Before you begin:
To condition electrode, immerse electrode in electrode storage solution or tap water for at least
30 minutes before use. DO NOT use de-ionized water.
Calibration
Calibration is not necessary unless exact readout agreement with a work standard and
at a specific ORP value is needed. To achieve this, use the following procedure:
1. Switch unit on (ON/OFF button)
2. Select a known solution sample from the actual process as near the critical ORP value as
possible, for example, 470 mV.
3. Dip the conditioned electrode into this solution for 2 to 5 minutes until the readings stabilize.
Ensure the whole sensor is immersed.
4. Press CAL button to enter Calibration mode. The screen will switch to a dual display with the
upper display flashing the relative mV value to be calibrated and the lower display showing the
default measurement value of the sample. The ‘CAL’ indicator will also be shown on the screen
replacing ‘MEAS’ to indicate that the tester is in calibration mode.
Note: To abort calibration, press the ‘CAL’ button to return to the measurement mode.
5. Immediately press and hold the HOLD/ENT button to scroll the reading to match value of
the known solution sample. The offset adjustment is up to +/- 150mV from the default measurement.
6. To complete the calibration, release the HOLD/ENT button and wait for 5 seconds for automatic
confirmation. The primary display will stop flashing and the secondary display will automatically
show ‘CO’ before the screen return to the single display measurement mode. The mV annunciator will now changed to ‘R.mV’ indicating the value displayed is a relative mV value of the
default measurement.
7. Rinse electrode in tap water before measuring other solutions.
ORP Measurement
1. Remove cap from the electrode and press the ON/OFF button to switch the tester on.
2. Dip the electrode about 2 to 3 cm into the test solution. Stir once and let the reading stabilize
for 2 to 5 minutes.
3. Note the ORP value or press HOLD/ENT button to freeze the reading. Press HOLD/ENT again
to release the reading.
4. Press ON/OFF to turn off tester. If you do not press a button for 8.5 minutes, the tester will
automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation
1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be displayed and
the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be displayed
anymore indicating the held measurement is released.
User Reset
You can reset the ORP calibration to the factory default by using the user reset function.
1. Switch off the tester.
2. While pressing the ‘Cal’ button, press and release the ON/OFF button to enter the ‘User
Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with a flashing ‘nO’
selection.
3. Use the ‘CAL’ button to toggle between ‘nO’ and ‘YES’ selection.
• nO deactivates reset selection
• YES activates the reset selection
4. Press the HOLD/ENT button to confirm the selection made.
Electrode Maintenance
1. Rinse the electrode with tap water or electrode storage solution after each measurement.
2. In aggressive chemicals, dir ty or viscous solutions, and solutions with heavy metals or proteins,
take readings quickly and rinse electrode immediately afterward.
3. If possible, keep a small piece of paper or sponge in the electrode cap – moistened with
clean water or electrode storage solution (NOT de-ionized water) – and close the cap over the
electrode.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity (shown in diagram below).