Lovibond PM 620 Instruction Manual

Page 1
Instruction manual
Photometer-System
P
620
GB
Page 2
Page 3
Please carry out the following steps as described in the Instruction manual. Become familiar with your new pho­tometer before starting with the first tests:
• Unpacking and inspection of delivery contents, see page 162.
• Install the batteries, see page 112, 113.
Perform the following settings in the Mode-Menu; Instruction manual from page 123 and following:
• MODE 10: select language
• MODE 12: set date and time
• MODE 34: perform “Delete data“
• MODE 69: perform “User m. init” to initialise the user polynomial system
If required set other functions.
Page 4
Page 5
Wichtige Information
Um die Qualität unserer Umwelt zu erhalten, beschützen und zu verbessern
Entsorgung von elektronischen Geräten in der Europäischen Union
Aufgrund der Europäischen Verordnung 2012/19/EU darf Ihr elektronisches Gerät
nicht mit dem normalen Hausmüll entsorgt werden!
Tintometer GmbH entsorgt ihr elektrisches Gerät auf eine professionelle und für die Umwelt verantwortungsvolle Weise. Dieser Service ist, die Transportkosten
nicht inbegriffen, kostenlos. Dieser Service gilt ausschließlich für elektrische Geräte
die nach dem 13.08.2005 erworben wurden. Senden Sie Ihre zu entsorgenden
Tintometer Geräte frei Haus an Ihren Lieferanten.
Important Information
To Preserve, Protect and Improve the Quality of the Environment
Disposal of Electrical Equipment in the European Union
Because of the European Directive 2012/19/EU your electrical instrument must not
be disposed of with normal household waste!
Tintometer GmbH will dispose of your electrical instrument in a professional
and environmentally responsible manner. This service, excluding the cost of
transportation is free of charge. This service only applies to electrical instruments
purchased after 13th August 2005. Send your electrical Tintometer instruments for
disposal freight prepaid to your supplier.
Notice importante
Conserver, protéger et optimiser la qualité de l’environnement
Élimination du matériel électrique dans l’Union Européenne
Conformément à la directive européenne nº 2012/19/UE, vous ne devez plus jeter
vos instruments électriques dans les ordures ménagères ordinaires !
La société Tintometer GmbH se charge d’éliminer vos instruments électriques de
façon professionnelle et dans le respect de l’environnement. Ce service, qui ne
comprend pas les frais de transport, est gratuit. Ce service n’est valable que
pour des instruments électriques achetés après le 13 août 2005. Nous vous prions
d’envoyer vos instruments électriques Tintometer usés à vos frais à votre fournisseur.
Belangrijke informatie
Om de kwaliteit van ons leefmilieu te behouden, te verbeteren en te
beschermen is voor landen binnen de Europese Unie de Europese richtlijn
2012/19/EU voor het verwijderen van elektronische apparatuur opgesteld.
Volgens deze richtlijn mag elektronische apparatuur niet met het huishoudelijk afval
worden afgevoerd.
Tintometer GmbH verwijdert uw elektronisch apparaat op een professionele en
milieubewuste wijze. Deze service is, exclusief de verzendkosten, gratis en alleen geldig voor elektrische apparatuur die na 13 augustus 2005 is gekocht. Stuur uw te
verwijderen Tintometer apparatuur franco aan uw leverancier.
DE
GB
FR
NL
Page 6
Información Importante
Para preservar, proteger y mejorar la calidad del medio ambiente
Eliminación de equipos eléctricos en la Unión Europea
Con motivo de la Directiva Europea 2012/19/UE, ¡ningún instrumento eléctrico
deberá eliminarse junto con los residuos domésticos diarios!
Tintometer GmbH se encargará de dichos instrumentos eléctricos de una manera
profesional y sin dañar el medio ambiente. Este servicio, el cual escluye los gastos de transporte, es gratis y se aplicará únicamente a aquellos instrumentos eléctricos
adquiridos después del 13 de agosto de 2005. Se ruega enviar aquellos instrumentos
eléctricos inservibles de Tintometer a carga pagada a su distribuidor.
Informazioni importanti
Conservare, proteggere e migliorare la qualità dell’ambiente
Smaltimento di apparecchiature elettriche nell’Unione Europea
In base alla Direttiva europea 2012/19/UE, gli apparecchi elettrici non devono essere
smaltiti insieme ai normali rifiuti domestici!
Tintometer GmbH provvederà a smaltire i vostri apparecchi elettrici in maniera
professionale e responsabile verso l’ambiente. Questo servizio, escluso il trasporto,
è completamente gratuito. Il servizio si applica agli apparecchi elettrici acquistati
successivamente al 13 agosto 2005. Siete pregati di inviare gli apparecchi elettrici
Tintometer divenuti inutilizzabili a trasporto pagato al vostro rivenditore.
Informação Importante
Para Preservar, Proteger e Melhorar a Qualidade do Ambiente
Remoção de Equipamento Eléctrico na União Europeia
Devido à Directiva Europeia 2012/19/UE, o seu equipamento eléctrico naõ deve ser
removido com o lixo doméstico habitual!
A Tintometer GmbH tratará da remoção do seu equipamento eléctrico de forma
profissional e responsável em termos ambientais. Este serviço, não incluindo
os custos de transporte, é gratuito. Este serviço só é aplicável no caso de
equipamentos eléctricos comprados depois de 13 de Agosto de 2005. Por favor,
envie os seus equipamentos eléctricos Tintometer que devem ser removidos ao seu
fornecedor (transporte pago).
Istotna informacja
Dla zachowania, ochrony oraz poprawy naszego środowiska
Usuwanie urządzeń elektronicznych w Unii Europejskiej
Na podstawie Dyrektywy Parlamentu Europejskiego 2012/19/UE nie jest
dozwolone usuwanie zakupionych przez Państwo urządzeń elektronicznych wraz z
normalnymi odpadami z gospodarstwa domowego!
Tintometer GmbH usunie urządzenia elektrycznego Państwa w sposób
profesjonalny i odpowiedzialny z punktu widzenia środowiska. Serwis ten jest,
za wyjątkiem kosztów transportu, bezpłatny. Serwis ten odnosi się wyłącznie do
urządzeń elektrycznych zakupionych po 13.08.2005r. Przeznaczone do usunięcia
urządzenia rmy Tintometer mogą Państwo przesyłać na koszt własny do swojego
dostawcy.
ES
IT
PT
PL
Page 7
Wichtiger Entsorgungshinweis zu Batterien und Akkus
Jeder Verbraucher ist aufgrund der Batterieverordnung (Richtlinie 2006/66/
EG) gesetzlich zur Rückgabe aller ge- und verbrauchten Batterien bzw. Akkus
verpflichtet. Die Entsorgung über den Hausmüll ist verboten. Da auch bei Produkten
aus unserem Sortiment Batterien und Akkus im Lieferumgang enthalten sind, weisen
wir Sie auf folgendes hin:
Verbrauchte Batterien und Akkus gehören nicht in den Hausmüll, sondern können
unentgeltlich bei den öffentlichen Sammelstellen Ihrer Gemeinde und überall
dort abgegeben werden, wo Batterien und Akkus der betreffenden Art verkauft
werden. Weiterhin besteht für den Endverbraucher die Möglichkeit, Batterien und
Akkus an den Händler, bei dem sie erworben wurden, zurückzugeben (gesetzliche
Rücknahmepflicht).
Important disposal instructions for batteries and accumulators
EC Guideline 2006/66/EC requires users to return all used and worn-out batteries
and accumulators. They must not be disposed of in normal domestic waste. Because
our products include batteries and accumulators in the delivery package our advice is
as follows :
Used batteries and accumulators are not items of domestic waste. They must be
disposed of in a proper manner. Your local authority may have a disposal facility;
alternatively you can hand them in at any shop selling batteries and accumulators.
You can also return them to the company which supplied them to you; the company
is obliged to accept them.
