LogTag TREL30-16 User Manual

LogTag Recorders
TREL30-16
Ultra-Low
Temperature
Recorder with
Display
Product User
Guide
Published 27. February 2019
Copyright © LogTag Recorders, 2004-2019
www.logtagrecorders.com
TREL30-16 User Guide 2
Contents
Safety Information 3 Liability 3 Battery Life 3 Disclaimer 3 Typographical Conventions 3 Introduction 4 Features 4
Case 4 Buttons 4 Display 5 Sensors 5
What You Need 5
Required Equipment 5 Optional Items 5
Configuring the TREL30-16 6 Standard Configuration Options 6
TREL30-16 Start Options 7 Audible Alarm 7 Alarm Configuration Options 8 Advanced Configuration Options 9
Display Overview 10 Real Time Clock 12 Adjusting the Display Clock 12 Starting the Logger 13
Push button start 13 Push button start, quickstart disabled 14 Push Button Start with Start Delay 15 Automatic date/time start 15
During Recording 16
Marking a reading with an inspection mark 16 Clearing an Alarm 17 Paused Readings 17 Sensor is disconnected 18 Power Save 18 Total Number of Days Statistics Collected 19
Reviewing Day Statistics directly on the TREL30-16 20
Trip Minimum/Maximum Temperatures 23 Resetting the Trip Minimum/Maximum Temperatures 23
Stopping the TREL30-16 25
Automatically 25 Push-button stop 25 Push-button stop, quick stop disabled 26
Resetting the Logger 27 Hibernating a TREL30-16 27 Technical Specifications 29 Appendix 1 - Battery Replacement 30
Safety Information TREL30-16 User Guide 3

Safety Information

The TREL30-16 temperature logger contains a user-replaceable Lithium Battery. When the battery indicates “LOW”, you can replace it as per the instructions in Appendix 1 - Battery Replacement on page30. The empty battery should be recycled or disposed of according to your local regulations. Do not expose the logger to extreme temperatures as it may lead to the destruction of the battery and may cause injuries. Keep out of the reach of children.

Liability

LogTag Recorders’ standard warranty terms apply. A copy can be requested by emailing support@logtagrecorders.com. In addition, LogTag Recorders shall not be held liable:
l If the device was used beyond LogTag Recorders’ stated limitations. l For any claims due to the improper storage and use of the device. l For any problems with refrigeration units. l For the bad quality of the monitored goods, if any. l For incorrect readings if the device was used with a low battery. l For consequential loss.

Battery Life

The main battery in the TREL30-16 is designed to power the device for up to 12 months, provided ...
… a fresh battery from a reputable manufacturer is used, … the device is not downloaded excessively (more than once a week), … the recording interval is not shorter than 5 minutes, … the acoustic alarm is not active over long periods … the device is stored and operated according to LogTag Recorders’ recommendations.

Disclaimer

The TREL30-16 monitors temperature exposure and not the quality of the goods it accompanies. Its purpose is to signal if product quality evaluation/testing is required.

Typographical Conventions

Text in this font refers to buttons on the TREL30-16. Text in this font refers to option settings, dialogue boxes or actions to be taken in LogTag® Analyzer.
Text in this font describes features of the product.
Introduction TREL30-16 User Guide 4

Introduction

The LogTag® TREL30-16 low temperature logger features a data logging memory storing up to 15,905 temperature readings and a separate statistical memory, storing maximum and minimum reading as well as alarm duration for each of the last 30 days. During recording the display shows the temperature of the most recent reading, whether or not this is within or outside the acceptance range, an alarm trigger summary of the last 30 days (today and 29 days previous), the current time and battery status. Alarm events can be triggered when a number of readings are outside preset Alarm thresholds and a “day alarm indicator” appears on the display. Logged temperature data can be downloaded via a standard LogTag® Interface to the free companion software LogTag® Analyzer, where you can display data in chart, list or summary formats. The software also allows electronic archiving and exporting or transmitting data in support of sophisticated data management systems.

Features

The TREL30-16 temperature logger features the familiar LogTag® case layout. 
Figure 1: TREL30-16 features

Case

l Mounting lug for secure fastening of logger to fixtures l Gold-plated, high-quality temperature sensor socket l Robust polycarbonate case, IP64 l Durable communications contacts

Buttons

l START/CLEAR/STOP button; can be used to start and stop the unit or to clear an alarm. It is also used
to exit the statistics review.
l REVIEW/MARK button; can be used to enter the statistics review and to scroll through the statistical
data directly on the display. It is also used to place an inspection mark in the data listing.
What You Need TREL30-16 User Guide 5

Display

The display shows ‘at a glance’ if alarm events have occurred for both the current day and up to 29 days in the past. Details of any alarm event can be checked directly on the unit by inspecting the statistics history on the logger's display or in more detail by downloading the logged data.

