Please read operating manual carefully before operation.
Warning:
Over-push or over-draw the syringe may result in the fluid
sprayed. Use appropriate measures to protect operator
and equipment. Be careful during operation.
When the fluid sprays out on the drive unit please shut
down the power supply immediately and clean the drive
unit, then turn on the power supply.
If a trouble happens please contact us or our dealer. Don't
repair the equipment by yourself.
Be careful when inserting or extracting the connection
wire between controller and drive unit to prevent the plug
from damaging.
If the power line or the plug are worn or damaged please
pull out the plug.
Please shut down the power supply before connecting the
external control equipment.
TableofContents
Introduction
Operation Panel
Basic Operation
Start/Stop Key
Fast Forward Key
Fast Reverse Key
Rotary Coded Switch
Channels Switch Key
Exit Key
Menu Frame Work
Multi-channel Running Interface
Multi-channel Functions Setting
Select Drive Unit
Select Copy Parameters
Set Delay Time
Controller I.D.
Single Channel Running Interface
Single Channel Functions Setting
TS series products are controllers of four channels and multi-function
syringe pumps. They need to be connected with drive units to constitute
syringe pump systems of strong functions and easy-to-operate. They are
widely used in biological laboratory. There are 3 types:
Controller Type
TS-1A
TS-2A
TS-1B
S2
S1
F1 Power Switch
F2 Power Socket
F3DriveUnitSocket
F4 External Control Input and Communication Interface
F5 External Control Output Interface
Note:
Controller and drive unit are connected through 8 pin S terminal.
Be careful when insertion or extraction. Please see page 21 for
detail information aboutcontrol function.External
DriveUnitType
L0107-1A
L0107- A
W0109-1B
F1
F2
F3
Working Mode
Infusion-only
2
Infusion/Withdrawal
Infusion/Withdrawal
S3
S1 Controller
S2 Drive Unit
S3 Glass Syringe
F5
F4
Operation Panel
Fast Forward Key
Exit Key
Channels Switch Key
Fast Reverse Key
Start/Stop Key
Rotary Coded Switch
Basic Operation
Start/Stop Key
Press the Start/Stop Key to start or stop the pump. In multichannel working mode, to pause the running state, the pump will
prompt if continue the current working state.
Fast Forward Key
In multi-channel working mode, pressing the Fast Forward Key all
the enabling channels infuse at the max. speed. In single channel
working mode, pressing the Fast Forward Key this channel infuses
at the max. speed. Loose this key the pump stops. Other keys are
invalid when pressing this key.
Fast Reverse Key
In multi-channel working mode, pressing the Fast Reverse Key all
the enabling channels withdraw at the max. speed. In single
channel working mode, pressing the Fast Reverse Key this channel
withdraws at the max. speed. Loose this key the pump stops. Other
keys are invalid when pressing this key.
Note: Fast Forward Key and Fast Reverse Key are invalid when the
pump is in running state.
Rotary Coded Switch (RCS)
Turn the Rotary Coded Switch for menu selection or parameter
setting. Press the Rotary Coded Switch for confirmation.
Note: Put the finger in the center of the switch to press the RCS.
Put the finger on the small concave which is on the side of the
switch to turn the RCS.
Channels Switch Key
Switch between different channels.
Exit Key
Cancel current operation and return to previous menu.
Switch on the pump. First initializing, and then display the running
interface.
a
2#:
1#:
e
3#:
b
a. Enabling symbol: Symbol of enabling or disabling the channels.
Means enabling the current channel.means disabling the
current channel. The disabling channel is invalid when the pump is
in running state or switching the channels.
b. Channel number: Channel serial number.
c. Start/Stop state: Display the state of current channel.
Start displaysStop displaysPause displays
d. Direction: Display running direction. Infusion displays.
Withdrawal displays.
e. Working mode: Display current working mode. Displayin
Select the parameters of one drive unit as copy parameters, all
enabling channels will run according to copy parameters. Not setting
copy parameters each enabling channel will run according to its own
parameters setting.Means copy channel;means non-copy
channel. Only one channel can be selected as copy channel, or noncopy channel is acceptable. The setting method is as below:
1#:
2#:
3#:
4#:
Set Delay Time
Press and
turn the RCS
Press the
Exit Key
Select Unit:
1# 2# 3# 4#
Copy Para Sel:
1# 2# 3# 4#
Select “Yes”
or “No”
Save change?
