Ludlum Measurements 177 User Manual

Page 1
LUDLUM MODEL 177 ALARM RATEMETER
October 2018
Serial Number 322945 and Succeeding
Serial Numbers
Page 2
LUDLUM MODEL 177 ALARM RATEMETER
Serial Number 322945 and Succeeding
October 2018
Serial Numbers
Page 3
Model 177 Technical Manual
STATEMENT OF WARRANTY
Ludlum Measurements, Inc. warrants the products covered in this manual to be free of defects due to workmanship, material, and design for a period of twelve months from the date of delivery. The calibration of a product is warranted to be within its specified accuracy limits at the time of shipment. In the event of instrument failure, notify Ludlum Measurements to determine if repair, recalibration, or replacement is required.
This warranty excludes the replacement of photomultiplier tubes, G-M and proportional tubes, and scintillation crystals which are broken due to excessive physical abuse or used for purposes other than intended.
There are no warranties, express or implied, including without limitation any implied warranty of merchantability or fitness, which extend beyond the description of the face there of. If the product does not perform as warranted herein, purchaser’s sole remedy shall be repair or replacement, at the option of Ludlum Measurements. In no event will Ludlum Measurements be liable for damages, lost revenue, lost wages, or any other incidental or consequential damages, arising from the purchase, use, or inability to
use product.
RETURN OF GOODS TO MANUFACTURER
If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information. Your cooperation will expedite the return of your equipment.
LUDLUM MEASUREMENTS, INC.
ATTN: REPAIR DEPARTMENT 501 OAK STREET SWEETWATER, TX 79556
800-622-0828 325-235-5494
FAX 325-235-4672
Page 4
Page 5
Page 6
Model 177 Technical Manual
Table of Contents
Introduction 1
Getting Started 2
Unpacking and Repacking 2-1
Preparing the Instrument for Use 2-1
Operating the Instrument 2-1
Specifications 3
Description of Controls and Functions 4
Front Panel 4-1
Back Panel 4-2
CAL Control 4-3
Internal Controls (Overhaul Only) 4-4
Safety Considerations 5
Environmental Conditions for Normal Use 5-1
Cleaning Instructions and Precautions 5-1
Warning Markings and Symbols 5-2
Electrical Safety Precautions 5-2
Calibration and Maintenance 6
Calibration 6-1
Establishing an Operating Point 6-2
Maintenance 6-3
Troubleshooting 7
Troubleshooting Electronics that Utilize a GM or Scintillation Detector 7-1
Troubleshooting GM Detectors 7-2
Troubleshooting Scintillators 7-3
Technical Theory of Operation 8
Amplifier 8-1
Page 7
Model 177 Technical Manual
Discriminator 8-1
Digital Analog Conversion 8-1
Alarm 8-2
Reset 8-2
Audio 8-2
Time Constant 8-2
High Voltage (HV) 8-3
Low Voltage 8-3
Battery Charge 8-3
High Voltage Test 8-3
Alarm Set Voltage 8-3
Battery Test Voltage 8-4
Recycling 9
Parts List 10
Model 177 Alarm Ratemeter 10-1
Main Board, Drawing 347 × 496 10-1
Calibration Board, Drawing 347 × 132 10-4
Wiring Diagram, Drawing 347 × 492 10-5
Drawings and Diagrams 11
Page 8
Model 177 Technical Manual Section 1
Section
1
T
Introduction
he Ludlum Model 177 Alarm Ratemeter may be used with GM (Geiger-Mueller) or scintillation detectors for contamination monitoring, surveying and area monitoring. The unit provides four
ranges (in decades) of the analog meter, enabling measurement from 0 to 500,000 counts per minute (CPM) on the standard meter dial; others are available. Detector high voltage is adjustable from 400 to 1500 volts.
The unit incorporates an adjustable alarm set point. The alarm setting may be checked by depressing the front-panel TEST switch. Audible and visual enunciators are triggered when the meter reading rises above the alarm set point. Accessory outputs include: Unbuffered Output, Supply Voltage, Negative Pulse Output, Recorder, and Alarm Sink for Remote Relay. The unit may be operated from an internal rechargeable battery or by external 9 Vdc wall transformer.
Ludlum Measurements, Inc. Page 1-1 October 2018
Page 9
Model 177 Technical Manual Section 2
Section
2
T
Getting Started
he Ludlum Model 177 Alarm Ratemeter is designed for use with
GM (Geiger Mueller) or scintillation detectors, which operate from
400 to 1500 V. Typical applications include contamination
monitoring, surveying, and area monitoring.

Unpacking and Repacking

Remove the calibration certificate and place it in a secure location. Remove the instrument and ensure that all of the items listed on the packing list are in the carton. Check individual item serial numbers and ensure calibration certificates match between instruments and detectors (if applicable). The Model 177 serial number is located on the back panel of the instrument.
To return an instrument for repair or calibration, provide sufficient packing material to prevent damage during shipment.
Every returned instrument must be accompanied by an Instrument Return Form, which can be downloaded from the Ludlum website at
www.ludlums.com. Find the form by clicking the “Support” tab and
selecting “Repair and Calibration” from the drop-down menu. Then choose the appropriate Repair and Calibration division where you will find a link to the form.

