Ludlum Measurements 2241-3, 2241-3i User Manual

LUDLUM MODEL 2241-3 & 2241-3i
SURVEY METER
January 2016
Serial Number 299171 and Succeeding
Serial Numbers
LUDLUM MODEL 2241-3 & 2241-3i
SURVEY METER
January 2016
Serial Number 299171 and Succeeding
Serial Numbers
STATEMENT OF WARRANTY
Ludlum Measurements, Inc. warrants the products covered in this manual to be free of defects due to workmanship, material, and design for a period of twelve months from the date of delivery. The calibration of a product is warranted to be within its specified accuracy limits at the time of shipment. In the event of instrument failure, notify Ludlum Measurements to determine if repair, recalibration, or replacement is required.
This warranty excludes the replacement of photomultiplier tubes, G-M and proportional tubes, and scintillation crystals which are broken due to excessive physical abuse or used for purposes other than intended.
There are no warranties, express or implied, including without limitation any implied warranty of merchantability or fitness, which extend beyond the description of the face there of. If the product does not perform as warranted herein, purchaser’s sole remedy shall be repair or replacement, at the option of Ludlum Measurements. In no event will Ludlum Measurements be liable for damages, lost revenue, lost wages, or any other incidental or consequential damages, arising from the purchase, use, or inability to product
.
use
RETURN OF GOODS TO MANUFACTURER
If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information. Your cooperation will expedite the return of your equipment.
ATTN: REPAIR DEPARTMENT 501 OAK STREET SWEETWATER, TX 79556
LUDLUM MEASUREMENTS, INC.
800-622-0828 325-235-5494
FAX 325-235-4672
Table of Contents
Introduction 1
Information on Model 2241-3i 1-2
Getting Started 2
Unpacking and Repacking 2-1 Battery Installation 2-1 Connecting a Detector 2-2 Operational Check 2-2
Specifications 3
Removable Switchboard Adjustable Parameters 3-2 Internal Detector Specifications for Model 2241-3i 3-4
Identification of Controls and Functions 4
Display 4-1 Display Status Definitions 4-1 Front Panel Controls 4-2 Front Panel Calibration Controls 4-3 Main Board Controls 4-4 Switch Board Controls 4-5
Safety Considerations 5
Environmental Conditions for Normal Use 5-1 Detector Connector 5-1 Warning Markings and Symbols 5-1
Maintenance 6
Operational Check 6-1 Recalibration 6-1 Batteries 6-3
Technical Principle of Operation 7
Main Board 7-1 Switch Board 7-2 Display Board 7-3
Ludlum Measurements, Inc. January 2016
Model 2241-3 & 3i Technical Manual
Instrument Setup & Calibration 8
Entering or Changing Switch Board Parameters 8-1 Loading Default Parameters 8-1 The Function Switch 8-2 Function Switch Position Descriptions and Variables 8-2 Calibration 8-7 General Detector Setup Information 8-8 Counts per minute (C/m) Calibration 8-9 R/hr Calibration 8-11 Determining CC and DTC 8-12 Software License Agreement 8-19
Recycling 9
Parts List 10
Model 2241-3 Survey Meter 10-1 Main Circuit Board, Drawing 408 × 285 10-1 Calibration Board, Drawing 408 × 98 10-3 Display Board, Drawing 408 × 259 10-4 Switch Board, Drawing 408 × 45 10-5 Chassis Wiring Diagram, Drawing 408 × 101 10-5
Drawings and Diagrams 11
RS-232 Output Formats - Appendix A A
Binary Output Format A-1 ASCII Output Format A-4
Ludlum Measurements, Inc. January 2016
Model 2241-3 & 3i Technical Manual Section 1
Section
1
Introduction
he Model 2241-3 is a portable microprocessor-based digital Scaler/ Ratemeter designed for use with scintillation, Geiger-Mueller (GM),
T
mode) Liquid Crystal Display (LCD) with moving decimal point. A toggle switch labeled SCA/RATE selects the desired operating mode for the
instrument. A five-position switch labeled ˝OFF/DET 1/DET 2/DET 3/DET 4˝ allows for a quick change of operating parameters when changing out
detectors.
Programmable display units (RATEMETER mode only) can be represented in R/hr, Sv/h, cpm, or cps with multipliers of micro (µ) or milli (m) for R/hr and Sv/h and kilo (k) for cpm or cps. The display units are auto-ranging, enabling the readout to display a broad range of radiation levels. The display also offers lower limit capability. For example, the display can be set to show only values that are greater than or equal to1 µR/hr.
and proportional-type detectors for measurement of ionizing radiation. Data is presented on a four-digit (six digits in the scaler
This instrument incorporates independent adjustable alarms for RATEMETER and SCALER operating modes. The RATEMETER mode has two alarm
indicators. The first-level alarm is indicated by display of the word ˝
on the LCD. The second-level alarm is indicated by display of the word
˝ALARM˝ and by the emitting of a continuous audible tone. The SCALER alarm condition will also display the word ˝
audible tone. Both audible alarms may be silenced (acknowledged) by depressing the RESET switch. All alarms are concurrent.
Other features include: 1) Dead Time Correction (DTC) to compensate for detector dead time; 2) audible click-per-event with programmable 1, 10, 100, and 1000 divide-by; 3) LCD backlight with programmable ON time; 4) programmable fixed or variable response time; and 5) count overflow visual alarm, indicative of the counting circuitry nearing the maximum counting capability.
Ludlum Measurements, Inc. Page 1-1 January 2016
ALARM˝ and produce the same
ALERT˝
Model 2241-3 & 3i Technical Manual Section 1
All of the features described above may be programmed manually using the internal switch board or by computer through the RS-232 port. Four different detector operating parameters may be stored in non-volatile memory. The switch board can be removed after entering or changing parameters to prevent tampering with setup parameters.
A regulated high-voltage power supply, four independent set-point controls adjustable from 400 to 2500 volts, and adjustable discrimination levels, add versatility to the instrument. This supports operation for a broad range of detectors and connecting cable lengths. All of the calibration controls are covered to prevent any inadvertent adjustment to the detector operating parameters.
The instrument is powered by two standard ˝D˝ cell batteries. The unit body
is made of cast-and-drawn aluminum with beige powder coating, which aids in the decontamination of surfaces.
The Model 2241-3i is identical to the Model 2241-3 except for the following:
The five-position selector switch located on the front panel has been modified to provide for selection of an internal detector. When selected, the INT position (formerly DET4) switches detector operation to an internal, energy-compensated Geiger-Mueller detector. Appropriate wiring changes to accommodate the internal detector have been made; however, physical dimensions are the same as for a Model 2241-3.
A drawing of the front-panel of the 2241-3i is included at the beginning of this manual. A wiring diagram for the instrument is included in the back of the manual with the other drawings. The internal detector specifications can be found at the end of Section 3 of this manual.
Ludlum Measurements, Inc. Page 1-2 January 2016
Model 2241-3 & 3i Technical Manual Section 2
Section
2
Getting Started
Unpacking and R epacking
Remove the calibration certificates and place them in a secure location. Remove the instrument, detectors, and accessories (batteries, cable, etc.), and ensure that all of the items listed on the packing list are in the carton. Check individual item serial numbers and ensure calibration certificates match. The Model 2241-3 serial number is located on the front panel below the battery compartment. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification.
Important!
If multiple shipments are received, ensure that the detectors and instruments are not interchanged. Each instrument is calibrated to specific detectors, and are therefore, not interchangeable.
To return an instrument for repair or calibration, provide sufficient packing material to prevent damage during shipment. Also provide appropriate warning labels to ensure careful handling. Include detector(s) and related cable(s) for calibration.
Every returned instrument must be accompanied by an Instrument Return Form, which can be downloaded from the Ludlum website at
www.ludlums.com. Find the form by clicking the “Support” tab and
selecting “Repair and Calibration” from the drop-down menu. Then choose the appropriate Repair and Calibration division where you will find a link to the form.
Battery Installation
Ensure the OFF/DET 1/DET 2/DET 3/DET 4 switch is in the OFF position. Open the battery lid by turning the quarter-turn thumb screw counter-
Ludlum Measurements, Inc. Page 2-1 January 2016
Model 2241-3 & 3i Technical Manual Section 2
clockwise. Install two ˝D˝ size batteries in the compartment. Note the (+)
and (–) marks inside the battery door. Match the battery polarity to these marks. Close the battery box lid.
Note:
The center post of “D” size battery is positive.
Connecting a Detector
Caution!
The detector operating voltage (HV) is supplied to the detector by way of the input connector. A mild electric shock may occur if contact is made with the center pin of the input connector. Switch the Model 2241-3 to the OFF position before connecting or disconnecting the cable or detector.
Connect a detector to the Model 2241-3 by using the cable provided; firmly pushing the connectors together while twisting clockwise until the connector latches (1/4 turn). The diagram to the left illustrates how this is done. Next, place the detector selector switch in the appropriate position.
Operational Check
Place the SCA/RATE (scaler/ratemeter) switch in the RATE position. Notice that the display goes through an initialization sequence. The display will show all “8”s with decimal points. Check to make sure all segments display, as illustrated in the diagram to the left.
The LCD then displays the firmware number in the format “P-XX YY.” The “XX” is the firmware number, and the “YY” is the firmware version. (The figure to the left is for example only; to illustrate location of display.)
The minimum displayable value (for example 00.0 µR/hr) should be shown. When switched to the SCA position, a single “0” will be displayed.
Ludlum Measurements, Inc. Page 2-2 January 2016
Model 2241-3 & 3i Technical Manual Section 2
The display will auto-range to the current level (see figure at left). When auto-ranging down, the Model 2241-3 uses multiples of 5. This technique keeps the decimal point from jumping between numbers when viewing values around multiples of 10.
Check for a proper background reading:
If using a Ludlum Model 44-9 detector, a typical reading would be 25-50
cpm or 8-15 µR/hr.
If using a Ludlum Model 44-2 detector, a typical reading would be 1.4-2.6
kcpm or 8-15 µR/hr.
