Lord MSLINK0004 Users Manual [en, fr]

Page 1
LORD QUICKSTARTGUIDE
®
V-Link®-200
Wireless 8 Channel Analog Input Sensor Node
The V- Link- 200 wireless sensor node features eight analog input channels designed to accommodate a wide range of Wheatstone bridge and analog sensors, including strain, load cell, torque, pressure, acceleration, vibration, magnetic field, displacement, and geophones. There are four channels for single­ended sensor measurement, four channels for differential sensor measurement, and an on-board internal temperature sensor.
To acquire sensor data, the V- Link-200 is used with a LORD Sensing data gateway such as the WSDA-101 or WSDA-1500 Base. V-Link-200 sensor nodes come with the following components.
Item Description Quantity
A V-Link-200 WirelessSensor Node B Antenna withright angleadapter
C Node Tester Board 1 D
AA Lithium Batteries (3.6 V dc, 2.4 Ah) 4
Table 1 - V-Link-200 Components List
1 1
The LED indicators on the V-Link-200 include operational modes showing when the node is booting up, idle and waiting for a command, sampling, resynchronizing, or if there is an error. see Table 2 - below describing basic indicator behavior.
Figure 1 - Interface and Indicators
Indicator Behavior Node Status
OFF Node isOFF
Device status indicator
Rapid green flashing on start-up
1 (slow) green pulse per second
1 green blink every 2 seconds
Blue LED during sampling Node isresynchronizing Red LED Error or low battery
Table 2 - Indicator Behaviors
Node isbooting up
Node isidle and waiting for a command
Node issampling
Wireless Simplicity, Hardwired Reliability™
Page 2
V-Link-200 Wireless Sensor Node QuickStart Guide
SYSTEM OPERATION
Sensor nodes have three operational modes: node is sampling, it is in active mode. When sampling stops, the node is switched into idle mode, which is used for configuring node settings, and allows toggling between sampling and sleeping modes. The node will automatically go into the ultra low-power sleep mode after a user- determined period of inactivity. The node will not go into sleep mode while sampling.
active,sleep
, and
idle
. When the
1. Install Software
Install the SensorConnect software on the host computer before connecting any hardware. The SensorConnect software is available on the LORD Sensing website for free download:
http://www.microstrain.com/software/software
2. Make System Connections
To acquire sensor data, the V- Link-200 is used with a LORD Sensing data gateway such as the WSDA-101-Base or WSDA-1500-LXRS.
Figure 2 - Node Operational Modes
2
Page 3
V-Link-200 Wireless Sensor Node QuickStart Guide
3. Establish Gateway Communication
The WSDA-101-Base USB gateway is used in this example. For information on how to use other gateways, refer to the gateway or SensorConnect user manual.
Drivers for the USB gateways are included the SensorConnect software installation. With the software installed, the USB gateway will be detected automatically whenever the gateway is plugged in.
1. Power is applied to the gateway through the USB connection. Verify the gateway status indicator is illuminated, showing the gateway is connected and powered on.
2. Open the SensorConnect software.
3. The gateway should appear in the Controller window automatically with a communication port assignment. If the gateway is not automatically discovered, verify the port is active on the host computer, and then remove and re-insert the USB connector.
4. Connect to Nodes
The node can be connected with the automatic node discovery feature and by manually entering the node address and then searching for it on the current gateway communication frequency.
1. If the base and node are on the same operating frequency, the node will populate below the Base Station listing when powering on the V­Link-200.
Figure 4 - Automatic Node Discovery
2. If a red circle with a number appears next to Base Station, the node is operating on a separate radio channel. Select the Base Station and then select the Nodes on Other Frequencies tile.
Figure 3 - USB Gateway Communication
Figure 5 - Nodes On Other Frequencies
3
Page 4
V-Link-200 Wireless Sensor Node QuickStart Guide
3. Highlight the new node being added. Select Move Node to Frequency (#).
Figure 6 - Move Node
4. When manually entering the node address, select Manual Add Node, enter Node Address, last known Frequency (factory default is
15), and select Add Node.
5. Configure Node
Node settings are stored to non-volatile memory and may be configured using SensorConnect. For details related to node configuration, see Wireless Sensor
Configuration on page 1.
For this example the V-Link-200 tester board is on channel 1.
