Lovibond SensoDirect pH110 Instruction Manual

Page 1
GB
SensoDirect pH110
pH - Instruction Manual
Lovibond® Water Testing
Tintometer® Group
Page 2
2
SensoDirect pH110
CONTENTS
1. GENERAL DESCRIPTION
2. SPECIFICATION
3. FUNCTIONAL DESCRIPTION
4. CALIBRATION 4-1 Overview 4-2 Equipment required 4-3 Two Point Calibration 4-4 Single Point Calibration
5. TAKING MEASUREMENTS
6. CHANGING THE BATTERY
7. ACCESSORIES
1. GENERAL DESCRIPTION
* General purpose pH meter with broad application including AQUARIA & FISH
HATCHERIES, FOOD & BEVERAGE PROCESSING, PHOTOGRAPHY, LABORATORY,
QUALITY CONTROL, EDUCATION, SWIMMING POOLS & WATER CONDITIONING * High quality, compact unit that is designed for easy operation * Water resistant front panel with easy to read LCD display and rubberised function
keys
Unpacking
Please check that the shipment includes the following items:
• Lovibond
®
SensoDirect pH110
• Lovibond
®
SensoDirect pH sensor plastic/gel-type with BNC plug
• 9V Battery
• Protective cover
• 7 Screws
• Screwdriver
• Plastic Cover Calibration Screws
• Instruction Manual
Battery installation
Prior to first use take the instrument out of the protective cover open the battery compartment and insert the 9V battery.
Ensure polarity is correct.
Page 3
3
SensoDirect pH110
Protective Cover
The instrument is equipped as standard with the protective cover. Prior to the first measurement please fix the cover by using the enclosed srews. The
protective armouring ensures reliable operation even in harsh environments.
When the instrument is used without the protective cover please use the enclosed plastic cover to protect the calibration screws against dust and humidity.
2. SPECIFICATION
Display LCD, 21.5 mm (0.7“) digit height Measurement Range 0 - 14 pH Resolution 0.01 pH Accuracy ± 0.07 pH (pH 5 - pH 9)
± 0.1 pH (pH 4 - pH 10)
± 0.2 pH (pH 1 - pH 3.9, pH 10.1 - pH 13) Main instrument only 23 ± 5°C, after calibration Input Impedance 10
12
Ohms
Temperature Compensation No adjustment
Sample Time Approx. 0.4 seconds Operating Temperature 0 - 50°C (32 - 122°F) Operating Humidity Maximum 80% relative humidity Calibration External, pH 4.0 (slope adjustment) & pH 7.0 (Cal.
adjustment)
Power Supply 006P DC 9V battery Power Consumption Approx. 2.0 mA Dimension 208 x 110 x 34 mm (L x W x H) Weight Approx. 380 g pH electrode pH plastic type, gel electrolyte with BNC connector
Page 4
4
SensoDirect pH110
3. FUNCTIONAL DESCRIPTION
Figure 1
1 Display 2 Power ON Button 3 Power OFF Button 4 Data Hold Button 5 BNC socket for pH electrode 6 Calibration pH 7 Adjust 7 Slope pH 4 pH10 Adjust 8 Battery compartment/cover 9 Stand 10 Protective cover
4. CALIBRATION
4-1 Overview The meter is supplied with calibration by mV signal that simulates an ideal pH
electrode mV output (0 mV at pH 7, based on 25°C working temperature). Ho­wever, because most pH electrodes do not exactly replicate the ideal electrode and because the working temperature may differ from 25°C, it is necessary to calibrate the instrument to ensure measurement accuracy of the combined pH electrode.
Page 5
5
SensoDirect pH110
4-2 Equipment Required i) Combination pH electrode ii) pH 4.0 & pH 7.0 buffer solutions
4-3 Two Point Calibration i) Connect the pH electrode to the BNC socket (Figure 1, 5) and immerse
in pH 7.0 buffer solution. ii) Switch the instrument on by pressing the Power ON Button (Figure 1, 2). iii) Adjust „pH 7 Adjust „ (Figure 1, 6) until the display reads exactly pH 7.0. iv) Rinse the electrode in distilled water. v) Place the electrode into buffer pH 4.0 solution. vi) Adjust „pH 4 Adjust „ (Figure 1, 7) until the display reads exactly pH 4.0. vii) Rinse the electrode in distilled water. viii) Repeat steps iii) to vii) at least twice.
4-4 Single Point Calibration i) Connect the pH electrode to the BNC socket and immerse in a standard
buffer solution (eg pH 7.0 or higher if available). ii) Switch the instrument on by pressing the Power ON Button (Figure 1, 2). ii i) Adjust the “pH 7 Adjust „ (Figure 1, 7) until the display reads the same value as the
standard buffer solution used.
5. TAKING MEASUREMENTS
After calibration, the unit is ready for taking measurements. i) Connect the pH electrode to the BNC socket. ii) Switch the instrument on by pressing the Power ON Button. ii i) Immerse the electrode in the solution being measured; the instrument will display the
pH value. iv) During measurement, pressing the Data Hold Button (Figure 1, 4) will freeze the
display. At the same time the display will show a Data Hold indicator. To release the display freeze function, press the Data. Hold Button again; the Data Hold indicator will disappear from the display.
v) After taking a measurement, rinse the electrode in distilled water.