If using a Ludlum Model 133-7 detector, a typical reading would be 0 cpm
or 8-15 µR/hr.
A reference reading (or readings) with a check source should be obtained with the detector(s) in a constant and reproducible manner at the time of calibration or at the time the instrument is received in the field.
If at any time the instrument fails to read within 20% of the reference reading when using the same check source, it should be sent to a calibration facility for recalibration and/or repair. If desired, multiple readings may be taken at different distances and/or with different sources so that other ranges or scales are checked.
Switch the
AUD ON/OFF switch to the ON position and confirm that the
external unimorph speaker produces an audible click for each event detected (audio divide-by 1 parameter). The AUD ON/OFF switch will silence the clicks if in the OFF position, however, an audible alarm condition will still be heard.
Increase the source activity or lower the alert and alarm points to initiate an
ALERT and ALARM condition. (Refer to Section 8, Subsection “Entering or
Changing Switchboard Parameters.”) Depress the RESET switch to acknowledge the audible alarm. Decrease the radiation activity below the
ALERT and ALARM threshold and depress the RESET switch to clear the alarm
conditions. If an alarm condition is not present, depressing the RESET switch the first time will reset the alert condition and zero the ratemeter.
Position a check source to produce a ratemeter reading of 100 to 2000 counts per minute or 10-100 µR/hr. While observing the ratemeter fluctuations, select between the fast and slow response time (
F/S) positions
to observe variations in the display. The S position should respond
Ludlum Measurements, Inc. Page 2-3 January 2016
Model 2241-3 & 3i Technical Manual Section 2
approximately five times slower than the F position (for fixed response mode) and three times slower when in variable response mode. The slow response position is normally used when the Model 2241-3 is displaying low numbers, which require a more stable display. The fast response position is used at high count levels.
Place the SCA/RATE switch to the SCA position. Depress the COUNT switch located in the end of the carrying handle in order to initiate a count cycle. The word “COUNTING” should be flashing on the LCD during the count cycle and should disappear at the end of the predetermined count time. If a scaler ALARM condition occurs, the RESET switch can be depressed to acknowledge the alarm; however, the COUNT switch must be depressed to clear the visual ALARM and to restart the count cycle.
Depress and release the
LIGHT switch. The backlight located behind the
LCD should illuminate (for pre-programmed ON time). Select the desired
F/S, AUD ON/OFF, and RATE/SCA parameters and proceed to use the
instrument.
Ludlum Measurements, Inc. Page 2-4 January 2016
Model 2241-3 & 3i Technical Manual Section 3
Section
3
Specifications
0
Linearity: Readings are within 10% of true value with a detector connected.
Display: a four-digit Liquid Crystal Display (LCD) with digits 1.3 cm (0.5
in.) in height. Two additional 0.5 cm (0.2 in.) high digits are used for the overflow counter (SCALER mode) and exponential powers (parameter setup). Enunciators are provided for display units, ALERT, ALARM, low battery, detector OVERLOAD (not used), counting OVERFLOW, and scaler COUNTING.
Warm-up Time: Unit may be used immediately after the LCD initialization
sequence is completed (approximately five seconds after power-up).
RATEMETER: Depending upon how the instrument was calibrated, the
RATEMETER can display in either R/hr, Sv/h, cpm, or cps when the control
switch is in the RATEMETER position.
SCALER: activated by pushbutton in handle when the three-position switch
is in the SCALER position. Count time is adjustable.
Calibration Controls: accessible from the front of the instrument
(protective cover provided). These controls are preset at the factory or calibration lab and should not be adjusted by field personnel.
Discriminator / Input Sensitivity: adjustable from 2 to 100 mV;
negative pulse response
High Voltage: four independent controls, each adjustable from
400-2500 volts; four separate set points for each detector setup-
HV 1, HV 2, HV 3, and HV 4, regulated within 0.2% at 1000 Vdc;
maximum load of 50 µA.
RESET: a pushbutton for zeroing the display, acknowledging
and/or resetting the alarm.
Ludlum Measurements, Inc. Page 3-1 January 2016
Model 2241-3 & 3i Technical Manual Section 3
Note:
The RESET button only silences the alarm in the current mode that the instrument is in. For example, the RESET button will not affect the scaler alarm if the instrument is in the ratemeter mode.
Light: display backlight activated by pushbutton
Audio: built-in audio speaker (unimorph) with AUD ON/OFF switch; at
maximum volume > 60 dB at 2 feet; internal adjustable volume
Alert/Alarm: indicated by either an ALERT or ALARM enunciator on the
display (RATEMETER mode only) and by an audible tone
Power: two each, ˝D˝ cell batteries housed in an externally accessible sealed
compartment. Current draw is approximately 35 mA with the backlight OFF. Minimum battery voltage is 2.2 ±0.1 Vdc.
Battery Dependence: Meter readings vary by less than 3% from fully
charged batteries until the battery symbol appears, indicating the need for recharge or replacement.
Battery Life: typically 200 hours with alkaline batteries (display indicates
low-battery condition). Instrument will operate for approximately 24 hours after the battery symbol first appears.
Size: 16.5 x 8.9 x 21.6 cm (6.5 x 3.5 x 8.5 in.) (H x W x D)
Weight: 1.6 kg (3.5 lb), including batteries.
Removable Switchboard Adjustable Parameters:
Backlight ON Time
: 5, 15, 30, 60, 90, 120, 180, or 240 seconds for the
backlight to stay on when activated by the pushbutton; factory set at 5.
Set Minimum Display: allows lower limit of the auto-ranging display to be
fixed. For example, the display can be set to only show values above or
equal to 1 µR/hr.
Ludlum Measurements, Inc. Page 3-2 January 2016
Model 2241-3 & 3i Technical Manual Section 3
RS-232 Data Dump Mode: enables or disables dump mode to the RS-232
port (˝D˝ type connector). When enabled, the data will be dumped every two seconds.
RS-232 Detector Setup Mode: allows for input of detector parameters via
the RS-232 port
Baud Rate: selects either 150, 300, 600, 1200, 2400, 4800, 9600, or 19200
bps
Detector Dead Time Compensation (DTC): adjustable from 0 to 9999
microseconds
Calibration Constant: adjustable from 0.001 to 280 × 10
9
counts/display
unit
Display Units: can display in R/hr, Sv/h, cpm, or cps
Time Base: can display in seconds or minutes
Audio Divide: 1, 10, 100, or 1000 events per click
Response Time: variable or fixed ratemeter response (All stated times
correspond to a range of 10% to 90% of the final reading). Factory default is variable so that the instrument will automatically adjust the response time to the best setting for the current count rate.
Variable Response: dependent on the number of counts
present, typically 4 to 25 seconds for FAST, and 4 to 60 seconds for SLOW
Fixed Response: The parameter is adjustable from 1-25
seconds resulting in a FAST response from approximately 2-50 seconds. The SLOW response is approximately 10-250 seconds. For MDA-type measurements, the fixed response mode is recommended.
Ratemeter Alert/Alarm: set at any point corresponding to the pre-selected
ratemeter range
Scaler Alarm: adjustable from 1 to 999999 counts
Scaler Count Time: adjustable from 1 to 9999 seconds
Ludlum Measurements, Inc. Page 3-3 January 2016
Model 2241-3 & 3i Technical Manual Section 3
Internal Detector Specifications for Model 2241-3i:
Operating Voltage
Input Sensitivity: -30 mV ±10 mV
Dead Time: typically 80 microseconds
Energy Reponse: within ±15%
Tube: 30 mg/cm
Size: 2.3 x 10.2 (0.9 x 4 in.) (Dia x L)
Temperature Range: -20 to 50 °C (-4 to 122 °F)
: recommended 550 volts
2
stainless steel (Halogen quench) GM
Ludlum Measurements, Inc. Page 3-4 January 2016
Model 2241-3 & 3i Technical Manual Section 4
Section
Identification of Controls and
4
Functions
Display
The Model 2241-3 utilizes a four-digit liquid crystal display (LCD) with a two-digit overflow (SCALER mode) and moving decimal point. The two
smaller digits located in the lower right corner of the display indicate counter OVERFLOW when in the scaler counting mode (equivalent to a six-digit scaler) or exponential power when in the parameter setup mode. The upper right corner of the LCD displays units and multiplier(s) - R/hr, mR/hr, or µR/hr; Sv/h, mSv/h or µSv/h; C/m, kC/m, C/s or kC/s. The bottom part of the readout displays the
ALARM, ALERT, OFLOW, OVERLOAD (not used) enunciators and the
low-battery icon. COUNTING indicates that the scaler mode has been initiated and is in the counting process.
Display Status Definitions
ALARM: Ratemeter or scaler count has increased above the preset alarm
threshold. An audible continuous tone will accompany the latching ALARM condition. Depressing RESET will acknowledge the audible ratemeter and/or scaler alarm. Depressing RESET a second time will reset the ratemeter reading and ratemeter alarm. To reset the scaler ALARM, depress the COUNT switch located in the carrying handle to re-initiate the scaler count cycle.
ALERT: Ratemeter count has increased above the preset alert threshold. To
reset an ALERT condition, press RESET once if in the non-alarm condition and twice if in an alarm condition. (The first depression in the alarm condition acknowledges the audible alarm.) The ratemeter will reset to the minimum displayable reading each time the alert is reset.
OFLOW (Overflow) [RATEMETER mode]: Indicates that the incoming count
exceeds the capability to display stable or reliable readings corresponding to the radiation level being measured. The overflow symbol will appear when the ratemeter exceeds 100 kcps or if the dead time correction is greater than
OFLOW will appear in the SCALER mode when the six-digit display
75%. (four digits display and two overflow digits in right corner) reaches “999999” and starts to roll over again.
Ludlum Measurements, Inc. Page 4-1 January 2016
Model 2241-3 & 3i Technical Manual Section 4
Low Battery icon: indicates that the batteries have decreased to the
minimum operating voltage of 2.2 ±0.1 Vdc. Instrument will continue to operate for approximately 24 hours thereafter.