Information importante pour l'élimination des piles et des accumulateurs
En vertu de la Directive européenne 2006/66/CE relative aux piles et accumulateurs,
chaque utilisateur est tenu de restituer toutes les piles et tous les accumulateurs
utilisés et épuisés. L'élimination avec les déchets ménagers est interdite. Etant donné
que l'étendue de livraison des produits de notre gamme contient également des piles
et des accumulateurs, nous vous signalons ce qui suit :
les piles et les accumulateurs utilisés ne sont pas des ordures ménagères, ils peuvent
être remis sans frais aux points de collecte publics de votre municipalité et partout où sont vendus des piles et accumulateurs du type concerné. Par ailleurs, l'utilisateur final a la possibilité de remettre les piles et les accumulateurs au commerçant auprès
duquel ils ont été achetés (obligation de reprise légale).
Belangrijke mededeling omtrent afvoer van batterijen en accu’s
Ledere verbruiker is op basis van de richtlijn 2006/66/EG verplicht om alle gebruikte
batterijen en accu’s in te leveren. Het is verboden deze af te voeren via het huisvuil.
Aangezien ook onze producten geleverd worden met batterijen en accu’s wijzen wij
u op het volgende; Lege batterijen en accu’s horen niet in het huisvuil thuis. Men
kan deze inleveren bij inzamelpunten van uw gemeente of overal daar waar deze
verkocht worden. Tevens bestaat de mogelijkheid batterijen en accu’s daar in te
leveren waar u ze gekocht heeft. (wettelijke terugnameplicht)
DE
GB
FR
NL
Page 8
Indicación importante acerca de la eliminación de pilas y acumuladores
Basado en la norma relativa a pilas/ baterías (directiva 2006/66/CE), cada consumidor,
está obligado por ley, a la devolución de todas las pilas/ baterías y acumuladores usados y
consumidos. Está prohibida la eliminación en la basura doméstica. Ya que en productos de
nuestra gama, también se incluyen en el suministro pilas y acumuladores, le sugerimos lo
siguiente:
Las pilas y acumuladores usados no pertenecen a la basura doméstica, sino que pueden
ser entregados en forma gratuita en cada uno de los puntos de recolección públicos de su
comunidad en los cuales se vendan pilas y acumuladores del tipo respectivo. Además, para
el consumidor final existe la posibilidad de devolver las pilas y baterías recargables a los
distribuidores donde se hayan adquirido (obligación legal de devolución).
Indicazioni importanti sullo smaltimento di pile e accumulatori
In base alla normativa concernente le batterie (Direttiva 2006/66/CE) ogni consumatore
è tenuto per legge alla restituzione di tutte le batterie o accumulatori usati ed esauriti. È
vietato lo smaltimento con i rifiuti domestici. Dato che anche alcuni prodotti del nostro assortimento sono provvisti di pile e accumulatori, vi diamo di seguito delle indicazioni:
Pile e accumulatori esauriti non vanno smaltiti insieme ai rifiuti domestici, ma depositati
gratuitamente nei punti di raccolta del proprio comune o nei punti vendita di pile e
accumulatori dello stesso tipo. Inoltre il consumatore finale può portare batterie e
accumulatori al rivenditore presso il quale li ha acquistati (obbligo di raccolta previsto per
legge).
Instruções importantes para a eliminação residual de pilhas e acumuladores
Os utilizadores finais são legalmente responsáveis, nos termos do Regulamento relativo
a pilhas e acumuladores (Directiva 2006/66/CE), pela entrega de todas as pilhas e
acumuladores usados e gastos. É proibida a sua eliminação juntamente com o lixo
doméstico. Uma vez que determinados produtos da nossa gama contêm pilhas e/ou
acumuladores, alertamos para os seguintes aspectos:
As pilhas e acumuladores usados não podem ser eliminados com o lixo doméstico, devendo sim ser entregues, sem encargos, junto dos pontos de recolha públicos do seu município, ou em qualquer ponto de venda de pilhas e acumuladores. O utilizador final
dispõe ainda da possibilidade de entregar as pilhas e/ou acumuladores no estabelecimento
comerciante onde os adquiriu (dever legal de aceitar a devolução).
Istotna wskazówka dotycząca utylizacji baterii i akumulatorów
Każdy użytkownik na mocy rozporządzenia w sprawie baterii (wytyczna 2006/66/WE)
jest ustawowo zobowiązany do oddawania wszystkich rozładowanych i zużytych baterii
lub akumulatorów. Utylizacja wraz z odpadkami domowymi jest zabroniona. Ponieważ
także w produktach z naszego asortymentu zawarte są w zakresie dostawy baterie i
akumulatory, zwracamy uwagę na poniższe zasady:
zużyte baterie i akumulatory nie mogą być wyrzucane wraz z odpadkami domowymi,
lecz powinny być bezpłatnie przekazywane w publicznych miejscach zbiórki
wyznaczonych przez gminę lub oddawane w punktach, gdzie sprzedawane są baterie i
akumulatory danego rodzaju. Poza tym użytkownik końcowy ma możliwość zwrócenia baterii i akumulatorów do przedstawiciela handlowego, u którego je nabył (ustawowy
obowiązek przyjęcia).
ES
IT
PT
PL
Page 9
1
PM620_4c 07/2015
Safety precautions
CAUTION
Reagents are formulated exclusively for chemical analysis and must not be used for
any other purpose. Reagents must not get into the hands of children. Some of the
reagents contain substances which are not entirely harmless environmentally. Be
aware of the ingredients and take proper care when disposing of the test solution.
CAUTION
Please read this instruction manual before unpacking, setting up or using the
photometer. Please read the method description completely before performing
the test. Be aware of the risks of using the required reagents by reading the MSDS
(Material Safety Data Sheets). Failure could result in serious injury to the operator or
damage to the instrument.
MSDS: www.lovibond.com
CAUTION
The accuracy of the instrument is only valid if the instrument is used in an
environment with controlled electromagnetic disturbances according to DIN 61326.
Wireless devices, e.g. wireless phones, must not be used near the instrument.
Revision_4c 07/2015
Page 10
2
PM620_4c 07/2015
Table of contents
Part 1 Methods ................................................................................................... 7
1.1 Table of Methods ..................................................................................................... 8
Acid demand to pH 4.3 ................................................................................................... 12
Alkalinity-total (Alkalinity-m, m-Value) ............................................................................. 14
Alkalinity-total HR (Alkalinity-m HR, m-Value HR) ............................................................ 16
Aluminium with tablets ................................................................................................... 18
Aluminium (powder pack) ............................................................................................... 20
Ammonia with tablet ...................................................................................................... 22
Bromine .......................................................................................................................... 24
Chlorine .......................................................................................................................... 26
Chlorine with tablet ........................................................................................................ 28
free Chlorine ............................................................................................................... 28
total Chlorine .............................................................................................................. 29
differentiated determination (free, combined, total) .................................................... 30
Chlorine HR with tablet ................................................................................................... 32
free Chlorine ............................................................................................................... 32
total Chlorine .............................................................................................................. 33
differentiated determination (free, combined, total) .................................................... 34
Chlorine with liquid reagent ............................................................................................ 36
free Chlorine ............................................................................................................... 36
total Chlorine .............................................................................................................. 37
differentiated determination (free, combined, total) .................................................... 38
Chlorine (powder pack) ................................................................................................... 40
free Chlorine ............................................................................................................... 40
total Chlorine .............................................................................................................. 41
differentiated determination (free, combined, total) .................................................... 42
Chlorine HR (powder pack) ............................................................................................. 44
free Chlorine ............................................................................................................... 44
total Chlorine .............................................................................................................. 45
differentiated determination (free, combined, total) .................................................... 46
Chlorine dioxide .............................................................................................................. 48
in presence of Chlorine ............................................................................................... 50
in absence of Chlorine ................................................................................................ 53
Chlorite in presence of Chlorine and Chlorine dioxide ..................................................... 54
Copper with tablet .......................................................................................................... 58
differentiated determination (free, combined, total) .................................................... 59
Page 11
3
PM620_4c 07/2015
free Copper ................................................................................................................ 60
total Copper ............................................................................................................... 61
Copper (powder pack) .................................................................................................... 62
CyA-TEST (Cyanuric acid) ................................................................................................ 64
H2O2 HR with liquid reagent (16 mm Ø) .......................................................................... 66
Hardness, Calcium with Calcio tablet .............................................................................. 68
Hardness, total ................................................................................................................ 70
Hardness, total HR .......................................................................................................... 72
Iodine ............................................................................................................................. 74
Iron with tablet ............................................................................................................... 76
Oxygen, active ................................................................................................................ 78
