Ludlum Measurements 4500 Operator's Manual

LUDLUM MEASUREME NTS, INC.
501 OAK ST., PO BOX 810 SWEETWATER, TX 79556 915/235-5494 FAX: 915/235-4672
LUDLUM MODEL 4500
RADIATION PORTAL MONITOR
Version 2.00
April 2003
Model 4500 Operator’s Manual
April 2003
page
i
TABLE OF CONTENTS
1. GENERAL
................................
................................
................................
..............................
2
2. BLOCK DIAGRAM
................................
................................
................................
...............
2
3. GETTING STARTED
................................
................................
................................
............
2
3.1 Power
................................
................................
................................
................................
.....
2
3.2 On/Off Switch
................................
................................
................................
........................
3
3.3 Fuses
................................
................................
................................
................................
......
3
4. MAINTENANCE
................................
................................
................................
....................
3
5. TROUBLESHOOTING
................................
................................
................................
.........
3
5.1 Infrared (IR) Sensors
................................
................................
................................
..............
3
5.2 Instrument Failure
................................
................................
................................
..................
5
5.3 Testing Battery, Line Power and Fuse
................................
................................
...................
6
6. SELF-DIAGNOSTICS
................................
................................
................................
...........
6
7. REMOTE STATUS DISPLAY
................................
................................
.............................
7
8. SOFTWARE OVERVIEW
................................
................................
................................
....
8
9. CONFIGURING LANES
................................
................................
................................
.....
10
9.1 Assigning Model 4500s to a Lane
................................
................................
........................
10
10. SETTING PARAMETERS
................................
................................
..............................
10
11. CHANGING THE PASSWORD
................................
................................
.....................
12
12. ALARM WINDOW
................................
................................
................................
..........
12
13. DATA ARCHIVING
................................
................................
................................
.........
13
DRAWINGS OR DIAGRAMS - Block Diagram
................................
.............
following page 2
Model 4500 Operator’s Manual
April 2003
page
2
1. GENERAL
The Model 4500 Radiation Portal Monitor has very sensitive gamma and neutron sensors to detect small amounts of radiation. These passive detectors are arranged to detect very low levels of radiation coming from personnel or vehicles that pass through the detectors.
It is important to note that the levels of radiation detected by this instrument are quite small. A radiation alarm does not
automatically indicate dangerous levels of radiation. Everyday products and materials, because of small amounts of naturally occurring radio active materials, will activate many radiation alarms. Medical patients having undergone tests involving radioactive materials may also activate the alarms for weeks after the test or treatment.
2. BLOCK DIAGRAM
The block diagram on the following page illustrates the major components of the radiation portal system.
Detector: passive gamma and neutron sensitive detectors.
Infrared (IR) Sensor: infrared sensor detects vehicle presence.
Remote Status Display: provides visual display of status, audible warning of alarm and reset capability.
Control Box: Main Electronics functions as the “brain”, receiving input from Detectors
and Object Sensors, powering the Remote Status Display, and sending information through the Ethernet to the computer.
Optional Video: network camera can be used to supply a picture of each vehicle passing through the portal system.
Computer: running NetSupervisor software, is used display multiple lanes of video and radiation data, to log data on every ve hicl e, to pop up alarm windows with vehicle profiles, and to remotely cha nge parameters within each control box.
3. GETTING STARTED
When power is applied for the first time, the Model 4500 has a warm-up time of approximately 1 minute. During this time, the control box initializes, and a 30-second background radiation measurement is taken.
3.1 Power
Power input to the control box is normally a hard-wired 120VAC mains connection. Backup power is supplied by way of an internal trickle-charged gell-cell battery ins ide the control box. The backup battery will supply power for several hours if the mains power is out. Because of the
Model 4500 Operator’s Manual
April 2003
page
3
internal backup battery, switching off the mains power (by switch or breaker) will NOT turn the system off. If remote ON/OFF operation is required, the battery fuse should be removed.
7
NOTE: Disconnect or remove power to instrument before checking or replacin g fuses. If necessary, have an electrician (or otherwise qualified person) perform this task.
3.2 On/Off Switch
The ON/OFF switch is a black rocker switch located in the center of the control box. Switching it to the down or OFF position disconnects both the mains power and the battery power.
3.3 Fuses
The fuses for both the mains power and the battery are located i n the control box next to the ON/OFF switch. The fuses are a 1-amp AGC-1 (main AC fuse) and a 5-amp ABC-5 (battery fuse).
4. MAINTENANCE
The Ludlum Model 4500 portal system is built to last for many years. The only regular maintenance that is required is to ensure that the vehicle sensors are clean. If the front lenses of the sensors are dirty or blocked by ice or snow, the system might think a vehicle is inside the portal. As a
result, the CHECKING indicator of the Remote Status Display might remain on all the time. The instrument does monitor the infrared sensors and will activate a FAIL alarm if the sensors remained blocked for several minutes.
