Lovibond MD 610 Instruction Manual

Page 1
Photometer
1
Instruction manual
GB
Lovibond® Water Testing
Tintometer® Group
www.lovibond.com
Page 2
Page 3
Important steps before using the photometer
Please carry out the following steps as described in the Instruction manual. Become familiar with your new pho­tometer before starting with the first tests:
• Unpacking and inspection of delivery contents, see page 344.
• Install the batteries, see page 288, 289.
Perform the following settings in the Mode-Menu; Instruction manual from page 301 and following:
• MODE 10: select language
• MODE 12: set date and time
• MODE 34: perform “Delete data“
• MODE 69: perform “User m. init” to initialise the user polynomial system
If required set other functions.
Page 4
Page 5
Wichtige Information
Um die Qualität unserer Umwelt zu erhalten, beschützen und zu verbessern
Entsorgung von elektronischen Geräten in der Europäischen Union
Aufgrund der Europäischen Verordnung 2012/19/EU darf Ihr elektronisches Gerät
nicht mit dem normalen Hausmüll entsorgt werden!
Tintometer GmbH entsorgt ihr elektrisches Gerät auf eine professionelle und für die Umwelt verantwortungsvolle Weise. Dieser Service ist, die Transportkosten
nicht inbegriffen, kostenlos. Dieser Service gilt ausschließlich für elektrische Geräte
die nach dem 13.08.2005 erworben wurden. Senden Sie Ihre zu entsorgenden
Tintometer Geräte frei Haus an Ihren Lieferanten.
Important Information
To Preserve, Protect and Improve the Quality of the Environment
Disposal of Electrical Equipment in the European Union
Because of the European Directive 2012/19/EU your electrical instrument must not
be disposed of with normal household waste!
Tintometer GmbH will dispose of your electrical instrument in a professional
and environmentally responsible manner. This service, excluding the cost of
transportation is free of charge. This service only applies to electrical instruments
purchased after 13th August 2005. Send your electrical Tintometer instruments for
disposal freight prepaid to your supplier.
Notice importante
Conserver, protéger et optimiser la qualité de l’environnement
Élimination du matériel électrique dans l’Union Européenne
Conformément à la directive européenne nº 2012/19/UE, vous ne devez plus jeter
vos instruments électriques dans les ordures ménagères ordinaires !
La société Tintometer GmbH se charge d’éliminer vos instruments électriques de
façon professionnelle et dans le respect de l’environnement. Ce service, qui ne
comprend pas les frais de transport, est gratuit. Ce service n’est valable que
pour des instruments électriques achetés après le 13 août 2005. Nous vous prions
d’envoyer vos instruments électriques Tintometer usés à vos frais à votre fournisseur.
Belangrijke informatie
Om de kwaliteit van ons leefmilieu te behouden, te verbeteren en te
beschermen is voor landen binnen de Europese Unie de Europese richtlijn
2012/19/EU voor het verwijderen van elektronische apparatuur opgesteld.
Volgens deze richtlijn mag elektronische apparatuur niet met het huishoudelijk afval
worden afgevoerd.
Tintometer GmbH verwijdert uw elektronisch apparaat op een professionele en
milieubewuste wijze. Deze service is, exclusief de verzendkosten, gratis en alleen geldig voor elektrische apparatuur die na 13 augustus 2005 is gekocht. Stuur uw te
verwijderen Tintometer apparatuur franco aan uw leverancier.
DE
GB
FR
NL
Page 6
Información Importante
Para preservar, proteger y mejorar la calidad del medio ambiente
Eliminación de equipos eléctricos en la Unión Europea
Con motivo de la Directiva Europea 2012/19/UE, ¡ningún instrumento eléctrico
deberá eliminarse junto con los residuos domésticos diarios!
Tintometer GmbH se encargará de dichos instrumentos eléctricos de una manera
profesional y sin dañar el medio ambiente. Este servicio, el cual escluye los gastos de transporte, es gratis y se aplicará únicamente a aquellos instrumentos eléctricos
adquiridos después del 13 de agosto de 2005. Se ruega enviar aquellos instrumentos
eléctricos inservibles de Tintometer a carga pagada a su distribuidor.
Informazioni importanti
Conservare, proteggere e migliorare la qualità dell’ambiente
Smaltimento di apparecchiature elettriche nell’Unione Europea
In base alla Direttiva europea 2012/19/UE, gli apparecchi elettrici non devono essere
smaltiti insieme ai normali rifiuti domestici!
Tintometer GmbH provvederà a smaltire i vostri apparecchi elettrici in maniera
professionale e responsabile verso l’ambiente. Questo servizio, escluso il trasporto,
è completamente gratuito. Il servizio si applica agli apparecchi elettrici acquistati
successivamente al 13 agosto 2005. Siete pregati di inviare gli apparecchi elettrici
Tintometer divenuti inutilizzabili a trasporto pagato al vostro rivenditore.
Informação Importante
Para Preservar, Proteger e Melhorar a Qualidade do Ambiente
Remoção de Equipamento Eléctrico na União Europeia
Devido à Directiva Europeia 2012/19/UE, o seu equipamento eléctrico naõ deve ser
removido com o lixo doméstico habitual!
A Tintometer GmbH tratará da remoção do seu equipamento eléctrico de forma
profissional e responsável em termos ambientais. Este serviço, não incluindo
os custos de transporte, é gratuito. Este serviço só é aplicável no caso de
equipamentos eléctricos comprados depois de 13 de Agosto de 2005. Por favor,
envie os seus equipamentos eléctricos Tintometer que devem ser removidos ao seu
fornecedor (transporte pago).
Istotna informacja
Dla zachowania, ochrony oraz poprawy naszego środowiska
Usuwanie urządzeń elektronicznych w Unii Europejskiej
Na podstawie Dyrektywy Parlamentu Europejskiego 2012/19/UE nie jest
dozwolone usuwanie zakupionych przez Państwo urządzeń elektronicznych wraz z
normalnymi odpadami z gospodarstwa domowego!
Tintometer GmbH usunie urządzenia elektrycznego Państwa w sposób
profesjonalny i odpowiedzialny z punktu widzenia środowiska. Serwis ten jest,
za wyjątkiem kosztów transportu, bezpłatny. Serwis ten odnosi się wyłącznie do
urządzeń elektrycznych zakupionych po 13.08.2005r. Przeznaczone do usunięcia
urządzenia rmy Tintometer mogą Państwo przesyłać na koszt własny do swojego
dostawcy.
ES
IT
PT
PL
Page 7
Wichtiger Entsorgungshinweis zu Batterien und Akkus
Jeder Verbraucher ist aufgrund der Batterieverordnung (Richtlinie 2006/66/
EG) gesetzlich zur Rückgabe aller ge- und verbrauchten Batterien bzw. Akkus
verpflichtet. Die Entsorgung über den Hausmüll ist verboten. Da auch bei Produkten
aus unserem Sortiment Batterien und Akkus im Lieferumgang enthalten sind, weisen
wir Sie auf folgendes hin:
Verbrauchte Batterien und Akkus gehören nicht in den Hausmüll, sondern können
unentgeltlich bei den öffentlichen Sammelstellen Ihrer Gemeinde und überall
dort abgegeben werden, wo Batterien und Akkus der betreffenden Art verkauft
werden. Weiterhin besteht für den Endverbraucher die Möglichkeit, Batterien und
Akkus an den Händler, bei dem sie erworben wurden, zurückzugeben (gesetzliche
Rücknahmepflicht).
Important disposal instructions for batteries and accumulators
EC Guideline 2006/66/EC requires users to return all used and worn-out batteries
and accumulators. They must not be disposed of in normal domestic waste. Because
our products include batteries and accumulators in the delivery package our advice is
as follows :
Used batteries and accumulators are not items of domestic waste. They must be
disposed of in a proper manner. Your local authority may have a disposal facility;
alternatively you can hand them in at any shop selling batteries and accumulators.
You can also return them to the company which supplied them to you; the company
is obliged to accept them.
Information importante pour l'élimination des piles et des accumulateurs
En vertu de la Directive européenne 2006/66/CE relative aux piles et accumulateurs,
chaque utilisateur est tenu de restituer toutes les piles et tous les accumulateurs
utilisés et épuisés. L'élimination avec les déchets ménagers est interdite. Etant donné
que l'étendue de livraison des produits de notre gamme contient également des piles
et des accumulateurs, nous vous signalons ce qui suit :
les piles et les accumulateurs utilisés ne sont pas des ordures ménagères, ils peuvent
être remis sans frais aux points de collecte publics de votre municipalité et partout où sont vendus des piles et accumulateurs du type concerné. Par ailleurs, l'utilisateur final a la possibilité de remettre les piles et les accumulateurs au commerçant auprès
duquel ils ont été achetés (obligation de reprise légale).
Belangrijke mededeling omtrent afvoer van batterijen en accu’s
Ledere verbruiker is op basis van de richtlijn 2006/66/EG verplicht om alle gebruikte
batterijen en accu’s in te leveren. Het is verboden deze af te voeren via het huisvuil.
Aangezien ook onze producten geleverd worden met batterijen en accu’s wijzen wij
u op het volgende; Lege batterijen en accu’s horen niet in het huisvuil thuis. Men
kan deze inleveren bij inzamelpunten van uw gemeente of overal daar waar deze
verkocht worden. Tevens bestaat de mogelijkheid batterijen en accu’s daar in te
leveren waar u ze gekocht heeft. (wettelijke terugnameplicht)
DE
GB
FR
NL
Page 8
Indicación importante acerca de la eliminación de pilas y acumuladores
Basado en la norma relativa a pilas/ baterías (directiva 2006/66/CE), cada consumidor,
está obligado por ley, a la devolución de todas las pilas/ baterías y acumuladores usados y
consumidos. Está prohibida la eliminación en la basura doméstica. Ya que en productos de
nuestra gama, también se incluyen en el suministro pilas y acumuladores, le sugerimos lo
siguiente:
Las pilas y acumuladores usados no pertenecen a la basura doméstica, sino que pueden
ser entregados en forma gratuita en cada uno de los puntos de recolección públicos de su
comunidad en los cuales se vendan pilas y acumuladores del tipo respectivo. Además, para
el consumidor final existe la posibilidad de devolver las pilas y baterías recargables a los
distribuidores donde se hayan adquirido (obligación legal de devolución).