Ozone ............................................................................................................................. 80
in presence of Chlorine ............................................................................................... 82
in absence of Chlorine ................................................................................................ 84
PHMB (Biguanide) ........................................................................................................... 86
Phosphate, ortho LR with tablet ...................................................................................... 88
pH value LR with tablet ................................................................................................... 90
pH value with tablet ........................................................................................................ 92
pH value with liquid reagent ........................................................................................... 94
pH value HR with tablet .................................................................................................. 96
Sodium hypochlorite ....................................................................................................... 98
Sulfate with tablet ........................................................................................................ 100
Sulfate (powder pack) ................................................................................................... 102
Urea ............................................................................................................................. 104
1.2 Important notes ......................................................................................... 106
1.2.1 Correct use of reagents ................................................................................ 106
1.2.2 Cleaning of vials and accessories for analysis ................................................ 107
1.2.3 Guidelines for photometric measurements ................................................... 107
1.2.4 Sample dilution techniques ........................................................................... 109
1.2.5 Correcting for volume additions ................................................................... 109
Part 2 Instrument Manual ....................................................................... 111
2.1 Operation ................................................................................................... 112
2.1.1 Set up .......................................................................................................... 112
2.1.2 Saving data – Important Notes ..................................................................... 112
2.1.3 Replacement of batteries .............................................................................. 112
2.1.4 Instrument (explosion drawing) .................................................................... 113
Page 12
4
PM620_4c 07/2015
2.2 Overview of function keys ....................................................................... 115
2.2.1 Overview ...................................................................................................... 115
2.2.2 Displaying time and date .............................................................................. 116
2.2.3 User countdown ........................................................................................... 116
2.2.4 Display backlight .......................................................................................... 116
2.3 Operation mode ........................................................................................ 117
2.3.1 Automatic switch off .................................................................................... 117
2.3.2 Selecting a method ...................................................................................... 117
2.3.2.1 Method Information (F1) .............................................................................. 117
2.3.2.2 Chemical Species Information ....................................................................... 118
2.3.3 Differentiation .............................................................................................. 118
2.3.4 Performing Zero ........................................................................................... 118
2.3.5 Performing Tests ........................................................................................... 119
2.3.6 Ensuring reaction periods (countdown) ......................................................... 119
2.3.7 Changing chemical species ........................................................................... 120
2.3.8 Storing results .............................................................................................. 120
2.3.9 Printing results (Infra-Red Interface Module) (optional) .................................. 121
2.3.10 Perform additional measurements ................................................................ 121
2.3.11 Selecting a new method ............................................................................... 122
2.3.12 Measure absorbance .................................................................................... 122
2.4 Photometer settings: Table of Mode Functions ...................................... 123
2.4.1 blank because of technical requirements ...................................................... 124
2.4.2 Instrument basic settings 1 ........................................................................... 124
2.4.3 Printing of stored results ............................................................................... 128
2.4.4 Recall / delete stored results.......................................................................... 133
2.4.5 Calibration ................................................................................................... 138
2.4.6 Lab function ................................................................................................. 144
Profi-Mode ................................................................................................... 144
One Time Zero ............................................................................................. 145
2.4.7 User operations ............................................................................................ 146
User method list ........................................................................................... 146
User Concentration Methods ........................................................................ 148
User Polynomials .......................................................................................... 150
Delete User Methods .................................................................................... 153
Print Data of User Methods .......................................................................... 154
Initialise User Method System ....................................................................... 155
2.4.8 Special functions .......................................................................................... 156
Page 13
5
PM620_4c 07/2015
Langelier Saturation Index ............................................................................ 156
2.4.9 Instrument basic settings 2 ........................................................................... 158
2.4.10 Instrument special functions /service ............................................................. 159
2.5 Data transfer .............................................................................................. 160
2.5.1 Data Printing ................................................................................................ 160
2.5.2 Data transfer to a personal computer ........................................................... 160
2.5.3 Internet Updates .......................................................................................... 160
Part 3 Enclosure .............................................................................................. 161
3.1 Unpacking .................................................................................................. 162
3.2 Delivery contents ...................................................................................... 162
3.3 blank because of technical requirements ................................................ 162
3.4 Technical data ............................................................................................ 163
3.5 Abbreviations ............................................................................................ 164
3.6 Troubleshooting ........................................................................................ 165
3.6.1 Operating messages in the display / error display .......................................... 165
3.6.2 General ........................................................................................................ 167
3.7 Declaration of CE-Conformity .................................................................. 168
Page 14
6
PM620_4c 07/2015
Page 15
7
PM620_4c 07/2015
Part 1
Methods
Page 16
8
PM620_4c 07/2015
1.1 Table of Methods
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range;
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
20 Acid demand to
pH 4.3 T
tablet 0.1-4 mmol/l
Acid/Indicator
1,2,5
610 12
30
Alkalinity, total T
tablet 5-200 mg/l CaCO3Acid/Indicator
1,2,5
610 14
31
Alkalinity HR, total T
tablet 5-500 mg/l CaCO3Acid/Indicator
1,2,5
610 16
40 Aluminium T tablet 0.01-0.3 mg/l Al Eriochrome
Cyanine R
2
530 18
50 Aluminium PP PP + liquid 0.01-
0.25
mg/l Al Eriochrome
Cyanine R
2
530 20
60 Ammonia T tablet 0.02-1 mg/l N
Indophenol blue
2,3
610 22
80 Bromine T tablet 0.05-13 mg/l Br
2
DPD
5
530 24
100 Chlorine T * tablet 0.01-6 mg/l Cl
2
DPD
1,2,3
530 26, 28
103 Chlorine HR T * tablet 0.1-10 mg/l Cl
2
DPD
1,2,3
530 26, 32
101 Chlorine L * liquid 0.02-4 mg/l Cl
2
DPD
1,2,3
530 26, 36
110 Chlorine PP * PP 0.02-2 mg/l Cl
2
DPD
1,2
530 26, 40
111 Chlorine HR PP * PP 0.1-8 mg/l Cl
2
DPD
1,2
530
26, 44
120 Chlorine dioxide T tablet 0.02-11 mg/l ClO
2
DPD, Glycine
1,2
530 48
150 Copper T * tablet 0.05-5 mg/l Cu Biquinoline
4
560 58
153 Copper PP PP 0.05-5 mg/l Cu Bicinchoninate 560 62
160
CyA-TEST T
tablet 0-160 mg/l CyA Melamine 530 64
214 H2O2 HR L liquid 40-500 mg/l H2O2Titanium
tetrachloride/acid
530
66
191 Hardness, Calcium 2Ttablet 0-500 mg/l CaCO3Murexide
4
560 68
200
Hardness, total T
tablet 2-50 mg/l CaCO3Metallphthalein 3560 70
201 Hardness, total
HR T
tablet 20-500 mg/l CaCO3Metallphthalein 3560 72
215 Iodine T tablet 0.05-3.6 mg/l I DPD
5
530 74
220 Iron T tablet 0.02-1 mg/l Fe PPST
3
560
76
Page 17
9
PM620_4c 07/2015
1.1 Table of Methods
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range;
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
290
Oxygen, active T
tablet 0.1-10 mg/l O
2
DPD 530 78
300
Ozone (DPD) T
tablet 0.02-2 mg/l O
3
DPD/Glycine
5
530 80
70 PHMB T tablet 2-60 mg/l PHMB Buffer/Indicator 560 86
319 Phosphate, T
ortho LR
tablet 0.05-4 mg/l PO
4
Ammonium­molybdate
2,3
660 88
329
pH-Value LR T
tablet 5.2-6.8
Bromocresolpurple5560 90
330 pH-Value T tablet 6.5-8.4 Phenolred
5
560 92
331 pH-Value L liquid 6.5-8.4 Phenolred
5
560 94
332
pH-Value HR T
tablet 8.0-9.6 Thymolblue
5
560 96
212 Sodium
hypochlorite T
tablet 0.2-16 % NaOCl Potassium
iodide
5
530 98
355 Sulfate T tablet 5-100 mg/l SO
4
Bariumsulfate­Turbidity
610 100
360 Sulfate PP PP 5-100 mg/l SO
4
Bariumsulfate­Turbidity
2
530 102
390 Urea T tablet +
liquid
0.1-2.5 mg/l Urea Indophenol/ Urease
610 104
Page 18
10
PM620_4c 07/2015
1.1 Methods
The precision of Lovibond® Reagent Systems (tablets, powder packs and tube tests) is identical to the precision specified in standards literature such as American Standards (AWWA), ISO etc.
