The UTRED-16 USB temperature logger contains a non-replaceable
Lithium Battery. When the battery indicates “LOW”, the logger
should be replaced, and the battery recycled or disposed of
according to your local regulations.
Do not expose the logger to extreme temperatures as it may lead to
the destruction of the battery and may cause injuries.
Keep out of the reach of children.
Liability
LogTag Recorders’ standard warranty terms apply. A copy can be requested by emailing
. In addition, LogTag Recorders shall not be held liable:
l if the device was used beyond LogTag Recorders’ stated limitations
l for any claims due to the improper storage and use of the device
l for any problems with refrigeration units
l for the bad quality of the monitored goods, if any
l for incorrect readings if the device was used with a low battery
l for consequential loss
Battery Life
The main battery in the UTRED-16 is designed to power the device for up to 24 months of
operation, provided
l the device was not stored for more than 2 years prior to activation;
l the device is not downloaded excessively to a PC (more than once a week);
l the recording interval is not shorter than 15 minutes;
l the device is stored and operated according to LogTag Recorders’ recommendations.
The UTRED-16 is unable to record readings when connected to USB. You cannot
use a USB power supply to power the device!
Disclaimer
The UTRED-16 monitors temperature exposure and not the quality of the goods it
accompanies. Its purpose is to signal if product quality evaluation/testing is required.
LogTag’s UTRED-16 is a fully configurable, multi-use USB PDF temperature logger with an
external temperature probe and a built-in display, that can create temperature reports
without the need to install proprietary software or hardware at the destination. Select
statistical data can be reviewed directly on the display without the need for a PC. The
logger can be configured using LogTag® Analyzer, and is then placed with the goods to be
monitored. At the destination the UTRED-16 can be plugged straight into a computer’s
USB port and generates a PDF file, which can be accessed using PDF software such as
Acrobat Reader.
Features
The UTRED-16 USB PDF temperature logger features the familiar LogTag case layout with
an additional USB plug at the bottom.
Case
l Mounting lug for secure fastening of logger to fixtures
l USB plug with protective cap - shields USB connector from moisture and dirt
l Gold-plated, high-quality MCX socket for external temperature probe
l Durable polycarbonate case, IP61
l START/Mark button; used to start the unit, clear alarms or place an inspection mark in
the data
l STOP/Review button; used to stop the unit or to review data recorded during the trip
Display
l Shows last recorded temperature and information about alerts
l Allows reviewing statistical data
Probes
The UTRED-16 will accept any probe from the ST100 product range.
PDF
The UTRED-16 will generate a detailed PDFreport when plugged into a USB port of a PC.
The PDF report shows a summary of the trip, presents the data in chart and list format. A
CSV file of the data list is also available.
The PDF can be shown in several different languages:
l English
l French
l German
l Russian
l Spanish
The default ex-factory language is English. You can select which language to use during
configuration in the File Settings tab.
You can only change the language during configuration, but not when viewing the
PDF.
In addition to your LogTag® UTRED-16 USB PDF Temperature Recorder with Remote Probe
and Display you will need the following:
l an external probe of the ST100 series
l a PC running Windows 7 or later and LogTag® Analyzer 3 installed if you wish to
configure and download the logger
l a PC with PDF reader software installed for viewing the generated PDF files
Optional Items
In addition to the above, following items are useful accessories:
l a glycol buffer, which simulates environmental behaviour of a vaccine vial
Software Requirements
Software requirements vary depending on the type of file you want the UTRED-16 to
generate when plugged into a USB port.
If you wish to configure UTRED-16 products you will need to download the LogTag®
Analyzer software from LogTag®’s web site at
https://logtagrecorders.com/software/lta3/download/. Follow the instructions to install
and start the software.
If you only plan to view a report, plug the UTRED-16 into a computer’s USB port.
Depending on the settings made during configuration, a number of on-board files will be
generated1 and made available to you in a new drive:
l a PDF file
you can open the file directly from the logger‘s USB memory storage with Acrobat
Reader 4.0 or later, or any other compatible PDF reader software of your choice
1 Depending on the configuration, the UTRED-16 may produce all, some or none of the files.
2 PDF files generated by any USB logger can be opened directly in LogTag
this file can be imported into a spreadsheet program such as Microsoft Excel
l an LTD file
LTD files are LogTag® Analyzer's native, encrypted data files. You can open this file
in LogTag® Analyzer, where you can analyze data in detail, generate report files or
combine data from multiple recorders for comparison
You can also download data directly into LogTag® Analyzer without accessing the logger‘s
Your LogTag® data logger is typically configured, placed
with goods that require monitoring and then
downloaded once the goods reach their intended
destination. Other applications may involve stationary
loggers placed in warehouses or to monitor fridges.
