DRU CK & TEM PERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
[email protected] •
• www.druck-temperatur.de
www.LR-Cal.net
MANUAL
Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
36
• Enter the 3. Funcon Menu Level and select „P2". Use the arrow-keys to enter the temperature
value, indicated by the pecision reference thermometer and press „E“-key to confirm. Your
input is confirmed by symbol „*“ on the display, appearing aer about 5 seconds.
• Select „CAL: INT“, set the parameter to „Y“ (yes) and confirm with „E“-key.
The internal calibraon procedure is now running for a couple of seconds. When this procedure
is finished, the temperature calibrator gives an audible „beep“.
10.2 Calibraon of the „EXT“ and „REF“ inputs with a signal calibrator -2I- only for versions
The purpose of the re-calibraon is to correct the EXT and REF inputs error together.„“ „“
To calibrate the two inputs, it is necessary to have a Pt100 calibrator and/or a thermocouples
calibrator depending on what is to be calibrated, e.g. or .LTC 100 LEC 200LR- LR-Cal Cal
C„“ „“alibraon of the EXT input automacally reproduces the same calibraon on the REF input:
• 2. Funcon Menu Level „ “On the , select the type of EXT input to calibrate (Pt100, Tc K, Tc J, Tc N,
„“-TcR, Tc S) following the instrucons in the manual. Press E key to confirm.
• 3. Funcon Menu Level arrow-keys „ “ „ “-keyEnter the and press the to set Cal_chnl = EXT. Press E
to accept.
• Choose two calibraon points depending on the appliance range or the field where one wishes
;to carry out calibraon. (For example 0 and 450°C for PT100 200 and 800°C for thermocouples).
•„“Connect the signal generator to the EXT input, generang the first calibraon value. See the
instrucons for the connecon.
• „ " arrow-keys „ “-Select P1 and press to set the first value (for example 0°C). Press E Key to
„“confirm. Confirmaon is indicated by the symbol * which appears on the display aer about 5
seconds.
• Generate the second calibraon value with the signal generator. See the instrucons for the
connecons.
• „ " arrow-keys „ “-kSelect P2 and press to set the second value (for example 450°C). Press E ey to
„“confirm. Confirmaon is indicated by the symbol * which appears on the display aer about 5
seconds.
•„ “, sto„y“ „“-Select CAL: EXT et es and confirm pressing E Key. The procedure takes a few seconds.
At the end there is a Beep.
10.3 Calibraon of the „REF“ input with a reference probe connected - only for -2I versions
This operaon adapts the value indicated by the REF input to the value indicated by the„“
reference probe connected to it, compensang its errors.
T„“o carry out the calibraon it is necessary to connect the probe to the REF terminals and to have
a precision reference thermometer, e.g. model .LRT 750LR-Cal
•„“Connect the probe to the REF input following the instrucons in the manual.
• a or in the bath of temperature calibratorInsert the probe in suitable hole the .
• precision reference a suitable hole or in the bath of the temperatureInsert the thermometer in
calibrator.
• wait for „ “ .Set the first calibraon point and wait for the appliance to be stable ( symbol ÷ )