GENERAL INFORMATION
This section details the general information concerning the standard LTS SCALE DR2100AN
indicator. This indicator was specifically designed for industrial on-board weighing system
applications. The rugged design of this unit will provide reliability and durability required in these
environments.
LTSscale.comPage 2 of 28 REV F – 2018.04.09
Page 5
LTS DR2100A/AN Indicator
STANDARD FEATURES
Standard operating voltage is 12VDC. For trucks having voltages greater than 12
VDC, a DC-DC converter is required.
The versatile DR2100 digital indicator offers the following standard features:
New Smaller Enclosure Size
Rugged Steel Enclosure
Water and Dust Resistant
1.0 Inch LCD Display
2 Line Alphanumeric Display
16 Characters/Line
Back-lit Display
Universal Mounting Bracket
12 VDC Operation
Low Power Indication
Zero Tracking
Push Button On/Off
Auto Shut Off Timer (Optional)
Motion Detection
Push-button Zero
Push-button Tare
Dummy Zero
Push-button Weight Accumulation
Push-button Print
NET/GROSS Switching
Keyboard ID Number Entry
Pounds/Kilograms Switching
Average Part Weight (APW) Entry
Piece Count Capabilities
Function Lockout Capabilities
Custom Printout Setup Capabilities
Keyboard Calibration
Four-Point Linearity Calibration
INDICATOR OPTIONS
The following options are available for the DR2100 indicator:
Custom Software
Time & Date
Remote Push-button Operation
Multiple Setpoint Controls
RS232C Output
Barcode Scanning
Radio Frequency Operation
Ticket Printing
Roll Tape Printing
Barcode Printer
LTSscale.comPage 3 of 28 REV F – 2018.04.09
Page 6
LTS DR2100A/AN Indicator
DISPLAY DESCRIPTION
R
LTS SCALE TWINSBURG, OHIO
DR2100A
0123
45
6
7
89
MENU
ENTER
WEIGH
COUNT
NET
GROSS
SCALE
ZERO
STORE
TARE
APWIDACCUMPRINTDEL
OFFON
The display has two display lines. Each line is capable of displaying 16 alphanumeric characters.
The display is equipped with a backlight.
DR2100 KEYBOARD LAYOUT
Figure: 1
LTSscale.comPage 4 of 28 REV F – 2018.04.09
Page 7
LTS DR2100A/AN Indicator
KEYBOARD CONTROLS
ON
SCALE
ZERO
NET
GROSS
STORE
TARE
WEIGH
COUNT
APW
ID
ACCUM
PRINT
DEL
MENU
ENTER
OFF
This section examines the functions of the various keyboard operations that can be performed on
the DR2100 indicator. There are 21 keys on the keyboard; 10 numeric key (0-9), 10 function keys
plus RESET Key.
Label Keyboard Function
Optional key
Press to zero scale (with no load on the scale and forks off of the floor).
Press to display Net or Gross weight (only displays net when a tare is entered).
Enter the weight of the empty container. Press STORE TARE key.
-Or- pick up empty container press STORE TARE key.
To clear tare: press 0 key. Press STORE TARE key.
Press to toggle between weigh and count modes.
Used to enter the Average Part Weight for piece counting purposes.
Press APW key. Key in the average part weight. Press ENTER Key.
See WEIGH/COUNT key.
To enter an ID number, key in an ID number. Press ID key.
Recall the ID number, press and hold ID key.
To clear the ID number, press 0. Press ID key.
Lift load, press ACCUM key. Weight and count are stored. Remove load.
Lift second load, press ACCUM key. Weights and counts will be added to previous
values. Press and hold ACCUM key to display accumulated weight.
To clear, press 0. Press ACCUM key.
Press PRINT key to send data out the data port.
Press DEL key to backspace one digit when entering a numeric value, also used to
toggle between selections.
Clears erroneous numbers that were keyed in. For lb/kg switching press once and then
press DEL key. Press twice to recall a stored tare weight.
Used to reset the indicator.
To reset, hold down DEL and the R keys, push and release R key, wait 2 seconds to
reset
LTSscale.comPage 5 of 28 REV F – 2018.04.09
Page 8
LTS DR2100A/AN Indicator
OPERATION SEQUENCES
Gross Weighing
Raise empty forks off ground and level
Press SCALE/ZERO key
Press STORE/TARE key (clears previous tare value)
Lower forks and lift load
Gross weight displayed
Net Weighing - Push-button Tare
Raise empty forks off ground and level
Press SCALE/ZERO key
Press STORE/TARE key (clears previous tare value)
Lower forks and lift load
Press STORE/TARE key
Add additional weight to lift
Net weight displayed
Net Weighing - Keyboard Tare
Raise empty forks off ground and level
Press SCALE/ZERO key
Press STORE/TARE key (clears previous tare value)
Lower forks and lift load
Key in tare value
Press STORE/TARE key
Net weight displayed
Piece Counting
Raise empty forks off ground and level
Press SCALE/ZERO
Press STORE/TARE key (clears previous tare value)
Lower forks and lift load
Press APW key
Key in APW value or accept displayed value
Press APW key
Press WEIGHT/COUNT key
Read count on top line and weight on bottom line
Gross Weight Accumulation
Raise empty forks off ground and level
Press SCALE/ZERO
Press STORE/TARE key (clears previous tare value)
Lower forks and lift load
Read weight
Press ACCUM key (weight value added)
Repeat for additional loads
To view: Press and hold ACCUM key
To Clear: Press 0 key
Press ACCUM key
LTSscale.comPage 6 of 28 REV F – 2018.04.09
Page 9
LTS DR2100A/AN Indicator
Net Weight Accumulation
Raise empty forks off ground and level
Press SCALE/ZERO
Press STORE/TARE key (clears previous tare value)
Key in tare value
Press STORE/TARE key
Lower forks and lift load
Read weight
Press ACCUM key (weight value added)
Repeat for additional loads
To view: Press and hold ACCUM key
To Clear: Press 0 key
Press ACCUM key
Counting Accumulation
Raise empty forks off ground and level
Press SCALE/ZERO
Press STORE/TARE key (clears previous tare value)
Press APW key
Key in APW value or accept displayed value
Press APW key
Press WEIGHT/COUNT key
Key in tare value
Press STORE/TARE key
Lower forks and lift load
Read weight
Press ACCUM key (count value added)
Repeat for additional loads
To view: View displays weight total only
Total data can be sent to printer
To Clear: Press 0 key. Press ACCUM key
Printing
Press PRINT key at any time to send data to printer.