Each application requires several steps so you can
deploy your data logger and retrieve the recorded
measurements. You may be able to skip some of the
steps, depending on the way the logger was prepared
when it was delivered to you and how you wish to
analyze the data.
Figure 1: LogTag® recording cycle
Things to do before you start logging
1. Download and install the latest version of LogTag® Analyzer 3.
You can do this from the LogTag Recorders Software Download page.
2. Configure the unit using LogTag® Analyzer.
3. Connect a probe and place the unit in the monitoring location.
4. Start the logger.
Things to do while recording data
5. Clear an active alarm.
6. Review Minimum and Maximum values during the trip, as well as alarm threshold
violations.
7. Reset Minimum and Maximum values.
8. Place a mark in the recorded readings.
9. Stop the logger recording data.
Things to do when you have finished
10. Plug the unit into a USB port.
11. View the PDF report generated by the UTRED-16.
12. Download the unit to LogTag® Analyzer for analysis.
13. Re-configure or reset the logger, ready for a new trip.
Configuring the UTRED-16 for loggingUTRED-16 User Guide11
This is done using LogTag® Analyzer software, which can also be used for downloading and
analyzing data.
Preparing the logger
l Start the LogTag® Analyzer software.
l Remove the protective cap and insert the UTRED-16 into a USB port.
You will receive a message that a new USB device has been found, and a generic driver
will be installed3.
Once the driver is installed and the UTRED-16 is ready, it can receive configuration
data. Click LogTag - Configure; LogTag® Analyzer will scan all USB ports and
display the configuration options for connected UTRED-16‘s. Although you can
configure many devices at the same time4, it is practical to limit the number of
devices to about 10, using a powered USB hub. It is not possible to configure different
models at the same time.
l Click Configure from the LogTag menu; LogTag® Analyzer will display the
configuration options for connected loggers.
l Remove the protective cap from the logger's USB connector, plug it into a USB
socket on your computer and wait for the drivers to be installed5.
Standard Configuration Options
The standard configuration options include settings such as User ID, start method, pre-
start recording, logging interval and duration, start delay, and password.
3 Depending on the factory configuration additional drivers may be installed at this time.
4 You can configure more than one UTRED-16 at the same time, however it is practical to limit the number of units to about 6,
using a powered USB hub.
5 You can configure more than one UTRED-16 at the same time, however it is practical to limit the number of units to about 6,
Configuring the UTRED-16 for loggingUTRED-16 User Guide12
For detailed information about each parameter please read the section about Configuring a LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.
UTRED-16 Start Options
During configuration with LogTag® Analyzer you can decide when the UTRED-16 starts
taking temperature readings:
Push button start
The logger will start taking temperature readings as soon as you have pressed the
START/Mark button (see Starting the Logger on page20).
When you choose the push button start option, you can Enable pre-start logging (or
disable it) and also Begin recording after a delay.
Pre-start readings
If you enable pre-start readings, the UTRED-16 starts recording as soon as it is unplugged
from the USB port6 and will continue to do so until you start the unit. No alarms are
processed while pre-start readings are being taken, pre-start readings will not appear in
the PDF file after the logger was started, and no PDF file will be generated if only pre-start
readings are present in the logger. Using pre-start readings is a good way to avoid data
loss if you forget to start the unit, as you can still access the data using LogTag®
Analyzer.
Start delay
If you configure the UTRED-16 to start after a delay period, the logger will not
immediately record temperature readings after you have pressed START/Mark, but start a
countdown timer instead, and record readings only after the timer has ended. The value for
the timer is set during configuration. If pre-start readings are enabled, these will continue
to be recorded during the delay period.
Date/Time start
The logger will start taking temperature readings at the date and time you specified during
configuration (local time). You cannot combine a date/time start with pre-start readings or
the start delay function.
