UTRID-16 Single-Use
UTRID-16 Multi-Use
ENGLISH
User Instructions
®
The
LogTag
temperature logger, which monitors and records the temperature
of the surrounding environment and shows any alarm conditions
on the built-in display. Alarm events can be reviewed on the
display or downloaded to a PC via the built-in USB plug and
analyzed using PDF soware such as Acrobat Reader.
Preparing the Logger
The UTRID-16 is shipped to you uncongured and must be set
up with the parameters required for starting and recording
temperature values. This is done using LogTag Analyzer soware,
which you can download for free from https://logtagrecorders.
com/soware/lta3 (you can also use this soware for
downloading and analyzing data in case the PDF report doesn’t
contain sucient detail). Please refer to the separate LogTag
Analyzer Quickstart guide for information on how to congure
and download the logger, and how to analyze the data.
Once congured, replace the protective cap. The logger is now
ready to start.
Starting the Logger
UTRID-16 is a fully congurable USB PDF
The display must show
READY
before the logger can be
started. To start the logger, press
the
STA RT/Ma rk
If the logger was congured with a
start delay, you will now see the
DELAY
starts a countdown timer, during
which no temperatures are
recorded.
button.
symbol. The UTRID-16
When the countdown has completed, the logger will start
recording temperatures at the congured intervals and monitor
alarm conditions.
begins.
If no start delay was congured,
the logger will immediately start
recording.
The logger should now be placed
with the goods, so it reaches their
temperature when the recording
During Recording
While the UTRID-16 is recording, the display shows:
• The last recorded temperature
• so you can identify it is recording
• The current time in hours and minutes
• A tick if no alarm events were triggered
• The alarm indicator if an alarm event occurred, and one of
the limit markers ( or ) so you can see if an upper or lower
alarm was triggered.
• One or more threshold arrows ( or ) to show if the
current temperature is above or below any of the alarm
thresholds
Example Screens
While the temperature readings stay within the pre-congured
limits, the OK symbol is displayed, as shown in the image
above.
outside the limits for the time set during conguration, an alarm
event is triggered.
acceptable levels, the Reject symbol and the limit marker remain
shown to indicate a previous alarm event, while the threshold
arrows turn o.
This symbol
shows ...
If the most recently recorded
temperature is above or below one
of the alarm thresholds, a
threshold arrow will be shown on
the display.
If the temperature continues to be
Once an alarm event is triggered,
the Reject symbol is displayed.
The limit marker
( or ) indicates the direction
of the alarm.
When the temperature returns to
... if the most recently recorded
temperature was
above the primary upper alarm
threshold, but below the secondary
above the secondary upper alarm
threshold, but below the tertiary
above the tertiary upper alarm
threshold (highest alarm)
below the primary lower alarm
threshold, but above the secondary
below the secondary lower alarm
threshold, but above the tertiary
below the tertiary lower alarm
threshold (lowest alarm)
Placing a Mark in the Readings
Each time you press the
Mark
button a mark is recorded in
the data. This is shown on the PDF
and in the data le and can be
used to
inspection. The
next reading is recorded.
Clearing and Alarm
You can clear an active alarm by pressing and holding the
STA RT/Ma rk
limit markers turn o.
is recorded in the data. The option to clear an active Alarm is set
during conguration.
MARK
button until the cross changes to a tick, and the
identify events such as a vaccine
symbol is shown on the display until the
MARK
is shown, and an inspection mark
STA RT/
Paused Function
Pressing either button excludes the next X readings from the
alarm and statistics calculations; during this time
be shown. X can be between 0 (feature disabled) and 15, and is
set during conguration. This allows you to review the statistics
or clear an alarm without causing an invalid reading, alarm or
statistic.
PAUSED
will
Stopping the Logger
When the shipment has reached its destination, you must
retrieve the UTRID-16 from the package and immediately stop
it, so the device does not generate false alarms. To do this, press
and hold the
STOP/Review
symbol changes from ashing to permanently on (aer approx.