Program setup determines what will be printed.
See Technical Manual for programming data output.
Note: Unit must have optional RS232 option installed.
LTSscale.comPage 7 of 28 REV F – 2018.04.09
Page 10
LTS DR2100A/AN Indicator
DISPLAY MESSAGES
ERROR Messages
DESCRIPTION
SOLUTION
*ERROR*
OUT OF LEVEL
Scale is tipped beyond the
maximum weighing angle.
Tilt mast until forks are level. If off left to
right then move truck to a level position.
1) Bad Load cell cable.
2) Bad Angle sensor.
3) Incorrect Angle settings
OUT OF RANGE
Indicates an overload condition
exists. The weight signal is far
beyond the scale capacity.
If there is weight on the scale it should
be removed immediately.
If no weight is on the scale check for:
Indicates an overload condition.
The Gross weight as measured
from the zero value 105% of scale
capacity.
Weight in question should be
immediately removed from the scale.
If the total weight on the scale is known
to be less than capacity then check
scale calibration.
DR21-190B
Displayed on power-up. Indicates
Model and program number.
A number may have been
entered and not cleared.
Bad Power.
If indicator display does not continue
past this display then:
1) Press DEL then ENTER key
2) Check power supply.
3) Turn truck off and power up if OK
check truck ignition wires and
distributor cap.
EXCEEDS CAPACITY
Tried to enter a tare larger than
the Scales Capacity.
Tare value must be less than the
calibrated weighing capacity of the
scale.
INHIBIT
Remote Input 1 or 2 is closed and
Input function has been
configured to prevent all scale
weighing functions.
Consult factory.
ITEM LOCKED OUT
Displayed if functions have been
disabled (LOCKED OUT).
Function will not operate without
enabling function.
LOW BATTERY
Input power to the DR2100AN
has dropped below 11 VDC.
Correct low voltage condition.
Reset indicator by holding DEL and R
keys
LTSscale.comPage 8 of 28 REV F – 2018.04.09
Page 11
LTS DR2100A/AN Indicator
MAIN PC BOARD LAYOUT
J11 Pin
RS485
1
485+
2
485-
J5 Pin
LOAD CELL
1
+EX
2
+SIG
3
-SIG
4
-EX
5
SHIELD
J6
Power Pin
1
+DC IN, 12 VDC
2
GND
3
GND
4
+DC IN, 12 VDC
J2 Pin
Clino Inputs
1
+5 VDC
2
X-axis
3
Y-axis
4
GND
5
SHIELD
J8 Pin
RS232
1
PA6 (TTL INPUT)
2
RS232 OUT
3
GND
4
RS232 IN
5
+5VDC
NTEP LC Cable– Black Coiled Cord
Pin
Color
Description
1
Red
+ Excitation
2
Green
+ Signal
3
White
- Signal
Shield
Shield
5
Brown
X Axis
6
Black
- Excitation
7
Blue
Y Axis
Figure: 2
Main Board Layout (P# P10014)
Connector Pin Outs - Main Board
LTSscale.comPage 9 of 28 REV F – 2018.04.09
Page 12
LTS DR2100A/AN Indicator
SET-UP CODE SCREENS 1
KEYBOARD 2100
KEYBOARD 2104
KEYBOARD 9080
KEYBOARD 9081
KEYBOARD 9083
LOCK KEYBOARD?
YES = LOCK OUT
KGS ONLY
YES *NO
DEL FACTORY CAL?
YES *NO
REMOTE INPUT 1
PRINT
KEYBOARD TARE
YES *NO
ZERO TRACK
*YES NO
SELECT COUNTBY
NUMBER = 5
ARE YOU SURE ?
YES *NO
REMOTE INPUT 2
ACCUM
I.D.
YES *NO
EXPAN= 0.0
GROSS= 0 LB
SELECT DECIMAL
NUMBER= 0
CAL STEP # 1
WEIGHT = XXXX
SCALE ZERO
YES *NO
SELECT FIXED ZRO
YES *NO
CAL STEP # 2
NUMBER= XXXX
PUSH BUTTON TARE
YES *NO
INCREMENTS
NUMBER = 1000
CAL STEP # 3
WEIGHT = XXXX
PRINT
YES *NO
USE ENTER OR DEL
NUMBER= 5005
CAL STEP # 4
NUMBER= XXXX
ACCUM
YES *NO
LEGAL FOR TRADE
*YES NO
MENU
YES *NO
SELECT MOTION #D
NUMBER = 1
SETUP
YES *NO
SELECT MOT DELAY
SECONDS= 1.0
CALIBRATE
YES *NO
DISPLAY MOTION ?