5. TROUBLESHOOTING
5.1 Infrared (IR) Sensors
SENSOR EMITTER
The emitters are the sensors that are located on the opposite side of the lane from the electronics box. These have a small pinhole for the IR beam to be emitted through.
There is a red power LED that is located on the side of the emitter that can be seen
when the unit is powered up. This red LED can be seen by looking th rough the hole on the side or by looking underneath the shroud. It may be difficult to see the red LED under bright light conditions. Shading the unit may help in seeing the LED.
NOTE: The LED will always be
illuminated, as long as the emitter has power.
Model 4500 Operator’s Manual
April 2003
page
4
Emitter Front Emitter Side
SENSOR RECEIVER
The receivers are the sensors that are located on the same side of the lane as the electronics box. They receive and acknowledge the beam from the emitter.
The receiver also has an LED on the side. It will only be illuminated when the
unit is powered up
and
the beam is present. This red LED can be seen by looking through the hole on the side or through the front of the shroud. (Take care to not block the beam during inspection.)
Receiver Front Receiver Side
SENSOR SETUP AND ALIGNMENT
1. Mount the sensors on the side of the
units.
2. The sensor set that vehicles reach first (front) should be mounted facing each other on opposite sides of the lanes. Both emitter and receiver should be 19 inches above the road surface.
3. The second set of sensors (back) will be mounted at an angle. The emitter should be 19 inches above the road, and the receiver should be 72 inches above the road surface.
4. Aim the receiver toward the emitter. Do this by sighting along the top of the receiver and adjusting it up and down until it is pointing directly at the pinhole on the emitter. A laser pointer or string may be useful as an alignment aid.
5. Aim the emitter at the receiver. Adjust the emitter until the LED light on the receiver is illuminated.
6. Move the emitter up until the light turns off.
7. Note the angle.
Model 4500 Operator’s Manual
April 2003
page
5
8. Move the emitter down until the light
turns off.
9. Note the angle.
10. Move the emitter to the middle of the
two angle readings.
11. Tighten the emitter and receiver
hold-down bolts.
12. Re-inspect for receiver LED.
5.2 Instrument Failure
The FAIL status, which activates warnings at both the remote status display and at the computer, can be caused by th ree conditions:
1) Sensor Blockage-when the sensors
are blocked by di rt , ice, or snow for several minutes or when a vehicle remains in the portal too long.
2) Hi Background - A radiation sensor
indicates radiation levels above the normal limits.
3) Lo Background - A radiation sensor
indicates radiation levels below the normal limits.
The most common failure mode is the sensor blockage, so checking the sensors for blockage should be the first response. (Refer to more details later in this section.) Once the blockage is corrected, the reset button should be pressed and held for three seconds to reboot the system and establish a new background.
The Hi or Lo Background conditions usually indicate detector failure, although the Hi Background condition could be associated with some external radiation source, such as an X-ray tru ck. If this is the case, remove the radiation source, and reboot the system. If it is determined that no external radiation source is causing the Hi Background, then the detector should be disabled (see section 10. “Setting Parameters”), and technical support should
be called. Disabling the detector allows the system to run with the other working detectors.
SENSOR BLOCKAGE
1. Check emitter for proper operation. (Refer to section 5.1 to identify the sensor emitter.) Its LED should be on and stay on. If not, it has a problem and technical support should be called.
2. Check that there is not a vehicle or other object in the lane. If there is, remove it.
3. Check the pinhole on the emitter. Clean and/or remove any blockage.
4. Check lenses on emitter and receiver. If lenses are dirty, clean them with a clean damp cloth. If they are iced over, use a commercial de-icer and then clean them.
5. Check alignment of sensors. (Refer to SENSOR SETUP AND ALIGNMENT, Section 5.1).
6. Check receiver for proper operation. (Refer to text in section 5.1 to identify the sensor receiver.). Its LED should be on and stay on until an object passes through the lane. If steps 1 through 5 have been completed and there is still a problem, technical support should be called.
BACKGROUND TOO HIGH
1. Check that VACIS source is not too close to portal.
2. Use a portable radiation instrument to determine if a foreign source of radiation is present. (A person who has recently undergone a medical procedure using radiopharmaceuticals can be an unknowing radiation source.)
3. Reinitialize instrument. (When portal is clear, hold reset button 3 seconds.)
4. If failure persists, contact Supervisor.
BACKGROUND TOO LOW
1. Reinitialize instrument. (When portal is clear, hold reset button 3 seconds.)
2. If failure persists, contact Supervisor.
Model 4500 Operator’s Manual
April 2003
page
6
5.3 Testing Battery, Line Power and Fuse
Battery Test:
1. Turn the power off to the unit.
2. Remove the Line (1A) fuse.
3. Power the unit ON ( 0= off, 1= on).
NOTE: When the fuse is removed, the unit is running on BATTERY only. If the unit fails to operate, the battery power is not functioning properly.