Sensors

The TREL30-16 will accept any sensor from the ST10 product range.

What You Need

Required Equipment

In addition to your LogTag® TREL30-16temperature logger you will need the following items:
l A USB interface for communication to the PC l An external probe of the ST10 series l For configuration - a PC running Windows 7 or later and LogTag® Analyzer installed.

Optional Items

In addition to the above, following items are useful accessories:
l A wallmount bracket, which can be used, for example, to attach the TREL30-16 to the side of a
cabinet.
Configuring the TREL30-16 TREL30-16 User Guide 6

Configuring the TREL30-16

Before a TREL30-16 logger can be used, it must be configured with the parameters required for starting and recording temperature values . This is done using LogTag® Analyzer software, which is also used for downloading and analyzing data.
l Start the LogTag® Analyzer software. l Plug a LogTag® interface cradle into a USB port and wait for the drivers to be installed. You can use
more than one cradle to configure multiple devices at the same time, however it is practical to limit the number of units to about 6, using a powered USB hub.
l Insert the TREL30-16 into the interface with the three contacts towards the back until it clicks in
place.
l Select LogTag - Configure; LogTag® Analyzer will display the configuration options for any connected
TREL30-16 loggers.

Standard Configuration Options

The standard configuration options are very similar to those of other LogTag® logger products and include settings such as User ID, start method, pre-start recording, logging interval and duration, start delay, and
password.
For detailed information about each parameter please read the section about Configuring a LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.
Standard Configuration Options TREL30-16 User Guide 7

TREL30-16 Start Options

During configuration with LogTag® Analyzer you can decide when the TREL30-16 starts taking temperature readings.
l ... via Date/Time start : The logger will start taking temperature readings at the date and time you
specify (local time).
When you choose the push button start option, you can select to record pre-start readings or Begin
recording after a delay.
l If you configure the TREL30-16 to start after a delay period, the logger will not immediately record
temperature readings after you have pressed START/CLEAR/STOP, but start a countdown timer instead, and record readings only after the timer has ended. It will, however, continue to take pre-start readings, if enabled.
You cannot combine a date/time start with pre-start readings or the start delay function.

Audible Alarm

The TREL30-16 is fitted with a buzzer, which can be turned on when an alarm event has been registered to provide extra feedback. This function is enabled or disabled when configuring the logger with LogTag® Analyzer.
Please note, that continual activation of the audible alarm will reduce the working life of the battery. When an alarm event is triggered, the alarm should be cleared as soon as possible.
The alarm will sound once every four seconds for the first 24 hours. To preserve battery life, the interval at which the alarm sounds will increase to 8 seconds when the clock passes midnight after the initial 24 hour period (i.e. the alarm will sound every 4 seconds between 24 and 48 hours, depending on when during the day the alarm was first triggered). When the clock passes midnight again, the interval changes to 12 seconds. It will remain at 12 seconds until the alarm is cleared, the unit stops or is re-configured.
Note: The interval will increase, even if the alarm is re-triggered! Please see more information about
The beeper will temporarily turn off when you are reviewing data.
Standard Configuration Options TREL30-16 User Guide 8

Alarm Configuration Options

The TREL30-16 can display an alarm if either of the configured alarm trigger conditions have been met. This is indicated on the display by showing the Alarm Indicator ( ) and the Day Alarm Marker for today ( ).
Each alarm trigger condition consists of a threshold temperature value, an activation type (which can be instant, consecutive or accumulative1) and a delay time, if it is not an instant alarm. If an alarm trigger condition requires readings to exceed an upper threshold temperature it is called an
upper alarm. If an alarm trigger condition requires readings to go below a lower threshold it is called a lower alarm.
All alarm trigger conditions are configured in the Alarm Settings tab during configuration of the logger with LogTag® Analyzer. The screen shows an example where:
l the upper alarm is triggered when the temperature is -70.0°C or above for an accumulative time of
10 hours.
l the lower alarm is triggered when the temperature is -80.0°C or below continuously for 1 hour.
Figure 2: Sample Alarm Configuration Settings for the TREL30-16
Once an alarm has triggered, the alarm indicator ( ) remains shown until the alarm is cleared (see Clearing
an Alarm on page17) or the unit is reconfigured. The day alarm marker remains shown until midnight,
then it turns off and the marker for the previous day is shown ( ) to indicate the alarm was registered against what is now the previous day. When midnight passes next, this marker will move to and so on.
Note: As soon as an alarm is triggered, the corresponding delay time (but not any others) resets to zero and starts again. The alarm processing for all other alarm delays is not affected. Therefore, accumulative or consecutive alarms will re-trigger, if the alarm conditions are met again, and the Alarm Indicator ( ) and the Day Alarm Marker ( ) will be shown, even if any previous alarm was cleared. Clearing an Alarm does not reset any of the delay values.
Alarm Re-triggering
1
l
Instant = one temperature reading is above (below) the threshold
l
Consecutive = temperature readings are above (below) the threshold for the time defined in the activation delay without interruption
l
Accumulative = temperature readings are above (below) the threshold for the total time defined in the activation delay time, but may not necessarily be sequential.
Standard Configuration Options TREL30-16 User Guide 9

