DISPLAY> takes you back to the previous display. Setting "0" takes you to display R10. Setting >0 takes you to display R8. Press <enter> to use the response displayed in the upper-right corner of the LCD.
R8
06: SEQUENCE OF INJECTIONS 1->2 1->2 2->1 LAST DISPLAY
If you have selected 2 injectors, you have to define the sequence of the injections here: injector 1 first and then injector 2 or vice versa. Make your selection by pressing the button below the option you want. The following display appears:
Page 68
Chapter 5: Raw Data Measurements LB 9507
5-6
R9
06: DELAY TIME INJ1/INJ2 1.2 (1.2 - 300.0) LAST DISPLAY
Enter the desired delay in seconds (1.2 to 300 sec) between both injections and confirm with <enter>. Only if you have selected one or two injectors:
R10
06: MEASURE BACKGROUND NO YES NO LAST DISPLAY
Select here if the background r ate is to be measured prior to the injections and the actual measurement of each tube. If you select <NO>, display R14. <YES> takes you to the next display:
R11
06: MEASURING TIME BKG [s] 0.5 (0.1 - 200.0) LAST DISPLAY
Enter between 0.1 and 200.0 seconds for the background measurement and pres s <enter>. Reasonable times start at half a second, allowing statistical variations to be compensated for. Otherwise, the background values may be too high or too low which either distorts the result of the m easurem ent or leads to an exceeding of the threshold.
Remember that the background effect is measured in order to obtain an alarm threshold for background rates that ar e too high, as with phosphorescence effects for example.
R12
06: AUTOMATIC BKG SUBTRACTION NO YES NO LAST DISPLAY
Here you can select whether the background effect is to be subtracted automatically from the measured results.
Remember that the statis tic al signif ic ance of the meas ured values is dependent upon the measuring time.
Page 69
LB 9507 Chapter 5: Raw Data Measurements
5-7
R13
06: MAX. BACKGROUND [RLU/s] 50 (0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this threshold is exceeded, you will be alerted accordingly on the display and the printout (BKG). Entering “0” means “no background threshold“. As soon as you confirm the entry with <enter> you move to the next display.
R14
06: DELAY LAST INJ./MEAS. [s] 1.0 (0.0 - 300 s) LAST DISPLAY
Enter the delay between the last injection and the start of the measurement in the range from 0.0 to 300 seconds and press <enter>.
R15
06: MEASURING TIME [s] 2.0 (0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each tube and press <enter>.
R16
06: NUMBER OF REPLICATES 2 (1 - 10) LAST DISPLAY
Enter the number of replicates and pres s <enter>. You can define a maximum of 10 sample tubes as replicates whose measured values are averaged. “1” means 1 tube, i.e. no replicates.
R17
06:KINETICS PRINTOUT DATA YES YES NO LAST DISPLAY
The measurement results can be printed in the form of kinetics data on the built-in thermal printer.
Page 70
Chapter 5: Raw Data Measurements LB 9507
5-8
R18
06:KINETICS PRINTOUT CURVE YES YES NO LAST DISPLAY
The measurement results can be printed in the form of kinetics curve on the built-in thermal printer.
R19
06: INPUT ALL CORRECT ? YES YES NO LAST DISPLAY
If you select <YES> the entered parameters are stored and the program returns to the main menu (R1). <NO> takes you back to the 1st display of this protocol (R5).
Note:
The protocol is not stored if you exit this protocol sequence via <exit>!
Page 71
LB 9507 Chapter 5: Raw Data Measurements
5-9
5.3 Editing/Deleting Protocols
R1
READY MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the button located directly below. Then follows:
R2
PROTOCOL MASTER CURVE CREATE CHANGE
Select <CHANGE>.
R2a
ENTER PROTOCOL NO. 6 PRINT LIST
You can print the list of stored protocols by selecting <PRINT LIST>. Enter the required protocol number and press <enter>.
R2b
PROTOCOL # 6: ATP-10SEC-1INJ CHANGE OTHER PROT. PRINT PROT.