4 seconds), then release the button. Holding the button for
longer than 6 seconds will abort this process and the logger will
continue to record.
The UTRID-16 will also stop automatically if you have congured
a xed recording length.
When stopped, the display will show:
•
STOPPED
to show the Logger is no longer recording
temperature
• The current time in hours and minutes
• A tick if no alarm events were triggered
• The alarm indicator if an alarm event occurred, and one of
the limit markers ( or ) so you can see if an upper or lower
alarm was triggered.
Once stopped, no additional readings will be taken or processed.
( ) button until the
STOPPED
Reviewing the Data
You can review the trip data on the display, either during logging,
or aer the recorder has stopped.
To show the rst review screen, press the
It shows the maximum temperature reached during the trip.
Pressing
STOP/Review
reached during the trip.
Each subsequent press of the
6 additional review screens, depending on the number of alarm
trigger conditions set during conguration.
STOP
Review
STOP
Review
STOP
Review
STOP
Review
These screens show each congured alarm threshold
temperature, and the time recorded during the trip above this
temperature, in descending order.
You can disable reviewing the alarm thresholds during
conguration of the logger.
When the last review screen is shown, pressing
shows the initial review screen again.
If the
STA RT/Ma rk
review, or neither button is pressed for 30 seconds, the or
STOPPED
screen is shown.
again shows the minimum temperature
START
Mark
button is pressed at any time during the
STOP/Review
STOP
Review
STOP/Review
button shows up to
STOP
Review
STOP
Review
STOP
Review
STOP
Review
STOP/Review
button.
PDF
You can view a PDF of the recorded data by plugging the logger
into a USB socket of any PC capable of displaying PDF les. PDF
reader soware is required, such as Acrobat Reader or similar.
At this stage, make sure no other LogTag Soware is running
on your PC.
The PDF contains a summary of the trip, alarm details, a chart
and a list of the recorded temperatures. Which details are shown
on the PDF is set during conguration.
It is possible to plug the UTRID-16 into a USB socket while it is
still recording, but not recommended. You will be able to view
the PDF le, but during this time the Logger will not record any
temperature data, and will be shown on the display.
Getting more information
For more information, please read the complete UTRID-16
Product User Guide, available from
https://logtagrecorders.com/product/utrid-16/
This guide contains additional content such as:
• How to interpret the PDF report and data list
• Which other symbols you may encounter on-screen
• How to reset a multi-use logger for another trip
• How to use pre-start readings for peace of mind
Attention: The UTRID-16 monitors temperature exposure
and not the product quality. Its purpose is to signal if product
quality evaluation/testing is required.
Battery
The UTRID-16 contains a Lithium Battery. If this symbol (
) is shown, the battery is low. A logger with a low battery cannot
be started but will have sucient capacity to complete an already
started trip.
Dispose or recycle the battery/logger in accordance with your
local regulations.
Do not expose the logger to extreme temperatures as it may lead
to the destruction of the battery and may cause injuries. Keep out
of reach of children.
Liability
The manufacturer shall not be held liable:
a. if the device was used beyond the manufacturer’s given
limitations;
b. for any claims due to the improper storage and use of the
device;
c. for any problems with refrigeration units;
d. for the bad quality of the monitored goods, if any;
e. for incorrect readings if the device was used with an activated
low battery sign; or
f. for consequential loss.
Useful Life
The operational life of the UTRID-16 is 1 year of operation ( 2
years for the re-usable model) subject to following conditions:
• The logger was not stored for more than 24 months prior to
activation.
• The logger’s display is not excessively activated (for example,
reviewing alarms several times a day).
• The logger is stored and operated inside the recommended
operating parameters specied by the manufacturer.
UTRID-16 Quick Start Guide, English, Revision A (220615)
Copyright © 2022 LogTag North America Inc. All Rights Reserved.
LOGTAG is a registered trademark of LogTag North America, Inc.