COUNTING: indicates that the scaler COUNT switch has been depressed and
that the scaler is accumulating counts for the pre-determined count time
Front Panel Controls
OFF/DET 1/DET 2/DET 3/DET 4 Switch: a five-position rotary switch that
applies power to the instrument and selects the desired set of detector parameters
SCA/RATE (Mode) Switch: a protected (locking) toggle switch that toggles
between ratemeter or scaler counting modes. The user pulls up and then over to change modes.
AUD ON/OFF Switch: The click-per-event audio may be silenced or enabled
via this front-panel toggle switch. The audible alarm is independent of the
AUD ON
audible alarm can only be silenced by depressing the RESET button.
/OFF switch and will override the audible click-per-event. An
F/S (Fast/Slow) Response Switch: a two-position toggle switch that selects
fast or slow counting response time
Variable Response: The F position allows the time constant (TC) to vary
from 1 to 10 seconds, while the S position varies from 1 to 30 seconds. The response time is automatically adjusted in proportion to the incoming count rate between the
Fixed Response: The F position corresponds to the selected fixed response
F/S TC variables.
time - TC. The S position is five times slower than the selected fast TC.
LIGHT (LCD Backlight): A pushbutton switch, when depressed, illuminates
the LCD for a pre-programmed time. The backlight ON time can be selected between 5 and 240 seconds during the parameter setup.
Ludlum Measurements, Inc. Page 4-2 January 2016
Model 2241-3 & 3i Technical Manual Section 4
RESET Pushbutton: In the non-alarm condition, depressing the RESET
switch resets the ratemeter display to the minimum display readout. In an alarm condition (ratemeter or scaler), depressing audible alarm. Depressing RESET a second time will reset the ratemeter alarm and/or alert condition. The scaler alarm can only be reset by depressing the scaler count switch located in the end of the Model 2241-3 handle.
Note:
The RESET button only silences the alarm in the current mode that the instrument is in. For example, the RESET button will not affect the scaler alarm if the instrument is in the ratemeter mode.
RESET will silence the
Scaler Count Switch: pushbutton switch located in the end of the Model
2241-3 carrying handle, which when depressed, initializes the start of the scaler count accumulation for the preset scaling time. The SCA/RATE switch must be in the SCA position to initiate the counting cycle. The scaler display uses the two digits in the lower right-hand corner for the two most significant digits of the six-digit readout. Scaling time can be set from 1 to 9999 seconds in the parameter setup by way of the switch board. Depressing the Count switch after a scaler alarm will reset the scaler display to 0, resetting the alarm condition.
F ront P anel Calibration Controls
Note:
Remove the front-panel calibration cover to expose the following calibration controls:
DISC (Discriminator): a multi-turn potentiometer (approximately 20
revolutions) used to vary the detector pulse-counting threshold from 2 to 100 millivolts. A Ludlum Model 500 Pulser or equivalent should be used in checking or adjusting the pulse discrimination parameter.
Ludlum Measurements, Inc. Page 4-3 January 2016
Model 2241-3 & 3i Technical Manual Section 4
Note:
When making adjustments to the HV potentiometers, make note of the following precautions: Use a Ludlum Model 500 Pulser or high-impedance voltmeter with a high-voltage probe to measure the high voltage at the detector connector. If a Ludlum Model 500 Pulser is not available, ensure that the impedance of voltmeter used is 1000 megohms or greater.
HV1, HV2, HV3, HV4: multi-turn potentiometers (approximately 20
revolutions) that vary the detector voltage from 400 to 2500 volts. The maximum high-voltage output is adjusted by the HV LIMIT potentiometer located on the internal main board.
Main Board Controls
Note:
To access the internal circuit boards, unlatch the latches at each end of the Model 2241-3. Carefully separate the top chassis from the bottom cover (referred to as a can). The can has the audio speaker (unimorph) with a two-conductor cable attached to the main board. The audio plug may be disconnected during the internal control adjustments.
HV LIMIT (R027): A multi-turn potentiometer (approximately 20
revolutions) sets the maximum HV limit with the front-panel adjusted to the maximum clockwise position. It is adjustable from 1250 to 2500 Vdc.
VOLUME (R002): A multi-turn potentiometer (approximately 20
revolutions) varies the audible click-per-event and alarm audio. Adjust the control to the maximum clockwise position for maximum volume. If the
VOLUME control is adjusted to the maximum counterclockwise position, the
clicks-per-event, or the audible alarm(s), will not be audible when active.
HV control
Ludlum Measurements, Inc. Page 4-4 January 2016
Model 2241-3 & 3i Technical Manual Section 4
Switch Board Controls
The switch board utilizes a 16-position rotary switch (FUNCTION) to select the 16 setup parameters. (Refer to schematics and component layout drawing near the end of the manual.) All of the setup parameters are stored in the non-volatile EEPROM, which will retain data even after the Model 2241-3 batteries are removed. After the parameters are entered, the switch board can be removed and the Model 2241-3 will continue to operate from the previously programmed information. Changing parameters and information on switchboard controls are covered in detail in Section 8 of this manual.
Ludlum Measurements, Inc. Page 4-5 January 2016
Model 2241-3 & 3i Technical Manual Section 5
Section
5
Safety Considerations
Environmental Conditions for Normal Use
Indoor or outdoor use
No maximum altitude
Temperature range of –20 to 50 °C (-4 to 122 °F)
Maximum relative humidity of less then 95% (non-condensing)
Pollution Degree 3 (as defined by IEC 664). (Occurs when conductive pollution or dry nonconductive pollution becomes conductive due to condensation. This is typical of industrial or construction sites.)
Detector Connector
Caution!
The detector operating voltage (HV) is supplied to the detector by way of the input connector. A mild electric shock may occur if contact is made with the center pin of the input connector. Switch the Model 2241-3 to the OFF position before connecting or disconnecting the cable or detector.
Warning Markings and Symbols
Ludlum Measurements, Inc. Page 5-1 January 2016
Caution!
The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc.
Model 2241-3 & 3i Technical Manual Section 5
Caution!
Verify instrument voltage input rating before connecting to a power converter. If the wrong power converter is used, the instrument and/or power converter could be damaged.
The Model 2241-3 Survey Meter is marked with the following symbols:
CAUTION, RISK OF ELECTRIC SHOCK (per ISO 3864, No. B.3.6):
designates a terminal (connector) that allows connection to a voltage exceeding 1 kV. Contact with the subject connector while the instrument is on or shortly after turning off may result in electric shock. This symbol appears on the front panel.
CAUTION (per ISO 3864, No. B.3.1): designates hazardous live voltage and risk of electric shock. During normal use, internal components are hazardous live. This instrument must be isolated or disconnected from the hazardous live voltage before accessing the internal components. This symbol appears on the front panel.
Warning!
The operator is strongly cautioned to take the following precautions to avoid contact with internal hazardous live parts that are accessible using a tool:
1. Turn the instrument power OFF and remove the batteries.
2. Allow the instrument to sit for one minute before accessing any internal components.
The “crossed-out wheelie bin” symbol notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding. Each material must be separated. The symbol is placed on the battery compartment. See Section 9, “Recycling,” for further information.
The “CE” mark is used to identify this instrument as being acceptable for use within the European Union.
Ludlum Measurements, Inc. Page 5-2 January 2016
Model 2241-3 & 3i Technical Manual Section 6
Section
6
Maintenance
Instrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 2241-3 may be externally cleaned with a damp cloth (using only water as the wetting agent). Do not immerse the instrument in any liquid. Observe the following precautions when cleaning:
1. Turn the instrument OFF and remove the batteries.
2. Allow the instrument to sit for one minute before performing
any external cleaning or accessing internal components for maintenance.
Operational Check
An instrument operational check should be performed prior to each use by exposing the detector to the supplied check source, confirming the proper reading on each scale. See Section 2, Subsection “Operational Check,” for further details.
Recalibration
Recalibration should be accomplished after any maintenance or adjustment of any kind has been performed on the instrument. Battery and cable replacements (of same length) and exterior cleanings do not normally require the instrument be recalibrated.
Ludlum Measurements, Inc. Page 6-1 January 2016
Note:
Ludlum Measurements, Inc. recommends recalibration at intervals no greater than one year. Check the appropriate local state and federal regulations to determine required recalibration intervals.
Model 2241-3 & 3i Technical Manual Section 6
Ludlum Measurements offers a full-service repair and calibration department. We not only repair and calibrate our own instruments, but most other manufacturers’ instruments as well.
See Section 8 “Instrument Setup & Calibration,” for further details on instrument calibration.
Batteries
The batteries should be removed and the battery contacts cleaned of any corrosion at least every three months. If the instrument has been exposed to a very dusty or corrosive atmosphere, more frequent battery servicing should be used. Use a spanner wrench to unscrew the battery contact insulators, exposing the internal contacts and battery springs. Removing the handle will facilitate access to these contacts.
Note:
Never store the instrument over 30 days without removing the batteries. Although this instrument will operate at very high ambient temperatures, battery seal failure can occur at
temperatures as low as 38 °C (100°F).
Ludlum Measurements, Inc. Page 6-2 January 2016
Model 2241-3 & 3i Technical Manual Section 7
Section
7
Refer to the Main Board schematic for the following:
T echnical Principle of Operation
Detector Input/Amplifier
Negative-going detector pulses are coupled from the detector through C021 to amplifier U021. R024 and CR021 protect the input of U021 from inadvertent shorts. Self-biased amplifier U021 provides gain in proportion to R022, divided by R025. Transistor pins 4, 5, and 6 of U021, provide amplification. Pins 10-15 of U021 are coupled as a constant current source to pin 6 of U021. The output is self-biased to 2 Vbe (approximately 1.4 volts) at pin 7 of U021. This provides just enough bias current through pin 6 of U021 to conduct all of the current from the constant current source. Positive pulses from pin 7 of U021 are coupled to the discriminator (U011) through R031 and C012.