4. Replace fuse when testing is
completed.
Line Test:
1. Turn the power off to the unit.
2. Remove the Battery (5A) fuse.
3. Power the unit ON ( 0= off, 1= on).
NOTE: When the fuse is removed, the unit is running on LINE only. If the unit fails to operate, the line power is not functioning properly.
4. Measure LINE voltage. It must measure between 90-240VAC (50-60 Hz).
5. Replace fuse when testing is completed.
Fuse Test:
1. Remove the fuse from the unit.
2. Using a volt meter on Ohms (6),
measure across both ends of the fuse, and determine if the fuse is good, as follows:
reading: 0-5 = good reading: open or OL = bad
6. SELF-DIAGNOSTICS
The Ludlum Model 4500 portal system has many built-in self-diagnostics. The control box constantly checks the readings from the radiation detectors to ensure that they are within normal limits. The control box also ensures that the vehicle sensors have not been blocked for more than several minutes. If the control box determines that any detector is outside of normal limits, or that the vehicle sensors might be dirty or
blocked by ice or snow, then a FAIL condition will be activated.
When a FAIL condition is activated, audio and visual warning will be given at the Remote Status Display and also at the NetSupervisor computer display. Section
5.2 “Instrument Failure”, earlier in this manual, covers what to do when a FAIL condition occurs.
Model 4500 Operator’s Manual
April 2003
page
7
7. REMOTE STATUS DISPLAY
The Model 4500 Remote Status Display is the primary indicator of status to the operator. The following is a description of each indicator:
Power OK - a green LED indicates that the system is ON.
Checking - a green LED momentarily turns ON as each vehicle passes through the detectors.
Gamma Alarm - a red LED indicates that radiation detectors have detected an increase in the gamma radiation levels as the vehicle has passed through the detectors.
Neutron Alarm - a red LED indicates that radiation detectors have detected an increase in the neutron radiation levels as the vehicle has passed through the detectors.
Instrument Fail - an orange LED indicates a failure of the instrument.
Reset-a pushbutton with two functions:
1) If an alarm is present, it can be used
to reset the system for the next vehicle,
or 2) If pressed for 3 seconds, a reboot or restart of the system occurs. The reboot function may be necessary to correct the background radiation readings if the readings have been artificially altered because of the nearby operation of an x-ray system. NOTE: Reboot cannot take place if a vehicle is within the portal.
Volume Control – a recessed screw adjustment located on the bottom of the remote display whic h controls the volume of the voice annunciator.
Voice Annunciator – an audible signal that indicates an ALARM condition with the spoken word “alert” or a FAIL condition with the spoken words “system alert”.
Model 4500 Operator’s Manual
April 2003
page
8
8. SOFTWARE OVERVIEW
The Display Window is a tabbed window that can display 12 lanes. An indicator next to the Car Counter turns red when the infrared beams are broken. Each lane has a counter for the number of Gamma Low, Gamma High, and Neutron Alarms.
If camera for the lane is enabled, live video will be displayed in the image window. If both front and back cameras are enabled, clicking on the image will toggle between the front and back cameras.
A status message is displayed next to the lane number. “Connected” indicates that the software is connected to the lane.
“Timeout” indicates that no data was received from the lane for 20 seconds. “Retrying” indicates that the software is attempting to reconnect to the lane. “Missing” indicates that the retry failed. The software will keep retrying every 20 seconds to connect to the lane.
If there is a background failure, the Average fields will turn red, indicating there is a problem with the detectors. A fail window will pop up indicating which channel and what type of background alarm occurred. Low background alarms are displayed in purple, and high background alarms are displayed in red.
Model 4500 Operator’s Manual
April 2003
page
9
The Prefs menu has five password­protected options. They are Search Network, Assign Instruments, Set Parameters, Options, and Set Password.
The Monitor software operates just like the Supervisor software without the ability to change any par amet ers. The only option available to the Monitor Software is that of setting an Image Store Drive Letter. This is the drive letter mapped to the network location of the camera images.
Model 4500 Operator’s Manual
April 2003
page
10
9. CONFIGURING LANES
Use the Prefs/Search Network menu to search for all Model 4500 monitors on the network. After searching the network, use the “Get ID Numbers” button to return the Unique ID numbers for each Model 4500. Click on the “Save” button to save the list of Model 4500s found.
9.1 Assigning Model 4500s to a Lane
To assign an unassigned Model 4500 to a lane, select the lane and click on the “Unassigned” button. Then select the Model 4500 and click on the “OK” button.