6 The first pre-start reading is taken one log interval after you unplug the unit from the USB port.
Configuring the UTRED-16 for loggingUTRED-16 User Guide13
File Settings
Select the File Settings tab to select which files are generated when the UTRED-16 is
plugged into a computer's USB port, and what information these files contain. Select as
many file formats as you wish to generate.
Parameters that influence the appearance of all files are:
l setting the temperature units
The temperature unit for the PDF file is also used for the units shown on the
display.
l setting the date and time format
l configuring time zone and MKT values
l setting the language
Parameters specifically influencing the appearance of the PDF file are:
l scaling parameters for the chart
l showing or hiding grid lines
l showing or hiding alarm threshold lines
l generating the data list
For detailed information about each parameter please read the section about Configuring a LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.
The PDF file can not only be viewed in a PDF viewer, but can also be opened with
LogTag® Analyzer directly.
Advanced Configuration Settings
Select Advanced Settings for additional configuration settings. These settings decide how
some of the elements are displayed on the unit's own display and set certain options
specific to the UTRED-16.
Configuring the UTRED-16 for loggingUTRED-16 User Guide14
Figure 2: UTRED-16 advanced configuration screen in LogTag® Analyzer 3
Parameters that influence the appearance of the display are:
l Pause alarm/statistics processing (see Paused Readings on page31)
l Switch off display after 30 seconds (Power save, see Power Save on page31)
l allowing a user to reset the trip's minimum and maximum values on the display
during recording
l showing the alarm screens during review (Reviewing Alarms on page24)
Parameters influencing specific behavior of the UTRED-16 are:
l clearing and resetting alarms when the START/Mark button is pressed
l leaving the alarm turned on, even if readings return to the normal temperature range
again
l allowing the user to stop the logger with the STOP/Review button
l allowing the user to reset the logger with the START/Mark button
For detailed information about each parameter please read the section about Configuring a LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.
Alarm Configuration Options
LogTag® Analyzer allows up to 6 different alarm trigger conditions for temperature when
configuring a UTRED-16 for recording, depending on the model.
Each alarm trigger condition consists of a threshold temperature value, an activation type
(which can be instant, consecutive or accumulative) and a delay time, if it is not an instant
alarm.
If an alarm trigger condition requires readings to exceed an upper threshold temperature it
is called an
thresholds it is called a
An alarm event is generated, when either of the entered alarm conditions is triggered.
The UTRED-16 can have up to three upper alarms, and up to three lower alarms.
Configuring the UTRED-16 for loggingUTRED-16 User Guide16
Any alarm condition that is activated will be shown on the PDF, in the software and on the
display.
The section about Reviewing Alarms on page24 shows examples of how this is
represented and reviewed on the display.
Finalizing the configuration
Click Configure to upload the configuration data to the UTRED-16.
When the configuration is complete, unplug the UTRED-16 from the USB socket and
replace the protective cap.
If you wish to configure more UTRED-16 units with the same configuration, insert the next
loggers into USB sockets, wait until they are ready for configuration and click Repeat
Configure. Alternatively, you can use the Profile function to configure multiple units with
the same settings.
You can upload the configuration to a UTRED-16 logger as often as required.
The symbol is shown as soon as the UTRED-16 has registered an alarm event.While there are no
alarms, the symbol is shown.
Recording indicators
The recording indicators show what the UTRED-16 is currently recording.
l If READY is shown, the UTRED-16 is ready to be started with the START/Mark button.
Depending on the configuration it may already record pre-start readings.
l If ST ARTING is shown, the logger has been started, and a start delay is active. The word
DELAY is also shown, together with the time in hours and minutes until the start.
l If is shown, the UTRED-16 is recording temperature at the sample interval defined
during configuration.
l If is shown together with the word PAUSED, the product is also recording, but the
recorded values are not taken into account when calculating alarm events and durations.
l If the word STOP PED is shown, the UTRED-16 has finished recording temperature data.
Mark indicator
The word MARK is shown when the next temperature reading is identified with an inspection mark
in the recorded data.
MIN/MAX indicator
The word MAX is shown when the temperature on the display represents the maximum
recorded temperature for the current trip. The word MIN is shown when the temperature
on the display represents the minimum recorded temperature for the current trip7.
Temperature Units
Depending on the selected PDF temperature units, this shows either°F or°C.