Indicazioni importanti sullo smaltimento di pile e accumulatori
In base alla normativa concernente le batterie (Direttiva 2006/66/CE) ogni consumatore
è tenuto per legge alla restituzione di tutte le batterie o accumulatori usati ed esauriti. È
vietato lo smaltimento con i rifiuti domestici. Dato che anche alcuni prodotti del nostro assortimento sono provvisti di pile e accumulatori, vi diamo di seguito delle indicazioni:
Pile e accumulatori esauriti non vanno smaltiti insieme ai rifiuti domestici, ma depositati
gratuitamente nei punti di raccolta del proprio comune o nei punti vendita di pile e
accumulatori dello stesso tipo. Inoltre il consumatore finale può portare batterie e
accumulatori al rivenditore presso il quale li ha acquistati (obbligo di raccolta previsto per
legge).
Instruções importantes para a eliminação residual de pilhas e acumuladores
Os utilizadores finais são legalmente responsáveis, nos termos do Regulamento relativo
a pilhas e acumuladores (Directiva 2006/66/CE), pela entrega de todas as pilhas e
acumuladores usados e gastos. É proibida a sua eliminação juntamente com o lixo
doméstico. Uma vez que determinados produtos da nossa gama contêm pilhas e/ou
acumuladores, alertamos para os seguintes aspectos:
As pilhas e acumuladores usados não podem ser eliminados com o lixo doméstico, devendo sim ser entregues, sem encargos, junto dos pontos de recolha públicos do seu município, ou em qualquer ponto de venda de pilhas e acumuladores. O utilizador final
dispõe ainda da possibilidade de entregar as pilhas e/ou acumuladores no estabelecimento
comerciante onde os adquiriu (dever legal de aceitar a devolução).
Istotna wskazówka dotycząca utylizacji baterii i akumulatorów
Każdy użytkownik na mocy rozporządzenia w sprawie baterii (wytyczna 2006/66/WE)
jest ustawowo zobowiązany do oddawania wszystkich rozładowanych i zużytych baterii
lub akumulatorów. Utylizacja wraz z odpadkami domowymi jest zabroniona. Ponieważ
także w produktach z naszego asortymentu zawarte są w zakresie dostawy baterie i
akumulatory, zwracamy uwagę na poniższe zasady:
zużyte baterie i akumulatory nie mogą być wyrzucane wraz z odpadkami domowymi,
lecz powinny być bezpłatnie przekazywane w publicznych miejscach zbiórki
wyznaczonych przez gminę lub oddawane w punktach, gdzie sprzedawane są baterie i
akumulatory danego rodzaju. Poza tym użytkownik końcowy ma możliwość zwrócenia baterii i akumulatorów do przedstawiciela handlowego, u którego je nabył (ustawowy
obowiązek przyjęcia).
ES
IT
PT
PL
Page 9
1
MD 610_1c 09/2015
Safety precautions
CAUTION
Reagents are formulated exclusively for chemical analysis and must not be used for
any other purpose. Reagents must not get into the hands of children. Some of the
reagents contain substances which are not entirely harmless environmentally. Be
aware of the ingredients and take proper care when disposing of the test solution.
CAUTION
Please read this instruction manual before unpacking, setting up or using the
photometer. Please read the method description completely before performing
the test. Be aware of the risks of using the required reagents by reading the MSDS
(Material Safety Data Sheets). Failure could result in serious injury to the operator or
damage to the instrument.
MSDS: www.lovibond.com
CAUTION
The accuracy of the instrument is only valid if the instrument is used in an
environment with controlled electromagnetic disturbances according to DIN 61326.
Wireless devices, e.g. wireless phones, must not be used near the instrument.
Revision_1c 09/2015
Page 10
2
MD 610_1c 09/2015
Table of contents
Part 1 Methods ................................................................................................... 9
1.1 Table of Methods ................................................................................................... 10
Acid demand to pH 4.3 ................................................................................................... 16
Alkalinity-total (Alkalinity-m, m-Value) ............................................................................. 18
Alkalinity-total HR (Alkalinity-m HR, m-Value HR) ............................................................ 20
Alkalinity-p (p-Value) ....................................................................................................... 22
Aluminium with tablets ................................................................................................... 24
Aluminium (powder pack) ............................................................................................... 26
Ammonia with tablet ...................................................................................................... 28
Ammonia (powder pack) ................................................................................................. 30
Ammonia, low range (LR) ................................................................................................ 32
Ammonia, high range (HR) .............................................................................................. 34
Boron ............................................................................................................................. 36
Bromine with tablet ........................................................................................................ 38
Bromine (powder pack) ................................................................................................... 40
Chloride with tablet ........................................................................................................ 42
Chloride with Liquid Reagent .......................................................................................... 44
Chlorine .......................................................................................................................... 46
Chlorine with tablet ........................................................................................................ 48
free Chlorine ............................................................................................................... 48
total Chlorine .............................................................................................................. 49
differentiated determination (free, combined, total) .................................................... 50
Chlorine HR with tablet ................................................................................................... 52
free Chlorine ............................................................................................................... 52
total Chlorine .............................................................................................................. 53
differentiated determination (free, combined, total) .................................................... 54
Chlorine with liquid reagent ............................................................................................ 56
free Chlorine ............................................................................................................... 56
total Chlorine .............................................................................................................. 57
differentiated determination (free, combined, total) .................................................... 58
Chlorine (powder pack) ................................................................................................... 60
free Chlorine ............................................................................................................... 60
total Chlorine .............................................................................................................. 61
differentiated determination (free, combined, total) .................................................... 62
Chlorine HR (powder pack) ............................................................................................. 64
Page 11
3
MD 610_1c 09/2015
free Chlorine ............................................................................................................... 64
total Chlorine .............................................................................................................. 65
differentiated determination (free, combined, total) .................................................... 66
Chlorine dioxide with tablet ............................................................................................ 68
in presence of Chlorine ............................................................................................... 70
in absence of Chlorine ................................................................................................ 73
Chlorine dioxide (powder pack) ....................................................................................... 74
in absence of Chlorine ................................................................................................ 74
in presence of Chlorine ............................................................................................... 75
Chlorine HR (KI) .............................................................................................................. 78
Chlorite in presence of Chlorine and Chlorine dioxide ..................................................... 80
Chromium with Powder Pack .......................................................................................... 84
differentiated determination ....................................................................................... 86
Chromium (VI) ............................................................................................................ 88
Chromium, total (Cr(III) + Cr(VI)) .................................................................................. 89
COD, low range (LR) ....................................................................................................... 90
COD, middle range (MR) ................................................................................................. 92
COD, high range (HR) ..................................................................................................... 94
Colour, true and apparent ............................................................................................... 96
Copper with tablet .......................................................................................................... 98
differentiated determination (free, combined, total) .................................................... 99
free Copper .............................................................................................................. 100
total Copper ............................................................................................................. 101
Copper with liquid reagent and powder ........................................................................ 102
differentiated determination (free, combined, total) .................................................. 103
free Copper .............................................................................................................. 105
total Copper ............................................................................................................. 106
Copper (powder pack) .................................................................................................. 108
Cyanide ........................................................................................................................ 110
CyA-TEST (Cyanuric acid) .............................................................................................. 112
DEHA T ......................................................................................................................... 114
DEHA PP ....................................................................................................................... 116
Fluoride ........................................................................................................................ 118
H2O2 with tablet reagent ............................................................................................... 120
H2O2 LR with liquid reagent ........................................................................................... 122
H2O2 HR with liquid reagent .......................................................................................... 124
Hardness, Calcium with Calcheck tablet ........................................................................ 126
Page 12
4
MD 610_1c 09/2015
Hardness, Calcium with Calcio tablet ............................................................................ 128
Hardness, total .............................................................................................................. 130
Hardness, total HR ........................................................................................................ 132
Hydrazine ..................................................................................................................... 134
Hydrazine with liquid reagent ....................................................................................... 136
Hydrazine with Vacu-vials .............................................................................................. 138
Hydrogen peroxide see H2O2 ........................................................................................ 140
Iodine ........................................................................................................................... 140
Iron ............................................................................................................................... 142
Iron with tablet ......................................................................................................... 144
Iron (powder pack) .................................................................................................... 146
Iron TPTZ (powder pack) ............................................................................................ 148
Iron (Fe in Mo, powder pack) .................................................................................... 150
Iron LR with liquid reagent ........................................................................................ 152
Iron LR 2 with liquid reagent ..................................................................................... 156
Iron HR with liquid reagent ....................................................................................... 160
Manganese with tablet ................................................................................................. 164
Manganese LR (powder pack) ....................................................................................... 166
Manganese HR (powder pack) ...................................................................................... 168
Manganese with liquid reagent ..................................................................................... 170
Molybdate with tablet ................................................................................................... 172
Molybdate LR (powder pack) ......................................................................................... 174
Molybdate HR (powder pack) ........................................................................................ 176
Molybdate HR with liquid reagent ................................................................................. 178
Nickel with tablet .......................................................................................................... 180
Nitrate with tablet and powder ..................................................................................... 182
Nitrate (tube test) .......................................................................................................... 184
Nitrite with tablet .......................................................................................................... 186
Nitrite LR (powder pack) ................................................................................................ 188
Nitrogen, total LR .......................................................................................................... 190
Nitrogen, total HR ......................................................................................................... 192
Oxygen, active .............................................................................................................. 196