Most of the data referred to in these standard methods relates to Standard Solutions. Therefore they are not readily applicable to drinking-, boiler- or waste-water, since various interferences can have a major influence on the accuracy of the method. For this reason we don’t state such potentially misleading data.
Due to the fact that each sample is different, the only way to check the tolerances (‘precision’) is the Standard Additions Method.
According to this method, first the original sample is tested. Then further samples (2 to 4) are taken and small amounts of a Standard Solution are added, and further results are obtained. The amounts added range from approximately half, up to double the amount present in the sample itself.
These supplementary results make it possible to estimate the actual concentration of the original sample by comparison.
Literature
The reagent formulations are based on internationally recognised test methods. Some are described in national and/or international guidelines.
1. Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlammuntersuchung
2. Standard Methods for the Examination of Water and Wastewater; 18th Edition, 1992
3. Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart 1989
4. Photometrische Analyse, Lange / Vejdelek, Verlag Chemie 1980
5. Colorimetric Chemical Analytical Methods, 9th Edition, London
Notes for searching:
OTZ (OneTimeZero) switching on and off, see Mode 55, page 145
Active Oxygen -> Oxygen, activ Alkalinity-m -> Alkalinity, total Biguanide -> PHMB Calcium Hardness -> Hardness, Calcium Cyanuric acid -> CyA-TEST H2O2 -> Hydrogen peroxide Total Hardness -> Hardness, total m-Value -> Alkalinity, total total Alkalinity -> Alkalinity, total total Hardness -> Hardness, total
Langelier Saturation -> Mode function 70 Index (Water Balance)
Page 19
11
PM620_4c 07/2015
Page 20
12
PM620_4c 07/2015
1.1 Methods
Acid demand to pH 4.3 with Tablet
0.1 – 4 mmol/l
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as Acid demand to pH 4.3 in mmol/l.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
2
0
Page 21
13
PM620_4c 07/2015
1.1 Methods
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Acid demand to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
Reagent Form of reagent/Quantity Order-No.
ALKA-M-PHOTOMETER Tablet / 100 513210BT
Page 22
14
PM620_4c 07/2015
1.1 Methods
Alkalinity, total = Alkalinity-m = m-Value with Tablet
5 – 200 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as total Alkalinity.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
0
Page 23
15
PM620_4c 07/2015
1.1 Methods
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Alkalinity to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
3. Conversion table:
Acid demand to pH 4.3
DIN 38 409 (KS 4.3)
German
°dH*
English
°eH*
French
°fH*
1 mg/l CaCO
3
0.02 0.056 0.07 0.1
*Carbonate hardness (reference = Hydrogencarbonate-anions) Example: 10 mg/l CaCO3 = 10 mg/l x 0.056 = 0.56 °dH 10 mg/l CaCO3 = 10 mg/l x 0.02 = 0.2 mmol/l
4. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
ALKA-M-PHOTOMETER Tablet / 100 513210BT
Page 24
16
PM620_4c 07/2015
1.1 Methods
Alkalinity HR, total = Alkalinity-m HR = m-Value HR with Tablet
5 – 500 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-HR PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Press [ ] key.
Wait for a reaction period of 1 minute.
8. Remix the solution.
9. Place the vial in the sample chamber making sure that the marks are aligned.
10. Press TEST key.
The result is shown in the display as total Alkalinity.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
1
Countdown 1:00 start:
Page 25
17
PM620_4c 07/2015
1.1 Methods
Notes:
1. For verification of the result look carefully at the bottom of the vial. If a thin yellow layer forms, then mix the vial again. This ensures that reaction is complete. Reread the result.
2. Conversion table:
Acid demand to pH 4.3
DIN 38 409 (KS 4.3)
German
°dH*
English
°eH*
French
°fH*
1 mg/l CaCO
3
0.02 0.056 0.07 0.1
*Carbonate hardness (reference = Hydrogencarbonate-anions) Example: 10 mg/l CaCO3 = 10 mg/l x 0.056 = 0.56 °dH 10 mg/l CaCO3 = 10 mg/l x 0.02 = 0.2 mmol/l
3. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
ALKA-M-HR PHOTOMETER Tablet / 100 513240BT
Page 26
18
PM620_4c 07/2015
1.1 Methods
Aluminium with Tablet
0.01 – 0.3 mg/l Al
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALUMINIUM No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod (dissolve the tablet).
6. Add one ALUMINIUM No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Aluminium.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
4
0
Page 27
19
PM620_4c 07/2015
1.1 Methods
Notes:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%). Rinse them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The effect of this is generally insignificant unless the water has fluoride added artificially. In this case, the following table should be used:
Fluoride [mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. A special tablet ingredient prevents effects on the measurement due to iron and manganese.
5. Al
Al2O
3
Reagent Form of reagent/Quantity Order-No. Set ALUMINIUM No. 1 / No. 2 Tablet / per 100
inclusive stirring rod
517601BT
ALUMINIUM No. 1 Tablet / 100 515460BT
ALUMINIUM No. 2 Tablet / 100 515470BT
Page 28
20
PM620_4c 07/2015
1.1 Methods
Aluminium with Vario Powder Pack
0.01 – 0.25 mg/l Al
Use two clean vials (24 mm Ø) and mark one as blank for zeroing.
1. Fill 20 ml of the water sample in a 100 ml beaker.
2.
Add the contents of one Vario Aluminum ECR F20
Powder Pack straight from the foil to the water sample.
3. Dissolve the powder using a clean stirring rod.
4. Press [ ] key. Wait for a reaction period of 30 seconds.
After the reaction period is finished proceed as follows:
5.
Add
the contents of
one Vario Hexamine F20 Powder
Pack straight
from the foil to the same water sample.
6. Dissolve the powder using a clean stirring rod.
7. Add 1 drop of Vario Aluminum ECR Masking Reagent in the vial marked as blank.
8. Add 10 ml of the prepared water sample to the vial (this is the blank).
9. Add the remaining 10 ml of the prepared water sample in the second clean vial (this is the sample).
10. Close the vials tightly with the caps and swirl several times to mix the contents.
11. Press [ ] key. Wait for a reaction period of 5 minutes.
After the reaction period is finished proceed as follows:
Countdown 1 0:30 start:
Countdown 2 5:00 start:
1 2
3 4
5
0
Page 29
21
PM620_4c 07/2015
1.1 Methods
12. Place the vial (the blank) in the sample chamber making sure that the marks are aligned.
13. Press ZERO key.
14. Remove the vial from the sample chamber.
15. Place the vial (the sample) in the sample chamber making sure that the marks are aligned.
16. Press TEST key.
The result is shown in the display in mg/l Aluminium.
Notes:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%). Rinse them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The effect of this is generally insignificant unless the water has fluoride added artificially. In this case, the following table should be used:
Fluoride [mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. Al
Al2O
3
Reagent Form of reagent/Quantity Order-No. Set
VARIO Aluminium ECR F20 VARIO Aluminium Hexamine F 20 VARIO Aluminium ECR Masking Reagent
Powder Pack / 100 Powder Pack / 100 Liquid reagent / 25 ml
535000
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 30
22
PM620_4c 07/2015
1.1 Methods
Ammonia with Tablet
0.02 – 1 mg/l N
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one AMMONIA No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Add one AMMONIA No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 10 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Ammonia as N.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 10:00
6
0
Page 31
23
PM620_4c 07/2015
1.1 Methods
Notes:
1. The tablets must be added in the correct sequence.
2. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2 tablet has been added.
3. The temperature of the sample is important for full colour development. At a temperature below 20°C the reaction period is 15 minutes.
4. Sea water samples: Ammonia conditioning reagent is required when testing sea water or brackish water samples to prevent precipitation of salts. Fill the test tube with the sample to the 10 ml mark and add one level spoonful of Conditioning Powder. Mix to dissolve, then continue as described in the test instructions.