Preparing the Instrument for Use

Turn the power switch to the ON position. Depress the BAT TEST button. Check that the meter reads above the BAT TEST indication. If the battery does not check, the instrument will operate on 9 Vdc wall transformer only. The battery may be trickle charged from 9 Vdc wall transformer.
Ludlum Measurements, Inc. Page 2-1 October 2018
Page 10
Model 177 Technical Manual Section 2
Select the operating voltage. This voltage is set by the manufacturer for the detector shipped with the instrument and recorded on the Certificate of Calibration. For other detector/instrument setups, consult the detector manual or manufacturer. Then adjust the HV potentiometer accordingly.
Note:
Most GM detectors will operate at 900 volts. However, some smaller GM tubes operate at lower voltages.

Operating the Instrument

Connect a detector to the instrument. Obtain a meter reading from a check source or calibrated source, if available. Remove the source.
Set the instrument to the appropriate range with the RANGE selector switch.
If the alarm point is not already set, press the ALARM TEST switch and adjust
ALARM SET for the desired alarm point.
Note:
The meter displays the alarm set point when the ALARM TEST switch is depressed. Recheck the set point after locking the
ALARM SET control.
Increase the meter count to exceed the alarm threshold. Both the alarm lamp and audible alarm signal should activate.
Depress the RESET button. The meter needle should drive to zero and the alarm circuit should de-energize, shutting off both the visual and audible alarms.
Depress the HV TEST button and ensure that the high voltage is properly set.
Proceed with use.
Ludlum Measurements, Inc. Page 2-2 October 2018
Page 11
Model 177 Technical Manual Section 3
Section
3
Specifications
0
Power: 95-250 Vac wall transformer and 6 volt gell-cell (sealed lead-acid)
battery; typical battery life of approximately 50 hours in a non-alarming condition with a fully charged battery.
Fuse: internal resettable
Response Time: toggle switch control selects FAST (2.2 seconds) or
SLOW (22 seconds) response, for 90% of full-scale reading
Linearity: within 5% of full scale; typically ±2% of full scale reading when
measured with an electronic pulse generator
Battery Dependence: Meter readings vary less than 3% within battery
check limits.
High Voltage: variable from 400 to 1500 V
Input Sensitivity: adjustable from -10 through -100 mV
Connector: Series "C"
Audio: unimorph speaker with volume control located on the front panel
Meter: 1 mA, size 6.4 x 6.4 cm (2.5 x 2.5 in.), DC movement
Meter Scale: 0-500 CPM; 0-1.5 kV; BAT TEST
Ranges: four ranges of X1 through X1K
Recorder Output: adjustable from 0 to 1.25 V at 1 mA
Alarm Output: current sink to 200 mA DC, open circuit voltage not to
exceed 50 Vdc
Unbuffered Output: may be used to externally add to or subtract from the
meter reading
Ludlum Measurements, Inc. Page 3-1 October 2018
Page 12
Model 177 Technical Manual Section 3
Alarm Range: adjustable from 0 through 150% of full scale
Alarm Output: visual indicator (lamp), audible tone and remote current
sink
Alarm Control: factory set to latching, non-latching alarm available
through the removal of main board resistor R116
Finish: powder coat paint
Size: 12.7 x 20.3 x 15.2 cm (5 x 8 x 6 in.) (H x W x D), excluding handle
Weight: 1.9 kg (4.2 lb) with battery
Ludlum Measurements, Inc. Page 3-2 October 2018
Page 13
Model 177 Technical Manual Section 4
Section
4
Description of Controls and Functions

Front Panel

Power ON-OFF Switch: provides 9 Vdc through a wall transformer to the
instrument and trickle-charges the standby battery. In case of line power failure, the battery automatically comes on-line to power the instrument. The battery will provide up to 50 hours of operation.
Note:
To recharge the battery, the ON-OFF switch must be in the ON position.
Power-on Lamp: a red lamp that comes on when power is supplied to the
instrument.
VOLUME Control: varies the volume of the audio output through the
unimorph speaker. This control has minimal effect on the audio when the alarm is activated.
Audio Speaker: a unimorph speaker, located behind the front panel.
ALARM Lamp: a red lamp that comes on when the alarm threshold has
been exceeded. The lamp will remain on (unless the alarm is configured to “non-latching”) until the reset button is depressed, driving the meter needle below the alarm threshold.
Ludlum Measurements, Inc. Page 4-1 October 2018
RANGE Selector Switch: a four-position switch providing range
multipliers of X1K, X100, X10, and X1. With a scale (meter face) of 0-500 CPM, the full range of the instrument is 0 to 500,000 CPM .
Ratemeter: a four-decade linear meter with ranges of 0-500, 0-5000,
0-50,000, 0-500,000 CPM. Other meter faces are available depending on the application. Readout is on a 6.4 cm (2.5 in.) scale panel meter. A separate scale is provided for battery check and high voltage readout.
Page 14
Model 177 Technical Manual Section 4
Connector: Series "C" connector (Series BNC and MHV connectors are
also available). The connector is provided on the front of the instrument for connection to a detector.
RESET Button: This button, when depressed, provides a rapid means of
driving the meter needle to zero.
FAST-SLOW RESPONSE Toggle Switch: When in the FAST position, this
switch provides 90% of full-scale meter deflection in 2.2 seconds. With this switch in the SLOW position, 90% of full-scale meter deflection takes 22 seconds. If quick needle response and maximum deviation are desired, the FAST position should be used. For slow response and damped meter movement, the SLOW position should be used.
BAT TEST Button: When this button is depressed, the meter displays the
battery status. A sufficiently charged battery is indicated when the meter needle is on or within the BAT TEST range.
HV TEST Button: When this button is depressed, the meter displays the
detector high voltage.
ALARM TEST Button: When this button is depressed, the meter displays
the alarm calibration set point.
ALARM SET: used to adjust the alarm calibration set point. Note the
locking knob below the control.

