LUDLUM 44-7 Technical Manual

LUDLUM MODEL 44-7 Alpha, Beta, Gamma
Detector
July 2012
Serial No. PR090405
and succeeding Serial Numbers
LUDLUM MODEL 44-7 Alpha, Beta, Gamma
Detector
July 2012
Serial No. PR090405
and succeeding Serial Numbers
STATEMENT OF WARRANTY
Ludlum Measurements, Inc. warrants the products covered in this manual to be free of defects due to workmanship, material, and design for a period of twelve months from the date of delivery. The calibration of a product is warranted to be within its specified accuracy limits at the time of shipment. In the event of instrument failure, notify Ludlum Measurements to determine if repair, recalibration, or replacement is required.
This warranty excludes the replacement of photomultiplier tubes, G-M and proportional tubes, and scintillation crystals which are broken due to excessive physical abuse or used for purposes other than intended.
There are no warranties, express or implied, including without limitation any implied warranty of merchantability or fitness, which extend beyond the description of the face there of. If the product does not perform as warranted herein, purchaser’s sole remedy shall be repair or replacement, at the option of Ludlum Measurements. In no event will Ludlum Measurements be liable for damages, lost revenue, lost wages, or any other incidental or consequential damages, arising from the purchase, use, or inability to product
.
use
RETURN OF GOODS TO MANUFACTURER
If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information. Your cooperation will expedite the return of your equipment.
ATTN: REPAIR DEPARTMENT 501 OAK STREET SWEETWATER, TX 79556
LUDLUM MEASUREMENTS, INC.
800-622-0828 325-235-5494
FAX 325-235-4672
Table of Contents
General 1
Specifications 1
Prelimin ar y Instructions 2
Operating P rocedures 3
Parts List 4
Assembly Drawing 5
Response Curve 6
Model 44-7 Technical Manual
Page 1 July 2012
GENERAL
The Model 44·7 GM (end window) Detector will detect alpha, beta, and gamma radiation. Its configuration allows for easy surveys or leak wipe testing setup. The detector is energy dependent, over-responding by a factor of 6 in the 60-100 keV range when normalized to l37Cs.
The thin mica window is protected by a 79% open, stainless steel screen that can be removed. The GM tube can be easily replaced if necessary.
The G M detector operates between 850 and 1000 V. The tube manufacturer recommends operation at approximately 900 V. The recommended instrument input sensitivity is approximately 30 mV or higher to prevent the detector from double pulsing.
The GM tube face can rupture above 2438 m (8000 ft) altitude pressure. Consequently, detectors carried in unpressurized aircraft above this altitude would be subject to failure.
SPECIFICATIONS
DETECTOR: end window halogen quenched GM WINDOW: 1.7±0.3mg/cm
2
WINDOW AREA: active is 6 cm
mica
2
(0.93 in2); open is 5 cm2 (0.78 in2)
EFFICIENCY(4π geometry): 2% for
14
C; 10% for 90Sr/90Y; 7%
for 99Tc; 7% for
239
Pu; 0.1% for
125
SENSITIVITY: typically 2100 cpm/mR/hr
I
ENERGY RESPONSE: energy dependent BACKGROUND: 40 cpm DEAD TIME: typically 200 microseconds COMPATIBLE INSTRUMENTS: general purpose survey meters,
rate meters, and scalers OPERATING VOLTAGE: 900 V
Model 44-7 Technical Manual
Page 2 July 2012
CONNECTOR: series "C" (others available) CONSTRUCTION: anodized aluminum housing TEMPERATURE RANGE: -15 to 50 °C (5 to 122 °F); may be
certified to operate from -40 to 65 °C (-40 to 150 °F)
SIZE: 4.6 x 14.7 cm (1.8 x 5.8 in.) (Dia x L) WEIGHT: 0.5 kg (1 lb)
PRELIMINARY INSTRUCTIONS
UNPACKING AND REPACKING
Remove calibration certificate or detector functional check certificate and place it in a secure location. Remove the detector and accessories (cable, etc.) and ensure that all of the items listed on the packing list are in the carton. If more than one detector is in the carton, refer to the calibration certificate(s) for serial number (S/N) match. The Model 44-7 S/N is located on the side of the detector near the connector. To return the instrument for repair or calibration, provide sufficient packing material to prevent damage during shipment and appropriate warning labels to ensure careful handling.
The following items and information should also be included to ensure a quick turn-around time on your repair calibration:
instrument(s) and related cable(s)
brief information as to the reason for return
description of service requested
return shipping address
customer name and telephone number
NOTE: When shipping a Model 44-7 by air, it is necessary to ship the tube in a sealed container to avoid sudden atmospheric changes, which could rupture the tube.
Model 44-7 Technical Manual
Page 3 July 2012
OPERATING PROCEDURES
CONNECTING DETECTOR TO INSTRUMENT
Connect one end of the cable provided to the detector by firmly pushing the connectors together while twisting clockwise a quarter of a turn until it latches. Repeat the process in the same manner with the other end of the cable and the instrument.
TESTI N G THE DETECT OR
1. Ensure that the instrument HV is at the proper setting for the
detector.
2. Connect the detector to the instrument and check for a proper
background reading (typically 25 to 50 cpm at a background level of
8 to15 µR/hr.
3. Expose the detector to a check source and verify that the
instrument indicates within 20% of the check source reading obtained during the last calibration. Alternatively, expose the detector to a check source of a known value and verify that the detector gets greater than or equal to the efficiency listed in the specification section of this manual.
4. Instruments that meet these criteria are ready for use. Failure to
meet these criteria may indicate a malfunction in the detector.
Model 44-7 Technical Manual
Page 4 July 2012
PARTS LIST
Model 44-7 Alpha-Beta-Gamma Detector
Ref. No. Description Part No.
UNIT Completely Assembled 47-1536 Model 44-7 alpha-beta-gamma detector 1 ea. Detector Body 7002-065-01 1 ea. End Cap 7002-065-02 1 ea. GM Tube 01-5006 (LND 723, TGM N210) 2 ea. Sponge 7002-065-04 1ea. Socket 01-5025 1 ea. Connector, UG 706/U 13-7751 1 ea. O-Ring 2-128 16-8284 3 ea. Screws, 4-40 x 3/16 FH 17-8811 1 ea. Red Protective Cap 03-5474 1 ea. Cap Model 44-7 Grid with O-Ring 4002-216
Model 44-7 Technical Manual
Page 5 July 2012
Model 44-7 Technical Manual
Page 6 July 2012
Loading...