5. Conversion: mg/l NH4 = mg/l N x 1.29 mg/l NH3 = mg/l N x 1.22
6. N NH4
NH
3
Reagent Form of reagent/Quantity Order-No. Set AMMONIA No. 1 / No. 2 Tablet / per 100
inclusive stirring rod
517611BT
AMMONIA No. 1 Tablet / 100 512580BT
AMMONIA No. 2 Tablet / 100 512590BT
Ammonia conditioning reagent (Sea water samples)
(approx. 100 tests) powder / 15 g
460170
Page 32
24
PM620_4c 07/2015
1.1 Methods
Bromine with Tablet
0.05 – 13 mg/l Br
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l Bromine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
8
0
Page 33
25
PM620_4c 07/2015
1.1 Methods
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances,the subsequent determination of Bromine may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) forone hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Bromine, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 22 mg/l Bromine can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Bromine. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Bromine.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
Page 34
26
PM620_4c 07/2015
1.1 Methods
Chlorine with Tablet
0.01 – 6 mg/l Cl
2
Chlorine HR with Tablet
0.1 – 10 mg/l Cl
2
Chlorine with Liquid Reagent
0.02 - 4 mg/l Cl
2
Chlorine with Vario Powder Pack
0.02 - 2 mg/l Cl
2
Chlorine HR with Vario Powder Pack
0.1 - 8 mg/l Cl
2
The following selection is shown in the display:
for the differentiated determination of free, combined and total Chlorine.
for the determination of free Chlorine.
for the determination of total Chlorine.
Select the desired determination with the arrow
keys [ ] and [ ]. Confirm with [ ] key.
Chlorine >> diff free total
>> diff
>> free
>> total
0
0
1
0
3
1
0
1
1
0
1
1
1
1
1
Page 35
27
PM620_4c 07/2015
1.1 Methods
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Chlorine may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. For individual testing of free and total Chlorine, the use of different sets of glassware is recommended (EN ISO 7393-2, 5.3)
3. Preparing the sample: When preparing the sample, the lost of Chlorine, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
4. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents therefore contain a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
5. Exceeding the measuring range: Concentrations above: 10 mg/l Chlorine using tablets (method 100) 4 mg/l Chlorine using liquid reagents (method 101) 2 mg/l using powder packs (method 110) 8 mg/l using powder packs (method 111) can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Chlorine. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
6. Turbidity (can lead to errors): The use of the DPD No. 1 tablet (method 100) in samples with high Calcium ion contents* and/or high conductivity* can lead to turbidity of the sample and therefore incorrect measurements. In this case, the reagent tablet DPD No. 1 High Calcium should be used as an alternative. If turbidity does occur after the DPD No. 3 tablet has been added, this can be prevented by using the DPD No. 1 High Calcium tablet and the DPD No. 3 High Calcium tablet. The DPD No. 1 High Calcium should only be used in combination with the DPD No.3 High Calcium.
* it is not possible to give exact values, because the development of turbidity depends on the nature of the sample.
7. If ??? is displayed at a differentiated test result see page 166.
8. Oxidizing agents such as Bromine, Ozone etc. interfere as they react in the same way as Chlorine.
Page 36
28
PM620_4c 07/2015
1.1 Methods
Chlorine, free with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes:
See page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
0
01
Page 37
29
PM620_4c 07/2015
1.1 Methods
Chlorine, total with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
Notes:
See page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
0
01
Page 38
30
PM620_4c 07/2015
1.1 Methods
Chlorine, differentiated determination with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it,
leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and
crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add one DPD No. 3 tablet straight from the foil to the
same water sample and crush the tablet using a clean stirring rod.
12. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
0
01
Page 39
31
PM620_4c 07/2015
1.1 Methods
13. Place the vial in the sample chamber making sure that the marks are aligned.
14. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes: See page 27
Reagent Form of reagent/Quantity Order-No. Set DPD No. 1 / No. 3 Tablet / per 100
inclusive stirring rod
517711BT
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
Kombi-Pack
DPD No. 1 HIGH CALCIUM / DPD No. 3 HIGH CALCIUM
Tablet / per 100 inclusive stirring rod
517781BT
DPD No. 1 HIGH CALCIUM Tablet / 100 515740BT
DPD No. 3 HIGH CALCIUM Tablet / 100 515730BT
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb Cl *,** mg/l total Cl
Page 40
32
PM620_4c 07/2015
1.1 Methods
Chlorine HR, free with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 HR tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes:
See page 27
3
01
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 41
33
PM620_4c 07/2015
1.1 Methods
Chlorine HR, total with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total Chlo-
rine.
Notes:
See page 27
3
01
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
Page 42
34
PM620_4c 07/2015
1.1 Methods
Chlorine HR, differentiated determination with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 HR tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add one DPD No. 3 HR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
3
01
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Page 43
35
PM620_4c 07/2015
1.1 Methods
12. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
13. Place the vial in the sample chamber making sure that the marks are aligned.
14. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes: See page 27
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 HR Tablet / 100 511500BT
DPD No. 3 HR Tablet / 100 511590BT
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb Cl *,** mg/l total Cl
Page 44
36
PM620_4c 07/2015
1.1 Methods
Chlorine, free with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty the vial.
5. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times to mix the contents.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlorine.
Notes (free and total Chlorine):
1. After use replace the bottle caps securely noting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C.
3. Also see page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
0
1
1
Page 45
37
PM620_4c 07/2015
1.1 Methods
Chlorine, total with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty
the vial.
5. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
3 drops of DPD 3 solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
to mix the contents.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
0
1
1
Page 46
38
PM620_4c 07/2015
1.1 Methods
Chlorine, differentiated determination with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty the vial.
5. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times to mix the contents.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add 3 drops of DPD 3 solution to the same water sample.
12. Close the vial tightly with the cap and swirl several times to mix the contents.
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
0
1
1
Page 47
39
PM620_4c 07/2015
1.1 Methods
13. Place the vial in the sample chamber making sure that
the marks are aligned.
14. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
1. After use replace the bottle caps securely noting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C.
3. Also see page 27
Reagent Form of reagent/Quantity Order-No. Set
DPD No. 1 buffer solution DPD No. 1 reagent solution DPD No. 3 solution
(approx. 300 tests) 3 x Liquid reagent / 15 ml 1 x Liquid reagent / 15 ml 2 x Liquid reagent / 15 ml
471056
DPD No. 1 buffer solution Liquid reagent / 15 ml 471010
DPD No. 1 reagent solution Liquid reagent / 15 ml 471020
DPD No. 3 solution Liquid reagent / 15 ml 471030
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
Page 48
40
PM620_4c 07/2015
1.1 Methods
Chlorine, free with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine FREE-DPD / F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l free Chlorine.
Notes:
See page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
0
1
1
Page 49
41
PM620_4c 07/2015
1.1 Methods
Chlorine, total with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine TOTAL-DPD /
F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
Notes:
See page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
Countdown 3:00
0
1
1
Page 50
42
PM620_4c 07/2015
1.1 Methods
Chlorine, differentiated determination with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine FREE-DPD/ F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
9. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times and then fill the vial with 10 ml of the water sample.
10. Add the contents of one VARIO Chlorine TOTAL-DPD / F10 Powder Pack straight from the foil to the water sample.
11. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
1 2
3 4
0
1
1
Page 51
43
PM620_4c 07/2015
1.1 Methods
12. Place the vial in the sample chamber making sure that
the marks are aligned.
13. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes: See page 27
Reagent Form of reagent/Quantity Order-No.
Vario Clorine Free-DPD/ F10 Powder Pack / 100 530100
VARIO Chlorine Total-DPD/F10 Powder Pack / 100 530120
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
T1 accepted prepare T2 press TEST
Countdown 3:00
Page 52
44
PM620_4c 07/2015
1.1 Methods
Chlorine HR, free with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine Free-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press the TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes: See page 27
1 2
3 4
1
1
1
10 ml
5 ml
3
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 53
45
PM620_4c 07/2015
1.1 Methods
Chlorine HR, total with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine TOTAL-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key. Wait for a reaction period of 3-6 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total Chlo-
rine.
Notes:
See page 27
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
Countdown 3:00
1
1
1
10 ml
5 ml
3
Page 54
46
PM620_4c 07/2015
1.1 Methods
Chlorine HR, differentiated determination with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine Free-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press the TEST key.
9. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times and then fill the vial with 5 ml of the water sample.
10. Add the contents of two Vario Chlorine TOTAL-DPD/ F10 Powder Pack straight from the foil into the water sample.
1 2
3 4
1
1
1
10 ml
5 ml
3
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Page 55
47
PM620_4c 07/2015
1.1 Methods
11. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
12. Place the vial in the sample chamber making sure that the marks are aligned.
13. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes: See page 27
Reagent Form of reagent/Quantity Order-No.
Vario Clorine Free-DPD/ F10 Powder Pack / 100 530100
VARIO Chlorine Total-DPD/F10 Powder Pack / 100 530120
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
T1 accepted prepare T2 press TEST
Countdown 3:00
Page 56
48
PM620_4c 07/2015
1.1 Methods
Chlorine dioxide with Tablet
0.02 – 11 mg/l ClO
2
The following selection is shown in the display:
for the determination of Chlorine dioxide in the presence of Chlorine.
for the determination of Chlorine dioxide in the absence of Chlorine.
Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key.
Chlorine dioxide >> with Cl without Cl
>> with Cl
>> without Cl
2
0
1
Page 57
49
PM620_4c 07/2015
1.1 Methods
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Chlorine dioxide may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Chlorine dioxide, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 19 mg/l Chlorine dioxide can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Chlorine dioxide. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. If ??? is displayed at a differentiated test result see page 166.
6. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Chlorine dioxide.
Page 58
50
PM620_4c 07/2015
1.1 Methods
Chlorine dioxide in the presence of Chlorine with Tablet
0.02 – 11 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample.
2. Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod.
3. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
4.
Fill a second clean vial with 10 ml of water sample
and
close tightly with the cap
.
5. Place the vial in the sample chamber making sure that the marks are aligned.
6. Press ZERO key.
7.
Remove the vial from the sample chamber and empty
the vial
.
8. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
9. Transfer the contents of the first vial (Glycine solu-
tion) into the prepared vial (point 8).
10. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
11. Place the vial in the sample chamber making sure that the marks are aligned.
12. Press TEST key.
Zero accepted prepare T1 press TEST
prepare Zero press ZERO
1
2 0
Page 59
51
PM620_4c 07/2015
1.1 Methods
13. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times. Fill with a few drops of water sample.
14. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
17. Place the vial in the sample chamber making sure that the marks are aligned.
18. Press TEST key.
19. Remove the vial from the sample chamber.
20. Add one DPD No. 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
21. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that the marks are aligned.
23. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
as Chlorine dioxide in mg/l ClO2.
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See next page.
T1 accepted prepare T2 press TEST
Countdown 2:00
T2 accepted prepare T3 press TEST
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
*,** mg/l ClO
2
Page 60
52
PM620_4c 07/2015
1.1 Methods
Notes: (Chlorine dioxide in the presence of Chlorine)
1. The conversion factor to convert Chlorine dioxide (display) to Chlorine dioxide as Chlorine units is 2.6315. mg/l ClO2 [Cl] = mg/l ClO2 · 2.6315 Chlorine dioxide displayed as Chlorine units ClO2 [Cl] has its origin in swimming poolwater treatment according to DIN 19643.
2. The total Chlorine result given includes the contribution of the chlorine dioxide as Chlorine units reading. For true Chlorine value add the free and combined Chlorine values.
3. See also page 49.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
GLYCINE Tablet / 100 512170BT
Page 61
53
PM620_4c 07/2015
1.1 Methods
Chlorine dioxide in absence of Chlorine with Tablet
0.02 – 11 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display
as Chlorine dioxide in mg/l ClO2.
Notes:
See page 49
prepare Zero press ZERO
Zero accepted prepare Test press TEST
*,** mg/l ClO2
2
0
1
Page 62
54
PM620_4c 07/2015
1.1 Methods
Chlorite in presence of Chlorine and Chlorine dioxide
0,01 – 6 mg/l Cl
2
Firstly, the glycine method is used to measure the concen­tration of Chlorine Dioxide. This is then followed by the determination of the free and total chlorine, from which the Combined Chlorine can be calculated. A third test is performed which measures the Total Chlorine concentration plus any Chlorite present. Finally, the Chlorite concentration can be calculated from the three recorded results.
The following selection is shown in the display:
select for the determination of free Chlorine.
1. Fill a clean vial with 10 ml of water sample.
2. Add one GLYCINE tablet straight from the foil to the
water sample and crush the tablet using a clean stirring rod.
3. Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved.
4. Fill a second clean vial with 10 ml of water sample, close tightly with the cap.
5. Place the vial in the sample chamber making sure that the marks are aligned.
6. Press ZERO key.
7. Remove the vial from the sample chamber and
empty the vial.
8. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
1
0 0
>> free
Chlorine >> diff free total
prepare Zero press ZERO
Page 63
55
PM620_4c 07/2015
1.1 Methods
9. Transfer the contents of the first vial (Glycine solu­tion) into the prepared vial (point 8).
10. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
11. Place the vial in the sample chamber making sure that the marks are aligned.
12. Press TEST key.
Record the displayed test result (G).
13. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times. Fill with a few drops of water sample.
14. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
17. Place the vial in the sample chamber making sure that the marks are aligned.
18. Press TEST key.
Record the displayed test result (A).
19. Remove the vial from the sample chamber.
20. Add one DPD No. 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
Zero accepted prepare Test press TEST
Zero accepted prepare Test press TEST
Page 64
56
PM620_4c 07/2015
1.1 Methods
21. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that the marks are aligned.
23. Wait for a reaction period of 2 minutes.
24. Press TEST key.
Record the displayed test result (C).
25. Remove the vial from the sample chamber.
26. Add one DPD ACIDIFYING tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
27. Wait for a reaction period of 2 minutes.
28. Add one DPD NEUTRALISING tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
29. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
30. Place the vial in the sample chamber making sure that the marks are aligned.
31. Press TEST key.
Record the displayed test result (D).
Zero accepted prepare Test press TEST
Zero accepted prepare Test press TEST
Page 65
57
PM620_4c 07/2015
1.1 Methods
Calculations:
mg/l Chlorine dioxide = result G x 1,9 mg/l free Chlorine = result A – result G mg/l combined Chlorine = result C – result A mg/l Chlorite = result D – ( result C + 4 x result G )
Tolerances:
1. By calculation of non direct analysable parameters it is necessary to consider the error propagation besed on the possible tolerances of the single test tesults.
2. see Notes Chlorine
Reagent Form of reagent/Quantity Order-No. Set DPD No. 1 / No. 3 Tablet / per 100
inclusive stirring rod
517711BT
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
GLYCINE Tablet / 100 512170BT
DPD ACIDIFYING Tablet / 100 512120
DPD NEUTRALISING Tablet / 100 511020BT
Page 66
58
PM620_4c 07/2015
1.1 Methods
Copper with Tablet
0.05 – 5 mg/l Cu
The following selection is shown in the display:
for the differentiated determination of free, combined and total Copper.
for the determination of free Copper.
for the determination of total Copper.
Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key.
Note:
1. If ??? is displayed at the diffentiated test result see page 166.
Reagent Form of reagent/Quantity Order-No. Set
COPPER No. 1 / No. 2
Tablet / per 100 inclusive stirring rod
517691BT
COPPER No. 1 Tablet / 100 513550BT
COPPER No. 2 Tablet / 100 513560BT
Copper >> diff free total
>> diff
>> free
>> total
5
0
1
Page 67
59
PM620_4c 07/2015
1.1 Methods
Copper, differentiated determination with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
9. Remove the vial from the sample chamber.
10. Add one COPPER No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
11. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
12. Place the vial in the sample chamber making sure that the marks are aligned.
13. Press TEST key.
The result is shown in the display in: mg/l free Copper mg/l combined Copper mg/l total Copper
*,** mg/l free Cu
*,** mg/l comb Cu *,** mg/l total Cu
Zero accepted prepare T1 press TEST
prepare Zero press ZERO
T1 accepted prepare T2 press TEST
5
0
1
Page 68
60
PM620_4c 07/2015
1.1 Methods
Copper, free with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l free Copper.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
5
0
1
Page 69
61
PM620_4c 07/2015
1.1 Methods
Copper, total with Tablet
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one COPPER No. 1 tablet and one COPPER No. 2 tablet straight from the foil to the water sample and crush the tablets using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l total Copper.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
5
0
1
Page 70
62
PM620_4c 07/2015
1.1 Methods
Copper, free (Note 1) with Vario Powder Pack
0.05 – 5 mg/l Cu
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Cu 1 F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (Note 3).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Copper
Countdown 2:00
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
5
3
1
Page 71
63
PM620_4c 07/2015
1.1 Methods
Notes:
1. For determination of total Copper digestion is required.
2. Extremely acid water samples (pH 2 or less) must be adjusted between pH 4 and pH 6 before the reagent is added (with 8 mol/l Potassium hydroxide solution KOH). Caution: pH values above 6 can lead to Copper precipitation.