Advanced Configuration Options

Select Advanced Options for additional configuration settings. These settings decide how some of the elements are displayed on the screen and set certain options specific to the TREL30-16.
Figure 3: TREL30-16 advanced configuration screen
Parameters that influence the appearance of the display are
l Pause alarm/statistics processing (see Paused Readings on page17)l Temperature display unit (see Display Overview on page10)l Switch off display after 30 seconds (Power save, see Power Save on page18)l showing the number of days on the display, for which statistical data have been collected (see Total
Number of Days Statistics Collected on page19)
Parameters influencing specific behaviour of the TREL30-16 are
l clear and reset alarm when the START/CLEAR/STOP button is pressedl leaving the alarm turned on, even if readings return to the normal temperature range again l allowing the user to stop the logger with the START/CLEAR/STOP buttonl allowing the user to reset the logger with the START/CLEAR/STOP buttonl enable or disable the quick start/stop mode (please see Push button start on page13 and Push
button start, quickstart disabled on page14).
Enter the desired options and click OK to accept the new values. This returns you to the standard configuration dialogue.
Display Overview TREL30-16 User Guide 10

Display Overview

Figure 4: Display Overview with all segments turned on
Day Alarm Markers
This grid shows 3 rows of 10 markers, named Today ( ) to Day -29 ( ), which are switched on when an alarm event occurred on that day.
Temperature
This shows the last recorded temperature while the TREL30-16 is recording. Once the logger has stopped, nothing will be displayed.
Day Number
During Review, this shows the day number of the currently displayed day statistic. Today is 0 0 DAYS, days in the past are represented between yesterday - 01 DAYS and - 29 DAYS. During configuration of the logger in LogTag® Analyzer you can also enable this to show the total number of days on which data were collected while recording.
Battery Low
A battery test is performed hourly. The battery low symbol will appear if the TREL30-16's battery is low and requires changing. Please follow the instructions provided in Appendix 1 - Battery Replacement on page30. If the symbol is not shown while the display is turned on, the battery is still OK.
ALARM/OK indicator
The symbol is shown as soon as the TREL30-16 has registered an alarm event. If no alarms have been registered, or if an existing alarm has been cleared, the symbol is shown.
Recording state indicators
The recording state indicators show if the TREL30-16 is currently recording data.
l
If the READY symbol is shown, the TREL30-16 is ready to be started. Depending on the configuration it may already record pre-start readings.
l
If the STARTING symbol is shown, the logger has been configured with a start delay. The word
DELAY is also shown, together with the time in hours and minutes until the start.
l If the symbol is shown, the TREL30-16 is recording temperatures at the sample interval defined
during configuration with LogTag® Analyzer.
l
If is shown together with the word PAUSED, the product is also recording, but the recorded values are not taken into account when calculating alarm events and durations.
l
If the word STOPPED is shown, the TREL30-16 has finished recording temperature data.
Display Overview TREL30-16 User Guide 11
Time Value and Time Indicators
The time value display is used to show one of the following:
l a clock with the current time, l time remaining to start of logging with a delayed start or l duration, for example of an alarm.
The time indicators identify, which of those is displayed:
l
If the word TIME is shown, the time value represents the current time in hours and minutes (24- hour format).
l
If the word DELAY is shown, the time value represents a start delay in hours and minutes.
l
If the word DURATION is shown, the time value represents the length of time in hours and minutes an alarm threshold was exceeded, for example the time above the upper alarm threshold.
The word d t st appears in place of a time value if the logger is configured for a date/time start, and the start time has not yet passed.
Reading Type
The word CURRENT is shown when the temperature on the display represents the last recorded temperature. The word MAX is shown in Review mode, when the temperature on the display represents the maximum recorded
temperature for the day displayed. The word MIN is shown in Review mode, when the temperature on the display represents the minimum recorded temperature for the day displayed.
Alarm Threshold Markers
The up-arrow is shown when the temperature displayed (i.e. last recorded) is above the specified upper temperature threshold. The down-arrow is shown when the temperature displayed is below the specified lower temperature threshold.
Temperature Units
Depending on the selected display temperature units during configuration, this shows either°F or°C.
Real Time Clock TREL30-16 User Guide 12

Real Time Clock

The time shown on the recording display is linked to the logger’s internal real time clock. A day change occurs when the display time rolls through midnight (i.e. 00:00). This is the primary function of the display clock. Each time the logger is configured with LogTag® Analyzer the display clock value is set to the PC's current local time (or timezone).
Note that the logger’s internal real time clock value is only updated when the recorder is configured with LogTag® Analyzer. This prevents the data logging becoming discontinuous, which would be the case if the real time clock were to be changed together with the display clock.
LogTag® Analyzer can display the logged readings in a number of different time zones, regardless of where it was configured. Please note, however, that the PDF's time zone has to be set at configuration and cannot be changed after the logger has been downloaded.