Select <PRINT PROT.> to print out the selected protocol. <OTHER PROT.> takes you back to the previous display and lets you enter a new protocol number. < CHANGE> takes you to the next display:
R2c
DELETE PROTOCOL NO. 6: NO YES
To delete this protocol, select <YES>. To edit the protocol, select <NO>. If you do this, the first entry of this protocol is displayed, i.e. the user name in display R5:
Page 72
Chapter 5: Raw Data Measurements LB 9507
5-10
R5
06: OPERATOR ID
This and all the following entries can be overwritten. T he following input sequence is the same as the one described in the last section starting with display R6. If you want to leave certain parameters as they are, just press <enter>.
Page 73
LB 9507 Chapter 5: Raw Data Measurements
5-11
5.4 Performing Raw Data Measurements
There are two ways of performing a raw data measurement:
1. You can use a protocol which has already been created and
stored. To do this, select the option <PROTOCOLS> in the <MEASURE> menu.
2. Alternatively, enter the param eters for the m easurement you
want to run. To do this, select the option <DIRECT ENTRY> in the <MEASURE> menu and enter the measurement parameters. These are the param eters listed in displays R5 to R19. Then start the measurement.
Measurement with stored protocol
R1
READY MEASURE PROTOCOL -OTHERS-
Select <MEASURE>.
R21
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
Select <PROTOCOLS> to open a stored protocol. Select <DIRECT ENTRY> to activate the input sequence for the measurement parameters of a direct measurement.
R22
ENTER PROTOCOL NO. 6 PRINT LIST
Enter the number of the protocol required and press <enter>. You can print out the list of stored protocols including num ber and name by pressing the button below <PRINT LIST>.
Page 74
Chapter 5: Raw Data Measurements LB 9507
5-12
R23
06: PROTOCOL ATP-10SEC-1INJ ? YES NO
You are prompted to confirm that this is the required pr otocol. If you select <NO> the program returns to display R22. <YES> leads to:
R24
COMMENT
You can enter a comment on the following measurement using the alphanumeric keyboard. Confirm it with <enter>. If you do not want to enter a comment, just press <enter>.
R28a
SAMPLE 1 TUBE # 1 INSERT TUBE...
Load the tube containing the first sample as defined in the protocol.
R29
SAMPLE 1 TUBE # 1
START
Press the button beneath the <START> function or the <start> button to the right of the rotor. Both buttons do the sam e thing. The tube will move back into the measurem ent cham ber, and the programmed measurement procedure will begin:
R30
BACKGROUND MEASURING...
Background measur ement is performed, assum ing you specified it in the protocol.
Page 75
LB 9507 Chapter 5: Raw Data Measurements
5-13
R31
BACKGROUND TOO HIGH !
This display only appears when the background threshold you have defined in the protocol is exceeded. If this happens the printout of measurement values is flagged BKG.
The steps which the instrum ent goes through are displayed and the measurement param eters f rom the active pr otocol are printed out at the same time.
R32
INJECTING ...
R33
MEASURING...
R34
SAMPLE-RLU 133
REMOVE TUBE
Once measurem ent is complete, the first line shows the r esult of the measurement in RLUs, the second line (if defined in the protocol) shows the result of the background measurement. At the same time the m easurement parameters defined in the protocol are printed out as a protocol header. The measured values are then printed:
Measured values output:
No kinetics curve, no replicates selected:
One measured value per sample (RLU)
No kinetics curve, but replicates:
One measured value per sample (RLU) Average for replicate group and coefficient of variation
Page 76
Chapter 5: Raw Data Measurements LB 9507
5-14
Kinetics curve (without data) selected without replicates:
Up to 2000 measurement points per sample displayed as a curve (according to the measurement time) Peak maximum and total integral of each sample
Kinetics curve (with data) with replicates:
20 to 50 measurement points per sample Up to 2000 measurement points per sample displayed as a curve Peak maximum and total integral of each sample Average for replicate group and coefficient of variation
When you remove the sa mple tube, the measured value display disappears and the following is requested:
R36
SAMPLE 2 TUBE # 2 INSERT TUBE ...
Load the next sample tube.