3. Accuracy is not affected by undissolved powder.
4. Interferences:
Cyanide, CN-Cyanide prevents full colour development.
Add 0.2 ml Formaldehyde to 10 ml water sample and wait for a reaction time of 4 minutes (Cyanide is masked). After this perform test as described. Multiply the result by 1.02 to correct the sample dilution by Formaldehyde.
Silver, Ag
+
If a turbidity remains and turns black, silver interference is likely. Add 10 drops of saturated Potassium chloride solution to 75 ml of
water sample. Filtrate through a fine filter. Use 10 ml of the filtered water sample to perform test.
Reagent Form of reagent/Quantity Order-No.
VARIO Cu 1 F10 Powder Pack / 100 530300
Page 72
64
PM620_4c 07/2015
1.1 Methods
CyA-TEST (Cyanuric acid) with Tablet
0 – 160 mg/l CyA
1. Fill a clean vial (24 mm Ø) with 5 ml of the water sample and 5 ml deionised water (Note 1), close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CyA-TEST tablet straight from the foil to the prepared water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved (Note 2, 3).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l Cyanuric acid.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
6
0
1
Page 73
65
PM620_4c 07/2015
1.1 Methods
Notes:
1. Use deionised water or tap water free of Cyanuric acid.
2. If Cyanuric acid is present a cloudy solution will occur. Small single particles are not necessarily caused by Cyanuric acid.
3. Dissolve the tablet completely (therefore swirl the vial approx. 1 minute). Un-dissolved particles of the tablet can cause results that are too high.
Reagent Form of reagent/Quantity Order-No.
CyA-TEST Tablet / 100 511370BT
Page 74
66
PM620_4c 07/2015
1.1 Methods
H2O2 (Hydrogen peroxide) HR with Liquid Reagent
40 – 500 mg/l H2O
2
Insert the adapter for 16 mm Ø vials.
1. Fill a clean vial (16 mm Ø) with 10 ml of the water sample, close tightly with the cap. (Note 1, 2)
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press the ZERO key.
4. Remove the vial from the sample chamber.
5. Fill the prepared vial with drops of the same size by holding the bottle vertically and squeeze slowly:
6 drops of H2O2-Reagent
6. Close the vial tightly with the cap and invert several times to mix the contents.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press the TEST key.
The result is shown in the display in mg/l H2O2.
1
4
2
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Ø 16 mm
Page 75
67
PM620_4c 07/2015
1.1 Methods
Notes:
1. The hydrogen peroxide is determined in the form of yellow/orange coloured peroxotitanic acids in strongly acidic media. In connection with neutral to weakly alkaline (~pH 10) samples, the acid in the reagent is sufficient in order to produce a medium suitable for measurement. In the case of strongly alkaline samples (pH > 10), the samples must be acidified before measurement otherwise the results may be deficient. This is achieved by diluting the sample with a 5% sulphuric acid solution, for example, at a ratio of 1:1.
In contrast to many other colour reactions, in connection with the presence of hydrogen peroxide, discoloration with long-term stability is achieved that can still be measured after 24 h. Particles in the sample solution or turbidity distort the analysis and must be eliminated by centrifuging or simply filtering the sample solution prior to performing the measurement. Falsification of the measurement results should also be expected in connection with coloured solutions.
2. Oxidising agents such as chlorine, bromine, chlorine dioxide and ozone do not distort the analysis. On the other hand, however, water discoloration does distort the analysis. In this case, proceed as described in the following:
Fill a clean vial (16 mm Ø) with 10 ml of the water sample and perform zero calibration
(see “Operation”).
Measure the sample solution without the addition of drops of reagent (result B).
Then the same sample solution, measured with the addition of the reagent drops
(result A).
Calculations: mg/l H2O2 = result A – result B
3. Attention: The reference reagent contains a 25% sulphuric acid solution. It is recommended to wear appropriate protective clothing (protective goggles/gloves).
Reagent Form of reagent/Quantity Order-No.
H2O2-reagent Liquid reagent / 15 ml 424991
Page 76
68
PM620_4c 07/2015
1.1 Methods
Hardness, Calcium 2T with Tablet
0 – 500 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CALCIO H No. 1 tablet straight from the
foil to the 10 ml water sample, crush the tablet using a clean stirring rod and dissolve the tablet completely.
6. Add one CALCIO H No. 2 tablet straight from the foil
to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several times until the tablet is completely dissolved.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display as Calcium Hardness.
Zero accepted prepare Test press TEST
prepare Zero press ZERO
Countdown 2:00
1
9 1
Page 77
69
PM620_4c 07/2015
1.1 Methods
Notes:
1. To optimise the readings an optional batch related calibration can be performed using Mode 40, see page 138.
2. Strong alkaline or acidic water samples must be adjusted to a pH-value between pH 4 and 10 before the tablets are added (use 1 mol/l Hydrochloride acid resp. 1 mol/l Sodium hydroxide).
3. For accurate test results exactly 10 ml of water sample must be taken for the test.
4. This method was developed from a volumetric procedure for the determination of Calcium Hardness. Due to undefined conditions, the deviations from the standardised method may be greater.
5. The tolerance of the method is increasing with higher concentrations. When diluting samples, this should be taken in account, always measuring in the first third of the range.
6. Interferences:
• Magnesium hardness up to 200 mg/l CaCO3 does not interfere.
• Iron concentration above 10 mg/l may cause low results.
• Zinc concentration above 5 mg/l may cause high results.
7. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No. Set
CALCIO H No. 1 / No. 2
Tablet / per 100 inclusive stirring rod
517761BT
Page 78
70
PM620_4c 07/2015
1.1 Methods
Hardness, total with Tablet
2 – 50 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one HARDCHECK P tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display as total Hardness.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
0
0
2
Page 79
71
PM620_4c 07/2015
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10 before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2. Conversion table:
mg/l CaCO
3
°dH °fH °eH
1 mg/l CaCO
3
---- 0.056 0.10 0.07
1 °dH 17.8 ---- 1.78 1.25
1 °fH 10.0 0.56 ---- 0.70
1 °eH 14.3 0.80 1.43 ----
3. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
HARDCHECK P Tablet / 100 515660BT
Page 80
72
PM620_4c 07/2015
1.1 Methods
Hardness, total HR with Tablet
20 – 500 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 1 ml of the water sample and 9 ml of deionised water, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one HARDCHECK P tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as total Hardness.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
0
1
2
Page 81
73
PM620_4c 07/2015
1.1 Methods
Notes:
1. Strong alkaline or acidic water samples must be adjusted between pH 4 and pH 10 before the tablet is added (use 1 mol/l Hydrochloric acid resp. 1mol/l Sodium hydroxide).
2. Conversion table:
mg/l CaCO
3
°dH °fH °eH
1 mg/l CaCO
3
---- 0.056 0.10 0.07
1 °dH 17.8 ---- 1.78 1.25
1 °fH 10.0 0.56 ---- 0.70
1 °eH 14.3 0.80 1.43 ----
3. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
HARDCHECK P Tablet / 100 515660BT
Page 82
74
PM620_4c 07/2015
1.1 Methods
Iodine with Tablet
0.05 – 3.6 mg/l I
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber, empty the
vial leaving a view drops in.