Adjusting the Display Clock

The display clock of the logger can be set to the current local time, either by using LogTag® Analyzer software or directly on the unit, using the buttons.
To set the clock, press and hold the button, then press and hold the button. Keep holding both buttons together continuously for a period of 8 seconds. Press the buttons firmly! During this period the CLOCKADJ icon flashes. Release the buttons when the flashing stops. The clock can now be adjusted.
Initially, the minutes digits flash.
Press the button to increment the minutes digit (once it reaches
59 it rolls back to 00 on the next press).
Accept the minutes value by pressing . The hours value now flashes.
Press the button to increment the hours digits (once it reaches
23 it rolls back to 00 on the next press).
Accept the hours value by pressing . The new clock value is now stored, and the display shows the normal screen.
NOTE: It is advised that the real time clock is only adjusted when the product is not recording (STOPPED or READY) to avoid large shifts in day boundary data.
If a display clock adjustment is made while in RECORDING mode, the next log taken will be identified in the downloaded data with a time change mark.
Changes to the display clock do not affect the internal real time clock value, so the logged data does not show time gaps. 
Starting the Logger TREL30-16 User Guide 13

Starting the Logger

Push button start

The logger must be in READY mode for it to be started. Current time and the READY symbol are shown.
Press and hold the START/CLEAR/STOP button. First, STARTING is shown in addition to READY:
Then READY disappears:
Once READY disappears, release the button within two seconds. STARTING will also disappear, and the
symbol will be shown. The TREL30-16 now records temperature data.
The logger will not start if you…
…
release the button before READY disappears
…
keep holding the button for more than 2 seconds after READY disappears.
Starting the Logger TREL30-16 User Guide 14

Push button start, quickstart disabled

Again, the logger must be in READY mode for it to be started.
Press and hold the START/CLEAR/STOP button. The STARTING symbol starts flashing for about 4 seconds. Then the symbol remains on permanently for 2 seconds. If you release the button during this time, The
symbol is shown, the STARTING symbol turns off and the logger starts taking temperature readings.
If the button is released earlier than 4 seconds or later than 6 seconds, the display shows READY again, and the logger does not start.
Starting the Logger TREL30-16 User Guide 15

Push Button Start with Start Delay

If the logger has been configured for push button start with a start delay (regardless of quickstart settings), the word DELAY is shown instead of the symbol.
The delay time is shown in hours and minutes. The time counts down and the TREL30-16 starts recording when it reaches 0:00.
The timer can be cancelled and the logger reset to READY. While STARTING and DELAY are shown, press and hold the START/CLEAR/STOP button. STOPPED will now show. Release the button when STARTING disappears.
STOPPED and DELAY will both disappear, and READY will show, together with the current time, and the
logger can now be started again as normal. The TREL30-16 will not be reset if you …
…
release the button before STARTING disappears, or
…
if you keep holding the button until STOPPED disappears.

Automatic date/time start

If you configured the TREL30-16 for a date/time start, it will start recording temperature values as soon as the entered start time is reached. The logger will display the following if a Date/Time Start has been chosen:
Hibernating the logger using LogTag® Analyzer will abort any previously configured Date/Time start.
During Recording TREL30-16 User Guide 16

During Recording

During normal operation the display shows the most recently recorded temperature. This temperature is updated at the same rate as the logging interval. The current time is also displayed (in 24 hour format). If an alarm event is registered, . At the bottom of the display you can see an alarm day summary, where any days on which an alarm was recorded are highlighted. Following are some sample display screens:
At 1:29 pm the display shows the following:
l Alarm events were recorded 7, 19, 20 & 25 days ago l These alarms were cleared by an inspector, as the display currently shows the OK tick ( ). l The current temperature is over the upper alarm threshold, as indicated by the upper alarm marker (
).
l the duration, however, of this temperature excursion has not yet triggered an alarm
At 1:49pm (20 minutes later) the display shows the following:
l The temperature has remained above the upper alarm threshold and has now triggered an alarm
event. l The alarm symbol is shown to indicate an alarm event occurred l The day marker for the current day ( ) is shown.
At 1:49am on the next day the display shows the following:
l The temperature has returned to within the accepted range (none of the alarm threshold markers are
visible), but the alarm remains present, as it has not been inspected and cleared l The day summary has shifted by 1 day as the display time has passed through midnight (00:00).