YesoCa ncelN
Turn the RCS to select
if save the change
Press the RCS
Press the
Exit Key
Select
“Cancel”
Select Unit:
1# 2# 3# 4#
Copy Para Sel:
1# 2# 3# 4#
Turn the RCS to select
the copy channel
Turn the RCS
Select Unit:
1# 2# 3# 4#
Copy Para Sel:
1# 2# 3# 4#
PresstheRCStoset
the copy channel
Set delay time to start each drive unit in turn. Time units are sec,
min and hour. Time numerical value is from 0.1 to 999.9. The
setting method is as below:
1#:
2#:
3#:
4#:
Press the RCS
Press the
Exit Key
Select Unit:
1# 2# 3# 4#
Copy Para Sel:
Turn the RCS
1# 2# 3# 4#
Save change?
YesoCa ncelN
Turn the RCS to select if save
the change
Press the RCS
Select “Yes”
or “No”
Select
“Cancel”
Delay time:
1#6.0 sec:
2#0.0 sec:
3#0.0 sec:
Press the
Exit Key
Delay time:
Press the RCS
1#6.0 sec:
2#0.0 sec:
3#0.0 sec:
Turn the RCS to select channel
Press the RCS
Controller I.D.
When control computer controls many pumps through RS485
interface, it must identify each controller's I.D.. This controller I.D.
should be unique. It's the identification of this controller.
The controller address serial number is from 1 to 30. The default is
1. The setting method is as below:
1#:
Press the RCS
2#:
3#:
4#:
Press the
Exit Key
Single Channel
Select Unit:
1# 2# 3# 4#
Copy Para Sel:
1# 2# 3# 4#
Save change?
YesoCa ncelN
Turn the RCS to select
if save the change
Turn the RCS
Select “Yes” or “No”
Select
“Cancel”
Press the RCS
Controller ID:
Controller ID:
Turn the RCS to set the
controller I.D.
Running Interface
Switch on the pump. First initializing, and then display the multichannel running interface. Press Channels Switch Key to enter single
channel running interface. Please see below for detail operation:
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Press the Exit Key
Press the
Exit Key
Note: TS-1A controller disables Mode Select. It only has infusion
working mode.
Working Mode Selection
1#1.860 L
Vol
:μ
8.3 sec10
Syringe
: 100 Lμ
Please see page16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
Press the RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn the RCS to select
working mode
Select “Yes”
or “No”
Save change?
YesoCa ncelN
Turn the RCS to select
if save the change
Prompts
Infusion: The pump infuses during the whole working course
Withdrawal: The pump withdraws during the whole working course
Infusion/Withdrawal: The pump infuses first and then withdraws during the
Withdrawal/Infusion: The pump withdraws first and then infuses during the
Continuous: The pump infuses first and then withdraws, and then cycles
Please see page 11 “Dispensing Parameters Setting” to set the parameters for
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
turn the
RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn the RCS to highlight
“Syringe Select”
Select “Yes” or “No”
Save change?
YesoCa ncelN
Turn theto select
RCS
if save the parameter
Press the RCS
Press the Exit Key
Unimetrics
Becton Dickinson
Hamilton
Barrel I.D.
Turn theto select
the type
Press the RCS
Size Select
5102550
Select
“Cancel”
100 250 500 1000
I.D.
Turn theto select
the specification
RCS
Press the RCS
()μL
: 1.13mm
RCS
Three syringe manufacturers are listed in the TS-1A,TS-2A menu.