Oxygen, dissolved ......................................................................................................... 198
Ozone ........................................................................................................................... 200
in presence of Chlorine ............................................................................................. 202
in absence of Chlorine .............................................................................................. 204
PHMB (Biguanide) ......................................................................................................... 206
Page 13
5
MD 610_1c 09/2015
Phosphate ..................................................................................................................... 208
Phosphate, ortho LR with tablet ................................................................................ 210
Phosphate, ortho HR with tablet ............................................................................... 212
Phosphate, ortho (powder pack) ............................................................................... 214
Phosphate, ortho (tube test) ...................................................................................... 216
Phosphate 1, ortho C ................................................................................................ 218
Phosphate 2, ortho C ................................................................................................ 220
Phosphate, hydrolyzable (tube test) ........................................................................... 222
Phosphate, total (tube test) ....................................................................................... 224
Phosphate LR with Liquid reagent ............................................................................. 226
Phosphate HR with Liquid reagent ............................................................................. 230
Phosphonates ............................................................................................................... 234
pH value LR with tablet ................................................................................................. 238
pH value with tablet ...................................................................................................... 240
pH value with liquid reagent ......................................................................................... 242
pH value HR with tablet ................................................................................................ 244
Polyacrylate with liquid reagent ..................................................................................... 246
Potassium ..................................................................................................................... 250
Silica ............................................................................................................................. 252
Silica LR PP .................................................................................................................... 254
Silica HR PP ................................................................................................................... 256
Silica with liquid reagent and powder ........................................................................... 258
Sodium hypochlorite ..................................................................................................... 260
Sulfate with tablet ........................................................................................................ 262
Sulfate (powder pack) ................................................................................................... 264
Sulfide .......................................................................................................................... 266
Sulfite ........................................................................................................................... 268
Suspended Solids .......................................................................................................... 270
Turbidity ....................................................................................................................... 272
Triazole / Benzotriazole (powder pack)........................................................................... 274
Urea ............................................................................................................................. 276
Zinc with tablet ............................................................................................................. 278
Zinc with liquid reagent and powder ............................................................................. 280
1.2 Important notes ......................................................................................... 282
1.2.1 Correct use of reagents ................................................................................ 282
1.2.2 Cleaning of vials and accessories for analysis ................................................ 283
Page 14
6
MD 610_1c 09/2015
1.2.3 Guidelines for photometric measurements ................................................... 283
1.2.4 Sample dilution techniques ........................................................................... 285
1.2.5 Correcting for volume additions ................................................................... 285
Part 2 Instrument Manual ....................................................................... 287
2.1 Operation ................................................................................................... 288
2.1.1 Set up .......................................................................................................... 288
2.1.2 Saving data – Important Notes ..................................................................... 288
2.1.3 Replacement of batteries .............................................................................. 288
2.1.4 Instrument (explosion drawing) .................................................................... 289
2.2 Overview of function keys ....................................................................... 291
2.2.1 Overview ...................................................................................................... 291
2.2.2 Displaying time and date .............................................................................. 292
2.2.3 User countdown ........................................................................................... 292
2.3 Operation mode ........................................................................................ 293
2.3.1 Automatic switch off .................................................................................... 293
2.3.2 Selecting a method ...................................................................................... 293
2.3.2.1 Method Information (F1) .............................................................................. 293
2.3.2.2 Chemical Species Information ....................................................................... 294
2.3.3 Differentiation .............................................................................................. 294
2.3.4 Performing Zero ........................................................................................... 294
2.3.5 Performing Tests ........................................................................................... 295
2.3.6 Ensuring reaction periods (countdown) ......................................................... 295
2.3.7 Changing chemical species ........................................................................... 296
2.3.8 Storing results .............................................................................................. 296
2.3.9 Perform additional measurements ................................................................ 297
2.3.10 Selecting a new method ............................................................................... 297
2.3.11 Measure absorbance .................................................................................... 298
2.4 Bluetooth® ................................................................................................. 299
2.5 Internet Updates ....................................................................................... 300
2.6 Mode Functions ......................................................................................... 301
Schema ....................................................................................................... 301
2.6.1 Instrument basic settings .............................................................................. 302
2.6.2 Data transfer of stored results....................................................................... 310
2.6.3 Recall / delete stored results.......................................................................... 314
2.6.4 Calibration ................................................................................................... 319
Calcium Hardness Method 191 .................................................................... 319
Page 15
7
MD 610_1c 09/2015
Fluoride Method 170 .................................................................................... 321
User Calibration ........................................................................................... 323
Store user calibration .................................................................................... 325
Delete user calibration .................................................................................. 326
2.6.5 Lab function ................................................................................................. 327
Profi-Mode ................................................................................................... 327
One Time Zero ............................................................................................. 328
2.6.6 User operations ............................................................................................ 329
User method list ........................................................................................... 329
User Concentration Methods ........................................................................ 331
User Polynomials .......................................................................................... 333
Delete User Methods .................................................................................... 336
Print Data of User Methods .......................................................................... 337
Initialise User Method System ....................................................................... 338
2.6.7 Langelier Saturation Index ............................................................................ 339
Selection of temperature unit ....................................................................... 340
2.6.8 Photometer-Information ............................................................................... 341
Part 3 Enclosure .............................................................................................. 343
3.1 Unpacking .................................................................................................. 344
3.2 Delivery contents ...................................................................................... 344
3.3 Technical data ............................................................................................ 345
3.4 Abbreviations ............................................................................................ 346
3.5 Troubleshooting ........................................................................................ 347
3.5.1 Operating messages in the display / error display .......................................... 347
3.5.2 General ........................................................................................................ 349
3.6 Declaration of CE-Conformity .................................................................. 350
3.7 Copyright and Trademark Notice ............................................................. 351
Page 16
8
MD 610_1c 09/2015
Page 17
9
MD 610_1c 09/2015
Part 1
Methods
Page 18
10
MD 610_1c 09/2015
1.1 Table of Methods
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range; Vacu-vial® is a registered trade mark of CHEMetrics Inc.
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
20 Acid demand to
pH 4.3 T
tablet 0.1-4 mmol/l
Acid/Indicator
1,2,5
610 16
30
Alkalinity, total T
tablet 5-200 mg/l CaCO3Acid/Indicator
1,2,5
610 18
31
Alkalinity HR, total T
tablet 5-500 mg/l CaCO3Acid/Indicator
1,2,5
610 20
35 Alkalinity-p T tablet 5-300 mg/l CaCO3Acid/Indicator
1,2,5
560 22
40 Aluminium T tablet 0.01-0.3 mg/l Al Eriochrome
Cyanine R
2
530 24
50 Aluminium PP PP + liquid 0.01-
0.25
mg/l Al Eriochrome
Cyanine R
2
530 26
60 Ammonia T tablet 0.02-1 mg/l N
Indophenol blue
2,3
610 28
62 Ammonia PP PP 0.01-0.8 mg/l N Salicylate
2
660 30
65
Ammonia LR TT
tube test 0.02-2.5 mg/l N Salicylate
2
660 32
66
Ammonia HR TT
tube test 1-50 mg/l N Salicylate
2
660 34
85 Boron T tablet 0.1-2 mg/l B Azomethine
3
430 36
80 Bromine T tablet 0.05-13 mg/l Br
2
DPD
5
530 38
81 Bromine PP PP 0.05-4.5 mg/l Br
2
DPD
1,2
530 40
90 Chloride T tablet 0.5 -25 mg/l Cl
-
Silver nitrate/ turbidity
530 42
92 Chloride L liquid 0.5-20 mg/l Cl
-
Mercurythiocyanate/
Iron nitrate
430 44
100 Chlorine T * tablet 0.01-6 mg/l Cl
2
DPD
1,2,3
530 46, 48
103 Chlorine HR T * tablet 0.1-10 mg/l Cl
2
DPD
1,2,3
530 46, 52
101 Chlorine L * liquid 0.02-4 mg/l Cl
2
DPD
1,2,3
530 46, 56
110 Chlorine PP * PP 0.02-2 mg/l Cl
2
DPD
1,2
530 46, 60
111 Chlorine HR PP * PP 0.1-8 mg/l Cl
2
DPD
1,2
530
46, 64
120 Chlorine dioxide T tablet 0.02-11 mg/l ClO
2
DPD, Glycine
1,2
530 68
122
Chlorine dioxide PP
PP 0.04-3.8 mg/l ClO
2
DPD
1,2
530 74
105 Chlorine HR (Kl) T tablet 5-200 mg/l Cl
2
KI/Acid
5
530 78
125 Chromium PP PP 0.02-2 mg/l Cr 1,5-Diphenyl-
carbohydrazide
1,2
530 84
130 COD LR TT tube test 0 -150 mg/l O
2
Dichromate/H2SO
4
1,2
430 90
131 COD MR TT tube test 0 -1500 mg/l O
2
Dichromate/H2SO
4
1,2
610 92
132 COD HR TT tube test 0 -15 g/l O
2
Dichromate/H2SO
4
1,2
610 94
204 Colour direct
reading
0-500 Pt-Co units Pt-Co-Scale
1,2
(APHA)
430 96
150 Copper T * tablet 0.05-5 mg/l Cu Biquinoline
4
560 98
151 Copper L* liquid +
powder
0.05-4 mg/l Cu Bicinchoninate 560 102
153 Copper PP PP 0.05-5 mg/l Cu Bicinchoninate 560 108
Page 19
11
MD 610_1c 09/2015
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range; Vacu-vial® is a registered trade mark of CHEMetrics Inc.
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
157 Cyanide Powder +
liquid
0.01-0.5 mg/l CN Pyridine­barbituric acid
1
580 110
160
CyA-TEST T
tablet 0-160 mg/l CyA Melamine 530 112
165 DEHA T tablet +
liquid
20-500 μg/l DEHA PPST
3
560 114
167 DEHA PP PP + liquid 20-500 μg/l DEHA PPST
3
560 116
170 Fluoride L liquid 0.05-2 mg/l F SPADNS
2
580 118
210 H2O2 T tablet 0.03-3 mg/l H2O2DPD/catalyst
5
530 120
213 H2O2 LR L liquid 1-50 mg/l H2O2 Titanium
tetrachloride/acid
430 122
214 H2O2 HR L liquid 40-500 mg/l H2O2Titanium
tetrachloride/acid
530 124
190
Hardness, Calcium T
tablet 50-900 mg/l CaCO3Murexide
4
560 126
191 Hardness, Calcium
2 T
tablet 0-500 mg/l CaCO3Murexide
4
560 128
200
Hardness, total T
tablet 2-50 mg/l CaCO3Metallphthalein 3560 130
201 Hardness, total
HR T
tablet 20-500 mg/l CaCO3Metallphthalein 3560 132
205 Hydrazine P powder 0.05-0.5 mg/l N2H
4
4-(Dimethyl­amino)­benzaldehyde
3
430 134
206 Hydrazine L liquid 0.005-
0.6
mg/l N2H
4
4-(Dimethyl­amino)­benzaldehyde
3
430 136
207 Hydrazine C Vacu-vial 0.01-0.7 mg/l N2H
4
PDMAB 430 138
215 Iodine T tablet 0.05-3.6 mg/l I DPD
5
530 140
220 Iron T tablet 0.02-1 mg/l Fe PPST
3
560
142, 144
222 Iron PP PP 0.02-3 mg/l Fe 1,10-Phenan-
troline
3
530
142, 146
223 Iron (TPTZ) PP PP 0.02-1.8 mg/l Fe TPTZ 580
142, 148
224 Iron (Fe in Mo) PP PP 0.01-1.8 mg/l Fe Fe in Mo 580
142, 150
225 Iron LR L liquid 0.03-2 mg/l Fe Ferrozine /
Thioglycolate
560
142, 152
226 Iron LR 2 L liquid 0.03-2 mg/l Fe Ferrozine /
Thioglycolate
560 142,
156
227 Iron HR L liquid 0.1-10 mg/l Fe Thioglycolate 530 142,
160
240 Manganese T tablet 0.2-4 mg/l Mn Formaldoxime 530 164 242 Manganese LR PP PP + liquid 0.01-0.7 mg/l Mn PAN 560 166
Page 20
12
MD 610_1c 09/2015
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range; Vacu-vial® is a registered trade mark of CHEMetrics Inc.