7 If MIN/MAX data were reset during the trip, these temperatures reflect only the period since the reset took place.
This shows the most recently recorded temperature while the UTRED-16 is recording. Once the
logger has stopped, nothing will be displayed. During review, this will show minimum or maximum
temperatures.
Time Value and Time Indicators
The time value display is used to show one of the following:
l the current time
l the time remaining until the logger starts recording (for a delayed start)
l a duration, for example of an alarm
The time indicators identify which of those is displayed as follows:
l If TIME is shown, the time value represents the current time in hours and minutes (24- hour
format).
l If DELAY is shown, the time value represents a start delay, or the time remaining until a
date/time start will occur.
l If DURATION is shown, the time value represents the duration, for example the time above
the upper alarm limit.
The word boot appears in place of the clock if new firmware is being uploaded to the UTRED-16.
The word USb appears when the unit is plugged into a USB port.
Above/Below Threshold Arrows
The up-arrow is shown when the temperature displayed (i.e. last recorded) is above the
specified upper temperature threshold. The down-arrow is shown when the temperature
displayed is below the specified lower temperature threshold.
The four smaller up/down-arrows are secondary and tertiary alarm symbols. They are used when
more than one upper or more than one lower alarm is defined and are shown then the current
temperature is above or below the corresponding alarm levels.
Alarm Threshold Markers
The upper alarm threshold marker () is shown if one of the upper temperature limits triggered
an alarm event. It is also displayed when the review screen shows one of the upper threshold
values.
The lower alarm threshold marker () is shown if an alarm event occurred at one of the lower
temperature limits. It is also displayed when the review screen shows one of the lower threshold
values.
The four smaller up/down-arrows are secondary and tertiary alarm symbols. They are displayed
when more than one upper or more than one lower alarm is defined and the current temperature
is above or below the corresponding alarm levels.
Battery Low
The battery low symbol will appear if the UTRED-16's battery is low. If the symbol is not shown
while the display is turned on, the battery is still OK.
Before the UTRED-16 can record temperature data you will
need to connect an external probe.
The UTRED-16 accepts any ST100 probe. Plug the sensor
into the socket located on the side of the device, as shown.
Starting the Logger
Before you carry out the following start procedure, please ensure you have plugged
the external probe firmly into the socket on the side of the logger.
Push button start, quickstart enabled
Press and hold the START/Mark button until READY disappears and the
symbolS T AR T I NG is shown. The UTRED-16 now records temperature data. This start
option works very similar to other LogTag® LED products.
Push button start, quickstart disabled
Press and hold the START/Mark button. READY immediately disappears, and the
symbol blinks for 4 seconds. The symbol then shows permanently for 2 seconds. If you
release the button during this time, the logger starts taking temperature readings. The
symbol remains shown. If the button is released earlier than 4 seconds or later than
6 seconds, the display shows READY again, and the logger does not start.
Push Button Start with Start Delay
If the logger was configured for push button start with a start delay (regardless of
quickstart settings), the word DELAY is shown instead ofthe symbol, once you have
started the logger.
The delay time is shown in hours and minutes, if it is less than 24:00 hours, and the word
The time counts down and the UTRED-16 starts recording when it reaches 0:00.
If the start delay is greater than 24 hours the word DAYS is shown along with the number
of full days remaining.
Each time a user presses the START/Mark button during the start delay phase, the start
delay value is reset to its originally configured value.
Automatic date/time start
If you configured the UTRED-16 for a date/time start, dtSt will be displayed instead of the
clock.
The logger will record temperature values as soon as the configured date and time have
passed.
If the logger is still plugged into a USB port when the start time passes, the recorded
readings will show USB Paused on the PDFand in the software's reports.
During normal operation the display shows the most recently recorded temperature. The
current time is also displayed in 24-hour format. A tick symbol is shown as long as no
alarm event has occurred.
Temperatures can be shown in Celsius (as on the left) or Fahrenheit (as on the right).
The temperature unit on the display is the same as used for the PDF and CSV files.
It is set during configuration.
Readings above or below alarm thresholds
If the most recently recorded temperature is above or below one of the alarm thresholds,
one or more alarm markers will be shown on the display.