1. Select Hardware > Input Range for channel 1, select +/-2 mV from the drop down menu.
2. Under Hardware Offset, select Balance Target for channel 1, select Mid (50%) from the drop down menu.
3. Select Auto- Balance. When auto- balance is complete, a blue information window will indicate the balance result.
Figure 8 - Auto-Balance
4. Select Calibration.
5. Select Microstrainfrom the Unit drop down menu, and select the Shunt Cal button enabled on the right.
Figure 7 - Adding a Node by Address
4
Page 5
V-Link-200 Wireless Sensor Node QuickStart Guide
Figure 9 - Node Configuration Menu
6. Use the following settings: a. Calibration Mode: Internal
b. Number of Active Gauges: 4
c. Gauge Factor: 2
d. Gauge Resistance: 1000
9. When the calibration is complete, the Wireless Node Configuration window will appear.
10. Select Apply Configuration to write to node memory.
See below for wiring of other strain gauges.
e. Shunt Resistance: 499000
7. Select Start Shunt Cal for Slope and Offset calibrations.
8. Select Accept Calibration.
Figure 10 - Channel Settings
Figure 11 - Full-Bridge (Standard)
Figure 12 - Half-Bridge
5
Page 6
V-Link-200 Wireless Sensor Node QuickStart Guide
Figure 13 - Two Wire Quarter-Bridge *
Figure 15 - Sampling Setting
3. Select Apply and Start Network.
4. Select Create QuickView Dashboard in the pop up window immediately to create a dashboard of the new data.
Figure 14 - Three Wire Quarter-Bridge *
* Requires optional on-board bridge completion
6. Configure Sampling Setting and Start Data Acquisition
1. Left click on the Base Station > Sampling, and indicate the nodes to be sampled by checking the box to the left of each node.
2. Under Sampling, select Sample Rate from the drop down menu, select Continuously to sample indefinitely, or For to set a limited time for sample sets.
6
Figure 16 - Quick View Dashboard
Page 7
V-Link-200 Wireless Sensor Node QuickStart Guide
7. Replacing Batteries
1. Remove the screws on both sides of the face plate to open the V-Link-
200.
2. It is important to replace all four of the batteries at the same time, observing the correct polarity orientation. The positive polarities are indicated on the batteries and the node by a "+" symbol.
3. Reassemble.
Figure 17 - Replace Batteries
8. Battery Hazards
The V- Link -200 contains internal, non-rechargeable lithium batteries . Lithium batteries are a fire and explosion hazard. Do not store or operate the node at temperatures above 212°F (100°C). Do not disassemble, short circuit, crush, puncture, or otherwise misuse the battery.
Lithium batteries contain toxic chemicals that are harmful to humans and the environment. Disposal is subject to federal and local laws. Do not discard the battery or the node in the trash. Follow proper battery disposal protocol, or contact LORD Sensing Technical Support for information on extracting the battery or returning the product for proper recycling and disposal.
9. Regulatory Compliance Information
FCC ID: XJQMSLINK0004
IC ID: 8505A-MSLINK0005
V-Link-200Wireless 8 Channel Analog Input SensorNode
This device complies with Part 15 of the United States FCC Rules, and Industry Canada’s license­exempt RSSs. Operation is subject to the following two conditions: 1) This device may not cause interference, and 2) This device must accept any interference, including interference that may cause undesired operation of the device. Changes or modifications, including antenna changes not expressly approved by LORD Corporationcould void theuser’s authority to operate the equipment.
Cet appareil est conforme à la Partie 15 des Règles de la FCC des États-Unis et aux RSSS exempts de licence d'Industrie Canada. Le fonctionnement est soumis aux deux conditions suivantes: 1) Cet appareil ne doit pas causer d'interférences et 2) Cet appareil doit accepter toute interférence, y compris les interférences pouvant entraîner un fonctionnement indésirable de l'appareil. Les changements ou modifications, y compris les changements d'antenne non expressément approuvés par LORD Corporation, pourraient annuler l'autorisation de l'utilisateur d'utiliser l'équipement.
LORDCo rporation Micro Strain®Sensing Systems 459 Hurricane Lane , Suite 102 Williston, VT 05495 USA
ph: 802-862-6629
Copyright © 2017 LORD Corporation Document 8501-0073 Revision D. Subject tochangewithout notice.
www.microstrain.com
Page 8
V-Link-200 Wireless Sensor Node QuickStart Guide
10. Power Supply
The V-Link-200 Wireless Sensor Node canbe powered by an external source.
Apply only the input voltage range specified for the V­Link-200. Connect to a power source that is near the device, is accessible, and adheres to all national wiring standards. Compliance with wiring standards is assumed in the installation of the power source and includes protection against excessive currents, short circuits, and ground faults. Failure to do so could result in personal injury and permanent damage to the device.
LORDCo rporation Micro Strain®Sensing Systems 459 Hurricane Lane , Suite 102 Williston, VT 05495 USA
ph: 802-862-6629
Copyright © 2017 LORD Corporation Document 8501-0073 Revision D. Subject tochangewithout notice.
www.microstrain.com
Loading...