R37
SAMPLE 2 TUBE # 2
START
Press the <Start> button. The 2nd tube is measured. If it is a replicate, “SAMPLE 1” will remain on the display until all tubes of this group have been measured, while the “TUBE #” is incremented continuously.
After pressing the <Start> button, the procedur e is the same as described for steps R29 to R34.
Page 77
LB 9507 Chapter 6: CUT-OFF-Measurements
6-1
6. CUT-OFF Measurements
6.1 Introduction
The aim of CUT -OFF measurem ents is to divide patient sam ples into 3 qualitative classes separated by 2 CUT-OFF thresholds (LOW and HIGH CUT-OFF) using fixed parameters and controls.
The samples are marked as follows:
RLU value below LOW CUT-OFF ”NEG” RLU value between LOW and HIGH CUT-OFF ”+/-” RLU value above HIGH CUT-OFF “POS”
Both CUT-OFF thresholds are calculated as follows:
LOW CUT-OFF = A
L
+ BL·(neg. std) + CL (1st pos. ctrl)
HIGH CUT-OFF = A
H
+ BH·(neg. std) + CH (1st pos. ctrl)
where the following are defined in the protocol:
A
L
= LOW CUT-OFF offset (RLU) Display no. C22
A
H
= HIGH CUT-OFF offset (RLU) C25
B
L
= LOW CUT-OFF FACTOR C23
influence of the neg. control on LOW CUT-OFF
B
H
= HIGH CUT-OFF FACTOR C26
influence of the neg. control on HIGH CUT-OFF
C
L
= LOW CUT-OFF LPOS. FACTOR C24
influence of the pos. control on LOW CUT-OFF
C
H
= HIGH CUT-OFF LPOS. FACTOR C27
influence of the pos. control on HIGH CUT-OFF
This program can be flexibly used with these parameters.
Example 1 Setting the number of controls to zero results in the CUT-OFF
thresholds being defined only from the OFFSET values.
Example 2 Setting the number of the pos. controls to zero res ults in the CUT-
OFF threshold being defined from the neg. control and - when entered - from the OFFSET values.
Example 3 Setting B
H
and CL to zero results in the CUT-OFF thresholds being
defined only from the neg. resp. pos. controls and - when entered
- from the OFFSET values.
Page 78
Chapter 6: CUT-OFF-Measurements LB 9507
6-2
Since negative controls often represent an important quality criterion, limits can be s et in the protocol outside which a flag is printed. (LOW/HIGH RLU FLAG NEG. STD.)
Also, with replicate measurem ents, the values may be flagged if their variations are too high. (% CV FLAG FOR NEG./POS. CTRLS./SAMPLES)
These options can each be turned off by entering zero.
The displays are numbered consecutively C1, C2 ...CX for clarity.
The illustration on the next page shows a flow chart from creating protocols to starting a measurement.
Page 79
LB 9507 Chapter 6: CUT-OFF-Measurements
6-3
C1
READY MEASURE PROTOCOL -OTHERS-
C2
PROTOCOL MASTER CURVE CREATE CHANGE
C3 C2a
C2b
C4 R2b
C2c
C5
07: OPERATOR ID.
(…)
C33
07: INPUT ALL CORRECT? YES YES NO
C1
READY MEASURE PROTOCOL -OTHERS-
C41
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
C42
ENTER PROTOCOL NO. 7 PRINT LIST
Fig. 6-1: Flow chart showing protocol setup and measurement
Create protocol
Change protocol
07: PROTOCOL NAME
07: PROTOCOL NAME RAW DATA CUTOFF -OTHERS-
07: PROTOCOL NUMBER 7 PRINT LIST
PROTOCOL 7 CHANGE OTHERPROT.PRINT PROT.
DELETE PROTOCOL? NO YES
Measurements
Start of measurement
Page 80
Chapter 6: CUT-OFF-Measurements LB 9507
6-4
6.2 Creating CUT-OFF Protocols
Start by entering the parameters for the measur ement procedure and the evaluation method in a measurement protocol in the <PROTOCOL> menu, and then save them. Later, you can run a measurement simply by calling up one of these stored protocols in the <MEASURE> menu and pressing the <START> button.