Please see below table for specifications:
Unimetrics
Hamilton
SGE
Spec L()μ
I.D. mm()
μ
Spec L()
I.D. mm()
Spec L()μ
I.D. mm()
10
0.46
25
0.73
50
1.03
100
1.46
250
2.3
102550100250500
0.48
25
0.73
0.73
50
1.03
1.03
100
1.46
1.462.33.26
250
2.3
500
3.26
500
3.26
1000
4.61
Three syringe manufacturers are listed in the TS-1B menu. Please see
below table for specifications:
Unimetrics
Becton
Dickinson
Hamilton
Spec L()μ
I.D. mm()
Spec cc()
I.D. mm()
Spec m()L
I.D. mm()
10
0.46
1
4.7
1
4.61
25
0.73
3
8.59
2.5
7.28
50
1.03
5
11.99
5
10.3
100
1.46
250
2.3
500
3.26
1000
4.61
10
14.482019.053021.596026.6
10
14.572523.035032.57
Setting and Select the Inner Diameter of the Syringe
1. Setting the Inner Diameter of the Syringe
If the syringe used is not listed in the menu the inner diameter of
the syringe needs to be measured and entered (0.01 mm - 40 mm).
The pump can save 4 inner diameters of the syringes.
1#
Vol
:
1.860 Lμ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
1#8.00mm
:
:
2#10.00mm
:
3#10.00mm
:
4#10.00mm
Select “Yes”
Save change?
YesoCa ncelN
Turn theto select
if save the change
RCS
Press and
turn the
RCS
Press the
Exit Key
Select “No”
Press the RCS
Select
“Cancel”
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theto highlight
RCS
the Syringe Select
“”
Press and
turn the
RCS
Unimetrics
Becton Dickinson
Hamilton
Barrel I.D.
1#10.00mm
:
:
2#10.00mm
:
3#10.00mm
:
4#10.00mm
1#8.00mm
:
:
2#10.00mm
:
3#10.00mm
:
4#10.00mm
Turn theto set
the inner diameter
Press the
Exit Key
Press the RCS
Press the RCS
RCS
Select “Yes” or “No”
Press the
Exit Key
Select
“Cancel”
Save change?
YesoCa ncelN
Turn theto select
if save the change
RCS
2.Select the Inner Diameter of the Syringe
Select the inner diameter of the syringe after setting the inner diameter.
The dispense parameters include dispense volume, infusion time,
withdrawal time, copy number and pause time.
Dispense volume: It is the volume for one time infusion or withdrawal
The volume range is 1 nL to 99.99 mL
There are volume units of nL, 0.01nL, 0.1µL, µL and 0.01mL available.
Press and
1#L
Vol
:
1.860 μ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
Note:
turn the
RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theand highlight
the “Dispense Para”
RCS
Select “Yes” or “No”
Select “Cancel”
Save change?
YesoCa ncelN
Turn theto select
RCS
if save the parameter
Press the RCS
Press the RCS
Press the
Exit Key
Disp Vol Unit
Infusion Unit
Disp Vol Unit
Infusion Unit
Turn theand set the unitRCS
Disp Vol Unit
Infusion Unit
Turn theto
set the volume
860.01 L
1μ
8.5sec
Press the RCS
860.01 L
1μ
8.5sec
Press the RCS
860.01 L
1μ
8.5sec
RCS
When the one time dispensing volume or the total dispensing volume
(multiply one time dispensing volume by copy number) are more than
the volume of the syringe the running interface will display “Vol
Overflow” (see below). Please change the syringe or adjust the
dispensing volume or the copy number.
Vol overflow!
8.3 sec10
Syringe: 100 Lμ
Infusion time: It is the time for one time infusion
The range of infusion time is from 0.1 second to 999.9 hours
There are time units of second, minute and hour available
Press and
1#
Vol
:
1.860 Lμ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
turn the
RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theand highlight
RCS
the Dispense Para
Select “Yes” or “No”
Select “Cancel”
Save change?
YesoCa ncelN
if save the change
RCS
Press and turn the RCS
Press the
Exit Key
Disp Vol Unit
860.01 L
1μ
Infusion Unit
8.5sec
Disp Vol Unit
860.01 L
1μ
Infusion Unit
8.5sec
Turn theand set the unitRCS
Press the RCS
Disp Vol Unit
860.01 L
1μ
Infusion Unit
8.5sec
TurnthetosetthetimeRCSTurn theto select
Press the RCS
Press the RCS
Note:
When the pump can't infuse the setting volume in the setting time
the running interface will display “I-speed Overflow” (see below).
Pleaseadjustthedispensingtimeorthedispensingvolume.
