Lovibond Scuba II, 21 61 00 Instruction Manual

Page 1
эксплуатации
Scuba II
GB DE
FR
IT
ES
NL
Betjeningsvejledning DK
SE
NO
FI
Инструкция по RU
Free Chlorine/
Freies Chlor
Total Chlorine/
Gesamtchlor
pH
Stabilizer
Alkalinity/
Alkalität
Page 2
2
7
9
15
15 sec
8
10
TEST
Fitting the battery/ Batteriewech­sel/ Remplacement des piles/ Sosti­tuzione delle batterie/ Cambio de batería/ Batterijwissel/ Batteriskifte/ Batteribyte/ Batteriskifte/ Pariston asennus/ Замена батарей
2
1
4
ZERO
5
MODE
6
Page 3
3
Technical Data/ Technische Daten/ Données techniques/ Dati tecnici/ Datos técnicos/ Technische gegevens/ Tekniske data/ Tekniska data/ Tekniske data/ Tekniset tiedot/ Технические данные:
Cl : 0.1 - 6.0 mg/l Cl
2
(0-1 mg/l ± 0.1; 1-2 mg/l ± 0.2;
2-3 mg/l ± 0.4; 3-6 mg/l ± 0.5) pH : 6.5 - 8.4 pH (± 0.2 pH) CyA : 1 - 50 mg/l Cys (± 10 mg/l), 50 - 160 mg/l Cys (± 20 mg/l) TA : 0 - 300 mg/l CaCO
3
(± 50 mg/l)
Light Source/ Optik/Optique/ Gruppo ottico/ Óptica/ Lens/ Optikk/ Optiikka/ Оптика:
LED: λ = 530 nm Battery/ Batterie/ Pile/ Batteria/ Batería/ Batterij/
Batteri/ Batterier/ Batteri/ Paristot/ Батарея: 2 x AAA (1.5 V), LR03
Auto-Off/ Arrêt automatique/ Spegnimento-auto­matico/ Automatisk sluk/ Automattinen sammutus/
Батарея:
300 sec Ambient Conditions/ Umgebungsbedingungen/
Conditions ambiantes/ Conditioni ambientali/ Condiciones-ambientales/ Omgevings-condities/ omgivelses-betingelser/ Miljöför-hållanden/ Ympä­ristötiedot Lämpötila/ Окружающие условия: T = 5-40°C
Waterproof/ Wasserdicht/ Étanchéité à l‘eau/ Impermeambile/ Resistente al agua/ Waterdicht/ Kapslingsklasse/ Vattentät/ Vanntett/ Vesitiiviys/
Водонепроницаемость:
Analog IP68, 1 h @ 0.1 m Conformity/ Konformität/ Conformité/ Conformi-
tà/ Conformidad/ Conformieit/ Overensstemmelse/ Märkning/ Konformitet/ Vaatmustenmukasisuus/ Соответствие нормам: CE
Page 4
4
GB
Operation
• Fitting the battery (2 x AAA (1.5 V), LR03)
Change the batteries with this display indication.
Check position of the O-Ring when opening/ closing the battery chamber lid! Switch the unit on using the ON/OFF key. The display shows: “0”. Make one Zero for all the methods: Swirl the instrument several times under the water surface allowing the sample chamber to completely fill. Take the instrument out of the water.
Close the blue lid tightly. Press the ZERO/TEST key. “000” flashes for
approx. 8 seconds. The display shows “0.0.0”. Select the test required using the MODE key: Cl (free Chlorine/ total Chlorine) → pH → TA → CyA → Cl ... (Scroll).
6
Open the blue lid and add the appropriate tablet straight from the foil. Method appropriate tablet Cl : Free chlorine DPD No.1 Cl : Total chlorine DPD No.1 + DPD No.3 pH : pH Value PHENOL RED PHOTOMETER CyA : Stabilizer CyA-Test
TA : Alkalinity-M ALKA-M-PHOTOMETER { Crush the tablet using a clean stirring rod.
Stir until the tablet is dissolved. | Close the lid and swirl the instrument gently for approx. 15 seconds. } Afterwards keep the instrument steady and wait for 15 seconds when testing ClF/pH/CyA and 2 minutes when testing Clt and TA.
Page 5
5
Press the “Zero/Test“-key. The “---“ symbol flashes for approx. 6 seconds. The result appears in the display. LOW
Result below the lowest limit of the measuring range HIGH
Measuring range exceeded Repeat Test Press the “Zero/Test“-key
Change Method
Press the “Mode“-key to scroll down.
New zero:
Press the “Zero/Test“-key for 2 sec.
Maintenance
1. Keep the instrument in a clean condition.
2. No solvent, abrasive materials or brushes should be used to clean the instrument. Clean the sample chamber using a soft tissue.
3. Dry the instrument when not in use, store under ambient environmental conditions.
4. Remove battery if instrument is being stored for an extended period.
• Trouble shooting
1. Use only blister tablets with black printing.
2. Do not touch the tablets. This can cause contamination and give pour results.
3. Close the blue lid on top of the sample chamber tightly before zeroing/reading.
4. Always ensure that the sample chamber, the lid and the stirring rod are thoroughly rinsed when changing from one test to the other.
Warranty
The waterproof Scuba ll is warranted to be free from manufacturing defects for two years.
Conversion Table
mmol/l KS4.3 °dh °e °f
1 mg/l CaCO
3
0.02 0.056 0.07 0.10
Page 6
Technical changes without notice Printed in Germany 10/2012
Delivery Content
1 Scuba II in a robust plastic case with hanger 2 Batteries (AAA) 20 DPD No.1 tablets 20 Phenolred tablets 10 DPD No.3 tablets 10 CyA tablets 10 Alka-M tablets 1 Stirring rod 1 Instruction Manual
Order Code 21 61 00
Reagents and Assessories
525600 Combipack, consisting of 20 DPD No.1 tablets and 10 tablets each DPD No.3, Phenolred, Cya, Alka-M
511050BT 100 DPD No.1 tablets 511080BT 100 DPD No.3 tablets 511770BT 100 Phenolred tablets 511370BT 100 CyA tablets 513210BT 100 Alka-M-tablets
19802218 Blue lid 364107 Stirring rod 1950021 Battery (AAA), 1.5 V (each)
Update: www.lovibond.com No.: 00 38 69 62