You can save up to 40 different measurement protocols which include all information such as injector control, definition of replicates, measuring time etc., as well as information on data evaluation. Each new protocol is given a number which is dependent upon the time the protocol was set up but not on the protocol type.
To create a protocol you have to run through displays C1 to C32.
Page 81
LB 9507 Chapter 6: CUT-OFF-Measurements
6-5
C1
READY MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the respective button. Then follows:
C2
PROTOCOL MASTER CURVE CREATE CHANGE
The <MASTER CURVE> is not relevant to CUT-OFF measurements. If you press the button below <CHANGE> you can open an existing protocol by entering its protocol number, and then edit it. To create a new protocol, press the button below <CREATE>. The following display appears:
C3
07: PROTOCOL NAME CUT-10SEC-1INJ
The next free number is displayed for the new protocol (in this example “07:”). Enter a name for this protocol using the alphanumeric keyboard. You should use a name which specifies the protocol type and other important features, so that it is easy to identify later. For example “CUT-10SEC-1INJ”. Confirm your entry with <enter>.
C4
07: PROTOCOL TYPE RAW DATA CUTOFF - OTHERS -
Select <CUTOFF>. Now you have to enter the parameters which are the same for all protocol types:
C5
07: OPERATOR ID
Enter the user name and press <enter>.
Page 82
Chapter 6: CUT-OFF-Measurements LB 9507
6-6
If you have no injectors selected (see 3.8.2) the display C15 follows. Otherwise C6 follows.
C6
07: ENTER INJECTOR VOLUME 1 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 1 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. Range: 25 µL - 300 µL in 5 µL steps. If the user enters the injector volume not within the 5 µL step range, the system rounds up or down. Values out of the range are adjusted to the limits (25 µL or 300 µL). Example: W hen entering 43 µL the system rounds up to 45 µL, when entering 102 µL the system rounds down to 100 µL. Press <enter> to use the response displayed in the upper-right corner of the LCD.
C7
07: ENTER INJECTOR VOLUME 2 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 2 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. <LAST DISPLAY> takes you back to the previous display. Setting "0" takes you to display C10. Setting >0 takes you to display C8. Press <enter> to use the response displayed in the upper-right corner of the LCD.
C8
07: SEQUENCE OF INJECTIONS 1->2 1->2 2->1 LAST DISPLAY
If you have selected 2 injectors, you have to define the sequence of the injections here: injector 1 first and then injector 2 or vice versa. Make your selection by pressing the button below the
Page 83
LB 9507 Chapter 6: CUT-OFF-Measurements
6-7
option you want. Press <enter> to use the response displayed in the upper-right corner of the LCD. The following display is shown:
C9
07: DELAY TIME INJ 1/INJ 2 1.2 (0.7 - 300.0) LAST DISPLAY
Enter the desired delay in seconds between both injections and confirm with <enter>.
C10
07: MEASURE BACKGROUND NO YES NO LAST DISPLAY
Select here whether the background rate is to be measured prior to the injections. If you select <NO>, display C14 appears. <YES> takes you to the next display:
C11
07: MEASURING TIME BKG [s] 0.5 (0.1 - 200.0) LAST DISPLAY
Enter between 0.1 and 200.0 seconds for the background measurement and pres s <enter>. Reasonable times start at half a second, allowing statistical variations to be compensated for. Otherwise, the background values may be too high or too low which either distorts the result of the m easurem ent or leads to an exceeding of the threshold. Remem ber that the back gr ound eff ect is measured in order to obtain an alarm threshold f or back ground rates that are too high, as with phosphorescence for example.
C12
07: AUTOMATIC BKG SUBTRACTION NO YES NO LAST DISPLAY
Here you can select whether the background effect is to be subtracted automatically from the measured results.
Page 84
Chapter 6: CUT-OFF-Measurements LB 9507
6-8
C13
07: MAX. BACKGROUND [RLU/s] 50 (0 = NONE) LAST DISPLAY
Here can you define the limit value for the background. If this threshold is exceeded, a direction is displayed and printed out. Entering “0” means “no background threshold“. As soon as you confirm the entry with <enter> you move to the next display.