*YES NO
CONFIG REMOTE IN
YES *NO
SELECT TRACK #D
NUMBER = 1
CONFIG PRINT
YES *NO
SET TIME & DATE
YES *NO
RAW A/D BAT TEST
YES *NO
OFF TIMERS
YES *NO
PIECE COUNT
YES *NO
PROGRAM NAME
YES *NO
****************
LTSscale.comPage 10 of 28 REV F – 2018.04.09
Page 13
LTS DR2100A/AN Indicator
SET-UP CODE SCREENS 2
KEYBOARD 9084
KEYBOARD 9085
KEYBOARD 9086
KEYBOARD 9087
KEYBOARD 9088
SELECT BAUD
9600
SET TIME & DATE
YEAR = XX
BACK LIGHT TIME
MINUTES= 30.0
AVERAGE PIECE WT
LB = 1.0000
# TOP LINE FEEDS
NUMBER = 2
SET TIME & DATE
MONTH = XX
12V BV = XXX
RAW A/D= XXXXXX
OFF TIMERS
MINUTES= 30.0
PRINT ITEM # 1
TIME & DATE
SET TIME & DATE
DAY = XX
FILTER
WEIGHT = 10
# SPACES OVER
NUMBER = 1
SET TIME & DATE
HOURS = XX
# CR LF
NUMBER = 1
SET TIME & DATE
MINUTES= XX
PRINT ITEM # 2
I.D.
KEYBOARD 9089
# SPACES OVER
NUMBER = 1
PROGRAMMING THRU
SERIAL PORT
# CR LF
NUMBER = 1
PRINT ITEM # 3
GROSS
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
PRINT ITEM # 4
TARE
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
PRINT ITEM # 5
NET
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
PRINT ITEM # 6
ACCUM
# SPACES OVER
NUMBER = 1
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
# CR LF
NUMBER = 1
PRINT ITEM # 7
PIECE COUNT
PRINT ITEM # 9
NAME
# SPACES OVER
NUMBER = 1
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
# CR LF
NUMBER = 1
PRINT ITEM # 8
TOTAL COUNT
# BTM LINE FEEDS
NUMBER = 1
LTSscale.comPage 11 of 28 REV F – 2018.04.09
Page 14
LTS DR2100A/AN Indicator
SET-UP CODE SCREENS 3
KEYBOARD 9090
KEYBOARD 9091
KEYBOARD 9092
KEYBOARD 9093
KEYBOARD 9094
SINGLE POINT CAL
YES *NO
LINEARITY STEP 1
OFFSET = 0
GROSS= 0
FB=-0.0 LR=-0.0
ZONE 1 FACTOR
NUMBER=1.0000000
CALIB ANGLE FTRS
YES *NO
ARE YOU SURE ?
YES *NO
LINEARITY STEP 1
ADD-WT = 0
SET ANGLE ZERO
YES *NO
ZONE 2 FACTOR
NUMBER=1.0000000
CALIB ANGLE FTRS
*YES NO
SET SCALE ZERO ?
YES *NO
LINEARITY STEP 1
FACTOR = 400000
FB ANGLE ZERO
NUMBER= 128
ZONE 3 FACTOR
NUMBER=1.0000000
TEST LOAD VALUE
WEIGHT = 1000
CAL MODE
WEIGHT = 0
LINEARITY STEP 2
OFFSET = 0
LR ANGLE ZERO
NUMBER= 128
ZONE 4 FACTOR
NUMBER=1.0000000
CAL MODE
WEIGHT = 1000
LINEARITY STEP 2
ADD-WT = 0
FB ANGLE SPAN
FACTOR= 42667
ZONE 5 FACTOR
NUMBER=1.0000000
LINEARITY STEP 2
FACTOR = 400000
LR ANGLE SPAN
FACTOR= 42667
ZONE 6 FACTOR
NUMBER=1.0000000
REAR UP 1000
FB= 3.0 LR= 0.0
LINEARITY STEP 3
OFFSET = 0
FRONT TO FRONT
YES *NO
ZONE 7 FACTOR
NUMBER=1.0000000
RIGHT UP 1000
FB= 0.0 LR=-3.0
LINEARITY STEP 3
ADD-WT = 0
ZONE 8 FACTOR
NUMBER=1.0000000
FRONT UP 1000
FB=-3.0 LR= 0.0
LINEARITY STEP 3
FACTOR = 400000
ZONE 1 ANGLE SP
NUMBER= 3.50
LEFT UP 1000
FB= 0.0 LR= 3.0
LINEARITY STEP 4
OFFSET = 0
ZONE 2 ANGLE SP
NUMBER= 3.50
LTS SCALE
GROSS= 1000 LB
LINEARITY STEP 4
ADD-WT = 0
ZONE 3 ANGLE SP
NUMBER= 3.50
LINEARITY STEP 4
FACTOR = 400000
ZONE 4 ANGLE SP
NUMBER= 3.50
SINGLE POINT CAL
FACTOR = 100000
ZONE 5 ANGLE SP
NUMBER= 3.50
NOMINAL ZERO
NUMBER= 0
ZONE 6 ANGLE SP
NUMBER= 3.50
SCALE ZERO
NUMBER= 0
ZONE 7 ANGLE SP
NUMBER= 3.50
PRINT
ZONE 8 ANGLE SP
NUMBER= 3.50
LTSscale.comPage 12 of 28 REV F – 2018.04.09
Page 15
LTS DR2100A/AN Indicator
INDICATOR SETUP
Code Listing
Function NTEP
2100
Keyboard Lockout
2104
Expanded Test Mode ( X10 )
2105
Restore Name On Top Line
2169
Load Defaults
6969
Initialize Clock Chip
9080
Setup
9081
Four-Point Calibration
9083
Configure Remote Inputs
9084
Configure Print
9085
Set Time & Date
9086
Raw A/D, Battery Test
9087
Auto Shut Off and Backlight Timer
9088
APW Recall/Store
9089
Program Name Through Serial Port
9090
Single-Point Calibration
9091
View / Store Calibration Factors
9092
View / Set Angle Zero and Span
9093
View / Set Angle Factors
9094
Out-of-Level Calibration
This section provides setup and calibration instructions for the DR2100 indicator. The indicator
setup and calibration is accomplished by entering various configuration codes through the
keyboard.
Configuration Codes
Note: The indicator must be placed in the program mode prior to entering a configuration code.