This symbol is displayed ...... when the most recent temperature recorded is
above the primary upper alarm threshold, but below the
secondary
above the secondary upper alarm threshold, but below the
tertiary
above the tertiary upper alarm threshold
below the primary lower alarm threshold, but above the
secondary
below the secondary lower alarm threshold, but above the
tertiary
below the tertiary lower alarm threshold
The presence of an alarm marker on the display does NOT indicate an alarm event.
It only shows that the most recent temperature recorded was above or below the
corresponding alarm threshold value. If an alarm marker is shown, but the display
still shows a tick, a delayed alarm is configured.
In this example, the most recently recorded temperature
is above the primary upper alarm threshold, but below a
secondary threshold and no alarm was triggered:
Here, the temperature was below the secondary alarm threshold, but above the tertiary
threshold; an alarm event has not been triggered:
An alarm was triggered
If an alarm event occurred the alarm indicator is
displayed in the upper left corner of the display.
Additionally, one of the alarm limit markers or is
displayed, to show if the event was an upper or a lower
alarm.
The alarm marker in this screen does not indicate which alarm threshold
caused the alarm, only that the most recent temperature recorded, which you see
on the display (18.8°C) was above the secondary threshold.
To identify which of the alarm conditions caused the alarm event you will need to
review the data on screen as described in Reviewing Alarms on the next page, or
open the PDF file on your PC.
Marking a reading with an inspection mark
When you press the button while the UTRED-16 is recording, the next reading taken will
be identified in the downloaded data and on the PDF report with an inspection mark.
The MARK symbol appears and remains shown until the next reading is taken.
If the Paused Readings function is enabled, you will also see the PAUSED symbol.
An inspection mark will also be recorded if you clear an alarm.
At any time you can review the recorded statistics by briefly pressing the STOP/Revie w
button. You can do this while the logger is still taking temperature readings, or when it
has already stopped, but not when it is plugged into a USB port.
After each subsequent press of STOP/Review the next review screen is shown. The number
of screens available for review depends on the number of configured alarms, and whether
or not the alarm review screens have been enabled. If no alarms are configured, or the
alarm review screens are disabled, only the MIN and MAX screens appear.
Figure 5: Review cycle with threshold screens disabled
Figure 7: Alarm configuration settings in LogTag® Analyzer
On first press of the STOP/Rev iew button the Maximum Reading screen appears.
The clock disappears, and the MAX symbol is shown along with the maximum temperature
recorded. If this temperature was above or below one of the alarm thresholds, the
corresponding alarm threshold arrow is also shown. Here, the maximum was above the
primary threshold.
When you press STOP/Review again, the Minimum Reading screen appears.
If this temperature was above or below one of the alarm thresholds, the corresponding
alarm marker is also shown. Here, the minimum was below the tertiary threshold.
If the alarm review screens are enabled, the next screens show the enabled alarm
threshold values in descending order, starting with the upper most alarm threshold. Up to
six screens will be shown, depending on how many alarms were enabled during
configuration.
1. In this example the tertiary upper alarm is enabled, indicated by the alarm threshold
marker shown in combination with the two small alarm arrows. The DURATION
symbol is shown, and the clock shows the total number of hours and minutes above
the threshold, instead of the time.
You can see how long the recorded temperature was above this threshold. Here, the
l Releasing the button within the 6-second period after the initial press switches to
the first review screen.
A min/max reset mark will be recorded in the logged data. The logger will now track new
min/max values, with the first update shown as soon as the next reading is taken.
You will be able to see in the chart in LogTag® Analyzer and in the PDF report when the
min/max values were cleared, but you will not be able to review previous min/max values
on screen once they have been cleared.
Clearing an Alarm
A UTRED-16 can be configured so you can clear an alarm from the display. To do this,
press and hold the STOP/Rev iew button.
After 2 seconds the alarm will be cleared and the will change to a . If you keep holding
the button for a further 2 seconds, the will show again and the alarm will not be cleared.
An inspection mark will be recorded against the next temperature reading taken, whether
you clear the alarm or not.
An alarm cannot be cleared once the UTRED-16 has stopped recording. Alarms
resulting from an accumulation of readings can also not be cleared.
During configuration of the UTRED-16 you can set the option to ignore up to 15 readings
for alarm and statistics calculations after either button is pressed. The readings are still
shown on the graph and in the data listing, but they are labelled as paused, and their
value is ignored when determining alarm trigger conditions, minimum/maximum values
and other statistical calculations.