Back Panel

9 VDC INPUT: Input for external 9 Vdc.
Data: a 9-pin type “D” data plug with connections as follows:
PIN 1: Battery terminal. This is a direct connection and does
not go through the front panel ON-OFF switch. Use to parallel the battery.
PIN 2: Unregulated supply from approximately 6 volts, battery
only at 8.7 volts with wall transformer plugged in. Limit current drain to 50 milliamperes.
Ludlum Measurements, Inc. Page 4-2 October 2018
Page 15
Model 177 Technical Manual Section 4
PIN 3: instrument common (ground).
PIN 4: alarm sink. The open collector of a 2N7002L. Limits
sink current to 200 milliamperes with open circuit voltage limited to a range of 0 to +50 volts. Unit conducts when in alarm.
PIN 5: Pulse out. A negative pulse connected to the
discriminator output through a 0.001 µF capacitor, typically -4.0 volts.
PIN 6: Unbuffered output ties directly to the meter drive
circuit. (R124/C122). Approximately 1.3 volts at full scale. Using an external constant current sink will allow background subtract. At full scale, draws out approximately 3.3 microamperes to zero the meter.
PIN 7: Recorder output, adjustable from 0 to 1.0 volts at
1 milliamperes.
PIN 8 and PIN 9: Spares

CAL Control

Remove the calibration (cal) cover plate to access the following calibration potentiometers:
DISCR: Discrimination Control. For pancake-type GM detectors (i.e.
Model 44-9), adjust DISCR for 80 ±10 millivolts. All other GM detectors and scintillators should be set to 35 ±10 millivolts. If a lower operating voltage for a scintillator is desired, adjust the discriminator to 10 ± 2 millivolts. This control has an adjustable range of 10 to 100 millivolts. A Ludlum Model 500 Pulser may be used to determine the discrimination level.
Calibration Controls: X1K through X1 calibration controls used to
calibrate ranges.
HV ADJ: used to set detector operating voltage.
RCDR: used to calibrate the recorder output.
Ludlum Measurements, Inc. Page 4-3 October 2018
Page 16
Model 177 Technical Manual Section 4

Internal Controls (Overhaul Only)

The following controls are located internally, on the main circuit board:
BAT C:
BAT T: used to adjust meter test voltage reading to 5.97 volts at the BAT
OK
HV T: used to adjust the high-voltage test reading to correspond with the
used to adjust charge voltage to 6.825 volts.
line.
actual high-voltage output.
Ludlum Measurements, Inc. Page 4-4 October 2018
Page 17
Model 177 Technical Manual Section 5
Section
5
Safety Considerations

Environmental Conditions for Normal Use

Indoor use only
No maximum altitude
Temperature range of -20 to 50 °C (-4 to 122 °F)
Maximum relative humidity of less then 95% (non-condensing)
Wall transformer supply voltage range of 95-250 Vac, 50/60 Hz single phase (less than 100 mA)
Maximum transient voltage of 1500 Vac
Installation Category II (Overvoltage Category as defined by IEC 1010-1)
Pollution Degree 2 (as defined by IEC 664) (Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected.)

Cleaning Instructions and Precautions

The Model 177 Alarm Ratemeter may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning:
1. Turn the instrument OFF and disconnect the instrument power
cord.
Ludlum Measurements, Inc. Page 5-1 October 2018
2. Allow the instrument to sit for one minute before cleaning.
Page 18
Model 177 Technical Manual Section 5

Warning Markings and Symbols

Caution!
The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc.
Caution!
Verify instrument voltage input rating before connecting to a power converter. If the wrong power converter is used, the instrument and/or power converter could be damaged.

Electrical Safety Precautions

Warning!
Failure to comply with the following instructions could result in a hazardous situation, which, if mishandled, could result in death or serious personal injury.

When installing the unit:

Do not expose the unit to rain or an environment where it may be
splashed by water or other liquids, as doing so may result in fire or electric shock.
Use the unit only with the voltage specified on the unit. Using a
voltage higher than that which is specified may result in fire or electric shock.
Ludlum Measurements, Inc. Page 5-2 October 2018
Page 19
Model 177 Technical Manual Section 5
Do not cut, kink, otherwise damager nor modify the power supply
cord. In addition, avoid using the power cord in close proximity to heaters, and never place heavy objects – including the unit itself – on the power cord, as doing so may result in fire or electric shock.
Avoid installing or mounting the unit or its power supply in
unstable locations, such as a rickety table or slanted surface. Doing so may result in the unit falling down and causing personal injury and/or property damage.