See the following list of configuration codes that may be accessed. To enter the program mode,
Key 9080 and The display will indicate [ PRESS CAL SWITCH ], Press CAL switch once. This
switch is located on the main board and can be accessed by removing the rear cover of the
indicator.
Note: if display shows “LOCKED OUT” use configuration Code 2100 to unlock specific
function or key to enable viewing or changing.
LTSscale.comPage 13 of 28 REV F – 2018.04.09
Page 16
LTS DR2100A/AN Indicator
Code 2100 - Keyboard Lockout
Available Lockout Functions
Keyboard Tare
Code 9080; Setup
I.D.
Code 9081; Calibration
Scale Zero
Code 9083; Configure Remote Inputs
Push Button Tare
Code 9084; Configure Print
Print
Code 9085; Set Time & Date – Option
Accumulate
Code 9085; Raw A/D, Battery Test
Menu ( lb./kg, Recall Tare )
Code 9087; Auto Shut Off and Backlight Timer
Code 9088; Piece Count/ APW
Code 9089; Program Name
LOCK KEYBOARD?
YES = LOCKOUT
KEYBOARD TARE
*YES NO
ZERO TRACK
*YES NO
EXPAN= 0.0
GROSS= 0 LB
!@#%^&*&^%$#@#@!
GROSS= 0 LB
LTS SCALE
GROSS= 0 LB
Note: Keyboard functions and the 90xx codes may be individually locked out.
Press keys 2100. Press ENTER key.
Top display line shows the item to be enabled or disabled. Bottom
display line shows the current status indicated by the asterisk.
Disabled or lockout = YES
Enabled or accessible = NO
To lock or unlock a function:
Use DEL key to toggle selection between YES and NO.
Press ENTER to accept selection and proceed to the next function.
Note: in the event that the indicator locks up or does not respond properly, hold
down DEL and R keys.
Code 2104 - Expand Test Mode X10
Expand mode is used for testing with a finer graduation size than normally displayed.
Example: if the Graduation size is set for 5 lb. then expanded mode will display 0.5 lb. graduations.
To enter expand mode:
Press keys 2104. Press ENTER.
ZERO TRACK
Disables Zero Tracking in test mode. Re-enabled upon expand mode exit.
Press DEL key to select mode.
Press ENTER key to accept the selected setting.
Top display line shows the expanded weight.
Bottom display line shows the normal weight.
To exit and return to normal operation, simply press ENTER key.
Code 2105 - Restore Default Name
Restore the default name [ LTS SCALE ] to the top display line.
Press numeric keys 2105. Press ENTER key.
The top display line will display the default name in the Gross weigh mode.
LTSscale.comPage 14 of 28 REV F – 2018.04.09
Page 17
LTS DR2100A/AN Indicator
Code 2169- Load Default Settings
PRESS CAL SWITCH
KGS ONLY
YES *NO
CTBY 1 , 2 , 5
NUMBER = 2
SELECT DECIMAL
NUMBER = 0
SELECT FIXED ZRO
YES *NO
CAUTION! This code clears calibration and other setup settings.
Consult factory prior to entering the load default settings mode.
The load default setting code is used to initialize the main board.
Default values loaded may not be the settings required for your application.
Press numeric keys 2169. Press ENTER key.
The indicator display will briefly flash when loading is complete.
Configure indicator to required specifications. Four-Point Calibration must be
done before the Single-Point calibration can be performed.
Code 6969 - Initialize Clock Chip
Initializes the clock on units equipped with the Time & Date option.
See Setup – Code 9085 to set time and date.
To initialize the clock:
Press keys 6969. Press ENTER key.
Units ordered with the Time & Date option are initialized before being shipped.
Code 9080 - Indicator Setup
Note: indicator setup has been performed at the factory. This section is provided in
the event that these settings need to be changed.
Press keys 9080. Press ENTER key.
Press CAL switch at the rear end of board
KGS ONLY
Calibrate and weigh in Kilograms only select YES.
Press DEL key to select mode.
Press ENTER key to accept the selected setting.
If changing mode (LB/KG to KGS ONLY) or (KGS ONLY to LB/KG) the
indicator MUST be reconfigured and Four-Point calibration MUST be done.
COUNTBY
Available counts by values are 1 25
Press DEL key to select Count-By value.
Press ENTER key to accept the selected setting.
DECIMAL POSITION
Available decimal positions are 012 3
Key in decimal positions value.
Press ENTER key to accept setting.
FIXED ZERO
Select YES to count by 10, 20, of 50’s.
Press DEL key to select the number of fixed zeroes.
Press ENTER key to accept setting.
Note: if there are decimals selected, the fixed zeros are not available.
LTSscale.comPage 15 of 28 REV F – 2018.04.09
Page 18
LTS DR2100A/AN Indicator
INCREMENTS
500
600
1000
1200
1500
2000
2500
3000
4000
5000
6000
10000
SELECT INCREMENT
NUMBER = 1000
USE ENTER OR DEL
5005
LEGAL FOR TRADE
*YES NO
SELECT MOTION #D
NUMBER = 1
SELECT MOT DELAY
SECONDS = 1.0
DISPLAY MOTION
*YES NO
DISPLAY TOP LINE
*YES NO
Display shows:
Standard Values for Number of increments are:
Press DEL key to select increments value.
Press ENTER key to accept setting.
Displays shows capacity, plus the count-by factor.
Example:
Count by factor: = 5
Decimal positioning value: = 0
Increments: = 1000
2500 (increments) x 2 (count-by) = 5000
Display shows: 5005
If the capacity and count-by factor are correct then, press ENTER key to accept these
settings. If not press DEL, and return to the beginning.
LEGAL FOR TRADE
Press DEL key to select YES or NO.
Press ENTER key to accept setting.
If set to YES, further calibration will be disabled until the back cover is removed to access
the calibration switch.