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
243 Manganese HR PP PP + liquid 0,1-18 mg/l Mn Periodate
oxidation
2
530 168
245 Manganese L liquid 0.05-5 mg/l Mn Formaldoxime 430 170 250 Molybdate T tablet 1-50 mg/l MoO4Thioglycolate
4
430 172
251 Molybdate LR PP PP 0,05-5 mg/l MoO4Mercaptoacetic
acid
610 174
252 Molybdate HR PP PP 0.5-66 mg/l MoO4Mercaptoacetic
acid
430 176
254 Molybdate HR L liquid 1-100 mg/l MoO4Thioglycolate 430 178 257 Nickel T tablet 0.1-10 mg/l Ni Nioxime 560 180
260 Nitrate tablet +
powder
0.08-1 mg/l N Zinc reduction / NED
530 182
265 Nitrate TT tube test 1-30 mg/l N
Chromotropic acid
430 184
270 Nitrite T tablet 0.01-0.5 mg/l N N-(1-Naphthyl)-
ethylendiamine
2,3
560 186
272 Nitrite LR PP PP 0.01-0.3 mg/l N Diazotization 530 188 280 Nitrogen,
total
LR TT
tube test 0.5-25 mg/l N Persulfate
digestion method
430 190
281 Nitrogen,
total
HR TT
tube test 5-150 mg/l N Persulfate
digestion method
430 192
290
Oxygen, active T
tablet 0.1-10 mg/l O
2
DPD 530 196
292 Oxygen, dissolved Vacu-vial 10-800 μg/l O
2
Rhodazine D
TM
530 198
300
Ozone (DPD) T
tablet 0.02-2 mg/l O
3
DPD/Glycine
5
530 200
70 PHMB T tablet 2-60 mg/l PHMB Buffer/Indicator 560 206
320 Phosphate, T
ortho LR
tablet 0.05-4 mg/l PO
4
Ammonium­molybdate
2,3
660 208,
210
321 Phosphate, ortho
HR T
tablet 1-80 mg/l PO
4
Vanado
-
molybdate
2
430 208,
212
323 Phosphate, PP
ortho
PP 0.06-2.5 mg/l PO
4
Molybdate/ Ascorbic acid
2
660 208,
214
324 Phosphate, ortho TTtube test 0.06-5 mg/l PO
4
Molybdate/ Ascorbic acid
2
660 208,
216
327 Phosphate 1 C,
ortho
Vacu-vial 5-40 mg/l PO
4
Vanado­molybdate
2
430 208,
218
328 Phosphate 2 C,
ortho
Vacu-vial 0.05-5 mg/l PO
4
Stannous chloride
2
660 208,
220
325 Phosphate, hydr. TTtube test 0.02-1.6 mg/l P
Acid digestion, Ascorbic acid
2
660 208,
222
326 Phosphate, total TTtube test 0.02-1.1 mg/l P
Acid persulf digestion, Ascorbic acid
2
660 208,
224
334 Phosphate LR L liquid 0.1-10 mg/l PO
4
Phosphomolybdic acid/Ascorbic acid
660 208,
226
Page 21
13
MD 610_1c 09/2015
No. Analysis Reagent Range Displayed asMethod
[nm]
OTZ Page
335 Phosphate HR L liquid 5-80 mg/l PO
4
Vanado­molybdate
430 208,
230
316
Phosphonate PP
PP 0-125 mg/l Persulfate
UV-Oxidation
660 234
329
pH-Value LR T
tablet 5.2-6.8
Bromocresolpurple5560 238
330 pH-Value T tablet 6.5-8.4 Phenolred
5
560 240
331 pH-Value L liquid 6.5-8.4 Phenolred
5
560 242
332
pH-Value HR T
tablet 8.0-9.6 Thymolblue
5
560 244
338 Polyacrylate L liquid 1-30 mg/ l
Polyacryl
Turbidity 660 246
340 Potassium T tablet 0.7-12 mg/l K
Tetraphenylborate­Turbidity
4
430 250
350 Silica T tablet 0.05-4 mg/l SiO
2
Silicomolybdate
2,3
660 252
351 Silica LR PP PP 0.1-1.6 mg/l SiO
2
Heteropolyblue2660 254
352 Silica HR PP PP 1-90 mg/l SiO
2
Silicomolybdate2430 256
353 Silica L liquid +
powder
0.1-8 mg/l SiO
2
Heteropolyblue2660 258
212 Sodium
hypochlorite T
tablet 0.2-16 % NaOCl Potassium
iodide
5
530 260
355 Sulfate T tablet 5-100 mg/l SO
4
Bariumsulfate­Turbidity
610 262
360 Sulfate PP PP 5-100 mg/l SO
4
Bariumsulfate­Turbidity
2
530 264
365 Sulfide tablet 0.04-0.5 mg/l S DPD/Catalyst
3,4
660 266
370 Sulfite T tablet 0.1-5 mg/l SO
3
DTNB 430 268
384 Suspended Solids direct
reading
0-750 mg/l TSS photometric 660 270
386 Turbidity direct
reading
0-1000 FAU
Attenuated Radiation Method
530 272
388 Tolyltriazole PP PP 1-16 mg/l Benzo
triazole
Catalysed UV photolysis
430 274
390 Urea T tablet +
liquid
0.1-2.5 mg/l Urea Indophenol/ Urease
610 276
400 Zinc T tablet 0.02 -1 mg/l Zn Zincon
3
610 278
405 Zinc L liquid +
powder
0.1 -2.5 mg/l Zn Zincon / EDTA 610 280
* = free, combined, total; PP = powder pack; T = tablet; L = liquid; TT = tube test; LR = low range; MR = middle range; HR = high range; Vacu-vial® is a registered trade mark of CHEMetrics Inc.
Page 22
14
MD 610_1c 09/2015
1.1 Methods
The precision of Lovibond® Reagent Systems (tablets, powder packs and tube tests) is identical to the precision specified in standards literature such as American Standards (AWWA), ISO etc.
Most of the data referred to in these standard methods relates to Standard Solutions. Therefore they are not readily applicable to drinking-, boiler- or waste-water, since various interferences can have a major influence on the accuracy of the method. For this reason we don’t state such potentially misleading data.
Due to the fact that each sample is different, the only way to check the tolerances (‘precision’) is the Standard Additions Method.
According to this method, first the original sample is tested. Then further samples (2 to 4) are taken and small amounts of a Standard Solution are added, and further results are obtained. The amounts added range from approximately half, up to double the amount present in the sample itself.
These supplementary results make it possible to estimate the actual concentration of the original sample by comparison.
Literature
The reagent formulations are based on internationally recognised test methods. Some are described in national and/or international guidelines.
1. Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlammuntersuchung
2. Standard Methods for the Examination of Water and Wastewater; 18th Edition, 1992
3. Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart 1989
4. Photometrische Analyse, Lange / Vejdelek, Verlag Chemie 1980
5. Colorimetric Chemical Analytical Methods, 9th Edition, London
Notes for searching:
OTZ (OneTimeZero) switching on and off, see Mode 55, page 328
Active Oxygen -> Oxygen, activ Alkalinity-m -> Alkalinity, total Biguanide -> PHMB Calcium Hardness -> Hardness, Calcium Cyanuric acid -> CyA-TEST H2O2 -> Hydrogen peroxide Monochloramine -> Chloramine, mono m-Value -> Alkalinity, total Total Hardness -> Hardness, total p-Value -> Alkalinity-p Silicon dioxide -> Silica total Alkalinity -> Alkalinity, total total Hardness -> Hardness, total
Langelier Saturation -> Mode function 70 Index (Water Balance)
Page 23
15
MD 610_1c 09/2015
Page 24
16
MD 610_1c 09/2015
Acid demand to pH 4.3 with Tablet
0.1 – 4 mmol/l
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as Acid demand to pH 4.3 in mmol/l.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
2
0
Page 25
17
MD 610_1c 09/2015
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Acid demand to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
Reagent Form of reagent/Quantity Order-No.
ALKA-M-PHOTOMETER Tablet / 100 513210BT
Page 26
18
MD 610_1c 09/2015
Alkalinity, total = Alkalinity-m = m-Value with Tablet
5 – 200 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display as total Alkalinity.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
0
Page 27
19
MD 610_1c 09/2015
Notes:
1. The terms total Alkalinity, Alkalinity-m, m-Value and Alkalinity to pH 4.3 are identical.
2. For accurate results exactly 10 ml of water sample must be taken for the test.
3. Conversion table:
Acid demand to pH 4.3
DIN 38 409 (KS4.3)
German
°dH*
English
°eH*
French
°fH*
1 mg/l CaCO
3
0.02 0.056 0.07 0.1
*Carbonate hardness (reference = Hydrogencarbonate-anions) Example: 10 mg/l CaCO3 = 10 mg/l x 0.056 = 0.56 °dH 10 mg/l CaCO3 = 10 mg/l x 0.02 = 0.2 mmol/l
4. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
ALKA-M-HR PHOTOMETER Tablet / 100 513240BT
Page 28
20
MD 610_1c 09/2015
Alkalinity HR, total = Alkalinity-m HR = m-Value HR with Tablet
5 – 500 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-M-HR PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Press [ ] key.
Wait for a reaction period of 1 minute.
8. Remix the solution.
9. Place the vial in the sample chamber making sure that the marks are aligned.
10. Press TEST key.
The result is shown in the display as total Alkalinity.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
1
Countdown 1:00 start:
Page 29
21
MD 610_1c 09/2015
Notes:
1. For verification of the result look carefully at the bottom of the vial. If a thin yellow layer forms, then mix the vial again. This ensures that reaction is complete. Reread the result.
2. Conversion table:
Acid demand to pH 4.3
DIN 38 409 (K
S4.3)
German
°dH*
English
°eH*
French
°fH*
1 mg/l CaCO
3
0.02 0.056 0.07 0.1
*Carbonate hardness (reference = Hydrogencarbonate-anions) Example: 10 mg/l CaCO3 = 10 mg/l x 0.056 = 0.56 °dH 10 mg/l CaCO3 = 10 mg/l x 0.02 = 0.2 mmol/l
3. CaCO3 °dH °eH °fH
°aH
Reagent Form of reagent/Quantity Order-No.
ALKA-M-PHOTOMETER Tablet / 100 513210BT
Page 30
22
MD 610_1c 09/2015
Alkalinity-p = p-value with Tablet
5 – 300 mg/l CaCO
3
1.
Fill a clean vial (24 mm Ø) with 10 ml of the water
sample,
close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALKA-P-PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display as Alkalinity-p.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
3
5
Countdown 5:00
Page 31
23
MD 610_1c 09/2015
Notes
1. The terms Alkalinity-p, p-Value and Alkalinity to pH 8.2 are identical.
2. For accurate test results exactly 10 ml of water sample must be taken for the test.
3. This method was developed from a volumetric procedure for the determination of Alkalinity-p. Due to undefined conditions, the deviations from the standardised method may be greater.