Withdrawaltime:Itisthetimeforonetimewithdrawal
The range of withdrawal time is from 0.1 second to 999.9 hours
There are time units of second, minute and hour available
Press and
1#
Vol
:
1.860 Lμ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
turn the
RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theand highlight
the “Dispense Para”
RCS
Select “Yes” or “No”
Save change?
YesoCa ncelN
Turn theto select
RCS
if save the parameter
Press and turn the RCS
Select “Cancel”
Press the RCS
Press the
Exit Key
Withdrawal Unit
8.3sec
Copy No.:
1
Press the RCS
Withdrawal Unit
8.3sec
Copy No.:
1
Turn theand set the unitRCS
Press the RCS
Withdrawal Unit
8.3sec
Copy No.:
1
TurnthetosetthetimeRCS
Note:
When the pump can't withdrawal the setting volume in the setting
time the running interface will display “W-speed Overflow” (see
below). Please adjust the dispensing time or the dispensing volume.
W-speed Overflow!
8.3 sec10
Syringe: 100 Lμ
Copy Number: It is the total infusion and withdrawal numbers of the
whole working course. The range is 1 - 999. In Infusion/Withdrawal
(I/W) and Withdrawal/Infusion (W/I) modes, one time working
course is finishing one time infusion and withdrawal.
1#
Vol
:
1.860 Lμ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
Note:
When the one time dispensing volume or the total
dispensing volume (multiply one time dispensing volume
by copy number) are more than the volume of the syringe
the running interface will display “Vol Overflow” (see
right). Please change the syringe or adjust the dispensing
volume or the copy number.
Press and
turn the
RCS
Press the
Exit Key
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theand highlight
RCS
the“DispensePara”
Turn theto select
if save the parameter
Select “Yes” or “No”
Save change?
YesoCa ncelN
RCS
Select “Cancel”
Press and turn the RCS
Press the Exit Key
Withdrawal Unit
60.0sec
Copy No.:
Press the RCS
Withdrawal Unit
60.0sec
Copy No.:
Turn theand set
the copy number
Vol overflow!
8.3 sec10
Syringe: 100 Lμ
1
Press the RCS
1
RCS
Pause Time: It is the stop time between two dispensing operation.
Therangeis0.1- 999.9min.
1#
Vol
:
1.860 Lμ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
When the accuracy of the dispensing volume can not reach 0.5% (in
the condition of30% of max. infusion distance), calibration is
≥
needed to adjust the linear speed of the drive unit.
Press and
1#
Vol 1.860 L:μ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Save change?
YesoCa ncelN
Turn theto select
RCS
if save the parameter
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
turn the
RCS
Press the
Exit Key
Select “Yes” or “No”
Press the RCS
Select “Cancel”
Mode Select
yringe Select
S
Dispense Para
Calibration
Turn theand
RCS
highlight the “Calibration”
Interface A
10% 50% 80%
Vol:
Time60sec
:
Act10.00 Lμ
:
Turn theto enter
the calibration volume
Turn theto select
if save the parameter
RCS
Select “Yes” or “No”
Save change?
YesoCa ncelN
RCS
Press the RCS
Press the Exit Key
Press the
Key
Press the RCS
Press the RCS
Select “Cancel”
Turn theto enter
the actual volume
Vol10.00:
:
Time60sec
:
Act10.00 L
Start/Stop
to start the test
Vol10.00:
:
Time60sec
:
Act10.00 L
Press the RCS
Vol10.00:
:
Time60sec
:
Act10.00 L
RCS
Volume (Vol): The required volume for calibration
Enter volume setting interface (see Interface A) and turn the
to set the calibration volume. There are 10%, 50% and 80% of the
syringe volume available. When selecting the user-defined syringe
the default calibration volume is 50% of the syringe volume. Please
see page 9 “Syringe Select” to select suitable syringe.
Time: The calibration time for the pump is 1 minute.
Actual volume (Act): Press theafter setting the
Start/Stop Key
calibration volume. Enter the actual volume.
μμL
Press the RCS
μμL
μμL
RCS
Display Select
Press and
1#
Vol 1.860 L:μ
8.3 sec10
Syringe
: 100 Lμ
Please see page 16 “Display
Select” for “Speed” and
“Flow” display.