The Model 177 Alarm Ratemeter is marked with the following symbols:

CAUTION (per ISO 3864, No. B.3.1) – designates hazardous live voltage and risk of electric shock. During normal use, internal components are hazardous live. This instrument must be isolated or disconnected from the hazardous live voltage before accessing the internal components. This symbol appears on the front panel. Note the following precautions:
Warning!
The operator is strongly cautioned to take the following precautions to avoid contact with internal hazardous live parts that are accessible using a tool:
1. Turn the instrument power OFF and disconnect the power cord.
2. Allow the instrument to sit for one minute before accessing internal components.
CAUTION, RISK OF ELECTRIC SHOCK (per ISO 3864, No. B.3.6) – designates a terminal (connector) that allows connection to a voltage exceeding 1 kV. Contact with the subject connector while the instrument is on or shortly after turning off may result in electric shock. This symbol appears on the front panel.
The “crossed-out wheelie bin” symbol notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding; each material must be separated. The symbol is placed on the back panel. See section 9, “Recycling,” for further information.
The “CE” mark is used to identify this instrument as being acceptable for use within the European Union.
Ludlum Measurements, Inc. Page 5-3 October 2018
Page 20
Model 177 Technical Manual Section 6
Section
6
Calibration and Maintenance

Calibration

Note:
Local procedures may supersede the following.
Connect the instrument to a Ludlum Model 500 Pulser (Pulse Generator) or equivalent.
The ratemeter may be calibrated by adjusting the calibration controls labeled 1, 10, 100, and 1K. Starting with the 1000 range, apply 400,000 CPM from the pulser. Adjust the 1K calibration control for a meter reading of 400. Drop the pulse rate to 100,000 CPM and ensure a meter reading of 100 ±10.
Repeat this procedure for the lower scales with scaled pulse rates.
For pancake-type GM detectors (i.e. Model 44-9), adjust DISCR for 80 ±10 millivolts. All other GM detectors and scintillators should be set to 35 ±10 millivolts. To lower the scintillation detector operating voltage, decrease the input sensitivity to 10 ±2 millivolts. Adjustment is made by setting the pulse generator amplitude to the desired pulse height. Adjust DISCR until the meter reaches 75% of the generated incoming count rate.
Connect the Model 177 to an external voltmeter. Adjust the rear panel HV control for a reading of 1000 Vdc on the voltmeter. Depress HV TEST. On the main board, adjust HV for a meter reading of 1.0 kV. Using the rear­panel HV control, vary the high-voltage output from 500 to 1500 Vdc and ensure that the high-voltage meter reads within 10% of the Model 177 meter reading.
Adjust RCDR (recorder output) for 1 volt output (equivalent to full scale).
Ludlum Measurements, Inc. Page 6-1 October 2018
Adjust ALARM SET to the desired set point.
Page 21
Model 177 Technical Manual Section 6
Establishing an Operating Point
The operating point for the instrument and detectors is established by setting the detector voltage and instrument sensitivity (HV and DIS). The proper selection of this point is the key to instrument performance.
Efficiency, background sensitivity, and noise are fixed by the physical makeup of the given detector and rarely vary from unit to unit. However, the selection of the operating point makes a significant difference in the contribution of these three sources of count.
The purpose of setting the operating point is to establish the system gain so that the desirable signal pulses (including background) are above the discrimination level, and the unwanted pulses from noise are below the discrimination level. The pulses above the discrimination level are counted by the instrument, while those below are not.
The total system gain is controlled by adjusting the instrument gain or the high voltage. Voltage affects the output of the detector. Amplifier gain is controlled by the DIS (discriminator) control.
In special cases of GM detectors, a minimum voltage must be applied to establish the Geiger-Mueller characteristic. Further changes in gain will not affect this type of detector.
The operating point for each detector is set at a compromise point between sensitivity, stability, and background contribution. These operating points are best for general monitoring. In application, these arbitrarily selected points may not be a better operating point. The following guidelines are presented:
GM Detectors: The output pulse height of the GM detector is not
proportional to the energy of the detected radiation. Adjusting DIS will have minimal effect on the observed count rate unless the setting is so low that the instrument double pulses.
For most GM detectors, set DIS for 30-40 millivolts and adjust HV to the GM tube recommended high voltage. Most GM detectors operate at 900 volts; however, some miniature detectors operate at 400-600 volts. If a recommended setting is unavailable, run a plateau of HV setting vs. count rate. Then set the high voltage on the low side of “center.”
Ludlum Measurements, Inc. Page 6-2 October 2018
Page 22
Model 177 Technical Manual Section 6
Scintillators: Set DIS for 10 millivolts. Carefully increase HV until the
instrument plateaus on the background count. This provides the most stable operating point for the detector.

Maintenance

Instrument maintenance consists of keeping the instrument clean and periodically checking the battery and calibration.
An instrument operational check should be performed prior to each use by exposing the detector to a known source and confirming the proper reading on each scale.
Recalibration should be accomplished after any maintenance or adjustment has been performed on the instrument. Ludlum Measurements recommends recalibration at intervals no greater than one year. Local regulations may have precedence over this recommendation.
To maintain the life of the battery, it is recommended that the instrument be constantly connected to external power with the power switch in the ON position, even when the instrument is not in use. This will keep the internal battery fully charged.
When the instrument is used without external power, adequate charge time must be allowed for the internal battery to recharge. If possible, leave the instrument on with line power applied overnight and weekends. At a minimum, allow one hour of charge time for each hour of use. If the battery is inadvertently allowed to fully discharge, and is left in that state, constant charging for 500 hours (3 weeks) may be required for battery recovery.
Note:
The ON-OFF switch must be in the ON position to charge the batteries. If the unit is out of service for extended periods of time, charge the battery every six months.
It is recommended that the internal GEL-CELL battery be replaced every four years.
Ludlum Measurements, Inc. Page 6-3 October 2018
Page 23
Model 177 Technical Manual Section 7
Section
7
O
Troubleshooting
ccasionally, you may encounter problems with your LMI instrument or detector that may be repaired or resolved in the field, saving turn-around time and expense in returning the
instrument to us for repair. Toward that end, LMI electronics technicians offer the following tips for troubleshooting the most common problems. Where several steps are given, perform them in order until the problem is corrected. Keep in mind that with this instrument, the most common problems encountered are detector cables and sticky meters.
Note that the first troubleshooting tip is for determining whether the problem is with the electronics or with the detector. A Ludlum Model 500 Pulser is invaluable at this point because of its ability to simultaneously check high voltage, input sensitivity or threshold, and the electronics for proper counting.
We hope these tips will prove to be helpful. As always, please call if you encounter difficulty in resolving a problem or if you have any questions.