Discriminator
Positive pulses from amplifier U021 are coupled to pin 2 of U011 comparator. The discrimination level is set by the DISC control connected to pin 3 of U011. As the positive pulses at pin 2 of U011 increase above DISC reference at pin 3, pin 1 goes low, producing a low pulse. Pin 1 of U011 is normally held high (+5 volts) by R014.
The low pulse from pin 1 of U021 is coupled to univibrator U001. U001 shapes and fixes the pulse width to approximately 10 µs. The univibrator is configured in the non-retriggerable mode. Negative pulses from pin 9 of U001 are coupled to the µP for counting.
Low V oltage Supply
Battery voltage is coupled to DC-DC converter U231. U231 and related components provide +5 V to power the µP, op-amps, and logic circuitry. R135 and R136 provide voltage division for low-battery detection. Pin 6 of U231 provides a low signal when the battery voltage decreases to +2.2 ±0.1 Vdc. U121 provides the +2.5 Vdc reference for the HV and DISC control references.
Ludlum Measurements, Inc. Page 7-1 January 2016
Model 2241-3 & 3i Technical Manual Section 7
High V oltage Supply
High voltage is developed by blocking oscillator Q241, T141, and C244 and rectified by voltage multiplier CR041-CR043, C041-C043, and C141. High voltage increases as current through R241 increases, with maximum output voltage with Q241 saturated.
High voltage is coupled back through R034 to op-amp pin 2 of U131. Resistor network R027, R132 completes the HV division circuit to ground. R027 provides HV limit from 1250-2400 when the HV control on the calibration board is at maximum. The regulated HV output is controlled by the HV1 and HV2 potentiometers located under the CAL cover on the front panel. This control provides the reference for comparator pin 3, U131. During stable operation, the voltage at pin 2 of U131 will equal the voltage at pin 3 of U131. Pin 1 of U131 will cause conduction of Q141 to increase or decrease until the HV finds a level of stability.
Microprocessor (µP)
U111 controls all of the data, control inputs, and display information. The clock frequency is crystal-controlled by Y221 and related components at
6.144 MHz. The µP incorporates internal memory (ROM), storing the program information. U1 resets the µP at power-up to initiate the start of the program routine. During the program loop, the µP looks at all the input switches for initiation or status changes and responds accordingly.
U122 is a 256 x 8 bit EEPROM used to store the setup parameters. The information is transferred serially from the µP. The EEPROM is non-volatile, meaning it retains memory even after power is removed.
Audio
Click-per-event, divide-by, and alarm audio pulse frequency is generated by the µP and coupled to Q101. Q101 then inverts the pulses and drives the bottom of T101. Bias voltage is provided by the volume control (R002) to the top of T101.
Refer to the Switch Board schematic for the following:
Ludlum Measurements, Inc. Page 7-2 January 2016
S1 (FUNCTION)
S1 is a 16-position binary rotary switch, which selects the programmable parameters for the Model 2241-3. The switch selects the parameters using the hexadecimal numbering system via buss lines SW1-SW4.
Model 2241-3 & 3i Technical Manual Section 7
S2-S4
S2-S4 are pushbutton switches which enter/change the variables for each of the 16 parameters.
U1
U1 is a +5 V powered RS-232 driver/receiver used to interface the Model 2241-3 to a computer.
Refer to Display Board schematic for the following:
LCD Driv e
U111 and U211 are serial input 32-bit LCD drivers. The data is loaded serially into the 32-bit shift registers (internal) via the “D” IN input. The
LOAD input instructs the shift register to receive data while the CLOCK input
shifts the data through the 32-bit registers. After all the data is loaded, the
LOAD line is pulsed by the µP, instructing the registers to transfer the data to
the LCD drivers. The backplane (BP) signal from U211 provides the reference signal (approximately 125 Hz at 5 Vdc) to the LCD (DSP1) BP connection. When a segment is illuminated, the signal to that segment will be out-of-phase with the BP signal. If the segment is OFF, the signal will be in-phase with the BP signal.
Backlight Driv e
Depressing the LIGHT button instructs the µP to set the BACKLIGHT line, pin 31 on µP, "low" for the predetermined backlight ON time. (Refer to main board schematic for details.) A "low" condition on pin 31 causes Q212 to conduct sending +3 V to P8-3 on the display board with +3 V at P8-3 (refer to display board schematic). Backlight oscillator Q011, T011, and related components start to oscillate, producing a 2.5 kHz, sine wave signal. The signal is amplified by T011 to 150 volts peak-to-peak to drive the LCD backlight.
Ludlum Measurements, Inc. Page 7-3 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Section
8
Instrument Setup & Calibration
Entering or Changing Switch Board Parameters
On the switch board, select the desired parameter to enter or change by using the corresponding FUNCTION switch position. Depress the ENTER button, and a character on the LCD will start to flash. The flashing character indicates that the program is in the parameter change mode.
To change the character, press the UP button until the desired variable is reached. To shift to another character, increment the LEFT pushbutton until the desired character is reached. The LEFT pushbutton switch enables the operator to sequence through all the characters on the LCD associated with a particular parameter.
Once the desired data is entered, depress the ENTER button. The LCD characters should stop flashing and the new parameter data should display.
Note:
Loading Default Parameters
To load the default parameters for all detector setups, hold down the UP pushbutton on the switch board while turning the instrument on until DEF is displayed on the LCD. The table on the following page shows the default values.
Ludlum Measurements, Inc. Page 8-1 January 2016
The DETECTOR SELECT switch allows the Model 2241-3 to have four sets of operating parameters.
Model 2241-3 & 3i Technical Manual Section 8
Model 2241-3 Cpm uR/hr R/Hr mR/hr
Setup 01 Setup 02 Setup 03 Setup 04 Dead Time
Cal Const 100e-2 105e+8 108e+4 198e+6
Rate Alarm 50.0 kcpm 50 µR/hr 5 R/hr 5 mR/hr
Scaler Alarm 85000 85000 85000 85000 Count Time 12 s 12 s 12 s 12 s Time Base Minutes Seconds Seconds Seconds Units cpm R/hr R/hr R/hr Audio Divide-By 1 1 1 1 Response 0 0 0 0 Check Source 0 0 0 0 Percent CS 0 0 0 0
Rate Alert 20.0 kcpm 20 µR/hr 2 R/hr 2 mR/hr
Minimum Display 0.00 cpm
Baud Rate 9600 LCD Time Off 5 s Detector 0
0 µs0 µs0 µs 0 µs
00.0
µR/hr
00.0
µR/hr
00.0
µR/hr
The Function Switch
FUNCTION Switch: a 16-position rotary switch labeled “0-9” and “A-F.”
This switch selects a parameter setup mode for the Model 2241-3. If the board is not installed, the normal operation mode (counting mode) is selected. If the switch board is installed, the selector switch must be set to the 0 position for normal instrument operation. The following may be changed using the switch board, and are discussed in detail in this section:
Detector parameters Durrent detector setup in use RS-232 communication baud rate RS-232 data dump mode RS-232 detector parameters set/read mode
Function Switch P osition Descriptions and Variables
POSITION 0: NORMAL OPERATION places the Model 2241-3 in the normal
(counting) operating mode. Unplugging the switch board from the Model 2241-3 main board defaults to the normal operating mode.
Ludlum Measurements, Inc. Page 8-2 January 2016
Model 2241-3 & 3i Technical Manual Section 8
POSITION 1: DEAD TIME (µs) allows changing the detector dead time
correction for the current detector setup. Setting this parameter to “0” disables dead time correction. The dead time adjusts from 0 to 9999 microseconds (µs). The incoming counts are adjusted for dead time using the following formula:
n =
1-m
cps x time base
CC =
m
rate
Where,
n = corrected counts per second
m = incoming count per second
= system dead time
POSITION 2: calibration constant allows changing the calibration constant
for the current detector setup. The calibration constant (CC) adjusts from
0.001 to 280 × 109. The calibration constant converts counts/time base to units/time base. The CC must be set to 1 to readout in cps (counts per second) or cpm (counts per minute).
CC CONVERSION TABLE
Conversion Rate Multiply by to get CC
cps/µR/hr 3.6 x 109 cps/mR/hr 3.6 x 106 cps/R/hr 3.6 x 103 cpm/µR/hr 6.0 x 107
4
cpm/mR/hr 6.0 x 10 cpm/R/hr 6.0 x 101 cps/µSv/h 3.6 x 107 cps/mSv/h 3.6 x 104 cps/Sv/h 3.6 x 101 cpm/µSv/h 6.0 x 105 cpm/mSv/h 6.0 x 102 cpm/Sv/h 0.6
Ludlum Measurements, Inc. Page 8-3 January 2016
Example:
The Model 44-9 GM detector produces approximately 3300 cpm/mR/hr for
137
Cs: 6.0 x 104 x 3300 = 198 x 106 for
CC.
Model 2241-3 & 3i Technical Manual Section 8
POSITION 3: DISPLAY UNITS selects the display units for the associated
detector setup number. The Model 2241-3 and detector may be calibrated in either exposure rate (R/hr or Sv/h) by entering the appropriate Calibration Constant (position 2) and Dead Time correction (position 1). The Model 2241-3 will automatically convert to the correct reading when switching between R and Sv.
The time base for count “C” is set independently in position 4. The display units may be set to:
R/hr (Roentgens per hour)
Sv/h (Sieverts per hour)
C/time base (Counts per time)
The display is auto-ranging with the appropriate multiplier symbol appearing in front of the “R,” “Sv,” or “C,” indicating the range:
µR/hr, mR/hr, R/hr
µSv/h, mSv/h, Sv/h
C/s, kC/s, C/m, kC/m
POSITION 4: (TIME BASE) CPS or CPM selects the display time base for the
current detector setup. This time base only applies if the units are set to C/ (Counts/time). The time base for R/hr and Sv/h is fixed in “hr.” For “true” reading (pulser calibration) cpm or cps calibrations, set the Calibration Constant (CC, parameter 2) to read “1.” For geometry calibrations, the detector efficiency can be entered for CC.