4. Conversion table:
mg/l CaCO
3
°dH °fH °eH
1 mg/l CaCO
3
---- 0.056 0.10 0.07
1 °dH 17.8 ---- 1.78 1.25
1 °fH 10.0 0.56 ---- 0.70
1 °eH 14.3 0.80 1.43 ----
5. CaCO3 °dH °eH °fH
°aH
6. By determining Alkalinity-p and Alkalinity-m it is possible to classify the alkalinity as Hydroxide, Carbonate and Hydrogencarbonate. The following differentiation is only valid if: a) no other alkalis are present and b) Hydroxide und Hydrogen are not present in the same water sample. If condition b) is not fulfilled please get additional information from ”Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlammuntersuchung, D 8”. Case 1: Alkalinity-p = 0 Hydrogen carbonate = m Carbonate = 0 Hydroxide = 0 Case 2: Alkalinity-p > 0 and Alkalinity-m > 2p Hydrogen carbonate = m – 2p Carbonate = 2p Hydroxide = 0 Case 3: Alkalinity-p > 0 and Alkalinity-m < 2p Hydrogen carbonate = 0 Carbonate = 2m – 2p Hydroxide = 2p – m
Reagent Form of reagent/Quantity Order-No.
ALKA-P-PHOTOMETER Tablet / 100 513230BT
Page 32
24
MD 610_1c 09/2015
Aluminium with Tablet
0.01 – 0.3 mg/l Al
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one ALUMINIUM No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod (dissolve the tablet).
6. Add one ALUMINIUM No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Aluminium.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
4
0
Page 33
25
MD 610_1c 09/2015
Notes:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%). Rinse them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The effect of this is generally insignificant unless the water has fluoride added artificially. In this case, the following table should be used:
Fluoride [mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. A special tablet ingredient prevents effects on the measurement due to iron and manganese.
5. Al
Al2O
3
Reagent Form of reagent/Quantity Order-No. Set
ALUMINIUM No. 1 / No. 2
Tablet / per 100 inclusive stirring rod
517601BT
ALUMINIUM No. 1 Tablet / 100 515460BT
ALUMINIUM No. 2 Tablet / 100 515470BT
Page 34
26
MD 610_1c 09/2015
Aluminium with Vario Powder Pack
0.01 – 0.25 mg/l Al
Use two clean vials (24 mm Ø) and mark one as blank for zeroing.
1. Fill 20 ml of the water sample in a 100 ml beaker.
2.
Add the contents of one Vario Aluminum ECR F20
Powder Pack straight from the foil to the water sample.
3. Dissolve the powder using a clean stirring rod.
4. Press [ ] key. Wait for a reaction period of 30 seconds.
After the reaction period is finished proceed as follows:
5.
Add
the contents of
one Vario Hexamine F20 Powder
Pack straight
from the foil to the same water sample.
6. Dissolve the powder using a clean stirring rod.
7. Add 1 drop of Vario Aluminum ECR Masking Reagent in the vial marked as blank.
8. Add 10 ml of the prepared water sample to the vial (this is the blank).
9. Add the remaining 10 ml of the prepared water sample in the second clean vial (this is the sample).
10. Close the vials tightly with the caps and swirl several times to mix the contents.
11. Press [ ] key. Wait for a reaction period of 5 minutes.
After the reaction period is finished proceed as follows:
12. Place the vial (the blank) in the sample chamber making sure that the marks are aligned.
Countdown 1 0:30 start:
Countdown 2 5:00 start:
1 2
3 4
5
0
Page 35
27
MD 610_1c 09/2015
13. Press ZERO key.
14. Remove the vial from the sample chamber.
15. Place the vial (the sample) in the sample chamber making sure that the marks are aligned.
16. Press TEST key.
The result is shown in the display in mg/l Aluminium.
Notes:
1. Before use, clean the vials and the accessories with Hydrochloric acid (approx. 20%). Rinse them thoroughly with deionised water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates. The effect of this is generally insignificant unless the water has fluoride added artificially. In this case, the following table should be used:
Fluoride [mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. Al
Al2O
3
Reagent Form of reagent/Quantity Order-No. Set
VARIO Aluminium ECR F20 VARIO Aluminium Hexamine F 20 VARIO Aluminium ECR Masking Reagent
Powder Pack / 100 Powder Pack / 100 Liquid reagent / 25 ml
535000
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 36
28
MD 610_1c 09/2015
Ammonia with Tablet
0.02 – 1 mg/l N
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one AMMONIA No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Add one AMMONIA No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 10 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Ammonia as N.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 10:00
6
0
Page 37
29
MD 610_1c 09/2015
Notes:
1. The tablets must be added in the correct sequence.
2. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2 tablet has been added.
3. The temperature of the sample is important for full colour development. At a temperature below 20°C the reaction period is 15 minutes.
4. Sea water samples: Ammonia conditioning reagent is required when testing sea water or brackish water samples to prevent precipitation of salts. Fill the test tube with the sample to the 10 ml mark and add one level spoonful of Conditioning Powder. Mix to dissolve, then continue as described in the test instructions.
5. Conversion: mg/l NH4 = mg/l N x 1.29 mg/l NH3 = mg/l N x 1.22
6. N NH4
NH
3
Reagent Form of reagent/Quantity Order-No. Set
AMMONIA No. 1 / No. 2
Tablet / per 100 inclusive stirring rod
517611BT
AMMONIA No. 1 Tablet / 100 512580BT
AMMONIA No. 2 Tablet / 100 512590BT
Ammonia conditioning reagent (Sea water samples)
(approx. 100 tests) powder / 15 g
460170
Page 38
30
MD 610_1c 09/2015
Ammonia with Vario Powder Pack
0.01 – 0.8 mg/l N
Use two clean vials (24 mm Ø) and mark one as blank for zeroing.
1. Fill a clean vial (24 mm Ø) with 10 ml of deionised water (this is the blank).
2. Fill the other clean vial (24 mm Ø) with 10 ml of the water sample (this is the sample).
3. Add the contents of one Vario Ammonia Salicylate F10 Powder Pack straight from the foil to each vial.
4. Close the vials with the caps and shake to mix the contents.
5. Press [ ] key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished proceed as follows:
6. Add the contents of one Vario Ammonia Cyanurate F10 Powder Pack straight from the foil to each sam­ple.
7. Close the vials tightly with the caps and shake to mix the contents.
8. Press [ ] key.
Wait for a reaction period of 15 minutes.
After the reaction period is finished proceed as follows:
9. Place the vial (the blank) in the sample chamber making sure that the marks are aligned.
10. Press ZERO key.
11. Remove the vial from the sample chamber.
12. Place the vial (the sample) in the sample chamber making sure that the marks are aligned.
13. Press TEST key.
The result is shown in the display in mg/l Ammonia as N.
1 2
3 4
Countdown 1 3:00 start:
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2 15:00 start:
6
2
Page 39
31
MD 610_1c 09/2015
Notes:
1. Extremely basic or acidic water samples should be adjusted with 0.5 mol/l (1 N) Sulfuric acid solution or 1 mol/l (1 N) Sodium hydroxide solution to pH 7.
2. Interferences:
Interfering substance Interference levels and treatments
Calcium greater than 1000 mg/l CaCO
3
Iron Interferes at all levels. Correct as follows:
a) determine the concentration of iron present in the sample by performing a total Iron test
b) add the same iron concentration as determined to the deionised water (step 1).
The interference will be blanked out successfully.
Magnesium greater than 6000 mg/l CaCO
3
Nitrate greater than 100 mg/l NO3-N
Nitrite greater than 12 mg/l NO2-N
Phosphate greater than 100 mg/l PO4-P
Sulfate greater than 300 mg/l SO
4
Sulfide intensifies the colour
Glycine, Hydrazine, Colour, Turbidity
Less common interferences such as Hydrazine and Glycine will cause intensified colours in the prepared sample. Turbidity and colour will give erroneous high values. Samples with severe interferences require distillation.
3. N NH4
NH
3
Reagent Form of reagent/Quantity Order-No. Set
VARIO Ammonia Salicylate F10 VARIO Ammonia Cyanurate F10
Powder Pack / per 100 PP 535500
Page 40
32
MD 610_1c 09/2015
Ammonia LR with Vario Tube Test
0.02 – 2.5 mg/l N
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml deionised water (this is the blank).
2. Open another white capped reaction vial and add 2 ml of the water sample (this is the sample).
3. Add the contents of one Vario Ammonia Salicylate F5 Powder Pack straight from the foil into each vial.
4. Add the contents of one Vario Ammonia Cyanurate F5 Powder Pack straight from the foil into each vial.
5. Close the vials tightly with the caps and swirl several times to dissolve the powder.
6. Press [ ] key. Wait for a reaction period of 20 minutes.
After the reaction period is finished proceed as follows:
7. Place the vial (the blank) in the sample chamber making sure that the marks are l aligned.
8. Press ZERO key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber making sure that the marks are l aligned.
11. Press TEST key.
The result is shown in the display in mg/l Ammonia as N.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 1 20:00 start:
1 2
3 4
Ø 16 mm
6
5
Page 41
33
MD 610_1c 09/2015
Notes:
1. Strong alkaline or acidic water samples must be adjusted to approx. pH 7 before analysis (use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. Iron interferes with the test. The interferences will be eliminated as follows: Determine the amount of total iron present in the water sample. To produce the blank add an iron standard solution with the same iron concentration to the vial (point 1) instead of deionised water
3. Conversion: mg/l NH4 = mg/l N x 1.29 mg/l NH3 = mg/l N x 1.22
4. N NH4
NH
3
Reagent Form of reagent/Quantity Order-No. Set
VARIO Ammonia Salicylate F5 VARIO Ammonia Cyanurate F5 VARIO Am Diluent Reagent LR VARIO deionised water
Set
Powder Pack / 50 Powder Pack / 50 Reaction tube / 50 100 ml
535600
Page 42
34
MD 610_1c 09/2015
Ammonia HR with Vario Tube Test
1 – 50 mg/l N
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 0.1 ml deionised water (this is the blank).
2. Open another white capped reaction vial and add
0.1 ml of the water sample (this is the sample).