Troubleshooting Electronics that Utilize a GM or Scintillation Detector

SYMPTOM
No power (or meter does not reach BAT
TEST
or BAT OK
mark)
POSSIBLE SOLUTION
1. Check battery and charge if necessary.
2. Check for loose or broken wires,
especially between the main board and the calibration board.
Ludlum Measurements, Inc. Page 7-1 October 2018
Page 24
Model 177 Technical Manual Section 7
SYMPTOM
Non-linear Readings
Meter goes full-scale or “pegs out”
POSSIBLE SOLUTION
1. Check the high voltage (HV) by
pressing the HV TEST button. If a Multimeter is used to check the HV, ensure that one with high impedance is used; as a standard multimeter could be damaged in this process.
2. Check for noise in the detector cable
by disconnecting the detector and placing the instrument on the lowest range setting. Wiggle the cable and observe the reading for significant changes.
3. Check for “sticky” meter movement.
Does the reading change when you tap the meter? Does the meter needle “stick” at any spot?
4. Check the “meter zero.” Turn the
power OFF. The meter should come to rest on “0.”
1. Replace the detector cable to see if it
has failed, causing excess noise.
2. Check the HV and, if possible, the
input threshold for proper setting.
3. Check for loose wires, especially
between the main board and the calibration board.

Troubleshooting GM Detectors

1. If the tube has a thin mica window, check for window breakage. If damage is evident, the tube must be replaced.
Ludlum Measurements, Inc. Page 7-2 October 2018
Page 25
Model 177 Technical Manual Section 7
2. Check the HV. For most GM tubes, the voltage is normally 900 Vdc, or 460-550 Vdc for “peanut” tubes (Ludlum Model 133 series).
3. If the input sensitivity is too low, the user could see some double-pulsing. See Page 4-3, “DISCR” for further information on sensitivity/discrimination control.
4. Wires to the tube may be broken or the crimped connector could have a loose wire.

Troubleshooting Scintillators

1. Alpha or alpha/beta scintillators are prone to light leaks. They
can be tested for this problem in a dark room or with a bright light. If a light leak is determined, changing the aluminized polyester window assembly will usually fix the problem.
Note:
When replacing the window, make sure to use a window made with the same thickness of aluminized polyester and the same number of layers as the original window.
2. Verify that the HV and input sensitivity are correct. Alpha and
gamma scintillators typically operate from 10-35 mV. High voltage varies with photomultiplier tubes (PMT), from as low as 600 Vdc to as high as 1400 Vdc.
3. On a gamma scintillator, visually inspect the crystal for
breakage or humidity leakage. Water inside the crystal will turn it yellow and gradually degrade performance.
4. Check the PMT to see if the photocathode still exists. If the
end of the PMT is clear (not brownish), this indicates a loss of vacuum, which will render the PMT useless.
Ludlum Measurements, Inc. Page 7-3 October 2018
Page 26
Model 177 Technical Manual Section 8
Section
8
Technical Theory of Operation

Amplifier

Negative detector pulses are coupled through C124 to emitter follower Pin U121. R127 protects the input from inadvertent high-voltage shorts. R129 couples the detector to the high-voltage supply.
Negative pulses from emitter, Pin 2 of U121, are coupled through C121 to amplifier Pin 5 through Pin 7 of U121. This amplifier is self-biased and provides gain in proportion to R029 divided by R0210. Transistor (pins 4, 5, 6, U121) provides amplification. Pins 12 and 15 of U121 are coupled as a current mirror to provide a load for Pin 6 of U121. The output self-biases to 2 Vbe (approximately 1.4 volts) at Pin 7 of U121. This provides just enough bias current through Pin 6 of U121 to conduct all of the current from the current mirror.
Positive pulses from Pin 7 of U121 are coupled to the discriminator.

Discriminator

Comparator U021 provides discrimination. The discriminator is set by the DIS (Discriminator) control located on the rear panel, coupled to Pin 5 of U021. Negative pulses (approximately 5 volts) at Pin 7 of U021 are coupled to Pin 5 of U011 for meter drive and Pin 11 of U011 for audio.