SELECT MOTION
Available motion settings are:
Legal-for-Trade = YES: 12 3
Legal-for-Trade = NO: 1 2 3 4 5 6 7 8 9
Press DEL key to select motion setting.
Press ENTER key to accept setting.
MOTION DELAY
Available motion delay settings are:
0.5 1.0 1.5 2.0 2.5 3.0
Press DEL key to select the motion delay setting.
Press ENTER key to accept setting.
Note: an analog filter is available on the main board. Potentiometer (R19) is
located at the bottom left-hand corner of the main board. Adjusting this
potentiometer may require re-calibration after adjustment. Counterclockwise
will slow the update rate while clockwise will speed the rate up.
DISPLAY MOTION
YES Weight is displayed during motion conditions.
NO Weight is only displayed when within the motion band.
Press DEL key to select YES or NO.
Press ENTER key to accept setting.
DISPLAY ON TOP LINE
YES Motion is displayed on top line
NO Motion is not displayed on top line
LTSscale.comPage 16 of 28 REV F – 2018.04.09
Page 19
LTS DR2100A/AN Indicator
ZERO TRACK DIGITS
SELECT TRACK #D
NUMBER = 1
REMOTE INPUT 1
PRINT
REMOTE INPUT 2
ACCUM
SELECT BAUD
9600
# TOP LINE FEEDS
NUMBER = 2
Numbers of graduations for zero tracking are:
.613
Press DEL key to select the motion delay setting.
Press ENTER key to accept setting.
End of Indicator Setup - Code 9080.
Enter another configuration code to review or change parameters or
Code 9081 – Four-Point Calibration
A Four-Point linearity calibration is generally not required after initial installation. Single-Point
calibration should be used. The Four-Point calibration may be required in the event of load cell
replacement or weight loads of increasing value are not linear.
Code 9083 - Configure Remote Inputs
Press keys 9083. Press ENTER key.
REMOTE INPUT 1
Functions available for Remote Input #1 are:
PRINT ACCUMULATE INHIBIT
SCALE ZERO PUSH-BUTTON TARE
Press DEL key to select function for Remote Input #1.
Press ENTER key to accept setting.
REMOTE INPUT 2
Functions available for Remote Input #2 are:
PRINT ACCUMULATE INHIBIT
SCALE ZERO PUSH-BUTTON TARE
Press DEL key to select function for Remote Input #2.
Press ENTER key to accept setting.
Code 9084 - Configure Print
Note: The RS232 data format is 8 data bits, 1 stop bit and No parity.
Data, Stop and Parity are not selectable.
Press keys 9084. Press ENTER key.
SELECT BAUD
Available baud rates are: 120024004800 9600
For IR or RF load cell cables set to 9600.
Press DEL key to select baud rate.
Press ENTER to accept setting.
# TOP LINE FEEDS
Set to print blank lines (feed paper down) prior to printing selected data fields.
Use the numeric keypad to enter the value of top line feeds.
Press ENTER key to accept setting.
LTSscale.comPage 17 of 28 REV F – 2018.04.09
Page 20
LTS DR2100A/AN Indicator
PRINT ITEM # 1
Gross weight
ID (Identification Number)
Piece count
Tare weight
Accumulate
Total count
Net weight
Name
No print
Time & Date (option)
# SPACES OVER
NUMBER = 1
# CR LF
NUMBER = 1
# BTM LINE FEEDS
NUMBER = 1
SET TIME & DATE
YEAR = 3
SET TIME & DATE
MONTH = 1
SET TIME & DATE
DAY = 1
SET TIME & DATE
HOUR = 1
SET TIME & DATE
MINUTES = 1
Available items to be printed are:
Press DEL key to select item.
Press ENTER key to accept setting.
# SPACES OVER
Sets the starting point on the left-hand margin for the selected print field.
Key in the value for the starting point for Print Item 1.
Press ENTER key to accept setting.
# CR LF
Sets the vertical spacing between lines of data.
0 = next line of data will be printed on the same line
1 = data will be printed on the next line
2 = double spacing, one blank line between print field
More line feeds results in a bigger vertical space between data lines.
Use the numeric keypad to enter the value.
Press ENTER key to accept setting.
PRINT ITEM #1 to PRINT ITEM #9 must be configured to print any of the ten print items in
any order desired. Repeat Print Item, # spaces and # CR LF for all nine Print Items before
next prompt is displayed.
# BTM LINE FEEDS
Set to print blank lines (feed paper down) after printing selected data fields.
Use the numeric keypad to enter the required value.
Press ENTER key to accept setting.
Code 9085 - Setting Time & Date (Optional)
Note: this indicator utilizes the 24-hour clock (military time).
Press keys 9085. Press ENTER key.
YEAR
1 or 2 digit year entry (0 to 99).
Use the numeric keypad to enter the Year.
Press ENTER to accept setting.
MONTH
Use the numeric keypad to enter the Month.
Press ENTER to accept setting.
DAY
Use the numeric keypad to enter the Day.
Press ENTER to accept setting.
HOURS
Use the numeric keypad to enter the Hour.
Press ENTER to accept setting.
MINUTES
Use the numeric keypad to enter the Minutes.
Press ENTER to accept setting.
LTSscale.comPage 18 of 28 REV F – 2018.04.09
Page 21
LTS DR2100A/AN Indicator
Code 9086 – Raw A/D, Battery Test
12V BV = XXX
RAW A/D= XXXXXXX
AVERAGE PIECE WT
LB. = X.XX
PROGRAMMING THRU
SERIAL PORT
BACK LIGHT TIME
MINUTES= 30.0
OFF TIMERS
MINUTES= 30.0
Troubleshooting tool used to view the voltage value and the raw A/D counts.
Press keys 9086. Press ENTER.
Press ENTER to exit.
Press ENTER key to exit.
Display lower line shows raw A/D counts. Reading should be positive and increase
as weight is applied. Maximum is approximately 400,000 counts
Code 9087 – Auto Shut Off and Backlight Timer, Filter Weight
Press keys 9087. Press ENTER key.