Page 6
6
SensoDirect pH110
6. CHANGING THE BATTERY
i) „ “ in the left corner of the display indicates that it is necessary to replace the battery. However, accurate measurements may still be made for several hours after the Low Battery” indicator first appears.
ii) Remove the protective cover before replacing the battery. To replace the battery, remove the Battery Compartment Cover (Figure 1, 9) on the rear of the meter.
iii) Remove the battery, install a replacement one (006P DC 9V) and replace the battery compartment cover.
7. ACCESSORIES
721330 pH Electrode, plastic type, gel electrolyte with BNC connector 721247 pH Buffer Solution pH 4 (25°C) red, 90 mL, traceable to N.I.S.T. 721248 pH Buffer Solution pH 7 (25°C) red, 90 mL, traceable to N.I.S.T. 721249 pH Buffer Solution pH 10 (25°C) red, 90 mL, traceable to N.I.S.T.
8. IMPORTANT INFORMATION FOR pH ELECTRODES
General information:
Non-refillable gel electrodes are wear parts. The service life and the measurement accuracy depend significantly on application, storage and maintenance. Improper storage as well as special samples with e.g. aggressive chemicals, high contamina­tion potential and high temperatures can reduce the service life to few months or even weeks. Samples with low conductivity (very low salinity) extend the response time of electrodes. Electrodes are subjected to a natural ageing effect, which in­duces shift of the offset point and gradient of the electrode. For slowing down the ageing effect and for maintaining the performance and accuracy, please note the following considerations:
Page 7
7
SensoDirect pH110
NotesStorage:
Never store electrodes in distilled water!
- Short-term storage (few hours): in tap water
- Medium-term storage (some days): in pH 4 buffer solution
- Longer-term storage: in 3 M KCl solution In case of drying out (KCl solution is crystallized), please store the electrode for
minimum 5 hours in 3 M KCl solution (order code 2418609) or tap water in such a way that water covers the glass part completely. Afterwards, the electrode has to be stored in 3 M KCl solution.
Handling of measurement and calibration
1. Prior to each immersion into sample or buffer solution please rinse the electrode with distilled water/tap water for removal of possible residues of salt and to mini­mize carryover effects.
2. In case of air bubbles, remove them by shaking.
3. During measurement/calibration, a flow rate is necessary (water flow/slewing of the electrode).
4. After each use, the electrode shall be rinsed with distilled water/tap water and covered by a destined cap (filled with KCl solution).
Cleaning:
For longer service life, electrodes shall be cleaned as following:
- Water soluble contaminations: distilled water (tap water)
- Chalk and hydroxide deposits: Hydrochloric acid (10 %)
- Fats and oils: Warm water and mild detergents (dishwashing detergent) Attention: Do not use petroleum, ether or similar chemicals for plastic-covered
electrodes.
Page 8
Technical changes without notice Printed in Germany 02/16 No.: 00 38 61 41 Lovibond® and Tintometer® are Trademarks of the Tintometer Group of Companies
Tintometer India Pvt. Ltd.
B-91, A.P.I.E. Sanath Nagar, Hyderabad, 500018 Tel: +91 (0) 40 4647 9911 Toll Free: 1 800 102 3891 indiaoffice@tintometer.com www.lovibondwater.in
India
Tintometer GmbH
Lovibond
®
Water Testing Schleefstraße 8-12 44287 Dortmund Tel.: +49 (0)231/94510-0 Fax: +49 (0)231/94510-30 verkauf@tintometer.de www.lovibond.com
Germany
Tintometer AG
Hauptstraße 2 5212 Hausen AG Tel.: +41 (0)56/4422829 Fax: +41 (0)56/4424121 info@tintometer.ch www.tintometer.ch
Switzerland
The Tintometer Limited Lovibond House / Solar Way Solstice Park / Amesbury, SP4 7SZ Tel.: +44 (0)1980 664800 Fax: +44 (0)1980 625412 water.sales@tintometer.com www.lovibond.com
UK
Tintometer South East Asia Unit B-3-12, BBT One Boulevard, Lebuh Nilam 2, Bandar Bukit Tinggi, Klang, 41200, Selangor D.E Tel.: +60 (0)3 3325 2285/6 Fax: +60 (0)3 3325 2287 lovibond.asia@tintometer.com www.lovibond.com
Malaysia
Tintometer China
Room 1001, China Life Tower 16 Chaoyangmenwai Avenue, Beijing, 100020 Tel.: +86 10 85251111 App. 330 Fax: +86 10 85251001
China
Tintometer Inc
6456 Parkland Drive Sarasota, FL 34243 Tel.: +1 941-756-6410 sales@tintometer.us www.lovibond.us
USA
Tintometer Brazil
Caixa Postal: 271 CEP: 13201-970 Jundiaí – SP ­Tel.: +55 (11) 3230-6410 sales@tintometer.com.br www.lovibond.com.br
Brazil
Loading...