C14
07: DELAY LAST INJ./MEAS. [s] 1.0 (0.0 - 300) LAST DISPLAY
Enter the delay between the last injection and the start of the measurement in a range from 0.0 to 300 seconds and press <enter>.
C15
07: MEASURING TIME [s] 2.0 (0.1 - 200.0) LAST DISPLAY
Enter the desired measuring time (0.1 to 200.0 seconds) for each tube and press <enter>.
C16
07: REPL. OF NEG. CTRL. (0-10) 3
LAST DISPLAY
Enter the number of replicates of the negative control and press <enter>. You can define a maximum of 10 replicates whose measured values are averaged.
C17
07: NUMBER OF POS. CTRL (0-3) 2
LAST DISPLAY
Enter the number of positive controls (0 to 3 allowed). If you enter 0, the displays C18 and C19 are skipped.
Page 85
LB 9507 Chapter 6: CUT-OFF-Measurements
6-9
C18
07: REPL. OF POS. CTRL.1 (1-10) 2
LAST DISPLAY
Enter the number of replicates of the first positive control (up to 10 allowed).
C19
07: REPL. OF POS. CTRL.2 (1-10) 2
LAST DISPLAY
Enter the number of replicates of the second positive control (up to 10 allowed).
C20
07: REPL. OF SAMPLES (1-10) 2
LAST DISPLAY
Enter the replicate number for all patient samples (1-10).
C21
07: LOW CUTOFF OFFSET 120
LAST DISPLAY
C22
07: LOW CUTOFF FACTOR 1.0
LAST DISPLAY
C23
07: LOW CUTOFF LPOS FACT. 0.0
LAST DISPLAY
These values are used to calculate the low CUT-OFF values: Low Cut-off = Offset + m ean value (NEG) * Low Cut-of f factor + mean value (1st POS.CTRL.) * LPOS. FACTOR (see section 6.1).
C24
07: HIGH CUTOFF OFFSET 120
LAST DISPLAY
Page 86
Chapter 6: CUT-OFF-Measurements LB 9507
6-10
C25
07: HIGH CUTOFF FACTOR 0
LAST DISPLAY
C26
07: HIGH CUTOFF LPOS FACT. 1
LAST DISPLAY
The formula for the HIGH CUT-OFF threshold is similar to this one.
C27
07: %CV FLAG FOR NEG. CTRL. 15
LAST DISPLAY
The negative control mean value is labelled “ HICV” in the result printout if its coefficient of variation is higher than the value entered here.
C28
07: %CV FLAG FOR POS. CTRL. 20
LAST DISPLAY
Positive standard mean values are labelled “HICV” if their coefficients of variation are higher than the value entered here.
C29
07: %CV FLAG FOR SAMPLES
LAST DISPLAY
If this CV limit is exc eeded by patient samples, a flag appears in the printout.
C30
07: LOW RLU FLAG NEG. CTRL. 200
LAST DISPLAY
Page 87
LB 9507 Chapter 6: CUT-OFF-Measurements
6-11
C31
07: HIGH RLU FLAG NEG. CTRL. 600
LAST DISPLAY
Lower and upper limit for marking negative controls: a FLAG appears in the printout when the value is not within the specified value range. “LOW” if measured value < LOW RLU FLAG “HIGH” if measured value > HIGH RLU FLAG
C32
07: INPUT ALL CORRECT ? YES YES NO LAST DISPLAY
This display appears at the end of the input sequence. If you press the button below <YES> the entered parameters are stored and the program returns to the m ain menu (C1) . <NO> takes you back to the 1st display of this protocol (C5).
Note:
The protocol is not stored if you exit this protocol sequence via <exit>!
Page 88
Chapter 6: CUT-OFF-Measurements LB 9507
6-12
6.3 Editing/Deleting Protocols
C1
READY <DATE> <TIME> MEASURE PROTOCOL -OTHERS -
Select the <PROTOCOL> menu by pressing the respective button. Then follows:
C2
PROTOCOL MASTER CURVE CREATE CHANGE
Select <CHANGE>.