BACK LIGHT TIME
Use numeric keypad to enter the time value.
Press ENTER key to accept value.
0.1 minutes = 6 seconds. Jumper JP5 must be set for backlight to turn off.
OFF TIMER
Use numeric keypad to enter the time value.
Press ENTER key to accept value.
If keyboard ON/OFF is used unit will turn off automatically when time expires.
Note: Any motion outside the motion band will reset the timers.
This keeps the scale from turning off during usage.
FILTER WEIGHT
This sets the allowed band of the ABSOLUTE weights that can be used to generate a
CALCULATED weight. Default is set at 10. The DR calculates a ROLLING AVERAGE of
consecutive ABSOLUTE weight observations. This number entered for this option is the
amount of variation from a mean weight that is allowed, both up or
down. Example, selecting 5 would mean the acceptable band is +/- 5
lbs. If any of these observations falls outside the selected range, the
calculation is thrown out and the DR continues seeking an allowed rolling average until it has
a sufficient number of consecutive observations within the selected band.
Use numeric keypad to enter the time value.
Press ENTER key to accept value.
Code 9088 - Entry Of Average Part Weight
Press keys 9088. Press ENTER.
Use the numeric keys to enter an Average Part Weight (APW).
Press ENTER key to accept the displayed value.
Code 9089 - Program Name Through Serial Port
Allows a 16-character alpha-numeric string to enter for a custom name.
This name is displayed on the top line of the display when in the Gross weigh mode. This
header can also be selected to be printed, see Code 9084.
You need the RS232 option installed, data cable, and a PC running a terminal emulation program.
Configure PC COMM port to match baud rate selected in Code 9084 (default 9600 baud), 8 data
bits, NO parity, 1 stop bit.
Press keys 9089. Press ENTER.
LTSscale.comPage 19 of 28 REV F – 2018.04.09
Page 22
LTS DR2100A/AN Indicator
Display on PC will show:
[ LTS SCALE ...CHANGE NAME ? Y/N]
To change: Send capitol “Y” from the PC to DR2100A.
Indicator sends:
[ENTER ALL 16 CHARACTERS ]
[---------------- ]
Send all 16 characters. Fill to end with spaces if needed.
[YOUR NAME HERE ...CHANGE NAME ? Y/N]
If name is correct, then press any key to end, if not return to step I).
NOTE: To exit this mode from the keyboard: Press and hold DEL and R keys,
LTSscale.comPage 20 of 28 REV F – 2018.04.09
Page 23
LTS DR2100A/AN Indicator
Code 9090 – Single-Point Calibration
Single-Point Calibration can only be done after the Four-Point calibration has been completed at least
one time. If the display shows ERROR when this code is entered then the Four-Point calibration must
be completed first.
See Calibration section.
Code 9091 - View / Store Calibration Factors
View / Store Calibration Factors
This is not a substitute for test weight calibration.
Recorded Calibration factors can be entered to put the indicator back to a known
working condition. The Calibrations Factors stored will change after every Single-
Point and Four-Point Calibration. Recording of these factors is the last thing to be
done before putting the scale in service.
See Calibration section.
Code 9092 - View / Set Angle Zero and Span NTEP
See Out-of-Level Calibration.
View and set Angle Zero and Angle Span values.
Code 9093 – View / Set Angle Factors
See Calibration section.
Code 9094 – Out-of-Level Calibration
See Calibration section.
LTSscale.comPage 21 of 28 REV F – 2018.04.09
Page 24
LTS DR2100A/AN Indicator
CALIBRATION
SINGLE POINT CAL
YES *NO
ARE YOU SURE ?
YES *NO
SET SCALE ZERO ?
YES *NO
CAL MODE
WEIGHT = 0
Note: A Four-Point Linearity Calibration (code 9081) has been performed at the factory. In most cases,
the Single-Point Calibration (code 9090) is the only calibration that will be required after installation.
Code 9090 – Single-Point Calibration
Single-Point Calibration should be used to make minor calibration adjustments.
Four-Point Linearity Calibration should be done if the load cell is being replaced.
Allow ample time for the indicator to warm up, generally 15-20 minutes.
1.1. Unload scale, ensure that they are not touching the floor and are level.
Press SCALE ZERO key.
1.2. Remove rear cover plate from the indicator.
1.3. Enable calibration by pressing the cal button, located in the upper right hand corner of
the main board. This button is usually white.
1.4. Press keys 9090. Press ENTER key.
1.5. SINGLE POINT CAL
Press DEL to select YES. Press ENTER key.
1.6. ARE YOU SURE?
Press DEL key to select YES. Press ENTER key.
1.7. SET SCALE ZERO?
Press DEL key to select YES. Press ENTER key.
Note: If emptying the scale is impossible select NO otherwise reset
the scale zero by selecting YES.
1.8. WEIGHT =
Apply test load (for best results apply at least 25% of scale capacity).
Wait until motion ceases.
Enter the value of the applied test load. Press ENTER key.
Single-Point Calibration is finished.
LTSscale.comPage 22 of 28 REV F – 2018.04.09
Page 25
LTS DR2100A/AN Indicator
Code 9081 - Four-Point Linearity Calibration
DEL FACTORY CAL
YES *NO
ARE YOU SURE?
YES *NO
CAL STEP # 1
WEIGHT = XXXX
CAL STEP # 2
WEIGHT = XXXX
CAL STEP # 3
WEIGHT = XXXX
CAL STEP # 4
WEIGHT = XXXXX
A Four-Point Linearity Calibration is generally not required after initial installation. Single-Point
Calibration should be used. The Four-Point Calibration may be required in the event of load cell
replacement or weight loads of increasing value are not linear.
1.9. Unload scale, ensure that they are not touching the floor and are level.
Press SCALE ZERO key.