Marking a reading with an inspection mark

When you press the button, the next reading taken by the TREL30-16 will be identified in the downloaded data with an inspection mark.
During Recording TREL30-16 User Guide 17

Clearing an Alarm

During configuration with LogTag® Analyzer you can allow a user to clear an alarm on the display. This is a useful function for an inspector, so repeated alarms can be recognised easier. 
This display screen shows an existing alarm that occurred yesterday, but has not yet been cleared. To clear an alarm (quickstart enabled), press and hold the
START/CLEAR/STOP button.
If the quickstart option is disabled, the symbol will flash for approx. 6 seconds. When the symbol remains permanently on, release the button within 2 seconds. The symbol now turns off, and the alarm is cleared. The alarm will not be cleared if you …
… release the button before the symbol remains
permanently on, or
… if you keep holding the button after the symbol
remains permanently on for more than 2 seconds.
 In this example the paused function was activated, and the paused symbol is shown on the display. For more information about paused readings please see the section about Paused
Readings below.
Note: Only the Alarm Indicator ( ) can be cleared! The day alarm marker ( in the above example) remains shown, as it is part of the statistic summary. Clearing an alarm also does not reset any of the delay values. Please see Alarm Configuration for additional information.

Paused Readings

During configuration of the TREL30-16 you can set the option to ignore up to 15 readings for alarm- and statistics calculations after either button is pressed. The readings are still shown on the graph and in the data listing, but they are labelled as paused, and their value is ignored when determining alarm trigger conditions, minimum/maximum values and other statistical calculations.
This function is useful, for example, when you wish to temporarily remove the sensor from the monitored location to inspect goods, but you do not wish to trigger an alarm due to your handling.
It also allows the logger to acclimatise to the environmental conditions again, before further readings are processed.
During Recording TREL30-16 User Guide 18
After a button press the display shows the word PAUSED next to the symbol.
The option is set in the Advanced Settings tab during configuration with LogTag® Analyzer and is expressed in number of readings after the last button press. Paused readings are specially marked in the graph and data listings. .
PAUSED will turn off as soon as the last ignored reading has been recorded. It will show longer, if your last
button press was just after the logger took a temperature reading, but will be displayed for a shorter period if you pressed the button just before a reading is taken. For example, if you configure a logging interval of 10 minutes and 2 paused readings, the time for PAUSED to show could be as short as 10 minutes, but as long as 20 minutes.

Sensor is disconnected

The display shows --.-, if the external sensor of the TREL30-16 is disconnected. This function is useful if the sensor is permanently mounted , but you wish to take the TREL30-16 to a computer for downloading its data.
Any values recorded during the time the sensor is disconnected will not be taken into account when calculating statistics values or alarms. If, for example, the temperature was above 10°C for 12 hours (which would normally trigger an alarm event), but the sensor was disconnected for 3 hours during this time, no alarm would be generated. , any readings taken while the sensor is disconnected are shown as --.- in the data list. The graph will show a gap during this period. If the sensor was disconnected for a complete day, the minimum and maximum statistics for that day will show --.­on the display and --.- on the list.

Power Save

When Power Save is enabled, the display will automatically switch off if none of the buttons have been pressed for 30 seconds. This function is appropriate in applications where you don't need to look at the display frequently, such as in transit monitoring applications, as the logger uses less battery power when the display is not turned on. Pressing any button will re-activate the display.
Power save is enabled or disabled when configuring the TREL30-16 via LogTag® Analyzer in the Advanced
Settings tab.
During Recording TREL30-16 User Guide 19

Total Number of Days Statistics Collected

The logger can be configured to show the total number of days for which statistics data was collected. This feature is enabled when configuring the TREL30-16 via LogTag® Analyzer and can be located in the
Advanced Options dialogue (see Advanced Configuration Options on page9).
When enabled, the total number of days for which statistics data was collected are shown on both the
Recording Display and the Stopped Display.
Figure 5: TREL30-16 recording, showing ‘18’ days of data has been collec ted.
Figure 6: TREL30-16 stopped, showing ‘27’ days of data has been collected.
Note, this is not always the number of days the logger has collected data. Depending on the sampling interval the logger can hold more than 30 days of temperature readings, but since the maximum number of days for which statistics can be collected is 30, the highest number that can be shown here is 30.