Example:
For alpha scintillation detector with 25% 2π efficiency; enter “250 10-3” in the CC parameter setup.)
The display time base may be set to:
seconds (s) minutes (m)
Ludlum Measurements, Inc. Page 8-4 January 2016
Model 2241-3 & 3i Technical Manual Section 8
POSITION 5: AUDIO DIVIDE-BY selects the audible click-per-event division
rate for the current detector setup. If the AUD ON/OFF switch is in the OFF position, no audible click-per-event will be heard.
This parameter ranges from:
0 = Divide By 1
1 = Divide By 10
2 = Divide By 100
3 = Divide By 1000
POSITION 6: RESPONSE TIME allows changing the time constant (TC) for
the current detector setup. If the response is set to 0, the Model 2241-3 automatically calculates (for variable mode) the time constant based on the incoming cps. If a variable of 1-199 is entered for TC, the response time becomes fixed.
Variable Response - Response time is varied in proportion to
the incoming count rate. The two-position F/S (Fast/Slow) toggle switch selects the maximum time constant (TC) for the variable mode. The fast position varies the TC from 4-25 seconds, and the slow position varies from 4-60 seconds.
Fixed Response - The Fast (F) response position is
programmable from 2-50 seconds, and the slow response is 5 times slower than the fast TC. For MDA-type measurements, the fixed response time mode is recommended.
POSITION 7: RATEMETER ALARM/ALERT allows changing the ratemeter
alarm for the current detector setup. The units of this alarm are the same as the units for the ratemeter display. The fifth push of the left button allows the decimal point to be moved. The ratemeter alarm adjusts from 1 to 999 R/hr (or Sv/h) or 1 to 999 kcpm or 1 to 100 kcps. The units of the alarm are determined by the units for the ratemeter.
POSITION 8: SCALER ALARM/COUNT TIME sets the scaler alarm variable
from 1-999999, corresponding to the accumulated scaler count. After the scaler alarm variable is entered, the scaler count time is prompted. The scaler count time is adjustable from 1-9999 seconds.
POSITION 9: NOT USED
Ludlum Measurements, Inc. Page 8-5 January 2016
Model 2241-3 & 3i Technical Manual Section 8
POSITION A: NOT USED
POSITION B
backlight will stay on after pressing the front panel switch labeled LIGHT. This value is stored in EEPROM.
POSITION C: SET MINIMUM DISPLAY sets the ratemeter minimum
displayable reading. Depressing the RESET button displays the minimum ratemeter units. The readout will auto-range up to the maximum displayable but will display 0 for ratemeter readings below the user-programmed minimum variable.
: LCD Backlight ON TIME is the amount of time that the LCD
Available values are:
5 seconds
30 seconds
60, 90 seconds
180, 240 seconds.
Minimum displayable values are:
00.0 µ, 000 µ, 0.00 m, 00.0 m, 000 m, 0.00, 00.0, 000 R/hr
.000 µ, 000 µ, 00.0 m, 000 m, 0.00, 00.0, 000 Sv/h
0.00, 00.0, 000, 0.00 k, 00.0 k, 000 kcpm, or cps
POSITION D: RS-232 DATA DUMP MODE allows the RS-232 port to dump
ratemeter data every two seconds. The Model 2241-3 is fully functional during RS-232 data dump with the exception of the audio function. The LCD will alternate between display of the ratemeter and the word “dUP” (representing "dump").
POSITION E: RS-232 DETECTOR PARAMETERS SETUP MODE allows the RS-
232 port to accept/send a string of parameters corresponding to the current detector setup values.
POSITION F: BAUD RATE configures the RS-232 port for the following baud:
150, 300, 600, 1200, 2400, 4800, 9600, and 19200. The data is eight data bits, one stop bit with no parity bit. This value is stored in EEPROM. The baud rate can only be programmed through the switch board.
Ludlum Measurements, Inc. Page 8-6 January 2016
Model 2241-3 & 3i Technical Manual Section 8
RS-232 PORT CONNECTOR: This nine-pin “D” type connector is designed
as a DCE port. A straight wire cable (extension cable) connects the Model 2241-3 to a computer’s nine-pin RS-232 port.
RS-232 CONNECTOR PIN OUT:
PIN FUNCTION
1 NC (No Connection) 2 DATA OUT 3 DATA IN 4 NC 5 NC 6 NC 7 HANDSHAKING IN 8 HANDSHAKING OUT 9 NC
Note:
Ludlum Measurements, Inc. offers a PC compatible software program, which incorporates the read/write commands necessary to communicate between the PC and the Model 2241-3. The program also incorporates an algorithm to calculate the detector Calibration Constant and Dead Time Constant. The software is offered in a DOS version (part number 1370-025) or a WINDOWS version (part number 1370-024). Read the Software License Agreement at the end of this section before installing any LMI software.
Calibration
The Model 2241-3 calibration routine consists of entering detector parameters into memory by way of the switch board and adjusting the CAL controls (HV and DISC) for the specific detector operating requirements.
The first subsection of calibration will give a general overview of detector setup, including the determination of various detector operating voltages (HV) and the adjustment of counter input sensitivity (DISC).
The next subsection deals with pulse generator counts-per-minute calibration. The counts-per minute-parameter setup is used in the initial
Ludlum Measurements, Inc. Page 8-7 January 2016
Model 2241-3 & 3i Technical Manual Section 8
instrument checkout procedure and the variables are saved under detector setup number “1” when shipped from Ludlum Measurements, Inc.
The following subsection deals with exposure rate calibration. The detector Calibration Constant (CC) and Dead Time Correction (DTC) are the two primary parameters used in the exposure rate calibrations (R/hr and Sv/h). These two constants are alternately varied to achieve linearity at the detector non-linear operating regions. An example of the Ludlum Model 44-9 GM detector calibration is given at the end of this section to illustrate the algorithm used in determining the CC and DTC variables.
G
ENERAL DETECTOR SETUP INFORMATION
The operating point for the instrument and probes is established by setting the probe voltage and instrument sensitivity (HV and DISC). The proper selection of this point is the key to instrument performance. Efficiency, background sensitivity, and noise are fixed by the physical makeup of the given detector and rarely vary from unit to unit. However, the selection of the operating point makes a marked difference in the apparent contribution of these three sources of count.
In setting the operating point, the final result of the adjustment is to establish the system gain so that the desirable signal pulses (including background radiation) are above the discrimination level and the unwanted pulses from noise are below the discrimination level and are therefore not counted.
The total system gain can be controlled by adjusting either the instrument sensitivity or the high voltage.
HV controls the gain of the detector; and DISC
(Discriminator) controls the instrument counting threshold (sensitivity).
In the special case of GM detectors, a minimum voltage must be applied to establish the Geiger-Mueller characteristic. Further changes in
HV will have
little effect on this type of detector.
GM Detectors: The output pulse height of the GM detector is not
proportional to the energy of the detected radiation. Adjusting DISC will have minimal effect on observed count rate unless the DISC setting is so low that the instrument will double-pulse.
Ludlum Measurements, Inc. Page 8-8 January 2016
Model 2241-3 & 3i Technical Manual Section 8
For most GM detectors, set DISC for 30-40 millivolts and adjust HV to the GM detector recommended high voltage. Most GM detectors operate at 900 volts, although some miniature detectors operate at 450-550 volts. If a recommended setting is unavailable, plot count rate versus HV to produce a plateau graph. Adjust the HV for 25-50 volts above the knee or start of the plateau. For mixed detector use, both sensitivity and high voltage may be tailored for other detectors as long as the GM detector is operated within the recommended voltage range. Caution must be observed in lowering the input sensitivity to ensure that the counter does not double or multi-pulse.
Alpha Air-Proportional Detectors: For air proportional alpha detectors, set
the DISC for 2 millivolt discrimination. Adjust HV until the detector just breaks down (shown by a rapid increase of count rate without a source present). Measure the HV output, then decrease the 100 volts below breakdown.
Proportional Detectors: For proportional detectors, set the DISC control for
2-millivolt discrimination (near maximum clockwise). Expose the detector to a check source and plot count rate versus HV, similar to the one in the figure below. Refine the HV adjustment for optimum source efficiency with a minimum acceptable background count.
HV setting to operate
Scintillators: Set the DISC for 10 millivolts. Plot background and
source counts versus HV to produce a plateau graph similar to the one in the figure. Adjust the HV to 25-50 volts above the knee or start of the plateau. This provides the most stable operating point for the detector.
COUNTS PER MINUTE (C/M) CALIBRATION
This procedure will set up the Model 2241-3 for the Counts/minute (C/m) mode of operation. Refer to Section 8, (Page 8-2 and following) for more information on setup parameter variables.
A Ludlum Model 500 Pulser or equivalent is required. If the pulser does not have a high-voltage display, use a high-impedance voltmeter with at least 1000 megohms input resistance to measure the detector high voltage.
Ludlum Measurements, Inc. Page 8-9 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Switch the SCA/RATE switch to the RATE position. Select position DET1 on the detector selector switch located on the front panel.
Select FUNCTION switch positions 1-6 and adjust for the following parameters:
Switch Pos. Parameter Function
1 0000 s-6 Dead Time 2 0100 -2 Calibration Constant 3 c/ Display Units 4 m Timebase 5 1 Audio Divide-by 6 000 s Response Time
Position 7 selects the desired ratemeter
ALERT and ALARM trip points.
If the parameters are undetermined, arbitrarily choose “0050 kC/m” for the alarm and “0045 kC/m” for the alert to confirm operation of the alert/alarm function.
Position 8 selects the scaler alarm parameter and the scaler count time.
If the values are unknown, set the scaler alarm to “4500
ALARM00
and the count time to “0060” (60 second count time).
Position 9 is not used, and position A is not used.
Switch to position B and enter “15” for a 15-second backlight ON time.
Switch to position C and enter “00.0 C/m” for the minimum displayable value.
Select position 0 to return to normal operation.
Connect the Model 500 Pulser to detector input and adjust HV and DISC to the specific detector operating parameters.