3. Add the contents of one Vario Ammonia Salicylate F5 Powder Pack straight from the foil into each vial.
4. Add the contents of one Vario Ammonia Cyanurate F5 Powder Pack straight from the foil into each vial.
5. Close the vials tightly with the caps and swirl several times to dissolve the powder.
6. Press [ ] key. Wait for a reaction period of 20 minutes.
After the reaction period is finished proceed as follows:
7. Place the vial (the blank) in the sample chamber making sure that the marks are l aligned.
8. Press ZERO key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber making sure that the marks are l aligned.
11. Press TEST key.
The result is shown in the display in mg/l Ammonia as N.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 1 20:00 start:
1 2
3 4
Ø 16 mm
6
6
Page 43
35
MD 610_1c 09/2015
Notes:
1. Strong alkaline or acidic water samples must be adjusted to approx. pH 7 before analysis (use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. If chlorine is known to be present, add one drop of 0.1 mol/l Sodium thiosulfate for each
0.3 mg/l Cl2 in a one litre water sample.
3. Iron interferes with the test. The interferences will be eliminated as follows: Determine the amount of total iron present in the water sample. Add an iron standard solution with the same concentration to the vial (point 1) instead of deionised water to produce the blank.
4. Conversion: mg/l NH4 = mg/l N x 1.29 mg/l NH3 = mg/l N x 1.22
5. N NH4
NH
3
Reagent Form of reagent/Quantity Order-No. Set
VARIO Ammonia Salicylate F5 VARIO Ammonia Cyanurate F5 VARIO Am Diluent Reagent HR VARIO deionised water
Set
Powder Pack / 50 Powder Pack / 50 Reaction tube / 50 100 ml
535650
Page 44
36
MD 610_1c 09/2015
Boron with Tablet
0.1 – 2 mg/l B
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one BORON No. 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod and dissolve the tablet.
6. Add one BORON No. 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 20 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Boron.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 20:00
8
5
Page 45
37
MD 610_1c 09/2015
Notes:
1. The tablets must added in the correct sequence.
2. The sample solution should have a pH value between 6 and 7.
3. Interferences are prevented by the presence of EDTA in the tablets.
4. The rate of colour development depends on the temperature. The temperature of the sample must be 20°C ± 1°C.
5. B
H3BO
3
Reagent Form of reagent/Quantity Order-No. Set
Bor No. 1 / No. 2
Tablet / per 100 inclusive stirring rod
517681BT
BORON No. 1 Tablet / 100 515790
BORON No. 2 Tablet / 100 515800BT
Page 46
38
MD 610_1c 09/2015
Bromine with Tablet
0.05 – 13 mg/l Br
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod (note 5).
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l Bromine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
8
0
Page 47
39
MD 610_1c 09/2015
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances,the subsequent determination of Bromine may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) forone hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Bromine, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 22 mg/l Bromine can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Bromine. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. Depending on the preparation of the dosed bromine, bromine compounds may not react completely with the DPD No.1 tablet. In this case, the DPD No.3 tablet should be added under observation with a reaction time of 2 minutes. Please follow the directions of the bromine compound manufacturer where necessary.
6. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Bromine.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
Page 48
40
MD 610_1c 09/2015
Bromine with Vario Powder Pack
0.05 – 4,5 mg/l Br
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine TOTAL-DPD / F10 Powder Pack straight from the foil to the water sample (note 5).
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key. Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Bromine.
prepare Zero press ZERO
1 2
3 4
8
1
Zero accepted prepare Test press TEST
Countdown 3:00
Page 49
41
MD 610_1c 09/2015
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances,the subsequent determination of Bromine may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) forone hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Bromine, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 4.5 mg/l Bromine can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Bromine. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. Alternatively a VARIO Chlorine FREE-DPD / F10 powder pack may be used for the determination of some bromine compounds. Please follow the directions of the bromine compound manufacturer where necessary.
6. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Bromine.
Reagent Form of reagent/Quantity Order-No.
VARIO Chlorine Total-DPD/F10 Powder Pack / 100 530120
Vario Clorine Free-DPD/F10 Powder Pack / 100 530100
Page 50
42
MD 610_1c 09/2015
Chloride with Tablet
0.5 – 25 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CHLORIDE T1 tablet straight from the foil to the water sample, crush the tablet using a clean stirring rod and dissolve the tablet.
6. Add one CHLORIDE T2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved (Note 1).
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l Chloride.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
9
0
Page 51
43
MD 610_1c 09/2015
Notes:
1. Ensure that all particles of the tablet are dissolved – Chloride causes an extremely fine distributed turbidity with a milky appearance. Heavy shaking leads to bigger sized particles which can cause false readings.
2. High concentrations of electrolytes and organic compounds have different effects on the precipitation reaction.
3. Ions which also form deposits with Silver nitrate in acidic media, such as Bromides, Iodides and Thiocyanates, interfere with the analysis.
4. Highly alkaline water should - if necessary - be neutralised using Nitric acid before analysis.
5. Conversion: mg/l NaCl = mg/l Cl- x 1,65
6. Cl-
NaCl
Reagent Form of reagent/Quantity Order-No. Set
CHLORIDE T1 / T2
Tablet / per 100 inclusive stirring rod
517741BT
CHLORIDE T1 Tablet / 100 515910BT
CHLORIDE T2 Tablet / 100 515920BT
Page 52
44
MD 610_1c 09/2015
Chloride with Liquid Reagent
0.5 – 20 mg/l Cl–
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
20 drops KS251 (Chloride Reagent A)
6. Close the vial tightly with the cap and invert several times to mix the contents.
7. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
20 drops KS253 (Chloride Reagent B)
8. Close the vial tightly with the cap and invert several times to mix the contents.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Chloride.
9
2
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
Page 53
45
MD 610_1c 09/2015
Notes:
1. Chloride causes an extremely fine distributed turbidity with a milky appearance. Heavy shaking leads to bigger sized particles which can cause false readings.
2. Conversion: mg/l NaCl = mg/l Cl- x 1,65
3. Cl-
NaCl
Reagent Form of reagent/Quantity Order-No.
KS251 (Chloride Reagenz A) Liquid reagent / 65 ml 56L025165
KS253 (Chloride Reagenz B) Liquid reagent / 65 ml 56L025365
Page 54
46
MD 610_1c 09/2015
Chlorine with Tablet
0.01 – 6 mg/l Cl
2
Chlorine HR with Tablet
0.1 – 10 mg/l Cl
2
Chlorine with Liquid Reagent
0.02 - 4 mg/l Cl
2
Chlorine with Vario Powder Pack
0.02 - 2 mg/l Cl
2
Chlorine HR with Vario Powder Pack
0.1 - 8 mg/l Cl
2
The following selection is shown in the display:
for the differentiated determination of free, combined and total Chlorine.
for the determination of free Chlorine.
for the determination of total Chlorine.
Select the desired determination with the arrow
keys [ ] and [ ]. Confirm with [ ] key.
Chlorine >> diff free total
>> diff
>> free
>> total
0
0
1
0
3
1
0
1
1
0
1
1
1
1
1
Page 55
47
MD 610_1c 09/2015
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Chlorine may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. For individual testing of free and total Chlorine, the use of different sets of glassware is recommended (EN ISO 7393-2, 5.3)
3. Preparing the sample: When preparing the sample, the lost of Chlorine, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
4. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents therefore contain a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
5. Exceeding the measuring range: Concentrations above: 10 mg/l Chlorine using tablets (method 100) 4 mg/l Chlorine using liquid reagents (method 101) 2 mg/l using powder packs (method 110) 8 mg/l using powder packs (method 111) can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Chlorine. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
6. Turbidity (can lead to errors): The use of the DPD No. 1 tablet (method 100) in samples with high Calcium ion contents* and/or high conductivity* can lead to turbidity of the sample and therefore incorrect measurements. In this case, the reagent tablet DPD No. 1 High Calcium should be used as an alternative. If turbidity does occur after the DPD No. 3 tablet has been added, this can be prevented by using the DPD No. 1 High Calcium tablet and the DPD No. 3 High Calcium tablet. The DPD No. 1 High Calcium should only be used in combination with the DPD No. 3 High Calcium.
* it is not possible to give exact values, because the development of turbidity depends on the nature of the sample.
7. If ??? is displayed at a differentiated test result see page 348.
8. Oxidizing agents such as Bromine, Ozone etc. interfere as they react in the same way as Chlorine.
Page 56
48
MD 610_1c 09/2015
Chlorine, free with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes:
See page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
0
01
Page 57
49
MD 610_1c 09/2015
Chlorine, total with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
Notes:
See page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
0
01
Page 58
50
MD 610_1c 09/2015
Chlorine, differentiated determination with Tablet
0.01 – 6 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it,
leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and
crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add one DPD No. 3 tablet straight from the foil to the
same water sample and crush the tablet using a clean stirring rod.
12. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
0
01
Page 59
51
MD 610_1c 09/2015
13. Place the vial in the sample chamber making sure that the marks are aligned.
14. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See page 47
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb Cl *,** mg/l total Cl
Reagent Form of reagent/Quantity Order-No. Set
DPD No. 1 / No. 3
Tablet / per 100 inclusive stirring rod
517711BT
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
Set
DPD No. 1 HIGH CALCIUM/ DPD No. 3 HIGH CALCIUM
Tablet / per 100 inclusive stirring rod
517781BT
DPD No. 1 HIGH CALCIUM Tablet / 100 515740BT
DPD No. 3 HIGH CALCIUM Tablet / 100 515730BT
Page 60
52
MD 610_1c 09/2015
Chlorine HR, free with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 HR tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes:
See page 47
3
01
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 61
53
MD 610_1c 09/2015
Chlorine HR, total with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet and one DPD No. 3 tablet straight from the foil and crush the tablets using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total Chlo-
rine.
Notes:
See page 47
3
01
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
Page 62
54
MD 610_1c 09/2015
Chlorine HR, differentiated determination with Tablet
0.1 – 10 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 HR tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add one DPD No. 3 HR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
12. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
3
01
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Page 63
55
MD 610_1c 09/2015
13. Place the vial in the sample chamber making sure that the marks are aligned.
14. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See page 47
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb Cl *,** mg/l total Cl
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 HR Tablet / 100 511500BT
DPD No. 3 HR Tablet / 100 511590BT
Page 64
56
MD 610_1c 09/2015
Chlorine, free with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty the vial.
5. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times to mix the contents.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display in mg/l free Chlorine.
Notes (free and total Chlorine):
1. After use replace the bottle caps securely noting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C.
3. Also see page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
0
1
1
Page 65
57
MD 610_1c 09/2015
Chlorine, total with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty
the vial.
5. Fill the vial with drops of the same size by holding the
bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
3 drops of DPD 3 solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times
to mix the contents.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 2:00
0
1
1
Page 66
58
MD 610_1c 09/2015
Chlorine, differentiated determination with Liquid Reagent
0.02 – 4 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty the vial.
5. Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times to mix the contents.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
10. Remove the vial from the sample chamber.
11. Add 3 drops of DPD 3 solution to the same water sample.
12. Close the vial tightly with the cap and swirl several times to mix the contents.
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
0
1
1
Page 67
59
MD 610_1c 09/2015
13. Place the vial in the sample chamber making sure that
the marks are aligned.
14. Press TEST key.
Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
1. After use replace the bottle caps securely noting the colour coding.
2. Store the reagent bottles in a cool, dry place ideally between 6°C and 10°C.
3. Also see page 47
T1 accepted prepare T2 press TEST
Countdown 2:00
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
Reagent Form of reagent/Quantity Order-No. Set
DPD No. 1 buffer solution DPD No. 1 reagent solution DPD No. 3 solution
(approx. 300 tests) 3 x Liquid reagent / 15 ml 1 x Liquid reagent / 15 ml 2 x Liquid reagent / 15 ml
471056
DPD No. 1 buffer solution Liquid reagent / 15 ml 471010
DPD No. 1 reagent solution Liquid reagent / 15 ml 471020
DPD No. 3 solution Liquid reagent / 15 ml 471030
Page 68
60
MD 610_1c 09/2015
Chlorine, free with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine FREE-DPD / F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l free Chlorine.
Notes:
See page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
0
1
1
Page 69
61
MD 610_1c 09/2015
Chlorine, total with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine TOTAL-DPD /
F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in mg/l total Chlorine.
Notes:
See page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
Countdown 3:00
0
1
1
Page 70
62
MD 610_1c 09/2015
Chlorine, differentiated determination with Vario Powder Pack
0.02 – 2 mg/l Cl
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine FREE-DPD/ F10 Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
9. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times and then fill the vial with 10 ml of the water sample.
10. Add the contents of one VARIO Chlorine TOTAL-DPD / F10 Powder Pack straight from the foil to the water sample.
11. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
1 2
3 4
0
1
1
Page 71
63
MD 610_1c 09/2015
12. Place the vial in the sample chamber making sure that
the marks are aligned.
13. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See page 47
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
T1 accepted prepare T2 press TEST
Countdown 3:00
Reagent Form of reagent/Quantity Order-No.
Vario Clorine Free-DPD/F10 Powder Pack / 100 530100
VARIO Chlorine Total-DPD/F10 Powder Pack / 100 530120
Page 72
64
MD 610_1c 09/2015
Chlorine HR, free with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine Free-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press the TEST key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes: See page 47
1 2
3 4
1
1
1
10 ml
5 ml
3
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Page 73
65
MD 610_1c 09/2015
Chlorine HR, total with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine TOTAL-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key. Wait for a reaction period of 3-6 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total Chlo-
rine.
Notes:
See page 47
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
Countdown 3:00
1
1
1
10 ml
5 ml
3
Page 74
66
MD 610_1c 09/2015
Chlorine HR, differentiated determination with Vario Powder Pack
plastic vial (type 3) 10 mm
0.1 – 8 mg/l Cl
2
1. Fill a clean vial (10 mm Ø) with 5 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of two Vario Chlorine Free-DPD/ F10 Powder Pack straight from the foil into the water sample.
6. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press the TEST key.
9. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times and then fill the vial with 5 ml of the water sample.
10. Add the contents of two Vario Chlorine TOTAL-DPD/ F10 Powder Pack straight from the foil into the water sample.
11. Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
1 2
3 4
1
1
1
10 ml
5 ml
3
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Page 75
67
MD 610_1c 09/2015
12. Place the vial in the sample chamber making sure that the marks are aligned.
13. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See page 47
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
T1 accepted prepare T2 press TEST
Countdown 3:00
Reagent Form of reagent/Quantity Order-No.
Vario Clorine Free-DPD/F10 Powder Pack / 100 530100
VARIO Chlorine Total-DPD/F10 Powder Pack / 100 530120
Page 76
68
MD 610_1c 09/2015
Chlorine dioxide with Tablet
0.02 – 11 mg/l ClO
2
The following selection is shown in the display:
for the determination of Chlorine dioxide in the presence of Chlorine.
for the determination of Chlorine dioxide in the absence of Chlorine.
Select the desired determination with the arrow keys [ ] and [ ]. Confirm with [ ] key.
Chlorine dioxide >> with Cl without Cl
>> with Cl
>> without Cl
2
0
1
Page 77
69
MD 610_1c 09/2015
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Chlorine dioxide may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Chlorine dioxide, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 19 mg/l Chlorine dioxide can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Chlorine dioxide. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. If ??? is displayed at a differentiated test result see page 348.
6. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Chlorine dioxide.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
GLYCINE Tablet / 100 512170BT
Page 78
70
MD 610_1c 09/2015
Chlorine dioxide in the presence of Chlorine with Tablet
0.02 – 11 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample.
2. Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod.
3. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
4.
Fill a second clean vial with 10 ml of water sample
and
close tightly with the cap
.
5. Place the vial in the sample chamber making sure that the marks are aligned.
6. Press ZERO key.
7.
Remove the vial from the sample chamber and empty
the vial
.
8. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
9. Transfer the contents of the first vial (Glycine solu-
tion) into the prepared vial (point 8).
10. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
11. Place the vial in the sample chamber making sure that the marks are aligned.
12. Press TEST key.
Zero accepted prepare T1 press TEST
prepare Zero press ZERO
1
2 0
Page 79
71
MD 610_1c 09/2015
13. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times. Fill with a few drops of water sample.
14. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
17. Place the vial in the sample chamber making sure that the marks are aligned.
18. Press TEST key.
19. Remove the vial from the sample chamber.
20. Add one DPD No. 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
21. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that the marks are aligned.
23. Press TEST key. Wait for a reaction period of 2 minutes.
After the reaction period is finished the measurement starts automatically.
The result is shown in the display in:
Chlorine dioxide in mg/l ClO2.
mg/l free Chlorine mg/l combined Chlorine mg/l total Chlorine
Notes:
See next page.
T1 accepted prepare T2 press TEST
Countdown 2:00
T2 accepted prepare T3 press TEST
*,** mg/l free Cl *,** mg/l comb. Cl *,** mg/l total Cl
*,** mg/l ClO
2
Page 80
72
MD 610_1c 09/2015
Notes: (Chlorine dioxide in the presence of Chlorine)
1. The conversion factor to convert Chlorine dioxide (display) to Chlorine dioxide as Chlorine is 2.6315. mg/l ClO2 [Cl] = mg/l ClO2 · 2,6315 Chlorine dioxide displayed as Chlorine units ClO2 [Cl] has its origin in swimming poolwater treatment according to DIN 19643.
2. The total Chlorine result given includes the contribution of the chlorine dioxide as Chlorine reading. For true Chlorine value add the free and combined Chlorine readings.
3. See also page 69.
Reagent Form of reagent/Quantity Order-No.
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
GLYCINE Tablet / 100 512170BT
Page 81
73
MD 610_1c 09/2015
Chlorine dioxide in absence of Chlorine with Tablet
0.02 – 11 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber and empty it, leaving a few drops remaining in the vial.
5. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
6. Add water sample to the 10 ml mark.
7. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
8. Place the vial in the sample chamber making sure that the marks are aligned.
9. Press TEST key.
The result is shown in the display
as Chlorine dioxide in mg/l ClO2.
Notes:
See page 69
prepare Zero press ZERO
Zero accepted prepare Test press TEST
*,** mg/l ClO2
2
0
1
Page 82
74
MD 610_1c 09/2015
Chlorine dioxide in absence of Chlorine with Vario Powder Pack
0.04 – 3.8 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine FREE-DPD / F10 Powder Pack straight from the foil to the water sample (Note 5).
6. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that the marks are aligned.
8. Press TEST key.
The result is shown in the display in mg/l Chlorine dioxide.
Notes:
See page 76
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
2
2
1
Page 83
75
MD 610_1c 09/2015
Chlorine dioxide in the presence of Chlorine with Vario Powder Pack
0.04 – 3.8 mg/l ClO
2
1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one GLYCINE tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved
7. Add the contents of one VARIO Chlorine FREE-DPD / F10 Powder Pack straight from the foil into the pre prepared vial.
8. Close the vial tightly with the cap and swirl several times to mix the contents (approx. 20 seconds).
9. Place the vial in the sample chamber making sure that the marks are aligned.
10. Press TEST key.
The result is shown in the display in mg/l Chlorine dioxide.
Notes:
See page 76
prepare Zero press ZERO
Zero accepted prepare Test press TEST
1 2
3 4
2
2
1
Page 84
76
MD 610_1c 09/2015
Notes:
1. Vial cleaning: As many household cleaners (e.g. dishwasher detergent) contain reducing substances, the subsequent determination of Chlorine dioxide may show lower results. To avoid any measurement errors, only use glassware free of Chlorine demand. Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample: When preparing the sample, the lost of Chlorine dioxide, e.g. by pipetting or shaking, must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagent tablet therefore contains a buffer for the pH adjustment. Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range: Concentrations above 3.8 mg/l Chlorine dioxide can lead to results showing 0 mg/l. In this case, the water sample must be diluted with water free of Chlorine dioxide. 10 ml of the diluted sample should be mixed with the reagent and the measurement repeated.
5. Oxidising agents such as Chlorine, Ozone etc. interfere as they react in the same way as Chlorine dioxide.
Reagent Form of reagent/Quantity Order-No.
Vario Clorine Free-DPD/F10 Powder Pack / 100 530100
GLYCINE Tablet / 100 512170BT
Page 85
77
MD 610_1c 09/2015
Page 86
78
MD 610_1c 09/2015
Chlorine HR (Kl) with Tablet
5 – 200 mg/l Cl
2
Insert the adapter for 16 mm Ø vials.
1. Fill a clean vial (16 mm Ø) with 8 ml of the water sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that the marks are l aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add one CHLORINE HR (KI) tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.
6. Add one ACIDIFYING GP tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
7. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
8. Place the vial in the sample chamber making sure that the marks are l aligned.
9. Press TEST key.
The result is shown in the display in mg/l Chlorine.
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Ø 16 mm
0
5
1
Page 87
79
MD 610_1c 09/2015
Notes:
1. Oxidizing agents interfere as they react in the same way as Chlorine.
Reagent Form of reagent/Quantity Order-No. Set ACIDIFYING GP/
CHLORINE HR (KI)
Tablet / per 100 inclusive stirring rod
517721BT
CHLORINE HR (KI) Tablet / 100 513000BT
ACIDIFYING GP Tablet / 100 515480BT
Page 88
80
MD 610_1c 09/2015
Chlorite in presence of Chlorine and Chlorine dioxide
0,01 – 6 mg/l Cl
2
Firstly, the glycine method is used to measure the concen­tration of Chlorine Dioxide. This is then followed by the determination of the free and total chlorine, from which the Combined Chlorine can be calculated. A third test is performed which measures the Total Chlorine concentration plus any Chlorite present. Finally, the Chlorite concentration can be calculated from the three recorded results.