Digital Analog Conversion

Pin 7 of U021 is connected to the dual univibrator, U011. For each low pulse for Pin 7 of U021, Pin 6 of U011 goes high. The pulse of Pin 6 of U011 is typically 5.0 volts for 6 milliseconds on X1 to 6 microseconds on X1K. This pulse is connected to the constant current drive U012. The pulse width control (R3-C2 on calibration board) is utilized for calibration adjustment. Controls R4 through R6 allow calibration on other scales.
For each positive pulse connected to Pin 8 of U012, a constant current pulse is sourced at Pin 15 of U012. This current pulse charges C122, which is
Ludlum Measurements, Inc. Page 8-1 October 2018
Page 27
Model 177 Technical Manual Section 8
discharged by R124. The average voltage on C122 is coupled through HV,
BAT, and ALARM TEST switch to voltage-follower Pin 5 of U311. Pin 7 of
U311 drives the meter and recorder output.

Time Constant

The meter time constant is determined by R124 and C122. For a slower time constant, C122 is paralleled by C101. When C101 is not used, it is connected to Pin 7 of U311 (voltage follower), maintaining the same voltage level as C122. This allows C101 to be switched in or out of the circuit without transients.

Alarm

An alarm is provided by U021, Pins 1, 2, and 3. The alarm set control biases the op-amp U021 for a low output. When the meter signal at Pin 3 exceeds the bias of Pin 2, the output at Pin 1 goes high. Q102 and Q103 saturate, allowing supply voltage to be coupled to:
Lamp voltage through R004.
Audio oscillator U16 through CR112.
R116 couples back to base of Q102, locking up the Alarm On.
Through CR113 to audio transformer T211, allowing full voltage for full volume.
Through R111 to saturating current sink Q101 for external use.

Reset

Reset is provided by coupling a voltage to the base of transistors U012 pins 1, 2, 3, and 4, 5, 6. Both transistors saturate. One discharges C122, causing the meter to zero. Pin 3 U012 turns Q102 off, allowing the alarm to reset.

Audio

A high on Pin 4 U111 turns the oscillator on, saturating Q111 with each positive swing of the oscillator. T211 couples the pulses to the unimorph. Audio volume is controlled by voltage, applied to Pin 2 of T211. This is either 4.3 volts from the alarm circuit or 0 to 4.3 volts from external volume control through emitter follower Q104.
Ludlum Measurements, Inc. Page 8-2 October 2018
Page 28
Model 177 Technical Manual Section 8
For counting, audio pulse width is set by R113/C111 of U011 with one pulse per count. For an alarm condition, Pin 4 U111 is held high through CR112 until alarm is reset. Alarm audio tone is controlled by R117 and C112.

High Voltage (HV)

The high-voltage power supply is a blocking oscillator utilizing Q401-T411 and quadrupler CR123, CR421, CR422, and through CR423. The HV output is controlled by conduction to ground through Q302. With Q302 saturated, the HV output is maximum.
The op amp, U311 Pins 1, 2, 3, is used as a comparator to compare the voltage reference at Pin 3 to the feedback voltage at Pin 2 through R322 for voltage control and regulation. High voltage is adjusted by HV control R311 changing bias on Pin 2 U311. With the HV control wiper at ground, high­voltage output is maximum.

Low Voltage

Low voltage is supplied by internal battery B1 (wiring diagram, 347 × 126) or external power T1. Unregulated power at C125 is coupled to voltage regulator VR211 and battery charger U201-Q301.
Regulated low voltage is supplied to the balance of the circuit through VR131 at 5.0 volts and U301 at 1.2 volts.

Battery Charge

Battery charge is provided by voltage regulator U201 and power transistor Q301. R402 limits charge current for discharged battery. A negative voltage coefficient of -0.0063 volts per degree F is provided by ratio of R013/R201. R013 sets output voltage to 6.825 volts.

High Voltage Test

High voltage test is supplied by R001 through HV TEST switch, BAT TEST switch, ALARM TEST, Pin 5 of U311, and then the meter. The HV readout is calibrated by R001.

Alarm Set Voltage

Alarm set voltage is coupled from ALARM SET control through the ALARM
TEST
switch, voltage follower Pin 5 of U311, and to the meter.
Ludlum Measurements, Inc. Page 8-3 October 2018
Page 29
Model 177 Technical Manual Section 8

Battery Test Voltage

Battery test voltage is controlled by R002 through BAT TEST switch, ALARM
switch, then voltage follower Pin 5 of U311 to the meter.
TEST
Ludlum Measurements, Inc. Page 8-4 October 2018
Page 30
Model 177 Technical Manual Section 9
Section
9
L
Recycling
udlum Measurements, Inc. supports the recycling of the electronics products it produces for the purpose of protecting the environment and to comply with all regional, national, and international agencies that promote economically and environmentally sustainable
recycling systems. To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies – public and private – involved in this pursuit, it becomes evident that a myriad of methods can be used in the process of recycling. Therefore, Ludlum Measurements, Inc. does not suggest one particular method over another, but simply desires to inform its consumers of the range of recyclable materials present in its products, so that the user will have flexibility in following all local and federal laws.
The following types of recyclable materials are present in Ludlum Measurements, Inc. electronics products, and should be recycled separately. The list is not all-inclusive, nor does it suggest that all materials are present in each piece of equipment:
Batteries Glass Aluminum and Stainless Steel
Circuit Boards Plastics Liquid Crystal Display (LCD)
Ludlum Measurements, Inc. products that have been placed on the market after August 13, 2005 have been labeled with a symbol recognized internationally as the “crossed-out wheelie bin,” which notifies the consumer that the product is not to be mixed with unsorted municipal waste when discarding. Each material must be separated. The symbol will be placed on the back panel, except for portable equipment where it will be placed on the battery lid.
The symbol appears as such:
Ludlum Measurements, Inc. Page 9-1 October 2018
Page 31
Model 177 Technical Manual Section 10
Section
10
Model Ratemeter
Main Board 347