Note:
Select “Yes” to save the
current setting
Select“No”nottosave
the setting
Select “Cancel” to return
to previous menu and
keep on selecting
turn the
RCS
Press the
Exit Key
Display Select
Turn theand highlight
RCS
the “Display Mode”
Select “Yes” or “No”
Press the RCS
Press the
Exit Key
Display Select
Volume
Speed
Flow
Turn theto select
the Display Mode
Press the RCS
Save change?
YesoCa ncelN
Turn theto select
if save the parameter
RCS
Select “Cancel”
RCS
There are volume, speed and flow rate display modes available.
Volume: Display current one time infusion or withdrawal volume
3. Lift up the pressing
knob and rotate the
press knob according
to the arrowhead
direction to fix the
glass syringe
1. Lift up and rotate
the pressing knob.
Rotate and loose
the tightening
knobs according to
the arrowhead
directions
TS-1B/W0109-1B
Plunger Fixing Board
Barrel Fixing Board
1.Rotate the tightening
knobs according to the
arrowhead directions.
Loose the plunger fixing
board and the barrel
fixing board
Pressing Knob
2.Lift up the pressing knob
and rotate the pressing
knob according to the
arrowhead direction.
Load the syringe. Rotate
the pressing knob to
press the syringe
2. See above drawing
to load the glass
syringe. Lift up the
press knob and
rotate the press
knob according to
the arrowhead
direction to press
the glass syringe
Tightening
Knob 2
Tightening
Knob 1
19
3. Rotate the tightening
knobs according to
the arrowhead
directions to fix the
glass syringe
3.Tighten the tightening
knobs according to the
arrowhead directions to
fix the syringe
External Control Input and Communication Interface
Port Definition:
1# +5V
2# COM
3# COM for external control start/stop
4# External control start/stop: change to high level, start and stop
states switch
5# NC
6# COM for Rs485
7# RS485 B
8# RS485 A
Note: 1 - 3# are RS485 communication bus ports. Can connect to
the control computer (computer, PLC, SCM).
Stop State: OC gate is open
Running State: OC gate is close
Infusion: OC gate is open
Withdrawal:OCgateisclose
Channel 2Channel 3 Channel 4
5678
9101112
13 14 15 16
(1 2)(5 6)(9 10)(13 14)
are 4 suits OC gates
(3 4)(7 8)(11 12)(15 16)
are 4 suits OC gates
12345678
Maintenance
Keep the surface of the pump clean and dry. Prevent the liquids
from entering the inner of the drive units.
The surface of drive units are not organic solvent and aggressive
liquids resistant. Please pay attention when using.
Note
If a trouble happens, please contact us or our dealers.
Warranty
The warranty period for this product is one year. If repair or
adjustment is necessary within the warranty period, the problem
will be corrected at no charge if it is not due to misuse or abuse on
your part, as determined by the manufacturer. Repair costs outside
the warranty period, or those resulting from product misuse or
abuse,maybeinvoicedtoyou.
TS-1A/L0107-1A、TS-2A/L0107-2A、TS-1B/W0109-1B
Technical Specifications
System Functions
Parameters Setting: The parameters of each channel can be different
Parameters Saving: The parameters of each channel can be saved
Running Control: Each drive unit can be controlled separately; Or four
drive units can be controlled to run simultaneously or run at
different time
Copy Channel: All drive units can be run according to the parameters of
one drive unit of them
Delaying Startup: Delaying startup time of each channel can be
controlled separately
Memory Function: Select resume operation or remain stopped when
power returns after an interruption
Block Protection: When one drive unit stops accidently, the system will
warm and stop
External Control: Each channel has two ways OC gate output to
indicate the start/stop and running direction of this channel
Communication: Realize computer control through RS485
communication interface
Main Functions for Each Channel
Syringe Selection: The syringe can be selected in the manufacturer
table which includes manufacturer and size
User-defined Syringe: 4 inner diameter of user-defined syringe barrel
canbeinputandsaved
Working Mode Setting: TS-1A has only infusion working mode. TS-
2A/TS-1B has 5 working modes available: infusion, withdrawal,
infusion/withdrawal, withdrawal/infusion, continuous
Parameters Setting: Set dispensing volume, infusion time, withdrawal
time, pause time and copy number
Display Mode Selection: Different parameters (volume, flow rate,
linear speed) can be selected in the main display interface