Reviewing Day Statistics directly on the TREL30-16

TREL30-16 User Guide 20
Reviewing Day Statistics directly on the TREL30-16
The review of day statistics history is accessed by pressing the REVIEW/MARK button . You can review the data regardless of whether the logger is still logging data, or has already been stopped. Following are some sample display screens you might see during a statistics review. All display screens
are based on the alarm settings made in Alarm Configuration Options on page8:
Today's data
Pressing the button displays the current day’s maximum statistic:
l
The Today marker flashes and 00 DAYS is shown to indicate that today's data are being
displayed. l Today, no temperature values were recorded above the upper alarm threshold, and no alarm event
was generated. l The maximum temperature recorded today was -75.6°C. l The paused function is enabled
Pressing the button again displays the current day’s minimum statistic:
l
still flashes and 00 DAYS is still shown, as the same day's data (today) is being displayed.
l Today, no temperature values were recorded below the lower alarm threshold, and no alarm event
was generated. l The minimum temperature recorded today was -77.3°C.
Yesterday's data
Pressing the button now displays yesterday’s maximum statistic:
l
The -1 marker flashes and -01 DAYS is shown to indicate that yesterday's data is being
displayed. l Yesterday, recordings above the upper threshold were recorded, indicated by the upper alarm
threshold marker l the duration value shows the amount of time recorded above the limit yesterday, which was 12 hours
30 minutes
Reviewing Day Statistics directly on the TREL30-16
l an alarm was triggered, indicated by the , as this duration was longer than the allowed time above
the threshold l The maximum temperature recorded yesterday was -58.9°C.
Pressing the button now displays yesterday’s minimum statistic:
l
still flashes and -01 DAYS still shows, as we are still looking at yesterday's data.l Yesterday, no temperature values were recorded below the lower alarm threshold. The alarm for the
day was generated by the upper alarm, not by the lower alarm, so . l The minimum temperature recorded yesterday was -74.8°C.
TREL30-16 User Guide 21
Data from the day before yesterday
Pressing the button now displays the maximum statistic from two days ago:
l
The flashes and -02 DAYS is shown to indicate that the data being displayed is from two days
ago. l Two days ago recordings above the upper threshold were recorded, indicated by the upper alarm
threshold marker l The duration was 1 hour 20 minutes, which was shorter than the allowed period, so no alarm event
was generated l The maximum temperature recorded on the day before yesterday was -64.8°C.
Pressing the button now displays the minimum statistic from two days ago:
l
still flashes and -02 DAYS still shows, as we are still looking at the data from 2 days ago.l Yesterday, no temperature values were recorded below the lower alarm threshold, and no alarm
event was generated . l The minimum temperature recorded on the day before yesterday was -73.5°C.
Data from 7 days ago
After pressing the button for a few times (skipping days -3 to -6) the maximum statistic from seven days ago is displayed:
Reviewing Day Statistics directly on the TREL30-16
l
The flashes and -07 DAYS is shown to indicate that the data being displayed is from seven days
ago. l Seven days ago recordings above the upper threshold were recorded, indicated by the upper alarm
threshold marker l The duration was 6 hour 30 minutes, which was shorter than the allowed period, so no alarm event
was generated l The maximum temperature recorded seven days ago was -61.8°C.
TREL30-16 User Guide 22
Data from 8 days ago
After pressing the button twice (skipping the minimum statistics for day -7) the maximum statistic from eight days ago is displayed:
l
The flashes and -08 DAYS is shown to indicate that the data being displayed is from eight days ago.l Eight days ago recordings above the upper threshold were recorded, indicated by the upper alarm
threshold marker l The duration was 5 hour 30 minutes l An alarm event was generated . Neither duration from day -7 or -8 by itself would cause an alarm
event, however the combined duration is 10 hours 50 minutes, which is more than the allowed 10
accumulative hours. l The maximum temperature recorded eight days ago was -64.2°C.
Tip: Pressing and holding the button will get you back to the previous screen, rather than advancing one screen.
Reviewing Day Statistics directly on the TREL30-16
TREL30-16 User Guide 23

Trip Minimum/Maximum Temperatures

You can configue the TREL30-16 so it tracks the minimum and maximum temperatures since it started logging. This feature needs to be enabled during configuration of the logger in the Advanced Configuration
Options on page9. Once enabled, the minimum and maximum values will be shown when the statistics data are reviewed, before any of the day statistic are shown.
Press the REVIEW/MAR K button . Instead of today's statistic data, the first screen shows the maximum temperature recorded during the trip to date, identified by the symbol (User resettable):
Pressing the REVIEW/MARK button again shows a screen with the minimum temperature recorded during the trip to date.
Pressing the REVIEW/MARK button again will then show the first day summary screen, and you can continue to review the remaining screens.