Disable: Disable the currently selected lane Enable All: Enable all lanes.
Disable All: Disable all lanes Unassigned: Displays a list of all
unassigned Model 4500s. The Search Network option must be used first before any Model 4500s display here. Move Up: Move the currently selected Model 4500 to the next lower lane number Move Down: Move the currently selected Model 4500 to the next higher lane number. Save: The current settings will be saved and the Display window will be updated with the new settings. Cancel: The current settings will not be saved.
10. SETTING PARAMETERS
The Set Parameters window allows the parameters to be set for all lanes or just a single lane. Refer to the Model 4500 Electronics Instruction Manual for a more detailed explanation of the parameters.
Select Lane: Determines which lane for which the parameters are changed. If set to All, the parameters are changed for all lanes.
Unit ID: An ID number that can be
assigned to each Model 4500. Valid values are from 0 to 255.
4500 Date/Time: The current date and
time when the parameters were downloaded. The computer’s current date and time is used when this parameter is set.
Alarm Setpoints: Sets the v alu e used i n
the alarm calculation in terms of Sigma and also an absolute value above background.
Model 4500 Operator’s Manual
April 2003
page
11
Background Alarms: Sets the high and low background values. If the background goes outside these limits, the Model 4500 will go into a fail condition.
Sum Alarms Active: If checked, Gamma Low Sum, Gamma High Sum, and Neutron Sums Alarms are active.
Individual Alarms: If checked, the individual channels are checked for an alarm.
Neutron Channels: If checked detectors are grouped as a Gamma Low, Gamma High, and Neutron channel. If unchecked all channels are Gamma channels.
Use Sigma for Alarm Calc: If checked, the alarms are based on sigma; otherwise, they are based on the absolute value.
Samples: Sets the number of samples for each vehicle and the number of pre- and post-samples that are checked.
Background Update: Determines how fast the background is updated.
Time Values: Sets the Count Time,
Alarm Hold Time, and the Incomplete Time. If the Alarm Hold Time is zero, the Model 4500 will alarm until the reset button is pressed. The Sensor Fail sets the amount of time in minutes that a continuous blocking of the sensors will cause a failure.
Enable/Disable Detectors: Detectors
that are checked are active, detectors that are not checked are not active.
Reset: Reset the Model 4500 Refresh: Re-download the parameters
from the Model 4500
Set: Saves the individual parameters to
the Model 4500.
Set All: Saves all the parameters to the
Model 4500.
Model 4500 Operator’s Manual
April 2003
page
12
11. CHANGING THE PASSWORD
Note: The default password is “4500”. After running the program for the first time, be sure to change the password to something else.
To change the password, you must enter the old password, type in a new password, and then confirm the new password. The password must be at least four (4) characters in length.
12. ALARM WINDOW
The Alarm window is automatically displayed, and an alarm report is printed when an alarm occurs. The alarm window displays the front and back camera images (if enabled), the background and maximum reading for each detector, and a profile graph of the vehicle. The values in the grid are displayed in red if an alarm occurs on that channel. The “Print” button will print another copy of the alarm report. The “Acknowledge” button will close this window.
The profile graph displays the sum of the Gamma Low, Gamma High, and Neutron channels in tenths of second intervals. The scale on the left is for the gamma channels
and the scale on the right is for the neutron channels. The profile can be used to locate the point along the vehicle that has the highest radiation reading. For example, if the peak is located at sample 15, and there are 30 samples, then the peak is in the middle of the vehicle.
The values in the grid are based on the number of samples set in the Parameters screen. The grid is based on the Parameters Screen. If the number of samples were set to 10, these values would be in counts per second. If the number of samples were 5, then the values would be in counts per half-second.
Model 4500 Operator’s Manual
April 2003
page
13
13. DATA ARCHIVING
There are three databases used in the Supervisor program. The alarms.mdb stores the data for the vehicles that alarmed. The noalarms.mdb stores the data for vehicles that did not alarm. The background.mdb stores the background readings every minute.
The database and any image files should be backed up regularly to prevent data loss. The Supervisor program contains an option under the “Prefs” menu called “Archive Databases”. This option moves the current databases to the archive folder and renames them with the current date and time and creates new blank databases to hold new data. The files are renamed using the following format:
YYYYMMDD_HHMMSS_name.mdb where “name” is either alarms, noalarms, or background.
The image files are not stored in the database. Only the path and filename are stored in the database. The images are stored as standard JPEG files. These files should be backed up and deleted from the computer to free storage space as needed. The images are stored in the following directory structure: Images_b\lane_xx\YYYYMMDD\ Images_f\lane_xx\YYYYMMDD\
The data can then be copied from the archive folder to a Direct-CD formatted cd for backup.
Loading...