The following selection is shown in the display:
select for the determination of free Chlorine.
1. Fill a clean vial with 10 ml of water sample.
2. Add one GLYCINE tablet straight from the foil to the
water sample and crush the tablet using a clean stirring rod.
3. Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved.
4. Fill a second clean vial with 10 ml of water sample, close tightly with the cap.
5. Place the vial in the sample chamber making sure that the marks are aligned.
6. Press ZERO key.
7. Remove the vial from the sample chamber and
empty the vial.
8. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
9. Transfer the contents of the first vial (Glycine solu-
tion) into the prepared vial (point 8).
1
0 0
>> free
Chlorine >> diff free total
prepare Zero press ZERO
Page 89
81
MD 610_1c 09/2015
10. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
11. Place the vial in the sample chamber making sure that the marks are aligned.
12. Press TEST key.
Record the displayed test result (G).
13. Remove the vial from the sample chamber, empty the vial, rinse vial and cap several times. Fill with a few drops of water sample.
14. Add one DPD No. 1 tablet straight from the foil and crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
17. Place the vial in the sample chamber making sure that the marks are aligned.
18. Press TEST key.
Record the displayed test result (A).
19. Remove the vial from the sample chamber.
20. Add one DPD No. 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.
21. Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
22. Place the vial in the sample chamber making sure that the marks are aligned.
Zero accepted prepare Test press TEST
Zero accepted prepare Test press TEST
Page 90
82
MD 610_1c 09/2015
23. Wait for a reaction period of 2 minutes.
24. Press TEST key.
Record the displayed test result (C).
25. Remove the vial from the sample chamber.
26. Add one DPD ACIDIFYING tablet straight from the foil
to the same water sample and crush the tablet using a clean stirring rod.
27. Wait for a reaction period of 2 minutes.
28. Add one DPD NEUTRALISING tablet straight from
the foil to the same water sample and crush the tablet using a clean stirring rod.
29. Close the vial tightly with the cap and swirl several times until the tablets are dissolved.
30. Place the vial in the sample chamber making sure that the marks are aligned.
31. Press TEST key.
Record the displayed test result (D).
Calculations:
mg/l Chlorine dioxide = result G x 1,9 mg/l free Chlorine = result A – result G mg/l combined Chlorine = result C – result A mg/l Chlorite = result D – ( result C + 4 x result G )
Zero accepted prepare Test press TEST
Zero accepted prepare Test press TEST
Page 91
83
MD 610_1c 09/2015
Tolerances:
1. By calculation of non direct analysable parameters it is necessary to consider the error propagation besed on the possible tolerances of the single test tesults.
2. see Notes Chlorine, page 47.
Reagent Form of reagent/Quantity Order-No. Set
DPD No. 1 / No. 3
Tablet / per 100 inclusive stirring rod
517711BT
DPD No. 1 Tablet / 100 511050BT
DPD No. 3 Tablet / 100 511080BT
GLYCINE Tablet / 100 512170BT
DPD ACIDIFYING Tablet / 100 512120
DPD NEUTRALISING Tablet / 100 511020BT
Page 92
84
MD 610_1c 09/2015
Chromium with Powder Pack
0.02 – 2 mg/l Cr
The following selection is shown in the display:
for the differentiated determination of Chromium (VI), Chromium (III) and total Chromium
for the determination of Chromium (VI)
for the determination of total Chromium (sum Cr (III) + Cr (VI))
Select the desired determination with the arrow keys [ ] and [ ]. Confirm with the [ ] key.
Note:
1. If ??? is displayed at the diffentiated test result see page 348.
Chrom >> diff Cr (VI) Cr (III + VI)
>> diff
>> Cr (VI)
>> Cr (III + VI)
1
2 5
Page 93
85
MD 610_1c 09/2015
Page 94
86
MD 610_1c 09/2015
Chromium, differentiated determination with Powder Pack
0.02 – 2 mg/l Cr
Digestion:
1.
Fill a clean vial (16 mm Ø) with 10 ml of water sample.
2. Add the contents of one PERSULF.RGT FOR CR Pow- der Pack straight from the foil into the vial.
3. Close the vial tightly with the cap and swirl several times to mix the contents.
4. Heat the vial for 120 minutes in a preheated thermo­reactor at a temperature of 100°C.
5. Remove the vial from the thermoreactor. (CAUTION: The vials are hot!). Invert the vial and allow to cool to room temperature.
Performing test procedure:
Insert the adapter for 16 mm Ø vials.
6. Place the pre prepared vial in the sample chamber making sure that the marks
are aligned.
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Add the contents of one CHROMIUM HEXAVALENT Powder Pack straight from the foil into the pre pre­pared vial.
10. Close the vial tightly with the cap and swirl several times to mix the contents.
11. Place the vial in the sample chamber making sure that the marks
are aligned.
12. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
Ø 16 mm
prepare Zero press ZERO
Zero accepted prepare T1 press TEST
Countdown 5:00
1
2 5
Page 95
87
MD 610_1c 09/2015
13. Fill a second clean vial (16 mm Ø) with 10 ml of the water sample.
14. Add the contents of one CHROMIUM HEXAVALENT Powder Pack straight from the foil to the water
sample.
15. Close the vial tightly with the cap and swirl several times to mix the contents.
16. Place the vial in the sample chamber making sure that the marks
are aligned.
17. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l Cr (Vl)
mg/l Cr (lll) mg/l Cr total Chromium
Notes:
1. Performing steps 1–12 determines concentration of total chromium and steps 13–17 determines concentration of Chromium (VI). The concentration of Chromium (III) results out of the difference.
2. pH value of the water sample should be between 3 and 9.
3. For information about interferences especially in waste water and chemical waste water through metals and reductive or oxidic agents see DIN 38 405 – D 24 and Standard Methods of Water and Wastewater, 20th Edition; 1998.
Reagent Form of reagent/Quantity Order-No.
PERSULF.RGT FOR CR Powder Pack / 100 537300
CHROMIUM HEXAVALENT Powder Pack / 100 537310
*,** mg/l Cr (VI) *,** mg/l Cr (III) *,** mg/l Cr tot.
T1 accepted prepare T2 press TEST
Countdown 5:00
Ø 16 mm
Page 96
88
MD 610_1c 09/2015
Chromium (Vl) with Powder Pack
0.02 – 2 mg/l Cr
Insert the adapter for 16 mm Ø vials.
1.
Fill a clean vial (16 mm Ø) with 10 ml of the water sample.
2. Place the vial in the sample chamber making sure that the marks
are aligned.
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one CHROMIUM HEXAVALENT Powder Pack straight from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times to mix the contents.
7. Place the vial in the sample chamber making sure that the marks
are aligned.
8. Press TEST key.
Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Chromium (Vl).
Notes: see previous page
Ø 16 mm
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
1
2 5
Reagent Form of reagent/Quantity Order-No.
PERSULF.RGT FOR CR Powder Pack / 100 537300
CHROMIUM HEXAVALENT Powder Pack / 100 537310
Page 97
89
MD 610_1c 09/2015
Chromium, total (Cr(lll) + Cr(Vl)) with Powder Pack
0.2 – 2 mg/l Cr
Digestion:
1.
Fill a clean vial (16 mm Ø) with 10 ml of water sample.
2. Add the contents of one PERSULF.RGT FOR CR Pow- der Pack straight from the foil into the vial.
3. Close the vial tightly with the cap and swirl several times to mix the contents.
4. Heat the vial for 120 minutes in a preheated thermo­reactor at a temperature of 100°C.
5. Remove the vial from the thermoreactor.
(CAUTION: The vials are hot!).
Invert the vial and allow to cool to room temperature.
Performing test procedure:
Insert the adapter for 16 mm Ø vials.
6. Place the pre prepared vial in the sample chamber mak­ing sure that the marks
are aligned.
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Add the contents of one CHROMIUM HEXAVALENT Powder Pack straight from the foil to the water sample.
10. Close the vial tightly with the cap and swirl several times to mix the contents.
11. Place the vial in the sample chamber making sure that the marks
are aligned.
12. Press TEST key. Wait for a reaction period of 5 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total
Chromium.
Ø 16 mm
prepare Zero press ZERO
Zero accepted prepare Test press TEST
Countdown 5:00
1
2 5
Page 98
90
MD 610_1c 09/2015
COD LR with Vario Tube Test
0 – 150 mg/l O
2
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml deionised water (this is the blank (Note 1)).
2. Open another white capped reaction vial and add 2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor at a temperature of 150°C.
5. (CAUTION: The vials are hot!) Remove the tubes from the heating block and allow them to cool to 60°C or less. Mix the contents by carefully inverting each tube several times while still warm. Then allow the tubes to cool to ambient temperature before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample chamber making sure that the marks are l aligned.
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample chamber making sure that the marks are l aligned.
10. Press TEST key.
The result is shown in the display in mg/l COD.
Zero accepted prepare Test press TEST
prepare Zero press ZERO
Ø 16 mm
3
0
1
Page 99
91
MD 610_1c 09/2015
Notes:
1. Run samples and blanks with the same batch of vials. The blank is stable when stored in the dark and can be used for further measurements with vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for final measurements.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is important to place the vials carefully in the sample chamber. The precipitate at the bottom of the sample should be not suspended.
4. Clean the outside of the vials with a towel. Finger prints or other marks will be removed.
5. Samples can be measured when the Chloride content does not exceed 1000 mg/l.
6. In exceptional cases, compounds contained in the water cannot be oxidized adequately, so results may be lower than reference methods.
Reagent Form of reagent/Quantity Order-No.
CSB VARIO LR 0 - 150 mg/l 1 Set (25 tests) 2420720
Page 100
92
MD 610_1c 09/2015
COD MR with Vario Tube Test
0 – 1500 mg/l O
2
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 2 ml deionised water (this is the blank (Note 1)).
2. Open another white capped reaction vial and add 2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor at a temperature of 150°C.
5. (CAUTION: The vials are hot!) Remove the tubes from the heating block and allow them to cool to 60°C or less. Mix the contents by carefully inverting each tube several times while still warm. Then allow the tubes to cool to ambient temperature before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample chamber making sure that the marks are l aligned.
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample chamber making sure that the marks are l aligned.
10. Press TEST key.
The result is shown in the display in mg/l COD.
Zero accepted prepare Test press TEST
prepare Zero press ZERO
Ø 16 mm
3
1
1
Loading...