Ludlum Measurements, Inc. Page 8-10 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Adjust the pulser amplitude to 1.5 times the Model 2241-3
discrimination level.
Adjust the pulser output to 800 cpm and confirm that the Model
2241-3 reads 800 cpm ±10% on the ratemeter setting.
Adjust the pulser output to 200 cpm and confirm that the Model
2241-3 reads 200 cpm ±10% on the ratemeter setting.
Adjust the pulser output to 800 cpm, take a one minute count and
confirm that the digital scaler readout displays 800 cpm ±2%.
Adjust the pulser output to 200 cpm, take a one minute count and
confirm that the digital scaler readout displays 200 cpm ±2%.
Confirm that the 20% and 80% readings for the upper decades are
within the pulser input by decading the pulser count output.
Confirm that the scaler readout is within 2% of the pulser input rate. Ensure that the ALERT and ALARMs function by inputting the preset
alarm levels as to initiate the alert and alarm conditions.
R/hr CALIBRATION
The following calibration procedure assumes that detector Calibration Constant (CC) and Dead Time Constant (DTC) are already known. If these constants must be determined, reference the following subsection, “Determining CC and DTC.”
Switch the toggle switch to
DET2. Detector setup number “1” is usually
reserved for the Counts/minute parameter calibration. Rotate the
FUNCTION switch counterclockwise to position 1 and enter the detector
Dead Time in µs. Rotate to position 2 and enter the Calibration Constant. Enter the desired parameters for positions 3-F. Switch to position 0 for normal operation.
Expose the detector to calibrated radiation fields extending from the lower to the upper operating range of the detector. Confirm that the linearity is within 10% of each respective reading. If the readings are off on the lower detector operating region, vary CC. If the readings are off at the upper end of the detector operating region, adjust DTC.
Ludlum Measurements, Inc. Page 8-11 January 2016
Model 2241-3 & 3i Technical Manual Section 8
DETERMINING CC AND DTC
This procedure contains the algorithm (hi-lo method) for determining the CC (Calibration Constant) and the DTC (Dead Time Correction). An example of the Ludlum Model 44-9 GM detector calibration is used in conjunction with the algorithm calculations to aid in solving the equations.
Note:
Ludlum Measurements, Inc. offers a PC-compatible software program, which incorporates the read/write commands necessary to communicate between a PC and the Model 2241-3. The program also incorporates the algorithm to calculate the detector CC and DTC. The software is offered in a DOS version (part number 1370-025) or a WINDOWS version (part number 1370-024). Read the Software License Agreement at the close of this section prior to installing any LMI software.
Hi-Lo Method
: The hi-lo method refers to the placement of the detector in a
radiation field using a two-point (CC and DT) calibration to make linear the detector response, even in the non-linear operating regions of the detector. The low-radiation field (CC) should be a field that yields from 2 to 5 % count loss. The high radiation field (DT) should be a field that yields from 30 to 60 % count loss. The algorithm ignores background counts, and therefore, the low field must be at least ten times the background count.
The following summary lists the calibration constraints.
Calibration and Dead Time Calibration Constraints
FIELD CONSTRAINT
BACKGROUND *10 times less than low field
LOW FIELD Yields from 2 to 5 % count loss
HIGH FIELD Yields from 30 to 60 % count loss
* This constraint only applies when using two sources (two fields) or a radiation range calibrated without background consideration.
Ludlum Measurements, Inc. Page 8-12 January 2016
Model 2241-3 & 3i Technical Manual Section 8
T
Preliminary CPS Setup
Refer to Section 8, Subsection “Function Switch Position Descriptions and Variables,” for cps readout variables.
Equation 1
LO
LO
2%
5%
CPS
Equation 2
CPS
Select position
DET1 on the detector selector switch located on the front
panel. Starting with FUNCTION switch position 1, enter the following variables:
SWITCH POS. PARAMETER FUNCTION
1 0000s-6 Dead Time 2 0100 -2 Calibration Constant 3 C/ Display Units 4 m Timebase 5 N/A Audio Divide-By 6 N/A Response Time 7 N/A Ratemeter Alm./Alert 8 0060 s ScalerAlm./Count Time
9 Not Used
1
=
DT x 49
A Not Used B N/A LCD Backlight C 000 C/s Set Minimum Display D-F N/A RS-232 Parameters
1
=
DT x 19
Equation 3
CPS
HI
30%
=
1
Equation 4
60%
=
1.5 D
CPS
HI
Ludlum Measurements, Inc. Page 8-13 January 2016
The equations to the left (Equations 1-4) determine the hi and lo radiation fields used to acquire counts for the CC and DTC algorithm. These calculations require an unknown variable, DT (Dead Time). Typical dead
DT x 2.3333
times for some of the standard LMI detectors are referenced in the table at the end of this section. The lo count field should be a field that yields between 2 and 5% count loss. The hi count field (CPS that yields between 30 and 60% count loss.
HI
) should be a field
Model 2241-3 & 3i Technical Manual Section 8
Reference the table at the end of this section to determine the cps/exposure rate (cps/ER). The conversion can be determined by placing the detector in
a radiation field, which produces from 50
cps
= cps/ER
to 200 cps. Calculate the count/exposure rate using the equation to the left.
radiation field in exposure rate units
For example, exposing an LMI Model 44-
9 to a 2 mR/hr
137
Cs field yields approximately 110 cps so that:
cps 110
mR/hr 2
Abbreviations used:
cps/mR/hr 55 =
The typical dead time for a Model 44-9 is approximately 85 µs. Therefore, using 85 µs for “DT” in equations 1-4, the lo field should be between 240 and 619 cps, and the hi field is between 5040-17,650 cps. Dividing the cps values by the 55 cps/mR/hr conversion equates to between 4-11 mR/hr for the lo field and 91-320 mR/hr for the hi field.
Select a calibrated field between the lo and hi data points determined above:
lo (CALlo) = 8 mR/hr
hi (CALhi) = 200 mR/hr
The following procedure outlines the hi-lo method
units = Sv, R, counts.
CALlo = lo field calibration point. CALhi = hi field calibration point.
CORRlo = recorded field at low calibration point. CORRhi = recorded field at high calibration point.
DT = dead time constant entered into Model 2241. CC = calibration constant entered into Model 2241.
fd and ad are intermediate steps in calculating DT. f
is an intermediate step in calculating CC.
cal
Ludlum Measurements, Inc. Page 8-14 January 2016
Model 2241-3 & 3i Technical Manual Section 8
e
s
60
m 60
h
CC and DTC Algorithm
Equations (5) and (6) convert units per time (R/hr Display Units) to units per second:
units
tim
Equation 5
CAL
Equation 6
CAL
(0.008 =
lo
(0.200 =
hi
units
second
R
( x )
h
R
( )x
Insert the cps lo data point (8 mR/hr for the Model 44-9 example) determined from equations (1) and (2):
h 1
m 60
m 1
( x )
s60
x 2.22 = )
10
6
s
Insert the cps hi data point (200 mR/hr for the Model 44-9 example) determined from equations (3) and (4):
h 1
m 1
( )x
x 55.6 =)
10
6
s
Place the detector in the low field and enter the counts per second:
Equation 7
CORR
lo
SAMPL
=
=
counts
s
Note:
lo
time count
The low field count sample should be 3000 counts. Use the scaler and adjust the count time to accumulate count 3000.
Ludlum Measurements, Inc. Page 8-15 January 2016
Model 2241-3 & 3i Technical Manual Section 8
As an example, assume a 60-second count sample in a low field of 8 mR/hr:
Example
CORR
Equation 8
CORR
Example
CORR
lo
hi
hi
26,427
=
SAMPL
=
5830
=
60
1
Place detector in the high field and enter the counts per second:
=
counts
hi
time count
Counts/second sample in high field of 200 mR/hr:
C/s 440 =
s
C/s 5830 =
Equation 9
=
f
Example
f
Equation 10
a
Example
a
CAL
d
d
( =
CAL
d
hi
x 55.6 =
10
x
CORR
x (55.6 =
10
Insert the values calculated in equations (5), (6), (7), and (8) and solve for fd:
x
CORR
­CORR
6
-
CAL
lo
lohi
x 2.22 x 5830
10
units
=
s
6
440
Solve for ad:
( - )
CAL
x
CORR
x (2.22 - 5830) x
10
= )
hilohihid
x 26.2 =
10
R
6
s
count units
2
s
x 31.1 = 5830) x
10
RCount
266
2
s
Ludlum Measurements, Inc. Page 8-16 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Enter the results of equations (9) and (10) into equation (11) to solve for DT:
Equation 11
Example
f
d
= DT
= DT
=
a
d
x
102.26
x
101.31
s
count
6
2
seconds
5
x =
104.8
or 84 x 10
-6
sec
count
Equation 12
cal
cal
=
CAL
CORR
= CC
f
Example
f
Equation 13
Example
= CC
f
440
x 2.22 =
cal
10
x 2.14
lo
10
( -
CAL
=
6
Solve for f
x
CORR
x (2.22-
10
:
cal
units
x 84 x 440 x
= DT) x
10
lololo
s
R
6666
x 2.14 = )
10
s
Enter the result of equation (12) into:
count
units
and solve for CC:
counts
6
x 206 =
10
R
Enter the CC and DT values (positions 1 and 2 of the FUNCTION switch), derived from the equations above. Perform an “R/hr calibration” as described in the previous subsection in order to ensure that the instrument and detector have been correctly calibrated.