Resetting the Trip Minimum/Maximum Temperatures

The currently stored Min/Max temperature values can be reset at any time while the unit is recording, but not once the unit has been stopped. To reset the values, follow this procedure:
l press REVIEW/MARK to display the trip maximum value
l press REVIEW/MARK again to display the trip minimum value
Reviewing Day Statistics directly on the TREL30-16
You can only complete the next steps if you are viewing the Minimum temperature screen!
l press and hold REVIEW/MAR K . After 1 second the following screen sequence will show, and then
repeat:
TREL30-16 User Guide 24
During the cycle, the buzzer will sound a series of short beeps. l After approx. 6 secondes, following screen will show:
Release the REVIEW/MARK button within 2 seconds. The values will be reset and the normal
recording screen will be shown. l if you keep holding the REVIEW/MARK button beyond the 2 seconds, the currently stored min/max
values will be retained. l releasing the button within the 4 second period after the initial press switches to the first review
screen.
The logger will now track new min/max values. A min/max reset mark will be recorded in the logged data. As soon as the next reading is taken, both minimum and maximum values will be updated. You will be able to see in the chart in LogTag® Analyzer when the min/max values were cleared, but you will not be able to review previous min/max values on screen once they have been cleared.
Stopping the TREL30-16 TREL30-16 User Guide 25

Stopping the TREL30-16

Automatically

The TREL30-16 automatically stops recording temperature when the maximum number of readings specified during configuration has been reached. Your unit can also be set up to stop automatically when it is downloaded. This option needs to be set up at the factory and cannot be changed during configuration with LogTag® Analyzer. Your distributor can supply more information about this option.

Push-button stop

You can configure a TREL30-16 so it can be stopped with the START/CLEAR/STOP button. This feature is useful when you take the logger out of a shipment and don’t want to falsify the statistics with readings taken after the shipment completion. The stop function is enabled in the Advanced Options dialogue during configuration. When enabled, following will stop the unit:
Press and hold the START/CLEAR/STOP button. The STOPPED symbol shows.
After approx 2 seconds the symbol turns off.
Release the button within 2 seconds. The logger will now stop taking readings.
The logger will not stop if you…
…
release the button before the STOPPED symbol remains permanently on, or
…
keep holding the button for more than 2 seconds after the STOPPE D symbol remains permanently
on.
If an ALARM is present ( ) and the Clear alarm by pressing STOP button function is enabled, you will need to clear the alarm first before you can stop the logger. Please see Clearing an Alarm on
page17 for more information on how to clear an alarm.
Once the logger has stopped, the alarm status will no longer be indicated with a cross or a tick symbol, and the audible alarm will turn off.
Stopping the TREL30-16 TREL30-16 User Guide 26

Push-button stop, quick stop disabled

Press and hold the START/CLEAR/STOP button. The STOPPED symbol starts flashing for 4 seconds.
Then the STOPP ED symbol turn on permanently:
Release the button within 2 seconds. The logger will now stop taking readings.
The logger will not stop if you…
…
release the button before the STOPPED symbol remains permanently on, or
…
keep holding the button for more than 2 seconds after the STOPPE D symbol remains permanently
on.
Resetting the Logger TREL30-16 User Guide 27

Resetting the Logger

You can reset a logger back to its original READY state. Once reset, recording can be started again as described in Starting the Logger on page13.
Warning: When performing this procedure all recordings and statistics stored in the logger are deleted and cannot be recovered. Please ensure your data have been saved!
For this process to work, the logger must be in the STOPPE D state.
l
Press and hold the button. The READY symbol will flash.
l
When the READY symbol remains permanently on, release the button within 2 seconds. The logger
is now ready to be started again.
The ability to reset a logger is enabled or disabled in the Advanced Options dialogue when configuring the TREL30-16 via LogTag® Analyzer (see Advanced Configuration Options on page9).

Hibernating a TREL30-16

When hibernated, the logger’s power consumption is near zero and the life of the battery approaches its shelf life (typ. 5 or more years). This is useful for conserving battery life when the logger is not used for extended periods . TREL30-16 loggers are placed into Hibernation using LogTag® Analyzer by clicking Hibernate on the LogTag menu. The display shows SLP for a few seconds, before turning itself off.
A hibernated logger has no active display, however a button press will wake the logger up briefly. The display will show:
l the low battery symbol if the battery requires replacing.
Hibernating a TREL30-16 TREL30-16 User Guide 28
A hibernated logger can be activated again by pressing and holding the and buttons together for a period of 6 seconds. During this time the READY symbol flashes. Release the buttons when the flashing
stops.
You can re-activate a logger with a low battery, however it is not recommended to commence another trip.
Note: The real time clock is not running in a hibernated logger and must be set up if the logger is manually activated. Therefore, the CLOCKADJ procedure is automatically invoked (see Adjusting
the Display Clock on page12).
Once the clock has been set, the logger is ready to be started again.