CAPACITORS

Parts List
Reference Description Part Number
177 Alarm
, Drawing
× 496

UNIT Completely Assembled Model 177 Alarm Ratemeter 48-1632

BOARD Completely Assembled Circuit Board 5347-502
C001 10µF, 25V, 04-5655 C011 47PF, 100V 04-5660 C012 68µF, 10V 04-5654 C021 0.1µF, 50V 04-5663 C022 0.001µF, 100V 04-5659 C023-C024 0.1µF, 50V 04-5663 C025 10PF, 100V 04-5673 C101 22µF, 20V 04-5672 C111 0.022µF, 50V 04-5667 C112 470PF, 100V 04-5668 C121 0.001µF, 100V 04-5659 C122 2.2µF, 20V 04-5671 C123 1µF, 35V 04-5656 C124 100PF, 3KV 04-5532 C125 3300µF, 35V 04-5675 C211 47µF, 10V 04-5666 C221 0.0056µF, 3KV 04-5522 C301-C302 47µF, 10V 04-5666 C311-C312 0.01µF, 50V 04-5664 C321 100PF, 3KV 04-5532 C322 0.0056µF, 3KV 04-5522 C323 0.0047µF, 3KV 04-5547
Ludlum Measurements, Inc. Page 10-1 October 2018
Page 32
Model 177 Technical Manual Section 10

TRANSISTORS

VOLTAGE REGULATOR

INTEGRATED CIRCUITS

DIODES

THERMISTOR

POTENTIOMETERS / TRIMMERS

Reference Description Part Number
C401-C402 1µF, 35V 04-5656 C403 0.1µF, 50V 04-5663 C411 47µF, 10V 04-5666 C421-C423 0.0047UF, 3KV 04-5547
Q101 2N7002LT1G 05-5840 Q102 MMBT3904T1G 05-5841 Q103 MJD210 05-5843 Q104 MMBT3904T1G 05-5841 Q111 MMBT3904T1G 05-5841 Q301 MJD200 05-5844 Q302 MMBT3904T1G 05-5841 Q401 MJD210 05-5843
VR211 LM2931AT 5.0 05-5813
U011 CD74HC4538M96 06-6297 U012 HFA3096BZ96 06-6468 U021 TLC372IDR 06-6290 U111 ICM7555CBA 06-6300 U121 HFA3096BZ96 06-6468 U201 ICL7663SCBA 06-6302 U301 LM285DR-1-2 05-5845 U311 TLC27M7IDR 06-6292
CR111-CR114 MMBD914LT1G 07-6353 CR201-CR202 CXSH-4LF 07-6358 CR321 1N4007 07-6274 CR401 MMBD914LT1G 07-6353 CR421-CR423 1N4007 07-6274
RT314 250K, 250mW 07-6366
R001 1M, HV TEST 09-6906 R002 50K, BAT T 09-6920 R013 50K, BAT C 09-6920
Ludlum Measurements, Inc. Page 10-2 October 2018
Page 33
Model 177 Technical Manual Section 10

RESISTORS

Reference Description Part Number
R003 475K, 1%, 250mW 12-7859 R004 150 ohm, 1%, 250mW 12-7062 R011 4.75K, 1%, 250mW 12-7858 R012 82.5K, 1%, 250mW 12-7849 R014-R015 100K, 1%, 250mW 12 7834 R016 100 ohm, 1%, 250mW 12-7840 R021-R022 10.0K, 1%, 250mW 12-7839 R023 100K, 1%, 250mW 12-7834 R024 1.00K, 1%, 250mW 12-7832 R025 10.0K, 1%, 250mW 12-7839 R026-R027 100K, 1%, 250mW 12-7834 R028 10.0K, 1%, 250mW 12-7839 R029 221K, 1%, 250mW 12-7845 R0210 5.62K, 1%, 250mW 12-7871 R1 10.0K, 1%, 250mW 12-7839 R101 1.00K, 1%, 250mW 12-7832 R102 10.0K, 1%, 250mW 12-7839 R103 100 ohm, 1%, 250mW 12-7840 R111 1.00K, 1%, 250mW 12-7832 R112 100K, 1%, 250mW 12-7834 R113 1.00M, 1%, 250mW 12-7844 R114 35.7K, 1%, 250mW 12-7911 R115 100 ohm, 1%, 250mW 12-7840 R116 56.2K, 1%, 250mW 12-7873 R117 1.00M, 1%, 250mW 12-7844 R118 3.32K, 1%, 250mW 12-7870 R121 3.92K, 1%, 250mW 12-7875 R122 10.0K, 1%, 250mW 12-7839 R123 100, 1%, 250mW 12-7840 R124 392K, 1%, 250mW 12-7841 R125 47.5K, 1%, 250mW 12-7872 R126 100K, 1%, 250mW 12-7834 R127 10.0K, 1%, 250mW 12-7839 R128 221K, 1%, 250mW 12-7845 R129 1M, 5%, 250mW 10-7028 R201 165K, 1%, 250mW 12-7877 R202 1.00K, 1%, 250mW 12-7832 R203 2.2 ohm, 5%, 250mW 12-7932 R221 1M, 5%, 250mW 10-7028 R301 10.0K, 1%, 250mW 12-7839
Ludlum Measurements, Inc. Page 10-3 October 2018
Page 34
Model 177 Technical Manual Section 10