C2a
ENTER PROTOCOL NO. 7 PRINT LIST
The list of stored protocols can be printed by selecting <PRINT LIST>. Enter the required protocol number and press <enter>.
C2b
PROTOCOL # 7: CUT-10SEC-1INJ CHANGE PROT. OTHER PROT. PRINT PROT.
Select <PRINT PROT.> to print out the selected protocol. <OTHER PROT.> takes you back to the previous display and lets you enter a new protocol number. <CHANGE PROT.> takes you to the next display:
C2c
DELETE PROTOCOL # 7: NO YES
To delete this protocol, select <YES>. To edit the protocol, select <NO>. If you do this, the first entry of this protocol is displayed, i.e. user name in display C5:
Page 89
LB 9507 Chapter 6: CUT-OFF-Measurements
6-13
C5
07: OPERATOR ID
This and all following entries can be overwritten. The following input sequence is the same as the one described in the last section starting with display C6.
Page 90
Chapter 6: CUT-OFF-Measurements LB 9507
6-14
6.4 Running Cut-off Measurements
C1
READY MEASURE PROTOCOL -OTHERS-
Select <MEASURE>.
C41
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
Select <PROTOCOLS> to open a stored protocol.
C42
ENTER PROTOCOL NO. 7 PRINT LIST
The list of stored protocols including number and name can be printed by pressing the button below <PRINT LIST>. Enter the number of the desired protocol and press <enter>.
C43
07: PROTOCOL CUT-10SEC-1INJ O.K. ? YES NO
You are prompted to confirm that this is really the protocol you want. If you select <NO> the program returns to display C42. <YES> leads to:
C44
COMMENT
You can enter a comment on the following measurement using the alphanumeric keyboard. Confirm it with <enter>. If you do not want to enter a comment, just press <enter>.
Page 91
LB 9507 Chapter 6: CUT-OFF-Measurements
6-15
C48
REMOVE TUBE...
Remove the tube from the sample holder if there is one. This display appears if there is a tube in the sample holder. So, if a tube already is in the correct position before you start measurement, you will have to lift the tube briefly. You will then see:
C48a
NEG. CTRL REPL.#1TUBE#1 INSERT TUBE ...
Load the tube containing the first negative control (as defined in the protocol).
C49
NEG. CTRL REPL.#1TUBE#1
START
Press the <start> button. The tube will move back into the measurement chamber, and the programmed measurement procedure will begin:
C50
BACKGROUND MEASURING... REMOVE TUBE...
A background measurement is performed, provided you have defined one in the protocol.
C51
BACKGROUND TOO HIGH !
This display only appears when the background threshold you have defined in the protocol is exceeded. If this happens the printout of measurem ent values is flagged BKG.The s teps which the instruments goes through are dis played and the measurement
Page 92
Chapter 6: CUT-OFF-Measurements LB 9507
6-16
parameters from the active protocol are printed out at the same time.
C51
INJECTING ...
C52
MEASURING ...
C53
SAMPLE-RLU 133 REMOVE BACKGR. (RLU/S) 30.0 TUBE
Once measurem ent is complete, the first line shows the r esult of the measurement in RLUs, the second line (if defined in the protocol) shows the result of the background measurement. The results are printed in 4 columns:
- Measured RLU; for replicates this represents the mean of the replicate group involved
- Coefficient of variation of the replicate group
-Flag
- Result: indication of the range
The results are classified ac cording to negative controls, positive controls and samples.
After removing the sample tube, the measured value display is cleared. You are prompted to:
C54
NEG. CTRL. REPL. # 2 TUBE # 2 INSERT TUBE ...
Load the next tube containing the sample def ined on the firs t line (in this case the 2nd negative control). The start request will then appear on the display:
Page 93
LB 9507 Chapter 6: CUT-OFF-Measurements
6-17
C55
NEG. CTRL. REPL. # 2 TUBE # 2
START
Start the measurement by pressing the <start> button. In accordance with the selected protocol, you are asked for the defined number of replicates for negative and then positive controls. Once a replicate group has been meas ured, the mean value and the coefficient of variation for the group are given.