Pausing readings is useful, for example, when you need to temporarily remove the
probe from the monitored location to inspect goods, but you do not wish to trigger
an alarm as a result of you handling the probe.
It allows the logger to acclimatize to the environment
again, before further readings are processed.
After a button press the display shows PAUSE D next to
the symbol. PAUSED will turn off as soon as the last
ignored reading has been recorded.
The option is set in the Advanced Settings tab during
configuration with LogTag® Analyzer and is expressed in
number of readings after the last button press.
Paused readings are specially marked in the graph and data
listings.
How long PAUSED is displayed depends on when between readings you press the
button. It will show longer, if the button is pressed just after the logger takes a
reading, but shorter if you press the button just before. For example, if you
configure a logging interval of 10 minutes and 2 paused readings, the time PAUSED
is shown could be as short as 10 minutes, but as long as 20 minutes.
Power Save
When Power Save is enabled, the display will automatically switch off if none of the
buttons have been pressed for 30 seconds.
This function is appropriate in applications where you don't need to look at the display
frequently, such as in transit monitoring applications, as the logger uses less battery
power when the display is not turned on.
Pressing any button will re-activate the display.
Power save is enabled or disabled when configuring the UTRED-16 via LogTag® Analyzer
If the battery is low while the UTRED-16 is still recording, the battery low symbol is
shown in addition to any other currently displayed information.
The product has reached the end of its life.
If you continue to use the UTRED-16 despite a low battery warning you may lose
readings, or the readings recorded may be inaccurate.
Probe is disconnected
The display shows instead of a temperature value if the logger's external
temperature probe is disconnected.
Any values recorded during this time will not be
taken into account when calculating statistics values
or alarms.
If, for example, the temperature was above 10°C for
12 hours (which would normally trigger an alarm
event), but the probe was disconnected for 3 hours
during this time, no alarm would be generated.
When evaluating data, any readings taken while the probe is disconnected are shown as -
-. - in the data list. The chart in LogTag® Analyzer will show a gap during this period.
The UTRED-16 automatically stops recording temperature when the maximum number of
readings specified during configuration has been reached. Your unit can also be set up to
stop when it is inserted into a USB socket. This option needs to be set up at the factory
and cannot be changed during configuration with LogTag® Analyzer. Your distributor can
supply more information about this option.
Manually
You can configure a UTRED-16 so it can be stopped with the button. This feature is useful
when you take the logger out of a shipment and don’t want to falsify the statistics with
readings taken after the shipment completion. The stop function is enabled in the
Advanced Options dialogue during configuration.
Press and hold the button. The STOPPED symbol shows in
addition to for 2 seconds.
The symbol turns off after 2 seconds. Release the
button and the logger stops taking temperature readings.
If the button is released while is still on, or you wait until STOPPED disappears, the
display shows again, and the logger continues to record data. In this case, an
inspection mark will be recorded in the data against the next temperature reading. The
MARK symbol will show, and the PAUSED symbol also shows if this feature is enabled.
Unlike many other LogTag® products with display, the alarm does not have to be
cleared before the logger can be stopped.
Logger has stopped
Once the logger has stopped, the display will show:
l the STOPPED symbol
l the OK indicator () if no alarm was recorded
l the alarm indicator () and one or both alarm threshold() indicators if an alarm
was recorded during the trip that has not been cleared
l the current time
l the battery low symbol () if the battery is low
The logger has stopped, and no alarms were recorded
during the trip.
Here, the logger has stopped, and an upper alarm was
recorded during the trip.
Resetting the Logger
You can reset a UTRED-16 back to its original READY state. Once reset, recording can be
started again.
When performing this procedure all recordings and statistics stored in the logger
are deleted and cannot be recovered. Please ensure your data has been saved!
All settings in the logger are retained.
For this process to work, the logger must be
STOPPED.
Press and hold the START/Mark button. The READY
symbol will turn on.
When the STOP PED symbol turns off, release the
button within 2 seconds. The logger is now ready to
be started again.
If the button is released while STOPPED is still on, or you wait until the READY symbol
disappears, the display shows STOP PED again, and the logger remains stopped. This
means, the data from the previous trip is still accessible.
The ability to reset a logger is enabled or disabled in the Advanced Settings tab when
configuring the UTRED-16 via LogTag® Analyzer.
