Lovibond SD 50 User Manual

Page 1
SD 50
pH
7. Delete Stored Data
Recall Data (see 6.) Press and hold (3 sec.)
Confirm deletion of stored data This erases all stored data
Automatic return to normal display
8. Select Buffer System
Power ON
Press and hold (3 sec.)
Press 1 x
Current pH system will flash
Press to confirm
Toggle to change / select (7.00 or 6.86)
Save and exit
9. Set Date/Time display type
Power ON
Press and hold (3 sec.)
Press 4 x
Date is displayed
Confirm
Toggle to 12 hour clock (am/pm) or 24 hour clock
Save and exit
4. Preparation of buffer solution
1. Fill a clean beaker with 20 ml distilled water.
2. Add one buffer tablet of the required value and shake or crush with a clean stirring rod until the tablet dissolves.
3. Allow to stand for about two minutes to ensure that the buffer is completely dissolved and to allow any insoluble particles to settle.
4. Immerse the electrode into the buf­fer solution and carry out calibration (see 3. Calibration)
5. Remove the electrode and rinse it tho­roughly
6. Use the buffer solution only once
Accuracy
The maximum tolerance of fresh buffer solu­tions is ±0.05 pH.
5. Selection of pH and mV mode
Toggle to select mode
Note: mode mV does NOT show ORP on the display, only millivolt.
6. Recall Data
Power ON
Press and hold (3 sec.)
Most recent memory setting will flash
Press to recall
Most recent memory setting is displayed
Toggle to other stored values
Press to exit
Press and hold (3 sec.) to start calibration
CAL 1 will flash
Wait until
disappears
Press 1 sec. Calibration will take place
CAL 2 will flash
Exit for 1 point calibration or go to 2nd point calibration
Rinse electrode with distilled water and immerse into pH 4
L
buffer solution.
CAL 2 will flash
Wait until
disappears
Press and hold (1 sec.)
Calibration will take place CAL 3 will flash
Exit for 2 point calibration or go to 3rd point calibration
Rinse electrode with distilled water and immerse into pH 10
H
buffer solution
CAL 3 will flash
Wait until
disappears
Press and hold (1 sec.)
Calibration will take place
3 point calibration is finished
LMH
will appear on screen and show type of calibration (1 point to 3 point calibration)
1. Power ON/OFF
Press briefly
If ! is displayed, press to deactivate
Automatically turns off after 8 minutes of inactivity
Press and hold (3 sec.) to turn the instrument off
Press to activate / de-activate backlight
2. Taking a reading
Remove protective cap and immerse the instrument into the sample to be measured so that the instrument is submersed to the sealing ring level.
Reading is displayed
Press briefly ! to freeze / unfreeze
Press and hold (3 sec.) to store data
While pressing button, memory listing (e.g. 01) appears
3. Calibration
Rinse electrode thoroughly with distilled water before using in each buffer solution.
Re-calibrate according to individual recommendations.
Use Lovibond
®
Standard Solutions or pH buffer tablets in accordance with the kit manual (see 4.)
Remove protective cap from the probe. Turn ON, rinse electrode with distilled water and immerse the probe into pH 7
M
solution.
Page 2
11. Set Temperature System
Power ON
Press and hold (3 sec.)
Press 3 x
Temperature system will flash
10. Set Date and Time
Power ON
Select Date/Time type first (see 9.)
Press and hold (3 sec.)
Press 2 x
Time will flash
Confirm
Select minutes
To increase
To decrease
Save and change to hours
To increase
To decrease
Save and change to Year
To increase
To decrease
Save and change to Day/Month
To increase
To decrease
Save and exit
Replacement of batteries
pH Buffer
Article No.
72 12 50 pH buffer-set, each 90 ml (25°C)
4.00/7.00/10.00 72 12 52 pH buffer 4.00 (25°C) 1 litre 72 12 54 pH buffer 7.00 (25°C) 1 litre 72 12 56 pH buffer 10.00 (25°C) 1 litre 51 56 00 buffer tablets pH 10 100 pc. 51 56 01 buffer tablets pH 10 250 pc. 51 56 10 buffer tablets pH 7 100 pc. 51 56 11 buffer tablets pH 7 250 pc. 51 56 20 buffer tablets pH 4 100 pc. 51 56 21 buffer tablets pH 4 250 pc.
Warranty
The waterproof SD 50 is warranted to be free from manufacturing defects for 2 years and electrode (probe) module for 6 months.
Error Code and Actions
Description Range Action
Electrode not installed Thermistor: Install or replace or faulty 0.5~100 kΩ electrode
pH out of range 0~14 pH Use new solution or replace electrode Electrode is not touching the Immerse electrode sample solution into solution Temperature out of 0~60 °C Bring solution to the range temperature within range Offset out of range -60~60 mV Use new buffer or replace electrode Slope out of range 85%~115%, Use new buffer or 50~68 (mV/pH) replace electrode mV out of range -1800~1800 mV Use new solution
SD 50 pH Specications/
Features
1 Operating range: 0-60 °C, 0-14 pH,
-1800 mV to +1800 mV 2 pH resolution 0.01 pH. accuracy ± 0,05 pH 3 Selectable buffer system (pH 7.00 or
pH 6.86): 1-, 2- or 3-point-calibration with auto-recognition 4 Automatic Temperature compensation 5 mV resolution 0.1 mV within ±1000 mV and 1 mV over ±1000 mV, accuracy ±20 mV 6 Temperature resolution and accuracy
0.1 °C, selectable °C or °F 7 T ime and date display with 25 sets of data storage (non-volatile) 8 22x22 mm LCD display, with backlight 9 2 x CR 2032 batteries 10 Battery life > 25 hr (continuous use, backlight OFF), low battery indicator on LCD display 11 Auto-power off to save battery-life (8 minutes non-use)
Instruction manuals available in further languages, visit www.lovibond.com • Printed in Germany 11/13
Storage of electrode
The sponge in the protective cap should be moistened with tap water.
Store electrodes dry when instrument is not being used for longer periods.
Moisten electrode in tap water for some hours before initial use and after dormant periods.
1 2 3
4 5 6
Replacement of electrode
new electrode
Confirm
Select Temperature style (°F/°C)
Save and exit
Accessories
19 50 017 Battery CR 2032 (2 Batteries required) 19 48 20 pH replacement electrode 38 48 01 Beaker, 100 ml
+
+
+
+
-
-
-
-