C56
ENTER REPL. NO. TO DISCARD NONE
If you want to delete a measured replicate, enter the num ber of the respective control that is indicated on the result printout. Otherwise press the button below <NONE>.
The printout shows when a replicate has been deleted. The deleted replicate is not removed from the memory. When you load the measured data again (<OPERATOR FUNCTIONS: DATA EVALUATION: STORED DATA>, see chapter 11.4), the original measurement data will be used.
C57
CHANGE NEG. CTRL ? YES NO
<YES>: Here you can edit the mean value. A correc tion is noted on the printout. The change is not stored to the mem ory, so that the original mean value will be displayed when calling the stored measurement data (<OPERATOR FUNCTIONS: DATA EVALUATION: STORED DATA>, see chapter 11.4).
<NO>: Continue the measurement.
Following mean value calculation and a possible correction, the
evaluation of the controls is output to the printer.
Then, the patient samples are requested and m ean value (RLU), variation coefficient, flag and range are printed out for each sample and each replicate group.
Page 94
Chapter 6: CUT-OFF-Measurements LB 9507
6-18
Page 95
LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-1
7. Quantitative Measurements (LIA...)
7.1 Introduction
This type of measurement is used for reagent kit specific measurements to quantify analyte concentrations (LIA, ILMA, etc.). Three measurement strategies are available:
1. Measurement with complete standard series
Known standard concentrations are measured before the unknown samples. After transformation of the calibration curve (either log-log or logit-log) interpolations between standards and unknown samples take place. The curve f it is based on a cubic spline function and is excellently suited to LIAs and ILMAs. If necessary, the curve can be smoothed either automatically or upon the user’s instruction.
2. Measurement with master curve
This measurement type uses 2 calibrators instead of standard concentrations, which are provided as sam ples by the assay kit manufacturer. Once this protocol has been c reated, the rem aining value pairs (RLU/conc) of a standard curve are entered in the submenu <MASTER CURVE> and are used as a logit-log transformed master curve to analyse the samples. The value pairs are included with the respective batch of reagent kits.
The displays are numbered consecutively L1, L2 ...LX for clarity.
Fig. 7-1 is a flow chart of the protocol setup procedure, entry of the master curve and start of the measurement.
Page 96
Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-2
Create Protocol
Master curve entry
L1
READY MEASURE PROTOCOL -OTHERS-
L2
PROTOCOL MASTER CURVE CREATE CHANGE
L2a L3
L4a
L3a R2b
L41 L4b
L5
L48 L30
L49 L38
L1
READY MEASURE PROTOCOL -OTHERS-
L50
SELECT TYPE OF MEASUREMENT ? PROTOCOLS DIRECT ENTRY RATEMETER
L51
ENTER PROTOCOL NO. 7 PRINT LIST
Fig. 7-1: Flow chart showing protocol setup and measurement
Enter Master Curve
PROTOCOL NUMBER PRINT LIST
PROTOCOL # O8 "LIA" RAW DATA CUTOFF -OTHERS-
08: PROTOCOL NAME
08: PROTOCOL TYPE RAW DATA CUTOFF -OTHERS-
08: PROTOCOL TYPE ? QUANT TUP/FTI -OTHERS-
Measurements
Start of measurement
08: CHARGE
08: HIGH CAL. CONC.
LAST DISPLAY
08: INPUT ALL CORRECT ? YES YES NO
08: OPERATOR ID.
08: FULL STANDARDIZATION YES YES NO LAST DISPLAY
08: INPUT ALL CORRECT ? YES YES NO
Page 97
LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-3
7.2 Setting Up Quantitative Protocols
Start by entering the parameters for the measur ement procedure and the evaluation strategy in a measurement protocol in the <PROTOCOL> menu, and then save them. Later, you can run a measurement simply by calling up one of these stored protocols in the <MEASURE> menu and pressing the <START> button.
You can save up to 40 different measurement protocols which include all information such as injector control, definition of replicates, measuring time etc., as well as information on data evaluation. Each new protocol is given a number which is dependent upon the time the protocol was set up but not on the protocol type.