As soon as you plug the UTRED-16 into a USB port, the word USb appears.
If you have configured the logger to generate files, USb remains turned on until all files
have been generated.
During the file generation a small animation will play on the display.
As soon as the file generation is finished, USb will blink every second to indicate these files
can now be accessed.
These files are generated each time you plug the logger into the USB port,
regardless of whether the logger is still recording readings or not.
What happens on your computer depends on the operating system of the computer, the
settings made during configuration and whether or not LogTag® Analyzer is running.
Microsoft Windows
Up to four drivers will now be installed, depending on the UTRED-16's configuration.
All drivers are part of the operating system and will typically not require administrator
privileges for your computer.
1. Mass Storage Device
2. LogTag Mass Storage USB Device
These two devices are required so you can access the data files in the same way as a
USB memory stick. These drivers will not be installed if the UTRED-16 does not
generate files.
3. USB Input Device (HID)
This device is used for communication to LogTag® Analyzer and its driver will always
be installed, even if LogTag® Analyzer is not present on the computer.
Typically, in these operating systems a new drive will be mounted, from which you can
open the PDF file. You will not be able to configure the UTRED-16 using either of these
operating systems, unless you use virtualization software such as Fusion or VirtualBox to
create a hosted Windows environment. You need to discuss these options with your
network administrator.
While a UTRED-16 is plugged into USB, no temperature readings are taken. The
graph will display a gap and the data list will show --- followed by the # symbol.
Accessing the files
If the logger was configured to generate files, a new drive letter or mounted device will
appear. The device name will be created from the serial number of the UTRED-16. You can
access the files by browsing to the newly created drive and double-clicking the PDF, CSV
or LTD files. For PDF files you need Adobe Acrobat Reader or a similar PDF viewer. To open
the LTD file you need to install the free LogTag® Analyzer software. CSV files can be
opened with a text editor, or imported into a spreadsheet program such as Microsoft®
Excel.
If a logger has only taken pre-start readings, PDF and CSV files will not be
available.
To retain the logger-generated files please copy them to a permanent storage
location on your computer, such as the Documents folder (they are not
automatically copied).
The data on the logger are retained, so each time you plug the UTRED-16 back into
the computer the files are re-generated, until the device is re-configured. Once the
battery is exhausted, the real time clock on the unit stops and dates and times for
the retained data may no longer be accurate.
A temperature scale is shown, which adjusts dynamically depending on the settings made
during configuration of the logger with LogTag® Analyzer.
Markers
The chart will show marks where the UTRED-16 has been downloaded (º) or where an
inspection mark has been placed with the START/Mark or STOP/Review button (•). It also
shows Paused marks (•) where a button press has prevented readings from being used for
statistics and alarm calculations.
USB Paused Readings
USB loggers cannot take a reading while plugged into a USB port. A gap is shown in the
graph where the UTRED-16 was plugged in at the time it would otherwise have taken a
reading. The list shows instead of the reading.
Recorded Data Chart
The chart shows a graphical representation of the data during the trip. As part of the
UTRED-16 configuration process you set the parameters that influence how the chart is
presented.
Legend
Shows the symbols for download marks, inspections marks and paused marks if they
appear in the readings.
File Information
This section shows general information about the PDF file, such as generation time, date
and time formats used in the chart and the data list as well as the file name, which is
compiled from information about the data it contains:
LogTag_[serial_number]_[trip number]_[file creation date]_file creation time]_[OK or ALM].pdf
Other files that may be generated have the extensions *.csv and *.ltd.
Logger Information and Configuration
This section shows general information such as serial and model numbers, trip number,
battery status and description. It shows how the logger was started, if a start delay was
active and the interval used for taking readings.
Logger Statistics Overview
This section gives a brief overview of the temperature data collected during the trip. It
shows minimum and maximum values, when these occurred and also shows average,
standard deviation and MKT values.
These two sections summarize the alarm trigger conditions and occurrences during the
trip, including:
l alarm threshold temperature value
l any delay value for consecutive or accumulative alarms
l the total time above or below a threshold
l how often an alarm occurred
l whether or not an alarm was generated for this alarm trigger
You may see up to three alarms for each direction, depending on the logger's
configuration.
Upper Threshold Lines
Upper alarm thresholds are shown with red dashed lines --- so you can see at a glance
where temperatures went above set limits. You may see up to three lines depending on
configuration and zoom settings.