1.10. Remove rear cover plate from the indicator
1.11. Press keys 9081. Press ENTER key.
Note: Pressing ENTER will use current value on the scale as zero value.
1.12. DEL FACTORY CAL
Press DEL key to select YES. Press ENTER key.
1.13. ARE YOUR SURE?
Press DEL key to select YES. Press ENTER key.
1.14. CALIBRATIONSTEP#1 - Apply first test load.
For best results apply approximately 25% of scale capacity.
Wait until motion ceases.
Key in the value of the applied test load. Press ENTER key.
1.15. CALIBRATIONSTEP#2 - Apply second test load.
For best results apply approximately 50% of scale capacity.
Wait until all motion ceases.
If the reading is correct, press ENTER key.
If not, key in the value of the test load. Press ENTER key.
1.16. CALIBRATIONSTEP#3 - Apply third test load.
For best results apply approximately 75% of scale capacity.
Wait until all motion ceases.
If the reading is correct, press ENTER key
If not, key in the value of the test load. Press ENTER key.
1.17. CALIBRATIONSTEP#4
Apply fourth test load.
For best results apply approximately 100% of scale capacity.
Wait until all motion ceases.
If the reading is correct, press ENTER key.
If not, key in the value of the test load. Press ENTER key.
Calibration Complete
This completes the Four-Point Linearity Calibration.
LTSscale.comPage 23 of 28 REV F – 2018.04.09
Page 26
LTS DR2100A/AN Indicator
Out-of-Level Calibration
GROSS = 0
FB= X.X LR= X.X
GROSS = 0
FB=+0.0 LR= +0.0
SET ANGLE ZERO
YES *NO
FB ANGLE ZERO
NUMBER = XXX
LR ANGLE ZERO
NUMBER = XXX
FB ANGLE SPAN
FACTOR = XXX
LR ANGLE SPAN
FACTOR = XXX
FRONT TO FRONT
YES *NO
CALIB ANGLE FTRS
YES *NO
TEST LOAD VALUE
WEIGHT = XXXXX
This procedure is to calibrate the DR2100AN indicator with a Clinometer for Out-of-Level weighing. The
scale base and DR2100AN indicator should be installed per the installation instructions prior to performing
this procedure. Out-of-Level calibration should be done after installation and before unit is placed into
service.
Warning: Once this procedure has been started it MUST be completed in all 4 directions. Before
starting, read through this procedure completely. Make sure you understand the
procedure and can tip the mast 3 degrees in all directions.
1.18. Apply power to the indicator. Allow 15-20 minutes for warm up.
1.19. Remove the cover from the back of the indicator.
1.20. Press keys 9092. Press ENTER key.
Top display line shows the gross weight.
Bottom display line shows: FB (front-to-back) angle
LR (left-to-right) angle.
No load should on the scale at this time.
Note: The clinometer is mounted on top of the scale base.
All references to the position of the truck while calibrating
are made while sitting in the driver seat.
1.21. Level truck so both FB and LR angles are within 0.0° +/- 0.2°.
Level is obtained by tilting forks for Front-to-Back
And moving to a flat location or tip truck for Left-to-Right.
Move truck, (level condition) until: FB = +0.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
1.22. Press DEL key.
Note: DO NOTselect “YES”. Angle zero has been set at the factory.Do not change any of the numbers in the following steps unless instructed.
1.23. Press ENTER key.
1.24. Press ENTER key.
1.25. Press ENTER key.
1.26. Press ENTER key.
1.27. Press ENTER key.
1.28. Press ENTER key.
1.29. Press keys 9094. Press ENTER key.
1.30. Press DEL key to place the asterisk next to YES.
Press ENTER key.
1.31. Key in the value of the test weight to be used.
Press ENTER key.
LTSscale.comPage 24 of 28 REV F – 2018.04.09
Page 27
LTS DR2100A/AN Indicator
RIGHT UP XXXX
FB=+0.0 LR=-3.0
LEFT UP XXXX
FB=+0.0 LR=+3.0
REAR UP 0
FB=+0.0 LR= 0.0
REAR UP XXXX
FB=+3.0 LR=+0.0
FRONT UP XXXX
FB=+0.0 LR=-3.0
Note: All references to position are made while sitting in the driver seat.
CAUTION: Care must be taken while applying weights in an out-of-level
condition. Bodily harm may occur from shifting weight.
Top display line shows the position the truck is to be placed in.
Bottom display line shows: FB (front-to-back) angle
LR (left-to-right) angle
Press SCALE ZERO key if weight value is not zero.
1.32. Lift test load (value entered above for TEST LOAD VALUE).
Tilt forks forward 3 degrees and tip truck so level left-to-right.
FB = +3.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
Weigh value shown is the non-corrected weight.
Wait for motion no motion. Press ENTER key.
1.33. Tilt forks until level front-to-back and tip truck so right side up 3 degrees.
FB = +0.0° (+/- 0.2°)
LR = -3.0° (+/- 0.2°)
Wait until all motion ceases. Press ENTER key.
1.34. Tilt forks backward 3 degrees and tip truck so level left-to-right.
FB = -3.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
Wait until all motion ceases. Press ENTER key.
1.35. Tilt forks until level front-to-back and tip truck so left side up 3 degrees.
FB = +0.0° (+/- 0.2°)
LR = +3.0° (+/- 0.2°)
Wait until all motion ceases. Press ENTER key.
Out-of-Level calibration is complete.
LTSscale.comPage 25 of 28 REV F – 2018.04.09
Page 28
LTS DR2100A/AN Indicator
Calibration Verification
GROSS = 0
FB=+0.0 LR=+0.0
GROSS = XXXXX
FB=+0.0 LR=+0.0
GROSS = XXXXX
FB=+3.0 LR=+0.0
GROSS = XXXXX
FB=+0.0 LR=-3.0
GROSS = XXXXX
FB=-3.0 LR=-0.0
GROSS = XXXXX
FB=-0.0 LR=+ 3.0
1.36. Press keys 9092.
Press ENTER key.