The differences between the three measurement strategies are evident in the protocol structure.
For measurem ents with a com plete standar d s eries, you will need the entire protocol that is described on the following pages.
For measurements with a master curve the protocol finishes at display L30 (see Fig. 7-1). The master curve is then entered in the submenu <MASTER CURVE> and assigned to the respective protocol.
Page 98
Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-4
L1
READY MEASURE PROTOCOL -OTHERS-
Select the <PROTOCOL> menu by pressing the respective button. Then follows:
L2
PROTOCOL MASTER CURVE CREATE CHANGE
The <MASTER CURVE> option can be used only when a “quantitative protocol” without a standard series has been created and stored. If you press the button below <CHANGE> you can open an existing protocol by entering its protocol number, and edit it. To create a new protocol, press the button below <CREATE>. The following display is shown:
L3
08: PROTOCOL NAME LIA-T1
The next free number is displayed for the new protocol (in this example “08:”). Enter a name for this protocol using the alphanumeric keyboard. You should use a name which specifies the protocol type and other important features, so that it is easy to identify later. For example “LIA-T1”. Confirm your entry with <enter>.
L4a
08: PROTOCOL TYPE RAW DATA CUTOFF -OTHERS-
Select <OTHERS>.
L4b
08: PROTOCOL TYPE QUANT. TUP/FTI -OTHERS-
Select <QUANT>. Now you have to enter the parameters which are the same for all protocol types:
Page 99
LB 9507 Chapter 7: Quantitative Measurements (LIA...)
7-5
L5
08: OPERATOR ID.
Enter the user name and press <enter>. If no injectors are activated (see chapter 3.8.2) the Display L15 follows.
L6
08: ENTER INJECTOR VOLUME 1 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 1 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. Range: 25 µL - 300 µL in 5 µL steps. If the user enters the injector volume not within the 5 µL step range, the system rounds up or down. Values out of the range are adjusted to the limits (25 µL or 300 µL). Example: W hen entering 43 µL the system rounds up to 45 µL, when entering 102 µL the system rounds down to 100 µL. Press <enter> to use the response displayed in the upper-right corner of the LCD.
L7
08: ENTER INJECTOR VOLUME 2 [µl]
LAST DISPLAY
Changing or confirmation of the displayed injector volume. Displayed are the standard settings for injector 2 entered in the <INSTRUMENT PARAMETERS> menu. These settings can be changed for this specific protocol. Setting "0" means the respective injector is disabled. <LAST DISPLAY> takes you back to the previous display. Setting "0" takes you to display L10. Setting >0 takes you to display L8. Press <enter> to use the response displayed in the upper-right corner of the LCD.
Page 100
Chapter 7: Quantitative Measurements (LIA...) LB 9507
7-6
L8
08: SEQUENCE OF INJECTIONS 1->2 1->2 2->1 LAST DISPLAY
If you have defined 2 injectors, you have to define the s equence of the injections here: injector 1 first and then injector 2 or vice versa. Make your selection by pressing the button below the option you want. Press <enter> to use the response displayed in the upper-right corner of the LCD. The following display appears:
L9
08: DELAY TIME INJ 1/INJ 2 [s] 1.2 (0.7 - 300.0) LAST DISPLAY
Enter the desired delay in seconds between both injections and confirm with <enter>.
L10
08: MEASURE BACKGROUND (BKG) NO YES NO LAST DISPLAY
Select here whether the background rate is to be measured prior to the injections. If you select <NO>, display L14 appears or - if no injector was selected - L15. <YES> takes you to the next display:
L11
08: MEASURING TIME BKG [s] 0.5 (0.1 - 200.0) LAST DISPLAY
Enter between 0.1 and 200.0 seconds for the background measurement and pres s <enter>. Reasonable times start at half a second, allowing statistical variations to be compensated for. Otherwise, the background values may be too high or too low which either distorts the result of the m easurem ent or leads to an exceeding of the threshold.
Remember that the background effect is measured in order to obtain an alarm threshold for background rates that are too high.
Loading...