5. Add one DPD No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l Iodine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1
5
2
Page 83
75
PM620_4c 07/2015
1.1 Methods
Notes:
1. Oxidizing reagents, such as Chlorine, Bromine, etc. interfere as they react in the same way as Iodine.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
Page 84
76
PM620_4c 07/2015
1.1 Methods
Iron (Note 1) with Tablet
0.02 – 1 mg/l Fe
Determination of total dissolved Iron Fe2+ and Fe
3+
*
*This information refers to analysis of the water sample without digestion.
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one IRON LR tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Iron (Fe
2+/3+
).
Notes:
1. This method determines the total dissolved Iron as Fe2+ and Fe3+.
2. The IRON (II) LR tablet is used for differentiation – as described above – instead of the IRON LR tablet. Fe3+ = Fe
2+/3+
– Fe
2+
3. For the determination of total dissolved and undissolved iron digestion is required. An example is described on page 77.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
2
0
2
Page 85
77
PM620_4c 07/2015
1.1 Methods
Notes:
Digestion procedure for the determination of total dissolved and undissolved iron.
1. Add 1 ml of concentrated sulfuric acid to 100 ml water sample. Heat and boil for 10 minutes or until all particles are dissolved. After cooling down, the sample is set to a pH­value of 3 to 6 by using ammonia solution. Refill with deionised water to the previous volume of 100 ml and mix well. 10 ml of this pre-treated solution is used for the following analysis. Perform as described by the selected test method.
2. Water which has been treated with organic compounds like corrosion inhibitors must be oxidised where necessary to break down the iron. Therefore add 1 ml concentrated sulfuric acid and 1 ml concentrated nitric acid to 100 ml water sample and boil to approx. half volume. After cooling down, proceed as described above.
Reagent Form of reagent/Quantity Order-No.
IRON LR Tablet / 100 515370BT
IRON (II) LR Tablet / 100 515420BT
result Bresult A
minus
= Fe
3+
Fe
2+
+ Fe
3+
Fe
2+
Fe
2+
+ Fe
3+
IRON LR
tablet
IRON (II) LR
tablet
IRON LR
tablet
dissolved Iron
dissolved and
undissolved Iron
Filtration Digestion
Iron Determination
Page 86
78
PM620_4c 07/2015
1.1 Methods
Oxygen, active* with Tablet
0.1 – 10 mg/l O
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one DPD No. 4 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l active Oxygen.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
9
0
2
Page 87
79
PM620_4c 07/2015
1.1 Methods
Notes: * Active Oxygen is a synonym for a common disinfectant (based on “Oxygen“)
in Swimming Pool Treatment.
1. When preparing the sample, the lost of Oxygen, e.g. by pipetting or shaking, must be avoided.
2. The analysis must take place immediately after taking the sample.
Reagent Form of reagent/Quantity Order-No.
DPD No. 4 Tablet / 100 511220BT
Page 88
80
PM620_4c 07/2015
1.1 Methods
Ozone with Tablet
0.02 – 2 mg/l O
3
The following selection is shown in the display:
for the determination of Ozone in the presence of Chlorine.
for the determination of Ozone in the absence of Chlorine.
Select the desired method with the arrow keys [ ] and [ ]. Confirm with [ ] key.
Ozon >> with Cl without Cl
>> with Cl
>> without Cl
0
0
3
Page 89
81
PM620_4c 07/2015
1.1 Methods
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Ozone may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Ozone, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 6 mg/l Ozone can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Ozone. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. If ??? is displayed at the diffentiated test result see page 166.
6. Oxidising agents such as Bromine, Chlorine etc. interfere as they react in the same way as Ozone.
Reagent Form of reagent/Quantity Order-No. Set
DPD No. 1 / No. 3
Tablet / per 100 inclusive stirring rod
517711BT
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
GLYCINE Tablet / 100 512170BT
Page 90
82
PM620_4c 07/2015
1.1 Methods
Ozone, in the presence of Chlorine with Tablet
0.02 – 2 mg/l O
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
10. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times.
11. Fill a second clean vial with 10 ml of water sample.
12. Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod.
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Countdown 2:00
3
0 0
Page 91
83
PM620_4c 07/2015
1.1 Methods
13. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
14. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil into the first cleaned vial and crush the tablets using a clean stirring rod.
15. Transfer the contents of the second vial (Glycine
solution) into the prepared vial (point 14).
16. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
17. Place the vial in the sample chamber making sure that the marks are aligned.
18. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l Ozone
mg/l total Chlorine
Notes:
See page 81
Countdown 2:00
T1 accepted prepare T2 press TEST
*,** mg/l O
3
*,** mg/l total Cl
Page 92
84
PM620_4c 07/2015
1.1 Methods
Ozone, in absence of Chlorine with Tablet
0.02 – 2 mg/l O
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in
mg/l Ozone.
Notes: See page 81.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
0
0
3
Page 93
85
PM620_4c 07/2015
Page 94
86
PM620_4c 07/2015
1.1 Methods
PHMB (Biguanide) with Tablet
2 – 60 mg/l PHMB
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one PHMB PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l PHMB.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
7
0
Page 95
87
PM620_4c 07/2015
1.1 Methods
Notes:
1. Clean vials with the brush immediately after analysis.
2. Vials and stirring rods may turn blue after prolonged use. In this case clean vials and stirring rods with a laboratory detergent (see chapter 1.2.2 Cleaning of vials and accessories for analysis). Rinse vials and caps thoroughly with tap water and then with deionised water.
3. The test result is influenced by Hardness and Total Alkalinity. The calibration of this method was done using water with the following concentration: Ca-Hardness: 200 mg/l CaCO3 Total Alkalinity: 120 mg/l CaCO
3
Reagent Form of reagent/Quantity Order-No.
PHMB PHOTOMETER Tablet / 100 516100BT
Page 96
88
PM620_4c 07/2015
1.1 Methods
Phosphate, ortho LR with Tablet
0.05 – 4 mg/l PO
4
1. Fill a clean vial (24 mm ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one PHOSPHATE No. 1 LR tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Add one PHOSPHATE No. 2 LR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the marks aligned.
9. Press TEST key.
Wait for a reaction period of 10 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display as ortho-Phosphate.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 10:00
3
91
Page 97
89
PM620_4c 07/2015
1.1 Methods
Notes
1. Only ortho-Phosphate ions PO
4
3-
react.
2. The tablets must be added in the correct sequence.
3. The test sample should have a pH-value between 6 and 7.
4. Interferences: Higher concentrations of Cu, Ni, Cr (III), V (V) and W (VI) interfere due to their colour. Silicates do not interfere (masked by Citric acid in the tablets).
5. Conversion: mg/l P = mg/l PO4 x 0.33 mg/l P2O5= mg/l PO4 x 0.75
6. PO4 P
P2O
5
Reagent Form of reagent/Quantity Order-No. Set
PHOSPHATE No. 1 / No. 2 LR
Tablet / per 100 inclusive stirring rod
517651BT
PHOSPHATE No. 1 LR Tablet / 100 513040BT
PHOSPHATE No. 2 LR Tablet / 100 513050BT
Page 98
90
PM620_4c 07/2015
1.1 Methods
pH value LR 5.2 – 6.8 with Tablet
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one BROMOCRESOLPURPLE PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as pH-value.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
2
9
3
Page 99
91
PM620_4c 07/2015
1.1 Methods
Notes:
1. For photometric determination of pH values only use BROMOCRESOLPURPLE tablets in black printed foil pack and marked with PHOTOMETER.
2. pH values below 5.2 and above 6.8 can produce results inside the measuring range. A plausibility test (pH-meter) is recommended.
3. The accuracy of the colorimetric determination of pH-values depends on various boundary conditions (buffer capacity of the sample, salt contents etc.).
4. Salt error Correction of test results (average values) for samples with salt contents of:
Indicator Salt content
Bromcresolpurple 1 molar
– 0.26
2 molar
– 0.33
3 molar
– 0.31
The values of Parson and Douglas (1926) are based on the use of Clark and Lubs buffers. 1 Mol NaCl = 58.4 g/l = 5.8 %
Reagent Form of reagent/Quantity Order-No.
BROMOCRESOLPURPLE PHOTOMETER Tablet / 100 515700BT
Page 100
92
PM620_4c 07/2015
1.1 Methods
pH value 6.5 – 8.4 with Tablet
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one PHENOL RED PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as pH-value.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
0
3
Loading...