Ludlum Measurements, Inc. Page 8-17 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Model 44-9 Detector Parameter Setup
FUNCTION PARAMETER
1 0084 s-6 2 0206 06 4 N/A 5-8 as desired B-C as desired D-F if applicable
Typical Count Rate and Dead Time for LMI Detectors
DEAD TIME
MODEL & TYPE COUNT RATE in µs (microseconds)
44-6, GM 20 cps/mR/hr 90-110 µs 44-9, GM 55 cps/mR/hr 80-90 µs 44-7, GM 35 cps/mR/hr 240-290 µs 133-2, GM 17.5 cps/mR/hr 40-55 µs 133-4, GM 2 cps/mR/hr 40-55 µs 133-6, GM 0.3 cps/mR/hr 40-55 µs 44-2, Gamma Scint. 2800 cps/mR/hr 8-12 µs* 44-10, Gamma Scint. 15,000 cps/mR/hr 18-20 µs 44-3, Low-Energy Gamma Scint. N/A, operated in Counts/units mode 8-12 µs* 44-21, Beta/Gamma Scint. N/A, operated in Counts/units mode 8-12 µs* 43-5, Alpha Scint. N/A, operated in Counts/units mode 20-28 µs
*The dead time values for these scintillation detectors are due to the dead time of the Model 2241-3 electronics.
Ludlum Measurements, Inc. Page 8-18 January 2016
Note:
The data represented in the table above is typical. Actual values may vary among detector and instrument combinations. This table represents some of the common detectors operated with the Model 2241-3. Consult the LMI sales department for information concerning detectors not listed in the table above.
Model 2241-3 & 3i Technical Manual Section 8
Ludlum Measurements, Inc. Page 8-19 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Ludlum Measurements, Inc. Page 8-20 January 2016
Model 2241-3 & 3i Technical Manual Section 8
Ludlum Measurements, Inc. Page 8-21 January 2016
Model 2241-3 & 3i Technical Manual Section 9
Section
9
Recycling
udlum Measurements, Inc. supports the recycling of the electronic products it produces for the purpose of protecting the environment
L
recycling systems. To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies – public and private – involved in this pursuit, it becomes evident that a myriad of methods can be used in the process of recycling. Therefore, Ludlum Measurements, Inc. does not suggest one particular method over another, but simply desires to inform its consumers of the range of recyclable materials present in its products, so that the user will have flexibility in following all local and federal laws.
The following types of recyclable materials are present in Ludlum Measurements, Inc. electronic products, and should be recycled separately. The list is not all-inclusive, nor does it suggest that all materials are present in each piece of equipment:
and to comply with all regional, national, and international agencies that promote economically and environmentally sustainable
Batteries Glass Aluminum and Stainless Steel
Circuit Boards Plastics Liquid Crystal Display (LCD)
Ludlum Measurements, Inc. products, which have been placed on the market after August 13, 2005, have been labeled with a symbol recognized internationally as the “crossed-out wheelie bin.” This notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding. Each material must be separated. The symbol will be placed near the AC receptacle, except for portable equipment where it will be placed on the battery lid.
The symbol appears as such:
Ludlum Measurements, Inc. Page 9-1 January 2016
Model 2241-3 & 3i Technical Manual Section 10
0
Section
1
Parts List
Reference Description Part Number
Model 2241-3 Survey Meter
Main Circuit Board, Drawing 408 × 285
CAPACITORS
UNIT Completely Assembled Model 2241-3 Survey Meter 48-2864
BOARD Completely Assembled Main Circuit Board 5408-285
C1 0.1µF, 50V 04-5663 C3 0.1µF, 50V 04-5663
C001-C002 47pF, 100V 04-5660
C011 0.001µF, 100V 04-5659 C012 0.1µF, 50V 04-5663
C021 100pF, 3KV 04-5532
C031 0.0047µF, 3KV 04-5547
C032 100pF, 3KV 04-5532
C033 0.0047µF, 3KV 04-5547 C041-C043 0.0047µF, 3KV 04-5547
C101 47uF, 10V 04-5666 C121 47uF, 10V 04-5666 C122-C123 27pF, 100V 04-5658
C131 0.0047µF, 3KV 04-5547 C132-C133 0.1µF, 50V 04-5663 C134 0.01µF, 50V 04-5664 C135 47µF, 10V 04-5666 C136 0.01µF, 50V 04-5664 C137 47µF, 10V 04-5666
C138 100pF, 100V 04-5661
C139 0.001µF, 100V 04-5659
C141 0.0047F, 3KV 04-5547
Ludlum Measurements, Inc. Page 10-1 January 2016
Model 2241-3 & 3i Technical Manual Section 10
Reference Description Part Number
C241 1µF, 35V 04-5656 C242 68µF, 10V 04-5654 C243 0.1µF, 50V 04-5663 C251 68µF, 10V 04-5654
TRANSISTORS
INTEGRATED CIRCUITS
DIODES
POTENTIOMETERS / TRIMMERS
RESISTORS
Q101 2N7002L 05-5840 Q141 MMBT3904LT1 05-5841 Q211 2N7002L 05-5840 Q212 MMBT4403LT1 05-5842 Q241 MJD210 RL 05-5843
U1 MAX810LEUR 06-6424 U001 CD74HC4538M 06-6297 U011 TLC372ID 06-6290 U021 CA3096M; 16=GND 06-6288 U111 N87C51FC 06-6303 U121 LM285MX-2.5 06-6291 U122 X24C02S8T5 06-6299 U131 LM358D 06-6312 U231 LT1073CS8-5 05-5852
CR021 MMBD7000LT1 07-6355 CR031 GI250-2 07-6266 CR041-CR044 GI250-2 07-6266 CR231 CXSH-4 EB33 07-6358 CR241 MMBD914LT1 07-6353 CR242 CXSH-4 EB33 07-6358
R002 10K; 3269X1-103, VOLUME 09-6921 R027 1M; 3269X1-105, HV LIMIT 09-6906
R001 100K, 1/4W, 1% 12-7834 R011-R012 10K, 1/4 W, 1% 12-7839 R013 1K, 1/4W, 1% 12-7832 R014 10K, 1/4W, 1% 12-7839 R015 100K, 1/4W, 1% 12-7834 R021 1M, 1/4W, 5% 10-7028 R022 392K, 1/8W, 1% 12-7841 R023 10K, 1/4W, 1% 12-7839 R024-R025 4.75K, 1/4W, 1% 12-7858
Ludlum Measurements, Inc. Page 10-2 January 2016
Model 2241-3 & 3i Technical Manual Section 10
Reference Description Part Number
R026 8.25K, 1/8W, 1% 12-7838 R031 4.75M, 1/4W, 5% 10-7030 R032 1M, 1/4W, 5% 10-7028 R033-R034 1G, FHV-1, 2% 12-7686 R111-R113 22.1K, 1/4W, 1% 12-7843 R121 100Ohm, 1/4W, 1% 12-7840 R122 6.81K, 1/4W, 1% 12-7857 R131 1M, 1/4W, 1% 12-7844 R132 511K, 1/8W, 1% 12-7896 R133 750K, 1/4W, 1% 12-7882 R134 1M, 1/4W, 1% 12-7844 R135 82.5K, 1/8W, 1% 12-7849 R136 10K, 1/4W, 1% 12-7839 R141 22.1K, 1/4W, 1% 12-7843 R211 2.21K, 1/4W, 1% 12-7835 R231 100Ohm, 1/4W, 1% 12-7840 R241 2.21K, 1/4W, 1% 12-7835 R242 200Ohm, 1/8W, 1% 12-7846
CRYSTALS
INDUCTOR
TRANSFORMERS
MISCELLANEOUS
Calibration Board, Drawing 408 × 98
POTENTIOMETERS
Y221 6.144 MHZ, 2=GND, 3=GND 01-5262
L231 100uH, CTX100-2 21-9740
T101 4275-083, AUDIO 4275-083 T141 L8050 40-0902
P1-P2 1-640456-3, MTA100X13 13-8100 P3 640456-6, MTA100X6 13-8095 P4 640456-2, MTA100X2 13-8073 P5 1-640456-2, MTA100X12 13-8061
BOARD Completely Assembled Calibration Board 5408-113
R115 1M, HV1 SET 09-6814 R116 100K, DISC. 09-6813 R117 1M, HV2 SET 09-6814 R118 1M, HV3 SET 09-6814 R119 1M, HV4 SET 09-6814
Ludlum Measurements, Inc. Page 10-3 January 2016
Model 2241-3 & 3i Technical Manual Section 10
Reference Description Part Number
RESISTORS
CONNECTOR
Display Board, Drawing 408 × 259
CAPACITORS
INTEGRATED CIRCUITS
RESISTORS
MISCELLANEOUS
R110 1M, 1/3W 12-7751 R111 10K, 1/3W 12-7748 R112 1M, 1/3W 12-7751 R113 1K, 1/3W 12-7750 R114 1M, 1/3W 12-7751 R120 1M, 1/3W 12-7751
P7 CONN-640456-7, MTA100 13-8115
BOARD Completely Assembled Display Board 5408-259
C1 27PF, 100V 04-5658
U1 AY0438-I/L 06-6358 U2 AY0438-I/L 06-6358
R001-R004 10.0K, 1%, 125mW 12-7839 R005 392 Ohm, 1%, 1/8 W 12-7054
J1 CONN-640456-8, MTA100 13-8039 DS1 EL-BACKLIGHT-LED 07-6527 DSP1 MAIN DISPLAY; LCD-8246-365-4E1-A/W-REV1 07-6383
Ludlum Measurements, Inc. Page 10-4 January 2016
Model 2241-3 & 3i Technical Manual Section 10
Reference Description Part Number
Switch Board, Drawing 408 × 45
CAPACITORS
INTEGRATED CIRCUITS
SWITCHES
RESISTORS
MISCELLANEOUS
Chassis Wiring Diagram, Drawing 408 × 101
SWITCHES
RESISTOR
CONNECTORS
BOARD Completely Assembled Switch Board 5408-052