Technical Specifications

Model Number
Temperature Sensor Measurement Range
-80°C to +40°C (-112°F to +104°F)
Logger Operating Temperature Range
-30°C to +60°C (-22°F to +140°F)
Storage Temperature Range
-10°C to +40°C (14°F to +104°F)
Ambient humidity range during transport and use
0 to 95%RH
Rated Temperature Resolution
l 0.1°C (0.2°F) for measurements from -80°C to 0°C (-112°F to 32°F) l 0.2°C (0.4°F) for measurements from 0°C to+18°C (32°F to 64.4°F) l 0.3°C (0.5°F) for measurements from +18°C to +30°C (64.4°F to
86°F)
l 0.5°C (0.9°F) for measurements from +30°C to +40°C (86°F to 104°F)
For more detailed information please refer to the published resolution charts
Rated Temperature Accuracy
l Better than ±0.5°C (±0.9°F) for -20°C to +10°C (-4°F to +50°F) l Better than ±0.7°C (±1.4°F) for -50°C to -20°C (-40°F to -4°F) and
+10°C to +30°C (50°F to 86°F)
l Better than ±1.0°C (±1.8°F) for -80°C to -50°C (-112°F to -58°F) and
+30°C to +40°C (86°F to 104°F)
For more detailed information please refer to the published accuracy charts
Sensor Compatibility
All ST10 precision thermistor external probes can be interchangeably connected to the TREL30-16. Please refer to the dedicated sensor page on the LogTag Recorders website.
Clock accuracy
Quartz crystal locked real time clock, rated accuracy ±25ppm @ 25°C (equiv to 2.5 seconds/day) Rated temperature coefficient is -0.034±0.006ppm/°C (i.e. typically +/-
0.00294seconds/day/°C)
Recording Capacity
15,905 real time temperature values, giving
l 66 days @ 6 minute logging interval or l 110 days @ 10 minute logging interval l 165 days @ 15 minute logging interval.
Supports continuous logging ("wrap-around") or specific recording period.
Statistics memory
for displaying statistics on LCD
l Max/Min values for the past 30 days l Alarm duration values for the past 30 days. l Min/Max trip temperature
Memory type
Non volatile
Sampling Interval
Configurable from 30 seconds to 18 hours.
Start options
l Push button start with optional configurable start delay from 1 minute to
18 hours l Push button start with quick start option l Date/time start up to 180 days in the future
Alarm functions
l one configurable upper alarm l one configurable lower alarm l OK tick and Alarm cross on display, linked to alarms
Vibration
Withstands vibration specification as detailed in EN12830:1999
Shock
l Withstands shock specification as detailed in EN12830:1999 l Withstands 5 drops from 1m onto smooth concrete floor without loss of
function or calibration
EMC compliance
l EC EMC directives (EN 61000-6-1:2005 & EN 61000-6-3:2006) l Includes electrostatic discharge as prescribed in EN 61000-4-2 l Complies with FCC Part 15 Subparts A and B
Environmental
IP61
Case Material
Polycarbonate
Power source
CR2032 3V Li-MnO2battery with extended temperature Chemistry – User replaceable, non-rechargeable
Battery life
Typically 1 year of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly) provided recorder is kept within the storage temperature range when not in use. The audible alarm reduces battery life significantly.
Size
93mm(H) x 54.5mm (W) x 8.6mm (T) (excluding probe)
Weight
43g (excluding probe)
Calibration
Calibration traceable to an ISO/IEC 17025 accredited testing laboratory
Download time
l Typically with full memory (15,905 readings) less than 10 seconds,
depending on computer or readout device used
Software requirements
l LogTag® Analyzer version 2.7 release 6 or later to configure and
download
Accessories
l Wall holder 200-000010 l LTI/USB Interface Cradle l Glycol Buffer Bottle
Page 29 of 29 TREL30-16 Product User Guide, Revision 1.0, copyright © 2004-2019, LogTag Recorders. All r ights reserved.

Appendix 1 - Battery Replacement

Prepare the Logger
l Download the unit to save any recorded data.
l Hibernate the unit using LogTag® Analyzer.
Failing to hibernate the logger may result in serious memory corruption. As a result you may no longer be able to use this product.
Remove the old battery
l Carefully remove the rear label covering the
battery door using a small-bladed screwdriver or
knife:
l Use a coin to turn the battery door anti-clockwise;
remove it from the case:
Insert the new battery
Only use batteries from a reputable manufacturer! Check the operational temperature range of the battery and make sure it covers the range of the TREL30-16.
l Put a new CR2032 cell into the battery
compartment as shown; the right hand edge of the battery is inserted first against the contact:
l Press firmly down on the left hand side of the
battery to click the battery into place:
l Remove the battery using a small-bladed
screwdriver, as shown:
l Replace the battery door; turn clockwise to lock it
into place.
l Cover the battery door with either a new label (Part
# 100-000502) or reuse the old one.
l Re-configure the logger with LogTag® Analyzer.
Please recycle or dispose of the old battery according to your local environmental regulations.
Page 30 of 29 TREL30-16 Product User Guide, Revision 1.0, copyright © 2004-2019, LogTag Recorders. All r ights reserved.