Lower Threshold Lines
Lower alarm thresholds are shown with blue dashed lines --- so you can see at a glance
where temperatures went below set limits. You may see up to three lines depending on
configuration and zoom settings.
Page information
The current page number and the total number of pages appear on every page.
This section shows general information about the PDF file, such as generation time, date
and time formats used in the chart and the data list as well as the file name, which is
compiled from information about the data it contains:
LogTag_[serial_number]_[trip number]_[file creation date]_file creation time]_[OK or ALM].pdf
Other files that may be generated have the extensions *.csv and *.ltd.
Recorded Data list
The Data list shows a single row for each recorded reading, along with the date, time and
temperature values, plus any special events that were recorded against this reading.
The current page number and the total number of pages appear on every page.
Marked readings
In the data list, each entry may be marked with one or more of the following symbols,
depending on which event occurred when the reading was taken:
l A + symbol will be shown, if the UTRED-16 was downloaded with LogTag® Analyzer
just prior to the reading being taken.
l A * symbol will be shown where an inspection mark has been placed.
l A $ symbol will be shown where readings were paused as a result of a button press.
l A # symbol will be shown where the logger was plugged into a USB port at the time
it would otherwise have taken a reading.
l A % symbol will be shown when the min/max readings were reset.
USB Paused Readings
USB loggers cannot take a reading while plugged into a USB port. A gap is shown in the
graph where the UTRED-16 was plugged in at the time it would otherwise have taken a
reading. The list shows instead of the reading.
Probe Disconnected
A gap is shown in the graph where the UTRED-16's probe was not plugged in at the time it
took a reading. The list shows instead of the reading.
Legend
Shows the symbols for download marks, inspections marks and paused marks if they
appear in the readings.
Size 93mm(H) x 54.5mm (W) x 8.6mm (T) including USB cap (excludi ng probe)
Weight 39g (excludi ng probe)
Cal ibration Factory cali bration using i nstruments traceable to an IS O/IEC 17025 accredited testing laborator y
PDF fea ture s
0 to 95%RH
Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4 °F to +104 °F), typically ±0.3 ° C (0.6 °F)
Better than ±0.8 °C (±1.4 °F) for -40 °C to -20 °C (-40 °F to -4 °F), typically ±0.5 ° C (0.9 °F)
Better than ±0.8 °C (±1.4 °F) for +40 °C to +70 ° C (104 °F to 1 58 °F), typically ±0.7 °C (1.3 ° F)
Better than ±1 .2 ° C (±2.2 °F) for +70 °C to +99 °C (158 ° F to 210 °F), typically ±1 °C (1.8 °F)
Please see resolution chart in Accuracy and R esolution
Better than 0.1 °C (0.2 °F) for - 30 °C to +60 °C (-22 °F to +140 °F)
Better than 0.2 °C (0.4 °F) for -40 °C to -30 °C (-40 ° F to -22 °F)
Better than 0.5 °C (0.9 °F) for +60 °C to +99 °C (+140 ° F to +210 °F)
Please see accuracy chart in Accuracy and Resol ution
Rated temperature coefficient is -0.034±0.006ppm/°C (i.e. typi cally +/-0.00294seconds/day/°C )
16,1 29 r eal time temperature values, givi ng
l 11 2 days @ 1 0min logging or
l 168 days @ 15min logging.
l Max/Mi n values
l Alarm duration v alues
l Push button start with optional configurable start delay from 1 minute to 72 hours
l Date/time start
l three config urable lower alarms
l OK tick and Alarm cross o n display, linked to alarms
l Withstands 5 drops from 1 m onto smooth concrete floor wi thout l oss of function o r calibr ation
l EC EMC directives (EN 6 1000-6-1 :2005 & EN 61000-6-3:2006)
l Includes electrostatic discharg e as prescribed in EN 61000-4-2
l Co mpli es with FCC Part 15 Subparts A and B
Typically 24 months of operation w ith nor mal use (15 mi nute log ging, statistics reviewed on the
display no mor e than once daily for no longer than 30 seconds each time, download data monthly)
provi ded recorder is kept w ithin the storage temperature range when not in use.
l compliant w ith standard 1.6 and later
l Single page report with tri p and alarm summary
l Multi page r eport with list of readings including date/time