C1-C2 4.7µF, 10V 04-5578 C3-C4 10µF, 20V 04-5592 C5 4.7µF, 10V 04-5578 C6 100µF, 10V 04-5576
U1 MAX220EPE 06-6359
S1 350134GSK; FUNCTION; 16 POS 08-6721 S2 LEFT 08-6716 S3 UP 08-6716 S4 ENTER 08-6716
R1-R2 22K 10-7070
P6 CONN-1-640456-3, MTA100 13-8100 P10 CONN-208006-2 13-8451
S1 30-1-PB GRAYHILL 08-6517 S2 PA-1012 08-6540 S3-S4 7101-SYZ-QE C&K 08-6511 S5 30-1-PB GRAYHILL 08-6517 S6 7201KZQE Toggle 08-6749 S7 MPS-103F 08-6699 * Switch Cap 08-6871
R1 10 M, 1/4W, 5% 10-7031
J1 J2 CONN-1-640442-3, MTA100 13-8138 J3 CONN-640442-6, MTA100 13-8171 J4 CONN-640442-2, MTA100 13-8178 J5 CONN-1-640442-2, MTA100 13-8407 J6 CONN-1-640442-3, MTA100 13-8138 J7 CONN-640442-7, MTA100 13-8172 J8 CONN-640442-8, MTA100 13-8184 J9 Series "C" -UG706/U 13-7777 J10 JACK-09-9011-1-4193 18-9080 P10 Handle Pin 7408-055
Ludlum Measurements, Inc. Page 10-5 January 2016
Model 2241-3 & 3i Technical Manual Section 10
Reference Description Part Number
AUDIO
BATTERY
MISCELLANEOUS
DS1 UNIMORPH 21-9251
B1-B2 ˝D˝ Duracell Battery 21-9313
* Model 2241 Switch Board Add On 4408-053 * Model 2241 Digital Bezel Assembly 4408-051 * Portable Battery Contact Set 2001-042 * Model 2241-3 Main Harness 8408-121 * Portable Harness Can Wires 8363-462 * Switch Board Harness 8408-027 * Portable Can Assembly (MTA) 4363-441 * Can Gasket 22-9773 * Portable Knob 08-6613 * Assembled Battery Lid 2363-191 * Portable Battery Gasket 7363-183 * Portable Calibration Cover with Screws 9363-200 * Model 2241-3 Flat Handle Assembly without Clip 4408-179 * Cable-C 1 meter (39 in.) only 40-1004 * Model 2241-3 Casting 7408-185
Ludlum Measurements, Inc. Page 10-6 January 2016
Model 2241-3 & 3i Technical Manual Section 11
Section
11
Drawings
Main Circuit Board, Drawings 408 × 285 (3 sheets)
Main Circuit Board Component Layout, Drawing 408 × 286A
Calibration Board, Drawing 408 × 98
Calibration Board Component Layout, Drawings 408 × 99 (2 sheets)
Display Board, Drawing 408 × 259
Display Board Component Layout, Drawings 408 × 260 (2 sheets)
Switch Board, Drawing 408 × 45
Switch Board Component Layout, Drawing 408 × 46
Wiring Diagram, Drawing 408 ×101
Wiring Diagram for Model 2241-3i, Drawing 408 × 263
Ludlum Measurements, Inc. Page 11-1 January 2016
Model 2241-3 & 3i Technical Manual Appendix A
A
Appendix
RS-232 Output Formats
The Ludlum Model 2241 series of instruments has an RS-232 serial communications port that can be used to log readings and read or set instrument parameters. There are two formats available. Most Model 2241 instruments have the binary format outlined below, but some newer Model 2241-2 units have an ASCII output, which is also outlined below following the binary format explanation.
The RS-232 port is configured at 9600 baud, 8 data bits, no parity, and 1 stop bit (9600,8,N,1). Ludlum Measurements can supply a Windows-based software that can be used to help calibrate the instruments, but note that it will not communicate with the newer ASCII output Model 2241-2 units.
Binary Output Format (15 Bytes)
BYTE01 RatemeterCPS+0 MSB BYTE02 RatemeterCPS+1 BYTE03 RatemeterCPS+2 BYTE04 RatemeterCPE+3 LSB BYTE05 Scaler+0 MSB BYTE06 Sclaer+1 BYTE07 Scaler+2 BYTE08 Scaler+3 BYTE09 Scaler+4 LSB BYTE10 CountTime+0 MSB BYTE11 CountTime+1 LSB BYTE12 CountTimeLeft+0 MSB BYTE13 CountTimeLeft+1 LSB BYTE14 Carriage Return (0DH) BYTE15 Line Feed (0AH)
The ratemeter value is in cps and is scaled by a factor of 256. To get the ratemeter reading in cpm, take the value and divide by 256, and then multiply by 60.
Ludlum Measurements, Inc. Page A-1 January 2016
Model 2241-3 & 3i Technical Manual Appendix A
RS-232 Commands
E – auto dump off
A – auto dump on
C – start scaler
F – set scaler count time
R – send parameters from instrument to computer
S – read parameters from computer to instrument
O – output once
All commands must be sent in upper case laters.
Output of “R” Command – Read Parameters
BYTE01 Detector+0 BYTE02 DeadConstant+0 MSB BYTE03 DeadConstant+1 LSB BYTE04 CaConstant+0 MSB BYTE05 CalConstant+1 BYTE06 CalConstant+2 BYTE07 CalConstant+3 BYTE08 CalConstant+4 BYTE09 CalConstant+5 LSB BYTE10 RateAlarm+0 MSB BYTE11 RateAlarm+1 BYTE12 RateAlarm+2 BYTE13 RateAlarm+3 BYTE14 RateAlarm+4 LSB
BYTE15 ScalerAlarm+0 MSB BYTE16 ScalerAlarm+1 BYTE17 ScalerAlarm+2 BYTE18 ScalerAlarm+3 BYTE19 ScalerAlarm+4 LSB BYTE20 CountTime+0 MSB BYTE21 CountTime+1 LSB BYTE22 Units+0 BYTE23 TimeBase+0 BYTE24 AudioDivideBy+0 BYTE25 Response+0
BYTE26 RateAlert+0 MSB BYTE27 RateAlert+1 BYTE28 RateAlert+2 BYTE29 RateAlert+3 BYTE30 RateAlert+4 LSB BYTE31 CheckSource+0 MSB BYTE32 CheckSource+1 BYTE33 CheckSource+2
Ludlum Measurements, Inc. Page A-2 January 2016
Model 2241-3 & 3i Technical Manual Appendix A
BYTE34 CheckSource+3 BYTE35 CheckSource+4 LSB BYTE36 PercentCS+0 BYTE37 MinDisplay+0 BYTE38 Carriage Return (0DH) BYTE39 Line Feed (0AH)
Input of “S” Command – Send Parameters
BYTE1 DeadCosntant+0 MSB BYTE2 DeadConstant+1 LSB BYTE3 CalConstant+0 MSB BYTE4 CalConstant+1 BYTE5 CalConstant+2 BYTE6 CalCosntant+3 BYTE7 CalConstant+4 BYTE8 CalConstant+5 LSB BYTE9 RateAlarm+0 MSB BYTE10 RateAlarm+1 BYTE11 RateAlarm+2 BYTE12 RateAlarm+3 BYTE13 RateAlarm+4 LSB
BYTE14 ScalerAlarm+0 MSB BYTE15 ScalerAlarm+1 BYTE16 ScalerAlarm+2 BYTE17 ScalerAlarm+3 BYTE18 ScalerAlarm+4 LSB BYTE19 CountTime+0 MSB BYTE20 CountTime+1 LSB BYTE21 Units+0 BYTE22 TimeBase+0 BYTE23 AudioDivdeBy+0 BYTE24 Response+0
BYTE25 RateAlert+0 MSB BYTE26 RateAlert+1 BYTE27 RateAlert+2 BYTE28 RateAlert+3 BYTE29 RateAlert+4 LSB BYTE30 CheckSource+0 MSB BYTE31 CheckSource+1 BYTE32 CheckSource+2 BYTE33 CheckSource+3 BYTE34 CheckSource+4 LSB BYTE35 PercentCS+0 BYTE36 MinDisplay+0
Ludlum Measurements, Inc. Page A-3 January 2016
Model 2241-3 & 3i Technical Manual Appendix A
Input of “F” Command – Set Count Time
BYTE1 CountTime+0 MSB BYTE2 CountTime+1 LSB
Units 0 = R 1 = Sv 2 = cpm
Timebase 0 = min 1 = seconds
AudioDivide 0 = Auto 1 = Manual
ASCII Output F ormat
This special firmware changes the format of the auto-dump from binary to ASCII. The output interval remains the same (2 seconds).
2241 Special Firmware 40804n10 08/06/2003 2241-2 Special Firmware 40806n13 04/04/2003 2241-3 Special Firmware 40806n16 07/07/2004
The old format was (binary): Byte 1 Ratemeter CPS + 0
Byte 2 Ratemeter CPS + 1 Byte 3 Ratemeter CPS + 2 Byte 4 Ratemeter CPS + 3 Byte 5 Scaler + 0 Byte 6 Scaler + 1 Byte 7 Scaler + 2 Byte 8 Scaler + 3 Byte 9 Scaler + 4 Byte 10 Count Time + 0 Byte 11 Count Time + 1 Byte 12 Count Time Left + 0 Byte 13 Count Time Left + 1 Byte 14 Carriage Return (0DH) Byte 15 Line Feed (0AH)
and is now replaced with (ASCII): byte 1 x Ratemeter
byte 2 x Raetemter byte 3 x Ratemeter byte 4 x Ratemeter byte 5 x Ratemeter byte 6 y Display units byte 7 Carriage Return (0DH)
Ludlum Measurements, Inc. Page A-4 January 2016
Model 2241-3 & 3i Technical Manual Appendix A
byte 8 Line Feed (0AH)
The ratemeter is displayed as 5 ASCII digits with a decimal, if necessary, and matches the LCD display on the 2241-2. The display mode is a value from 0 to 9 representing the display units.
CPS 0 KCPS 1 CPM 2 KCPM 3 µR 4 mR 5 R 6 µSv 7 mSv 8 Sv 9
Note: The LMI Model 2241 calibration software is not compatible with this firmware version.
Example output:
02.991 (=2.99 kcps)
01.801 (=1.80 kcps)
01.221 (=1.22 kcps)
00.831 (=0.83 kcps)
00.581 (=0.58 kcps) 004160 (=416 cps)
Ludlum Measurements, Inc. Page A-5 January 2016
Loading...