TRANSFORMERS

MISCELLANEOUS

Calibration Board, Drawing 347

CAPACITORS

POTENTIOMETERS

Reference Description Part Number
R302 1.00K, 1%, 250mW 12-7832 R303 2.21K, 1%, 250mW 12-7835 R311 750K, 1%, 250mW 12-7882 R312 301 ohm, 1%, 250mW 12-7863 R313 475 ohm, 1%, 250mW 12-7851 R315 22.1K, 1%, 250mW 12-7843 R316 1.00M, 1%, 250mW 12-7844 R321-R322 1G 12-7686 R323 1M, 5%, 250mW 10-7028 R401 200 ohm, 1%, 250mW 12-7846 R402 15 ohm, 5%, 1W 12-7738
T211 4275-083 4275-083 T411 40-0902, 9=GND, 10=GND 40-0902
P1 640445-3 MTA 156x3 13-8125 P2 1-640456-2 MTA 100x12 13-8061 P3 1-640456-0 MTA 100x10 13-8066 10 EA. CLOVERLEAF RECEPT. 011-6809-000-599 18-8771 1 EA. #26 Black Coax Ground 2.5 inch 21-9432
× 132
BOARD Completely Assembled Calibration Board 5347-189
C1 0.0047µF, 100V 04-5570 C2 0.047µF, 100V 04-5565
R1 10K, RECORDER 09-6787 R2 100K, HV 09-6813 R3 1M, X1 09-6814 R4 1M, X10 09-6814 R5 2M, X100 09-6834 R6 250K, X1000 09-6819 R7 10K, DISCRIMIINATOR 09-6787
Ludlum Measurements, Inc. Page 10-4 October 2018
Page 35
Model 177 Technical Manual Section 10

RESISTOR

RESISTOR NETWORK

MISCELLANEOUS

Wiring Diagram, Drawing

SWITCHES

POTENTIOMET ERS

CONNECTORS

Reference Description Part Number
R9 680, 1/3W 12-7885
RN1 10K 12-7720
P6 CONN 1-640457-1 MTA100 13 8397
347 × 492
S1 46206-LR SLIDE 08-6523 S2 PA-1002 08-6543 S3 7101-SYZ-QE 08-6511 S4 HV (#923 SWITCHCRAFT) 08-6518 S5 BAT (#923 SWITCHCRAFT) 08-6518 S6 ALARM (#923 SWITCHCRAFT) 08-6518 S7 RESET (30 1 PB GRAYHILL) 08-6517
R1 10K, VOLUME 09-6753 R2 100K, ALARM SET 09-6795
J1 MAIN BOARD 5347-496,
3 PIN SIP (CONN 640428-3 MTA 156) 13-8124
J2 MAIN BOARD 5347-496,
12 PIN SIP (CONN 1-640441-2) 13-8431
J3 MAIN BOARD 5347-496,
10 PIN SIP (CONN 1-640441-0MTA 100) 13-8197
J4 CONNECT BOARD 5347-489 13-7801
5 PIN SIP (CONN-3-640428-5 MTA 156)
J5 CONNECTOR BOARD 5347-489 13-8096
2 PIN SIP (CONN3-643193-2 MTA-156)
J6 CALIBRATION BOARD 13-8161
5347-189, 11 PIN SIP (CONN 1-640441-1 MTA 100)
J7 FLYING LEAD TO MAIN 4478-011 BOARD 5347-293
(RECPT-UG706)
Ludlum Measurements, Inc. Page 10-5 October 2018
Page 36
Model 177 Technical Manual Section 10

AUDIO

BATTERY

MISCELLANEOUS

J8 CONNECTOR BOARD 5347-489 13-8041 (used for modifications only)
J9 CONNECTOR BOARD 5347-489 13-8461 8 PIN SIP (CONN 3-640441-8) MTA-100
Reference Description Part Number
DS2 UNIMORPH (TEC3526PU) 21-9251
B1 6V (PS610 GELL CELL) 21-9385
F1 1 AMP (FUSE #312001 AGC-1) 21-9704 DS1 LAMP-PT 502-OUR-002F-W6 21-9169 DS3 LAMP-PT 502-OUR-002F-W6 21-9169 M1 Model 177 METER 4173-166 * KNOB PKG-50B 08-6601 * KNOB-407D2KI 08-6604 * KNOB-LOCK KL 701 08-6607 * KNOB-70-1-2G 08-6637 * WALL TRANSFORMER 21-8887 PX9025AWPL05 9 VDC 2.5A
Ludlum Measurements, Inc. Page 10-6 October 2018
Page 37
Model 177 Technical Manual Section 11
Section
11
Drawings
MAIN BOARD, Drawing 347 × 496 (4 sheets)
MAIN BOARD Component Layout, Drawing 347 × 503
CALIBRATION BOARD, Drawing 347 × 132
CALIBRATION BOARD Component Layout, Drawing 347 × 133
CONNECTOR BOARD, Drawing 347 x 489
CONNECTOR BOARD Component Layout, Drawing 347 x 490
WIRING DIAGRAM, Drawing 347 × 492
Ludlum Measurements, Inc. Page 11-1 October 2018
Page 38
Page 39
Page 40
Page 41
Page 42
Page 43
Page 44
Page 45
Page 46
Page 47
Loading...