1.37. Tilt forks front-to-back until level and position truck so left-to-right is level.
FB = +0.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
Press ZERO SCALE key if necessary.
1.38. Lift known test load.
Tilt forks front-to-back until level and position truck so left-to-right is level.
FB = +0.0°(+/- 0.2°
LR = +0.0° (+/- 0.2°)
Reading in tolerance? Repeat Single-Point calibration if necessary.
1.39. Tilt forks forward 3 degrees and tip truck so level left-to-right.
FB = +3.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
Reading in tolerance? Repeat Out-of-Level calibration if necessary.
1.40. Tilt forks until level front-to-back and tip truck so right side up 3 degrees.
FB = +0.0° (+/- 0.2°)
LR = -3.0° (+/- 0.2°)
Reading in tolerance? Repeat Out-of-Level calibration if necessary.
1.41. Tilt forks backward 3 degrees and tip truck so level left-to-right.
FB = -3.0° (+/- 0.2°)
LR = +0.0° (+/- 0.2°)
Reading in tolerance? Repeat Out-of-Level calibration if necessary.
1.42. Tilt forks until level front-to-back and tip truck so left side up 3 degrees.
FB = +0.0° (+/- 0.2°)
LR = +3.0° (+/- 0.2°)
Reading in tolerance? Repeat Out-of-Level procedure if necessary.
LTSscale.comPage 26 of 28 REV F – 2018.04.09
Page 29
LTS DR2100A/AN Indicator
Weight Calibration Numbers Standard and NTEP
1.43. Press 9091 then press ENTER Key.
9091
1.44. Record value.
LINEARITY STEP 1
0FFSET =
1.45. Press ENTER key. Record value.
LINEARITY STEP 1
ADD-WT =
1.46. Press ENTER key. Record value.
LINEARITY STEP 1
FACTOR =
1.47. Press ENTER key. Record value.
LINEARITY STEP 2
0FFSET =
1.48. Press ENTER key. Record value.
LINEARITY STEP 2
ADD-WT =
1.49. Press ENTER key. Record value.
LINEARITY STEP 2
FACTOR =
1.50. Press ENTER key. Record value.
LINEARITY STEP 3
0FFSET =
1.51. Press ENTER key. Record value.
LINEARITY STEP 3
ADD-WT =
1.52. Press ENTER key. Record value.
LINEARITY STEP 3
FACTOR =
1.53. Press ENTER key. Record value.
LINEARITY STEP 4
0FFSET =
1.54. Press ENTER key. Record value.
LINEARITY STEP 4
ADD-WT =
1.55. Press ENTER key. Record value.
LINEARITY STEP 4
FACTOR =
1.56. Press ENTER key. Record value.
SINGLE POINT CAL
FACTOR =
1.57. Press ENTER key. Record value.
SCALE ZERO
NUMBER =
This procedure can only be used after the initial calibration has been performed in the field and all numbers
in this section have been recorded. Entering calibration numbers can be performed in the event that an
electronics problem is encountered. This does not guarantee calibration is correct or within tolerance.
Scale should always be tested using known test weight before placing unit back into service.
This is not a substitute for a periodic calibration utilizing test weights.
1.58. Standard Unit, Recording complete.
NTEP unit, continue to Angle Calibration Numbers, next section.
LTSscale.comPage 27 of 28 REV F – 2018.04.09
Page 30
LTS DR2100A/AN Indicator
Angle Calibration Numbers – DR2100AN only
1.59. Press keys 9092. Press Enter Key.
9092
1.60. Press DEL key.
GROSS = 0
FB=+X.X LR=+X.X
1.61. Press ENTER key.
Note: DO NOT select “YES”. Angle Zero set at the factory.
SET ANGLE ZERO
YES *NO
1.62. Press ENTER key. Record value.
FB ANGLE ZERO
NUMBER =
1.63. Press ENTER key. Record value.
LR ANGLE ZERO
NUMBER =
1.64. Press ENTER key. Record value.
FB ANGLE SPAN
FACTOR =
1.65. Press ENTER key. Record value.
LR ANGLE SPAN
FACTOR =
1.66. Press ENTER key. Record value.
FRONT TO FRONT
*NO YES
1.67. Press 9093 keys. Press ENTER key.
9093
1.68. Record value:
ZONE 1 FACTOR
NUMBER
=1.0000000
1.69. Press ENTER key. Record value.
ZONE 2 FACTOR
NUMBER
=1.0000000
1.70. Press ENTER key. Record value.
ZONE 3 FACTOR
NUMBER
=1.0000000
1.71. Press ENTER key. Record value.
ZONE 4 FCTOR
NUMBER
=1.0000000
1.72. Press ENTER key. Record value.
ZONE 5 FACTOR
NUMBER
=1.0000000
1.73. Press ENTER key. Record value.
ZONE 6 FACTOR
NUMBER
=1.0000000
1.74. Press ENTER key. Record value.
ZONE 7 FACTOR
NUMBER
=1.0000000
1.75. Press ENTER key. Record value.
ZONE 8 FACTOR
NUMBER
=1.0000000
1.76. Press ENTER key. Record value.
ZONE 1 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.77. Press ENTER key. Record value.
ZONE 2 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.78. Press ENTER key. Record value.
ZONE 3 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.79. Press ENTER key. Record value.
ZONE 4 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.80. Press ENTER key. Record value.
ZONE 5 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.81. Press ENTER key. Record value.
ZONE 6 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.82. Press ENTER key. Record value.
ZONE 7 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
1.83. Press ENTER key. Record value.
ZONE 8 ANGLE SP
NUMBER =3.50
CHANGE TO 4.80
LTSscale.comPage 28 of 28 REV F – 2018.04.09